Baxter Flo-Gard 6201 Service manual

Page 1
Service Manual
Flo-Gard@ 6201
Volumetric Infusion Pump Product Code: 2M8063
07-l g-A1 -688
Baxter Healthcare Corporation Deerfield, IL 60015, USA
0 Copyright 1992, 1995, Baxter Healthcare Corporation. All rights reserved.
Page 2
Page 3

Table of Contents

General Description
1.1 Introduction
1.2 Features .............................
1.2.1 Nurse Call ........................
1.3 Technical Specifications .......................
1.4 Controls and Indicators
1.5 Configuration Option Feature
1.5.1 Reviewing the Configuration Option Settings
1.5.2 Modifying the Configuration Option Settings
l.6Alarms .............................
1.7Alerts .............................
Hospital Service Procedures
2.1 Replacement Of Main Power Fuse
2.2 Cleaning .............................
2.3 Battery Charging ..........................
2.4 Preventive Maintenance .......................
Problem Checklist . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
Theory of Operation
4.1CPUSystem ...........................
4.1.1CPUs
4.1.2 Programmable Timer Module (PTM)
4.1.3 Watchdog Function
4.1.4 I/O Controllers
4.1.5 Multiplexer ........................
4.1.6 Universal Pulse Processor
4.1.7 Communication Controller
4.1.8 Air Sensor Circuit
4.1.9 Occlusion Sensors
4.1.10 Tube Misloading Detector ..................
4.1.11 Slide Clamp Detector
4.1.12 Battery Low Alert/Alarm Detector
4.1.13 Interlock Switch
4.1.14 Panel Lock Circuit
4.1.15 Keypad
4.1.16 Displays .........................
4.1.17MotorDriver
4.1.18 Motor Rotation Detectors ..................
............................
............................
.......................
....................
..........
.......... 1-12
.........................
...................
............................
..........................
..............
.....................
....................... 4-3
..................
..................
.....................
..................... 4-4
....................
..............
......................
..................... 4-5
.........................
.......................
l-l
l-l l-l l-2 l-3
l-4 1- 12 1- 12
1-17 1-19
2-1 2-l 2-2 2-3 2-4
4-1 4-l 4-l 4-2 4-2
4-3 4-3 4-4 4-4
4-5 4-5 4-5 4-5
4-6 4-6 4-6 4-6
7-19-Al-688
. . .
111
Page 4
4.2 DC Power Supply and Power Control Circuit . . . . . . . . . . . . . . 4-7
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . .
5.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . .
5.2 Failure Identification Codes . . . . . . . . . . . . . . . . .
5.3 Automatic Test Modes . . . . . . . . . . . . . . . . . . .
5.3.1 Automatic Test Mode 1: Calibration Mode 1 . . . . . . . . . . . 5-2
5.3.2 Automatic Test Mode 2: Calibration Mode 2 . . . . . . .
5.3.3 Automatic Test Mode 3: Manufacturing Test Mode . . . .
5.3.4 Automatic Test Mode 4: Aging Mode . . . . . . . . . .
5.3.5 Automatic Test Mode 5: Display Check Mode . . . . . . . . . . 5-6
5.3.6 Automatic Test Mode 6: Time Information Display Mode . . . . . . 5-6
5.3.7 Automatic Test Mode 7: Pumping Sensor Monitoring Mode .
5.38 Automatic Test Mode 8: Air Sensor Test Mode . . . . . .
5.3.9 Automatic Test Mode 9: Elapsed Time Test Mode . . . . .
5.3.10 Automatic Test Mode 0: Downstream Occlusion Test Mode
5.4 Failure Identification Code Troubleshooting Table . . . . . . . .
5.5 Troubleshooting Chart . . . . . . . . . . . . . . . . . . .
Disassembly and Calibration
......................... 6-1
6.1 Introduction ............................
6.2 Preparation for Maintenance
.....................
6.2.1 Tools and Test Equipment ..................
6.2.2 Recording the Configuration Option Settings
6.3 Disassembly/Reassembly
...................... 6-3
...........
6.3.1 Separation of Front and Rear Housings .............
6.3.2 Separation of Printed Circuit Boards
6.3.3 Replacement of Front Panel Film
6.3.4 Replacement of Pump Head Door Cover
6.3.5 Replacement of Pump Door Latch
6.3.6 Replacement of the Pump Door or Pump Door Assembly
6.3.7 Replacement of Door Latch Pin
6.38 Replacement of Pump Head Assembly
6.3.9 Replacement of Upstream Occlusion Sensor
..............
...............
............ 6-7
............... 6-7
...... 6-8
................
.............
.......... 6-10
6.3.10 Replacement of Downstream Occlusion Sensor Assembly
6.3.11 Air Sensor Assembly Replacement .............
6.3.12 Replacement of (FSR**) Devices for Tube Misloading Sensors .
6.3.13 Replacement of Pump Motor
6.3.14 Replacement of Safety/Slide Clamp Assembly
................ 6-13
......... 6- 15
6.3.15 Spring Retainer Removal .................
6.3.16 Spring Retainer Installation
6.3.17 Replacement of Back Plate
6.3.18 Replacement of Battery
................
................ 6-18
................. 6-18
6.3.19 Replacement of Power Supply Board ............
6.3.20 Replacement of CPU Board
**Interlink Electronics
................
. . . .
5-l
. . . . 5-l
. . . . 5-l
. . . . 5-2
. . . . 5-4
. . . . 5-5
. . . . 5-5
. . . . 5-6
. . . .
. . . .
5-6
5-7
. . . . 5-7
. . . , 5-7 . . . 5-15
6-l 6- 1 6-l 6-2
6-3 6-5 6-6
6-8 6-8
.... 6-10
6- 11
. 6- 12
6-17 6-17
6-19 6-20
iv
7-19-Al-688
Page 5
6.3.21 Replacement of Display Board
6.3.22 Replacement of Sensor Board
6.3.23 Replacement of Audible Alarm/Alert Board
6.3.24 Replacement of Backup Audible Alarm/Alert Board
6.3.25 Replacement of Terminal Board
6.3.26 Replacement of Power Transformer
6.3.27 Replacement of PANEL LOCK Switch
6.3.28 Replacement of Lithium Backup Battery
6.3.29 Tube Guide Shim Installation
6.3.30 LCD Cushion Installation
6.3.3 1 Software Installation
6.4 Calibration
6.4.1 DC Line Voltages
6.4.2 Air Sensor Calibration
6.4.3 Downstream Occlusion Sensor Calibration
6.4.4 Upstream Occlusion Sensor Calibration
6.4.5 Slide Clamp Sensor Calibration
6.4.6 Slide Shaft Sensor Calibration
6.4.7 A/D Convertor Reference Voltage Calibration
...........................
....................
...............
...............
..............
...............
.................
...................
...................
...............
................
..........
.............
...........
...........
...........
............
.........
......
6-20
6-21 6-21 6-22
6-22
6-23
6-23
6-24
6-24
6-26
6-27
6-29
6-29
6-32
6-34
6-35
6-36
6-37
6-38
Checkout
................................
7.1 Introduction
...........................
7.2 Maintenance Flowchart
7.3 Operational Checkout
.......................
7.3.1 Administration Set Placement
7.3.2 Functional Testing
7.3.3 Door Open Alarm Testing
7.3.4 Air Alarm Testing
7.3.5 Drive Defect Test/Occlusion Check
7.3.6 Downstream Occlusion Testing
7.3.7 Upstream Occlusion Testing
7.3.8 Battery Check
7.3.9 Panel Lock Test
7.3.10 Safety Clamp Test
7.3.11 AlarmVolume
7.3.12 Slide Clamp Test
7.3.13 Electrical Safety Tests
7.3.14 Accuracy Tests
Illustrated Parts Breakdown
.........................
Warranty and Service Information
9.1 Warranty Information
9.2 Service Information
9.3 General Information
........................
........................
........................
......................
................
....................
.................
....................
................
......................
.....................
....................
.....................
....................
..................
.....................
......................
.............
...............
. 7-l . 7-l
. 7-l . 7-3 . 7-3 . 7-3 . 7-5 . 7-6 . 7-7 . 7-7 . 7-8 . 7-9
7-10 7-10 7-10 7-11 7-12 7-13
8-l 9-1
9- 1 9- 1 9-2
7-i9-~1-688
V
Page 6
Diagrams . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-l
Data Sheet, Flo-Gard@ 6201
Volumetric Infusion Pump (2M8063) . . . . . . . . . . . . . . . . . . . . . A-l
Multiple Key Combinations . . . . . . . . . . . . . . . . . . . . . . . . . B-l
vi
7-19-Al-688
Page 7
Page 8
Figure lo- 13. Power Supply Control Block (P/O Sensor Board)
Figure 10-14. Occlusion Sensing Block (P/O Sensor Board) ............
........... lo- 15
lo-16
Figure 10-15. Motor Control Block (P/O Sensor Board) .............. lo-17
Figure 10-16. PPI Block #1 (LCD, Keys) (P/O Sensor Board) ........... lo- 18
Figure 10-17. PPI Block#2 (P/O Sensor Board)
................. lo-19
Figure 10-18. Serial Communication Block (P/O Sensor Board) .......... lo-20
Figure lo- 19. Additional Sensor Board Circuitry ................ 10-2 1
Figure 10-20. Display Board ......................... lo-22
Figure 10-21. LCD Drivers (P/O Display Board) ................. lo-23
Figure 10-22. Liquid Crystal Displays (P/O Display Board)
............ lo-24
Figure 10-23. LED Lamps (P/O Display Board) ................. lo-25
Figure 10-24. Audible Alarm Board ...................... lo-26
Figure 10-25. Backup Buzzer Board Figure 10-26. Terminal Board
........................ lo-28
Figure 10-27. Accessories ..........................
..................... lo-27
lo-29
Figure 10-28. Pumphead Sensors ....................... lo-30
Figure 10-29. Front Panel Key Assignments .................. 10-3 1
Figure 10-30. Power Supply
......................... lo-32
. . .
VI11
7
7-19-Al-688
Page 9
Section
General Description

1 .I Introduction

This manual provides service information for the Flo-Gard@ 6201 Single Channel Volumetric Infusion Pump (product code 2M8063) for qualified hospital biomedical engineers and service personnel. See the pump’s Operator’s Manual for detailed operating instructions.

