BC Biomedical NIBP-1030, NIBP-1020, NIBP-1010 User Manual

NON-INVASIVE
BLOOD PRESSURE
SIMULATORS
NIBP-1030
USER MANUAL
NIBP-1010
BC BIOMEDICAL
NIBP-1000 SERIES
TABLE OF CONTENTS
WARNINGS, CAUTIONS, NOTICES ...............................................................................ii
DESCRIPTION............................................................................................................. 1
OVERVIEW.................................................................................................................. 4
KEYS............................................................................................................................ 7
SCREENS.................................................................................................................... 8
SETUP ......................................................................................................................... 13
OPERATIONS.............................................................................................................. 16
CONNECTING PRESSURE CUFFS............................................................... 16
CONNECTING PATIENT LEADS .................................................................... 17
IBP CONNECTOR (NIBP-1030 ONLY)............................................................ 17
TEMPERATURE CONNECTOR (NIBP-1030 ONLY)....................................... 18
AUX CONNECTOR.......................................................................................... 18
THEORY OF OPERATIONS........................................................................... 19
RUNNING A TEST.......................................................................................... 21
MANUAL REVISIONS.................................................................................................. 33
WARRANTY................................................................................................................. 33
SPECIFICATIONS ....................................................................................................... 34
NOTES......................................................................................................................... 40
i
WARNING - USERS
The NIBP-1000 Series Simulators are for use by
skilled technical personnel only.
WARNING - USE
The NIBP-1000 Series Simulators are intended
for testing only and they should never be used in
diagnostics, treatment or any other capacity
where they would come in contact with a patient.
WARNING - MODIFICATIONS
The NIBP-1000 Series Simulators are intended
for use within the published specifications. Any
application beyond these specifications or any
unauthorized user modifications may result in
hazards or improper operation.
WARNING - CONNECTIONS
All connections to patients must be removed
before connecting the Device Under Test (DUT)
to the Simulators. A serious hazard may occur
if the patient is connected when
testing with the Simulators.
Do not connect any leads from the patient
directly to the Simulators or DUT.
WARNING - ADAPTER
Turn Power Off and unplug any AC Power
Adapter before cleaning the surface
of the Simulators.
WARNING - LIQUIDS
Do not submerge or spill liquids on Simulators.
Do not operate the Simulators if internal
components may have been exposed to fluid,
as the internal leakage may have caused
corrosion and be a potential hazard.
ii
CAUTION - SERVICE
The NIBP-1000 Series Simulators are intended to
be serviced only by authorized service
personnel. Troubleshooting and service
procedures should only be performed by
qualified technical personnel.
CAUTION - ENVIRONMENT
The NIBP-1000 Series Simulators are intended to
function between 15 and 40 °C.
Exposure to temperatures outside this range can
adversely affect the performance
of the Simulators.
CAUTION - CLEANING
Do not immerse. The Simulators should be
cleaned by wiping gently with a damp,
lint-free cloth.
CAUTION - INSPECTION
The NIBP-1000 Series Simulators should be
inspected before each use for wear and
the Simulator should be serviced
if any parts are in question.
iii
NOTICE – CE
The NIBP-1000 Simulators bear the mark
Based on the following testing standards:
ELECTROMAGNETIC COMPATIBILITY DIRECTIVE
EMC – Directive 89/336/EEC and 2004/108/EC as amended by
92/31/EEC, 93/68/EEC and Directive 91/263/EEC [ TTE/SES ]
EN 61326-1:1997 + A1:1998 + A2:2001 + A3:2003
“Electrical equipment for measurement, control and
laboratory use – EMC requirements”
This equipment has been type tested by an independent, accredited testing laboratory
and compliance was demonstrated to the above standard to the extent applicable.
EMISSIONS
Radiated and Line Conducted Emissions
EN 61000-3-2 Harmonic Current Emissions EN 61000-3-3 Voltage Fluctuation and Flicker
IMMUNITY– CLASS C
EN 61000-4-2 Electrostatic Discharge EN 61000-4-3 Radiated Electric Field Immunity EN 61000-4-4 Electrical Fast Transients / Bursts EN 61000-4-5 Surge Voltage EN 61000-4-6 Conducted Disturbance EN 61000-4-11 Voltage Dips and Short Interrupts
LOW VOLTAGE DIRECTIVE
EC – Directive 73/23/EC
EN 61010-1:2001
“Safety requirements for electrical equipment for measurement, control, and
laboratory use – General requirements”
This equipment has been type tested and compliance was demonstrated
to the above standard to the extent applicable.
iv
NOTICE – SYMBOLS
Caution
(Consult Manual for Further Information)
Center Negative
Direct Current
Per European Council Directive 2002/95/EC, do not dispose of this product as unsorted municipal waste.
