BC Biomedical DA-2006P, DA-2006 User Manual

DEFIBRILLATOR
ANALYZERS
DA-2006P
ANALYZER
W/ PACER
USER MANUAL
BC BIOMEDICAL
DA-2006 SERIES
TABLE OF CONTENTS
WARNINGS ................................................................................................................. 2
DESCRIPTION............................................................................................................. 5
OVERVIEW.................................................................................................................. 13
DEFIBRILLATOR ANALYZER ..................................................................................... 21
MAIN SCREEN.................................................................................................. 21
ECG WAVEFORMS SCREEN .......................................................................... 27
PLAYBACK LAST PULSE SCREEN ................................................................. 31
START CHARGE TIMER SCREEN .................................................................. 33
PRINT HEADER................................................................................................ 35
SELF TEST WAVEFORM ................................................................................. 37
RUNNING A DEFIBRILLATION TEST .............................................................. 39
INTRODUCTION .................................................................................... 39
DIFIBRILLATION TEST .......................................................................... 41
CARDIOVERSION TEST........................................................................ 45
CHARGE TIME TEST............................................................................. 49
SHOCK ADVISORY ALGORITHM TEST ............................................... 51
TRANSCUTANEOUS PACEMAKER ANALYZER ....................................................... 53
PACE MAIN SCREEN ...................................................................................... 53
PACER MODE SETUP SCREEN...................................................................... 59
SENSITIVITY TEST .......................................................................................... 61
REFRACTORY PERIOD TEST ......................................................................... 63
PRINT MENU SCREEN .................................................................................... 65
MANUAL REVISIONS.................................................................................................. 67
WARRANTY................................................................................................................. 67
SPECIFICATIONS ....................................................................................................... 69
NOTES......................................................................................................................... 75
1
The DA-2006 and DA-2006P are intended for
testing only and they should never be used in
diagnostics, treatment or any other capacity
where they would come in contact with a patient.
All connections to patients must be removed
before connecting the Device Under Test (DUT)
to the Analyzer. A serious hazard may occur if
The DA-2006 and DA-2006P
skilled technical personnel only.
WARNING
are for use by
WARNING
WARNING
the patient is connected when testing
with the Analyzer.
Do not connect any leads from the patient
directly to the Analyzer or DUT.
2
NOTICE
BC GROUP INTERNATIONAL, INC. RESERVES THE RIGHT TO MAKE
CHANGES TO ITS PRODUCTS OR SPECIFICATIONS AT ANY TIME, WITHOUT NOTICE, IN ORDER TO IMPROVE THE DESIGN OR PERFORMANCE AND TO SUPPLY THE BEST POSSIBLE PRODUCT. THE INFORMATION IN THIS MANUAL HAS BEEN CAREFULLY CHECKED AND IS BELIEVED TO BE ACCURATE. HOWEVER, NO RESPONSIBILITY
IS ASSUMED FOR INACCURACIES.
CONTACT INFORMATION
BC BIOMEDICAL
BC GROUP INTERNATIONAL, INC.
PO BOX 25125
9415 GENTRY AVE
ST. LOUIS, MO 63125
USA
1-800-242-8428
314-638-3800
www.bcgroupintl.com
sales@bcgroupintl.com
Manual DA-2006 Series Copyright © 2006
www.bcgroupintl.com Made in the USA
7/06 Rev 04
3
This page intentionally left blank.
4
BC GROUP
DA-2006 SERIES
DEFIBRILLATOR ANALYZER
The Model DA-2006 Series is a microprocessor-based instrument family that is used in the testing of defibrillators. They measure the energy output and provided information about the pulse. They are used on manual, semi-automatic and automatic defibrillators with monophasic or biphasic outputs.
The DA-2006P model additionally provides a Transcutaneous Pacemaker analysis function. It measures and displays pacer pulse information as well as performing Refractory Period, Sensitivity and Immunity testing.
All models have a built in 50 ohm human body simulation load as well as 12 lead ECG with arrhythmias and performance waveforms. Additionally, they have a Centronics printer port, a serial port, oscilloscope output, high level ECG output, as well as provision for a battery eliminator.