1.2 Features

The Flo-Gard@ 6201 Single Channel Volumetric Infusion Pump is an electromechanical pump used for the intravenous infusion of liquids at user-selected rates. The pump contains a linear
peristaltic pump head which is programmable and permits infusion of primary and secondary
medication programs. The secondary program automatically switches over to the primary pro­gram when secondary infusion is complete (automatic piggybacking).
1
The pump operates on standard 115 VAC 60 Hz electrical power, or on its self-contained re-
chargeable battery. It is portable and has a panel lock-out feature to prevent patient tampering. It is designed for use with Baxter’s standard administration sets which contain an “s” as the last
character of the code number, for example, 20537s. When infusing solutions through a central line catheter, sets with a Luer lock adapter should be used. Sets with a Flashball@ device are not recommended in these applications.
The primary rate of infusion is selectable from 1 to 1,999 mL in 1 mL/hr increments and 1 .O to
99.9 mL/hr in 0.1 mL/hr increments. The secondary rate is selectable from 1 to 999 mL in 1 mL/hr increments and 1 .O to 99.9 mL/hr in 0.1 mL/hr increments. The volume to be infused (VTBI) is also selectable from 1 to 9,999 mL in 1 mL increments and 1.0 to 99.9 mL in 0.1 mL increments.
The total volumes infused from primary and secondary programs are added together and accu­mulated and can be displayed on demand. The primary and secondary VTBIs are independently decremented and displayed. Upon completion of the primary VTBI, the pump automatically switches to a keep vein open (KVO) rate. If the pump is started on a secondary rate and VTBI, the pump will change to the primary rate when the secondary program is completed. The pump may be stopped at any time by pressing the STOP key unless the pump is in the lockout mode.
7-19-Al-688
l-l
Page 10
General Description
1-2
7-19-Al-688
The Programmed Delivery Profile (PDP) enables manual programming or ramped calculation (available only on pumps running software versions 1.09 or later) of up to 10 sequential infusion programs for situations where multiple flow rates are indicated. See the appropriate Operator’s Manual for additional information.
1.2.1 Nurse Call
The Nurse Call feature is enabled by installing a jumper wire to the Terminal Printed Circuit Board at the location labeled R421 on the board. When the jumper is connected, the Nurse Call relay will be energized only during an alarm condition and when the ALARM LED is lit. The re­lay contact points (normally closed, normally open, and common) are accessible from the COM­MUNICATIONS PORT on the rear of the pump (pins 1,4, and 9, respectively). The Nurse Call feature may be used when the pump is connected to a computer, however, provisions must be made at the communications port to connect both options simultaneously. Specifications for the 9-pin D connector are listed under Technical Specifications.
Page 11
Page 12
General Description
,I Item Characteristic , / Air-in-Line Detection Factory set to NORM which causes the device to alarm on air bubbles
approximately 75 t.rL or larger. The MIN setting causes the device to alarm on bubbles approximately 50 PL or larger.
I
I Fuse
0.5 A Slo-Blo
1 Nurse Call I
g-pin D connector, Pin 1: N/C, Pin 4: N/O, Pin 9: Common
i Contact rating: 0.4A at 30 VAC, 2A at 30 VDC resistive load (internal
actrvatron required).

1.4 Controls and Indicators

All controls and indicators are shown in Figures l-l, 1-2, and l-3. Service personnel should be familiar with the pump’s features and operation before servicing the pump.
The pump’s serial number is recorded on a label on the back of the pump.
ITEM
1. Message Display
2. BACKLIGHT key
FUNCTION
Shows all messages.
Provides backlighting of the displays when the pump is used in darkened areas. The key toggles the backlighting on and off. If the pump is operating on battery power, the backlight remains on for 60 seconds and turns off automat­ically to conserve battery power.
3. SILENCE Key
Temporarily silences an audible alarm or alert for two min­utes. All visual alarm or alert information remains dis­played.
4. NEXT Legend When the NEXT legend is lit, the TOT VOL/STATUS key functions as a scrolling key to advance to the next step of a programmed delivery profile (PDP) or configuration op­tion. See Section 1.5.
1-4
7-19-Al-688
Page 13
General Description
6
/
27
26
25\
24\
23\
7
\
/
\
/
8
9
/
10
/
/
11
22
21 /
20 /
19 ’
1
/
/I2
Flo-Gard@ 6201
VOLUMETRIC INFUSION PUMP
Figure l-l. Front View of Pump
7-19-Al-688
l-5
Page 14
General DescriDtion
1-6
7-19-Al-688
5. TOT VOL/STATUS Key
6. CLEAR TOT VOL Key
7. TIME Key
8. STOP Key
9. Door Latch
Dual-function key. During operation, this key causes total volume delivered and current settings to display when pressed. It is also used to select a next step in Review Con­figuration, Modify Configuration and Programmed Deliv­ery Profile modes. The word “NEXT” above the key is illuminated when the key is functioning as a NEXT key.
Resets the total volume delivered to zero when the pump is
stopped.
Enters desired time interval for an infusion during Volume­Time or Rate-Time programming.
Stops the pump. The
STOPPED
message appears when the key is pressed. An alert will sound if pump is stopped for more than two minutes. Clears all programming alerts while pump is running.
Opens and closes pump door.
10. ON-OFF/CHARGE Key
11. Battery Icon
12. Plug Icon
13. MONITOR Legend
14. COMPUTER CONTROL Legend
Turns the pump on and off. The internal battery charger re­mains on regardless of the ON-OFF/CHARGE key as long as the pump is plugged in.
Yellow LED, always lit while the pump is operating on bat­tery power.
Green LED, always lit while the pump is plugged in and the battery is charging.
Lights for at least 2 seconds each time the host computer communicates with the pump when the pump is in monitor mode.
Flashes when the pump is initiating communications with
a host computer to enter the computer control mode. It is constantly illuminated when the pump is in computer con­trol mode.
15. SEC START Key
Starts the delivery of the secondary solution.
16. SEC VTBI Key Allows programming of the secondary VTBI.
Page 15
General Description
17. OPTIONS Key/Legend
18. SEC RATE Key
19. CLR Key
20. Numerical Keypad
21. PRI VTBI Key
22. PRI RATE Key
23. PRI START Key
24. Main Display
Allows the pump to operate in special modes, if enabled. The OPTIONS legend lights when the key is enabled.
Allows programming of the secondary infusion rate.
Clears any locked in values and programming values cur­rently being entered.
The numerical values for rate, VTBI, and time are entered with these keys.
Allows programming of the primary VTBI.
Allows programming of the primary infusion rate.
Starts the primary infusion.
Shows rate, volume to be infused (VTBI), and total vol­ume infused for primary and secondary infusion programs.
25. ALARM LED
26. PUMPING LED
27. ALERT LED
Red LED that blinks on and off during an alarm, accompa­nied by a visual message display and a repeated sequence of three beeps. An alarm indicates that the pump requires immediate attention.
Green LED which is constantly lit while pump is running.
Yellow LED which lights during alerts, accompanied by a message display and a repeated single beep. An alert indi­cates that the pump needs timely attention.
7-19-Al-688
1-7
Page 16
General Description
See Figure l-2 for the location of the following items.
1. Upstream Occlusion Sensor
2. Pump Mechanism
3. Tube Misloading Sensor
4. Downstream Occlusion Sensor
5. Air Sensor
6. Safety Clamp
7. Slide Clamp Feature
Detects a complete tubing restriction upstream of the pump.
Linear peristaltic pump mechanism.
Detects misloaded tubing out of the channel guide slots.
Detects a complete tubing restriction downstream of the pump.
Detects air bubbles in the IV set.
Prevents accidental fluid flow when the IV set is properly
loaded and the pump door is opened.
Provides an additional means of preventing accidental
gravity fluid flow by occluding the tubing in the admini-
stration set with the slide clamp before the set can be re-
moved from the pump. The use of this feature is optional;
when used, the slide clamp must be loaded in the slide
clamp slot. The feature is selectable through the pump’s
configuration options. If alert mode is enabled, the pump
will operate without the slide clamp inserted and the
SERT SLIDECLAMP
message is displayed. An alert tone
IN-
will sound to notify the user that the slide clamp should be
inserted. If the alarm mode is enabled (software versions
1.09 or later), the pump will NOT start and an alarm tone
will sound if the slide clamp is not loaded.
8. Spring Retainer Insert
9. Channel Guide Ridges
1-8
The pump is shipped from the factory with this plastic in-
sert in the slide clamp slot. It prevents damage to the
mechanism during shipment and maintains the proper
spring tension. If your hospital does not plan on using the slide clamp feature, Baxter recommends that this plastic in­sert remain in place during use. See Section 6.3.15.
Function as a guide to keep the tubing properly aligned
over the pumping fingers.
7-19-Al-688
Page 17
Page 18
General Description
The following items are located on the rear of the pump and are shown in Figure l-3.
1. IV Pole Clamp
2. Power Cord Strap
3. Audio Speakers
4. Battery Compartment
5. VOLUME Knob
6. PANEL LOCK Switch
7. Power Cord
Secures the pump to the IV pole.
Stores power cord during battery operation and pump \t(~i-­age.
For generation of audible alarm and alert tones.
Allows authorized service personnel easy access to the bat­tery, EPROMS, and battery fuse.
Adjusts loudness of audible alarm and alert tones. The
audible alarm cannot be turned completely off.
Disables all front panel keys, except BACKLIGHT and TOT VOL/STATUS, while the pump is running without alerts.
Removable only by authorized service personnel.
8. Fuse Compartment
9. COMMUNICATIONS PORT
The power cord cover must be removed to access the fuses.
The communications port contains wiring for a nurse call
jack as well as RS-232 serial communications. See the
Technical Specifications listed previously in this section.
l-10
7-19-Al-688
Page 19
1
General Description
_--
\
lo
/=\
::*.-
..-
>
/4
3
7-19-Al-688
a
/
Figure 1-3. Rear View of Pump
l-11
Page 20
General Description

1.5 Configuration Option Feature

This section describes the configuration option feature of the pump, which allows qualified per­sonnel to inspect and/or modify certain pump operating parameters to suit customer require­ments.
These parameters and their setting options are shown in Table 1- 1. The factory settings made at the time of manufacture are also shown in the table.
Note: Although the configuration option data is stored in battery backed-up
RAM, it may be lost if the main battery connector (CN302) is discon­nected from the CPU board without turning off the device. The con­figuration option data is also lost if the lithium backup battery connector (CN304) is disconnected while the main battery is discon­nected. Therefore, we advise that the configuration options be re­corded before beginning repair procedures and reset when repairs are complete.
1.5.1 Reviewing the Configuration Option Settings
To review the configuration option settings, the pump must be powered on and stopped. Press TIME and TOT VOWSTATUS simultaneously for one second. The message REVlEW
CONFlG
will be displayed in the Message Display. The option description will be displayed in the first line of the Message Display when the NEXT or SEC START key is pressed, beginning with
CLUSION.
The current setting will be displayed on the second line.
OC-
To view the next setting, press the NEXT or SEC START key. Each press of the NEXT or SEC START key will cause the pump to advance to the next setting, in the order shown in Table l-l, starting with Occlusion Alarm Level. To exit the inspection mode, press the STOP key or press TIME and TOT VOLBTATUS simultaneously.
1.5.2 Modifying the Configuration Option Settings
1. Turn off the pump.
2. While pressing the STOP key and PANEL LOCK switch, press the ON-OFF/CHARGE key. The following will occur:
a. The
MODIFY CONFIG
message appears in the Message Display.
Note: If the message
LOCKED OUT
is displayed, the configuration option set­tings can be changed only via computer control. See the Program­mer’s Manual for additional information.
1-12 7-19-Al-688
Page 21
Page 22
Page 23
Table l-l. Configurable Settings
Parameter Description Setting Options Factory
Settings
9. Maximum Rate of Infusion Sets the maximum programmable primary infusion rate of both pumps. When the maximum primary rate is 999 or higher, the maximum secondary rate is 999. At values below 999, the maximum secondary rate matches the value of the maximum primary rate.
10. Maximum VTBI
The maximum volume that the pump can be
programmed to infuse.
11. Flow Check Determines whether the flow check display will 2: ON appear during pump operation, or only when
the decimal point key and TOT VOUSTATUS
keys are pressed simultaneously.
12. Baud Rate
Determines the baud rate for normal communications between the pump and a computer. The baud rate is 9600 when the
pump is in modify configuration mode.
13. Computer Control
Determines the type of computer control option available at power up.
1 - 1999 mL/hr 1999
1 - 9999 mL 9999 mL
1: OFF
1: 300 2: 1200 3: 2400
4: 4800
5: 9600
1: Disabled 2: Off with Alarm
The pump will drop out of
remote control when an alarm occurs. 3: On with Alarm
The pump will remain in remote control when an alarm occurs.
See the Programmer’s Manual
for additional information.
mL/hr
OFF
~ 9600
Disabled
14. Hospital Area Designator Determines the hospital area designator to be
displayed upon power ON for 3 seconds. These messages can be redefined using the computer communications feature.
7-19-Al-688
.
0: no message ~ 0 1: NICU
2: PICU
3: MEDKURGICAL
4: TRAUMA/BURN UNIT
~ 5: OPER ROOM
6: CARDIACXCU
1 7: SURGICAUICU / 8: ICU
9: ONCOLOGY
1
1-15
Page 24
General Description
Table l-l. Con
Parameter Description
15. Close Clamp Determines whether or not the CLOSE CLAMP message appears with the DOOR OPEN message.
16. Insert Clamp Determines whether or not the slide clamp loading feature is enabled. The spring retainer must be removed when this feature is enabled. See Sections 6.3.15 and 6.3.16. Note: The alert or alarm option is available only on pumps running software versions
1.09 or later.
[urable Settings
Setting Options
1: OFF
2: ON
-­1: OFF (The spring retainer
should be installed.) 2: ON for software versions earlier than 1.09 or ALERT for
software versions 1.09 or later (An audible alert occurs and the message INSERT SLIDECLAMP is displayed.
The pump continues pumping.
3: ALARM (An audible alarm occurs and the message
INSERT SLIDECLAMP is displayed. The pump will not operate.)
-\
Factory Settings
ON
OFF
17. Programmed Delivery Profile Determines the memory to store the programmed delivery profile. Note: Ramp PDP option is available only on pumps running software versions 1.09 or later.
18. Time Setting
Set the real time clock in hours and minutes.
(military time 0O:OO - 2359)
19. Date Setting Set the date using the Month/Day/Year format.
-
1: Disabled 2: 5 hour memory 3: Semi-permanent memory 4: Permanent memory 5: Ramp PDP (Profile is saved
in permanent memory. If a new
PDP is entered, the pump will
revert to the PDP in permanent
memory once the cycle has
been completed.)
-
Disabled
Central Standard Time-CST
Current date-CST
1-16
7-19-Al-688
Page 25
General Description

