BC Biomedical NIBP-1030, NIBP-1020, NIBP-1010 User Manual

NON-INVASIVE
BLOOD PRESSURE
SIMULATORS
NIBP-1030
USER MANUAL
NIBP-1010
BC BIOMEDICAL
NIBP-1000 SERIES
TABLE OF CONTENTS
WARNINGS, CAUTIONS, NOTICES ...............................................................................ii
DESCRIPTION............................................................................................................. 1
OVERVIEW.................................................................................................................. 4
KEYS............................................................................................................................ 7
SCREENS.................................................................................................................... 8
SETUP ......................................................................................................................... 13
OPERATIONS.............................................................................................................. 16
CONNECTING PRESSURE CUFFS............................................................... 16
CONNECTING PATIENT LEADS .................................................................... 17
IBP CONNECTOR (NIBP-1030 ONLY)............................................................ 17
TEMPERATURE CONNECTOR (NIBP-1030 ONLY)....................................... 18
AUX CONNECTOR.......................................................................................... 18
THEORY OF OPERATIONS........................................................................... 19
RUNNING A TEST.......................................................................................... 21
MANUAL REVISIONS.................................................................................................. 33
WARRANTY................................................................................................................. 33
SPECIFICATIONS ....................................................................................................... 34
NOTES......................................................................................................................... 40
i
WARNING - USERS
The NIBP-1000 Series Simulators are for use by
skilled technical personnel only.
WARNING - USE
The NIBP-1000 Series Simulators are intended
for testing only and they should never be used in
diagnostics, treatment or any other capacity
where they would come in contact with a patient.
WARNING - MODIFICATIONS
The NIBP-1000 Series Simulators are intended
for use within the published specifications. Any
application beyond these specifications or any
unauthorized user modifications may result in
hazards or improper operation.
WARNING - CONNECTIONS
All connections to patients must be removed
before connecting the Device Under Test (DUT)
to the Simulators. A serious hazard may occur
if the patient is connected when
testing with the Simulators.
Do not connect any leads from the patient
directly to the Simulators or DUT.
WARNING - ADAPTER
Turn Power Off and unplug any AC Power
Adapter before cleaning the surface
of the Simulators.
WARNING - LIQUIDS
Do not submerge or spill liquids on Simulators.
Do not operate the Simulators if internal
components may have been exposed to fluid,
as the internal leakage may have caused
corrosion and be a potential hazard.
ii
CAUTION - SERVICE
The NIBP-1000 Series Simulators are intended to
be serviced only by authorized service
personnel. Troubleshooting and service
procedures should only be performed by
qualified technical personnel.
CAUTION - ENVIRONMENT
The NIBP-1000 Series Simulators are intended to
function between 15 and 40 °C.
Exposure to temperatures outside this range can
adversely affect the performance
of the Simulators.
CAUTION - CLEANING
Do not immerse. The Simulators should be
cleaned by wiping gently with a damp,
lint-free cloth.
CAUTION - INSPECTION
The NIBP-1000 Series Simulators should be
inspected before each use for wear and
the Simulator should be serviced
if any parts are in question.
iii
NOTICE – CE
The NIBP-1000 Simulators bear the mark
Based on the following testing standards:
ELECTROMAGNETIC COMPATIBILITY DIRECTIVE
EMC – Directive 89/336/EEC and 2004/108/EC as amended by
92/31/EEC, 93/68/EEC and Directive 91/263/EEC [ TTE/SES ]
EN 61326-1:1997 + A1:1998 + A2:2001 + A3:2003
“Electrical equipment for measurement, control and
laboratory use – EMC requirements”
This equipment has been type tested by an independent, accredited testing laboratory
and compliance was demonstrated to the above standard to the extent applicable.
EMISSIONS
Radiated and Line Conducted Emissions
EN 61000-3-2 Harmonic Current Emissions EN 61000-3-3 Voltage Fluctuation and Flicker
IMMUNITY– CLASS C
EN 61000-4-2 Electrostatic Discharge EN 61000-4-3 Radiated Electric Field Immunity EN 61000-4-4 Electrical Fast Transients / Bursts EN 61000-4-5 Surge Voltage EN 61000-4-6 Conducted Disturbance EN 61000-4-11 Voltage Dips and Short Interrupts
LOW VOLTAGE DIRECTIVE
EC – Directive 73/23/EC
EN 61010-1:2001
“Safety requirements for electrical equipment for measurement, control, and
laboratory use – General requirements”
This equipment has been type tested and compliance was demonstrated
to the above standard to the extent applicable.
iv
NOTICE – SYMBOLS
Caution
(Consult Manual for Further Information)
Center Negative
Direct Current
Per European Council Directive 2002/95/EC, do not dispose of this product as unsorted municipal waste.