The DA-2006 Series makes viewing and selecting the desired waveforms and test data quick and intuitive, with all operational information being available on the 240 by 64 pixel graphic display, allowing for easy maneuvering through parameters and scrolling through available options.
NOTE: The instrument is intended for use by trained service technicians.
5
DA-2006
The following are highlights of some of the main features:
GENERAL
SIMPLE TO OPERATE
GRAPHICS DISPLAY WITH SIMULTANEOUS DETAILED STATUS OF
PARAMETERS AND SCROLLING CONTROL OF OPTIONS
ON SCREEN VIEWING OF DEFIBRILLATOR WAVEFORM
DROP DOWN CHOICE SCREENS LIST ALL OPTIONS FOR PARAMETERS
MONOPHASIC AND BIPHASIC COMPATABLE
5000 V, 1000 JOULE CAPACITY
HIGH AND LOW RANGES
CARDIOVERSION DELAY MEASURMENT
CHARGE TIME MEASUREMENT
WAVEFORM STORAGE AND PLAYBACK
10 UNIVERSAL PATIENT LEAD CONNECTORS
25 PIN CONNECTOR FOR CENTRONICS PRINTER
9 VOLT BATTERY POWER
LOW BATTERY INDICATOR
AVAILABLE BATTERY ELIMINATOR
DISPLAY BACKLIGHT
FULL REMOTE OPERATION VIA RS-232
FLASH PROGRAMMABLE FOR UPGRADES
PACEMAKER OPTION
26 SELECTABLE INTERNAL LOADS
FULL PULSE ANALYSIS
DEMAND SENSITIVITY TEST
REFRACTORY PERIOD TESTS
50/60 Hz INTERFERENCE TEST SIGNALS
6
Description
ENERGY OUTPUT MEASUREMENT GENERAL
The unit measures the energy in the output pulse of both monophasic and biphasic defibrillators.
PULSE TYPE: Monophasic or Biphasic
LOAD RESISTANCE: 50 Ohms +/- 1%, non-inductive (<1 µH)
DISPLAY RESOLUTION: 0.1 Joules
MEASUREMENT TIME WINDOW: 100 ms
ABSOLUTE MAX PEAK VOLTAGE: 6000 Volts
CARDIOVERSION DELAY: 0 to 6000 ms
CARDIOVERSION RESOLUTION: 0.1 ms
ENERGY OUTPUT MEASUREMENT HIGH RANGE
The high range allows for a large pulse with high voltage and current.
VOLTAGE: <5000 Volts
MAX CURRENT: 120 Amps
MAX ENERGY: 1000 Joules
TRIGGER LEVEL: 100 Volts
PLAYBACK AMPLITUDE: 1 mv / 1000 V Lead I
TEST PULSE: 125 Joules +/- 20%
ENERGY OUTPUT MEASUREMENT LOW RANGE
The low range allows greater resolution on smaller pulses.
VOLTAGE: <1000 Volts
MAX CURRENT: 24 Amps
MAX ENERGY: 50 Joules
TRIGGER LEVEL: 20 Volts
PLAYBACK AMPLITUDE: 1 mv / 1000 V Lead I
TEST PULSE: 5 Joules +/- 20%
7
DA-2006
ENERGY OUTPUT MEASUREMENT OTHER
OSCILLIOSCOPE OUTPUT
HIGH MEASURE RANGE: 1000:1 amplitude-attenuated
LOW MEASURE RANGE: 200:1 amplitude-attenuated
WAVEFORM PLAYBACK
OUTPUT – LEAD 1 & PLATES
GRAPHICS SCREEN
200:1 Time Base Expansion
SYNC TIME MEASUREMENTS
TIMING WINDOW: Starts 40 ms before each R-wave peak
TEST WAVEFORMS: All waveform simulations available
CHARGE TIME MEASUREMENT
From 0 .1 to 99.9 sec
ECG FUNCTIONS
The unit can produce a wide variety of ECG simulations. The user simply selects the parameters that match the desired output.