1.6 Alarms

The pump has a number of built-in safety features. Should a situation occur which requires op­erator attention or intervention, the pump stops infusing and sounds an audible alarm. The fol­lowing are brief descriptions of these alarms.
A/R
An ultrasonic sensor in the pump head detects air in the administration set. Detection of an air bubble stops the infusion and illuminates the red ALARM LED. sage Display and the audible alarm is activated.
AIR
is displayed in the Mes-
OCCLUSlON
The downstream occlusion sensor senses an increase in back-pressure between the patient and the pump, indicating an occlusion or closed clamp. When an occlusion is sensed, the pump stops,
OCCLUSION
is displayed in the Message Display, the red ALARM LED is illumi-
nated and the audible alarm is activated.
UPSTREAM OCCLUSloN
blockage upstream of the pump. When sensed, the pump stops,
The upstream occlusion sensor senses a closed clamp or complete
UPSTREAM OCCLUSION
is dis-
played in the Message Display, the red ALARM LED is illuminated, and the audible alarm is ac-
tivated.
BATTERY LOW
tion,
BATTERY LOW
When approximately 15 minutes of running time remain during battery opera-
is displayed in the Message Display, the yellow ALERT LED is illumi-
nated, and the audible alert is activated. After approximately 15 minutes, the pump stops,
BATTERY LOW
is displayed on the Message Display, the red ALARM LED is illuminated, and
the audible alarm is activated. Table l-2 lists the alarms and the possible causes of each. In all cases, fluid infusion is halted.
See Section 5 for an explanation of how to troubleshoot the pump.
Table 1-2. Alarm Messages
Alarm Message
AIR
bCCLUSlON
j
Possible Cause
, a. Air bubble at sensor.
b. Empty fluid container. c. Improper set loading. d. A START key was pressed with no set in pump.
a. Closed distal clamp, stopcock, clogged filter, kinked tubing or other blockage downstream of the pump. b. Ambient and/or solution temperature is too low.
a. Closed clamp or other blockage upstream of the pump. b. Pinched or kinked tube loaded in the pump.
7-19-Al-688 1-17
Page 26
General Description
Table 1-2. Alarm Messages
Alarm Message
FAlLURE with code number
BATTERY LOWwith rapid three­beep alarm tone
CHECK SET/LOADING The tubing is pinched between the pump head door and base
COM TIMEOUT
EXT COMM ERROR
INSERT SLIDECLAMP
Note: This alarm option is available only on pumps running software versions 1.09 or later.
Possible Cause
Turn the pump off and back on to reset. If FAILURE does not
clear, the microprocessor has detected a pump malfunction. After recording the code number, remove the pump from use and have it serviced. See Section 5 for more detailed service instructions.
Battery power has been exhausted. Plug pump into AC outlet immediately to restore operation.
plate. Load set properly in the IV set loading path. There has been no communication between the device and
the computer for the specified time period. Check for a disconnected cable or computer problem. To clear this alarm,
press the OPTIONS key to return the device to computer control, or press SILENCE to use the pump without the computer.
The controlling computer is sending multiple queries or commands to the device without waiting for the device’s
replies. The condition has been caused by the computer, not the device. To clear the alarm, press the OPTIONS key to
return the device to computer control, or press control the pump manually. Notify the technical personnel
responsible for the computer. If the alarm recurs, disconnect the cable from the communications port. Reprogram the
pump(s) for manual operation. The slide clamp loading option is enabled and the slide clamp
is not loaded in the slide clamp slot. To clear this alarm, open
the pump door and insert the slide clamp into the slot. Close
the door.
SILENCE
to
1-18
7-19-Al-688
Page 27

1.7 Alerts

Alert messages call attention to a condition which will require operator intervention in the near future, or indicate that a procedure has been initiated which requires that the operator complete a
sequence of keystrokes. These alerts are generally cleared by the operator taking the appropriate
action. Table l-3 lists the various alerts and possible causes.
r
Alert Message STOPPED
KVO PRI VTBI = 0
NEW RATE
PRIRATE=O
LOW with audible alert SEC PROGRAM
SEC RATE = 0
SEC VTBI = 0
SEC COMPLETE
LED
Yellow
Green, Yellow
Green, Yellow
Yellow
Green, Yellow
Green,
Yellow
Yellow
Yellow
Yellow
Table 1-3. Alert Ic essages
Flow Status No flow
KVO rate None
No change until procedure is
completed
No flow
No change SEC
No flow SEC
No flow
Key
Pressed
+
None
+
PRI or SEC RATE
c
PRI or SEC
START
None No change BATTERY
c
RATE or
SEC VTBI
c
START
SEC START
None No change
Alert Condition
The pump has been left in the STOPPEI
mode for more than two minutes. Primary VTBI has been delivered and th pump has switched to 5 mL/hr KVO rate (or programmed rate, whichever is lower
Primary or Secondary flow rate is being changed while pump is running. Pump will not exit this alert condition until the appropriate START key is pressed.
The pump cannot be started without entering a primary flow rate.
Battery needs recharging. The pump will stop operating in approximately fifteen minutes or longer unless it is plugged int an AC source. Secondary (piggyback) information is being programmed into the pump while i is running in primary mode. Pump will nc exit this alert condition until SEC STAR1 is pressed.
A secondary (piggyback) infusion cannoi
be started unless a secondary flow rate i input.
A secondary (piggyback) infusion cannoi
be started unless a secondary volume to be infused has been input.
The pump has completed infusing the secondary VTBI and has switched overt the primary infusion settings. Pump will
not exit this alert condition until any
enabled key is pressed.
7-19-Al-688 1-19
Page 28
General Description
Table 1-3. Alert Messages
Alert 1 LED Flow Message Status FLOW RATE Yellow No flow
CHECK VTBI Yellow No flow PRI or Enter a VTBI within the range selected
INSERT Yellow No flow or Closed Slide clamp is not loaded into the slide SLIDECLAMP flow if door, PRI clamp slot although the set is loaded into
started or SEC the pump.
COMM TIME Yellow No change None Communication timeout period has OUT
PGM DELIV Yellow No flow PRI or An attempt was made to start the PDP ENTER PGM SEC before a program was entered. Enter a
PGM DELIV REVIEW PGM
PGM DELIV Yellow No flow CLR The CLR key was pressed after entering CLEAR ALL? the PDP mode to clear all steps. Do one
Yellow No flow j PRI or An attempt was made to start a PDP
or no flow
Key Pressed
PRI or SEC START pump cannot be started when Hior Lo is
I
SEC through the configuration option. The START pump cannot be started when Hi is
START
START , program or press OPTIONS to leave the
SEC profile prior to reviewing all the START programmed steps. Press NEXT until the
Alert Condition
Enter a rate within the range selected
I through the configuration option. The
displayed in a rate display.
displayed in a VTBI display.
elapsed. No communication has occurred
I
between the pump and the host computer , during the power-up default time period I (60 seconds) or during the time period
most recently specified by the host
computer (l-300 seconds). The computer
and the pump must maintain periodic and
successful communication in order to
avoid this timeout alert. See the
Programmer’s Manual for further
information.
j PDP mode.
first step is again displayed or press
OPTIONS to leave the PDP mode.
of the following: press CLR to erase the
profile; press NEXT to review the rest of
the profrle; press PRI RATE or PRI VTBI to modify this step; press OPTIONS to
~ leave the PDP mode. ~ ~ ~
l-20 7-19-Al-688
Page 29
_-
,-
Table 1-3. Alert Messages
Alert LED Flow Message Status Pressed
RAMP PDP ENTER DATA to be entered. Either enter the data (as Note: Alert described in the Operator’s manual, or messages or press the OPTIONS key to leave the Ramp PDP PDP mode. can only ap­pear on pumps run-
ning soft­ware ver­sions 1.09 or
later.
RAMP PDP
RATES HIGH
Note: Alert that exceeds the pump’s capabilities. To
messages or correct this alert condition, either reduce
Ramp PDP the volume for total VTBI or increase the can only ap­pear on pumps run­ning soft­ware ver­sions 1.09 or later.
RESUME Yellow No flow PRI The pump is asking if it should resume
PDP Ramp START or the Ramp PDP where it was interrupted.
Profile Phase STOP Do one of the following: Note: Alert messages or Ramp PDP can only ap­pear on pumps run­ning soft-
ware ver-
sions 1.09 or later.
Yellow No flow None
Yellow No flow None
Key
Alert Condition
The pump is waiting for Ramp PDP data
The Programmed Ramp PDP parameters
have resulted in a calculated flow rate
Run Time.
Press PRI START to start Fluid delivery
or press STOP to cancel the alert.
7-19-Al-688 1-21
Page 30
Page 31
Section 2
Hospital Service Procedures
This section contains a table describing preventive maintenance which should be performed on
the Flo-GardB 6201 Single Channel Volumetric Infusion Pump. The maintenance procedures outlined in this section may be performed in the hospital. If an abnormal condition occurs which is not correctable by performing the following procedures, remove the pump from service and troubleshoot it in accordance with Section 5, or return it to Product Service for repair.

