B&C Electronics CL 7685 Instruction Manual

0 (0)
B&C Electronics CL 7685 Instruction Manual

U N I E N I S O 9 0 0 1 : 2 0 0 0

C E R T . N . 9 1 1 5 B C E L

INSTRUCTION MANUAL

CL 7685

RESIDUAL CHLORINE

CHLORINE DIOXIDE - D. OZONE MICROPROCESSOR CONTROLLER * potentiostatic *

Range:

0/2.000/20.00 PPM

Temperature range:

-2/+52 °C

Power supply:

110/220 Vac

Software:

R:2.4 x

Option

S/N

REP N°

Cod. 28006765 - Rev. D -

11/06

 

 

 

B&C Electronics Srl -

Via per Villanova 3 - 20040 Carnate (Mi) - Italy

- P.IVA 00729030965

 

Tel +39 039 63

1721 -

Fax +39 039 607 6099 – bc@bc-electronics.it

- www.bc-electronics.it

B&C Electronics

CL 7685

 

 

Index

 

1

GENERAL.........................................................................................................................................

4

 

1.1

Functional specifications ..........................................................................................................

5

 

1.2

Physical description..................................................................................................................

8

2

SPECIFICATIONS............................................................................................................................

9

3

INSTALLATION ............................................................................................................................

14

 

3.1

Physical installation................................................................................................................

14

 

3.2

Electrical installation ..............................................................................................................

14

 

3.3

Operating the system ..............................................................................................................

17

4

KEYBOARD ...................................................................................................................................

18

5

READOUT SEQUENCES ..............................................................................................................

19

6

CALIBRATION SEQUENCES ......................................................................................................

23

 

6.1

Operating mode selection .......................................................................................................

23

 

6.2

Zero and sensitivity calibration ..............................................................................................

24

 

6.3

Temperature calibration..........................................................................................................

28

 

6.4

Set-point A/B setting ..............................................................................................................

29

 

6.5

Alarm setting ..........................................................................................................................

33

 

6.6

Cleaning function ...................................................................................................................

34

7

CONFIGURATION.........................................................................................................................

37

 

7.1

Keyboard locked/unlocked .....................................................................................................

37

 

7.2

LCD display contrast ..............................................................................................................

37

 

7.3

Access number........................................................................................................................

38

 

7.4

Type of measuring ..................................................................................................................

38

 

7.5

Input range..............................................................................................................................

38

 

7.6

Autorange ...............................................................................................................................

39

 

7.7

Software filter .........................................................................................................................

39

 

7.8

Cell polarization voltage.........................................................................................................

39

 

7.9

Calibration mode ....................................................................................................................

39

 

7.10

Temperature coefficient..........................................................................................................

40

 

7.11

Input related to analog output n°1 (option 091.3711) ............................................................

40

 

7.12

Analog output n°1 range.........................................................................................................

40

 

7.13

Input related to analog output n°2 (option 091.3711) ............................................................

41

 

7.14

Analog output n°2 range.........................................................................................................

41

 

7.15

Set-point A operating mode....................................................................................................

42

 

7.16

Set-point A function ...............................................................................................................

42

 

7.17

Set-point B operating modE ...................................................................................................

42

 

7.18

Set-point B function................................................................................................................

43

 

7.19

Alarm on Set-point A .............................................................................................................

43

 

7.20

Alarm on Set-point B..............................................................................................................

43

 

7.21

Alarm relay contact function ..................................................................................................

44

 

7.22

Alarm type ..............................................................................................................................

44

 

7.23

alarm flashing frequency ........................................................................................................

44

 

7.24

Cleaning function ...................................................................................................................

45

 

7.25

Cleaning time (relay D on) .....................................................................................................

45

Instruction Manual - Rev. D - 11/06

- 2

B&C Electronics

CL 7685

 

7.26

Holding time ...........................................................................................................................

45

 

7.27

New access number ................................................................................................................

45

8

CALIBRATION...............................................................................................................................

47

 

8.1

Electrical calibration...............................................................................................................

47

 

8.2

Chemical calibration...............................................................................................................

47

9

PREVENTIVE MAINTENANCE...................................................................................................

48

WARRANTY CERTIFICATE...............................................................................................................

