BC Biomedical SA-2010S, SA-2010S-INTL, SA-2010S-AUS User Manual

Page 1
SA-2010S-INTL SA-2010S-AUS
SAFETY ANALYZER
ELECTRICAL
SA-2010S
USER MANUAL
Page 2
Page 3
i
BC BIOMEDICAL
TABLE OF CONTENTS
This User Manual covers the following units:
SA-2010S SERIES
WARNINGS, CAUTIONS, NOTICES ............................................................................ ii
DESCRIPTION ............................................................................................................. 1
LAYOUT ....................................................................................................................... 4
TESTING ...................................................................................................................... 10
PATIENT SIMULATOR ................................................................................................ 15
COMMUNICATION PROTOCOL ................................................................................ 16
MANUAL REVISIONS .................................................................................................. 19
LIMITED WARRANTY .................................................................................................. 19
SPECIFICATIONS ....................................................................................................... 20
NOTES ......................................................................................................................... 24
SA-2010S & SA-2010S-R SA-2010S-INTL & SA-2010S-INTL-R SA-2010S-AUS & SA-2010S-AUS-R
Page 4
ii
WARNING - USE
or any other capacity where they would come in
contact with a patient.
WARNING - MODIFICATIONS
beyond these specifications or any unauthorized
improper operation.
WARNING - CONNECTIONS
not connect any leads from the patient directly to
SA-2010S.
WARNING - CLEANING
2010S should be cleaned by wiping gently with a
used if desired.
The SA-2010S is intended for testing only and
should never be used in diagnostics, treatment
All connections to patients must be removed
before connecting the DUT to the SA-2010S. A
serious hazard may occur if the patient is
connected when testing with the SA-2010S. Do
the SA-2010S or DUT while it is powered by the
The SA-2010S is intended for use within the
published specifications. Any application
user modifications may result in hazards or
Disconnect Line Power to the SA-2010S before
attempting to clean it. Do not immerse. The SA-
damp, lint-free cloth. A mild detergent can be
Page 5
iii
WARNING - LIQUIDS
2010S. In the event of a spill onto the SA-2010S,
type.
CAUTION - ENVIRONMENT
30 minutes before attempting to operate it.
CAUTION - USAGE
The SA-2010S is not a continuous duty device, it current limits and duty periods specified. Do not
Do not to drop the SA-2010S.
WARNING - VOLTAGE
voltage is applied to the Patient lead connectors
and/or External test leads. Although this voltage
depressed.
Do not submerge or spill liquids on the SA-
Do not operate the SA-2010S regardless of fluid
When the SA-2010S is in Lead Isolation mode
and the ISO key is depressed, 110% of line
is applied through an internal current limiting
resistance of 121 kΩ (per standard test
specifications), Do not touch the test leads,
connections or DUT while the ISO key is
is intended for short duration testing within the
leave the DUT connected to the SA-2010S for
extended time periods.
Exposure to environmental conditions outside
the specifications can adversely affect the
performance of the SA-2010S. Allow SA-2010S
to acclimate to specified conditions for at least
Page 6
iv
NOTICE – SYMBOLS
Symbol
Description
Caution
(Consult Manual for Further Information)
Electrical Caution
(Consult Manual for Further Information)
Per European Council Directive 2002/95/EC, do not dispose of this product as unsorted municipal waste.
CAUTION - SERVICE
The SA-2010S is intended to be serviced only by
performed by qualified technical personnel.
CAUTION - INSPECTION
serviced if any parts are in question.
CAUTION - FUSE
specified type and rating.
The SA-2010S should be inspected before each
use for obvious signs of abuse or wear. The SA-
2010S should not be used and should be
authorized service personnel. Troubleshooting
and service procedures should only be
Only replace the SA-2010S fuse with the
Page 7
v
NOTICE – ABBREVIATIONS
Amp
Ampere(s)
AHA
American Heart Association
Association for the Advancement of Medical Instrumentation
BPM
Beats Per Minute
C
Celsius
cm
centimeter(s)
°
degree(s)
DUT
Device Under Test
ECG
Electrocardiogram
Euro
European
ft
feet
FS
Full Scale
Hz
hertz
IEC
International Electrotechnical Commission
ISO
Isolation
kg
kilogram(s)
kHz
kilohertz
kΩ
kilohm(s)
LED
Light Emitting Diode
MAP
Mains on Applied Parts
MHz
Megahertz
µA
microampere(s)
mA
milliampere(s)
mm
millimeter(s)
National Electrical Manufacturers Association
Ω
Ohm(s)
p-p
peak-to-peak
PC
Personal Computer
Lbs
pounds
RH
Relative Humidity
RMS
Root Mean Square
USA
United States of America
V
Volt(s)
VA
Volt-Ampere(s)
VAC
Volt(s) Alternating Current
W
Watt(s)
AAMI
NEMA
Page 8
vi
NOTICE – DISCLAIMER
BC GROUP INTERNATIONAL, INC. WILL NOT BE RESPONSIBLE FOR ANY INJURIES SUSTAINED DUE TO UNAUTHORIZED EQUIPMENT MODIFICATIONS OR APPLICATION OF EQUIPMENT OUTSIDE OF THE PUBLISHED INTENDED USE AND SPECIFICATIONS.