2.1 Replacement Of Main Power Fuse

1.
Plug the pump into an AC power outlet.
Check if the Plug Icon is illuminated.
2.
3.
If it is not, replace the fuse. Remove the power plug from the AC power outlet.
4.
Remove the power cord cover on the back of the pump and unscrew the fuse caps with a small screwdriver.
Remove the fuses and check them for electrical continuity with an ohmmeter.
5.
If necessary, replace with a new fuse of the same value, type and voltage.
6.
7.
Replace and tighten the fuse caps with a screwdriver. Over tightening the fuse caps may cause the fuse holders to break.
Replace the power cord cover.
8.
9.
Perform the Electrical Safety Test to verify proper grounding impedance. See Section
7.3.13.
7-19-Al-688
2-1
Page 32
Hospital Service Procedures

2.2 Cleaning

The pump should be cleaned as soon as possible after each use to minimize the accumulation
and hardening of spilled solutions. The case and front panel may be cleaned with a soft cloth or
cotton swabs dampened with a properly diluted cleaning agent listed in Table 2- 1. Be sure to follow the manufacturer’s dilution instructions for concentrated cleaners where appli-
cable. Do not spray cleaning agents directly onto the inside of the pump door, the pump mecha­nism, and the front panel film. If these areas require cleaning, wipe carefully with a soft cloth, sparingly dampened with a cleaning agent listed in Table 2- 1. If solution spillage onto the pump­ing mechanism or front panel occurs, it should be cleaned immediately. If necessary, contact Product Service at 1-(800)-THE-PUMP.
-~
CLEANER
LpH, Septisol Cidex 7 Super Edisonite
Bafix Tor
Hi-Tor Plus
10% bleach and water Soapy water isopropyl alcohol (up to 95%)
Table 2-1. Cleaning Solutions
.-
MANUFACTURER Vestal Labs, Inc. Surgikos Inc. Edison Chemical Co. Hysan Corp. Huntington Labs. Huntington Labs.
Caution: Attempts to clean or disinfect internal parts, autoclaving or steriliza-
tion by ethylene oxide gas will damage the pump and void the war­ranty.
Caution: The following chemicals may damage the plastic front panel and tube
misloading sensors: Acetoldehyde, Acetone, Ammonia, Benzene, Hy­droxytoluene, Methylene Chloride, n-Alkyl Dimethyl Ethyl Benzyl
Ammonium Chloride, and Ozone.
For a pump that has been in an Isolation Area, select those agents from Table 2-l that both clean and disinfect. Only external parts of the pump should be disinfected. The following are proce­dures for cleaning accessible areas of the pump. Do not use hard instruments for cleaning.
1. Lift the door latch to the open position. Open the door and press the safety clamp latch until it locks in the open position.
2-2
7-19-Al-688
Page 33
Page 34
Hospital Service Procedures
2-4

2.4 Preventive Maintenance

Table 2-2 lists preventive maintenance for the pump, which should be performed at the intervals shown.
Table 2-2. Preventive Maintenance Procedures
CHECK ACTION Schedule: As required, but recommended after every use.
Pump mechanism
Case
Clean with an agent listed in Table 2-l. Clean with an agent listed in Table 2-l.
I
Rear Panel
Connector
(comm port)
Loose or missing hardware
Main Battery
Schedule: Every 12 months or as required
Back plate and
safety clamp
Pole clamp
Power cord
Clean with an agent listed in Table 2-l.
Replace the connector if its shell is damaged.
Check that plastic cover is in place.
1
Replace in accordance with Section 6.
Recharge by plugging into a 115 VAC outlet for at least 8 hours.
Check that the Plug Icon is illuminated during this time.
L
If the safety clamp or back plate does not operate smoothly, clean or replace in accordance with Section 6.3.14 or 6.3.17.
If operation is not smooth, apply one drop of high grade general purpose
machine oil to the screw threads.
Replace power cord if the pins are bent or the insulation is damaged.
Preventive Perform the appropriate tests as detailed in Section 7.3 Operational
maintenance tests 1 Checkout.
7-19-Al-688
Page 35
,-
Section 3
Problem Checklist
Table 3-l is a list of problems, checks and corrections to aid in the diagnosis of possible pump malfunctions. Corrections contained in the table can be performed without opening the pump housing. Review this list whenever a condition exists that does not appear to be normal. Perform the specified checks and corrections. If the problem cannot be corrected, remove the pump from service. Troubleshoot it in accordance with Section 5 and repair it in accordance with Section 6.
Table 3-1. Problem Checklist
PROBLEM
The plug icon is not lit when the pump is plugged in,
or the battery icon is lit when the pump is plugged in.
The pump fails to run on the
internal battery (No LCD dis-
plays appear).
The pump stops with BAT-
TERY LOW alarm.
CHECKS Check the tightness of the
power olua into the AC outlet. Check the rear power fuses
under the power cord cap. Check the AC outlet for proper
voltage.
Check the line cord for
continuity.
After recharging the battery for 24 hours with the pump turned off, check the battery charging voltage, MB, per Section 53.1.
No check required.
CORRECTIVE ACTION
Press the power plug firmly into
the grounded AC outlet.
Replace the fuse(s) if it has
failed and recharge the battery.
If the voltage is below 105 VAC, connect the pump to the correct supply voltage.
Connect the power terminals of the power plug to an ohmmeter. The ohmmeter should indicate continuity.
If the battery charge voltage is normal and the problem still persists, replace the battery. See Section 6.3.18.
Recharge the battery.
7-19-Al-688
3-1
Page 36
Problem Checklist
1
PROBLEM The pump door will not open
or close smoothly.
The audible alarm volume is not loud enough.
The interval between audible alarm tones is too long.
The backlight is off when the pump is running on internal battery power.
A RATE, VTBI, or START key is not accepted by the
pump Hi or Lo is displayed during
Volume-Time programming.
A FLOW RATE or CHECK
VTBl alarm occurs when a
START kev is pressed.
Table 3-1. Problem Checklist
CHECKS
Check the positioning and seating of the administration set tubing and the slide clamp.
Check the administration set for type and code.
Check for solution spills (liquids or residues).
Check that the door latch roller pin turns smoothly.
­Check for possible damage to
the door latch, latch pin roller, or
door hinge.
r
Check if the front panel is locked (Lot appears in Main Display).
L
Calculate the rate and verify that it is within the allowable range set by the configuration option.
Check that the rate or VTBI are within the limits set by the configuration option.
CORRECTIVE ACTION
Position the tubing and the slide clamp properly and make certain they are seated in the
guides.
Replace with a Baxter’s “s”
suffix administration set if
required.
Clean all accessible areas with cotton swabs dampened with one of the cleaners listed in
Section 2. Remove fibers or foreign particles. Do not use
hard instruments for cleaning. Clean the roller with an
approved cleaner. See Section 6.3.5 for
instructions on replacing the door latch and Section 6.3.7 for
replacing the latch pin.
Replace door as described in Section 6.3.6.
Turn the VOLUME knob on the
rear of the pump clockwise until the desired volume is obtained.
Change the interval for alert and/or alarm tones to the desired value through the configuration option.
Press the BACKLIGHT key as long as required to view the pump settings.
Press the PANEL LOCK switch
to remove the panel lock.
Enter a rate within the range set by the configuration option or
change the maximum rate setting in the configuration option if appropriate.
Change the maximum rate and/or maximum VTBI setting through the configuration option.
3-2 7-19-Al-688
Page 37
PROBLEM
An AIR alarm occurs with no
air in the tubing or with the pump door closed and the START key pressed.
An OCCLUSION alarm or
an UPSTREAM
OCCLUSION alarm occurs
with the pump door closed
and the START key pressed.
An INSERT SLIDECLAMP
alert or alarm occurs when
the pump door is closed.
Note: The alarm option is available only on pumps running software versions
1.09 or later.
Table 3-1. Problem Checklist
CHECKS Check the positioning and
seating of the tubing. Check the tubing for surface
scratches and for tube
roundness.
Check the administration set for type and code.
Check for solution spills (liquids
or residues).
Check the positioning and seating of the tubing.
Check that there are no
obstructions upstream or downstream of the pump.
Check the administration set for
type and code.
Check that ambient and solution
temperatures are above 60” F.
Check for solution spills (liquids
or residues) on the inside of the
door and/or on the baseplate.
Check that the slide clamp is in
the slide clamp slot.
Check for solution spills (liquids or residues) on the slide clamp or safety/slide clamp assembly.
Check if the administration set is equipped with the slide clamp designed for use with this pump. See the instructions accompanying the administration set.
Problem Checklist
CORRECTIVE ACTION Position the tubing fully into the
air sensor. Replace or reposition the tubing
if surface scratches are significant or if the tubing has become flattened or oval in shape.
Replace with Baxter’s “s” suffix administration set.
Clean the sensor with cotton swabs dampened with one of
the agents listed in Table 2-l. Remove fibers or foreign particles. Do not use hard instruments for cleaning.
Position the tubing properly into
the sensor and safety clamp.
Correct any pinched or kinked tubing in the pump.
Remove obstructions and/or
open the roller clamp.
Replace with Baxter’s “s” suffix
administration set if required.
Raise ambient and/or solution
temperatures. Clean all accessible areas with
cotton swabs dampened with one of the cleaning agents listed
in Table 2-1. Remove fibers or foreign particles. Do not use hard instruments for cleaning.
Push the slide clamp all the way
into the slide clamp slot. Clean the safety/slide clamp
assembly.
If not, use an administration set which has the proper slide clamp for use with the pump.
Inset-t the slide clamp into the slide clamp slot before closing the pump door.
7-19-Al-688
3-3
Page 38
Problem Checklist
PROBLEM
The safety clamp will not
latch open.
A CHECK SET/LOADING alarm occurs when the pump door is closed.
Table 3-1. Problem Checklist
CHECKS Check the positioning and Push the slide clamp all the way
seating of the slide clamp in the into the slide clamp slot. slide clamp slot.
Check if the administration set is If not, use an administration set equipped with the slide clamp which has the compatible slide designed for use with this pump. clamp. Insert the slide clamp See the instructions into the slide clamp slot before accompanying the closing the pump door. administration set.
Make sure the safety clamp arm cover is in the full open position.
Check for solution spills (liquids or residues).
Open the door and check the position of the tubing in the guide channel.
CORRECTIVE ACTION
Exercise the safety clamp by opening and closing it several times.
Clean with cotton swabs dampened with one of the cleaning agents listed in Table 2-l. Remove fibers or foreign
particles. Do not use hard
instruments for cleaning.
Load the set properly in the guide channel.
3-4 7-19-Al-688
Page 39
Section
4
Theory of Operation
This section covers the operating principles of the pump. The theory of operation does not cover the specific circuitry in great detail, but provides general information needed to perform fault isolation. Active-low signals on all schematic diagrams in Section 10 are denoted by an ex­clamation point (!) preceding the signal name. Figure lo- 1 is a block diagram of the major components in the pump. The numbers at the upper left of each block refer to the number of the figure in Section 10 in which the major components of that block are shown in greater detail.

