BC Biomedical DA-2003P User Manual

User Manual
DA-2003P
Defibrillator Transutaneous
Pacemaker Analyzer
CONTENTS
II. INTRODUCTION.................................................................. 3
2.1 Features
2.2 General Information
III. SPECIFICATIONS ............................................................... 4
3.1 Defibrillator Analyzer
3.2 Transcutaneous Pacemaker Analyzer
IV. INSTALLATION ................................................................... 8
4.1 Receipt, Inspection and Return
4.2 Setup
4.3 Power
4.4 Internal Paddles
4.5 Special Contacts
V. OPERATING ...................................................................... 10
5.1 Control Switches and Connections
5.2 Menu and Function Keys
5.3 Menu and Messages: Defibrillator Mode
5.4 Menu and Messages: Transcutaneous Pacemaker Mode
5.5 Test Result Printouts
VI. DEFIBRILLATOR MODE TESTING .................................. 16
6.1 Introduction
6.2 Test Preparation
6.3 Energy Test
6.4 Cardioversion Test
6.5 Maximum Energy Charging Time Test
6.6 Shock Advisory Algorithm Test
VII. TRANSCUTANEOUS PACEMAKER MODE TESTING .... 21
7.1 Introduction
7.2 Test Preparation
7.3 Demand Sensitivity Test
7.4 Refractory Period Test
VIII. WARRANTY ...................................................................... 25
IX. TECHNICAL SUPPORT .................................................... 25
I. MANUAL REVISION RECORD
This record page is for recording revisions to your DA-2003P User Manual that have been published by BC Biomedical or its authorized representatives. We recommend that only the management or facility representative authorized to process changes and revisions to publications:
Make the pen changes or insert the revised pages;
Ensure that obsolete pages are withdrawn and either disposed of immediately, or marked as
superseded and placed in a superseded document file, and;
Enter the information below reflecting that the revisions have been entered.
Rev No Date Entered Reason Signature of Person Entering Change
0 - Initial Release
1 11/12/04 Part Number Change KLB
2
BC Biomedical DA-2003P User Manual 11/04
2.1 Features
II. INTRODUCTION
The DA-2003P Analyzer is a precision instrument for testing defibrillators and transcutaneous pacemakers, and is designed to be used by trained service technicians.
The defibrillator function of the DA-2003P measures the energy output, and ensures that the defibrillator complies with specified requirements. DA-2003P has a built-in load resistance of 50 ohm, which roughly corresponds to the impedance of the human body. The defibrillator pads are placed on the DA-2003P contact plates. Thus, the defibrillator is connected through the load resistance. When the defibrillator is discharged, DA-2003P calculates and displays the energy delivered.
In the pacer function the DA-2003P tests all types of transthoracic pacemakers. The testing is menu driven, and simple to operate. DA-2003P measures and displays a pulse’s amplitude, rate, energy and width. It also conducts demand sensitivity tests, measuring and displaying refractory periods, and immunity tests, which determine the pacemaker’s susceptibility to 50/60 Hz interference.
2.2 General Information
Temperature Requirements +15°C to +35°C when operating
0°C to +50°C in storage Display Type LCD graphic display Alphanumeric format 6 lines, 40 characters
Data Input/ Output (2)
Parallel printer port (1); Bi-directional RS -232C (1) for Computer control
Power
2 x 9 volt alkaline Battery Duracell
equivalent) for 20 -25 operational hours, or 240
VAC (Battery Eliminator), 115 VAC for US. Mechanical Specifications Housing High impact plastic case Height 9.8 cm 3.9 in. Width 24.8 cm 9.8 in. Depth 28.0 cm 11.0 in. Weight 2.06 kg (with battery) 4.5 lbs
Recommended Printer Standard Accessories
HP DeskJet 500C / 550C and Canon BJ -10SX.
110 V or 220 V AC Adapter (P/N BC20-00429) Internal paddle-contact adapter (P/N BC20-00430) Snap-to-banana adapters (10 pk) (P/N BC20-00427) DA-2003P User Manual Protective Cover (P/N BC20-30108)
Additional Accessories
Defib. paddle adapter (specify defibrillator type) Pacemaker External Load Cable (specify pacemaker type) Soft Carrying Case (P/N BC20-30108) Storage
Store in the carrying case in dry surroundings within the temperature range specified, without battery. There are no other storage requirements.
Periodic Inspection
The unit should be calibrated every 12 months.
BC Biomedical DA-2003P User Manual 11/04
3
MN1604 (or
3.1 Defibrillator Analyzer
1. Energy Output Measurement
High Range
Voltage <5000 volts Maximum current 120 amperes Maximum energy 1000 Joules Accuracy
Trigger level 100 volts Playback amplitude 1 mV/1000 V Lead I Test pulse 100 + 4 Joules
Low Range
Voltage <1000 volts Maximum current 24 amperes Maximum energy 50 Joules
Accuracy
Trigger level 20 volts Playback amplitude 1 mV/200 V Lead I Test pulse Approx. 4 Joules Load Resistance Display Resolution 0.1 Joules Measurement Time Window Absolute Max. Peak Voltage Pulse Width 100 ms Cardioversion
Oscilloscope Output High measure range 1000:1 amplitude-attenuated
Low measure range 200:1 amplitude-attenuated
III. SPECIFICATIONS
± 2 % of reading for >100 Joules ± 2 Joule of reading for <100 Joules
± 2% of reading for >20 Joules ± 2 Joule of reading for <20 Joules
50 ohms ± 1%, non-inductive (<1 µH)
100 ms
6000 volts
Measured time delay ± 2 ms
2. ECG Wave
Waveform Storage And Playback
Discharge can be viewed via ECG outputs and paddles. Output: 200:1 Time Base expansion.
