BC Biomedical DA-2006, DA-2006P User manual

DEFIBRILLATOR
ANALYZERS
DA-2006P
ANALYZER
W/ PACER
SERVICE MANUAL
BC BIOMEDICAL
SERVICE MANUAL
DA-2006 SERIES
TABLE OF CONTENTS
WARNINGS, CAUTIONS, NOTICES ...............................................................................ii
MANUAL REVISIONS.....................................................................................................iv
WARRANTY....................................................................................................................iv
MANUALS
DA-2006 Series User Manual (Rev 06)............................................. SECTION A
DA-CS-06 (Rev 02) ........................................................................... SECTION A
OPERATIONAL THEORY.............................................................................. SECTION B
VALIDATION / CALIBRATION .......................................................................SECTION C
PARTS ...........................................................................................................SECTION D
This Service Manual covers the following units:
DA-2006
DA-2006P
i
The DA-2006 Series Analyzers are intended to be
serviced only by authorized service personnel.
Troubleshooting and service procedures
WARNING - MODIFICATIONS
The DA-2006 Series Analyzers are intended for
use within the published specifications. Any
application beyond these specifications or any
unauthorized user modifications may result in
CAUTION - SERVICE
should only be performed by
qualified technical personnel.
hazards or improper operation.
ii
USER ASSUMES FULL RESPONSIBILITY FOR UNAUTHORIZED EQUIPMENT MODIFICATIONS OR APPLICATION OF EQUIPMENT OUTSIDE OF THE PUBLISHED INTENDED USE AND SPECIFICATIONS. SUCH MODIFICATIONS OR APPLICATIONS MAY RESULT IN EQUIPMENT DAMAGE OR PERSONAL INJURY.
BC GROUP INTERNATIONAL, INC. RESERVES THE RIGHT TO MAKE CHANGES TO ITS PRODUCTS OR SPECIFICATIONS AT ANY TIME, WITHOUT NOTICE, IN ORDER TO IMPROVE THE DESIGN OR PERFORMANCE AND TO SUPPLY THE BEST POSSIBLE PRODUCT. THE INFORMATION IN THIS MANUAL HAS BEEN CAREFULLY CHECKED AND IS BELIEVED TO BE ACCURATE. HOWEVER, NO RESPONSIBILITY IS ASSUMED FOR INACCURACIES.
NOTICE – CONTACT INFORMATION
NOTICE – DISCLAIMER
NOTICE – DISCLAIMER
BC BIOMEDICAL
BC GROUP INTERNATIONAL, INC.
PO BOX 25125
9415 GENTRY AVE
ST. LOUIS, MO 63125
USA
1-800-242-8428
314-638-3800
www.bcgroupintl.com
sales@bcgroupintl.com
Service Manual DA-2006 Series Copyright © 2007
www.bcgroupintl.com Made in the USA
5/07 Rev 02
iii
A
A
MANUAL REVISIONS
Revision # Engineering # Revisions Made
Rev 01 7395 Origination
Rev 02 395 Drawing scans inserted
LIMITED WARRANTY
WARRANTY
FROM DEFECTS IN MATERIALS AND WORKMANSHIP UNDER THE SERVICE FOR WHICH THEY
RE INTENDED. THIS WARRANTY IS EFFECTIVE FOR TWELVE MONTHS FROM THE DATE OF
SHIPMENT.
EXCLUSIONS: THIS WARRANTY IS IN LIEU OF ANY OTHER WARRANTY EXPRESSED OR IMPLIED, INCLUDING, BUT NOT LIMITED TO ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE.
BC GROUP INTERNATIONAL, INC. IS NOT LIABLE FOR ANY INCIDENTAL OR CONSEQUENTIAL DAMAGES.
NO PERSON OTHER THAN AN OFFICER IS AUTHORIZED TO GIVE ANY OTHER WARRANTY OR
SSUME ANY LIABILITY.
