BC Biomedical PS-2010 User Manual

Page 1
ECG PATIENT SIMULATOR
TEN LEAD
PS-2010
USER MANUAL
Page 2
Page 3
BC BIOMEDICAL
PS-2010
TABLE OF CONTENTS
WARNING.................................................................................................................... 2
DESCRIPTION............................................................................................................. 3
OVERVIEW.................................................................................................................. 4
MANUAL REVISIONS.................................................................................................. 8
WARRANTY................................................................................................................. 8
SPECIFICATIONS ....................................................................................................... 9
NOTES........................................................................................................................ 11
BC GROUP INTERNATIONAL, INC. RESERVES THE RIGHT TO MAKE
CHANGES TO ITS PRODUCTS OR SPECIFICATIONS AT ANY TIME, WITHOUT NOTICE, IN ORDER TO IMPROVE THE DESIGN OR PERFORMANCE AND TO SUPPLY THE BEST POSSIBLE PRODUCT. THE INFORMATION IN THIS MANUAL HAS BEEN CAREFULLY CHECKED AND IS BELIEVED TO BE ACCURATE. HOWEVER, NO RESPONSIBILITY
IS ASSUMED FOR INACCURACIES.
NOTICE
1
Page 4
All connections to patients must be removed
before connecting the Device Under Test (DUT)
to the Simulator. A serious hazard may occur if
the patient is connected when testing
Do not connect any leads from the patient
directly to the Simulator or DUT.
CONTACT INFORMATION
BC GROUP INTERNATIONAL, INC.
WARNING
with the Simulator.
BC BIOMEDICAL
PO BOX 25125
9415 GENTRY AVE
ST. LOUIS, MO 63125
USA
1-800-242-8428
314-638-3800
www.bcgroupintl.com
sales@bcgroupintl.com
Manual PS-2010 Copyright © 2006
www.bcgroupintl.com Made in the USA
9/06 Rev 02
2
Page 5
BC GROUP
PS-2010
PATIENT SIMULATOR
The Model PS-2010 is a Microprocessor based Patient Simulator. It provides ECG Simulation with four waveforms with constant QRS duration and six machine performance testing waveforms. The following are highlights of some of the main features:
10 UNIVERSAL PATIENT LEAD CONNECTORS
ECG: 30, 60, 120 AND 240 BPM
SINE: 10, 60 AND 100 HZ
SQUARE: 0.125 AND 2.0 HZ
TRIANGLE: 2 HZ
HIGH LEVEL OUTPUT (1 V p-p)
AMPLITUDE ACCURACY: +/- 2%
FREQUENCY ACCURACY: +/- 0.5% OF SETTING
9 VOLT BATTERY POWER
LOW BATTERY INDICATOR
% BATTERY LIFE INDICATOR
OPTIONAL BATTERY ELIMINATOR
LEAD TEST FUNCTION
HIGH IMPACT PLASTIC CASE
LIGHT TOUCH KEYS – NO KNOBS OR SWITCHES
OPTIONAL ACCESSORIES
BC20 - 30103 BC BIOMEDICAL SMALL SOFT SIDED CARRYING CASE
BC20 - 21100 BATTERY ELIMINATOR (US Version) BE20 - 21101 BATTERY ELIMINATOR (Euro Version)
3
Page 6
OVERVIEW
p
)
This section looks at the layout of a PS-2010 and gives descriptions of the elements that
are present.
High Level
Out
ut (+
Lead Test Terminals
Universal Patient Lead Connectors: RA R LA L RL(-) N LL F V1 C1
Universal Patient Lead Connectors: V2 C2 V3 C3 V4 C4 V5 C5 V6 C6
10 Waveforms with LED Indicators for Selection: ECG--30, 60, 120, 240 BPM Sine--10, 60, 100 HZ Square--0.125,
2.000 HZ Triangle--2 HZ
4 Light Touch Keys for Power and Waveform Selection: On Off Up Down
Battery Eliminator Socket
Low Battery Indicator LED
Lead Test Indicator LED
9V Battery Compartment (Back)
The unit is controlled by 4 light touch keys. They allow the user to select waveforms and
control the power for the unit. There is an audio click when a key is depressed.
There are 10 LEDs to provide the user with information about waveforms that are
generated 1 LED for Lead Test confirmation and 1 LED for Low Battery Indication.
4
Page 7
Universal Patient Lead Connectors
The 10 Universal Patent Lead Connectors allow for 12 lead ECG simulation with
independent outputs. AHA and IEC color-coded labels are located on the face of the unit to
aid in connecting the corresponding U.S. and International Patient Leads.
AHA Label IEC Label Description
RA R Right Arm
LA L Left Arm
RL N
LL F Left Leg
V1 V2 V3 V4 V5 V6
High Level Output (+)
C1 C2 C3 C4 C5 C6
Right Leg (reference or ground)
V Leads (V1-V6) (U.S. and Canada)
also referred to as pericardial, precordial or unipolar chest leads
Chest Leads (C1-C6) (International)
An additional Universal Patient Lead Connector is located on the top of the unit for
connecting the high level ECG output signal (1 Volt p-p ). The connection is between High
Level (+) and RL (-).
