I. MANUAL REVISION RECORD ........................................... 2
II. INTRODUCTION.................................................................. 3
2.1 Features
2.2 General Information
III. SPECIFICATIONS ............................................................... 4
3.1 Defibrillator Analyzer
3.2 Transcutaneous Pacemaker Analyzer
IV. INSTALLATION ................................................................... 8
4.1 Receipt, Inspection and Return
4.2 Setup
4.3 Power
4.4 Internal Paddles
4.5 Special Contacts
V. OPERATING ...................................................................... 10
5.1 Control Switches and Connections
5.2 Menu and Function Keys
5.3 Menu and Messages: Defibrillator Mode
5.4 Menu and Messages: Transcutaneous Pacemaker Mode
5.5 Test Result Printouts
VI. DEFIBRILLATOR MODE TESTING .................................. 16
6.1 Introduction
6.2 Test Preparation
6.3 Energy Test
6.4 Cardioversion Test
6.5 Maximum Energy Charging Time Test
6.6 Shock Advisory Algorithm Test
VII. TRANSCUTANEOUS PACEMAKER MODE TESTING .... 21
7.1 Introduction
7.2 Test Preparation
7.3 Demand Sensitivity Test
7.4 Refractory Period Test
VIII. WARRANTY ...................................................................... 25
IX. TECHNICAL SUPPORT .................................................... 25
I. MANUAL REVISION RECORD
This record page is for recording revisions to your DA-2003P User Manual that have been published by
BC Biomedical or its authorized representatives. We recommend that only the management or facility
representative authorized to process changes and revisions to publications:
• Make the pen changes or insert the revised pages;
• Ensure that obsolete pages are withdrawn and either disposed of immediately, or marked as
superseded and placed in a superseded document file, and;
• Enter the information below reflecting that the revisions have been entered.
Rev No Date Entered Reason Signature of Person Entering Change
0 - Initial Release
1 11/12/04 Part Number Change KLB
2
BC Biomedical DA-2003P User Manual 11/04
2.1 Features
II. INTRODUCTION
The DA-2003P Analyzer is a precision instrument for testing defibrillators and transcutaneous
pacemakers, and is designed to be used by trained service technicians.
The defibrillator function of the DA-2003P measures the energy output, and ensures that the defibrillator
complies with specified requirements. DA-2003P has a built-in load resistance of 50 ohm, which roughly
corresponds to the impedance of the human body. The defibrillator pads are placed on the DA-2003P
contact plates. Thus, the defibrillator is connected through the load resistance. When the defibrillator is
discharged, DA-2003P calculates and displays the energy delivered.
In the pacer function the DA-2003P tests all types of transthoracic pacemakers. The testing is menu
driven, and simple to operate. DA-2003P measures and displays a pulse’s amplitude, rate, energy and
width. It also conducts demand sensitivity tests, measuring and displaying refractory periods, and
immunity tests, which determine the pacemaker’s susceptibility to 50/60 Hz interference.
2.2 General Information
Temperature Requirements +15°C to +35°C when operating
0°C to +50°C in storage
Display
Type LCD graphic display
Alphanumeric format 6 lines, 40 characters
Data Input/ Output (2)
Parallel printer port (1); Bi-directional RS -232C
(1) for Computer control
Power
2 x 9 volt alkaline Battery Duracell
equivalent) for 20 -25 operational hours, or 240
VAC (Battery Eliminator), 115 VAC for US.
Mechanical Specifications
Housing High impact plastic case
Height 9.8 cm 3.9 in.
Width 24.8 cm 9.8 in.
Depth 28.0 cm 11.0 in.
Weight 2.06 kg (with battery) 4.5 lbs
Recommended Printer
Standard Accessories
HP DeskJet 500C / 550C and Canon BJ -10SX.