RATE: 30,40,45,60,80,90,100,120,140,160,180,200,220,240,260,280,300 BPM
AMPLITUDE: 0.50,1.0,1.5,2.0 mV (Lead II)
ECG-PERFORMANCE FUNCTIONS
The unit can generate Sine, Square, Triangular, and Pulse waveforms with adjustable amplitudes for performance testing.
SINE: 0.1,0.2,0.5,5,10,40,50,60,100 Hz
SQUARE: 0.125,2 Hz
TRIANGLE: 2,2.5 Hz
PULSE: 30,60,120 BPM; 60 ms WIDTH
AMPLITUDE: 0.5,1.0,1.5,2.0 mV (Lead II)
8
Description
ARRHYTHMIA FUNCTIONS
The unit can simulate 12 different arrhythmias.
VENTRICULAR FIBRILLATION
ATRIAL FIBRILLATION
SECOND DEGREE A-V BLOCK
RIGHT BUNDLE BRANCH BLOCK
PREMATURE ATRIAL CONTRACTION
EARLY PVC
STANDARD PVC
R ON T PVC
MULTIFOCAL PVC
BIGEMINY
RUN OF 5 PVC
VENTRICULAR TACHYCARDIA
SHOCK ADVISORY TESTS
The unit can simulate 8 different waveforms to test the shock algorithm of advanced defibrillators:
ASYSTOLE
COARSE VENTRICULAR FIBRILLATION
FINE VENTRICULAR FIBRILLATION
MULTIFOCAL VENTRICULAR TACHYCARDIA @ 140 BPM
MULTIFOCAL VENTRICULAR TACHYCARDIA @ 160 BPM
POLYFOCAL VENTRICULAR TACHYCARDIA @ 140 BPM
POLYFOCAL VENTRICULAR TACHYCARDIA @ 160 BPM
SUPRAVENTRICULAR TACHYCARDIA @ 90 BPM
9
DA-2006
TRANSCUTANEOUS PACER ANALYZER
The unit can test external transcutaneous pacemakers. It has a wide variety of loads and can measure the Pacer Pulse, Demand Sensitivity and Refractory Periods (Pacing and Sensing):
LOAD:
RANGE: 50,100,150,200,300,400,500,600,700,800,900,1000,1100,
1200,1300,1400,1500,1600,1700,1800,1900,2000,2100, 2200,2300 Ohms
PULSE:
AMPLITUDE: 4 TO 300 mA (100 Ohm load)
RATE: 30 TO 800 ppm
WIDTH: 0.6 to 80 ms
DEMAND SENSITIVITY:
WAVEFORMS:
SELECTIONS: SQUARE, TRIANGLE, HAVERSINE
WIDTH: 10,25,40,100,200 ms
ECG:
AMPLITUDE – OUT: 0 to 4 mv
PACER INPUT (50 TO 400 OHMS):
AMPLITUDE – OUT: 0 to 10 mv / 50 Ohms
RATE – IN: 30 to 120 ppm
PACER INPUT (500 TO 2300 OHMS & OPEN):
AMPLITUDE – OUT: 0 to 100 mv
RATE – IN: 30 to 120 ppm
DEFIBRILLATOR PLATES:
AMPLITUDE – OUT: 0 to 10 mv
RATE – IN: 30 to 120 ppm
REFRACTORY PERIOD:
PACING: 20 to 500 ms
SENSING: 20 to 500 ms
50/60 HZ INTERFERENCE TEST SIGNAL:
ECG OUTPUT: 0,0.4,0.8,1.2,1.6,2.0,2.4,2.8,3.2,3.6,4.0 mv
PACER INPUT 50 OHMS: 0,1,2,3,4,5,6,7,8,9,10 mv
PACER INPUT 100 OHMS: 0,2,4,6,8,10,12,14,16,18,20 mv
PACER INPUT 150 OHMS: 0,3,6,9,12,15,18,21,24,27,30 mv
PACER INPUT 200 OHMS: 0,4,8,12,16,20,24,28,32,26,40 mv
PACER INPUT 300 OHMS: 0,6,12,18,24,30,36,42,48,54,60 mv
PACER INPUT 400 OHMS: 0,8,16,24,32,40,48,56,64,72,80 mv
PACER INPUT > 500 OHMS: 0,10,20,30,40,50,60,70,80,90,100 mv
DEFIBRILLATOR PLATES: 0,1,2,3,4,5,6,7,8,9,10 mv
10
Description
ACCESSORIES
BC20 - 40032 INTERNAL PADDLE ADAPTERS (2 adapters) BC20 - 21103 BATTERY ELIMINATOR (US Version) BC20 - 21101 BATTERY ELIMINATOR (Euro Version) BC20 - 00427 PLASTIC ELECTRODE PLATES (2 plates)
OPTIONAL ACCESSORIES
