BC Biomedical PS-2110 User Manual

MULTI-PARAMETER
PATIENT SIMULATOR
PS-2110
USER MANUAL
BC BIOMEDICAL
PS-2110
TABLE OF CONTENTS
WARNINGS ................................................................................................................. 2
DESCRIPTION............................................................................................................. 3
OVERVIEW.................................................................................................................. 7
ECG--NORMAL SINUS RHYTHM................................................................................ 15
ECG--ARRHYTHMIAS ................................................................................................. 17
ECG--PERFORMACE.................................................................................................. 21
BLOOD PRESSURE .................................................................................................... 23
RESPIRATION............................................................................................................. 25
TEMPERATURE .......................................................................................................... 27
SETUP ......................................................................................................................... 29
POWER UP SETTINGS............................................................................................... 33
SpO2 (OPTION)............................................................................................................ 35
OUTPUT CONNECTIONS ........................................................................................... 37
MANUAL REVISIONS.................................................................................................. 39
WARRANTY................................................................................................................. 39
SPECIFICATIONS ....................................................................................................... 41
NOTES......................................................................................................................... 45
All connections to patients must be removed
before connecting the Device Under Test (DUT)
to the Simulator. A serious hazard may occur if
the patient is connected when testing
Do not connect any leads from the patient
directly to the Simulator or DUT.
iii
BC GROUP INTERNATIONAL, INC. RESERVES THE RIGHT TO MAKE
CHANGES TO ITS PRODUCTS OR SPECIFICATIONS AT ANY TIME,
WITHOUT NOTICE, IN ORDER TO IMPROVE THE DESIGN OR
PERFORMANCE AND TO SUPPLY THE BEST POSSIBLE PRODUCT.
THE INFORMATION IN THIS MANUAL HAS BEEN CAREFULLY CHECKED AND IS BELIEVED TO BE ACCURATE. HOWEVER, NO RESPONSIBILITY IS
ASSUMED FOR INACCURACIES.
WARNING
with the Simulator.
N O T I C E
iii
Manual PS-2110 Copyright © 2004
www.bcgroupintl.com Made in the USA
12/04 Rev 04
BC BIOMEDICAL
PS-2110
PATIENT SIMULATOR
The Model PS-2110 is a Microprocessor based Patient Simulator. It provides ECG, Blood Pressure, Respiration and Temperature Simulation. There are 12 arrhythmias, a pacemaker rhythm, a Fetal/Maternal rhythm, seven waveforms with constant QRS duration and 12 machine performance testing waveforms.
The PS-2110 makes viewing and selecting the desired waveforms and parameters quick and intuitive, with all operational information being available at one time on a cursor-based graphic display, allowing for easy maneuvering through parameters and scrolling through available options.
The following are highlights of some of the main features:
GENERAL
SIMPLE TO OPERATE
NO CODES TO REMEMBER OR ENTER
GRAPHICS DISPLAY WITH SIMULTANEOUS DETAILED STATUS OF
PARAMETERS AND SCROLLING CONTROL OF OPTIONS
DROP DOWN CHOICE SCREENS LIST ALL OPTIONS FOR PARAMETERS
SPECIAL POWER UP FEATURE ALLOWS THE USER TO CHOOSE TO USE
DEFAULT, LAST OR CUSTOM SETTINGS
AUTO SEQUENCES FOR BPM, STATIC-PRESSURE LEVELS AND PERFORMANCE
10 UNIVERSAL PATIENT LEAD CONNECTORS
MINI-DIN CONNECTOR FOR BP CABLES
9 VOLT BATTERY POWER
% BATTERY LIFE DISPLAY
LOW BATTERY INDICATOR
AVAILABLE BATTERY ELIMINATOR
DISPLAY BACKLIGHT
FLASH PROGRAMMABLE FOR UPGRADES
PS-2110
ECG FUNCTIONS
The unit can produce a wide variety of ECG simulations. The user simply selects the parameters that match the desired output.
RATE: 30,60,80,120,180,240,300 BPM
AMPLITUDE:
0.5,1.0,1.5,2.0 mV (Lead II)
AUTOMATIC MODE
PACEMAKER FUNCTION
A pacemaker waveform may be simulated.
WAVEFORM: ASYNCHRONOUS
PULSE HEIGHT:
1.0 mV
PULSE WIDTH:
1.0 ms
FETAL/MATERNAL
A combination of fetal and maternal ECG waveforms may be simulated.