REMEDIES: THE PURCHASER'S SOLE AND EXCLUSIVE REMEDY SHALL BE: (1) THE REPAIR OR REPLACEMENT OF DEFECTIVE PARTS OR PRODUCTS, WITHOUT CHARGE. (2) AT THE OPTION OF BC GROUP INTERNATIONAL, INC., THE REFUND OF THE PURCHASE PRICE.
: BC GROUP INTERNATIONAL, INC. WARRANTS ITS NEW PRODUCTS TO BE FREE
P:\Manuals\BCGroup\…\DA-2006_Series_SM_Rev02.doc
iv
SECTION A
USER MANUAL
DA-2006 SERIES
REV 06
DEFIBRILLATOR
ANALYZERS
DA-2006P
ANALYZER
W/ PACER
USER MANUAL
BC BIOMEDICAL
DA-2006 SERIES
TABLE OF CONTENTS
WARNINGS, CAUTIONS, NOTICES ........................................................................... 2
DESCRIPTION............................................................................................................. 7
OVERVIEW.................................................................................................................. 15
DEFIBRILLATOR ANALYZER ..................................................................................... 23
MAIN SCREEN.................................................................................................. 23
ECG WAVEFORMS SCREEN .......................................................................... 31
PLAYBACK LAST PULSE SCREEN ................................................................. 35
START CHARGE TIMER SCREEN .................................................................. 37
PRINT HEADER................................................................................................ 39
SELF TEST WAVEFORM ................................................................................. 41
RUNNING A DEFIBRILLATION TEST .............................................................. 43
INTRODUCTION .................................................................................... 43
DEFIBRILLATION TEST......................................................................... 45
CARDIOVERSION TEST........................................................................ 49
CHARGE TIME TEST............................................................................. 53
SHOCK ADVISORY ALGORITHM TEST ............................................... 55
TRANSCUTANEOUS PACEMAKER ANALYZER ....................................................... 57
PACE MAIN SCREEN ...................................................................................... 57
PACER MODE SETUP SCREEN...................................................................... 63
SENSITIVITY TEST .......................................................................................... 65
REFRACTORY PERIOD TEST ......................................................................... 67
PRINT MENU SCREEN .................................................................................... 69
PLAYBACK LAST PULSE SCREEN ................................................................. 71
1
DA-2006 Series
MESSAGES ................................................................................................................. 73
SYSTEM SETUP.......................................................................................................... 75
POWER UP SETTINGS............................................................................................... 77
AUTO SEQUENCE FUNCTION................................................................................... 79
VIEW MODE ..................................................................................................... 81
RUN MODE ...................................................................................................... 85
PROGRAMMING AUTO SEQUENCES ....................................................................... 95
MANUAL REVISIONS................................................................................................ 107
WARRANTY............................................................................................................... 107
SPECIFICATIONS ..................................................................................................... 109
NOTES....................................................................................................................... 115
2
DA-2006 Series
WARNING - USERS
The DA-2006 Series Analyzers are for use by
skilled technical personnel only.
WARNING - USE
The DA-2006 Series Analyzers are intended for
testing only and they should never be used in
diagnostics, treatment or any other capacity
where they would come in contact with a patient.
WARNING - MODIFICATIONS
The DA-2006 Series Analyzers are intended for
use within the published specifications. Any
application beyond these specifications or any
unauthorized user modifications may result in
hazards or improper operation.
WARNING - CONNECTIONS
All connections to patients must be removed
before connecting the Device Under Test (DUT)
to the Analyzer. A serious hazard may occur
if the patient is connected when
testing with the Analyzer.
Do not connect any leads from the patient
directly to the Analyzer or DUT.
WARNING - POWER ADAPTOR
Unplug the Power Adaptor before
cleaning the surface of the Analyzer.
WARNING - LIQUIDS
Do not submerge or spill liquids on the Analyzer.
Do not operate the Analyzer if internal
components not intended for use with fluids may
have been exposed to fluid, as the internal
leakage may have caused corrosion and be a
potential hazard.
3
DA-2006 Series
CAUTION - SERVICE
The DA-2006 Series Analyzers are intended to be
serviced only by authorized service personnel.