Symbol Description
NOTICE – ABBREVIATIONS
AC Alternating Current
BP Blood Pressure
Bpm Beats per minute
brpm breaths per minute
C Celsius
° degree
DUT Device Under Test
Dias Diastolic
ECG Electrocardiogram
Euro European
HR Heart Rate Hz hertz
k kilo- (10 Kg Kilogram
µ micro- (10
µA microampere
m milli- (10
mA milliampere
mm millimeter
mmHg millimeter mercury
mV millivolt
min minutes
Neo Neonatal
NSR Normal Sinus Rhythm
ohm
lbs pounds
R Respiration
Sec seconds
Syst Systolic
US United States
VDC Direct Current Voltage
3
)
-3
-6
)
)
v
BC GROUP INTERNATIONAL, INC. WILL NOT BE RESPONSIBLE FOR ANY INJURIES SUSTAINED DUE TO UNAUTHORIZED EQUIPMENT MODIFICATIONS OR APPLICATION OF EQUIPMENT OUTSIDE OF THE PUBLISHED INTENDED USE AND SPECIFICATIONS.
BC GROUP INTERNATIONAL, INC. RESERVES THE RIGHT TO MAKE CHANGES TO ITS PRODUCTS OR SPECIFICATIONS AT ANY TIME, WITHOUT NOTICE, IN ORDER TO IMPROVE THE DESIGN OR PERFORMANCE AND TO SUPPLY THE BEST POSSIBLE PRODUCT. THE INFORMATION IN THIS MANUAL HAS BEEN CAREFULLY CHECKED AND IS BELIEVED TO BE ACCURATE. HOWEVER, NO RESPONSIBILITY IS ASSUMED FOR INACCURACIES.
NOTICE – CONTACT INFORMATION
NOTICE – DISCLAIMER
NOTICE – DISCLAIMER
BC BIOMEDICAL
BC GROUP INTERNATIONAL, INC.
PO BOX 25125
9415 GENTRY AVE
ST. LOUIS, MO 63125
USA
1-800-242-8428
314-638-3800
www.bcgroupintl.com
sales@bcgroupintl.com
User Manual NIBP-1000 Series Copyright © 2008
www.bcgroupintl.com
1/08 Rev 10
Made in the USA
vi
BC GROUP
NIBP-1000 SERIES
NON-INVASIVE BLOOD PRESSURE SIMULATORS
The Model NIBP-1000 Series is a family of Microprocessor based, High Precision Non­Invasive Blood Pressure (NIBP) Simulators. The units are small, easy to use and have multiple features to fit many different applications. The NIBP-1020, offers ECG waveforms that are full QRS and respiration waveforms that look real. The NIBP-1030 offers Invasive Blood Pressure, Temperature, Arrhythmias and a Leak Rate test mode.
The graphics display not only provides multiple screens that give the pressure digitally in mmHg, but also offers views of the plot of the overall pressure or a close-up of the BP waveform.
The following are highlights of some of the main features:
NIBP-1010 (BASIC FEATURES, NIBP ONLY):
LARGE GRAPHICS DISPLAY WITH CURSOR SELECTION OF OPTIONS AND SETUP OF PARAMETERS
FULL RANGE MANOMETER
ADULT, NEONATAL HYPERTENSIVE AND HYPOTENSIVE MODES
0-500 mmHg PRESSURE RANGE
+/- 1% OF READING PRESSURE ACCURACY
DIGITAL PRESSURE ENVELOPE OFFSET
OPTIONAL PEAK PRESSURE DETECT WITH SIMPLE RESET
SpO2 READY – COMPATIBLE WITH MSP-2100 MODULE
DIGITAL CALIBRATION – NO POTS TO TURN
SELECTABLE DISPLAY OPTIONS AND DIGIT SIZES
SOFTWARE ADJUSTABLE CONTRAST
FLASH PROGRAMMABLE
BATTERY LIFE DISPLAY (0 TO 100%)
OPTIONAL RECHARGEABLE NiMH BATTERIES
BATTERY ELIMINATOR
RS232 INTERFACE
NIBP-1020 (NIBP, ECG PACE & RESP)
HAS ALL THE BASIC MODEL FEATURES PLUS:
ECG OUTPUT WITH FULL NSR WAVEFORM
SINUSOIDAL RESPIRATION SIMULATION
ECG TEST WAVEFORMS
PACE WAVEFORM
OPTIONAL PEAK PRESSURE DETECT WITH ECG ALARM TEST
ECG SYNCRONIZED WITH BLOOD PRESSURE
1
NIBP-1030 (NIBP, IBP, TEMPERATURE, ECG PACE & RESP)
HAS ALL THE 1020 MODEL FEATURES PLUS:
BACKLIT LCD DISPLAY
SYNCHRONIZED INVASIVE BLOOD PRESSURE OUTPUT
SELECTABLE IBP SENSITIVITY 5/40 µV/V/mmHg
STATIC IBP SIMULATION –10mmHg TO 400 mmHg
LEAK RATE TEST
ECG ARRHYTHMIA WAVEFORMS
ECG ARRHYTHMIA SEQUENCE
YSI 400 AND 700 SERIES TEMPERATURE SIMULATION
AVAILABLE MODELS:
NIBP1010 BASIC UNIT NIBP1010-P + PEAK NIBP1010-BP + BATTERY & PEAK NIBP1010-KIT NIBP1010-BP, 5 ADAPTERS & SOFT CARRYING CASE NIBP1020 BASIC UNIT WITH ECG NIBP1020-PA + PEAK & ALARM NIBP1020-BPA + BATTERY, PEAK & ALARM NIBP1020-KIT NIBP1020-BPA, 5 ADAPTERS & SOFT CARRYING CASE NIBP1030 BASIC UNIT WITH ECG, IBP, TEMP, PEAK & ALARM + MANY MORE NIBP1030-B + BATTERY NIBP1030-KIT NIBP1030-B, 5 ADAPTERS & SOFT CARRYING CASE
ACCESSORIES:
BC20-40614 AC POWER ADAPTER (EURO)
BC20-40613 AC POWER ADAPTER (US)
OPTIONAL ACCESSORIES:
BC20-30111 SOFT CARRYING CASE
BC20-40714 DINAMAP/CRITIKON ADAPTER
BC20-40715 QUICK DISCONNECT ADAPTER
BC20-40716 LUER ADAPTER
BC20-40717 MARQUETTE ADAPTER
BC20-40608 BULB ADAPTER
BC20-40602 NIBP FITTING KIT (HP CUFF ADAPTER,
INFLATION BULB, CPC CONNECTOR, QUICK COUPLER WITH 1/8” MNPT AND SILICONE TUBING AND TEE)
BC20-40605 NIBP ADAPTER KIT (11 CONNECTORS)
2
INVASIVE BLOOD PRESSURE CABLES FOR A COMPLETE LIST CONSULT THE BC GROUP WEBSITE AT WWW.BCGROUPINTL.COM
TEMPERATURE CABLES BC20 - 41333 UT-1 YSI-400 BC20 - 41334 UT-2 YSI-700
3
OVERVIEW
This section looks at the layout of the NIBP-1030 and gives descriptions of the elements that are present.
