BC Biomedical PS-2110 User Manual

MULTI-PARAMETER
PATIENT SIMULATOR
PS-2110
USER MANUAL
BC BIOMEDICAL
PS-2110
TABLE OF CONTENTS
WARNINGS ................................................................................................................. 2
DESCRIPTION............................................................................................................. 3
OVERVIEW.................................................................................................................. 7
ECG--NORMAL SINUS RHYTHM................................................................................ 15
ECG--ARRHYTHMIAS ................................................................................................. 17
ECG--PERFORMACE.................................................................................................. 21
BLOOD PRESSURE .................................................................................................... 23
RESPIRATION............................................................................................................. 25
TEMPERATURE .......................................................................................................... 27
SETUP ......................................................................................................................... 29
POWER UP SETTINGS............................................................................................... 33
SpO2 (OPTION)............................................................................................................ 35
OUTPUT CONNECTIONS ........................................................................................... 37
MANUAL REVISIONS.................................................................................................. 39
WARRANTY................................................................................................................. 39
SPECIFICATIONS ....................................................................................................... 41
NOTES......................................................................................................................... 45
All connections to patients must be removed
before connecting the Device Under Test (DUT)
to the Simulator. A serious hazard may occur if
the patient is connected when testing
Do not connect any leads from the patient
directly to the Simulator or DUT.
iii
BC GROUP INTERNATIONAL, INC. RESERVES THE RIGHT TO MAKE
CHANGES TO ITS PRODUCTS OR SPECIFICATIONS AT ANY TIME,
WITHOUT NOTICE, IN ORDER TO IMPROVE THE DESIGN OR
PERFORMANCE AND TO SUPPLY THE BEST POSSIBLE PRODUCT.
THE INFORMATION IN THIS MANUAL HAS BEEN CAREFULLY CHECKED AND IS BELIEVED TO BE ACCURATE. HOWEVER, NO RESPONSIBILITY IS
ASSUMED FOR INACCURACIES.
WARNING
with the Simulator.
N O T I C E
iii
Manual PS-2110 Copyright © 2004
www.bcgroupintl.com Made in the USA
12/04 Rev 04
BC BIOMEDICAL
PS-2110
PATIENT SIMULATOR
The Model PS-2110 is a Microprocessor based Patient Simulator. It provides ECG, Blood Pressure, Respiration and Temperature Simulation. There are 12 arrhythmias, a pacemaker rhythm, a Fetal/Maternal rhythm, seven waveforms with constant QRS duration and 12 machine performance testing waveforms.
The PS-2110 makes viewing and selecting the desired waveforms and parameters quick and intuitive, with all operational information being available at one time on a cursor-based graphic display, allowing for easy maneuvering through parameters and scrolling through available options.
The following are highlights of some of the main features:
GENERAL
SIMPLE TO OPERATE
NO CODES TO REMEMBER OR ENTER
GRAPHICS DISPLAY WITH SIMULTANEOUS DETAILED STATUS OF
PARAMETERS AND SCROLLING CONTROL OF OPTIONS
DROP DOWN CHOICE SCREENS LIST ALL OPTIONS FOR PARAMETERS
SPECIAL POWER UP FEATURE ALLOWS THE USER TO CHOOSE TO USE
DEFAULT, LAST OR CUSTOM SETTINGS
AUTO SEQUENCES FOR BPM, STATIC-PRESSURE LEVELS AND PERFORMANCE
10 UNIVERSAL PATIENT LEAD CONNECTORS
MINI-DIN CONNECTOR FOR BP CABLES
9 VOLT BATTERY POWER
% BATTERY LIFE DISPLAY
LOW BATTERY INDICATOR
AVAILABLE BATTERY ELIMINATOR
DISPLAY BACKLIGHT
FLASH PROGRAMMABLE FOR UPGRADES
PS-2110
ECG FUNCTIONS
The unit can produce a wide variety of ECG simulations. The user simply selects the parameters that match the desired output.
RATE: 30,60,80,120,180,240,300 BPM
AMPLITUDE:
0.5,1.0,1.5,2.0 mV (Lead II)
AUTOMATIC MODE
PACEMAKER FUNCTION
A pacemaker waveform may be simulated.
WAVEFORM: ASYNCHRONOUS
PULSE HEIGHT:
1.0 mV
PULSE WIDTH:
1.0 ms
FETAL/MATERNAL
A combination of fetal and maternal ECG waveforms may be simulated.