Troubleshooting and service procedures
should only be performed by
qualified technical personnel.
CAUTION - ENVIRONMENT
The DA-2006 Series Analyzers are intended to
function between 15 and 40 °C.
Exposure to temperatures outside this range can
adversely affect the performance of the Analyzer.
CAUTION - CLEANING
Do not immerse. The Analyzer should be
cleaned by wiping gently with a damp, lint-free
cloth.
CAUTION - INSPECTION
The DA-2006 Series Analyzers should be
inspected before each use for wear and
the Analyzer should be serviced
if any parts are in question.
NOTICE – SYMBOLS
Symbol Description
Caution
(Consult Manual for Further Information)
Center Positive
Direct Current
4
DA-2006 Series
NOTICE – ABBREVIATIONS
A, Amps Amperes
BPM Beats Per Minute
-2
c centi- (10
C Celsius
° degree
dt Delta Time, Change in Time
DUT Device Under Test
E Energy
ECG Electrocardiogram
Euro European
Hz hertz J Joules
k kilo- (10 kg kilograms
lbs pounds
µ micro- (10
µA microampere
µH microhertz µV microvolt
µsec microsecond
m milli- (10
mA milliampere
mm millimeter ms, mS,
msec
millisecond
mV millivolts
ohm
P Power
ppm pulse per minute
R Resistance, ohms
Sec, S seconds
US United States
V volt
VDC Direct Current Voltage
)
3
)
-6
)
-3
)-
5
DA-2006 Series
USER ASSUMES FULL RESPONSIBILITY FOR UNAUTHORIZED EQUIPMENT MODIFICATIONS OR APPLICATION OF EQUIPMENT OUTSIDE OF THE PUBLISHED INTENDED USE AND SPECIFICATIONS. SUCH MODIFICATIONS OR APPLICATIONS MAY RESULT IN EQUIPMENT DAMAGE OR PERSONAL INJURY.
BC GROUP INTERNATIONAL, INC. RESERVES THE RIGHT TO MAKE CHANGES TO ITS PRODUCTS OR SPECIFICATIONS AT ANY TIME, WITHOUT NOTICE, IN ORDER TO IMPROVE THE DESIGN OR PERFORMANCE AND TO SUPPLY THE BEST POSSIBLE PRODUCT. THE INFORMATION IN THIS MANUAL HAS BEEN CAREFULLY CHECKED AND IS BELIEVED TO BE ACCURATE. HOWEVER, NO RESPONSIBILITY IS ASSUMED FOR INACCURACIES.
NOTICE – CONTACT INFORMATION
NOTICE – DISCLAIMER
NOTICE – DISCLAIMER
BC BIOMEDICAL
BC GROUP INTERNATIONAL, INC.
PO BOX 25125
9415 GENTRY AVE
ST. LOUIS, MO 63125
USA
1-800-242-8428
314-638-3800
www.bcgroupintl.com
sales@bcgroupintl.com
Manual DA-2006 Series Copyright © 2007
www.bcgroupintl.com Made in the USA
5/07 Rev 06
6
BC GROUP
DA-2006 SERIES
DEFIBRILLATOR ANALYZER
The Model DA-2006 Series is a microprocessor-based instrument family that is used in the testing of defibrillators. They measure the energy output and provide information about the defibrillation pulse. They are used on manual, semi-automatic and automatic defibrillators with monophasic or biphasic outputs.
The DA-2006P model additionally provides a Transcutaneous Pacemaker analysis function. It measures and displays pacer pulse information as well as performing Refractory Period, Sensitivity and Immunity testing.
All models have a built in 50 ohm human body simulation load as well as 12 lead ECG with arrhythmias and performance waveforms. Additionally, they have a Centronics printer port, a serial port, oscilloscope output, high-level ECG output, as well as provision for a battery eliminator.
The DA-2006 Series makes viewing and selecting the desired waveforms and test data quick and intuitive, with all operational information being available on the 240 by 64 pixel graphic display, allowing for easy maneuvering through parameters and scrolling through available options.
