BC Biomedical DA-2000 User Manual

User Manual
DA-2000
Defibrillator Analyzer
Notices
BC Group International Inc.
9415 Gentry Avenue St. Louis, MO 63125 USA
Phone: 314-638-3800 Toll-Free: 888-223-6763 Fax: 314-638-3200
E-mail: sales@bcgroupintl.com Internet: www.bcgroupintl.com
All Rights Reserved
This publication is protected by copyright and all rights are reserved. No part of this manual may be reproduced or transmitted in any form, or by any means electronic or mechanical, including photocopying and recording, for any purpose other than the purchaser's personal use without the written permission of the product manufacturer.
Restrictions and Liabilities
Information in this document is subject to change, and does not represent a commitment by BC Group International Inc. (BC Group). Changes made to the information in this document will be incorporated in new editions of the publication. No responsibility is assumed by BC Group for the use or reliability of software or equipment that is not supplied by BC Group.
Trademarks
BC Biomedical and DA2000 are trademarks of BC Group International Inc.
Notices / Contents
Contents
Section 1 Introduction 1-1
Description 1-2 Front Panel 1-2 Back Panel 1-3 Specifications 1-4 Accessories 1-8
Section 2 Installation & Operation 2-1
Operation Warnings 2-1 Getting Acquainted 2-2 Measuring Defibrillator Energy 2-7 Defib Pulse Playback 2-8 Oscilloscope Viewing of Output 2-8 Measuring Synchronization 2-9 Generating Waveforms 2-11 Procedure 2-11 Testing High Level Out 2-12 Measuring Peak Voltage, Current & Overshoot 2-12 Charge Time (DA2000W, P) 2-13 Pacer (Non-Invasive) Testing 2-14 Pacer Refractory Period Testing 2-15 Autosequencing (DA2000P) 2-16
Programming an Automatic Test Sequence 2-17
Running an Automatic Test Sequence 2-19 Resetting the DA2000 for Factory Defaults 2-19
Section 3 Data Transfer & Utilities 3-1
Serial Port 3-1 BC Biomedical Serial Cable (optional accessory) 3-2 Printing 3-2 The Utilities Menu 3-5 Adjusting RS232 Parameters 3-5 Adjusting Display Contrast 3-6 Computer Control Commands 3-6 Communications Protocol 3-6 Initializing Computer Control 3-6
Section 4 Troubleshooting & Service 4-1
Warranty 4-1 Storage and Shipping 4-1 Troubleshooting 4-2 Returning the DA2000 for Service 4-2
ii
DA2000 User's Guide – Rev. 7/03
Introduction
Section 1
Regular testing of defibrillators and pacemakers is critical in order to ensure safe and effective operation. The BC Biomedical DA2000 Defibrillator Analyzer accurately verifies the output characteristics of all defibrillators as well as testing the parameters of non-invasive pacemakers. The DA2000 is battery operated and completely portable. Simple-to-use menu soft keys allow quick access to tests.
The DA2000 measures the delivered energy in Joules (watt-seconds) from a defibrillator by simulating the human body’s resistance. The analyzer then measures the flow of current through that resistance. The standard resistance used by the DA2000 is 50 ohms. Defibrillator energy is measured in one of two ranges: 0-100 Joules, or 0-1000 Joules.
The defibrillator pulse waveform can be replayed via the ECG jacks or paddle plates for viewing on a recorder, or on an oscilloscope for greater detail.
Synchronization time in milliseconds is measured by timing the firing delay from either the Q-wave (base) or R­wave (peak) simulated by the DA2000. The simulated waveform is present at both the ECG jacks and the paddle plates. Peak voltage and peak current (amps) of the defibrillator pulse can be measured. Overshoot voltage and current measurements of the defibrillator pulse are calculated and displayed.
The DA2000 also measures the defibrillator’s charge time (the time it takes for a defibrillator to reach its maximum charge setting).
Waveforms, including ECG, arrhythmias, and performance, help verify monitor and recorder accuracy, and also test the automatic defibrillator’s ability to recognize and fire.
All waveforms are present at the ECG jacks, the paddle plates and scope output. Utilities allow the setting of Serial RS-232 communication parameters to download results to printer, or computer. Display contrast can be adjusted to obtain the best view of the LCD display.
Upgrading the DA2000
A number of pre-configured DA2000 models are available. In addition, older models may be upgraded by contacting BC Biomedical. The following DA2000 models are available:
DA2000: Base unit. Features output energy, synchronization time, peak measurements, bi-directional RS-232.
DA2000W: “Waveform” version. Features output energy, sync time, peak measurements, overshoot, bi- directional RS-232, waveforms, charge time measurements, 28 programmable autosequences.
DA2000P: “Pacer” version. Output energy, sync time, peak measurements, overshoot, bi-directional RS- 232, waveforms, charge time measurements, 28 programmable autosequences, pacer output measurements and pacer refractory period measurements.