NOTICE – DISCLAIMER
BC GROUP INTERNATIONAL, INC. RESERVES THE RIGHT TO MAKE CHANGES TO ITS PRODUCTS OR SPECIFICATIONS AT ANY TIME, WITHOUT NOTICE, IN ORDER TO IMPROVE THE DESIGN OR PERFORMANCE AND TO SUPPLY THE BEST POSSIBLE PRODUCT. THE INFORMATION IN THIS MANUAL HAS BEEN CAREFULLY CHECKED AND IS BELIEVED TO BE ACCURATE. HOWEVER,
NO RESPONSIBILITY IS ASSUMED FOR INACCURACIES.
NOTICE – CONTACT INFORMATION
sales@bcgroupintl.com
NOTICE – PERFORMING TESTS
SA-2010S Series User Manual Copyright © 2012
www.bcgroupintl.com Made in the USA
06/12 Rev 11
REFER TO DUT MANUFACTURER’S SERVICE MANUAL FOR
TEST PROCEDURES AND MEASUREMENT LIMITS.
BC BIOMEDICAL
BC GROUP INTERNATIONAL, INC.
3081 ELM POINT INDUSTRIAL DRIVE
ST. CHARLES, MO 63301
1-800-242-8428 1-314-638-3800
www.bcgroupintl.com
USA
Page 9
1
BC BIOMEDICAL
WITH PATIENT SIMULATOR
SA-2010S SERIES
ELECTRICAL SAFETY ANALYZER
The Model SA-2010S Series is a Microprocessor based Electrical Safety Analyzer with a built-in Patient Simulator. It allows for a multitude of tests to be performed on a device using the same unit and lead connections. It provides a full Electrical Safety Analyzer, as well as, ECG Simulation with four waveforms with constant QRS duration and six machine performance testing waveforms. There is Patient Lead testing for 10 inputs. The following are highlights of some of the main features:
SA-2010S:
LED STATUS INDICATORS AUDIO FEEDBACK TOUCH CONTROL KEYS – NO KNOBS 10 UNIVERSAL PATIENT LEAD INPUTS HIGH IMPACT PLASTIC CASE
SAFETY ANALYZER:
LINE VOLTAGE MEASUREMENT DEVICE UNDER TEST CURRENT MEASUREMENT EARTH / GROUND LEAD RESISTANCE EARTH / GROUND LEAKAGE CURRENT ENCLOSURE / CHASSIS LEAKAGE CURRENT EXTERNAL RESISTANCE EXTERNAL LEAKAGE CURRENT SOURCE WIRING INTEGRITY MONITOR TRUE RMS MEASUREMENTS AAMI ES1-1993 or IEC 601 SELECTABLE TEST LOADS 90 TO 264 VAC OPERATION 20 AMP RATING (SA-2010S and SA-2010S-INTL Models) 10 AMP RATING (SA-2010S-AUS Models) SELF TEST POINTS EXTERNALLY REPLACEABLE GROUND FUSE AUTOMATIC LOAD REVERSAL DELAY PATIENT LEAD TO LEAD LEAKAGE CURRENT PATIENT LEAD TO EARTH / GROUND LEAKAGE CURRENT PATIENT ISOLATION LEAKAGE CURRENT EXTERNAL ISOLATION LEAKAGE CURRENT
Page 10
2
PATIENT SIMULATOR:
ECG: 30, 60, 120 AND 240 BPM SINE: 10, 60 AND 100 Hz SQUARE: 0.125 AND 2.0 Hz TRIANGLE: 2 Hz HIGH LEVEL OUTPUT (1 V p-p) AMPLITUDE ACCURACY: +/- 2% FREQUENCY ACCURACY: +/- 0.5%
AVAILABLE MODELS:
SA-2010S – STANDARD MODEL WITH HOSPITAL-GRADE NEMA 5-15P LINE
PLUG AND HOSPITAL-GRADE NEMA 5-20R DUT TEST RECEPTACLE FOR USE IN THE US AND OTHER COMPATIBLE COUNTRIES
SA-2010S-INTL – INTERNATIONAL MODEL, IEC C20 RECEPTACLE PIGTAIL
(MUST USE COUNTRY-SPECIFIC LINE CORD – SEE ACCESSORIES SECTION) AND UNIVERSAL DUT TEST RECEPTACLE THAT WORKS WITH THE FOLLOWING COUNTRY-SPECIFIC PLUGS:
o NEMA 5-15P, NEMA 5-20P, NEMA 6-15P and NEMA 6-20P (US/NORTH
AMERICA)
o UK1-13P and UK3-5P (UK) o SW1-10P (SWITZERLAND) o IT1-10P (ITALY) o IS1-16P (ISRAEL) o JA1-15P (JAPAN) o EU1-16P (EURO) CEE 7/7 “SCHUKO” (NOTE: MUST USE SCHUKO
GROUNDING ADAPTER TO PERFORM LEAKAGE MEASUREMENTS, SEE OPTIONAL ACCESSORIES SECTION)
o DE1-13P (DENMARK) (NOTE: NO EARTH/GROUND PIN, THEREFORE
LEAKAGE MEASUREMENTS NOT APPLICABLE)
o EUROPLUG CEE 7/16 (NOTE: NO EARTH/GROUND PIN, THEREFORE
LEAKAGE MEASUREMENTS NOT APPLICABLE)
SA-2010S-AUS – AUSTRALIAN MODEL, AU1-10P LINE PLUG AND AU1-10R
