BCBiomedical ESU 2000A User Manual

ELECTRO-SURGERY UNIT
ANALYZER
ESU-2000A
USER MANUAL
BC BIOMEDICAL
ESU-2000A
TABLE OF CONTENTS
DESCRIPTION............................................................................................................. 3
OVERVIEW.................................................................................................................. 4
SAFETY PRECAUTIONS ............................................................................................ 5
OPERATING INSTRUCTIONS .................................................................................... 6
MANUAL REVISIONS.................................................................................................. 9
WARRANTY................................................................................................................. 9
SPECIFICATIONS ....................................................................................................... 10
NOTES......................................................................................................................... 12
1
BC GROUP INTERNATIONAL, INC. RESERVES THE RIGHT
WARNING
WARNING
All connections to patients must be removed before
All connections to patients must be removed before
connecting the Device Under Test (DUT) to the
connecting the Device Under Test (DUT) to the
Analyzer. A serious hazard may occur if the patient
Analyzer. A serious hazard may occur if the patient
is connected when testing with the Simulator.
is connected when testing with the Simulator.
Do not connect any leads from the patient directly to
Do not connect any leads from the patient directly to
the Analyzer or DUT.
the Analyzer or DUT.
CAUTION
Always turn the ESU under test off before making or
removing connections. Never touch exposed
metal surfaces on test leads or other current
carrying parts while ESU is activated. Never open
the ESU-2000A when connected to an ESU.
Electro-Surgery units generate
high frequency/RF voltages at high levels.
N O T I C E
TO MAKE CHANGES TO ITS PRODUCTS OR SPECIFICATIONS AT ANY TIME, WITHOUT NOTICE, IN ORDER TO IMPROVE THE DESIGN OR PERFORMANCE AND TO SUPPLY THE BEST POSSIBLE PRODUCT. THE INFORMATION IN THIS MANUAL HAS BEEN CAREFULLY CHECKED AND IS BELIEVED TO BE ACCURATE. HOWEVER, NO RESPONSIBILITY IS ASSUMED FOR INACCURACIES.
Manual ESU-2000A Copyright © 2005
www.bcgroupintl.com Made in the USA
10/05 Rev 02
2
BC GROUP ESU-2000A
ELECTRO-SURGERY UNIT ANALYZER
The Model ESU-2000A is a precision Electro-Surgery Unit Analyzer (ESU). It tests the RF power output of Electro-Surgery Units. The power reading is taken from a large 4 ½ inch analog display. It offers six resistive loads for RF power testing from 50 to 500 ohms, along with a 200 ohm load for RF leakage current testing. The patient return electrode resistance simulation is adjustable in one ohm increments from 0 to 999 ohms.
The following are highlights of some of the main features:
LARGE 4 ½ INCH ANALOG DISPLAY
+/- 5% OF FULL SCALE ACCURACY
RETURN CONTACT MONITOR TEST
RF LEAKAGE TEST
PORTABLE
OSCILLOSCOPE ISOLATED OUTPUT
0-10 MHz BANDWIDTH
EASY OPERATION
ACCESSORY KIT INCLUDED
HIGH IMPLACT PLASTIC ENCLOSURE