FETAL HEARTRATE: 120 BPM
ARRHYTHMIA FUNCTIONS
The unit can simulate 12 different arrhythmias. Where applicable, both manual and automatic triggering of the waveform is available.
12 DIFFERENT ARRHYTHMIAS
MANUAL AND AUTOMATIC TRIGGERING
Description
ECG-PERFORMANCE FUNCTIONS
The unit will generate Sine, Square and Triangular waveforms with adjustable amplitudes for performance testing. A special Automatic mode is available to auto sequence through the entire range of waveforms.
SINE:
0.1, 0.5, 5, 10, 40, 50, 60,100 Hz
SQUARE:
0.125, 2 Hz
TRIANGLE: 2 Hz
AMPLITUDE:
0.5,1.0,1.5,2.0 mV (Lead II)
AUTOMATIC MODE
RESPIRATION
Respiration is simulated at 8 different rates plus apnea, with the ability to select from 2 Baseline Impedances, the Lead in which it will appear, and the Delta Ohms (amplitude) of the signal.
RATE: 15,20,30,40,60,80,100,120 BrPM
BASELINE IMPEDANCE: 500,1000 OHMS
LEAD: LA or LL
DELTA IMPEDANCE:
0.1,0.2,0.5,1.0,2.0,3.0 OHMS
APNEA: CONTINUOUS (0 BrPM)
PS-2110
BLOOD PRESSURE
Both static and dynamic invasive pressures are simulated. In the static mode the BP output is fixed at the selected level or sequenced through the list using the Automatic mode selection. In the Dynamic mode the selected waveform is synchronized with the ECG and provides a continuous output.
STATIC: 0,20,40,80,100,200,250,300 mmHg
AUTOMATIC STATIC PRESSURE MODE
6 DYNAMIC WAVEFORMS
SENSITIVITY:
5uV/V/mmHg and 40uV/V/mmHg
TEMPERATURE
The unit will simulate three temperatures. This is done by providing the necessary ohmic levels for both the YSI 400 and 700 Series thermistors.
YSI SERIES 400 and 700 SIMULATION
SELECTIONS:
30,37,40 °C (86.0,98.6,104.0 °F)
LEAD TEST FUNCTION
The unit provides a set of test terminals to quick check leads. It will determine if a lead has less than 1000 Ohms resistance.
SPO2 SIMULATION (Option)
When used with the MSP-2100 external module and FingerSim family of SpO2 finger simulators, the system will provide a pulse synchronized SpO2 output for all NSR rates.
RATE: 30,60,80,120,180,240,300 BPM
SpO
OUTPUT:
2
80,90,97 %
OVERVIEW
This section looks at the layout of a PS-2110 and gives descriptions of the elements that
are present.
Lead Continuity
10 Universal Patient Lead Connectors: RA R LA L RL(-) N LL F V1 C1 V2 C2 V3 C3 V4 C4 V5 C5 V6 C6
7 Light Touch Keys for Selecting Parameters and Settings: LEFT and RIGHT Curved Arrows for Moving through Parameters UP and DOWN Arrows for Scrolling through Options ENTER for Selecting Option CHOICES for Displaying Submenu of All Options for a Given Parameter QUIT for Returning to Previous Status
4 Light Touch Keys for Category Selection: Normal Sinus Rhythm Arrhythmias Performance Setup
Test Connections
9V Battery Compartment (Back)
LCD Graphical Display: Shows Parameters for ECG, Respiration, Blood Pressure and Temperature
6-Pin Mini-DIN Plug Connector for Blood Pressure Cable
8-Pin Mini-DIN Plug Connector for Temperature Cable
7-Pin Mini-DIN Plug Connector for Aux Functions
AC Adaptor/ Battery Eliminator
Power Key for Turning Unit On and Off
Back Light Key for turning on and off the backlight
PS-2110
General Operation
The unit is controlled by 13 light touch keys. They allow the user to move around within the
displayed parameters, select the desired options, choose a specific category and control
the setup and power for the unit. When a key is depressed there is an audio click when it is
accepted, or a razz tone if the key is invalid.
A graphics LCD display provides the user with information about the current status of the
ECG, Respiration, Blood Pressure and Temperature settings. The keys
move the block cursor through the displayed information; highlighting the parameter
available for selection. The keys change the options for the highlighted
parameter. The cursor begins flashing if the parameter has been changed. The
key selects the changed option. The key returns to the previous state,
without any changes being made.