DA2000 User's Guide 1-1
Introduction
Description
Figure 1.1. DA2000 Isometric View
Front Panel
The front panel of the DA2000 (Figure 1.2) includes the following operating features:
1. Universal ECG jacks that utilize AHA and International color coding, allowing for waveform output to
monitor/recorder.
2. A “High Level ECG” Banana jack provides 1 Volt peak output of the selected waveform.
3. The “Common” Banana jack provides ground for the “High Level ECG” and “Defib Scope Out” jacks.
Figure 1.2. DA2000 Front Panel Layout
4. The “Defib Scope Out” Banana jack provides pulse output to an oscilloscope.
5. A range switch allows for defibrillator settings from 0 to 1000 Joules (high), for power below 0-100 Joules
(low) for increased accuracy, and a PACER range setting for pacer output measurements.
6. The Power switch enables the DA2000 (I = ON, O = OFF).
1-2 DA2000 User's Guide
Introduction
7. LCD display: 24 characters x 2 lines. The upper line of the LCD display provides messages and test
results, while the bottom line displays menu choices.
8. Five “soft keys” can be used to select the desired function highlighted on the lower line of the display.
9. Two nickel-plated Defibrillator Paddle Plates are available for defibrillator paddle contact. All waveforms
are present at the paddle plates simultaneously with the ECG jacks.
Back Panel
The Back Panel includes a battery holder that houses a 9-Volt alkaline battery, and a DC battery eliminator jack. An RS-232 D-9 pin Serial Port allows communications to a computer, serial printer or other test equipment.
DA2000 User's Guide
1-3
Specifications
OUTPUT POWER MEASUREMENT
Load Resistance: 50 ohms +/- 1% non-inductive
160 watts
Range: 1000 J: 0-1000.0 Joules 100 J: 0-100.0 Joules
Resolution: 0.1 Joule
Max. Voltage: 1000 J: 5500 Volts
100 J: 1750 Volts
Max. Current: 1000 J: 110 Amps 100 J: 35 Amps
Measurement Trip Levels: 1000 J: 66 +/- 5 Volts
100 J: 20 +/- 5 Volts
Pulse Width: 1-50 ms
Accuracy:
1000 J Range: +/- 2% of reading 100-1000 J: +/- 2 Joules 100 J Range: +/- 2% of reading +/- 0.1 Joule
SYNCHRONIZATION MEASUREMENTS
Range: 0-199.9 ms
Measurement: From peak of R-wave
From base of R-wave
Accuracy: 1% of fullscale or +/- 2 ms
QRS complex: Rates: 30, 60, 120, 180, 240 BPM
Rate Accuracy: +/- 1% of setting
Amplitude: Fixed at 1 mV Lead II
Amplitude Accuracy: +/- 2% (RA-LL)
Introduction
ECG WAVEFORMS
(RA-LL) Fixed at 1.1 mV (Apex- Sternum)
+/- 10% (Apex-Sternum)
1-4 DA2000 User's Guide
Introduction
PERFORMANCE WAVEFORMS
Pulse: 30, 60 BPM
pulse width -60 ms
Triangle Wave: 2 Hz
Square Wave: 0.125 Hz, 2 Hz, 50% duty cycle
Sine Waves: 10, 40, 50, 60, 100 Hz
Time Base Accuracy: +/- 1% of setting
Amplitude: Fixed at 1 mV Lead II (RA-LL)
(Triangle wave 2 mV Lead II (RA-LL) Fixed at 1.1 mV (Apex-Sternum)
Amplitude Accuracy: +/- 2% (RA-LL)
+/- 10% (apex-sternum)
DEFIB WAVEFORM PLAYBACK
Time Base Expansion: 100:1 @25 mm/sec paper
speed; each division equals 40 ms
Amplitude Scaling Lead II (RA-LL): 1000 J Range: 1 mV = 3000 Volts 100 J Range: 1 mV = 900 Volts
ECG Output 1000 J Range: 0.5 V = 3000 Volts 100 J Range: 0.5 V = 900 Volts
ARRHYTHMIAS
Afib, Vfib, Atach, Vtach, Aflutter, RUN, PVC, R on T, Idioventricular
Rate Accuracy: +/- 1%
Amplitude: Fixed at 1 mV Lead II (RA-LL)
Fixed at 1.1 mV (Apex-Sternum)
Amplitude Accuracy: +/- 2 (RA-LL)
+/- 10% (Apex-Sternum)
DA2000 User's Guide
1-5
Introduction
SCOPE OUTPUTS
ECG Hi-Level: Fixed at 1 Volt
Accuracy: +/- 2%
Defib Output: Real Time
Pacer Range: 1V = 3.11 V
1000 J Range: 1V = 1450 V 100 J Range: 1V = 440 V
Amplitude Accuracy: +/-2% of scale
EXTERNAL NON-INVASIVE PACER MEASUREMENTS
Load: 50 ohms +/- 1%
(Apex-Sternum)
R-wave Amplitude: 1.1mV +/- 10% (Apex-Sternum) 1 mV +/- 2% Lead II (RA-LL)