Symbol Description
NOTICE – ABBREVIATIONS
AC Alternating Current
BP Blood Pressure
Bpm Beats per minute
brpm breaths per minute
C Celsius
° degree
DUT Device Under Test
Dias Diastolic
ECG Electrocardiogram
Euro European
HR Heart Rate Hz hertz
k kilo- (10 Kg Kilogram
µ micro- (10
µA microampere
m milli- (10
mA milliampere
mm millimeter
mmHg millimeter mercury
mV millivolt
min minutes
Neo Neonatal
NSR Normal Sinus Rhythm
ohm
lbs pounds
R Respiration
Sec seconds
Syst Systolic
US United States
VDC Direct Current Voltage
3
)
-3
-6
)
)
v
BC GROUP INTERNATIONAL, INC. WILL NOT BE RESPONSIBLE FOR ANY INJURIES SUSTAINED DUE TO UNAUTHORIZED EQUIPMENT MODIFICATIONS OR APPLICATION OF EQUIPMENT OUTSIDE OF THE PUBLISHED INTENDED USE AND SPECIFICATIONS.
BC GROUP INTERNATIONAL, INC. RESERVES THE RIGHT TO MAKE CHANGES TO ITS PRODUCTS OR SPECIFICATIONS AT ANY TIME, WITHOUT NOTICE, IN ORDER TO IMPROVE THE DESIGN OR PERFORMANCE AND TO SUPPLY THE BEST POSSIBLE PRODUCT. THE INFORMATION IN THIS MANUAL HAS BEEN CAREFULLY CHECKED AND IS BELIEVED TO BE ACCURATE. HOWEVER, NO RESPONSIBILITY IS ASSUMED FOR INACCURACIES.
NOTICE – CONTACT INFORMATION
NOTICE – DISCLAIMER
NOTICE – DISCLAIMER
BC BIOMEDICAL
BC GROUP INTERNATIONAL, INC.
PO BOX 25125
9415 GENTRY AVE
ST. LOUIS, MO 63125
USA
1-800-242-8428
314-638-3800
www.bcgroupintl.com
sales@bcgroupintl.com
User Manual NIBP-1000 Series Copyright © 2008
www.bcgroupintl.com
1/08 Rev 10
Made in the USA
vi
BC GROUP
NIBP-1000 SERIES
NON-INVASIVE BLOOD PRESSURE SIMULATORS
The Model NIBP-1000 Series is a family of Microprocessor based, High Precision Non­Invasive Blood Pressure (NIBP) Simulators. The units are small, easy to use and have multiple features to fit many different applications. The NIBP-1020, offers ECG waveforms that are full QRS and respiration waveforms that look real. The NIBP-1030 offers Invasive Blood Pressure, Temperature, Arrhythmias and a Leak Rate test mode.
The graphics display not only provides multiple screens that give the pressure digitally in mmHg, but also offers views of the plot of the overall pressure or a close-up of the BP waveform.
The following are highlights of some of the main features:
NIBP-1010 (BASIC FEATURES, NIBP ONLY):
LARGE GRAPHICS DISPLAY WITH CURSOR SELECTION OF OPTIONS AND SETUP OF PARAMETERS
FULL RANGE MANOMETER
ADULT, NEONATAL HYPERTENSIVE AND HYPOTENSIVE MODES
0-500 mmHg PRESSURE RANGE
+/- 1% OF READING PRESSURE ACCURACY
DIGITAL PRESSURE ENVELOPE OFFSET
OPTIONAL PEAK PRESSURE DETECT WITH SIMPLE RESET
SpO2 READY – COMPATIBLE WITH MSP-2100 MODULE
DIGITAL CALIBRATION – NO POTS TO TURN
SELECTABLE DISPLAY OPTIONS AND DIGIT SIZES
SOFTWARE ADJUSTABLE CONTRAST
FLASH PROGRAMMABLE
BATTERY LIFE DISPLAY (0 TO 100%)
OPTIONAL RECHARGEABLE NiMH BATTERIES
BATTERY ELIMINATOR
RS232 INTERFACE
NIBP-1020 (NIBP, ECG PACE & RESP)
HAS ALL THE BASIC MODEL FEATURES PLUS:
ECG OUTPUT WITH FULL NSR WAVEFORM
SINUSOIDAL RESPIRATION SIMULATION
ECG TEST WAVEFORMS
PACE WAVEFORM
OPTIONAL PEAK PRESSURE DETECT WITH ECG ALARM TEST
ECG SYNCRONIZED WITH BLOOD PRESSURE
1
NIBP-1030 (NIBP, IBP, TEMPERATURE, ECG PACE & RESP)
HAS ALL THE 1020 MODEL FEATURES PLUS:
BACKLIT LCD DISPLAY
SYNCHRONIZED INVASIVE BLOOD PRESSURE OUTPUT
SELECTABLE IBP SENSITIVITY 5/40 µV/V/mmHg
STATIC IBP SIMULATION –10mmHg TO 400 mmHg
LEAK RATE TEST
ECG ARRHYTHMIA WAVEFORMS
ECG ARRHYTHMIA SEQUENCE
YSI 400 AND 700 SERIES TEMPERATURE SIMULATION
AVAILABLE MODELS:
NIBP1010 BASIC UNIT NIBP1010-P + PEAK NIBP1010-BP + BATTERY & PEAK NIBP1010-KIT NIBP1010-BP, 5 ADAPTERS & SOFT CARRYING CASE NIBP1020 BASIC UNIT WITH ECG NIBP1020-PA + PEAK & ALARM NIBP1020-BPA + BATTERY, PEAK & ALARM NIBP1020-KIT NIBP1020-BPA, 5 ADAPTERS & SOFT CARRYING CASE NIBP1030 BASIC UNIT WITH ECG, IBP, TEMP, PEAK & ALARM + MANY MORE NIBP1030-B + BATTERY NIBP1030-KIT NIBP1030-B, 5 ADAPTERS & SOFT CARRYING CASE