FETAL HEARTRATE: 120 BPM
ARRHYTHMIA FUNCTIONS
The unit can simulate 12 different arrhythmias. Where applicable, both manual and automatic triggering of the waveform is available.
12 DIFFERENT ARRHYTHMIAS
MANUAL AND AUTOMATIC TRIGGERING
Description
ECG-PERFORMANCE FUNCTIONS
The unit will generate Sine, Square and Triangular waveforms with adjustable amplitudes for performance testing. A special Automatic mode is available to auto sequence through the entire range of waveforms.
SINE:
0.1, 0.5, 5, 10, 40, 50, 60,100 Hz
SQUARE:
0.125, 2 Hz
TRIANGLE: 2 Hz
AMPLITUDE:
0.5,1.0,1.5,2.0 mV (Lead II)
AUTOMATIC MODE
RESPIRATION
Respiration is simulated at 8 different rates plus apnea, with the ability to select from 2 Baseline Impedances, the Lead in which it will appear, and the Delta Ohms (amplitude) of the signal.
RATE: 15,20,30,40,60,80,100,120 BrPM
BASELINE IMPEDANCE: 500,1000 OHMS
LEAD: LA or LL
DELTA IMPEDANCE:
0.1,0.2,0.5,1.0,2.0,3.0 OHMS
APNEA: CONTINUOUS (0 BrPM)
PS-2110
BLOOD PRESSURE
Both static and dynamic invasive pressures are simulated. In the static mode the BP output is fixed at the selected level or sequenced through the list using the Automatic mode selection. In the Dynamic mode the selected waveform is synchronized with the ECG and provides a continuous output.
STATIC: 0,20,40,80,100,200,250,300 mmHg
AUTOMATIC STATIC PRESSURE MODE
6 DYNAMIC WAVEFORMS
SENSITIVITY:
5uV/V/mmHg and 40uV/V/mmHg
TEMPERATURE
The unit will simulate three temperatures. This is done by providing the necessary ohmic levels for both the YSI 400 and 700 Series thermistors.
YSI SERIES 400 and 700 SIMULATION
SELECTIONS:
30,37,40 °C (86.0,98.6,104.0 °F)
LEAD TEST FUNCTION
The unit provides a set of test terminals to quick check leads. It will determine if a lead has less than 1000 Ohms resistance.
SPO2 SIMULATION (Option)
When used with the MSP-2100 external module and FingerSim family of SpO2 finger simulators, the system will provide a pulse synchronized SpO2 output for all NSR rates.
RATE: 30,60,80,120,180,240,300 BPM
SpO
OUTPUT:
2
80,90,97 %
OVERVIEW
This section looks at the layout of a PS-2110 and gives descriptions of the elements that
are present.
Lead Continuity
10 Universal Patient Lead Connectors: RA R LA L RL(-) N LL F V1 C1 V2 C2 V3 C3 V4 C4 V5 C5 V6 C6
7 Light Touch Keys for Selecting Parameters and Settings: LEFT and RIGHT Curved Arrows for Moving through Parameters UP and DOWN Arrows for Scrolling through Options ENTER for Selecting Option CHOICES for Displaying Submenu of All Options for a Given Parameter QUIT for Returning to Previous Status
4 Light Touch Keys for Category Selection: Normal Sinus Rhythm Arrhythmias Performance Setup
Test Connections
9V Battery Compartment (Back)
LCD Graphical Display: Shows Parameters for ECG, Respiration, Blood Pressure and Temperature
6-Pin Mini-DIN Plug Connector for Blood Pressure Cable
8-Pin Mini-DIN Plug Connector for Temperature Cable
7-Pin Mini-DIN Plug Connector for Aux Functions
AC Adaptor/ Battery Eliminator
Power Key for Turning Unit On and Off
Back Light Key for turning on and off the backlight
PS-2110
General Operation
The unit is controlled by 13 light touch keys. They allow the user to move around within the
displayed parameters, select the desired options, choose a specific category and control
the setup and power for the unit. When a key is depressed there is an audio click when it is
accepted, or a razz tone if the key is invalid.
