BC GROUP INTERNATIONAL, INC. RESERVES THE RIGHT TO MAKE
CHANGES TO ITS PRODUCTS OR SPECIFICATIONS AT ANY TIME,
WITHOUT NOTICE, IN ORDER TO IMPROVE THE DESIGN OR
PERFORMANCE AND TO SUPPLY THE BEST POSSIBLE PRODUCT. THE
INFORMATION IN THIS MANUAL HAS BEEN CAREFULLY CHECKED AND
IS BELIEVED TO BE ACCURATE. HOWEVER, NO RESPONSIBILITY
The SPO-2000 is a microprocessor based pulse generator capable of calibrating any
phantom finger based heart rate monitor.
The unit has six available pulse rates ranging from 30 to 180 BPM. Each pulse rate has a
unique LED indicator.
The SPO-2000 is designed to work in conjunction with the FingerSim manufactured by
BC Biomedical and available from BC Group International. The FingerSim is used to
verify the operation of pulse oximeters. The SPO-2000 adds calibrated pulse rates to the
functionality of the FingerSims.
3
OVERVIEW
This section looks at the layout of the SPO-2000 and gives descriptions of the elements
that are present.
2 Light Touch Keys
Connector for Power
Transformer
ON/OFF Switch
ACTIVE/STANDBY for
Controlling the Power
Generator.
SCROLL for selecting
the Pulse Rate.
6 LEDs for Indicating
Pulse Rate (BPM)
FingerSim Holder
2 LEDs for Indicating Status
ACTIVE for showing when the
Power Generator is active.
POWER for showing when the
Unit is On.
4
KEYS
2 soft touch keys are provided for unit operation:
– This key is used to turn the pulse generator output ON and OFF.
– This key is used to select the desired pulse rate.
5
LED INDICATORS
There are eight LEDs to indicate the status of the system. They are marked POWER,
ACTIVE, 30, 60, 90, 120, 150, AND 180.
POWER – This LED indicates that the power switch is turned ON and the on board power
supply is functioning properly.
ACTIVE– This LED indicates that the system is in the ACTIVE mode and the solenoid is
being pulsed at the currently selected heart rate.
30 – This LED indicates that the currently selected heart rate is 30 BPM. This is the heart
rate that is selected when the system is first turned ON.
60 – This LED indicates that the currently selected heart rate is 60 BPM.
90 – This LED indicates that the currently selected heart rate is 90 BPM.
120 – This LED indicates that the currently selected heart rate is 120 BPM.
150 – This LED indicates that the currently selected heart rate is 150 BPM.
180 – This LED indicates that the currently selected heart rate is 180 BPM.
6
OPERATIONS
When the system is first turned on the SPO-2000 will start in the standby mode with a
heart rate of 30 BPM.
Insert the desired FingerSim into the rectangular opening on the SPO-2000.
NOTE: It is recommended that the FingerSim be inserted with the printed side down to
minimize wear on the printing by the mechanism inside the SPO-2000.
Use the key to select the desired pulse rate. Then the output is enabled by
depressing the key. The ACTIVE LED will turn on and the FingerSim will pulse
at the desired rate.
The pulse rate can be changed by depressing the key, in either the ACTIVE or
STANDBY mode.
Connect the finger probe of the pulse oximeter being tested to the FingerSim. Run the
probe’s cable underneath the SPO-2000.
The output is disabled by depressing the key. The ACTIVE LED will turn off and
the Phantom Finger will stop pulsing.
7
NOTE: The accuracy of the saturation percentage on the pulse oximeter being tested is a
function of the FingerSim - not the SPO-2000. The SPO-2000 only controls the pulse
rate. See the user information included with the FingerSim for accuracy and related
data.
8
WARRANTS ITS NEW PRODUCTS TO BE FREE
FROM DEFECTS IN MATERIALS AND WORKMANSHIP UNDER THE SERVICE FOR WHICH THEY ARE
INTENDED. THIS WARRANTY IS EFFECTIVE FOR TWELVE MONTHS FROM THE DATE OF
ANY OTHER WARRANTY EXPRESSED OR IMPLIED,
OR FITNESS
DENTAL OR CONSEQUENTIAL
NO PERSON OTHER THAN AN OFFICER IS AUTHORIZED TO GIVE ANY OTHER WARRANTY OR
THE PURCHASER'S SOLE AND EXCLUSIVE REMEDY SHALL BE: (1) THE REPAIR OR
, WITHOUT CHARGE. (2) AT THE OPTION
MANUAL REVISIONS
Revision # Program # Revisions Made
Rev 01 DT7300CA Original BC Manual
Rev 02 DT7300PGA Updated to new format
LIMITED WARRANTY
WARRANTY: BC GROUP INTERNATIONAL, INC.
SHIPMENT.
EXCLUSIONS: THIS WARRANTY IS IN LIEU OF
INCLUDING, BUT NOT LIMITED TO ANY IMPLIED WARRANTY OF MERCHANTABILITY
FOR A PARTICULAR PURPOSE.
BC GROUP INTERNATIONAL, INC. IS NOT LIABLE FOR ANY INCI
DAMAGES.
ASSUME ANY LIABILITY.
REMEDIES:
REPLACEMENT OF DEFECTIVE PARTS OR PRODUCTS
OF BC GROUP INTERNATIONAL, INC., THE REFUND OF THE PURCHASE PRICE.