52

REPAIRS................................................................................................................................................

52

TECHNICAL SUPPORT .......................................................................................................................

53

Instruction Manual - Rev. D - 11/06

- 3

B&C Electronics

CL 7685

1 GENERAL

This series of controllers with built-in microprocessor forms an advanced system of measurement and regulation of pH - ORP - Conductivity - Dissolved Oxygen - Chlorine in industrial processes.

These instruments are the result of all of B&C Electronics know-how built up over more than 30 years of experience in the field of industrial electrochemical analysis.

We have continually developed and updated these products keeping in mind the maximum consideration for quality, reliability, completeness of functions, simplicity of handling, and also the cost.

The computing capability and the versatility of the microprocessor are accompanied by friendly software specially designed to make the use of the instrument, the programming of its functions, routine checks and calibrations easily accessible even for an untrained operator.

The alpha-numeric backlit liquid crystal display helps the operator supplying him with all the information on functioning and on operating, while the software suggests possible steps for calibration and set-up.

The display supplies, at the same time as the measurement, an indication of present status of the output relays and of the output current.

For a higher level of checks, the display gives an indication of the operating conditions of the electrochemical sensors, the current status of the calibrations, set points and analog output.

The user friendly operation of the instrument saves the operator feats of memory and continual reference to the instruction manual.

All operations are carried out using the five key pads on the front panel, for the mode selection, the input of calibration and set point data and the setting-up. (Fig. 1)

There are no knobs or switches to manipulate in order to operate the unit. This makes the instrument rugged and more corrosion resistant.

A non-volatile EEPROM memory assures measurement parameters are maintained in the event of a power interruption.

The software is provided with a "watch-dog" check for correct functioning of the programs.

The electrical circuit is protected by a device which comes back into operation automatically following an overload or a wiring error.

Instruction Manual - Rev. D - 11/06

- 4

B&C Electronics

CL 7685

1.1 FUNCTIONAL SPECIFICATIONS

Input

The instrument accepts input from potentiostatic sensor for Active Chlorine, Chlorine Dioxide and D.Ozone measuring.

A second input is provided for 3 wires Pt100 Temperature sensor.

Temperature compensation

The unit is supplied with manual or automatic Temperature compensation and Temperature information may be displayed on LCD.

The instrument detects the absence or malfunctioning of the Temperature sensor and automatically switches to manual operation mode.

Measuring ranges

The unit provides an input range which may be selected from 0/1.999 or 0/19.99 PPM. Auto ranging function may be activated for the measuring range 0/1.999 PPM.

Auto ranging allows the operator to calibrate the unit in the low range, against a high concentration of standard solution.

Analog output

Either a 0/20 mA or 4/20 mA isolated output may be selected, provided for interface with computer or data loggers.

A special routine allows selection of the analog output range.

If the instrument is programmed for high range, the output may be set anywhere from 0/20.00 PPM. If low range is selected, the output may be set anywhere from 0/2.000 PPM.

Control relays

The monitor is equipped with two SPDT control relays.

These output relays may be used in a variety of ways, and the function of each relay is programmable by operator.

Control relays can be used in one of 3 different output modes: on/off, pulse width modulation and pulse frequency modulation.

The on/off mode of operation is used for simple control or for alarm purposes. Each control relay may be programmed for set-point, high/low, hysteresis, delay time for actuation.

The proportional mode of operation is used for more accurate control.

Each control relay may be programmed for set-point, high/low, proportional band, Frequency or Width modulation.

The full display indicates the current setting and the current status of each relay.

Instruction Manual - Rev. D - 11/06

- 5

B&C Electronics

CL 7685

Alarm relay

The unit contains a SPST relay designated as an alarm relay.

This relay may be used to warn of various conditions that might indicate operational problems.

The relay will activate on either high or low concentration conditions, or on failure of the control relays to maintain proper control.

This relay may be programmed for either normal or fail-safe operation.

The software R:2.4x provides the additional function of the selectable continuous/flashing alarm and the selectable flashing frequency.

Cleaning function

The unit contains a SPST relay designed for the auto clean cycle.

It's possible manually activate the cleaner to test its operation.

The operator can change the frequency of cleaning.

The configuration routine selects:

-auto/manual operation

-the cleaning time

-the holding time.