BC20 - 30108 BC BIOMEDICAL MEDIUM SOFT SIDED CARRYING CASE BC20 - 41341 COMMUNICATION CABLE (DB 9 M to DB 9 F) BC20 - 00420 PHYSIO-CONTROL DEFIB / PACE TEST CABLE BC20 - 00421 MARQUETTE DEFIB / PACE TEST CABLE BC20 - 00423 ZOLL DEFIB/PACE TEST CABLE BC20 - 00424 PHYSIO-CONTROL PACE ONLY TEST CABLE BC20 - 00425 ZOLL PACE ONLY TEST CABLE BC20 - 00426 HP / AGILENT / LAERDAL / AAMI
DEFIB / PACE TEST CABLE
11
DA-2006
This page intentionally left blank.
12
p
OVERVIEW
This section looks at the layout of a DA-2006 and gives descriptions of the elements that
are present.
LCD Graphical Display:
10 Universal Patient Lead Connectors: RA R LA L RL(-) N LL F V1 C1 V2 C2 V3 C3 V4 C4 V5 C5 V6 C6
Back Light Key for turning on and off the backlight
Shows Parameters for Test Data and Waveforms
5 Light Touch Keys for Dynamic functions: These keys are labeled in the bottom portion of the screen and change function based on operating mode.
Range Key for selecting Defib input range high or low and
ace mode
9V Battery Large Defib Plates: Fixed 50 Ohm input load for Defibrillator testing
13
Compartment
(Back)
DA-2006 Series
p
This section looks at the layout of a DA-2006P and gives descriptions of the elements that
are present.
10 Universal Patient Lead Connectors: RA R LA L RL(-) N LL F V1 C1 V2 C2 V3 C3 V4 C4 V5 C5 V6 C6
LCD Graphical Display: Shows Parameters for Test Data and Waveforms
5 Light Touch Keys for Dynamic functions: These keys are labeled in the bottom portion of the screen and change function based on operating mode.
Back Light Key for turning on and off the backlight
Pacemaker Input Jacks Variable Load (optional): 50-2300 Ohms
Large Defib Plates: Fixed 50 Ohm input load for Defibrillator testing
Range Key for selecting Defib input range high or low and
ace mode
9V Battery Compartment (Back)
Pacer Input Jacks Fixed Load: 50 Ohms
14
Overview
A
cro
r
This section looks at the layout of the back and gives descriptions of the elements that are
present.
Oscilloscope output BNC connector for easy viewing of input waveforms
Power Switch
Parallel Port
C Adaptor/ Battery Eliminator
2.1 mm Mi
Serial Port High Level ECG output RCA Jack
Female 9-Pin
D-Sub
Connecto
Female 25-Pin D-Sub Connector
NOTE
The DA-2006 and the DA-2006P offer the same
features, with the DA-2006P having the addition
of a Transcutaneous Pacemaker Analyzer
function (See Pacemaker Analyzer section for
more details).
15
DA-2006 Series
General Operation
The unit is controlled by 7 light touch keys. They allow the user to move around within the
displayed parameters, select the desired options, choose a specific category and control
the setup for the unit. When a key is depressed there is an audio click when it is accepted,
or a razz tone if the key is invalid.
A large LCD graphics display with backlight provides the user with information about the
current status of the device configuration options, test results and more. The display
identifies the function of each key on a dynamic basis. As the operation mode changes,
the keys change to suit the operating mode.