A graphics LCD display provides the user with information about the current status of the
ECG, Respiration, Blood Pressure and Temperature settings. The keys
move the block cursor through the displayed information; highlighting the parameter
available for selection. The keys change the options for the highlighted
parameter. The cursor begins flashing if the parameter has been changed. The
key selects the changed option. The key returns to the previous state,
without any changes being made.
To make option selection even easier and to make memorizing and using codes
unnecessary, the key will bring up a screen that displays all the options for the
selected parameter. The and keys can then be used to quickly
scroll through the available options and select the desired setting.
Three category keys allow for quick setting of output waveforms. The , and
keys move the display directly to the selected category. The or
keys can then be used to scroll through and select the desired
settings.
Overview
The key opens a screen that allows the user to select the unit’s general output
settings, as well as setup for the system.
Category Keys
The category, and keys can be used to select ECG waveforms.
The key enters the NSR category and gives access to the seven BPM (Beats Per
Minute) settings with constant QRS duration.
The key enters the arrhythmia category and changes the first line in the display to
the first arrhythmia choice.
The key enters the machine performance testing category and changes the first
line in the display to the first performance waveform choice.
Power Key
The key turns the unit on and off. To turn off the unit, the key must be held for 1
second.
PS-2110
Backlight
The Graphic LCD display may be viewed with or without the backlight. Depressing any key
will activate the backlight. However, since the backlight will drain the battery if left on, it will
automatically shut off after a few seconds when running on battery power. (Note: This time
is selectable in the System Setup screen).
The intensity of the backlight can be adjusted in the System Setup screen to conserve
battery life.
The key is provided to toggle the backlight on or off at any time.
NOTE: The backlight parameter in the System Setup screen may be set to Off, 1-30 sec
Timed or Manual.
ECG Waveforms
The microprocessor has stored in its memory all of the digitalized waveforms. It sends the
individual lead waveforms to D/A converters, which generate accurate analog
representations. The waveforms are then sent through resistor networks, developing the
appropriate signals on the output terminals.
10
Overview
Respiration
Respiration waveforms are provided that have adjustable rates from 15 to 120 BrPM
(Breaths Per Minute) as well as an Apnea (0 BrPM) setting. The signal is generated by
a variation in the impedance in either the LL or LA lead (selectable). The amplitude is
settable from .5 to 3.0 ohms.
Blood Pressure
There is a 6-pin mini-DIN plug connector on the right side of the unit for connection of a
Blood Pressure cable. Full blood pressure simulation is available through this
connector. The circuit is totally isolated.
Temperature
There is an 8-pin mini-DIN plug connector on the right side of the unit for connection of
a Temperature cable. Temperatures are simulated for both YSI 400 and YSI 700 probe
types. There are three different temperatures selectable for each.
11
PS-2110
Universal Patient Lead Connectors
The 10 Universal Patent Lead Connectors allow for 12 lead ECG simulation with
independent outputs. AHA and IEC color-coded labels are located on the face of the unit to
aid in connecting the corresponding U.S. and International Patient Leads.
AHA Label IEC Label Description
RA R Right Arm
LA L Left Arm
RL N
Right Leg (reference or ground)
LL F Left Leg
V Leads (U.S. Canada),
V1 V2 V3 V4 V5 V6
C1 C2 C3 C4 C5 C6
also referred to as pericardial, precordial or unipolar chest
Chest Leads
leads
(International)
High Level Output (+)
A high level ECG output signal (200 x Amplitude Setting) is available in the BP 6-Pin
mini-DIN connector.
Lead Test Terminals
There are two test terminals on the top of the unit that allow for a quick test of the
continuity of the lead cables. Connecting one end of the cable to one terminal and the
other end to the other terminal will test the cable. If the cable is OK (less than 1000
ohms), the LEAD TEST ACTIVE screen will be displayed.
12
Overview
Auto Power Off
The unit may be programmed to automatically turn off after a selected number of minutes
of no key activity to conserve the battery. (Note: This time is selectable in the System
Setup screen).
Power Supply
The unit utilizes two 9 Volt Alkaline Batteries in the rear battery compartment. When the
unit detects a LOW BATTERY condition (5% Battery Life), a warning window will appear
once per minute to alert the user. The key may be used to clear this window and
continue use of the unit. If the battery is not replaced before the battery reaches a critical
level (0 % Battery Life), the unit will shut down. (The percentage of life left in the batteries
can be viewed in the System Setup screen.)