To make option selection even easier and to make memorizing and using codes
unnecessary, the key will bring up a screen that displays all the options for the
selected parameter. The and keys can then be used to quickly
scroll through the available options and select the desired setting.
Three category keys allow for quick setting of output waveforms. The , and
keys move the display directly to the selected category. The or
keys can then be used to scroll through and select the desired
settings.
Overview
The key opens a screen that allows the user to select the unit’s general output
settings, as well as setup for the system.
Category Keys
The category, and keys can be used to select ECG waveforms.
The key enters the NSR category and gives access to the seven BPM (Beats Per
Minute) settings with constant QRS duration.
The key enters the arrhythmia category and changes the first line in the display to
the first arrhythmia choice.
The key enters the machine performance testing category and changes the first
line in the display to the first performance waveform choice.
Power Key
The key turns the unit on and off. To turn off the unit, the key must be held for 1
second.
PS-2110
Backlight
The Graphic LCD display may be viewed with or without the backlight. Depressing any key
will activate the backlight. However, since the backlight will drain the battery if left on, it will
automatically shut off after a few seconds when running on battery power. (Note: This time
is selectable in the System Setup screen).
The intensity of the backlight can be adjusted in the System Setup screen to conserve
battery life.
The key is provided to toggle the backlight on or off at any time.
NOTE: The backlight parameter in the System Setup screen may be set to Off, 1-30 sec
Timed or Manual.
ECG Waveforms
The microprocessor has stored in its memory all of the digitalized waveforms. It sends the
individual lead waveforms to D/A converters, which generate accurate analog
representations. The waveforms are then sent through resistor networks, developing the
appropriate signals on the output terminals.
10
Overview
Respiration
Respiration waveforms are provided that have adjustable rates from 15 to 120 BrPM
(Breaths Per Minute) as well as an Apnea (0 BrPM) setting. The signal is generated by
a variation in the impedance in either the LL or LA lead (selectable). The amplitude is
settable from .5 to 3.0 ohms.
Blood Pressure
There is a 6-pin mini-DIN plug connector on the right side of the unit for connection of a
Blood Pressure cable. Full blood pressure simulation is available through this
connector. The circuit is totally isolated.
Temperature
There is an 8-pin mini-DIN plug connector on the right side of the unit for connection of
a Temperature cable. Temperatures are simulated for both YSI 400 and YSI 700 probe
types. There are three different temperatures selectable for each.
11
PS-2110
Universal Patient Lead Connectors
The 10 Universal Patent Lead Connectors allow for 12 lead ECG simulation with
independent outputs. AHA and IEC color-coded labels are located on the face of the unit to
aid in connecting the corresponding U.S. and International Patient Leads.
AHA Label IEC Label Description
RA R Right Arm
LA L Left Arm
RL N
Right Leg (reference or ground)
LL F Left Leg
V Leads (U.S. Canada),
V1 V2 V3 V4 V5 V6
C1 C2 C3 C4 C5 C6
also referred to as pericardial, precordial or unipolar chest
Chest Leads
leads
(International)
High Level Output (+)
A high level ECG output signal (200 x Amplitude Setting) is available in the BP 6-Pin
mini-DIN connector.
Lead Test Terminals
There are two test terminals on the top of the unit that allow for a quick test of the
continuity of the lead cables. Connecting one end of the cable to one terminal and the
other end to the other terminal will test the cable. If the cable is OK (less than 1000
ohms), the LEAD TEST ACTIVE screen will be displayed.
12
Overview
Auto Power Off
The unit may be programmed to automatically turn off after a selected number of minutes
of no key activity to conserve the battery. (Note: This time is selectable in the System
Setup screen).
Power Supply
The unit utilizes two 9 Volt Alkaline Batteries in the rear battery compartment. When the
unit detects a LOW BATTERY condition (5% Battery Life), a warning window will appear
once per minute to alert the user. The key may be used to clear this window and
continue use of the unit. If the battery is not replaced before the battery reaches a critical
level (0 % Battery Life), the unit will shut down. (The percentage of life left in the batteries
can be viewed in the System Setup screen.)
Battery Eliminator
The unit has a 2.1 mm micro jack for connecting a 9-Volt AC battery eliminator (Optional).
The adapter will power the unit, but will not charge the battery.
Power Up Settings
The unit may be setup to turn on using either the factory default settings, the same settings
that it had when last turned off or a custom set of parameters as previously saved by the
user (See Power Up Settings section for details).
13