NOTE: The instrument is intended for use by trained service technicians.
7
DA-2006
The following are highlights of some of the main features:
GENERAL
SIMPLE TO OPERATE
GRAPHICS DISPLAY WITH SIMULTANEOUS DETAILED STATUS OF
PARAMETERS AND SCROLLING CONTROL OF OPTIONS
ON SCREEN VIEWING OF DEFIBRILLATOR AND PACEMAKER WAVEFORMS
DROP DOWN CHOICE SCREENS LIST ALL OPTIONS FOR PARAMETERS
MONOPHASIC AND BIPHASIC COMPATIBLE
5000 V, 1000 JOULE CAPACITY
HIGH AND LOW RANGES
CARDIOVERSION DELAY MEASUREMENT
CHARGE TIME MEASUREMENT
WAVEFORM STORAGE AND PLAYBACK
10 UNIVERSAL PATIENT LEAD CONNECTORS
25 PIN CONNECTOR FOR CENTRONICS PRINTER
9 VOLT BATTERY POWER
LOW BATTERY INDICATOR
AVAILABLE BATTERY ELIMINATOR
DISPLAY BACKLIGHT
FULL REMOTE OPERATION VIA RS-232
FLASH PROGRAMMABLE FOR UPGRADES
AUTO SEQUENCE TESTING CAPABLE OF STORING 50 CUSTOM TEST
SEQUENCES
PACEMAKER OPTION
26 SELECTABLE INTERNAL LOADS
FULL PULSE ANALYSIS
DEMAND SENSITIVITY TEST
REFRACTORY PERIOD TESTS
50/60 Hz INTERFERENCE TEST SIGNALS
INPUT TERMINALS AND CIRCUITRY PROTECTED AGAINST ACCIDENTAL
DEFIBRILLATOR DISCHARGE INTO PACEMAKER TEST TERMINALS
8
Description
ENERGY OUTPUT MEASUREMENT GENERAL
The unit measures the energy in the output pulse of both monophasic and biphasic defibrillators.
PULSE TYPE: Monophasic or Biphasic
LOAD RESISTANCE: 50 ohm +/- 1%, non-inductive (<1 µH)
DISPLAY RESOLUTION: 0.1 Joules
MEASUREMENT TIME WINDOW: 100 ms
ABSOLUTE MAX PEAK VOLTAGE: 6000 Volts
CARDIOVERSION DELAY: 0 to 6000 ms
CARDIOVERSION RESOLUTION: 0.1 ms
ENERGY OUTPUT MEASUREMENT HIGH RANGE
The high range allows for a large pulse with high voltage and current.
VOLTAGE: <5000 Volts
MAX CURRENT: 120 Amps
MAX ENERGY: 1000 Joules
TRIGGER LEVEL: 100 Volts
PLAYBACK AMPLITUDE: 1 mV / 1000 V Lead I
TEST PULSE: 125 Joules +/- 20%
ENERGY OUTPUT MEASUREMENT LOW RANGE
The low range allows greater resolution on smaller pulses.
VOLTAGE: <1000 Volts
MAX CURRENT: 24 Amps
MAX ENERGY: 50 Joules
TRIGGER LEVEL: 20 Volts
PLAYBACK AMPLITUDE: 1 mV / 1000 V Lead I
TEST PULSE: 5 Joules +/- 20%
9
DA-2006
ENERGY OUTPUT MEASUREMENT OTHER
OSCILLOSCOPE OUTPUT
HIGH MEASUREMENT RANGE: 1000:1 amplitude-attenuated
LOW MEASUREMENT RANGE: 200:1 amplitude-attenuated
WAVEFORM PLAYBACK
OUTPUT – LEAD 1 & PLATES
GRAPHICS SCREEN
200:1 Time Base Expansion
SYNC TIME MEASUREMENTS
TIMING WINDOW: Starts at peak of each R-wave
TEST WAVEFORMS: All waveform simulations available
CHARGE TIME MEASUREMENT
From 0 .1 to 99.9 sec
ECG FUNCTIONS
The unit can produce a wide variety of ECG simulations. The user simply selects the parameters that match the desired output.