DUT TEST RECEPTACLE FOR USE IN AUSTRALIA/NEW ZEALAND AND OTHER COMPATIBLE COUNTRIES
OPTIONS
-R ADDS RS-232 SERIAL COMMUNICATIONS AVAILABLE MODELS:
o SA-2010S-R o SA-2010S-INTL-R o SA-2010S-AUS-R
Page 11
3
STANDARD ACCESSORIES
BC20-20110 8 FT CHASSIS TEST LEAD (BLACK) BC20-30107 SOFT CARRYING CASE UF-0250-01 REPLACEMENT GROUND LEG FUSE BC20-204XX (SA-2010S-INTL ONLY) – REFER TO PAGE 9
OPTIONAL ACCESSORIES
BC20-20111 8 FT EXTERNAL TEST LEAD (RED) BC20-20112 16 FT CHASSIS TEST LEAD (BLACK) BC20-20113 16 FT EXTERNAL TEST LEAD (RED) BC20-20200 REPLACEMENT SA-2010S-INTL DUT INTERNATIONAL
RECEPTACLE ADAPTER
BC20-20221 SCHUKO GROUNDING ADAPTER BC20-41337 RS-232 COMMUNICATIONS CABLE (DB-9M to DB-9F) BC20-41339 USB TO RS-232 ADAPTER CS-2000-U 1 AMP CURRENT SOURCE - USA/NORTH AMERICA CS-2000-E 1 AMP CURRENT SOURCE - EURO (SCHUKO)
SA-2010S-INTL LINE CORDS: (REFER TO PAGE 9 FOR DETAILS)
BC20-20400 NEMA 5-20P PLUG LINE CORD (USA/NORTH AMERICA) BC20-20401 JA1-15P PLUG LINE CORD (JAPAN) BC20-20402 UK1-13P PLUG LINE CORD (UK) BC20-20403 CEE 7/7 “SCHUKO” PLUG LINE CORD (EURO)
NOTE: INCLUDES BC20-20221 GROUNDING ADAPTER
BC20-20409 BSS546A PLUG LINE CORD (INDIA/SOUTH AFRICA) BC20-20410 SW1-10P PLUG LINE CORD (SWITZERLAND) BC20-20412 IT1-10P PLUG LINE CORD (ITALY) BC20-20416 AUSTRALIA PLUG LINE CORD
Page 12
4
LAYOUT
RS-232 Communication Port
(Note: Only with Option –R)
3-1/2 digit LCD
Display
ISO Test key
Load Selection
NEMA 5-20R DUT
Analyzer Mode keys &
Earth/Ground Leakage
Lead Isolation Leakage ECG Waveforms
LCD Unit Indicators:
Ohms
10 Universal Patient Lead Inputs
Chassis Connector
Patient Lead
• V1/C1 • V2/C2
• V3/C3 • V4/C4
• V5/C5 • V6/C6
• 2.0 Hz
Analyzer Receptacle Control
Polarity
Earth/Ground
Fuse
Test Snaps:
High Level (+)
1.0 Ω
External Lead
Connector
3 Neon indicators for
Incoming Power status
This section looks at the layouts of the SA-2010S Models and gives descriptions of the elements that are present.
For all models, the user controls all Analyzer functions via light touch keys. There is an audio click when any key is depressed, while a “Razz” or error tone consisting of a rapid succession of beeps is sounded if an invalid key is depressed. Analyzer modes and status are indicated by 33 LEDs as well as 3 Neon lamps. A 3-1/2 digit LCD display conveys DUT test results to the user (Not applicable for ECG Waveforms mode)
100 uA
switch:
Selection and ECG/Performance Simulator keys & indicators:
Lead Selection:
• RA/R • LA/L
• RL/N (-) • LL/F
ECG Selection:
30, 60, 120, 240 BPM
Sine Perf. Wave. Selection:
•10, 60, 100 Hz Sq. Perf. Wave. Selection:
0.125, 2.0 Hz Tria. Perf. Wave. Selection:
Receptacle for
SA-2010S Layout (SA-2010S-R shown)
Volts Amps µAmps
indicators:
Mains Voltage Device (DUT) Current Earth Resistance
Enclosure Leakage Lead to Earth Leakage Lead to Lead Leakage
External Leakage
keys & indicators:
Hot Neutral Ground
& indicator
Page 13
5
Universal International Receptacle for DUT
Australian 10A @ 240V Receptacle for DUT
SA-2010S-INTL Layout SA-2010S-AUS Layout
The SA-2010S-INTL and SA-2010S-AUS Model Layouts are shown above, note that the only differences from the standard model are the Line cords and the DUT test Receptacles.
Note: The RS-232 connector isn’t shown above for these two models, but is an available option for both (see previous section).