Battery Eliminator
The unit has a 2.1 mm micro jack for connecting a 9-Volt AC battery eliminator (Optional).
The adapter will power the unit, but will not charge the battery.
Power Up Settings
The unit may be setup to turn on using either the factory default settings, the same settings
that it had when last turned off or a custom set of parameters as previously saved by the
user (See Power Up Settings section for details).
13
PS-2110
Automatic Modes
The ECG NSR Rate, ECG Performance and Static Blood Pressure Parameters all allow for
an automatic setting. In each of these, the unit will sequence through the full range of
settings automatically at a fixed rate (as selected in the Auto Step Time Parameter). When
in this mode, the time remaining in each step is displayed.
The key may be used to manually advance to the next step. The key is used
to terminate the mode.
14
(
(
ECG – NORMAL SINUS RHYTHM
The PS-2110 can send waveforms to ECG machines in 3, 5 or 12 lead configurations. It has independent outputs for each signal lead, referenced to the right leg.
Normal Sinus Rhythm (NSR) occurs when the heartbeat is normal, beating at a rate
between 50 and 100 BPM with a standard QRS waveform shape and height. The PS-2110
simulates the NSR with a default pulse of 80 BPM, amplitude of 1.0 mV on Lead II and P-R
interval of 160 milliseconds.
The PS-2110 is placed into NSR mode by pressing the category key.
The display will resemble the following:
RATE
30 BPM 60 BPM 80 BPM* 120 BPM 180 BPM 240 BPM 300 BPM Auto
* Indicates Default Setting
See Power Up Settings)
PATIENT MODE
Adult
80 mS QRS)*
AMPLITUDE
0.5 mV
1.0 mV*
1.5 mV
2.0 mV
The rates and amplitudes can be selected by using to highlight the
parameter to change and using to scroll to the desired value. Then
is used to accept the new setting.
15
PS-2110
Alternately, to see a submenu of all the options for a highlighted parameter, use .
Use to scroll to the desired option. Then is used to accept the new
setting.
Auto Rate
If the BPM parameter is set to AUTO, the unit will automatically sequence through all of the
BPM settings, starting with 30 BPM, incrementing at a fixed interval. The interval may be
set in the System Setup Menu under “Auto Step Time”.
Displays time (seconds) remaining before advancing to next rate.
The key can be used to exit the Auto Mode during the sequence.
16
(
ECG – ARRHYTHMIAS
The PS-2110 can send arrhythmia waveforms to ECG machines in 3, 5 or 12 lead
configurations. It has independent outputs for each signal lead, referenced to the right leg.
There are 12 Arrhythmias available that model abnormal heartbeats, plus Paced and
Fetal/Maternal. The PS-2110 is placed into ARRHYTHMIA mode by pressing the
category key.
The display will resemble the following:
ARRHYTHMIAS
Atrial Fib – Coarse 2nd Deg Heart Block RT Bundle Branch Block Atrial PAC – Auto Atrial PAC – Man PVC 1 – Auto PVC 1 – Man PVC 1 Early – Auto PVC 1 Early – Man PVC 1 R on T – Auto PVC 1 R on T – Man Multifocal PVCs – Auto Multifocal PVCs – Man Bigeminy Run of 5 PVCs – Auto Run of 5 PVCs – Man Vent Tach Vent Fib – Coarse Paced Fetal/Maternal
* Indicates Default Setting
See Power Up Settings)
AMPLITUDE
0.5 mV
1.0 mV*
1.5 mV
2.0 mV
17
PS-2110
The arrhythmias and amplitude can be selected by using to highlight the
parameter to change and using to scroll to the desired option. Then
is used to accept the new setting.
Alternately, to see a submenu of all the options for a highlighted parameter, use
Use to scroll to the desired option. Then is used to accept the new
setting.
Auto/Manual
There are 6 arrhythmias that have both Automatic and Manual versions. Both versions
output the same waveform; however, in the Manual version, the arrhythmia is triggered
each time is depressed. In the Auto versions, the arrhythmia is automatically triggered
periodically.