RATE: 30,40,45,60,80,90,100,120,140,160,180,200,220,240,260,280,300 BPM
AMPLITUDE: 0.50,1.0,1.5,2.0 mV (Lead II)
ECG-PERFORMANCE FUNCTIONS
The unit can generate Sine, Square, Triangular, and Pulse waveforms with adjustable amplitudes for performance testing.
SINE: 0.1,0.2,0.5,5,10,40,50,60,100 Hz
SQUARE: 0.125,2 Hz
TRIANGLE: 2,2.5 Hz
PULSE: 30,60,120 BPM; 60 ms WIDTH
AMPLITUDE: 0.5,1.0,1.5,2.0 mV (Lead II)
10
Description
ARRHYTHMIA FUNCTIONS
The unit can simulate 12 different arrhythmias.
VENTRICULAR FIBRILLATION
ATRIAL FIBRILLATION
SECOND DEGREE A-V BLOCK
RIGHT BUNDLE BRANCH BLOCK
PREMATURE ATRIAL CONTRACTION
EARLY PVC
STANDARD PVC
R ON T PVC
MULTIFOCAL PVC
BIGEMINY
RUN OF 5 PVC
VENTRICULAR TACHYCARDIA
SHOCK ADVISORY TESTS
The unit can simulate 8 different waveforms to test the shock algorithm of advanced defibrillators:
ASYSTOLE
COARSE VENTRICULAR FIBRILLATION
FINE VENTRICULAR FIBRILLATION
MULTIFOCAL VENTRICULAR TACHYCARDIA @ 140 BPM
MULTIFOCAL VENTRICULAR TACHYCARDIA @ 160 BPM
POLYFOCAL VENTRICULAR TACHYCARDIA @ 140 BPM
POLYFOCAL VENTRICULAR TACHYCARDIA @ 160 BPM
SUPRAVENTRICULAR TACHYCARDIA @ 90 BPM
11
DA-2006
TRANSCUTANEOUS PACER ANALYZER
The unit can test external transcutaneous pacemakers. It has a wide variety of loads and can measure the Pacer Pulse, Demand Sensitivity and Refractory Periods (Pacing and Sensing):
LOAD:
RANGE: 50,100,150,200,300,400,500,600,700,800,900,1000,1100,
1200,1300,1400,1500,1600,1700,1800,1900,2000,2100, 2200,2300 ohm
PULSE:
PULSE CURRENT: 4 TO 300 mA (100 ohm load)
RATE: 30 TO 800 ppm
WIDTH: 0.6 to 80 ms
DEMAND SENSITIVITY:
WAVEFORMS:
SELECTIONS: SQUARE, TRIANGLE, HAVERSINE
WIDTH: 10,25,40,100,200 ms
ECG:
AMPLITUDE – OUT: 0 to 4 mV
PACER INPUT (50 TO 400 OHM):
AMPLITUDE – OUT: 0 to 10 mV / 50 ohms
RATE – IN: 30 to 120 ppm
PACER INPUT (500 TO 2300 OHM & OPEN):
AMPLITUDE – OUT: 0 to 100 mV
RATE – IN: 30 to 120 ppm
DEFIBRILLATOR PLATES:
AMPLITUDE – OUT: 0 to 10 mV
RATE – IN: 30 to 120 ppm
REFRACTORY PERIOD:
PACING: 20 to 500 ms
SENSING: 20 to 500 ms
50/60 HZ INTERFERENCE TEST SIGNAL:
ECG OUTPUT: 0,0.4,0.8,1.2,1.6,2.0,2.4,2.8,3.2,3.6,4.0 mV
PACER INPUT 50 OHM: 0,1,2,3,4,5,6,7,8,9,10 mV
PACER INPUT 100 OHM: 0,2,4,6,8,10,12,14,16,18,20 mV
PACER INPUT 150 OHM: 0,3,6,9,12,15,18,21,24,27,30 mV
PACER INPUT 200 OHM: 0,4,8,12,16,20,24,28,32,26,40 mV
PACER INPUT 300 OHM: 0,6,12,18,24,30,36,42,48,54,60 mV
PACER INPUT 400 OHM: 0,8,16,24,32,40,48,56,64,72,80 mV
PACER INPUT > 500 OHM: 0,10,20,30,40,50,60,70,80,90,100 mV
DEFIBRILLATOR PLATES: 0,1,2,3,4,5,6,7,8,9,10 mV
INPUT CIRCUITRY PROTECTION
INPUT CIRCUITRY IS PROTECTED AGAINST DAMAGE IN THE
EVENT OF AN ACCIDENTAL DEFIBRILLATOR DISCHARGE INTO THE PACEMAKER TEST INPUT TERMINALS