Page 14
6
AHA Label
IEC Label
Description
RA
R
Right Arm
LA
L
Left Arm
RL
N
Right Leg (reference or ground)
LL
F
Left Leg
V1 V2 V3 V4 V5 V6
C1 C2 C3 C4 C5 C6
V Leads (V1-V6) (U.S. and Canada)
also referred to as pericardial, precordial or unipolar chest leads
Chest Leads (C1-C6) (International)
Display
The main information in the system is presented in the 3-1/2 digit LCD display. This data is provided as simple meter readings with the measurement units indicated to the right by one of four LED indicators.
Universal Patient Lead Connectors
The 10 Universal Patient Lead Connectors allow for 12 lead ECG simulation with independent outputs. AHA and IEC color-coded labels are located on the face of the unit to aid in connecting the corresponding U.S. and International Patient Leads.
Function Selection
Ten LEDs and two keys make up the Function Selection Section. The keys are up and down arrows. When depressed, they step the Analyzer through the available options. The LED next to the currently selected option is illuminated.
Load Selection
The unit may either use the AAMI ES1-1993 or IEC 601 Test load for measurements. This is selected by the Load Selection switch.
Page 15
7
Analyzer Test Receptacle Control
There are four keys and 8 LEDs in the Analyzer Test Receptacle Control Section. They allow the manual control of the power connections that are made to the DUT. Internally, a series of relays are switched by the microprocessor based on the keys that are depressed. The LEDs indicate the current state of the power connections to the Receptacle.
Note: The Forward/Reverse Polarity key has a safety delay feature, preventing damage to the internal relays and the DUT. When the key is depressed, the DUT power is disabled and the safety delay is activated. When this delay is complete, the internal relays switch the polarity to the DUT and apply power. This delay allows any reactive power stored in the DUT to self-discharge before the polarity is reversed.
Note: The unit will power up with the Neutral and Ground Closed, in Forward Polarity and with the Hot Open. It is recommended that the unit be returned to this condition when plugging and unplugging the DUT.
Patient Lead Control
There are ten patient leads. During testing, it is necessary to select between these leads, select all of them and apply High Voltage to them. This section provides the control keys to do these test configurations and the LEDs to indicate the current state.
There is one LED and an internal relay for each lead. The LEDs and the markings to the right of each LED indicate when that relay is on, thus selecting the indicated lead. The Up and Down arrow keys sequentially select each lead in order and scroll through from None, to 1-10 to All and around again.
To apply High Voltage to the leads, the “ISO” Test key is depressed. It is only active in the Lead Isolation mode. It is a momentary key so that High Voltage is only applied while the key is held down.
Waveform Selection
When the ECG Waveforms function is selected, the two keys and 10 LEDs become the Waveform Selection Control Section. The markings on the left side of the LEDs are used during these tests. The LEDs indicate which waveform is generated. The Up and Down arrow keys sequentially select each waveform. The microprocessor sends the selected waveform to a Digital to Analog converter that generates an accurate analog representation. This waveform is then sent through a resistor network, developing the appropriate signals on the output terminals.
Power Outlet Indicators
Three Neon indicators help verify the polarity and wiring of the wall receptacle that the Safety Analyzer is plugged into.
Page 16
8
HI Level Output (+)
A HI Level connector is located on the side of the unit for the high level ECG output signal (1 V p-p). The connection is between the High Level (+) snap and the RL (-) ECG lead.
Self-Test Snaps
There are two snaps on the side of the unit that allow for a quick self-test of the Analyzer. They provide a fixed 1.0 Ohm resistance to Earth/Ground and a 100 µAmp leakage current source to Earth/Ground when the Chassis lead is applied and the Analyzer set to the proper mode.
Connectors
There are two connectors for test cables on the unit. One is for the Chassis lead and the other is for one of two different leads used for external testing. The test cables simply plug into the sockets. There is a release pin on the cable plug that must be depressed to remove the cable.
Fuse
There is a fuse in the ground leg of the Analyzer Test Receptacle. This is to help prevent damage from excess ground current. It is located on the face for ease of replacement.
Test Receptacle
This receptacle is for the connection of the DUT. The Receptacle Rating depends on the specific Analyzer model. SA-2010S models use a Hospital Grade North American/USA standard NEMA 5-20R receptacle rated 20 Amps @ 125 VAC. SA­2010S-INTL models use a universal international receptacle rated 20 Amps @ 250 VAC. SA-2010S-AUS models use an Australian-specific AU1-10R receptacle rated 10 Amps @ 240 VAC. An external patch cord may be necessary to connect devices utilizing different types of plugs to the Analyzer receptacle.
Power Cord
The Power Cord, which is connected internally, provides power to both the Safety Analyzer and the DUT through the Test Receptacle. The Power Cord varies depending on the Analyzer model. SA-2010S models use a NEMA 5-15P plug designed to plug into a NEMA 5-15R or 5-20R Receptacle. SA-2010S-INTL models have a short IEC C20 plug for which a country-specific adapter cable must be connected – See the following section for this information. SA-2010S-AUS models use an Australian AU1­10P plug intended for operation with Australian and New Zealand AU1-10R Receptacles.
Page 17
9
Adapter Cable with Standard Universal Power Cord which
Connects to Standard Product Plug
Communications (Optional)
On Analyzer models with the –R option, a DB-9 RS-232 Communications port is provided to interface the analyzer with a PC. Data from the Analyzer display may be transferred over this communications link as well as full control over all analyzer functions and settings. For further details, see the Communication Protocol section.