4.1 CPU System

4.1.1 CPUS See Figures 10-6 and 10-7. The pump uses two identical CPUs, UlOl and UOOl. Normally,
UlOl and UOOl act as master and slave, respectively. The master CPU controls all pump functions except motor control, which is handled by the
slave CPU. The master CPU sends rate information and motor start/stop messages to the slave
CPU and also monitors the motor control by the slave CPU. The master CPU gathers data from and/or controls the interlock switch, the power control cir-
cuit, the communication controller, two I/O controllers, the occlusion detection multiplexer, the RAM, the real time clock, the air sensor, the universal pulse processor and the alarm control cir­cuit.
The master CPU also handles RS-232 serial communication with an external computer through the communication controller.
The slave CPU controls the pump motor via the motor control circuit in three ways:
l
Generates pulses to rotate the motor.
l
Monitors motor skip step by checking the signal from the motor rotation detector.
-
l
Controls the motor current while minimizing current draw from the battery. It also controls the alarm control circuit.
7-19-Al-688
4-l
Page 40
Theory of Operation
The slave CPU outputs an interrupt signal to the master CPU through the universal pulse proces-
sor after every one-eighth of the liveband period to provide air bubble detection timing to the
master CPU. Both CPUs handle the watchdog function, which is the periodic communication between the
CPUs through two serial communication lines at 15,625 baud. The CPUs, in conjunction with the two 64K x 8 EPROMs, utilize 16 address lines and eight data
lines. The master CPU addresses an EPROM, RAM, Real Time Clock, universal pulse processor, two
I/O controllers and the communication controller. The slave CPU addresses an EPROM, RAM and programmable timer module. The software in the EPROMs for master and slave CPUs is
different.
4.1.2 Programmable Timer Module (PTM) See Figure 10-10. The programmable timer module (PTM) divides the 8 MHz system clock into
500 kHz for the oscillation of the air sensor and also generates a signal for pulse width modula­tion control of the motor driver.
The slave CPU calculates and outputs motor drive signals based on the rate information from the master CPU. It also sets a motor current level in the PTM from a reference table.
4.1.3 Watchdog Function The watchdog function is performed in two ways:
Both CPUs monitor each other’s status. The purpose of this watchdog is to detect a malfunction of either microprocessor and stop the pump with an alarm. See Figures 10-7 and 10-9. Both
CPUs communicate through the two serial communication lines, TX and Rx. Each CPU has a communication counter, which is initialized to a predetermined value by a signal from the other CPU. The counter is then decremented by one count every 32.768 mS. The counters are nor-
mally initialized again by the signal from the other CPU before they decrement to zero. If a counter reaches zero, it indicates that the watchdog signal from the other CPU was never re­ceived. This indicates a problem with the other CPU. The remaining functional CPU then stops the pump with visual and audible alarms.
If communication between the CPUs cannot occur, both CPUs stop the pump with visual and audible alarms.
Should both CPUs fail at the same time, this watchdog function does not work. The alarm con­trol circuit is provided as a backup watchdog function. See Figure 10-9. Signals from each CPU are the inputs to the alarm control circuit.
When both CPUs are functioning normally, the signals change their state periodically. The soft-
ware to control the signals is divided into several parts and located in different portions of the
4-2 7-19-Al-688
Page 41
main program. The state changes of the signals are considered normal only when all the individ­ual parts of the program are executed according to an expected sequence.
If either or both signals fail, the alarm control circuit is triggered and stops the pump with visual
and audible alarms.
The accompanying audible tone whenever either watchdog function is activated is continuous rather than intermittent.
4.1.4 I/O Controllers See Figures lo-16 and 10-17. The I/O controller U601 performs the following functions: activat-
ing backlight, addressing the keyboard and scanning the ON-OFF/CHARGE key, PANEL LOCK switch and LCD drivers, and writing display data from the master CPU into the display
drivers.
The other I/O controller, U602, performs the following functions: controlling the air and occlu-
sion sensors, and activating all LEDs and icons except ALARM and OPTIONS LEDs and key
beep. It also transfers the slide clamp sensor signals to the master CPU.
4.1.5 Multiplexer See Figure lo- 14. The multiplexer, U85 1, selects one of the two occlusion sensor outputs in ac-
cordance with the address signals from an I/O controller, and sends it to the master CPU.
4.1.6 Universal Pulse Processor
See Figure 10-8. The universal pulse processor (UPP) is controlled by the master CPU and con­verts the following analog signals into digital signals: air sensor outputs, tube misloading sensor outputs, battery voltages, motor currents and the voltage of the CPUs. The digital signals are pe­riodically read by the master CPU.
The UPP generates 2 kHz and 4 kHz signals for the audible alarm and key beep and a 17 kHz signal for the door open sensor, and the signal for backlight dimming.
The UPP also interrupts the master CPU each time a pulse is received from the motor rotation detector, when an interrupt signal from slave CPU is received or when the pump communicates with an external PC.
The UPP is used to select the baud rate for external communications. The baud rate is set at power-up according to the configuration setting.
7-19-Al-688
4-3
Page 42
Page 43
4.1 .I 0 Tube Misloading Detector
See Figures 10-8 and 10-28. The pump head has a Force Sensing Resistor* device (FSR**) at-
tached to each side of the tube loading channel. If the tube is misloaded over the FSR, its resis­tance decreases. The two FSR output voltages are converted into digital signals by the UPP and monitored by the master CPU. The CPU activates an alarm when the resistance decreases below
a specified level.
4.1 .I 1 Slide Clamp Detector
See Figures lo-17 and 10-28. The slide clamp sensor consists of two opto-interrupters contained in the safety/slide clamp assembly. If the slide clamp option is selected via the configuration op­tion settings, the administration set slide clamp should be loaded into the slide clamp slot in or­der to avoid an alert or alarm condition. If the slide clamp is not loaded into the slide clamp slot, or is loaded improperly, the opto-interrupters do not receive reflection signals. The interrupter output voltages are read and monitored by the master CPU via the I/O controller. The CPU acti­vates an alert or an alarm (software versions 1.09 or later) when the slide clamp is not loaded and the option has been selected.
4.1 .I2 Battery Low Alert/Alarm Detector
See Figure 10-8. The battery voltage is converted into a digital signal by the UPP and monitored
by the master CPU. The CPU activates the alert or alarm if the battery charge state falls too low. The
BATTERY
approximately 15 minutes of operation. The
LOW alert is triggered if the battery voltage drops below 11.4 V, which permits
BATTERY LOW
alarm is triggered when battery voltage drops below 10.4 VDC, which stops the pump with an alarm to prevent the battery from being damaged.
4.1 .I 3 Interlock Switch
See Figure 10-28. The interlock switch is a reed type, activated by a magnet attached to pump door latch. The switch opens the circuit when the pump door is opened. The master CPU moni-
tors the interlock switch, activates the
DOOR OPEN
alarm and stops the pump when the door is
opened.
4.1 .I 4 Panel Lock Circuit
See Figure 10-16. The panel lock circuit is initiated by the PANEL LOCK pushbutton switch lo­cated on the rear of the pump. The switch is connected to an I/O controller. The purpose of this circuit is to prevent patient tampering. After the PANEL LOCK switch has been pressed, the message
Lot
is displayed in the unused rate window and no keys except TOT VOLBTATUS
* Interlink Electronics **Interlink Electronics
7-19-Al-688
4-5
Page 44
Page 45
Theory of Operation

4.2 DC Power Supply and Power Control Circuit

See Figure 10-30. The DC power supply circuit is a switching type, which provides a regulated
13.65 V output at 0.85A maximum. The circuit will charge the battery as long as the pump is
connected to a specified AC outlet. The design continues to supply useful output down to a
105.0 VAC input. Two 500 mA slow blow replacement fuses are required for AC. A 2.OA fuse is required in the battery circuit. Whenever the supply is activated by AC voltage, the Plug Icon on the front panel is lit. The DC power supply has over-voltage and over-current protection.
The power supply circuit on the CPU board (Figure 10-13) generates several DC sources from the 13.65 V output of the DC power supply board, (when the pump is plugged into a specified AC source) or from the internal battery. The +5 VDC lines are shown below. See Section 6.4.1 for voltage specification limits.
Vcpu: This line is the output of a DC to DC convertor, HIC301, and is switched by the
master CPU. Vcpu is the main DC source voltage of the pump. Motor current does not flow while VcpU is off.
Vref: This is the reference voltage regulated by a filter and a reference diode (U502:
LM336M-5.0) for the A/D convertor in the UPP. This voltage is generated from Voth, which is switched by the master CPU.
Vkey: This is generated either from the output of the DC power supply board, or from
the internal battery voltage and is used for monitoring the ON-OFF/CHARGE key. This is the unswitched power source.
Other DC voltages used in the pump are as follows:
Vmtr: This is the DC source for the motor. It comes either from the DC power supply
output or the internal battery voltage. This is the unswitched power source.
Voth: This is used for backlights, icons and optional nurse call feature. The voltage
comes from the DC power supply output and is switched by the master CPU.
Vmem: This is the output of the lithium backup battery for the backup memory in the
RAM (U507) and for maintaining the time in real time clock (U506). This is the
unswitched power source. All the voltages except unswitched power sources are turned on when the ON-OFF/CHARGE key is pressed while the pump is off. However, turning these voltages off can only be accom­plished by the master CPU through the power control circuit.
The master CPU saves necessary data in the back up memory in the RAM (U507) and turns off the Vcpu, Vref and Voth voltages when the pump is turned off.
7-19-Al-688
4-7
Page 46
‘I heory of Operation
The power control circuit is contained in a hybrid chip, HIC302, which also includes a circuit to activate the backup audible alarm when the pump is shut down by a failure. The RAM (UO06) f~>r the slave CPU is not backed up.
-.
4-8
7-19-Al-688
Page 47
Page 48
Page 49
5. Check that the value displayed in the PRI VTBI window of the Main Display is between 2967 and 3039. No calibration is required if the values are within the specification.
If the value is out of specification, perform Section 6.4.3, steps 2 and 3.
6.
7. Place the thickness gauge used in step 2 onto the upstream occlusion sensor.
8. Close the pump door.
9. Check that the value displayed in the PRI RATE window of the Main Display is between 3242 and 3314. Remove the thickness gauge, close the door, and check that the value is 3080 or less.
10. If the value is out of specification, perform Section 6.4.4, steps 2 and 3.
11. Check that the voltage readings in the SEC RATE and SEC VTBI windows are within specification:
Note: The pump must be plugged in to obtain voltage readings within the
specifications below.
/--
Descriptor Window Displayed
Vmain = MB Vme,,, = BB SEC VTBI
SEC RATE
Specification
688-714 574-737
(a)
(b)
Note: (a) Actual voltage = Displayed data x 0.01953
(b) Actual voltage = Displayed data x 0.00488.
12. If the BB display is out of specification, replace the lithium backup battery per Section
6.3.28.
13. MB is a value relating to the voltage of the charging circuit. If the MB display is out of specification, turn the pump off and charge it for 8 hours. Repeat steps 1 and 11. If the MB display is still out of specification, perform the following procedure.
14. Remove the battery cover by removing 4 screws (Figure 8-4 item 27) on the battery cover located on the back of the pump. Disconnect the main battery at the battery termi­nals and perform steps 1 and 11.
15. If the MB display is out of specification, the DC power supply must be calibrated per
6.4.1.
7-19-Al-688
5-3
Page 50
Page 51
,-
10. Open the door and intentionally load the tubing over the tube misloading sensor 2L (Fig­ure 8-6, item 10). Close the door. Do not load hard components on the administration set between the door and the base plate. The FSRL value (at SEC VTBI window)
should display 450 or less.
11. Repeat above step on the tube misloading sensor 2R (Figure 8-6, item 11). The FSRR value (at SEC RATE window) should display 450 or less.
Note: The sensors are normal if the values in steps 8 through 11 are within
specification. If any of these values are out of specification, replace the tube misloading sensor per Section 6.3.12.
Press the TIME key and check that the voltage Vcpu reading in the SEC RATE window
12. of the display is within 472 - 579. Actual voltage = displayed data x 0.00976.
5.3.3 Automatic Test Mode 3: Manufacturing Test Mode
Warning: This is an automatic test mode for testing the pump during manufac-
turing only. This mode must not be used on patients because the downstream occlusion, upstream occlusion and air sensors are dis­abled.
The initial display in this automatic mode is for a PRI RATE of 1999 mL/hr, PRI VTBI of 3998 mL, SEC RATE of 500 m/hr, and SEC VTBI of 100 mL. Pressing the TIME key changes the primary settings as follows:
PRI RATE PRI VTBI SEC RATE SEC VTBI
999 1998 499 998
250 ..--.-
125 250 500
10 20 500 100
500 _ 500 100
In each combination, the primary and secondary rates and VTBI can be manually altered.
500
500
100 100
100
The pump will stop automatically one hour after the PRI START key was pressed. While the pump is running, the air sensor and the upstream and downstream occlusion sensors are disabled.
5.3.4 Automatic Test Mode 4: Aging Mode
Warning: This is an Automatic Test Mode for testing the pump during manu-
facturing only. This mode must not be used on patients because the rates and VTBI settings cannot be altered.
While the pump is stopped, pressing any numbered key (0 - 6) programs a preset rateNTB1 com-
bination as follows:
7-19-Al-688
5-S
Page 52
Page 53
Troubleshooting
sage display when the pump is started. All other keys function normally except for the TIME
key. Rate and VTBI information for both primary and secondary mode may be changed.
5.3.9 Automatic Test Mode 9: Elapsed Time Test Mode The elapsed pumping time is displayed in the first line of the message display in this test mode.
Momentarily pressing the TIME key selects two pre-programmed primary mode test conditions. The first is a PRI RATE of 125 mL/hr with a VTBI of 100 mL. The second is a PRI RATE of 499 mL/hr with a VTBI of 100 mL. The elapsed time counts only when the pump is running. Stopping the pump and then restarting it will reset the timer to zero. All keys except the TIME key work normally. When the VTBI has been delivered, the pump will enter KVO mode.
53.10 Automatic Test Mode 0: Downstream Occlusion Test Mode In this mode, pressing the TIME key will allow you to select the pump’s downstream occlusion
level sensitivity without entering the configuration mode. The selected occlusion sensor level (1, 2, or 3) is momentarily shown in the second line of the message display until the test mode is ex­ited. After exiting test mode, the downstream occlusion sensor level will revert back to the origi­nal configuration setting. Rate and VTBI information may be programmed manually.