12
Description
ACCESSORIES
BC20 - 40032 INTERNAL PADDLE ADAPTERS (2 adapters) BC20 - 21103 BATTERY ELIMINATOR (120 VAC) (US Version) BC20 - 21101 BATTERY ELIMINATOR (220 VAC) (Euro Version) BC20 - 00427 PLASTIC ELECTRODE PLATES (2 plates)
OPTIONAL ACCESSORIES
BC20 - 30108 BC BIOMEDICAL MEDIUM SOFT SIDED CARRYING CASE BC20 - 41341 COMMUNICATION CABLE (DB 9 M to DB 9 F) BC20 - 00420 PHYSIO-CONTROL DEFIB / PACE TEST CABLE BC20 - 00421 MARQUETTE DEFIB / PACE TEST CABLE BC20 - 00423 ZOLL DEFIB/PACE TEST CABLE BC20 - 00424 PHYSIO-CONTROL PACE ONLY TEST CABLE BC20 - 00425 ZOLL PACE ONLY TEST CABLE BC20 - 00426 HP / AGILENT / LAERDAL / AAMI
DEFIB / PACE TEST CABLE
13
DA-2006
This page intentionally left blank.
14
OVERVIEW
This section looks at the layout of a DA-2006 and gives descriptions of the elements that
are present.
LCD Graphical Display:
10 Universal Patient Lead Connectors: RA R LA L RL(-) N LL F V1 C1 V2 C2 V3 C3 V4 C4 V5 C5 V6 C6
Back Light Key for turning on and off the backlight
Shows Parameters for Test Data and Waveforms
5 Light Touch Keys for Dynamic functions: These keys are labeled in the bottom portion of the screen and change function based on operating mode.
Range Key for selecting Defib input range high or low
Large Defib Plates: Fixed 50 ohm input load for Defibrillator testing
9V Battery Compartment (Rear)
15
DA-2006 Series
This section looks at the layout of a DA-2006P and gives descriptions of the elements that
are present.
10 Universal Patient Lead Connectors: RA R LA L RL(-) N LL F V1 C1 V2 C2 V3 C3 V4 C4 V5 C5 V6 C6
LCD Graphical Display: Shows Parameters for Test Data and Waveforms
5 Light Touch Keys for Dynamic functions: These keys are labeled in the bottom portion of the screen and change function based on operating mode.
Back Light Key for turning on and off the backlight
Pacemaker Input Jacks Variable Load (optional): 50-2300 ohms
Large Defib Plates: Fixed 50 ohm input load for Defibrillator testing
Range Key for selecting Defib input range high or low and pace mode
9V Battery Compartment (Rear)
Pacer Input Jacks Fixed Load: 50 ohm
16
Overview
A
r
This section looks at the layout of the back and gives descriptions of the elements that are
present.
Oscilloscope output BNC connector for easy viewing of input waveforms
Power Switch
Parallel Port
C Adaptor/ Battery Eliminator
2.1 mm Micro
High Level ECG output RCA Jack
Serial Port Female 9-Pin D-Sub Connecto
Female 25-Pin D-Sub Connector
NOTE
The DA-2006 and the DA-2006P offer the same
features, with the DA-2006P having the addition
of a Transcutaneous Pacemaker Analyzer
function (See Pacemaker Analyzer section
for more details).