During cleaning and holding time:

-messages are flashing

-analog output is held constant

-set-points and alarm relays are deactivated.

Operating mode

The instrument is provided with 3 programmable modes of operation.

- Automatic operation (AUTO):

The Automatic mode is the normal operation mode of the unit.

- Measuring operation (MEAS.):

In this mode of operation the display indicates only the concentration, the analog output is active but the control relays are deactivated.

This would be the mode to use if the relays are not being used for alarm or control functions.

The measuring mode of operation is useful for start up or for manual operation of disinfection plants.

Instruction Manual - Rev. D - 11/06

- 6

B&C Electronics

CL 7685

- Simulated operation (SIM.):

In this operating mode the users choose the displayed value by means of up/down keys.

The unit maintains the set point, alarm and analog output parameters in order to test the plant.

The display does not indicate the measuring units (PPM) and access to calibration of the parameters is not allowed.

The message "SIM" is shown on the display, indicating the current operation mode of the instrument.

Calibration mode

The instrument may be programmed for the immediate or postponed calibration.

The immediate calibration mode allows the operator to calibrate the unit immediately against a field measurement on the same sample that the sensor is measuring.

The postponed calibration mode allows the operator to calibrate the unit against a laboratory measurement on the same sample that the sensor is measuring.

The calibration may be performed later even if the sample concentration has changed.

Software filter

The unit is provided with a dual programmable software filter, to be inserted when the readout is not stable.

The filter values can be selected separately for the small or large signal variations.

Configuration

The electronics for the monitor is designed to be as flexible as possible.

A number of programming functions are provided in the Configuration menu, protected by an access number, which must be entered to allow changes in this setting.

The routine allows the programming of a custom access number.

Front panel lock

The keys on the front panel of the monitor can be used for changing the display, for calibrations and set-point adjustments.

When the monitor is shipped, all functions are accessible.

However, the adjustment and calibration functions may be locked in order to prevent unauthorized adjustments to the instrument.

Instruction Manual - Rev. D - 11/06

- 7

B&C Electronics

CL 7685

Options

091.3711 Dual isolated and programmable output

The operator may select an output for Temperature.

091.404 24 VAC power supply.

091.4143 9/36 VDC power supply.

091.701 RS232 isolated output

The output sends the data (PPM,°C) to the serial port of the computer.

1.2 PHYSICAL DESCRIPTION

The controller enclosure is designed for surface or panel mounting.

It consists of an anodized aluminium case built according to the standard DIN 43700, with an aluminium panel coated with scratch-proof and non-corrosive polycarbonate membrane.

A transparent waterproof front door SZ 7601 can be added to the housing, in order to protect the unit from excessive moisture or corrosive fumes.

Signal and power connections are made by using two special extractable terminal blocks placed in the back of the instrument.

This makes wiring, installation and general maintenance of the probes and other devices easier.

The package is supplied complete with fixing clamps for panel-mounting.

Instruction Manual - Rev. D - 11/06

- 8

B&C Electronics

CL 7685

2 SPECIFICATIONS

The DEFAULT values are correspondent to the factory calibration values.

Parameters marked by " * " can be modified in the Configuration procedures.

 

 

Default

 

 

OPERATING MODE

 

 

Automatic/Measuring/Simulation

Auto

 

 

 

 

MEASURING TYPE

 

*

Chlorine/Chlorine dioxide/D. Ozone

Chlorine

 

 

 

 

INPUT SCALES

 

 

Input range: 2.000/20.00 PPM

20.00

 

Display resolution at 20 °C: 1/2000

 

 

Software filter 90% RT:

 

*

LARGE signal: 0.4/20,0 s

2.0 s

*

SMALL signal: 0,4/20,0 s

10,0 s

 

Current at 20 °C: 250/5000 nA/PPM

2000 nA/PPM

 

Cell Sensitivity: 12.5/250%

100%

 

Zero: adjustable

0 µA

*

Comp. Temp. Coefficient Cl/ClO2: 0/4.0 % / °C

2.0% / °C

*

Comp. Temp. Coefficient O3: 0/4.0 % / °C

2.5% / °C

*

Polarization Voltage: adjustable

-200 mV

 

 

 

 

TEMPERATURE

 