18
Arrhythmias
The following is a brief description of how the PS-2110 simulates the available Arrhythmias:
Abbreviation Arrhythmia Description
Absence of P-wave and irregular P-R
Atrial Fib
2nd Deg Heart Block
Rt Bundle Branch Block
Atrial PAC – Auto
Atrial PAC – Man
PVC 1 – Auto
PVC 1 – Man
PVC 1 Early ­Auto
PVC 1 Early ­Man
PVC 1 R on T – Auto
PVC 1 R on T – Man
Multifocal PVCS – Auto
Atrial Fibrillation
Second Degree Heart Block
Right Bundle Branch Block
Premature Atrial Contraction
Premature Atrial Contraction
Standard Type 1 Premature Ventricular Contraction
Standard Type 1 Premature Ventricular Contraction
Early Type 1 Premature Ventricular Contraction
Early Type 1 Premature Ventricular Contraction
R on T Type 1 Premature Ventricular Contraction
R on T Type 1 Premature Ventricular Contraction
Multifocal Premature Ventricular Contraction
19
interval rate (Continuous) 80 BPM with increasing P-R interval for four beats (160, 220, 400, 470 ms) followed by a P wave without a QRS (Continuous) 80 BPM with Normal P-wave and P-R interval but wider QRS complexes (Continuous) NSR of 80 BPM with Periodic Abnormal 25 % early P waves (PAC, 7 NSR) (Continuous) NSR of 80 BPM with Periodic Abnormal 25 % early P waves (One-Time event) NSR of 80 BPM with periodic left focus premature ventricular beats with 20% premature timing (PVC Type 1, 9 NSR) (Continuous) NSR of 80 BPM with periodic left focus premature ventricular beats with 20% premature timing (One-Time event) NSR of 80 BPM with periodic left focus premature ventricular beats with 33% premature timing (PVC Type 1, 9 NSR) (Continuous) NSR of 80 BPM with periodic left focus premature ventricular beats with 33% premature timing (One-Time event) NSR of 80 BPM with periodic left focus premature ventricular beats with 65% premature timing, placing R on the previous T (PVC Type 1, 9 NSR) (Continuous) NSR of 80 BPM with periodic left focus premature ventricular beats with 65% premature timing, placing R on the previous T (One-Time event) NSR of 80 BPM with Type 1 and Type 2 PVCs (PVC Type 1, 2 NSR, PVC TYPE 2, 2 NSR)
PS-2110
Multifocal PVCS – Man
Bigeminy
Run of 5 PVCs – Auto
Run of 5 PVCs – Man
Vent Tach
Vent Fib – Coarse
Paced
Fetal / Maternal
Multifocal Premature Ventricular Contractions
Bigeminal Rhythm
Run of 5 Premature Ventricular Contractions
Run of 5 Premature Ventricular Contractions
Ventricular Tachycardia
Ventricular Fibrillation
Paced Rhythm
Fetal/Maternal ECG
(Continuous) NSR of 80 BPM with Type 1 and Type 2 PVCs (PVC Type 1, 2 NSR, PVC TYPE 2) (One-Time event) NSR of 80 BPM with every other beat a Type 1 PVC (Continuous) NSR of 80 BPM with periodic group of 5 Type 1 PVCs (5 PVC Type 1, 36 NSR) (Continuous) NSR of 80 BPM with periodic group of 5 Type 1 PVCs (One-Time event) 160 BPM, No P-wave, Beats similar to Type 1 PVC (Continuous) Irregular waveform with no real P-wave or clear R-R interval (Continuous) Ventricular paced beats at 75 BPM with no P-waves (Continuous) Maternal NSR at 80 BPM with Fetal Heart Rate at 120 BPM (Continuous)
20
(
ECG – PERFORMANCE
The PS-2110 can send performance waveforms to ECG machines in 3, 5 or 12 lead
configurations. It has independent outputs for each signal lead, referenced to the right leg.
There are 11 Performance waves available for testing and verifying. The PS-2110 is
placed into PERFORMANCE mode by pressing the category key.
The display will resemble the following:
WAVEFORM
Square Wave .125 Hz Square Wave 2 Hz* Triangle Wave 2 Hz Sine Wave 0.1 Hz Sine Wave 0.5 Hz Sine Wave 5 Hz Sine Wave 10 Hz Sine Wave 40 Hz Sine Wave 50 Hz Sine Wave 60 Hz Sine Wave 100 Hz Auto Wave
* Indicates Default Setting
See Power Up Settings)
AMPLITUDE
0.5 mV
1.0 mV*
1.5 mV
2.0 mV
These waves and amplitudes can be selected by using to highlight the
parameter to change and using to scroll to the desired option. Then
is used to accept the new setting.
