This publication is protected by copyright and all rights are reserved. No part of this manual may be
reproduced or transmitted in any form, or by any means electronic or mechanical, including
photocopying and recording, for any purpose other than the purchaser's personal use without the
written permission of the product manufacturer.
Restrictions and Liabilities
Information in this document is subject to change, and does not represent a commitment by BC
Group International Inc. (BC Group). Changes made to the information in this document will be
incorporated in new editions of the publication. No responsibility is assumed by BC Group for the use
or reliability of software or equipment that is not supplied by BC Group.
Trademarks
BC Biomedical and DA2000 are trademarks of BC Group International Inc.
Notices / Contents
Contents
Section 1 Introduction 1-1
Description 1-2
Front Panel 1-2
Back Panel 1-3
Specifications 1-4
Accessories 1-8
Section 2 Installation & Operation 2-1
Operation Warnings 2-1
Getting Acquainted 2-2
Measuring Defibrillator Energy 2-7
Defib Pulse Playback 2-8
Oscilloscope Viewing of Output 2-8
Measuring Synchronization 2-9
Generating Waveforms 2-11
Procedure 2-11
Testing High Level Out 2-12
Measuring Peak Voltage, Current & Overshoot 2-12
Charge Time (DA2000W, P) 2-13
Pacer (Non-Invasive) Testing 2-14
Pacer Refractory Period Testing 2-15
Autosequencing (DA2000P) 2-16
Programming an Automatic Test Sequence 2-17
Running an Automatic Test Sequence 2-19
Resetting the DA2000 for Factory Defaults 2-19
Section 3 Data Transfer & Utilities 3-1
Serial Port 3-1
BC Biomedical Serial Cable (optional accessory) 3-2
Printing 3-2
The Utilities Menu 3-5
Adjusting RS232 Parameters 3-5
Adjusting Display Contrast 3-6
Computer Control Commands 3-6
Communications Protocol 3-6
Initializing Computer Control 3-6
Section 4 Troubleshooting & Service 4-1
Warranty 4-1
Storage and Shipping 4-1
Troubleshooting 4-2
Returning the DA2000 for Service 4-2
ii
DA2000 User's Guide – Rev. 7/03
Introduction
Section 1
Regular testing of defibrillators and pacemakers is critical in order to ensure safe and effective operation. The BC
Biomedical DA2000 Defibrillator Analyzer accurately verifies the output characteristics of all defibrillators as well as
testing the parameters of non-invasive pacemakers. The DA2000 is battery operated and completely portable.
Simple-to-use menu soft keys allow quick access to tests.
The DA2000 measures the delivered energy in Joules (watt-seconds) from a defibrillator by simulating the human
body’s resistance. The analyzer then measures the flow of current through that resistance. The standard
resistance used by the DA2000 is 50 ohms. Defibrillator energy is measured in one of two ranges: 0-100 Joules, or
0-1000 Joules.
The defibrillator pulse waveform can be replayed via the ECG jacks or paddle plates for viewing on a recorder, or
on an oscilloscope for greater detail.
Synchronization time in milliseconds is measured by timing the firing delay from either the Q-wave (base) or Rwave (peak) simulated by the DA2000. The simulated waveform is present at both the ECG jacks and the paddle
plates. Peak voltage and peak current (amps) of the defibrillator pulse can be measured. Overshoot voltage and
current measurements of the defibrillator pulse are calculated and displayed.
The DA2000 also measures the defibrillator’s charge time (the time it takes for a defibrillator to reach its maximum
charge setting).
Waveforms, including ECG, arrhythmias, and performance, help verify monitor and recorder accuracy, and also test
the automatic defibrillator’s ability to recognize and fire.
All waveforms are present at the ECG jacks, the paddle plates and scope output. Utilities allow the setting of Serial
RS-232 communication parameters to download results to printer, or computer. Display contrast can be adjusted to
obtain the best view of the LCD display.
Upgrading the DA2000
A number of pre-configured DA2000 models are available. In addition, older models may be upgraded by contacting
BC Biomedical. The following DA2000 models are available:
•DA2000: Base unit. Features output energy, synchronization time, peak measurements, bi-directional
RS-232.
•DA2000W: “Waveform” version. Features output energy, sync time, peak measurements, overshoot, bi-
directional RS-232, waveforms, charge time measurements, 28 programmable autosequences.
•DA2000P:“Pacer” version. Output energy, sync time, peak measurements, overshoot, bi-directional RS-
232, waveforms, charge time measurements, 28 programmable autosequences, pacer output
measurements and pacer refractory period measurements.
DA2000 User's Guide 1-1
Introduction
Description
Figure 1.1. DA2000 Isometric View
Front Panel
The front panel of the DA2000 (Figure 1.2) includes the following operating features:
1. Universal ECG jacks that utilize AHA and International color coding, allowing for waveform output to
monitor/recorder.
2. A “High Level ECG” Banana jack provides 1 Volt peak output of the selected waveform.
3. The “Common” Banana jack provides ground for the “High Level ECG” and “Defib Scope Out” jacks.
Figure 1.2. DA2000 Front Panel Layout
4. The “Defib Scope Out” Banana jack provides pulse output to an oscilloscope.
5. A range switch allows for defibrillator settings from 0 to 1000 Joules (high), for power below 0-100 Joules
(low) for increased accuracy, and a PACER range setting for pacer output measurements.
6. The Power switch enables the DA2000 (I = ON, O = OFF).
1-2 DA2000 User's Guide
Introduction
7. LCD display: 24 characters x 2 lines. The upper line of the LCD display provides messages and test
results, while the bottom line displays menu choices.
8. Five “soft keys” can be used to select the desired function highlighted on the lower line of the display.
9. Two nickel-plated Defibrillator Paddle Plates are available for defibrillator paddle contact. All waveforms
are present at the paddle plates simultaneously with the ECG jacks.
Back Panel
The Back Panel includes a battery holder that houses a 9-Volt alkaline battery, and a DC battery eliminator jack.
An RS-232 D-9 pin Serial Port allows communications to a computer, serial printer or other test equipment.