17
DA-2006 Series
General Operation
The unit is controlled by 7 light touch keys. They allow the user to move around within the
displayed parameters, select the desired options, choose a specific category and control
the setup for the unit. When a key is depressed there is an audio click when it is accepted,
or a razz tone if the key is invalid.
A large LCD graphics display with backlight provides the user with information about the
current status of the device configuration options, test results and more. The display
identifies the function of each key on a dynamic basis. As the operation mode changes,
the key functions change to suit the operating mode.
Range Key
The key scrolls through the ranges of the DA-2006 Series analyzers. Depressing
the key will allow the user to select between High Defibrillator Range (1000J max), Low
Defibrillator Range (50J max) and, with the DA-2006P, Pacemaker Range. The default
mode on power up is High Defibrillator Range.
Backlight Key
The Graphic LCD display may be viewed with or without the backlight. Depressing any key
will activate the backlight. However, since the backlight will drain the battery if left on, it will
automatically shut off after a user programmable delay when running on battery power.
The key is provided to toggle the backlight on or off at any time.
18
Overview
Function Keys
There are five keys that are used to provide general operational control. The
functions of the keys vary depending on the current screen. The section of the screen
just above the key indicates its current meaning.
NOTE: Only functions that are available to the user will be visible at any given time.
Sample Function Key Labeling
ECG Waveforms
The microprocessor has stored in its memory all of the digitalized waveforms. It sends the
waveforms to a D/A converter, which generates an accurate analog representation. The
waveform is then sent through resistor networks, developing the appropriate signals on the
output terminals.
19
DA-2006 Series
Universal Patient Lead Connectors
The 10 Universal Patient Lead Connectors allow for 12 lead ECG simulations. AHA and
IEC color-coded labels are located on the face of the unit to aid in connecting the
corresponding U.S. and International Patient Leads.
AHA Label IEC Label Description
RA R Right Arm
LA L Left Arm
RL N
LL F Left Leg
V1 V2 V3 V4 V5 V6
C1 C2 C3 C4 C5 C6
Right Leg (reference or ground)
V Leads (V1-V6) (U.S. and Canada)
also referred to as pericardial, precordial or unipolar chest leads
Chest Leads (C1-C6) (International)
20
Overview
High Level Output (+)
A high level ECG output signal (200 X Amplitude Setting) is available on the RCA jack
located on the rear of the unit.
Serial Port
A female 9-pin D-Sub connector is provided for the connection of the unit to a PC or laptop
serial port (e.g. Com 1). This link is then used for either remote control or flash
downloading of software upgrades.
Parallel Port
A female 25-pin D-Sub connector is provided for the connection of a printer via a
Centronics parallel interface.
Oscilloscope Output
A BNC connector is provided to connect an oscilloscope to the unit. This output is a 200:1
attenuated version of the input to the Defibrillator Plates.
21
DA-2006 Series
Power Switch
A rocker switch is provided on the rear of the unit to turn the power on and off.
Power Supply
The unit utilizes two 9 Volt Alkaline Batteries in the bottom battery compartments. When
the unit detects a LOW BATTERY condition (10% Battery Life), a warning window will
appear once per minute to alert the user.
Battery Eliminator
The unit has a 2.1 mm micro jack for connecting a 10-Volt AC battery eliminator. The
adapter will power the unit, but will not charge the battery.
22
DEFIBRILLATOR ANALYZER
When the DA-2006 is first powered up, the Defibrillator Analyzer MAIN SCREEN will be
displayed. This screen shows the current CONFIGURATION, the TEST RESULTS and the
available FUNCTION KEYS. All defibrillator tests are run from the MAIN SCREEN. When
the unit detects an input of greater than 100 Volts on the Defibrillator Plates (20 Volts in
Low Range), it will automatically begin a test.
The default configuration is the High Range Defibrillator mode. This mode allows for a
MAIN SCREEN
waveform of up to 1000 Joules to be analyzed.
The following is a sample screen for this mode:
23
Loading...
+ 141 hidden pages