Product Plug and Required International Connector (See Options Below)
Page 18
10
TESTING
NOTICE – PERFORMING TESTS
The SA-2010S Models allow the user a great deal of flexibility in testing. Any of the basic tests can be run and in almost any sequence. The information in this section presents a systematic approach that is just one way to proceed. It is only presented as a guide and it is the responsibility of the user to establish which tests are required.
REFER TO DUT MANUFACTURER’S SERVICE MANUAL FOR
TEST PROCEDURES AND MEASUREMENT LIMITS.
The Analyzer requires a good Earth/Ground connection for operation. It should be plugged into a “Hospital Grade” receptacle where available. This is necessary for both valid test results and personal safety.
Power Receptacle Confirmation
Once plugged in, the first step is to ensure that the wall receptacle the Safety Analyzer is plugged into is wired properly (120 VAC Non-Isolated Power Systems Only). There are three neon indicators in the unit that provide this confirmation. The REV indicator is red and the other two are green. If the two green indicators are on, the receptacle is wired correctly. If not, utilize the following patterns to help determine the problem. Do not proceed with any testing until you get only the two green lights.
NOTE: Neutral/Ground Reversal is not checked.
NOTE: These Indicators are valid only for 120 VAC Non-Isolated Power Systems.
Mains Voltage
With the Mains Voltage function selected, the display will show the Voltage that is present on the incoming power lines. This is measured from Line to Neutral. Note that the voltage may drop when the DUT is turned on. Ensure that this value is within the DUT specifications.
Page 19
11
Device Current
With the Device Current function selected, the display will show the current draw of the DUT. The Receptacle should be configured with HOT-CLOSED, NEUTRAL-CLOSED, GROUND-CLOSED and POLARITY-FWD. Refer to the Specifications section for current capacity and permitted duty cycle for this test mode.
Earth Resistance
With the Earth Resistance function selected, the display will show the resistance between the Chassis Test lead and Receptacle Earth/Ground. This resistance is a combination of the resistance within the DUT enclosure and the resistance of the Earth/Ground Lead in the DUT power cord.
NOTE: This test has no meaning for equipment that does not use a grounded cord. The test requires that the Chassis Test lead be plugged into the Chassis Connector.
The other end should be connected to a solid ground point on the DUT. The display is in hundredths of and will read to 19.99 . Over-range shows as
1_ _ _.
Earth/Ground Leakage Current
With the Earth/Ground Leakage function selected and the Ground-Open, the display will show the leakage current in the ground wire of the DUT.
NOTE: This test has no meaning for equipment that does not use a grounded cord. Selecting this function automatically opens the connection to Earth/Ground and passes
any leakage current through a 1000 load with either AAMI ES1-1993 or IEC 601 frequency compensation as selected by the Load Selection switch.
Enclosure Leakage
With the Enclosure function selected, the display will show the leakage current between the Enclosure (Chassis) and Earth/Ground.
The test requires that the Chassis Test lead be plugged into the Chassis Connector. The other end should be connected to a solid ground point on the DUT.
NOTE: If a non-conductive enclosure is used, a 200 cm2 conductive foil pad should be used. This foil is to be placed in close contact with the enclosure and connected to the Chassis Test lead.
Any leakage current will flow through the Chassis Test lead and then through a 1000 load with either AAMI ES1-1993 or IEC 601 frequency compensation as selected by the Load Selection switch.
Page 20
12
WARNING
The SA-2010S applies 110% of line voltage to the
Isolation test. Although this is current limited by
Isolation Test key.
Lead to Earth/Ground Leakage
With the Lead to Earth/Ground function selected, the display will show the leakage current between the selected lead and Earth/Ground.
Attach the patient leads to the connectors on the top of the Safety Analyzer. The Up and Down arrow keys may then be used to select any individual lead or all of the leads.
This test should be done for each lead individually and all leads together. This test measures the leakage current that would flow through the leads if the patient
were to come into contact with Earth/Ground.
Lead to Lead Leakage
With the Lead to Lead function selected, the display will show the leakage current between the selected Patient Lead and all other patient leads.
Attach the patient leads to the connectors on the top of the Safety Analyzer. The Up and Down arrow keys may then be used to select any individual lead. Internally, relays connect the leads as necessary. The LEDs indicate the selected lead.
This test should be done for each lead individually. This test measures the current that would flow from a lead to other leads. Normally
these are Auxiliary currents from bias, measurement and sensing circuits.
Lead Isolation
With the Lead Isolation function selected and the “ISO” Test key depressed, the display will show the leakage current between the selected Patient Lead(s) and Earth/Ground.
Attach the patient leads to the connectors on the top of the Safety Analyzer. The Up and Down arrow keys may then be used to select any individual lead and all of the leads.
Patient leads or External test leads during the
a 121 kΩ internal resistor, per standard test
specifications, care should be taken to prevent contact with this voltage. Do not touch the test
leads, connections or DUT while depressing the
Page 21
13
WARNING
The SA-2010S applies 110% of line voltage to the
Isolation test. Although this is current limited by
Isolation Test key.