5.4 Failure Identification Code Troubleshooting Table

Use Table 5- 1 to determine the corrective action necessary to resolve failure codes. The causes of each failure code are listed in the order in which they are most likely to occur. Perform the corrective action items in the order in which they are listed and retest the pump after each ac­tion. All replacement procedures are contained in Section 6.
There are two types of Failure Identification Codes: Alarm (codes 1-19) and Failure (codes 20
and above). Alarm codes report malfunctions generally correctable by the operator. Failure codes may require servicing of the pump. A code of 0 indicates that less than 10 alarms/failures have occurred since the log was cleared. The time and date will also be 0.
-
7-19-Al-688 5-7
Page 54
Table 5-l. Failure Identification Codes
CODE CAUSE CHECKS 0
1 Air bubble detected in a Check for air in tubing. Ensure set Expel air in tubing.
2 Downstream occlusion
3 Downstream occlusion
4 Upstream occlusion was
No alarms or failures. No action necessary.
1 Run mode. is correct type and is properly Replace or reposition set
loaded. Ensure set tubing is not if tubing is flattened or scratched or deformed. scratched. Clean air
Perform Section 5.3.2 steps 1 If values are not within
through 5. specification, recalibrate
Check for scratched or deformed Remove downstream was detected: The sensor output increased by a specified value.
was detected: The sensor output increased to an absolute value.
~ detected: The sensor
output decreased by a specified value.
tubing. Verify that solution and occlusion and reposition
ambient temperatures are between the tubing. Clean
60” and 100” F (15.5” and 37.7” occlusion sensors in
C). Ensure set is correct type and accordance with Section
is properly loaded. Check for 2.2 if dirty. Raise or lower
spilled solution in sensor region. j temperatures if
Perform Section 5.3.1, steps 1
through 6.
i See Failure Identification Code 2.
Check for spilled solution in sensor Remove upstream
region. Ensure set is proper type occlusion and reposition
and is properly loaded.
/
Ensure ambient and solution Raise or lower
temperatures are between 60” and temperatures.
100” F (15.5” and 37.7” C). Perform the appropriate steps from Reading displayed in
Section 5.3.1, related to upstream pump’s PRI RATE
CORRECTIVE ACTION
sensor in accordance with Section 2 if dirty.
or replace the air sensor.
necessary.
tubing. Clean the sensor with cotton swabs
dampened with one of the agents listed in Table
i 2-l. Remove fibers or i foreign particles. Do
not use hard
instruments for
cleaning.
5-8
7-19-Al-688
Page 55
Troubleshooting
-
r
CODE 5
CAUSE
Upstream occlusion was
Table 5-1. Failure Identification Codes
CHECKS
I
See Failure Identification Code 4.
L
CORRECTIVE ACTION
detected: The sensor output decreased to an
absolute value.
i + +
6
Low battery voltage (10.4
V or less) was detected.
If BATTERY LOWalarm is on and CHARGING LED is off, check: a. tightness of AC plug
b. AC voltage source
a. Plug in firmly. b. If AC voltage is below
105 VAC, connect pump to a correct voltage source.
c. power fuse d. battery fuse
Turn pump off and recharge for 8
hours.
c. Replace if it has failed. d. Replace if it has failed.
L
If alarm recurs after recharging, check and
repair loose connections at battery terminals. If problem persists, replace
c
7
Communication alarm was detected: A framing error, overrun error or parity error occurs during the communication with
Check that the pump is in computer control mode. Check for the loss of the communications
link. Verify that computer software is correctly loaded onto PC.
an external computer.
battery.
Place pump into computer control mode.
Ensure communication cable is plugged in and/or the cable is wired correctly. Correct problems with
PC software.
L
8
Communication time out alarm was detected: No
communication between the pump and an external
PC for a given period
See Failure Identification Code 7.
Place pump into computer control mode.
Ensure communication cable is plugged in and/or the cable is wired correctly.
Correct problems with
I L
9
Slide Clamp Alarm
(software versions 1.09
or later).
Slide clamp is not inserted or is improperly inserted in the slide clamp slot.
PC software.
Inset-t slide clamp
properly. Clean the slide clamp sensor with a cotton tipped swab dampened with water. If alarm recurs after
cleaning the sensor,
perform the slide clamp test in Section 7.
10 Misloaded tubing was
detected.
’ Check set loading in the pump
mechanism. Check any damage on the tube misloading sensors.
Load the tubing in the guides and channels correctly. Repair or
replace the FSR(s)* .
*Interlink Electronics
7-19-Al-688 5-9
Page 56
CODE
12
20
21
23
24
27
28
Table 5
CAUSE CHECKS CORRECTIVE ACTION
t
Upstream occlusion was detected in Revolution Counter Interrupt.
­Malfunction in Door Open detection circuitry: P54 of
the Master CPU remains
Malfunction in frequency converter circuitry for occlusion detection: P20 of master CPU remains high or low.
Malfunction in frequency converter circuitry for occlusion detection: P20 of master CPU remains high or low.
P50 (IRQ interrupt port) remains high.
UP and DOWN occlusion values agree.
Malfunction in slide Check battery voltage in clamp detection circuit. accordance with Section 5.3.1.
Two air sensor outputs (NORM and MIN) cannot be selected.
Two occlusion sensors (UP and DOWN) cannot be selected.
Malfunction in A/D converter circuitry for air detection: A/D conversion
did not complete in 50 @.
. Failure Identification Codes
See Failure Identification Code 4.
Check that magnet in pump door
latch is present.
Check connectors CN851 and CNBUSlENBUS2 for proper
connection and continuity:
See Failure Identification Code 21.
Check UlOl pin 10 for a high (+5 VDC) signal.
Check connector CNBUSl/ CNBUS2 for proper connection and continuity.
Check connector CN803 for proper connection.
Check connectors CN811 and CNBUSlKNBUS2 for proper connection and continuity.
Check tightness of connectors CN851 and CNBUSlENBUS2 for proper connection and continuity.
Turn pump off and back on.
Replace pump door latch
if magnet is missing.
Repair or replace CPU
board.
Repair or replace
connectors as necessary.
Replace CPU board. Replace Sensor board.
If signal is always high, replace CPU board.
Repair or replace CNBUSl/ CNBUS2 if necessary. Replace CPU board. Reolace Sensor board.
Clean contacts. Charge or replace the battery. Repair CN803 if necessary.
Replace safety/slide
clamp assembly.
Repair or replace CN811 or CNBUSlENBUS2 if necessary.
Replace CPU board.
Replace Sensor board.
Repair or replace CN851 or CNBUSlKNBUS2 if necessary. Replace CPU board.
Replace Sensor board.
If failure recurs, replace CPU board.
II
s-10 7-19-Al-688
Page 57
Troubleshooting
-
_-
Table 5-l. Failure Identification Codes
CODE CAUSE
31 Malfunction in air MIN
’ detection circuitry: Input
to A/D converter remains Replace Sensor board. high.
32 Malfunction in air NORM Turn pump off and back on.
detection circuitry: Input
to A/D converter remains
high.
33
34 Malfunction in air NORM Check tightness of connectors on Replace CPU board.
35 Front panel key was Check that key was not
37
38
41 Abnormal operation of Turn pump off and back on.
42 ~ Malfunction in alarm Turn pump off and back on.
46---
47 , The PWRONl signal is Turn pump off and back on. 1 If failure recurs, replace
48 Malfunction in real time j Turn pump off and back on. Replace U506 or CPU
Malfunction in air MIN detection circuitry: Input the CPU board. Replace Sensor board.
to A/D converter is not
0 V although the MIN air
transducer is not oscillating.
detection circuitry: Input the CPU board. Replace Sensor board.
, to A/D converter is not
0 V although the NORM air transducer is not oscillating.
1 pressed for more than 40 inadvertently pressed for 40
seconds. seconds. Check that front panel is
ON-OFF/CHARGE key Check that the key was not Replace front panel if
was pressed for more than 5 seconds when
pump power was off. Malfunction in tube Check connector CN803 for proper Clean contacts on mak
misloading detection connection. connector.
circuitry. Repair CN803 if
master CPU. CPU board.
control circuit.
i The PWRONl signal is
held low. CPU board.
, held high. ) CPU board.
clock: RTC does not count time correctly.
CHECKS CORRECTIVE ACT101 Turn pump off and back on. If failure recurs, replace
CPU board.
If failure recurs, replace CPU board. Replace Sensor board.
Check tightness of connectors on
not damaged.
/ inadvertently pressed for 5 necessary.
seconds. Check that key is not damaged.
Turn pump off and back on.
Replace CPU board.
Turn the pump off and back on. If failure code recurs, replace front panel.
necessary. Replace the FSR*.
If failure recurs, replace
If failure recurs, replace
CPU board.
If failure recurs, replace
board.
* Interlink Electronics
7-19-Al-688 5-11
Page 58
Troubleshooting
Table 5-l. Failure Identification Codes
CODE CAUSE CHECKS CORRECTIVE ACTlOh 49 Malfunction in real time Turn pump off and back on. If failure recurs, charge
clock: Abnormal RTC Check backup battery voltage in replace the battery. data. Section 53.1.
Reset all configuration option Replace U506 or CPU settings in Section 1.5.2 board.
50 Master CPU received a Turn pump off and back on. If failure recurs, replace
TRAP interrupt. the software (EPROMs)
Replace CPU board.
51 Master CPU received a Turn pump off and back on. If failure recurs, replace
CMI interrupt. CPU board.
52 Master CPU received a
SWI interrupt. CPU board.
53 Master CPU received a Turn pump off and back on. If failure recurs, replace
NMI interrupt. CPU board.
54 Master CPU received an Turn pump off and back on. If failure recurs, replace
OCI interrupt. CPU board.
55 Master CPU received an Turn pump off and back on. If failure recurs, replace
IRQ2 interrupt. CPU board.
58 Master CPU’s TIMER 1 Turn pump off and back on. If failure recurs, replace
interrupt does not occur. CPU board.
59 Program stuck in block 1 Turn pump off and back on. If failure recurs, replace
routine. CPU board.
60 Malfunction in battery Turn pump off and back on.
voltage detection CPU board. circuitry: Input to A/D converter remains high.
61 Malfunction in battery Turn pump off and back on. If failure recurs, replace
voltage detection CPU board. circuitry: Input to A/D converter remains low.
62
63
64
Malfunction in memory Turn pump off and back on. If failure recurs, replace back up battery voltage CPU board. replace detection circuitry: Input Sensor board. to A/D converter remains high.
Lithium backup battery Check connector CN304 for proper Plug CN304 in firmly. voltage is too low to back connection and continuity. Replace CPU board. up the memories. Replace Sensor board.
Malfunction in A/D converter circuitry. A/D conversion for 0 through 3 channels did not complete in 144 $S.
Turn pump off and back on. If failure recurs, replace
If failure recurs, replace
Replace lithium backup battery.
Turn pump off and back on. If failure recurs, replace
CPU board. Replace
Sensor board.
5-12
7-19-Al-688
Page 59
Troubleshooting
Table 5-l. Failure Identification Codes
CODE CAUSE CHECKS 65
66
67 Supply voltage of C Turn pump off and back on. If failure recurs, replace
68
69 Abnormal battery Check calibration of DC Power Recalibrate the board. If
70
71 Slave CPU received no Turn pump off and back on. If failure recurs, replace
72 ’ Slave CPU received Turn pump off and back on.
73 Overcurrent to motor. ’ Check for loose motor coupler.
