BC Biomedical PS-2005 User Manual

Page 1
ECG PATIENT SIMULATOR
FIVE LEAD
USER MANUAL
Page 2
Page 3
BC BIOMEDICAL
PS-2005
TABLE OF CONTENTS
WARNINGS ................................................................................................................. 2
DESCRIPTION............................................................................................................. 3
OVERVIEW.................................................................................................................. 4
MANUAL REVISIONS.................................................................................................. 8
WARRANTY................................................................................................................. 8
SPECIFICATIONS ....................................................................................................... 9
NOTES........................................................................................................................ 11
BC GROUP INTERNATIONAL, INC. RESERVES THE RIGHT TO MAKE
CHANGES TO ITS PRODUCTS OR SPECIFICATIONS AT ANY TIME, WITHOUT NOTICE, IN ORDER TO IMPROVE THE DESIGN OR PERFORMANCE AND TO SUPPLY THE BEST POSSIBLE PRODUCT. THE INFORMATION IN THIS MANUAL HAS BEEN CAREFULLY CHECKED AND IS BELIEVED TO BE ACCURATE. HOWEVER, NO RESPONSIBILITY
IS ASSUMED FOR INACCURACIES.
NOTICE
1
Page 4
All connections to patients must be removed
before connecting the Device Under Test (DUT)
to the Simulator. A serious hazard may occur if
the patient is connected when testing
Do not connect any leads from the patient
directly to the Simulator or DUT.
CONTACT INFORMATION
BC GROUP INTERNATIONAL, INC.
WARNING
with the Simulator.
BC BIOMEDICAL
PO BOX 25125
9415 GENTRY AVE
ST. LOUIS, MO 63125
USA
1-800-242-8428
314-638-3800
www.bcgroupintl.com
sales@bcgroupintl.com
Manual PS-2005 Copyright © 2006
www.bcgroupintl.com
09/06 Rev 02
Made in the USA
2
Page 5
BC GROUP
PS-2005
PATIENT SIMULATOR
The Model PS-2005 is a Microprocessor based Patient Simulator. It provides ECG Simulation with four waveforms with constant QRS duration and six machine performance testing waveforms. The following are highlights of some of the main features:
6 PATIENT LEAD SNAP CONNECTORS
ECG: 30, 60, 120 AND 240 BPM
SINE: 10, 60 AND 100 HZ
SQUARE: 0.125 AND 2.0 HZ
TRIANGLE: 2 HZ
AMPLITUDE ACCURACY: +/- 2%
FREQUENCY ACCURACY: +/- 0.5% OF SETTING
9 VOLT BATTERY POWER
LOW BATTERY INDICATOR
% BATTERY LIFE INDICATOR
LEAD TEST FUNCTION
HIGH IMPACT PLASTIC CASE
LIGHT TOUCH KEYS – NO KNOBS OR SWITCHES
3
Page 6
OVERVIEW
This section looks at the layout of a PS-2005 and gives descriptions of the elements that
are present.
Patient Lead Snap Connectors: LA L LL F V1 C1
10 Waveforms with LED Indicators for Selection: ECG--30, 60, 120, 240 BPM Sine--10, 60, 100 HZ Square--0.125,
2.000 HZ Triangle--2 HZ
Low Battery Indicator LED
9V Battery Compartment (Back)
The unit is controlled by 2 light touch keys. They allow the user to select waveforms and
Patient Lead Snap Connectors: RA R RL N
Lead Test Indicator LED
Lead Test Terminals
2 Light Touch Keys for Power and Waveform Selection
control the power for the unit.
There are 10 LEDs to provide the user with information about waveforms that are
generated, 1 LED for Lead Test confirmation and 1 LED for Low Battery Indication.
4
Page 7
Patient Lead Snap Connectors
AHA and IEC color-coded labels are located on the face of the unit to aid in connecting the
corresponding U.S. and International Patient Leads.
AHA Label IEC Label Description
RA R Right Arm
LA L Left Arm
RL N
LL F Left Leg
Right Leg (reference or ground)
V1 V2 V3 V4 V5 V6
C1 C2 C3 C4 C5 C6
V Leads (V1-V6) (U.S. and Canada)
also referred to as pericardial, precordial or unipolar chest leads
Chest Leads (C1-C6) (International)
Waveform Selection
There is one key and 10 LEDs in the Waveform Selection Control Section. The LEDs
indicate which waveform is generated. The key sequentially selects each
waveform. Internally, the microprocessor has stored in memory the digitalized waveforms.