21
PS-2110
Alternately, to see a submenu of all the options for a highlighted parameter, use .
Use to scroll to the desired option. Then is used to accept the new
setting.
NOTE: Respiration and Blood Pressure outputs are disabled during performance waves.
Auto Wave
If the Performance parameter is set to AUTO, the unit will automatically sequence through
all of the performance waves, starting with Square Wave .125 Hz, incrementing at a fixed
interval. The interval may be set in the System Setup Menu under “Auto Step Time”.
A countdown timer is shown in the display:
Displays time (seconds) remaining before advancing to next waveform.
The key can be used to exit the Auto Mode during the sequence.
22
(
BLOOD PRESSURE
NOTE: The Transducer Sensitivity (5 uV/V/mmHg or 40 uV/V/mmHg) must be set to
correlate with the monitoring equipment before simulation can begin. (See SETUP for
selection information).
The PS-2110 offers one Blood Pressure Channel and will simulate the set Blood Pressure
wave during ECG waveforms where it occurs.
There are 14 Blood Pressure settings available. Each of the six dynamic waveforms will
synchronize with the NSR rate or arrhythmia selection.
The display will resemble the following:
WAVEFORM
Arterial 120/80 Left Vent 120/0 Right Vent 25/0 Pulm Artery 25/10 CVP 15/10 PAW 10/2 Static 0 mmHg* Static 20 mmHg Static 40 mmHg Static 80 mmHg Static 100 mmHg Static 200 mmHg Static 250 mmHg Static 300 mmHg Auto Static Pressure
* Indicates Default Setting
See Power Up Settings)
23
PS-2110
These settings can be selected by using to highlight the parameter to
change and using to scroll to the desired option. Then is used to
accept the new setting.
Alternately, to see a submenu of all the options for a highlighted parameter, use .
Use to scroll to the desired option. Then is used to accept the new
setting.
Auto Static Pressure
If Auto Static Pressure is selected, the channel will automatically sequence through all of
the Static Pressure settings, starting with 0 mmHg, incrementing at a fixed interval. The
interval may be set in the System Setup Menu under “Auto Step Time”.
Displays time (seconds) remaining before advancing to next static pressure.
The key can be used to exit the Auto Mode during the sequence.
24
(
RESPIRATION
NOTE: The delta ohm Respiration Signal can be inserted in either the LL or LA lead. The
Baseline impedance can be set to either 500 or 1000 Ohms. These must be set to
correlate with the monitoring equipment before simulation can begin. (See SETUP for
selection information).
There are 9 rate settings available.
The display will resemble the following:
Apnea 15 BrPM 20 BrPM* 30 BrPM 40 BrPM 60 BrPM 80 BrPM 100 BrPM 120 BrPM
* Indicates Default Setting
See Power Up Settings)
AMPLITUDE
0.1 Ω
0.2 Ω
0.5 Ω
1.0 Ω*
2.0 Ω
3.0 Ω
25
PS-2110
These rates and Impedance Variations can be selected by using to
highlight the parameter to change and using to scroll to the desired option.
Then is used to accept the new setting.
Alternately, to see a submenu of all the options for a highlighted parameter, use .
Use to scroll to the desired option. Then is used to accept the new
setting.
26
(
TEMPERATURE
The PS-2110 simulates 3 temperatures that are independent from the rest of the functions
of the unit. The temperature setting can be selected at any time.
The output will simulate both YSI 400 and YSI 700 Temperature probes.
(Note: Both outputs are available at the output connector simultaneously.)
The display will resemble the following:
* Indicates Default Setting
See Power Up Settings)
These temperatures can be selected by using to highlight the parameter to
TEMPERATURE
30˚C 86.0˚F 37˚C 98.6˚F* 40˚C 104.0˚F
change and using to scroll to the desired option. Then is used to
accept the new setting.
Alternately, to see a submenu of all the options for a highlighted parameter, use .
Use to scroll to the desired option. Then is used to accept the new
setting.
27
PS-2110
This page intentionally left blank.
28
(
SETUP
The PS-2110 allows for setup of the Outputs and the System Parameters through the
category key. Depressing the key multiple times toggles between the Output and
System Setup screens.