Range Key
The key scrolls through the ranges of the DA-2006 Series analyzers. Depressing
the key will allow the user to select between High Defibrillator Range (1000J max), Low
Defibrillator Range (50J max) and Pacemaker Range. The default mode on power up is
High Defibrillator Range.
Backlight Key
The Graphic LCD display may be viewed with or without the backlight. Depressing any key
will activate the backlight. However, since the backlight will drain the battery if left on, it will
automatically shut off after 5 seconds when running on battery power.
The key is provided to toggle the backlight on or off at any time.
16
Overview
Function Keys
There are five keys that are used to provide general operational control. The
functions of the keys vary depending on the current screen. The section of the screen
just above the key indicates its current meaning.
NOTE: Only functions that are available to the user will be visible at any given time.
Sample Function Key Labeling
ECG Waveforms
The microprocessor has stored in its memory all of the digitalized waveforms. It sends the
waveforms to a D/A converter, which generates an accurate analog representation. The
waveform is then sent through resistor networks, developing the appropriate signals on the
output terminals.
17
DA-2006 Series
Universal Patient Lead Connectors
The 10 Universal Patient Lead Connectors allow for 12 lead ECG simulations. AHA and
IEC color-coded labels are located on the face of the unit to aid in connecting the
corresponding U.S. and International Patient Leads.
AHA Label IEC Label Description
RA R Right Arm
LA L Left Arm
RL N
LL F Left Leg
V1 V2 V3 V4 V5 V6
C1 C2 C3 C4 C5 C6
Right Leg (reference or ground)
V Leads (V1-V6) (U.S. and Canada)
also referred to as pericardial, precordial or unipolar chest leads
Chest Leads (C1-C6) (International)
18
Overview
High Level Output (+)
A high level ECG output signal (200 x Amplitude Setting) is available on the RCA jack
located on the rear of the unit.
Serial Port
A female 9-pin D-Sub connector is provided for the connection of the unit to a PC or laptop
serial port (e.g. Com 1). This link is then used for either remote control or flash
downloading of software upgrades.
Parallel Port
A female 25-pin D-Sub connector is provided for the connection of a printer via a
Centronics parallel interface.
Oscilloscope Output
A BNC connector is provided to connect an oscilloscope to the unit. This output is a 200:1
attenuated version of the input to the Defibrillator Plates.
19
DA-2006 Series
Power Switch
A rocker switch is provided on the rear of the unit to turn the power on and off.
Power Supply
The unit utilizes two 9 Volt Alkaline Batteries in the bottom battery compartments. When
the unit detects a LOW BATTERY condition (10% Battery Life), a warning window will
appear once per minute to alert the user. If the batteries are not replaced before they reach
a critical level (0 % Battery Life), the unit will shut down.
Battery Eliminator
The unit has a 2.1 mm micro jack for connecting a 10-Volt AC battery eliminator. The
adapter will power the unit, but will not charge the battery.
20
DEFIBRILLATOR ANALYZER
When the DA-2006 is first powered up, the Defibrillator Analyzer MAIN SCREEN will be
displayed. This screen shows the current CONFIGURATION, the TEST RESULTS and the
available FUNCTION KEYS. All defibrillator tests are run from the MAIN SCREEN. When
the unit detects an input of greater than 100 volts on the Defibrillator Plates (20 volts in Low
Range), it will automatically begin a test.
The default configuration is the High Range Defibrillator mode. This mode allows for a
MAIN SCREEN
waveform of up to 1000 Joules to be analyzed.
The following is a sample screen for this mode:
21
Defibrillator Analyzer
The key may be used to toggle the unit to the Low Range Defibrillator mode. This
mode allows for waveforms up to 50 joules to be analyzed. The Defibrillator Analyzer
works the same in both ranges. The lower range simply provides for a higher resolution for
smaller pulsed.
The following is a sample screen for this mode:
NOTE: The key will also put the DA-2006P into the Transcutaneous Pacemaker
Analyzer mode (See Pacemaker Analyzer section for more information).
22
Loading...
+ 53 hidden pages