Sync Time Measurements Timing window Starts - 40 ms at each R-wave peak.
Test waveforms All waveform simulations available. Delay time accuracy
± 1 ms
Charge Time Measurement From 0.1 seconds to 99.9 seconds.
ECG General Lead configuration 12-lead simulation. RL, RA, LA, LL, V1-6 Output impedance Limb leads 1000 ohms to RL
V Leads 1000 ohms to RL
4
BC Biomedical DA-2003P User Manual 11/04
All other signals are in relative proportion to Lead amplitude as follows: The amplitudes are shown for a Lead I amplitude by 1 mV: Lead I 1.0 mV (LA - RA) Lead II 1.5 mV (LL - RA) Lead III 0.5 mV (LL - LA) V Lead 1.5 mV (V - 1/3 (LL+LA+RA))
High Level Output (ECG Jack) 1/4" standard phone-jack with an amplitude of 1V/mV of low level Lead II signal
Defibrillator Contact Plates Same amplitude as Lead I low level ECG.
1 mV between contact surfaces.
Playback 200 to 1 time-base expansion of defibrillator pulse by playback to ECG Leads
Manual ECG Performance Test DC Pulse 4 seconds 1.0 mV
Square wave 2 Hz 1.0 mV p-p biphasic Triangular wave 2 Hz 1.0 mV Sine 0.1, 0.2, 0.5, 10, 40, 50, 60, and 100 Hz Amplitude 0.5, 1.0, 1.5, 2.0 mV (Lead II) Accuracy
± 5 % (Lead II 1.0 mV)
ECG Performance Test Gain/Damping 2 Hz square wave
Frequency Response Low Frequency 4 second DC pulse Band Pass 10 Hz sine Monitor -3dB point: 40 Hz sine Power Line Notch
50 Hz sine Filter Linearity 2 Hz triangle wave
Normal Sinus Rates 30, 60, 80, 120, 180, 240 and 300 BPM.
Accuracy
±1% of selection Amplitudes 0.5, 1.0, 1.5 and 2.0 mV (Lead II) Accuracy
±5 % (Lead II 1.0 mV)
Automatic ECG Rate Test
Arrhythmia Selections
vfib Ventricular Fibrillation afib Atrial Fibrillation blk II Second degree A-V block RBBB Right Bundle Branch Block PAC Premature Atrial Contraction PVC_E Early PVC PVC_STD PVC PVCRonT R on T PVC mfPVC Multifocal PVC bigeminy Bigeminy run5PVC Bigeminy Run of 5 PVCs vtach Ventricular Tachycardia
BC Biomedical DA-2003P User Manual 11/04
5
Shock Advisory Test Algorithms ASYS Asystole
SVTa_90 Supraventricular Tachycardia PVT_140 PVT_ 160 MVT_140 MVT_160 CVF Course Ventricular Fibrillation FVF Fine Ventricular Fibrillation
6
BC Biomedical DA-2003P User Manual 11/04
3.2 Transcutaneous Pacemaker Analyzer
1. TEST LOAD RANGE
50 to 2300 ohms in step of: 50 ohm up to 200 ohms 100 ohm from 200 up to 2300 ohms Accuracy
Oscilloscope Output 50 - 150 ohm 10.24:1 amplitude attenuation
200 - 500 ohm 41:1 amplitude attenuation 600 - 2300 ohm 164:1 amplitude attenuation
2. PULSE MEASUREMENTS
Amplitude 4 to 300 mA (100 ohm load) Accuracy Max. Amplitude 300 mA all loads Rate 30 to 800 Pm Accuracy Pulse width 0.6 to 80 ms Accuracy
50 - 1300 ohm ±1% 1400 - 2300 ohm ±1.5 %
±5 % or ±0.5 mA
±1% or 2 Pm
±1% or ±0.3 ms
3. DEMAND SENSITIVITY TEST
Waveforms Square(SQR), Triangle(TRI), and Havemine (SSQ) ECG output Amplitude 0 - 4 mV
Pacer input (Load depended) Amplitude (50 ohm) 0 10 mV Resolution (50 ohm)
Amplitude: (500 ohm) 0 - 100 mV Resolution: (500 ohm) 1 mV Defib. Pads Amplitude 0 10 mV Resolution 0.1 mV Waveform width 10, 25, 40, 100 and 200 ms Pacer rate 30 to 120 Pm
Immunity Test 50/60 Hz Interference Signal ECG output 0 - 4 mV peak in steps of 0.4 mV Pacer input (Load dependent) 0 - 10 mV peak in steps of 1 mV (50 ohm)
Defibrillator pads 0 - 10 mV peak in steps of 1 mV
Resolution 40 µV
40 µV
0 - 100 mV peak in steps of 10 mV (500 ohm)
4. Refractory Period Measurement
20 to 500 ms (both Pacing and Sensing) Accuracy: ±2 ms
BC Biomedical DA-2003P User Manual 11/04
7
Loading...
+ 18 hidden pages