Malfunction in A/D Turn pump off and back on. If failure recurs, replace converter circuitry. A/D CPU board. Replace conversion for 4 through Sensor board. 7 channels did not
complete in 144 us. Supply voltage of CPU Turn pump off and back on. If failure recurs, replace
circuitry is too high. Sensor board.
circuitry is too low. Sensor board.
Malfunction in A/D Turn pump off and back on. i If failure recurs, replace converter circuitry. A/D CPU board. conversion for channels 8 Replace Sensor board. and 9 did not complete in
72 j.6.
recharging voltage. Supply board. failure recurs, replace
Data in duplicate memory Turn pump off and back on. If failure recurs, replace areas does not agree.
I data from Master CPU. CPU board.
undefined data 4 times from Master CPU.
Check that connector CN201 is seated firmly.
Check that connector CN201 is
: seated firmly. Check for blown fuse
CORRECTIVE ACTION
U501 or CPU board.
CPU board.
If failure recurs, replace
CPU board.
Apply a trace of Loctite* 211 or 222 to coupling
) set screws and tighten
them to 5 kgf-cm (4.3 in-
lb). Plug CN201 firmly. Replace Sensor board. Replace CPU board.
~ g;;;;;;d. 1; ~ zTrno ~ ::“~umpo~andbackon~
revolution interrupt while Ensure that the encoder disk is
motor was rotating. properly secured.
Check photointerrupter outputs for Replace proper function. photointerrupters if
necessary. Replace Sensor board. Replace CPU board.
*Loctite, Inc.
7-19-Al-688 5-13
pi
Page 60
Troubleshooting
Table 5
CODE 78 See Failure Identification Code 77
79 Ensure that pump head is clean
82 -.-
83
86
87 LVDT reading of Check calibration of upstream
90
91
92
93 Write data and read data Turn pump off and back on.
94
CAUSE
­Drive defect -
overinfusion. High motor
rotation against the number of revolution
counter interrupts.
Drive defect ­underinfusion. Low motor rotation against the
i number of revolution ) counter interrupts.
No acknowledgment
CPU for three communication trials.
+
Undefined message type
received from Slave CPU. LVDT reading of
downstream occlusion occlusion sensors. sensor is too high.
upstream occlusion occlusion sensors.
( sensor is too high.
RATE, VTBI and VI data checksum is not 0.
L
Checksum error of Programmed Delivery Profile stored data.
Checksum error of
T
Programmed Delivery Profile working data.
of M5M5165 do not agree.
-+
Configuration option Reset all of the configuration settings checksum is not option settings in accordance with
0. Section 1.52.
I
. Failure Identification Co
CHECKS
c
-.
and rotates freely. Check coupling between pump mechanism and motor. Check encoder disk for secure mounting.
+
Turn pump off and back on. If failure recurs, replace
I
c
Turn pump off and back on.
Check calibration of downstream
+
Turn pump off and back on.
Reset all of PDP.
Reset all of PDP.
1
I Turn pump off and back on. 1 Check battery voltage in
de! 7 s
CORRECTIVE ACTION
Apply a trace of Loctite* 211 or 222 to coupling
set screws and tighten. Secure encoder disk. Replace Sensor board. Replace CPU board.
CPU board. message from Slave
If failure recurs, replace CPU board.
See Failure ldentificatior
I
Code 2.
1
See Failure ldentificatior Code 4.
If failure recurs, replace CPU board.
If failure recurs, replace CPU board.
If failure recurs, replace
CPU board.
If failure recurs, replace
CPU board.
If failure recurs, charge c replace the battery.
95 Checksum error of alarm
log data.
==l ~~~~
5-14
configuration mode.
7-19-Al-688
Page 61
Page 62
Troubleshooting
Table 5-2. Troubleshooting Chart
SYMPTOM CHECKS CORRECTIVE ACTION One or more front panel keys Turn pump off and back on and Replace front panel film.
are not accepted. try the key again. Replace Sensor board.
Check front panel cables for Replace Display board.
! proper connection and
continuity.
PANEL LOCK switch is not Check the operation of the 1 Replace PANEL LOCK
accepted. switch. j switch.
Check CN701, CNVF2 and CNBUSlKNBUS2 for proper
connection and continuity.
The pump does not run on AC Check CN301 and Repair or replace power cord. but runs on internal battery. CNBUSlKNBUS2 for proper Repair or replace the
connection and continuity. connectors. Replace CPU
Check that the AC power cord board. is firmly connected. Replace Sensor board.
The Battery Icon is lit while Check AC fuse (F2). Replace F2 with a 2A, slow-
pump is running on AC. Check connectors CNAC, blow fuse, if it has failed.
CN301 and CNBUSlKNBUS2 Replace CPU board. for proper connection and Replace Sensor board. continuity.
The Battery Icon is off while
pump is running on internal battery. board.
The LCDs are blank, indicate Check ICP301 and CNDISP for Install LCD Cushions, if
irrational display or have proper connection and necessary. See Section 6.3.30. missing segments. continuity using Test Mode 5 Replace CPU board.
The backlight does not light. Replace Display board. The backlight cannot be turned Replace Sensor board.
off.
The pump turns off or stops Check ICP301. Replace ICP301. with audible alarm for no Check CNBUSlKNBUS2 for Repair or replace
apparent reason. ~ proper connection and CNBUSlKNBUS2.
~ Turn pump off and back on. I If failure recurs, replace 1
(Section 53.5). Replace Sensor board.
continuity.
1 Replace Sensor board. ’
I Display board. Replace CPU
Replace Display board.
Replace Display board.
ALARM LED does not light.
connection and continuity.
5-16 7-19-Al-688
-
Page 63
.e 5-2. Troubleshooting Chart
Tal
SYMPTOM
ALERT LED is always lit.
ALERT LED does not light. Check CNDISP for proper
PUMPING LED is always lit. Check CNDISP for proper
PUMPING LED does not light. Check CNDISP for proper
ALARM and ALERT LEDs are lit but the pump will not operate.
Audible alarm is always on. Momentarily disconnect and
Audible alarm does not sound.
Beep does not sound when a key other than BACKLIGHT or SILENCE is pressed when the pump is stopped.
CHECKS Turn pump off and back on.
Check CNBUSlKNBUS2 for proper connection and continuitv.
I
connection and continuitv.
c I
connection and continuitv.
i
connection and continuity.
L
Check backup battery voltage in accordance with Section
53.1.
+
reconnect battery. Check F302 or ICP301. Check
CNBUWCNBUS2, CN302, CN701, CN751, and CNI/F for proper connection and
continuity.
+
Turn pump off and back on. If problem recurs, replace CPU
Turn pump off and back on. Check CN751, CN701, CNI/Fl, and CNBUSlKNBUS2 for proper connection and continuity.
L
DOOR OPEN alarm occurs while door is closed, or does not occur when door is opened.
Optional nurse call is always on.
Optional nurse call is always off.
Check magnet on door latch. Check CN851.
+
Check CN401, CNI/F2 and CNBUSlKNBUS2 for proper connection and continuity.
Verify that the Nurse Call feature has been enabled via the addition of a jumper (R421)
on the terminal PCB. Check
CN401, CNI/F2 and
CNBUSlKNBUS2 for proper connection and continuity.
Troubleshooting
CORRECTIVE ACTION
1-
If problem recurs, charge or replace battery. Replace CPU board. Replace Display board.
Replace CPU board. Replace Display board.
Replace Sensor board. Replace Display board.
Replace Sensor board. Replace Display board.
+
Sometimes this condition is
followed by a F49 and/or F94
code. Recharge or replace the batterv to clear the condition.
i
Charge or replace battery. Replace F302, ICP301, and/or
connectors as required. Replace CPU board.
board. Replace Sensor board. Replace speaker.
I
If problem recurs, replace CPU board. Repair or replace the
connectors.
Replace Sensor board. Replace speaker.
Replace door latch if magnet is
missing.
Replace CPU board. Replace CPU board.
Replace Sensor board.
Replace CPU board. Replace Sensor board.
7-19-Al-688
5-17
Page 64
Troubleshooting
SYMPTOM Configuration option cannot be
set via the remote computer.
Note: Computer monitoring and remote control are available only when used with user-generated software applications. See the Flo-Gard@ 6201 and 6301
Programmer’s Manual for details.
Pump operation is louder than
normal.
The rear clamp fails to hold the
pump on the IV pole.
Free flow occurs when the door is closed.
Table 5-2. Troubleshooting Chart
CHECKS
+
Check the cable connections between the computer and the
w-v. Check that the user-generated
software is properly loaded onto the computer. Check CN401, CNI/F2 and CNBUSlKNBUS2 for proper connection and continuity.
Check administration set for proper type and code.
Check for solution spills on the pumping fingers and around the back plate. Check the tightness of the motor mounting bracket screws. Check for a PC board or other assembly vibrating against the inside of the case.
Check for worn or missing friction pad. Check the clamp for worn threads and other worn or defective parts.
Check administration set for proper type and code.
Check position and seating of the tubing. Check that ambient and solution temperatures are between 60” and 100” F (15.5”and 37.7” C). Check for cracks in the door hinges. Check for deformation of the coil springs of the back plate. Check for worn cams and pumping fingers. Check for damaged or deformed door latch.
CORRECTIVE ACTION Repair connections between
computer and pump. Restart the program.
Repair or replace connectors as required. Replace CPU board. Replace Sensor board.
Replace with Baxter’s “s” suffix standard administration set, if required. Clean in accordance with Section 2.
Tighten as required.
Ensure all internal assemblies
are securely fastened.
Replace worn or defective parts on the clamp as necessary.
Replace with Baxter’s “s” suffix standard administration set if required.
Position tubing properly, without stretching or slack. Raise or lower the
temperatures.
Replace the pump head in accordance with Section 6.3.8. Replace the springs if required.
Replace worn parts as required Replace door latch in
accordance with Section 6.35
-
5-18
-.
7-19-Al-688
Page 65
Troubleshooting
Table 5-2. Troubleshooting Chart
SYMPTOM CHECKS CORRECTIVE ACTION Free flow occurs when the
door is opened. proper type and code. standard administration set if
Backup buzzer does not
activate during self test.
INSERT SLIDECLAMP alert or Ensure that the slide clamp Clean the slide clamp sensor. alarm is constantly on. sensor is clean. Note: The alarm option is Check connections. Disconnect and reconnect. available only on pumps Check calibration and Perform steps in Section 6.4.5. running software versions operation.
1.09 or later. Replace the safety/slide clamp
Check administration set for Replace with Baxter’s “s” suffix
required. Check for solution spills on and Clean in accordance with around safety clamp. Section 2. Check position and seating of Position tubing properly and tubing in safety clamp. make certain it is seated in
guides and channels. Check that ambient and Raise or lower the solution temperatures are temperatures. between 60” and 100°F (15.5” and 37.7%). Check for broken safety clamp Replace broken parts if spring and/or broken safety required. clamp.
Check that the buzzer wiring is Plug in connector. connected. Check for an open fuse on the Replace fuse. battery wire harness.
Replace backup buzzer.
assembly.
7-19-Al-688 5-19
Page 66
Troubleshooting
s-20
7-19-Al-688
Page 67
Section 6
Disassembly and Calibration

6.1 Introduction

This section of the manual includes a list of tools and test equipment required for performing maintenance, procedures for removing and replacing subassemblies once the cause of a malfunc­tion has been determined, and procedures for calibration after component or circuit board re­placement. Detailed exploded views of the pump are provided in Section 8.