ACCESSORIES:
BC20-40614 AC POWER ADAPTER (EURO)
BC20-40613 AC POWER ADAPTER (US)
OPTIONAL ACCESSORIES:
BC20-30111 SOFT CARRYING CASE
BC20-40714 DINAMAP/CRITIKON ADAPTER
BC20-40715 QUICK DISCONNECT ADAPTER
BC20-40716 LUER ADAPTER
BC20-40717 MARQUETTE ADAPTER
BC20-40608 BULB ADAPTER
BC20-40602 NIBP FITTING KIT (HP CUFF ADAPTER,
INFLATION BULB, CPC CONNECTOR, QUICK COUPLER WITH 1/8” MNPT AND SILICONE TUBING AND TEE)
BC20-40605 NIBP ADAPTER KIT (11 CONNECTORS)
2
INVASIVE BLOOD PRESSURE CABLES FOR A COMPLETE LIST CONSULT THE BC GROUP WEBSITE AT WWW.BCGROUPINTL.COM
TEMPERATURE CABLES BC20 - 41333 UT-1 YSI-400 BC20 - 41334 UT-2 YSI-700
3
OVERVIEW
This section looks at the layout of the NIBP-1030 and gives descriptions of the elements that are present.
10 Snap Patient Lead Connectors: RA R LA L RL N LL F V1 C1 V2 C2 V3 C3 V4 C4 V5 C5 V6 C6
LCD Graphical Display: Shows Test Data and Configuration Data
Male Quick Connect for
Pressure
Input
6 Light Touch Keys for Selecting Parameters and Settings: POWER for Turning Unit On and Off UP and DOWN Arrows for Scrolling Through Selected Options. (When No Options are Selected, for Changing Display Screen.) SELECT for Choosing Setting or Parameter to Change SETUP/RETURN for Entering and Exiting Setup Menu RESET for resetting peak pressure reading.
6 pin Mini-
Din
connector for
IBP
8-Pin Mini-DIN Plug Connector for Temperature
7 pin Mini-
Din
connector for
RS-232 and
Power
2.1 mm Jack
4
This section looks at the layout of the NIBP-1020 and gives descriptions of the elements that are present.
10 Snap Patient Lead Connectors: RA R LA L RL N LL F V1 C1 V2 C2 V3 C3 V4 C4 V5 C5 V6 C6
7 pin Mini-
Din
connector for
RS-232 and
Power
2.1 mm Jack
LCD Graphical Display: Shows Test Data and Configuration Data
6 Light Touch Keys for Selecting Parameters and Settings: POWER for Turning Unit On and Off UP and DOWN Arrows for Scrolling Through Selected Options. (When No Options are Selected, for Changing Display Screen.) SELECT for Choosing Setting or Parameter to Change SETUP/RETURN for Entering and Exiting Setup Menu RESET for resetting peak pressure reading.
Male Quick
Connect for
Pressure
Input
5
This section looks at the layout of the NIBP-1010 and gives descriptions of the elements that are present.
Male Quick LCD Graphical Display: Shows Test Data and Configuration Data
6 Light Touch Keys for Selecting Parameters and Settings: POWER for Turning Unit On and Off UP and DOWN Arrows for Scrolling Through Selected Options. (When No Options are Selected, for Changing Display Screen.) SELECT for Choosing Setting or Parameter to Change SETUP/RETURN for Entering and Exiting Setup Menu RESET for resetting peak pressure reading.
Connect for
Pressure
Input
7 pin Mini-
Din
connector for
RS-232 and
Power
2.1 mm Jack
6
KEYS
Six tactile-touch keys are provided for system operation:
– This key turns the unit off and on. The unit will return to the screen that was
active when it was turned off.
– In the DISPLAY MODE, these keys toggle the display through the available
main screens.
In the SELECT MODE, if a parameter has been highlighted, these keys will scroll through
the available settings.
– On the Main screen, this key sequences through the available NIBP or ECG
simulations. On the Setup screen, there are a number of parameters that may be selected
and changed. This key sequences the cursor (Highlight) through those parameters.
– This key is used to RESET the peak pressure reading or to begin a Leak Test.
– This key toggles the unit into and out of the Setup Mode. Depressing this key
will enter the Setup screen where the configuration can be viewed and adjusted.
Depressing the key again will exit the Setup Mode and return to the previously viewed
main screen.
7