Waveform Selection
There are two keys and 10 LEDs in the Waveform Selection Control Section. The LEDs
indicate which waveform is generated. The keys sequentially select each
waveform. Internally, the microprocessor has stored in memory the digitalized
waveforms. It sends the selected waveform to a D/A converter that generates an
accurate analog representation. This waveform is then sent through a resistor network,
developing the appropriate signals on the output terminals.
5
Page 8
Lead Test Terminals
There are two test terminals on the top of the unit that allow for a quick test of the
continuity of the lead cables. Connecting one end of the cable to one terminal and the
other end to the other terminal will test the cable. If the cable is OK (less than 1000
ohms), the LEAD TEST LED in the lower left of the face will light.
Power Keys
The keys control the power for the unit.
Auto Power Off
The unit will automatically turn off after 10 minutes of no key activity to conserve the
battery.
To override this feature and keep the unit on continuously, depress the key twice
after the unit is on. This will keep the unit on until it is manually turned off. The “Low Batt”
LED will illuminate for 3 sec to indicate that the Auto Power Off feature has been turned off.
Percent of Battery Life Indicator
The unit provides an indication of the Percent of battery life left on the 9 Volt Alkaline
Battery. An A/D converter monitors the battery voltage. Continuously holding down
the key will change the 10 waveform LEDs into a Percent of Battery Life display,
with each LED representing 10%. The stack will strobe up to the present level and flash.
This sequence will continue while the key is depressed.
6
Page 9
Power Supply
The unit utilizes a 9 Volt Alkaline Battery in the rear battery compartment. When the unit
detects a LOW BATTERY, the LED in the lower right of the face will light, indicating the
need to change the battery.
Power Input – A 2.1 mm jack is provided for the optional 9 VDC Battery Eliminator power
supply that may be used for continuous run applications. It bypasses the internal battery
when plugged in.
Note: The unit is shipped with a red plug in the Power Input that prevents it from turning on
and using up the battery during transport. This plug must be removed before use.
7
Page 10
MANUAL REVISIONS
A
A
Revision #
Rev 01 DT7345 Origination
Rev02 DT7345CA Format and Pictures Updated
WARRANTY
FROM DEFECTS IN MATERIALS AND WORKMANSHIP UNDER THE SERVICE FOR WHICH THEY
RE INTENDED. THIS WARRANTY IS EFFECTIVE FOR TWELVE MONTHS FROM THE DATE OF
SHIPMENT.
EXCLUSIONS: THIS WARRANTY IS IN LIEU OF ANY OTHER WARRANTY EXPRESSED OR IMPLIED, INCLUDING, BUT NOT LIMITED TO ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE.
BC GROUP INTERNATIONAL, INC. IS NOT LIABLE FOR ANY INCIDENTAL OR CONSEQUENTIAL DAMAGES.
NO PERSON OTHER THAN AN OFFICER IS AUTHORIZED TO GIVE ANY OTHER WARRANTY OR
SSUME ANY LIABILITY.
REMEDIES: THE PURCHASER'S SOLE AND EXCLUSIVE REMEDY SHALL BE: (1) THE REPAIR OR REPLACEMENT OF DEFECTIVE PARTS OR PRODUCTS, WITHOUT CHARGE. (2) AT THE OPTION OF BC GROUP INTERNATIONAL, INC., THE REFUND OF THE PURCHASE PRICE.
P:\MANUALS\BCGroup\…\PS2010_UM_Rev02.doc
Program # Revisions Made
LIMITED WARRANTY
: BC GROUP INTERNATIONAL, INC. WARRANTS ITS NEW PRODUCTS TO BE FREE
8
Page 11
SPECIFICATIONS
Normal Sinus Rhythm Rate 30,60,120,240 BPM Performance Sine Wave 10,60,100 Hz Square Wave 0.125, 2.000 Hz Triangle Wave 2.000 Hz
SCALES & RANGES
LEAD 1 1.75 mV LEAD 2 2.75 mV LEAD 3 1.00 mV LEAD TO LEAD
IMPEDENCE LEAD TEST
IMPEDENCE
OUTPUT
1000 Ohms
< 1000 Ohms
AMPLITUDE +/- 2 % Lead II FREQUENCY 0.5 %
ACCURACY
9
Page 12
ENCLOSURE
WEIGHT
FACE PLATE Lexan, Back printed OPERATING RANGE 15 to 40 degrees C STORAGE RANGE -20 to 65 degrees C
POWER
BATTERY ELIMINATOR (Optional)
PHYSICAL
6.7 x 3.35 x 1.33 Inches
(170.18 x 85.09 x 33.78 mm)
ABS Plastic
< 1Lbs
(< 0.45 Kg)
ELECTRICAL
Battery, 9 VDC
(NEDA 1604)
Alkaline
BE2000PU (120 VAC) – US
BE2000PE (220 VAC) – Euro
9V, 50 mA DC
10
Page 13
NOTES
11