6.2 Preparation for Maintenance

6.2.1 Tools and Test Equipment The following tools and test equipment are required to perform the procedures contained in this
section. Since all fasteners on this pump are metric, ensure that all tools used are for met­ric fasteners. Tightening torques on certain screws are specified in kgf-cm and in-lb for your convenience. The values in in-lb are approximate.
Note: Reset the configuration options in accordance with Section 1.52 after
performing any repairs. This is especially important after replacing the EPROMs, and after the occurrence of failure codes 49 or 94.
7-19-Al-688
6-1
Page 68
Disassembly and Calibration
Test Equipment
Digital Voltmeter DC Ammeter 10 ohm, 1 OW resistor with variable load (see Fig-
ure 6-l)
Thickness gauge, part number UKOGl 013.6
0.3 mm, flat- feeler gauge Grounded wriststrap
Anti-static mat or other anti-static work surface Calibration tubing, part no. 3-2-92-479 Precision 0.1 yL gas tight syringe,
part no. S9662-81 (or equivalent)
Tools
Razor blade Metric Phillips-head screwdriver, #0 Metric Phillips-head screwdriver, #1
Metric Phillips-head screwdriver, #2 UPP extractor, part no. UKOG1020.A
Wire harness connector extractor, part no. UKOGlO21 .A
Torque screwdriver, O-l 5 kgf-cm or O-20 in-lb Metric ball point Allen wrench, 1.5 mm Slide Clamp Resetting Tool,
part no. 020416132
Precision 250 PL gas tight syringe.
part no. S9662-82 (or equivalent) Oscilloscope (dual trace)
Soldering Iron, temperature-regulated, 600” ­7OO”F, 20 - 48 Watts, l/32” tip
6.2.2 Recording the Configuration Option Settings It is necessary to record the.configuration option settings and the alarm log data before begin-
ning maintenance procedures so that the pump can be reconfigured properly when maintenance is completed.
Note:
When the battery, EPROMs or a circuit board is disconnected or re­placed, or a calibration is performed, all the configuration options must be reset. See Section 1.5.2.
1. Turn on the pump. Enter the configuration review mode by pressing the TIME and TOT VOLDTATUS keys simultaneously. Hold the keys for 1 second.
2. Press the NEXT or SEC START key to access each parameter sequentially. Record the configuration settings.
3. Exit the configuration inspection mode by pressing TIME and TOT VOWSTATUS again or press the STOP key.
6-2
-,
7-19-Al-688
Page 69
Page 70
Disassembly and Calibration
b. Battery connector, CN302, from the sensor board (Figure 8-2, item 4) located in the
front half of the pump.
Note: The backup audible alarm may be activated. Continue with step c to
disable the alarm.
c. Battery connector, CN75 1, from the backup audible alarm board (Figure 8-3, item 3)
located in the rear half of the pump.
Note: Unlock the flat cable connector flange before disconnecting the cable.
d. Disconnect the flat cable connector from the back case half by first pulling the flange
forward about 2 mm (5/64”). (CN-IF2)
Note:
Later versions of the pump may have the ribbon cable tacked to the side of the case half.
5. Lay the front case half down on its front surface. Position it immediately in front of the rear half. Separate the two housings by disconnecting the ground wire from the main
ground plate (Figure 8-3, item 7) from the back case half.
6. Reassemble the housings in the following order:
a. Replace the ground wire.
b. Stand the front housing up. Make sure that the connector flange is unlocked and
reconnect the ribbon cable connector. Push in the flange to hold the cable in place. Use double sided tape as necessary.
c. Reconnect CNl, CN75 1 and CN302 in that order.
d. Route the following wires between the front and rear housings so that they will not
be pinched by internal components when the housings are closed:
6-4
Ground wire Battery harness Pump head harness Ribbon cable
e. Route the battery harness so that it is located over the battery compartment and
through the cut portion of the CPU and Power Supply PCBs (Figure 8-2, items 3 and
13).
7-19-Al-688
Page 71
Disassembly and Calibration
Caution: The CPU board (item 3) must be placed in the space between the bat-
tery compartment and the power transformer mounting bracket. Do not hit the EPROMs on the board against the battery compartment.
f. Close the two housings by slightly lifting and moving the rear housing slowly toward
the front housing.
g. Plug power cord into an AC outlet and check that the Plug Icon is lit. If not, check
for proper connection of CNl .
h. Press the ON-OFF/CHARGE key and check that the self test is performed correctly.
i. Be sure the front and rear housing surfaces are flush before tightening. Replace 6
screws (Figure 8-4, item 41) and tighten them to 9 kgf-cm (7.8 in-lb) with a torque screwdriver.
j. Replace 2 screws (Figure 8-4, item 52) at handle. Tighten them to 9 kgf-cm (7.8
in-lb) with a torque screwdriver.
k. If no other repairs are performed, perform the Operational Checkout procedures in
Section 7.
6.3.2 Separation of Printed Circuit Boards
1. Separate the front and rear housings completely (Section 6.3.1).
2. Disconnect the two connectors (CN601 and CN602) of the front panel film from the sen­sor board.
3. Disconnect four pump head connectors, CN201, CN85 1, CN803 and CN8 11, from the sensor board (Figure 8-2, item 4) with the wire harness connector extractor, part no.
UKOG1021.A.
4. Remove the outermost four screws affixing the sensor board to the front housing. They are marked by arrows. Remove the ground wire screw (Figure 8-2, item 15) beside mo­tor only.
5. Hold both top corners of the CPU board (Figure 8-2, item 3) and lift it until the sensor board (item 4) engages with the ground tab of the pump head mounting bracket. Then in­cline the boards toward the pump head assembly and slide the boards out of the side of the front half.
7-19-Al-688
6-5
Page 72
Disassembly and Calibration
6. Place the printed circuit boards with the LCD display face down on a dust free flat sur­face. Be careful not to touch or damage the LCD display surface. Keep them away from
dust. Take care not to place the LCD display on components (such as screws) which could damage the LCD display.
7. Reassemble in reverse order. Be sure to remove any dirt or dust on the LCD or its win­dows on the front housing before tightening the sensor board to the front housing. Be
sure to reconnect all the removed connectors. Tighten the four screws (item 18) securing
the sensor board to 7 kgf-cm (6.1 in-lb).
Note: The function of the four printed circuit boards can be checked out-
side of the front housing as long as all the connectors are properly connected.
8. Perform the Operational Checkout procedures in Section 7.
6.3.3 Replacement of Front Panel Film
1. Separate the front and rear housings completely (Section 6.3.1).
2. Separate the printed circuit boards. (Section 6.3.2).
3. Remove one screw (Figure 8-2, item 14) that grounds the front panel film.
Caution: Do not damage the front housing or the EM1 plating with the tool. Do
not reuse the removed front panel.
4. Remove the front panel film and the silicon rubber around the panel from the front hous­ing with a razor blade or similar tool.
5. Reassemble in reverse order. Tighten the front panel film ground screw (item 14) to 3 kgf-cm (2.6 in-lb). Remove the adhesive backing from the new front panel film edges and press firmly around front panel film edges and ensure that film is securely attached to the pump with no gaps.
6. Perform the Operational Checkout procedures in Section 7.
--.
7. Seal the edges of the new front panel as follows:
a. Place the pump on its back and apply a strip of masking tape along the four edges
and comers of the front panel, approximately .38 to .64 mm (0.015 to 0.025”) from
the edges and comers of the panel.
6-6
--X
7-19-Al-688
Page 73
Disassembly and Calibration
b. Using a dispenser with a small-diameter tip, apply a bead of Toray Dow Corning*
Silicone SE 9189 L (part no. UPAS0002.A) along the edges and corners of the panel. Apply the bead uniformly, within 40 to 50 seconds per edge.
c. Remove the masking tape before the silicone starts to form a skin.
d. Keep the pump on its back for 30 - 45 minutes.
6.3.4 Replacement of Pump Head Door Cover
Open the pump head door.
1.
2. Remove five screws that secure the pump head door cover to the pump head door. (Fig­ure 8-1, item 12) Do not remove the screw affixing the metal plate in the recess at the
bottom of the door (Figure 8-5, item 28).
Reassemble the cover in reverse order. Tighten each of the five screws to 4 kgf-cm (3.5
3. in-lb) with a torque screwdriver.
6.3.5 Replacement of Pump Door Latch
1. Remove the pump door cover in accordance with Section 6.3.4.
Close the pump door.
2.
3. Remove the E-ring (Figure 8-5, item 5) that secures the door latch.
4. Lift the door latch and carefully pull it out of the pump door. Use caution to avoid los­ing the associated parts on the door latch pin.
5. To replace the latch, lay the pump on its back.
6. Place one thin washer (Figure 8-5, item S), the spring (item 6), and the other thin washer (item 8) on the door latch pin.
7. Insert the pin into the hole in the pump door.
Open the pump door slightly.
8.
9. Press the door latch in toward the pump door so that the spring is compressed.
* Toray Dow Coming
7-19-Al-688
6-7
Page 74
Disassembly and Calibration
10. Install 4 mm washer (item 7) onto the pin.
11. Install the E-ring (item 5) onto the pin.
12. Calibrate in accordance with Sections 6.4.2 through 6.4.4.
13. Perform the Operational Checkout procedures in Section 7.
6.3.6
Replacement of the Pump Door or Pump Door Assembly
1. Remove the pump door cover in accordance with Section 6.3.4.
2. Remove the two set screws (Figure 8-5, item 29)
3. Slide the hinge pins (Figure 8-5, item 14) toward each other and remove the door.
4. Install the replacement door or door assembly in reverse order. Note the alignment of the hinge pins and make sure that the tapered ends are pointed toward each other.
5. Calibrate in accordance with Section 6.4.1 through 6.4.4.
6. Perform the Operational Checkout procedures in Section 7.
6.3.7 Replacement of Door Latch Pin
1. Remove one screw (Figure 8-6, item 17) that secures the latch pin mounting block to the pump head base plate .
2. Install replacement door latch pin and secure with 1 screw. Tightening torque is 7 kgf­cm (6.1 in-lb).
3. Calibrate in accordance with Sections 6.4.1 through 6.4.4.
4. Perform the Operational Checkout procedures in Section 7.
6.3.8 Replacement of Pump Head Assembly
1. Separate the front and rear housings (Section 6.3.1).
6-8
7-19-Al-688
-
Page 75
Disassembly and Calibration
2. Disconnect four pump head connectors, CN201, CN851, CN803 and CN811, from the sensor board (Figure 8-2, item 4) with the wire harness connector extractor, part no.
UKOG1021.A.
Remove the screw that secures the pump head ground wire to the pump head mounting
3.
plate (Figure 8-2, item 15, top ground wire only).
4. Remove the pump door cover (Section 6.3.4).
Open the pump door and remove one black flathead screw and Phillips screw at the top
5. and bottom right of pump head assembly. (Figure 8-1, items 5 and 6.)
Close the door and remove two black flat head screws (Figure 8-1, item 6) on the door
6.
hinge side that secure the pump head assembly to the front housing. These screws are ac-
cessible only when the door is closed. Do not open the door before completely remov-
ing the two screws from the pump head assembly.
Push the pump head assembly from the back and take it out of the front housing.
7.
Caution: Do not pull out the two flexible cables (Figure 8-6, items 10 and 11)
from their associated connector housings. This action could open the circuit. Do not remove the gasket (Figure 8-l item 4) from the front case. If resistance is felt, check any pinched wires between the pump head assembly and the mounting plate or remove tangled wiring harness. Do not force the assembly out. Insulation of pump head wire harness can be damaged.
8. