BC Biomedical DPM-2001, DPM-2001PLUS, DPM-2100 User Manual

DIGITAL PRESSURE METERS
DPM-2001
USER MANUAL
BC BIOMEDICAL
DPM-2001/2001 PLUS/2100
TABLE OF CONTENTS
DESCRIPTION............................................................................................................. 3
SYSTEM INDICATORS .............................................................................................. 5
KEYS:
DPM-2001 .......................................................................................................... 6
DPM-2001 PLUS................................................................................................ 8
DPM-2100 .......................................................................................................... 10
SETUP MODE ............................................................................................................. 12
INPUTS ........................................................................................................................ 13
STATUS DISPLAYS..................................................................................................... 14
COMMUNICATIONS.................................................................................................... 15
MANUAL REVISIONS.................................................................................................. 20
WARRANTY................................................................................................................. 20
SPECIFICATIONS ....................................................................................................... 21
DRAWINGS
DPM-2001 PHYSICAL DIAGRAM.................................................................... 23
DPM-2001 PLUS PHYSICAL DIAGRAM ......................................................... 24
DPM-2100 PHYSICAL DIAGRAM.................................................................... 25
NOTES......................................................................................................................... 26
1
iii
N O T I C E
iii
BC GROUP INTERNATIONAL, INC. RESERVES THE RIGHT TO MAKE CHANGES TO ITS PRODUCTS OR SPECIFICATIONS AT ANY TIME, WITHOUT NOTICE, IN ORDER TO IMPROVE THE DESIGN OR PERFORMANCE AND TO SUPPLY THE BEST POSSIBLE PRODUCT. THE INFORMATION IN THIS MANUAL HAS BEEN CAREFULLY CHECKED AND IS BELIEVED TO BE ACCURATE. HOWEVER, NO RESPONSIBILITY IS ASSUMED FOR INACCURACIES.
Manual DPM-2001, 2001 PLUS, 2100 Copyright © 2004 www.bcgroupintl.com MADE IN THE USA
10/04 Rev 06
2
BC GROUP
DPM-2001/2001 PLUS/2100
DIGITAL PRESSURE METERS
The Model DPM-2001 Series is a Microprocessor based Digital Pressure Meter family. They measure both gas and liquid pressures and provide multiple engineering unit displays for the results. The DPM-2100 Series adds temperature measurement. The following are highlights of some of the main features.
DPM-2001:
-13.50 TO 100.00 PSI RANGE
0.1% FS ACCURACY (Pressure)
PRESSURE SCALES INCLUDE PSI, inH
DIGITAL CALIBRATION – NO POTS TO TURN
5 DIGIT LCD PLUS SCALE INDICATION
BATTERY LIFE DISPLAY (0 to 100%)
PROGRAMMABLE DIGITAL FILTER
LCD CONTRAST IS SOFTWARE ADJUSTABLE
16 BIT PRESSURE MEASUREMENT
DIGITAL ZERO ADJUST
DPM-2001 PLUS: Includes all the features of the DPM-2001 and:
MAX and MIN PRESSURE VALUE STORAGE
RS232 SERIAL COMMUNICATIONS
DPM-2100: Includes all the features of the DPM-2001 and:
O, cmH2O, AND mmHg
2
MAX and MIN PRESSURE VALUE STORAGE
RS232 SERIAL COMMUNICATIONS
YSI 700 TEMPERATURE PROBE INTERFACE
0.0-100.0 C / 32.0-212.0 F TEMPERATURE RANGE
0.5% FS ACCURACY (Temperature)
MAX and MIN TEMPERATURE VALUE STORAGE
3
OPTIONAL ACCESSORIES:
BE2000PU 120 VAC BATTERY ELIMINATOR, U.S.
BE2000PE 220 VAC BATTERY ELIMINATOR, EUROPE
DPM2000C RS-232 CABLE
BC20-01004 HARD CARRYING CASE
BC20-01005 UNIVERSAL PRESSURE ADAPTER KIT
4
SYSTEM INDICATORS
Four indicators are provided to identify the Current Operating Mode.