110 V or 220 V AC Adapter (P/N BC20-00429)
Internal paddle-contact adapter (P/N BC20-00430)
Snap-to-banana adapters (10 pk) (P/N BC20-00427) DA-2003P User Manual
Protective Cover (P/N BC20-30108)
Store in the carrying case in dry surroundings within the temperature range
specified, without battery. There are no other storage requirements.
Periodic Inspection
The unit should be calibrated every 12 months.
BC Biomedical DA-2003P User Manual 11/04
3
MN1604 (or
3.1 Defibrillator Analyzer
1. Energy Output Measurement
High Range
Voltage <5000 volts
Maximum current 120 amperes
Maximum energy 1000 Joules
Accuracy
Trigger level 100 volts
Playback amplitude 1 mV/1000 V Lead I
Test pulse 100 + 4 Joules
Low Range
Voltage <1000 volts
Maximum current 24 amperes
Maximum energy 50 Joules
Accuracy
Trigger level 20 volts
Playback amplitude 1 mV/200 V Lead I
Test pulse Approx. 4 Joules
Load Resistance
Display Resolution 0.1 Joules
Measurement Time
Window
Absolute Max. Peak
Voltage
Pulse Width 100 ms
Cardioversion
Oscilloscope Output
High measure range 1000:1 amplitude-attenuated
Low measure range 200:1 amplitude-attenuated
III. SPECIFICATIONS
± 2 % of reading for >100 Joules
± 2 Joule of reading for <100 Joules
± 2% of reading for >20 Joules
± 2 Joule of reading for <20 Joules
50 ohms ± 1%, non-inductive (<1 µH)
100 ms
6000 volts
Measured time delay ± 2 ms
2. ECG Wave
Waveform Storage And Playback
Discharge can be viewed via ECG outputs and paddles.
Output: 200:1 Time Base expansion.
Sync Time Measurements
Timing window Starts - 40 ms at each R-wave peak.
Test waveforms All waveform simulations available.
Delay time accuracy
± 1 ms
Charge Time Measurement
From 0.1 seconds to 99.9 seconds.
ECG General
Lead configuration 12-lead simulation. RL, RA, LA, LL, V1-6
Output impedance Limb leads 1000 ohms to RL
V Leads 1000 ohms to RL
4
BC Biomedical DA-2003P User Manual 11/04
All other signals are in relative proportion to Lead amplitude as follows:
The amplitudes are shown for a Lead I amplitude by 1 mV:
Lead I 1.0 mV (LA - RA)
Lead II 1.5 mV (LL - RA)
Lead III 0.5 mV (LL - LA)
V Lead 1.5 mV (V - 1/3 (LL+LA+RA))
High Level Output (ECG Jack)
1/4" standard phone-jack with an amplitude of 1V/mV of low level Lead II signal
Defibrillator Contact Plates
Same amplitude as Lead I low level ECG.
1 mV between contact surfaces.
Playback
200 to 1 time-base expansion of defibrillator pulse by playback to ECG Leads
Manual ECG Performance Test
DC Pulse 4 seconds 1.0 mV
ECG Performance Test
Gain/Damping 2 Hz square wave
Frequency Response
Low Frequency 4 second DC pulse
Band Pass 10 Hz sine
Monitor -3dB point: 40 Hz sine
Power Line Notch
50 Hz sine
Filter
Linearity 2 Hz triangle wave
Normal Sinus
Rates 30, 60, 80, 120, 180, 240 and 300 BPM.
Accuracy
±1% of selection
Amplitudes 0.5, 1.0, 1.5 and 2.0 mV (Lead II)
Accuracy
±5 % (Lead II 1.0 mV)
Automatic ECG Rate Test
Arrhythmia Selections
vfib Ventricular Fibrillation
afib Atrial Fibrillation
blk II Second degree A-V block
RBBB Right Bundle Branch Block
PAC Premature Atrial Contraction
PVC_E Early PVC
PVC_STD PVC
PVCRonT R on T PVC
mfPVC Multifocal PVC
bigeminy Bigeminy
run5PVC Bigeminy Run of 5 PVCs
vtach Ventricular Tachycardia