As each lead and then All leads are selected, depress and hold the “ISO” Test key. This will apply 110% of the line voltage through a 121 kΩ resistor to the selected lead(s) and measure the current that flows to Earth/Ground through a 1000 load with either AAMI ES1-1993 or IEC 601 frequency compensation as selected by the Load Selection switch.
This test is to be done for each lead individually and All leads together. This test measures the leakage current that would flow through the lead(s) if the patient
were to come into contact with Line voltage. This is referred to as MAP (MAINS on Applied Parts).
Point to Point Measurements
The unit has the ability to measure Leakage Current, Isolation Leakage Current and Resistance between two points, utilizing two test leads. These tests are separated because they use a slightly different setup than the previous tests.
Point to Point Leakage Current
With the External function selected, the display will show the leakage current between the test leads.
The test requires that the Chassis Test lead be plugged into the Chassis Connector and the External Test lead be plugged into the External Connector. The other ends of the leads are then attached to the points of interest.
Any current flowing between the test points is passed through a 1000 load with either AAMI ES1-1993 or IEC 601 frequency compensation as selected by the Load Selection switch.
Point to Point Isolation Leakage Current
With the External function selected and the “ISO” Test key depressed, the display will show the isolation leakage current between the test leads.
Patient leads or External test leads during the
a 121 kΩ internal resistor, per standard test
specifications, care should be taken to prevent
contact with this voltage. Do not touch the test
leads, connections or DUT while depressing the
Page 22
14
The test requires that the Chassis Test lead be plugged into the Chassis Connector and the External Test lead be plugged into the External Connector. The other ends of the leads are then attached to the points of interest.
Depress and hold the Isolation Voltage key. This will apply 110% of the line voltage through a 121 kΩ resistor to the test leads and measure the current that flows through a 1000 load with either AAMI ES1-1993 or IEC 601 frequency compensation as selected by the Load Selection switch.
Point to Point Resistance
With the Earth Resistance function selected, the display will show the resistance between the two test leads.
The test requires two Chassis Test leads, one in the Chassis Connector and the second in the External Connector. The opposite cable ends should be connected to the points of interest.
NOTE: Remove any device plugged into the Analyzer Test Receptacle. If there is a DC voltage present between the two test points, the reading may contain an error. This can be checked by reversing the connections. If the readings differ, average the two to get the actual resistance value.
Page 23
15
PATIENT SIMULATOR
NOTICE – PERFORMING TESTS
The SA-2010S Models provide ECG and Performance Waveforms to aid in the verification of the basic operational characteristics of cardiac-monitoring equipment. The information in this section presents a listing of available waveforms. It is the responsibility of the user to establish which waveform verifications are required.
REFER TO DUT MANUFACTURER’S SERVICE MANUAL FOR
TEST PROCEDURES AND MEASUREMENT LIMITS.
Note: The LCD display is unused in these modes and will indicate an over-range condition shown as “1_ _ _”.
Pulse Waves
The SA-2010S simulates a Normal Sinus Rhythm pulse at 30, 60, 120 and 240 BPM. With ECG Waveforms function selected, use the arrow keys under the waveform LEDs to select the desired pulse. The LEDs will pulse with the waveform.
Performance Waves
Sine
The SA-2010S sends out Sine waves of selectable frequency - 10, 60 and 100 Hz. With ECG Waveforms function selected, use the arrow keys under the waveform LEDs to select the desired Sine wave output mode.
Square
The SA-2010S sends out Square waves of selectable frequency - 0.125 and 2 Hz. With ECG Waveforms function selected, use the arrow keys under the waveform LEDs to select the desired Square wave output mode.
Triangle
The SA-2010S sends out Triangle waves of frequency 2 Hz. With ECG Waveforms function selected, use the arrow keys under the waveform LEDs to select the Triangle wave output mode.
Page 24
16
COMMUNICATION PROTOCOL
The communication protocol provides a means to completely configure and use the Analyzer from a PC or other device with RS-232 communications interface. This provides for hands free or automated operation of the equipment.
Communication Port
The Serial port is configured as 115,200 Baud Rate, 8 Data Bits, 1 Stop Bit, and No Parity.
Command Syntax
The command description is broken into columns; the KEYWORD, the NODE and the VALUE.
The KEYWORD provides the name of the command. The actual name of the command consists of one or more keywords since SCPI commands are based on a hierarchical structure, also known as a tree system.
In such a system, associated commands are grouped together under a common node in the hierarchy, analogous to the way leaves at a same level are connected at a common branch. This and similar branches are connected to fewer and thicker branches, until they meet at the root of the tree. The closer to the root, the higher a node is considered in the hierarchy. To activate a particular command, the full path to it must be specified.
This path is represented in the following tables by placing the highest node in the left­most position. Further nodes are indented one position to the right, below the parent node.
The highest level node of a command is called the Keyword, followed by the Node, and then the value.
Some commands allow for reading and writing data and some commands are Read Only. To indicate a read function, a question mark (?) is placed at the end of the command path. For example, to change the mode to Earth Resistance measurement, “CONFigure:MODE ERESistance<cr>”, where <cr> indicates a carriage-return. To read the current operating mode of the device, a mode read command would be “CONFigure:MODE?<cr>”, which would return the current operating mode of the device.