 

Input: RTD Pt100

 

 

Connection: 2/3 wires

 

 

Measuring and compensation range: -2/+52 °C

 

 

Resolution: 0.1 °C

 

 

Zero adjustment: +/- 2°C

0 °C

 

Manual Temperature compensation: 0/50 °C

20 °C

 

 

 

 

 

 

Instruction Manual - Rev. D - 11/06

- 9

B&C Electronics

CL 7685

SET A/B

*Selectable actions: ON-OFF

PFM - Pulse frequency proportional PWM - Pulse width proportional

*

Action: ON-OFF

SET B

 

Value: 0/2.000 - 0/20.00 PPM (as selected scale)

0 PPM

 

Hysteresis: 0/0.200-0/2.00 PPM (as selected scale)

0.2 PPM

 

Delay: 0/99.9 s

0.0 s

*

Function: HI/LO (Max/min)

LO

*

Action: PFM

SET A

 

Value: 0/20.00 PPM (as selected scale)

0 PPM

 

Proportional band: 0/0.200 - 0/2.00 PPM

0.2 PPM

 

(as selected scale )

 

 

Pulse max. Frequency: 0/120 pulse/minute

100 p/m

 

Pulse width:

0.1 s

*

Function: HI/LO (Max/min)

LO

*Action: PWM

 

Value: 0/0.200 - 0/20.00 PPM (as selected scale)

0 PPM

 

Proportional band: 0/0.200 - 0/2.00 PPM

0.2 PPM

 

Pulse width: 0/99.9 s

(as selected scale)

20.0 s

 

 

 

Min. pulse width:

 

0.3"

*

Function: HI/LO (Max/min)

 

LO

 

Relay Contacts: SPDT 220 V

5 A Resistive load

 

 

 

 

 

 

 

ALARM (RELAY C)

 

 

 

High value: 0/2.000 - 0/20.00 PPM (as selected scale)

20.00 PPM

 

Low value: 0/2.000 - 0/20.00 PPM (as selected scale)

0.00 PPM

 

Delay: 0/99.9 s

 

0.0 s

*

Alarm on max. SA: ON/OFF

 

OFF

*

Max. time SA: 0/60 minutes

 

60 min

*

Alarm on max. SB: ON/OFF

 

OFF

*

Max. time SB: 0/60 minutes

 

60 min

*

Contact type: ACT/DEA

 

ACT

*

Alarm type: CONT./FLASH

 

CONT

*

Flashing frequency:

LO (approx. 0.3 Hz duty cycle)

ME

 

 

ME (approx. 0.6 Hz duty cycle)

 

 

 

HI (approx. 1.2 Hz duty cycle)

 

 

Relay Contact: SPST 220 V

5 A Resistive load

 

 

 

 

 

Instruction Manual - Rev. D - 11/06

- 10

B&C Electronics

CL 7685

 

 

 

CLEANING FUNCTION (RELAY D)

 

*

Action: Disable/Manual Clean/Auto+Manual Clean

Disable

 

Auto Clean:

 

 

Repetition time: 0.1/24.0 h

24.0 h

*

Cleaning time: 0.5/60.0 s

15.0 s

*

Hold time: 0.1/20.0 minutes

3.0 min.

 

Relay contacts: SPST

 

 

 

 

 

ANALOG OUTPUT Nr. 1

 

* Current range: 0-20/4-20 mA

0/20 mA

* Point 1 corresponding to 0 mA or 4 mA

 

 

RANGE 20.00 PPM: 0.00/20.00

0.00 PPM

 

RANGE 2.000 PPM: 0.000/2.000

0.000 PPM

 

TEMPERATURE: 0.0°C/50.0°C (Option 091.3711)

0.0 °C

* Point 2 corresponding to 20 mA

 

 

RANGE 20.00 PPM: 0.00/20.00

20.0 PPM

 

RANGE 2.000 PPM: 0.000/2.000

2.000 PPM

 

TEMPERATURE: 0.0°C/50.0°C (Option 091.3711)

50.0 °C

 

Response time: 10 s for 98 %

 

 

Isolation: 250 Vca

 

 

Rmax: 600 Ω

 

 

 

 

 

ANALOG OUTPUT N°2 (Only for Option 091.3711)