The Output Setup screen allows for the setting of the SpO2 Output, Respiration Baseline
Ohms, Respiration Lead and the Blood Pressure Sensitivity parameters. These should be
set according to the device under test.
Output Setup
The Output Setup screen allows for the setting of the parameters that control the placement
and level of the outputs.
The display will resemble the following:
500 Ω 1k Ω*
Enabled Disabled*
5 uV/V/mmHg* 40 uV/V/mmHg
* Indicates Default Setting
See Power Up Settings)
Lead LA* Lead LL
29
PS-2110
System Setup
The System Setup screen allows for the setting of the parameters controlling various
function of the unit as well as the viewing of Battery Life and Software information.
The display will resemble the following:
These settings can be selected by using to highlight the parameter to
change and using to scroll to the desired option. Then is used to
accept the new setting.
Use or , or to exit from the setup screens.
30
Setup
The following is a brief description of the parameters and the available range of settings:
Parameter Description Range
The elapsed time after which the unit will automatically power down. This
Auto Off Timer
timer is reset by each key depression.
0-30 min (Setting the value to 0 eliminates this feature.) Off – Always off 1-30 sec – The elapsed time after
Backlight Timed
which the backlight will automatically turn off.
Off, 1-30 sec, Manual
Manual – The backlight will be manually controlled by backlight key. Sets the intensity of the backlight.
Backlight Intensity
(Note: Lower intensities extend
0-100%
battery life.)
Contrast Adjust
Sets the contrast of the display screen.
0-20
Displays current life of the batteries.
Battery Life
At 5%, a warning screen will appear. At 0%, the unit will power down
5-100%
(Read Only) automatically. Selects the values that will be used
Power up with
when the unit is first turned on. It is also used to Set the Custom Defaults,
Default/Last/Custom/
Set Custom Defaults if used. (See Power Up Settings). Sets the interval that is used with the
Auto Step Time
Auto increment features in BPM, BP
1 to 60 sec
Rate and Performance.
Software Displays current software program. (Read Only)
31
PS-2110
This page intentionally left blank.
32
POWER UP SETTINGS
The PS-2110 allows the user to tailor the settings that the unit will have on Power Up. The
“Power Up With” parameter in the System Setup Menu allows for the selection of either
Default, Last or Custom selections.
Default
If this option is selected the following settings will be used every time the unit is turned on.
ECG – NSR: 80 BPM, 1.0 mV, Adult QRS, SpO2 Output Disabled
ECG – Performance: 2 Hz Square Wave, 1.0 mV
Respiration: 20 BrPM, delta 1.0 ohms, 1000 ohms baseline, LA lead
Blood Pressure: 0 mmHg, 5 uV/V/mmHg sensitivity
Temperature: 37 C (98.6 F)
System Setup:
Auto Timer Off 30 min
Backlight Time 5 sec
Backlight Intensity 100%
Contrast Adjust 10
Power Up With Default
Auto Step Time 5 sec
33
PS-2110
Last
If this option is selected, the unit will remember the settings that were being used when it
was turned off and bring them back when the power is turned on.
Custom
If this option is selected, the user may save a unique set of default parameters and the unit
will recall them every time the power is turned on.
Set Current as Custom
To create the set of custom default parameters, this fourth choice is provided in this
parameter. The user simply configures the unit to the desired default conditions, selects
this option and presses . The current configuration is then saved as the Custom
Power up values.
34
SPO2 (Option)
The PS-2110 has the ability to drive an external SpO2 module. This module (MSP-2100)
accepts the FingerSim family of SpO2 finger simulators (fingers are available with SpO2 of
80, 90 and 97 %). The output pulses the fingers at the NSR BPM rate. The output is off in
Arrhythmia and Performance Modes.
The module plugs directly into the AUX (7 pin mini din) connector and is powered from the
PS2100. The output is only functional when the unit is powered from the Battery Eliminator
provided with the MSP-2100 Module, since the batteries do no have enough power to run
this option.
The output is enabled and disabled in the Setup Output screen.
35
PS-2110
This page intentionally left blank.
36
OUTPUT CONNECTIONS
The following are representations of the socket connectors found on the unit. They are
viewed as if looking at the socket in the unit, not the cable pins.
37
PS-2110
This page intentionally left blank.