10 Snap Patient Lead Connectors: RA R LA L RL N LL F V1 C1 V2 C2 V3 C3 V4 C4 V5 C5 V6 C6
LCD Graphical Display: Shows Test Data and Configuration Data
Male Quick Connect for
Pressure
Input
6 Light Touch Keys for Selecting Parameters and Settings: POWER for Turning Unit On and Off UP and DOWN Arrows for Scrolling Through Selected Options. (When No Options are Selected, for Changing Display Screen.) SELECT for Choosing Setting or Parameter to Change SETUP/RETURN for Entering and Exiting Setup Menu RESET for resetting peak pressure reading.
6 pin Mini-
Din
connector for
IBP
8-Pin Mini-DIN Plug Connector for Temperature
7 pin Mini-
Din
connector for
RS-232 and
Power
2.1 mm Jack
4
This section looks at the layout of the NIBP-1020 and gives descriptions of the elements that are present.
10 Snap Patient Lead Connectors: RA R LA L RL N LL F V1 C1 V2 C2 V3 C3 V4 C4 V5 C5 V6 C6
7 pin Mini-
Din
connector for
RS-232 and
Power
2.1 mm Jack
LCD Graphical Display: Shows Test Data and Configuration Data
6 Light Touch Keys for Selecting Parameters and Settings: POWER for Turning Unit On and Off UP and DOWN Arrows for Scrolling Through Selected Options. (When No Options are Selected, for Changing Display Screen.) SELECT for Choosing Setting or Parameter to Change SETUP/RETURN for Entering and Exiting Setup Menu RESET for resetting peak pressure reading.
Male Quick
Connect for
Pressure
Input
5
This section looks at the layout of the NIBP-1010 and gives descriptions of the elements that are present.
Male Quick LCD Graphical Display: Shows Test Data and Configuration Data
6 Light Touch Keys for Selecting Parameters and Settings: POWER for Turning Unit On and Off UP and DOWN Arrows for Scrolling Through Selected Options. (When No Options are Selected, for Changing Display Screen.) SELECT for Choosing Setting or Parameter to Change SETUP/RETURN for Entering and Exiting Setup Menu RESET for resetting peak pressure reading.
Connect for
Pressure
Input
7 pin Mini-
Din
connector for
RS-232 and
Power
2.1 mm Jack
6
KEYS
Six tactile-touch keys are provided for system operation:
– This key turns the unit off and on. The unit will return to the screen that was
active when it was turned off.
– In the DISPLAY MODE, these keys toggle the display through the available
main screens.
In the SELECT MODE, if a parameter has been highlighted, these keys will scroll through
the available settings.
– On the Main screen, this key sequences through the available NIBP or ECG
simulations. On the Setup screen, there are a number of parameters that may be selected
and changed. This key sequences the cursor (Highlight) through those parameters.
– This key is used to RESET the peak pressure reading or to begin a Leak Test.
– This key toggles the unit into and out of the Setup Mode. Depressing this key
will enter the Setup screen where the configuration can be viewed and adjusted.
Depressing the key again will exit the Setup Mode and return to the previously viewed
main screen.
7
SCREENS
MAIN SCREENS
– There are five main screens: Pressure Only, Pressure with Output
Waveform, Pressure with Pressure Graph, ECG (optional) and Battery Indicator (optional).
The available screens can be toggled using .
PRESSURE ONLY SCREEN – This screen has a large pressure display, as shown below.