Lowercase letters indicate the long-form of the command (for example, CONFigure:MODE ECGwaveforms) and can be omitted for simplification. Uppercase letters indicate the abbreviated, or short-form, of the commands and must be included (for example, CONF:MODE ECG).
All commands sent to the unit are terminated with a Carriage Return. NOTE: Commands can be entered in either upper or lowercase or a mixture of the two,
Page 25
17
SA-2010S-R Communication Command Summary
Keyword
Node
Value
Description
CONFigure
MODE
MVOLtage DCURrent ERESistance EGROund ENCLosure LEARth LLEAd LISOlation EXTernal ECGwaveforms
Mains Voltage Device Current Earth Resistance Earth / Ground Leakage Enclosure Leakage Lead to Earth Leakage Lead to Lead Leakage Lead Isolation External Measurement ECG Waveform Simulation
ECG
NSR30 NSR60 NSR120 NSR240 SINE10 SINE60 SINE100 SQU0.125 SQU2 TRI2
NSR, 30BPM NSR, 60BPM NSR, 120BPM NSR, 240BPM Sine Wave, 10Hz Sine Wave, 60Hz Sine Wave, 100Hz Square Wave, 0.125 Hz Square Wave, 2Hz Triangle Wave, 2Hz
LEADSelect
NONe RA LA RL LL V1 V2 V3 V4 V5 V6 ALL
No Leads Selected RA Lead Only LA Lead Only RL Lead Only LL Lead Only V1 Lead Only V2 Lead Only V3 Lead Only V4 Lead Only V5 Lead Only V6 Lead Only All Leads Active
HOT
OPEN CLOSed
Receptacle Hot Lead Open Receptacle Hot Lead Closed
NEUtral
OPEN CLOSed
Receptacle Neutral Lead Open Receptacle Neutral Lead Closed
GROUnd
OPEN CLOSed
Receptacle Ground Lead Open Receptacle Ground Lead Closed
POLarity
FWD REV
Receptacle Polarity Forward Receptacle Polarity Reversed (reverse HOT/NEUTRAL)
uppercase and lowercase. Commands sent to the device are not case sensitive. Upper case characters in the command summary table below are used to show the shortened versions of the commands if available.
Page 26
18
SA-2010S-R Communication Command Summary
(Continued)
Keyword
Node
Value
Description
ISO
ON OFF
Isolation Test ON Isolation Test OFF
SYSTem
UNITs?
Returns the units of Measure (Volts, Amps, uA, Ohms) [Read Only]
MEASurement?
Returns the latest measurement [Read Only]
KEY
MUP
Changes Mode Up one position
MDN
Changes Mode Down one position
LEUP
Changes Output / Lead Select Up one position
LEDN
Changes Output / Lead Select Down one position
HOT
Duplicates the HOT key press
NEUtral
Duplicates the Neutral Key press
GROUnd
Duplicates the Ground key press
POLarity
Duplicates the Polarity key press
ISO
Duplicates the ISO Test key press
MODel?
Returns Safety Analyzer Model [Read Only]
VERsion?
Returns Firmware Version [Read Only]
Page 27
19
LIMITED WARRANTY
WARRANTY: BC GROUP INTERNATIONAL, INC. WARRANTS ITS NEW PRODUCTS TO BE
FREE FROM DEFECTS IN MATERIALS AND WORKMANSHIP UNDER THE SERVICE FOR WHICH THEY ARE INTENDED. THIS WARRANTY IS EFFECTIVE FOR TWELVE MONTHS FROM
ANY OTHER WARRANTY EXPRESSED OR
MERCHANTABILITY
IS NOT LIABLE FOR ANY INCIDENTAL OR
NO PERSON OTHER THAN AN OFFICER IS AUTHORIZED TO GIVE ANY OTHER WARRANTY
THE PURCHASER'S SOLE AND EXCLUSIVE REMEDY SHALL BE: (1) THE REPAIR
OR REPLACEMENT OF DEFECTIVE PARTS OR PRODUCTS, WITHOUT CHARGE. (2) AT THE
MANUAL REVISIONS
Revision # Revisions Made
Rev 01 Origination Rev 02 Photos Updated Rev 03 Specifications Updated Rev 05 Address Updated Rev 06 Communication Protocol Added Rev 07 Communication Protocol Edited Rev 08 Pictures Updated Rev 09 Miscellaneous Edits Rev 10 Format Updated, Specifications Updated, Misc. Edits Rev 11 Miscellaneous Edits
THE DATE OF SHIPMENT. EXCLUSIONS: THIS WARRANTY IS IN LIEU OF
IMPLIED, INCLUDING, BUT NOT LIMITED TO ANY IMPLIED WARRANTY OF OR FITNESS FOR A PARTICULAR PURPOSE.
BC GROUP INTERNATIONAL, INC.
CONSEQUENTIAL DAMAGES.
OR ASSUME ANY LIABILITY.
REMEDIES:
OPTION OF BC GROUP INTERNATIONAL, INC., THE REFUND OF THE PURCHASE PRICE.