 

*

Current range: 0-20/4-20 mA

0/20 mA

*

Point 1 corresponding to 0 mA or 4 mA

 

 

RANGE 20.00 PPM: 0.00/20.00

0.00 PPM

 

RANGE 2.000 PPM: 0.000/2.000

0.000 PPM

 

TEMPERATURE: 0.0 °C/50.0 °C

0.0 °C

* Point 2 corresponding to 20 mA

 

 

RANGE 20.00 PPM: 0.00/20.00

20.00 PPM

 

RANGE 2.000 PPM: 0.000/2.000

2.000 PPM

 

TEMPERATURE: 0.0 °C/50.0 °C

50.0 °C

 

Response time: 10 s for 98 %

 

 

Isolation: 250 Vca

 

 

Rmax: 600 Ω

 

 

 

 

Instruction Manual - Rev. D - 11/06

- 11

B&C Electronics

CL 7685

SERIAL COMMUNICATION (Option 091.701)

Baud Rate: 4800 bit/s

Nr. of bit: 8 bit

Nr. of stop bit: 1 bit

Parity: None

Isolated from measure circuits

Example of data sent: ± 20.00 PPM ± 50.0 °C

Data sent every: 0.4 s

 

 

PARAMETERS ON CONFIG. BLOCK (See for *)

 

Free calibration (access code not required):

 

Keyboard Lock/Unlock

Unlock

LCD contrast (0/7).

4

Under access code number: (0)

0

Type of measure:Cl/ClO2/O3

Cl

Measuring range: 2.000/20.00

20.00 PPM

Auto ranging: On/Off

Off

LARGE software filter response time

2.0 s

SMALL software filter response time

10.0 s

Polarization

- 200 mV

Immediate/postponed calibration mode

Immed.

Temperature Coefficient

2.0%

Input connected to the output N°1

PPM

Analog output N°1 range (0/20 4/20 mA)

0/20 mA

Point 1 corresponding to 0 mA or 4 mA

0.00 PPM

Point 2 corresponding to 20 mA

20.00 PPM

Input connected to the output N°2

PPM

Analog output N°1 range (0/20 4/20 mA)

0/20 mA

Point 1 corresponding to 0 mA or 4 mA

0.00 PPM

Point 2 corresponding to 20 mA

20.00 PPM

Action of relays A (On-Off/PFM/PWM)

PFM

Function of the A (HI/LO)

LO

Action of relays B (On-Off/PFM/PWM)

On-Off

Function of the B (HI/LO)

LO

Alarm on max. operating time of SA

Off

Max. operating time of SA for alarm

60 m

Alarm on max. operating time of SB

Off

Max. operating time of SB for alarm

60 m

Alarm relay status (ACT/DEA)

ACT

Alarm type

Cont

Flashing frequency

ME

Cleaning function (Auto/Manual/Disabled)

Disabled

Cleaning time: 0.5/60.0 s

15.0 s

Holding time: 0.1/60.0 minutes

3.0 min.

Access number: 0/999

0

 

 

Instruction Manual - Rev. D - 11/06

- 12

B&C Electronics

CL 7685

GENERAL SPECIFICATIONS

Alphanumeric display: 1 line x 16 characters Response time to 98% of value changing, with TC=2% / °C - T=20 °C - S=100% :

<5 s for HI range

<15 s for LO range

Operating Temperature: 0/50 °C

Humidity: 95% without condensate

Power supply: 110/220 Volt ac +/- 10% 50/60 Hz

Isolation: 4000 V between primary and secondary (IEC 348)

Power: 5 VA max.

Terminal block: extractable

Weight: 850 g

Dimensions: 96 x 96 x 155 mm

Instruction Manual - Rev. D - 11/06

- 13

B&C Electronics

CL 7685

3 INSTALLATION

3.1 PHYSICAL INSTALLATION

The controller may be installed close to the areas being monitored, or it may be located some distance away in a control area.

The enclosure is designed for panel-mounting.

It should be mounted on a rigid surface, in a position protected from the possibility of damage, excessive moisture and corrosive fumes.

The cable from the probe must be protected by a sheath and not installed near power cables.

Interruption on cables must be avoided or carried out by high insulation terminals.