PRESSURE SCALE – The pressure scale is indicated by an identifier bar. The RANGE key will toggle the pressure units among PSI, mmHG, inH2O and cmH2O. The following is a breakdown of the available pressure scales and the measurement range for each scale:
Identifier Bar Pressure Units Pressure Range
PSIG -13.50 to 100.00
mmHg -701 to 5190
inH2O -374 to 2773
bar no bar
(flashing)
cmH2O -951 to 7043
NOTE: If the measured pressure is outside of the range of the instrument, the display will display -HI- or -Lo-.
NOTE: InH20, cmH20 and mmHg ranges are calibrated for 20 degrees Celsius.
TEMPERATURE SCALE (MODEL DPM-2100 ONLY) – The temperature can be displayed
in either degrees Celsius or Fahrenheit. When temperature is displayed, the RANGE key will toggle the temperature units between Fahrenheit and Celsius.
Temperature Units Temperature Range
Degrees C 0.0 to 100.0 Degrees F 32.0 to 212.0
NOTE: If the measured temperature is outside of the range of the instrument, the display will display -HI- or -Lo-. If a probe is not connected, the display will show “no probe”.
LOW BATTERY – When the battery life reaches 10 percent, the display will show the message “lo bat” once every minute.
LINE POWER – When meter is running from the optional Battery Eliminator (BE2000PU, BE2000PE), the display will show “Line” instead of the battery life remaining.
5
KEYS – DPM-2001
Four tactile-touch keys are provided for system operation.
ON/SETUP – The function of this key is dependant on the Current Operating Mode as follows:
POWER OFF – If this key is pressed while the power is turned OFF, the power will
be turned ON
PRESSURE MEASUREMENT – If this key is pressed while the pressure is being
displayed, the Setup Mode will be entered.
SETUP MODE – Pressing this key while in the Setup Mode will sequence the
display through the available parameters.
OFF/RESET – The function of this key is dependant on the Current Operating Mode as follows:
PRESSURE MEASUREMENT – If this key is pressed while the pressure is being
displayed, the power will be turned OFF
SETUP MODE – Pressing this key while in the Setup Mode will exit the Setup Mode
and automatically save all settings.
6
RANGE/UP – The function of this key is dependant on the Current Operating Mode as follows:
PRESSURE MEASUREMENT – If this key is pressed while the pressure is being
displayed, the unit will step through the available pressure ranges (PSI, inH20, cmH20 and mmHg).
SETUP MODE – If this key is pressed in the Setup Mode, the value of the displayed
setting will increment. Pressing and holding this key will cause the rapid automatic incrementing of the displayed setting.
BATTERY/DOWN – The function of this key is dependant on the Current Operating Mode as follows:
PRESSURE MEASUREMENT – If this key is pressed while the pressure is being
displayed, the unit will display the percent (0 to 100) of battery life remaining.
NOTE: If the optional wall mount power supply is used, the unit will display “Line”.
SETUP MODE – If this key is pressed in the Setup Mode, the value of the displayed
setting will decrement. Pressing and holding this key will cause the rapid automatic decrementing of the displayed setting.
ZERO – This function is a combination of two keys (RANGE/UP and BATTERY/DOWN). If both keys are depressed and held for 5 seconds, the pressure display will be zeroed.
NOTE: If the BATTERY/DOWN key is depressed first, it will prevent the range from
being changed in the process.
7
KEYS – DPM-2001 PLUS
Six tactile-touch keys are provided for system operation.
ON/SETUP – The function of this key is dependant on the Current Operating Mode as follows:
POWER OFF – If this key is pressed while the power is turned OFF, the power will
be turned ON
PRESSURE MEASUREMENT – If this key is pressed while the pressure is being
displayed, the Setup Mode will be entered.
SETUP MODE – Pressing this key while in the Setup Mode will sequence the
display through the available parameters.
OFF/RESET – The function of this key is dependant on the Current Operating Mode as follows:
PRESSURE MEASUREMENT – If this key is pressed while the pressure is being
displayed, the power will be turned OFF
SETUP MODE – Pressing this key while in the Setup Mode will exit the Setup Mode
and automatically save all settings.
MIN/MAX MODE – Pressing this key when a Min or Max Value is being displayed
(MIN or MAX key is held down) will cause that value in that capture register to be reset to the current reading.
8
Loading...
+ 18 hidden pages