P:\MANUALS\BCGroup\\SA2010S_Series_UM_Rev11.docx
Page 28
20
MAINS VOLTAGE
RANGE
90 to 264 VAC, 50/60 Hz
ACCURACY
± 3% of Reading, ± 1 digit
DEVICE CURRENT
RANGE
0 to 19.99 Amps, RMS
ACCURACY
± 5% of Reading, ± 1 digit
EARTH RESISTANCE
RANGE
0 to 19.99 Ω
ACCURACY
0 to 1.99 Ω
± 1% Range
2.00 to 19.99 Ω
± 1% Range
CURRENT SOURCE
10 mA
(Note: 1 Amp Current Source Available as an optional
accessory, see description section for ordering
information on the CS-2000 Series)
LEAKAGE CURRENT
RANGE
0 to 1999 µA, RMS
ACCURACY
DC
± 1% Reading, ±1 digit
25 Hz up to 1 kHz
± 1% Reading, ±1 digit
1 kHz up to 100 kHz
± 2.5% Reading, ±1 digit
100 kHz to 1 MHz
± 5% Reading, ± 1 digit
LEAKAGE LOAD
1000 Ω
AAMI ES1-1993 or IEC 601
User Selectable
LEAD ISOLATION TEST
(MAP)
110% Line Voltage, ± 5%
Internally Current-Limited by 121 kΩ resistor
FUSE
250 mA, 250 V, 5x20 mm, Fast Acting
(Receptacle Ground Leg)
BC Part No. UF-0250-01
SPECIFICATIONS
Page 29
21
PATIENT SIMULATOR
NORMAL SINUS RHYTHM
30, 60, 120, 240 BPM
PERFORMANCE
WAVEFORMS
SINE – 10, 60, 100 Hz
SQUARE – 0.125, 2.0 Hz
TRIANGLE – 2.0 Hz
ACCURACY
FREQUENCY
± 0.5 %
AMPLITUDE
± 2 % Lead II
OUTPUTS
LEAD 1
1.75 mV
LEAD 2
2.75 mV
LEAD 3
1.00 mV
IMPEDANCE
LEAD TO LEAD
1000 Ω
LEAD TEST
< 1000 Ω
ELECTRICAL
OPERATING LINE
VOLTAGE
90 to 264 VAC, 50/60 Hz
LINE PLUG
SA-2010S MODELS
NEMA 5-15P
SA-2010S-INTL MODELS
Universal IEC C20
Receptacle Pigtail Cord
Must use Country-specific
Line Cord (see
Accessories section)
SA-2010S-AUS MODELS
AU1-10P (AS/NZS 3112)
DUT CURRENT CAPACITY
SA-2010S & SA-2010S-INTL
MODELS
15 Amps, 30 minutes
20 Amps, 5 minutes
SA-2010S-AUS MODELS
10 Amps, 30 minutes
DUT RECEPTACLE
SA-2010S MODELS
20 Amps @ 125 VAC
Capacity
Hospital Grade
NEMA 5-20R
Compatible with:
NEMA 5-15P and 5-20P
Plugs
Page 30
22
ELECTRICAL (continued)
DUT RECEPTACLE
(continued)
SA-2010S-INTL MODELS
20 Amps @ 250 VAC
Capacity
International Receptacle
Compatible with:
o NEMA 5-15P, NEMA
5-20P, NEMA 6-15P and NEMA 6-20P (US/NORTH AMERICA)
o UK1-13P and UK3-5P
(UK)
o SW1-10P
(SWITZERLAND)
o IT1-10P (ITALY) o IS1-16P (ISRAEL) o JA1-15P (JAPAN) o EU1-16P (EURO)
CEE 7/7 “SCHUKO”
(NOTE: MUST USE BC20-20221 GROUNDING ADAPTER)
o DE1-13P
(DENMARK)
o EUROPLUG CEE
7/16
SA-2010S-AUS MODELS
10 Amps @ 240 VAC
Capacity
AU1-10R (AS/NZS 3112)
Compatible with:
AU1-10P Plugs
POWER CONSUMPTION
5 VA (5 W)
PHYSICAL & ENVIRONMENTAL
DISPLAY
Non-Backlit 3½ Digit LCD
0.5 inches (12.7 mm) Digit Height
CONSTRUCTION
ENCLOSURE
ABS Plastic
OVERLAY
Back-printed Lexan
Page 31
23
PHYSICAL & ENVIRONMENTAL (continued)
SIZE
8.65 x 5.73 x 2.40 Inches (219.7 x 145.5 x 61.0 mm)
WEIGHT
3.5 Lbs (1.6 kg)
OPERATING RANGE
15 to 40 °C (59 to 104 °F)
10 to 90% RH, Non-Condensing
STORAGE RANGE
-20 to 65 °C (-4 to 149 °F)
Page 32
24
NOTES
Page 33
25
NOTES
Page 34
26
NOTES
Page 35
Page 36
BC GROUP INTERNATIONAL, INC.
3081 ELM POINT INDUSTRIAL DRIVE
ST. CHARLES, MO 63301
USA
1-800-242-8428 1-314-638-3800
www.bcgroupintl.com
sales@bcgroupintl.com
SA-2010S Series User Manual
06/12 – Rev 11
Copyright © 2012
Made in the USA
Loading...