It sends the selected waveform to a D/A converter that generates an accurate analog
representation. This waveform is then sent through a resistor network, developing the
appropriate signals on the output terminals.
5
Page 8
Lead Test Terminals
There are two test terminals on the side of the unit that allow for a quick test of the
continuity of the lead cables. Connecting one end of the cable to one terminal and the
other end to the other terminal will test the cable. If the cable is OK (less than 1000 ohms),
the LEAD TEST LED in the lower right of the face will light.
Power Key
The key turns the unit off and on.
Auto Power Off
The unit will automatically turn off after 10 minutes of no key activity to conserve the
battery.
To override this feature and keep the unit on continuously, press and hold the SELECT key
while turning the unit on. This will keep the unit on until it is manually turned off. The “Low
Batt” LED will illuminate for 3 sec to indicate that the Auto Power Off feature has been
turned off.
Power Supply
The unit utilizes a 9 Volt Alkaline Battery in the rear battery compartment. When the unit
detects a LOW BATTERY, the LED in the lower left of the face will blink, indicating the
need to change the battery.
6
Page 9
Percent of Battery Life Indicator
The unit provides an indication of the Percent of battery life left on the 9 Volt Alkaline
Battery. An A/D converter monitors the battery voltage. Continuously holding down the
SELECT key will change the 10 waveform LEDs into a Percent of Battery Life display, with
each LED representing 10%. The stack will strobe up to the present level and flash. This
sequence will continue while the SELECT key is depressed.
7
Page 10
A
A
MANUAL REVISIONS
Revision #
Rev 01 DT7344 Origination
Rev 02 DT7344CB Format and Pictures Updated
WARRANTY
FROM DEFECTS IN MATERIALS AND WORKMANSHIP UNDER THE SERVICE FOR WHICH THEY
RE INTENDED. THIS WARRANTY IS EFFECTIVE FOR TWELVE MONTHS FROM THE DATE OF
SHIPMENT.
EXCLUSIONS: THIS WARRANTY IS IN LIEU OF ANY OTHER WARRANTY EXPRESSED OR IMPLIED, INCLUDING, BUT NOT LIMITED TO ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE.
BC GROUP INTERNATIONAL, INC. IS NOT LIABLE FOR ANY INCIDENTAL OR CONSEQUENTIAL DAMAGES.
NO PERSON OTHER THAN AN OFFICER IS AUTHORIZED TO GIVE ANY OTHER WARRANTY OR
SSUME ANY LIABILITY.
REMEDIES: THE PURCHASER'S SOLE AND EXCLUSIVE REMEDY SHALL BE: (1) THE REPAIR OR REPLACEMENT OF DEFECTIVE PARTS OR PRODUCTS, WITHOUT CHARGE. (2) AT THE OPTION OF BC GROUP INTERNATIONAL, INC., THE REFUND OF THE PURCHASE PRICE.
P:\MANUALS\BCGroup\…\PS-2000\PS2005_UM_Rev02.doc
Program # Revisions Made
LIMITED WARRANTY
: BC GROUP INTERNATIONAL, INC. WARRANTS ITS NEW PRODUCTS TO BE FREE
8
Page 11
SPECIFICATIONS
Normal Sinus Rhythm Rate 30,60,120,240 BPM Performance Sine Wave 10,60,100 Hz Square Wave 0.125, 2.000 Hz Triangle Wave 2.000 Hz
SCALES & RANGES
LEAD 1 1.75 mV LEAD 2 2.75 mV LEAD 3 1.00 mV LEAD TO LEAD
IMPEDENCE LEAD TEST
IMPEDENCE
OUTPUT
1000 Ohms
< 1000 Ohms
AMPLITUDE +/- 2 % Lead II FREQUENCY 0.5 %
ACCURACY
9
Page 12
ENCLOSURE
PHYSICAL
5.12 x 2.56 x 0.97 Inches (130 x 65 x 25 mm)
ABS Plastic
WEIGHT
FACE PLATE Lexan, Back printed OPERATING RANGE 15 to 40 degrees C STORAGE RANGE -20 to 65 degrees C
> ½ Lb
(> 0.23 Kg)
ELECTRICAL
POWER
Battery, 9 VDC
(NEDA 1604)
Alkaline
10
Page 13
NOTES
11