Pulse Width: 1-50 ms
Peak Voltage: 0-12.5 Volts
Peak Current: 4-250 mA < 4 mA = 0.0 mA
Rate: 25-400 ppm < 25 ppm = 0 ppm
Refractory Period:
Sensed: 110-500 ms < 110 ms = 110 ms Pulsed: 70-500 ms < 70 ms = 70 ms
Accuracy: +/- 2% of fullscale for pulsewidth, peak
voltage & current +/- 1% of fullscale for rate and refractory period measurements
CALIBRATION SCREEN
Load: 50 ohms +/- 1% (Apex-Sternum)
Amplitude scaling: Apex (+) to Sternum (-)
Pacer Range: 318 counts/Volt 1000 J Range: 0.683 counts/Volt 100 J Range: 2.252 counts/Volt
Accuracy: +/- 15 counts
Measurement Range: Apex (+) to Sternum (-)
Pacer Range: (0-12.86) = (0-4095) 1000 J Range: (0-5995) = (0-4095) 100 J Range: (0-1814) - (0-4095)
Zero Voltage Input: 0 +/- 2 counts
PEAK / OVERSHOOT
1-6 DA2000 User's Guide
Introduction
Voltage Accuracy: 1000 J Range: +/- 10 Volts 100 J Range: +/- 25 Volts
Current Accuracy: +/- 1 Amp
RS-232 OUTPUT / COMPUTER CONTROL
Computer control allows the user to operate the DA2000 remotely via a serial RS-232 interface.
Requires a BC Biomedical RS-232 cable and a Bi-directional, D-9 connector.
Selectable Communications parameters: Baud Rate: 300, 600, 1200, 2400, 9600 Parity: None, Even, Odd Stop Bits: 1 or 2 Data Bits: 7 or 8
Display: 2-line x 24-character LCD supertwist alphanumeric
Power: One 9-Volt Alkaline (DuracellMN1604 or equivalent);
12 hours continuous operation; low battery indication; 120/240 Volts battery eliminator input.
Weight: 4.54 lbs
Dimensions: 26.67 x 24.13 x 10.16 cm
ENVIRONMENTAL OPERATING SPECS
Storage Temperature: -25 to 50°C
Operating Temperature: 0 to 40°C
Maximum Humidity: 90% Relative Humidity
DA2000 User's Guide
1-7
Accessories
Introduction
DESCRIPTION SUPPLIED
Carrying Case (optional)
User’s Guide 1
RS-232/Printer Cable
Serial/Parallel (optional)
Printer (DPM 411-21) (optional)
Printer paper for
DPM 411-21 (optional)
Converter Data Serial- Parallel 110V (optional)
Power Supply for DPU- 411-21 Printer (220 V) (optional)
Internal Paddle Adapters 2
Automatic Paddle Adapters (optional) Hewlett Packard Marquette Laerdhal Physio Control (Automatic Defibrillation) Physio Control (Pacer) Zoll Cable Asby
9 VDC 300 mA Adapter (optional)
1-8 DA2000 User's Guide
Installation & Operation
Use the following checklist when unpacking the DA2000 to check the instrument for shipping damage.
Perform a visual inspection to ensure the front panel or case has not been damaged during shipping.
DD
Check the LCD display to ensure that it is intact. If the DA2000 has been damaged, call your BC
DD
Biomedical representative immediately.
Place the DA2000 on a level surface and power up the instrument by turning the power switch to I (ON).
DD
Check the display. If the message, Warning - Low Battery!! appears, the instrument's battery requires replacement.
Operation Warnings
Section 2
The operator must be familiar with, and follow the safety precautions listed detailed below, before using the DA2000 with a defibrillator:
Inspect the defibrillator daily. Examine the paddles, lead wires, and power cord for cracks and frays.
If the defibrillator is line powered, be sure that it is plugged into a grounded receptacle. Do not touch the
electrical contact surfaces of the defibrillator paddles.
Grip one paddle handle firmly in each hand. Apply to the DA2000 plates. Keep the paddles firmly
depressed to prevent arcing that can cause injury to the operator and/or damage the DA2000 or defibrillator.
Do not touch the contact plates on the DA2000 when the defibrillator paddles are being pressed onto the
plates. Do not use any electrical paste or pads when testing a defibrillator with the DA2000.
DA 2000 User's Guide 2-1
Installation & Operation
2-2 DA 2000 User's Guide
Loading...
+ 29 hidden pages