Also displayed on this screen is the peak pressure and selected output waveform
The display will resemble the following:
PEAK PRESSURE
PRESSURE
(Optional)
Systolic/Diastolic (mean) Mean Heart Rate
Adult 120/80 93 80 35
Adult 120/80 w/Pace 93 Pacemaker 70 35
Adult High 190/120 150 80 35
Adult Low 80/40 53 80 35
Neonatal 70/40 55 94 47
Alarm Test
(Alternates every 30 sec)
Arrhythmia Seq
(NIBP-1030 Only)
Leak Test
Manometer - - -
(Optional)
(NIBP-1030 Only)
OUTPUT SELECTION
Rate (Optional)
-
- 80 35
- - -
47
160
Respiration
00 80
PRESSURE ENVELOPE
OFFSET
This line is only displayed if a non-zero value is programmed into the Systolic &/or Diastolic Shift. (See Setup for more information).
8
OUTPUT WAVEFORM SCREEN – This screen shows the output waveform, as well as, the
pressure, peak pressure and output selection.
NOTE
This waveform is not intended to be
physiologically correct.
The display will resemble the following:
OUTPUT SELECTION
Systolic/Diastolic (mean) Mean Heart Rate
Adult 120/80 93 80 35 Adult 120/80 w/Pace 93 Pacemaker 70 35 High (Adult) 190/120 150 80 35
Low (Adult) 80/40 53 80 35
Neo (Neonatal) 70/40 55 94 47
Alarm Test
(Alternates every 30 sec)
Arrhythmia Seq
(NIBP-1030 Only)
Leak Test
Manometer - - -
(Optional)
(NIBP-1030 Only)
-
- 80 35
- - -
47
160
Rate (Optional)
PRESSURE
PEAK PRESSURE
(Optional)
OUTPUT
WAVEFORM
(as generated by
the NIBP-1000
Dependant on
pressure
Respiration
00 80
9
(Op
)
PRESSURE GRAPH SCREEN – This screen provides a graph of the pressure, as well as,
the pressure, peak pressure and output selection
OUTPUT SELECTION
Systolic/Diastolic (mean) Mean Heart Rate
Adult 120/80 93 80 35 Adult 120/80 w/Pace 93 Pacemaker 70 35 High (Adult) 190/120 150 80 35
Low (Adult) 80/40 53 80 35
Neo (Neonatal) 70/40 55 94 47
Alarm Test
(Alternates every 30 sec)
Arrhythmia Seq
(NIBP-1030 Only)
Leak Test
Manometer - - -
(Optional)
(NIBP-1030 Only)
-
- 80 35
- - -
47
160
PRESSURE
PEAK PRESSURE
tional
PRESSURE
GRAPH
2mmHg per
pixel
Respiration
Rate (Optional)
00 80
10
A
ECG OUTPUT SCREEN – This screen shows the selected ECG output mode.
(NIBP-1020 & NIBP-1030 units only)
NOTE: While in this mode, the NIBP simulation does not run.
NOTE: NSR ECG output is active during NIBP testing at the rate stipulated for the selected
test.
The display will resemble the following:
SELECTED
OUTPUT
LIST OF AVAILABLE
OUTPUTS
NSR 30 Bpm NSR 60 Bpm NSR 120 Bpm NSR 240 Bpm Sine Wave 10 Hz Sine Wave 60 Hz Sine Wave 100 Hz Square Wave 0.125 Hz Square Wave 2 Hz Triangle Wave 2 Hz PVC Arrhythmia* Pair of PVC’s* Run of 5 PVC’s* Ventricular Tachycardia* Ventricular Fibrillation*
systole*
*NIBP-1030 Only
11
BATTERY INDICATOR SCREEN – This screen shows the status of the battery voltage
level with Built-In Battery Option only.
NOTE: It is only an estimate of the battery life remaining.
When the level reaches 10%, the BP simulation mode will be disabled; however, the
Manometer and ECG (optional) will continue to function. Once the battery level reaches
0%, the unit will automatically turn itself off to avoid damaging the batteries.
The display will resemble the following:
Batteries charge from provided AC power adapter. The charge time is about 4 hours from
full discharge. While the batteries are charging, the display will flash “charging.”
USE ONLY THE AC POWER ADAPTER PROVIDED
WITH THE NIBP-1010, NIBP-1020 AND NIBP-1030.
CAUTION
12
SETUP
The Setup Mode allows the user to adjust the configuration of the simulator. The Setup
screen can be entered using the key. The parameters can be changed by using
key to highlight the line and to toggle the available options.
The Setup screen can be exited using the key.
NIBP 1010 and NIBP 1020
NIBP 1030
13
The following is a breakdown of the parameters available in the configuration of the unit
and their available options:
System Setup Configuration
Parameter Description Range
Systolic Shift
Diastolic Shift
Static BP
(NIBP-1030 Only)
IBP Sens
(NIBP-1030 Only)
Temp
SpO2 Output
Auto Off Timer (Min)
Contrast Adjust
Backlight (Sec)
(NIBP-1030 Only)
Battery Life
Software
Adjusts the Systolic Output of the NIBP Simulation. This is not a direct mmHg adjustment.
Adjusts the Diastolic Output of the NIBP Simulation. This is not a direct mmHg adjustment.
Adjusts the Static Blood Pressure Output.
Selects the Invasive Blood Pressure Output sensitivity.
Selects the simulated Temperature output.
This parameter determines whether the SpO2 output pulse is active. The output drives an MSP­2100 FingerSim Module.
Determines the period of inactivity before the unit is turned OFF. A timer is started when the unit is turned ON and is reset each time a key is pressed. When the timer reaches the value set in this parameter, the power is automatically turned OFF. (NOTE: Setting this parameter to 0 disables the Auto Off timer. When running from line power, the unit does not automatically shut off. Auto Off timer is inactive during a test.)
Sets the contrast of the display screen.
Off – Always off 1-30 sec – The elapsed time after which the backlight will automatically turn off. ON – Always ON. The Default setting is 30 seconds.
Available only with Battery Option installed. Displays current life of the battery. At 10%, a warning screen will appear. At 0%, the unit will power down automatically.
Displays current software program.
±50.0
±50.0
-10 to 400 mmHg
5 µV/V/mmHg
40 µV/V/mmHg
0, 24, 30, 35, 37, 40,
42 C
Off/On
0-30 Minutes
0-20
Off, 1-30 sec, ON
0-100%
(Read Only)
(Read Only)
14
SYSTOLIC AND DIASTOLIC SHIFT – The NIBP-1000 Series is equipped with the option
to shift test results to compensate for different methods of measuring Oscillometric NIBP by
various manufacturers and models of devices under test.
There are no absolute standards for Oscillometric NIBP readings; therefore, for a number
of reasons (including patents, technology, etc.), each manufacturer has established a
different method for evaluating the oscillometric pulses. Due to these varying methods,
precisely the same waveforms will give different results on different manufacturer’s units.
CAUTION
These adjustments must be used with caution as they will allow the user to adjust the output results to invalid values.
These adjustments should only be used to aid in the
simplification of testing and with documented controls.
The normal technique used is to run the monitor against a fixed source like the NIBP-1000,
with the understanding that each manufacturer has a predictable error from this norm.
While this is generally the most direct method, users have asked for a method to correct
for this difference, making the monitors read the same as the test unit. The Systolic and
Diastolic Shift settings allow for just such correction.
These adjustments are indicated in a line added to the main display to inform the user of
any shift that has been programmed into the system. This is done so there is no
misunderstanding of the meaning of the results.
15
OPERATIONS
CONNECTING PRESSURE
The NIBP-1010, NIBP-1020 and NIBP-1030 Blood Pressure Simulators are connected
between the blood pressure cuff and the monitor (Device Under Test, DUT).
The blood pressure cuff should be disconnected from the DUT and a ‘T’ adapter inserted
between the cuff and the DUT. The NIBP simulator is then connected to the open side.
Blood
Pressure
Cuff
NIBP-
Roll the cuff tightly on itself or around a mandrel.
10xx
Monitor
(Device Under Test)
(DUT)
16
For the models with ECG output, test snaps are provided along the sides and are identified
by the markings on the overlay.
CONNECTING PATIENT LEADS
IBP CONNECTOR
(NIBP-1030 ONLY)
For the models with IBP output, connect a standard BP test cable (mini-din to monitor
connector) between the NIBP-1030 and the DUT (Device Under Test). The pin-out shown
below for the mini-din is the industry standard configuration that is used on most modern
simulators. As such, there are many existing cables that may be used or you may obtain
the proper cable from BC Group.
17
TEMPERATURE CONNECTOR
(NIBP-1030 ONLY)
For the models with Temperature output, connect a standard Temperature test cable (mini-
din to monitor connector) between the NIBP-1030 and the DUT (Device Under Test). The
pin-out is shown below.
AUX CONNECTOR
The Aux connector is used for both serial communications and SpO2. The serial
connection is made using the BC20-41337 cable. The MSP-2100 module plugs directly
into the Aux port and provides the SpO2 function.
18
THEORY OF OPERATIONS
NIBP (Non-Invasive Blood Pressure) Function
A digital stepping motor and piston assembly is used to generate the output waveforms. A
differential pressure sensor is used to measure the pressure of the cuff. The pressure
sensor is read by a 16-bit differential Analog to Digital converter.
The cuff pressure is monitored and the motor/piston assembly reacts to provide the correct
response based on the selected mode.
IBP (Invasive Blood Pressure) Function
A 12-bit Digital to Analog converter is used to generate the IBP waveforms. The output
circuit is fully isolated and capable of switching between the two standard sensitivities (5
µV/V/mmHg and 40 µV/V/mmHg).
Temperature
There is an 8-pin mini-DIN plug connector on the right side of the unit for connection of
a Temperature cable. Temperatures are simulated for both YSI 400 and YSI 700 probe
types. There are seven different temperatures selectable for each.
19
ECG Function
The ECG output is performed by a 12-bit Digital to Analog converter. All waveforms for the
standard ECG and Arrhythmias are generated via data stored in the program.
Respiration Function
A 12-bit Digital to Analog converter is used to generate the respiration waveform. The
respiration output is present on the LA ECG lead only. Refer to Monitor manual to
determine which lead the Monitor uses for respiration. It may be necessary to reverse the
LA and LL leads for respiration to be detected.
20
RUNNING A TEST
Depending on the model, there are up to 9 Basic Test Modes and 3 Sub-Test Modes. This
section will walk through each of the tests and their basic operation.
The Main Tests are accessible with a single key. The key will scroll through the
following tests in a continuous loop:
Adult 120/80 Adult 120/80 w/Pace Adult High 190/120 Adult Low 80/40 Neonatal 70/40 Alarm Test Arrhythmia Seq Leak Test Manometer
BASIC TEST MODES
NIBP:
The first 5 test modes deal with various NIBP setups. To run an NIBP simulation, the cuff
and monitor are connected to the pressure input. Then the measurement is initiated by the
monitor and the NIBP-1000 unit will output the proper waveform based on the cuff pressure
provided by the monitor and the selected simulation.
The NIBP output mode can be changed by pressing the key. Once the desired
operating mode is selected, the output will automatically begin when the correct pressure is
detected.
21
(Op
)
There are 5 selectable NIBP Basic Test Modes; Adult, Adult w/pace, Adult High, Adult Low
and Neonatal. The displays will resemble the following examples:
NOTE: The screens for the first test will have the individual components labeled. The
component labels for subsequent tests are the same.
Adult 120/80
(Typical Screens)
PRESSURE
PEAK PRESSURE
(Optional)
PRESSURE ENVELOPE
NIBP TEST
Type Adult 120/80 Heart Rate 80 Respiration Rate 35 (Optional)
This line is only displayed if a non-zero value is programmed into the Systolic &/or Diastolic Shift. (See Setup for more information).
OFFSET
(Pressure Only Screen)
PRESSURE
NIBP TEST
Type Adult 120/80 Heart Rate 80 Respiration Rate 35 (Optional)
PRESSURE
NIBP TEST
Type Adult 120/80 Heart Rate 80 Respiration Rate 35 (Optional)
PEAK PRESSURE
(Optional)
OUTPUT
WAVEFORM
(as generated by
the NIBP-1000
Dependant on
pressure
(Output Waveform Screen)
PEAK PRESSURE
tional
PRESSURE
GRAPH
2mmHg per
pixel
(Pressure Graph Screen)
22
Adult 120/80 with Pace
(Typical Screens)
Adult High 120/80
(Typical Screens)
23
Adult Low 80/40
(Typical Screens)
Neonatal 70/40
(Typical Screens)
24
,
)
ALARM TEST:
Use the key to cycle through the Test Modes until Alarm Test appears. The
purpose of this function is to help test the alarms of the monitor under test. The ECG
output will alternate from a 47 BPM NSR with apnea respiration (0 brpm) to 160 BPM NSR
with 80 brpm respiration. The time interval between alternations is 30 seconds. A count
down timer is displayed to indicate the time still remaining for the current output.
NOTE: While in this mode, the NIBP simulation output is disabled.
The displays will resemble the following examples: (Typical Screens)
PRESSURE
(Information not relevant to this
test, NIBP is disabled.
PEAK PRESSURE (Information not relevant to this test
TIMER
NIBP is disabled.)
PRESSURE ENVELOPE
ALARM TEST
Heart Rate 47/160 Respiration Rate 0/80
This line is only displayed if a non-zero value is programmed into the Systolic &/or Diastolic Shift. (See Setup for more information).
OFFSET
(Pressure Only Screen)
(Output Waveform Screen) (Pressure Graph Screen)
25
,
y
)
ARRHYTHMIA SEQUENCE: (NIBP-1030 Only)
Use the key to cycle through the Test Modes until Arrhythmia Sequence appears.
The purpose of this function is to provide a timed sequence of some of the more common
Arrhythmias. The test will continually cycle through the 6 arrhythmias in the sequence listed
below. A count down timer is displayed to indicate the time still remaining in each step.
NOTE: While in this mode, the NIBP simulation output is disabled.
Step Display Total Time ECG Output
1 Single PVCs 18 sec 11 NSR, 1 PVC, 11 NSR, 1 PVC 2 Pair of PVCs 9 sec 10 NSR, Pair of PVCs 3 Single PVCs 18 sec 11 NSR, 1 PVC, 11 NSR, 1 PVC 4 Run of 5 PVCs 12 sec 11 NSR, Run of 5 PVCs 5 Vtach 30 sec 15 NSR, VTach 6 Vfib 30 sec 15 NSR, VFib 7 Asystole 30 sec 15 NSR, Asystole
The displays will resemble the following examples: (Typical Screens)
PRESSURE
(Information not relevant to this
test, NIBP is disabled.
PEAK PRESSURE (Information not relevant to this test
TIMER
NIBP is disabled.)
ARRHYTHMIA
Single PVCs
Pair of PVCs
Single PVCs
Run of 5 PVCs
Vtach
Vfib
As
stole
(Pressure Only Screen)
PRESSURE ENVELOPE
OFFSET
This line is only displayed if a non-zero value is programmed into the Systolic &/or Diastolic Shift. (See Setup for more information).
(Output Waveform Screen) (Pressure Graph Screen)
26
LEAK TEST: (NIBP-1030 Only)
Use the key to cycle through the Test Modes until Leak Test appears. The
purpose of this function is to provide a standard pressure leak test. The pressure input is
connected to the system that is to be monitored. Pressure is then applied to the system.
The Leak Test is initiated by pressing the key. The unit traces and displays the
pressure drop and the time since the test was initiated. It also, calculates and displays the
leak rate in mmHg per minute.
NOTE: While in this mode, the BP simulation outputs are disabled.
The displays will resemble the following examples: (Typical Screens)
PRESSURE
LEAK TEST
Drop in mmHg Time in minutes:seconds Leak Rate in mmHg/minute
(Pressure Only Screen)
PEAK PRESSURE
(Optional)
(Output Waveform Screen) (Pressure Graph Screen)
27
MANOMETER:
Use the key to cycle through the Test Modes until Manometer appears. The
purpose of this function is to allow the unit to be used strictly as a Manometer with a full
scale range of +/- 500 mmHg.
NOTE: The Graphic Screens are also available in this mode for maximum flexibility
The displays will resemble the following examples: (Typical Screens)
PRESSURE
PEAK PRESSURE
(Optional)
PRESSURE ENVELOPE
This line is only displayed if a non-zero value is programmed into the Systolic &/or Diastolic Shift. (See Setup for more information).
OFFSET
(Pressure Only Screen)
(Output Waveform Screen) (Pressure Graph Screen)
28
SUB TEST MODES
PEAK: (Optional)
The optional Peak test feature is available whenever the display is in a pressure mode. It
will continuously monitor the pressure input and display and hold the maximum value. This
value may be reset at any time by pressing the key.
The displays will resemble the following examples: (Typical Screens)
29
A
ECG PERFORMANCE TEST: (NIBP-1020 & NIBP-1030)
The ECG output is selected by pressing until the ECG Waveforms menu is
displayed. The key is used to choose the desired ECG output. There are 16 fixed
waveforms available.
NOTE: While in this mode, the NIBP simulation is disabled, but the IBP simulation is active
(NIBP-1030 only).
The display will resemble the following example:
LIST OF AVAILABLE
SELECTED
OUTPUT
NSR 30 Bpm NSR 60 Bpm NSR 120 Bpm NSR 240 Bpm Sine Wave 10 Hz Sine Wave 60 Hz Sine Wave 100 Hz Square Wave 0.125 Hz Square Wave 2 Hz Triangle Wave 2 Hz PVC Arrhythmia* Pair of PVC’s* Run of 5 PVC’s* Ventricular Tachycardia* Ventricular Fibrillation*
systole*
OUTPUTS
*NIBP-1030 Only
30
DYNAMIC IBP: (NIBP-1030 Only)
The IBP output matches the NIBP setting in the NIBP mode or the ECG setting in the ECG
Waveform mode.
NOTE: The IBP sensitivity (5 µV/V/mmHg or 40 µV/V/mmHg) must be correctly selected in
the SETUP Mode before using IBP. (See Setup section for more information.)
STATIC IBP: (NIBP-1030 Only)
A static Blood Pressure output may be selected in the SETUP mode. This fixed value will
remain on the IBP output until changed.
NOTE: The IBP sensitivity (5 µV/V/mmHg or 40 µV/V/mmHg) must be correctly selected in
the SETUP Mode before using IBP. (See Setup section for more information.)
31
TEMPERATURE SIMULATION: (NIBP-1030 Only)
The temperature setting can be selected in the SETUP mode. This fixed value will remain
on the Temperature output until changed. The output will simulate both YSI 400 and YSI
700 Temperature probes.
NOTE: Both outputs are available at the output connector simultaneously.
32
A
A
MANUAL REVISIONS
Revision # Program # Revisions Made
Rev 01 DT7355CA Origination Rev 02 DT7355CA Specification Information Updated Rev 03 DT7355CA Miscellaneous Edits Rev 04 DT7355CF Add Systolic/Diastolic adjustments Rev 05 DT7355CG Output Selection edited Rev 06 DT7355CI Miscellaneous Edits Rev 07 DT7355CI Warnings, Cautions, Notices Updated Rev 08 DT7359CA NIBP-1030, IBP, ECG Arrhythmia Waveforms Added Rev 09 DT7359CA Temperature Added Rev 10 DT7359CA Temp Connector Updated, Static 120 Added
LIMITED WARRANTY
WARRANTY
FROM DEFECTS IN MATERIALS AND WORKMANSHIP UNDER THE SERVICE FOR WHICH THEY
RE INTENDED. THIS WARRANTY IS EFFECTIVE FOR TWELVE MONTHS FROM THE DATE OF
SHIPMENT.
EXCLUSIONS: THIS WARRANTY IS IN LIEU OF ANY OTHER WARRANTY EXPRESSED OR IMPLIED, INCLUDING, BUT NOT LIMITED TO ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE.
BC GROUP INTERNATIONAL, INC. IS NOT LIABLE FOR ANY INCIDENTAL OR CONSEQUENTIAL DAMAGES.
NO PERSON OTHER THAN AN OFFICER IS AUTHORIZED TO GIVE ANY OTHER WARRANTY OR
SSUME ANY LIABILITY.
REMEDIES: THE PURCHASER'S SOLE AND EXCLUSIVE REMEDY SHALL BE: (1) THE REPAIR OR REPLACEMENT OF DEFECTIVE PARTS OR PRODUCTS, WITHOUT CHARGE. (2) AT THE OPTION OF BC GROUP INTERNATIONAL, INC., THE REFUND OF THE PURCHASE PRICE.
: BC GROUP INTERNATIONAL, INC. WARRANTS ITS NEW PRODUCTS TO BE FREE
33
SPECIFICATIONS
NIBP-1010
RANGE 0-500 mmHg @ 20°C
ACCURACY +/- (1% of Reading + .5 mmHg)
RATE 80, 94 bpm
ACCURACY +/- 1%
DISPLAYS LCD Graphical 128 X 64 Pixels
ENCLOSURE
WEIGHT
FACE PLATE Lexan, Back printed OPERATING RANGE 15 to 40 C STORAGE RANGE -20 to 65 C
BATTERY ELIMINATOR
BATTERY (OPTIONAL)
BATTERY RUNTIME
BATTERY STORAGE LIFE
Blood Pressure
Physical
Electrical
500 Test Cycles between charges
7 x 5 x 4 Inches
(177.8 x 127.0 x 101.6 mm)
ABS Plastic
< 3 Lbs
(< 1.36 Kg)
12VDC, 500mA
6 AA NiMH Rechargeable
(Not user serviceable)
1 Year from full charge
34
NIBP-1020
RANGE 0-500 mmHg @ 20°C
ACCURACY +/- (1% of Reading + .5 mmHg)
RATE 80, 94 bpm (synchronized to ECG)
ACCURACY +/- 1%
RATE 30,60,120,240 BPM
ACCURACY +/- 1%
AMPLITUDE 2.75 mV
ACCURACY +/- 2% @ Lead II
SINE WAVE 10,60,100 Hz SQUARE WAVE 0.125, 2.000 Hz
TRIANGLE WAVE 2.000 Hz
RATE ACCURACY +/- 1% AMPLITUDE 2.75 mV AMPLITUDE
ACCURACY
AMPLITUDE 3 mV ACCURACY +/- 10% WIDTH 6 ms ACCURACY +/- 5%
Blood Pressure
ECG NSR
ECG Performance
+/- 2% @ Lead II
Pacemaker Waveforms
35
RATE ACCURACY +/- 1% IMPEDANCE DELTA 3.0 ohms ACCURACY +/- 10% BASELINE 1000 ohms ACCURACY +/- 5%
DISPLAYS
ENCLOSURE
WEIGHT
FACE PLATE Lexan, Back printed OPERATING RANGE 15 to 40 C STORAGE RANGE -20 to 65 C
AC POWER ADAPTER
BATTERY (OPTIONAL)
BATTERY RUNTIME
BATTERY STORAGE LIFE
Respiration
Physical
LCD Graphical 128 X 64 Pixels
White LED Backlight
7 x 5 x 4 Inches
(177.8 x 127.0 x 101.6 mm)
ABS Plastic
< 3 Lbs
(< 1.36 Kg)
Electrical
12VDC, 500mA
6 AA NiMH Rechargeable
(Not user serviceable)
500 Test Cycles between charges
1 Year from full charge
36
NIBP-1030
RANGE +/- 500 mmHg @ 20°C
ACCURACY +/- (1% of Reading + .5 mmHg)
RATE 80, 94 bpm (synched to ECG)
ACCURACY +/- 1%
STATIC PRESSURE
IMPEDANCE 300 ohms
EXCITATION RANGE 2 to 16 V RMS EXCITATION
FREQUENCY SENSITIVITY 5 or 40 µV/V/mmHg
RATE 30,60,120,240 BPM
ACCURACY +/- 1%
AMPLITUDE 2.75 mV
ACCURACY +/- 2% @ Lead II
ACCURACY
ACCURACY +/- 10%
Blood Pressure
Invasive Blood Pressure
-10,-5,0,20,40,50,60,80,100,120,150, 160,200,240,250,300,320,400 mmHg
+/- (1% Full Range + 1 mmHg) or
+/- (2% Setting + 2 mmHg)
whichever is better
DC to 5 KHz
ECG NSR
37
ECG Performance
SINE WAVE 10,60,100 Hz SQUARE WAVE 0.125, 2.000 Hz
TRIANGLE WAVE 2.000 Hz
RATE ACCURACY +/- 1% AMPLITUDE 2.75 mV AMPLITUDE
ACCURACY
+/- 2% @ Lead II
Pacemaker Waveforms
AMPLITUDE 3 mV ACCURACY +/- 10% WIDTH 6 ms ACCURACY +/- 5%
Respiration
RATE ACCURACY +/- 1% IMPEDANCE DELTA 3.0 ohms ACCURACY +/- 10% BASELINE 1000 ohms ACCURACY +/- 5%
Temperature
Temperature
Accuracy +/- 0.1 C Type YSI Series 400 and 700
0,24,30,37,40 C (32.0,
75.2,86.0,98.6,104.0 F)
38
P:\Manuals\BCGroup\…\NIBP1000_UM_Rev10.doc
DISPLAYS LCD Graphical 128 X 64 Pixels
ENCLOSURE
WEIGHT
FACE PLATE Lexan, Back printed OPERATING RANGE 15 to 40 C STORAGE RANGE -20 to 65 C
AC POWER ADAPTER
BATTERY (OPTIONAL)
BATTERY RUNTIME
BATTERY STORAGE LIFE
Physical
Electrical
500 Test Cycles between charges
7 x 5 x 4 Inches
(177.8 x 127.0 x 101.6 mm)
ABS Plastic
< 3 Lbs
(< 1.36 Kg)
12VDC, 500mA
6 AA NiMH Rechargeable
(Not user serviceable)
1 Year from full charge
39
NOTES
40
Loading...