When installing "in line" electrodes it is suggested to follow the specific instructions given by the sensor's manufacturer.

3.2 ELECTRICAL INSTALLATION

All connections within the controller are made on detachable terminal strips located on the rear side. (Fig. 2)

All power and output-recorder connections are made at the 13_ pin terminal strip, while input signal connections are made at the 12_ pin terminal strip.

The electrical installation consists of:

Connecting the power

- Connect ground to terminal 4;

 

- connect ac power to

1 - 2

terminals if power voltage is

110 Vac;

- connect ac power to

1 - 3

terminals if power voltage is

220 Vac;

- if 091.404 option is installed, connect 24 Vac to 1-3 terminals.

WARNINGS:

-Power the device by means of an isolation transformer;

-avoid mains voltage from an auto-transformer;

-avoid mains voltage from a branch point with inductive loads;

-separate power supply wires from signal cables;

-control the mains voltage value.

Instruction Manual - Rev. D - 11/06

- 14

B&C Electronics

CL 7685

Connecting the sensor

-Sensor cabling is a critical part of the whole system;

-use original cable supplied with the sensor;

-avoid interruption on the cable.

- Connect the White wire (Counter electrode) to the terminal 17 marked EL; - connect the Black wire (Measuring electrode) to the terminal 18 marked IN; - connect the shield (Reference electrode) to the terminal 19 marked R.

Connecting alarms, pumps, valves

The output connections referred to Set-point SA and Set-point SB are made at terminal strip and they consist of two independent SPDT relays corresponding to Regulator A and Regulator B .

The output connection referred to alarm consists of SPST relay corresponding to Alarm C.

The output connection referred to auto-clean consists of SPST relay corresponding to Auto clean D.

Control relay "A"

 

Set-point "SA"

Terminal

6_

 

 

marked

C_

common contact

Terminal

 

5_

 

 

marked

NO_

normal open contact

Terminal

 

7_

 

 

marked

NC_

 

normal closed contact

Control relay "B"

 

Set-point "SB"

Terminal

9_

 

 

marked

C_

common contact

terminal

 

8_

 

 

 

marked

NO

_

normal open contact

terminal

10_

 

marked

NC_

normal closed contact

Alarm relay "C"

 

 

 

 

 

Terminal

12_

 

marked

C_

common contact

Terminal

 

11_

 

marked

NO_

normal open contact

Auto clean relay "D"

 

 

 

 

 

Terminal

12_

 

marked

C_

common contact

Terminal

 

13_

 

marked

NO_

normal open contact

Instruction Manual - Rev. D - 11/06

- 15

B&C Electronics

CL 7685

Connecting a recorder

A current output for a remote recorder or P.I.D. regulators is available on terminals 14-16.

Connect the recorder high (+) to terminal 14

Connect the recorder low (-) to terminal 16

Series connection is required for driving more loads having a total input Resistance lower than 600 Ω.

If the 091.3711 dual output option is installed, a second isolated and programmable output is available between 15-16 terminals.

Output N°1 and Output N°2 are isolated and selectable 0/20 or 4/20 mA.

Connecting the RTD

The instrument has the automatic Temperature compensation carried out by means of RTD Pt100.

To operate the automatic Temperature compensation, connect the RTD as shown in the "connection" figure. (Fig. 2)

A three wire connection is suggested to achieve an accurate compensation over a long distance between the sensor and the controller.

3-wire connection

-connect the terminal of RTD to terminal 23 of the meter;

-connect the common terminal of RTD to terminals 24 - 25 of the meter;

-the 3 wire-cable must not be interrupted on the overall length.

If an extension is needed, the cable must be fastened to the high insulation terminal strip;

-keep the cable away from power wires.

The RTD connection as above described allows the controller to provide a digital readout of Temperature.

If the Temperature sensor is not connected or damaged, the unit will operate in manual Temperature compensation automatically.

2-wire connection

-connect the Pt100 to terminals 23 - 24 ;

-install a jumper to terminals 24 - 25.

Checking

Before connecting the system to the power supply:

-check that all cables are properly fastened to prevent strain on the connections

-check that all terminal-strip connections are mechanically and electrically sound.

Instruction Manual - Rev. D - 11/06

- 16

Loading...
+ 37 hidden pages