38
A
MANUAL REVISIONS
Revision # Program # Revisions Made
Rev 01 DT7346 Origination
Rev 02 DT7346 Preliminary Manual
Rev 03 DT7346 Updates to Preliminary Manual
Rev 04 DT7346 Overlay Additions
WARRANTY
FROM DEFECTS IN MATERIALS AND WORKMANSHIP UNDER THE SERVICE FOR WHICH THEY
RE INTENDED. THIS WARRANTY IS EFFECTIVE FOR TWELVE MONTHS FROM THE DATE OF
SHIPMENT.
EXCLUSIONS: THIS WARRANTY IS IN LIEU OF ANY OTHER WARRANTY EXPRESSED OR IMPLIED, INCLUDING, BUT NOT LIMITED TO ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE.
BC GROUP INTERNATIONAL, INC. IS NOT LIABLE FOR ANY INCIDENTAL OR CONSEQUENTIAL DAMAGES.
NO PERSON OTHER THAN AN OFFICER IS AUTHORIZED TO GIVE ANY OTHER WARRANTY OR ASSUME ANY LIABILITY.
REMEDIES: THE PURCHASER'S SOLE AND EXCLUSIVE REMEDY SHALL BE: (1) THE REPAIR OR REPLACEMENT OF DEFECTIVE PARTS OR PRODUCTS, WITHOUT CHARGE. (2) AT THE OPTION OF BC GROUP INTERNATIONAL, INC., THE REFUND OF THE PURCHASE PRICE.
P:\MANUALS\BCGroup\Published Versions\Ps2000\PS2110_UM_Rev04.doc
: BC GROUP INTERNATIONAL, INC. WARRANTS ITS NEW PRODUCTS TO BE FREE
LIMITED WARRANTY
39
PS-2110
This page intentionally left blank.
40
SPECIFICATIONS
ECG NSR
Rate 30,60,80,120,180,240,300 BPM Accuracy +/- 1% Amplitude 0.5, 1.0, 1.5, 2.0 mV Accuracy +/- 2% @ Lead II High Level 200 times Amplitude Accuracy +/- 5%
ECG Performance
Sine Wave 0.1,0.5,5,10,40,50,60,100 Hz Square Wave 0.125, 2.000 Hz Triangle Wave 2.000 Hz
Rate Accuracy +/- 1% Amplitude Accuracy +/- 2% @ Lead II
Static Pressure 0,20,40,80,100,200,250,300 mmHg Accuracy +/- (2% of Reading + 2 mmHg) Impedance 300 ohms Accuracy +/- 10% Excitation Range 2 to 16 V RMS Excitation Frequency DC to 5 KHz Sensitivity 5 or 40 uV/V/mmHg
Blood Pressure
41
PS-2110
Respiration
Rate
Accuracy +/- 1% Impedance delta 0.1,0.2,0.5,1.0,2.0,3.0 ohms Accuracy +/- 10% Baseline 500,1000 ohms Accuracy +/- 5% Lead LA or LL
Apnea,15,20,30,40,60,80,100,120
BrPM
Temperature
Temperature 30,37,40 C (86.0,98.6,104.0 F) Accuracy +/- 0.1 C Type YSI Series 400 and 700
Output
LEAD TO LEAD IMPEDENCE (RL, LL, RA, LA)
LEAD TO LEAD IMPEDENCE (V1-V6)
LEAD TEST IMPEDENCE
500,1000 Ohms
1000 Ohms
< 1000 Ohms
42
Specifications
Physical
DISPLAY LCD Graphical 128 X 64 Pixels,
Backlit
8.625 x 5.5 x 1.625 Inches
ENCLOSURE
(216 x 140 x 41 mm)
ABS Plastic
WEIGHT
FACE PLATE Lexan, Back printed OPERATING RANGE 15 to 40 C STORAGE RANGE -20 to 65 C
< 2 Lbs
(< 0.91 Kg)
Electrical
Battery, 9 VDC (2 required)
POWER
BATTERY ELIMINATOR (Optional)
BATTERY ELIMINATOR (Provided with SpO2 Module MSP-2100)
BE2000PU (120 VAC) – US
BE2000PE (220 VAC) – Euro
BE2100PU (120 VAC) – US
BE2100PE (220 VAC) – Euro
(NEDA 1604)
Alkaline
9V, 200 mA DC
12V, 500 mA DC
43
PS-2110
This page intentionally left blank.
44
NOTES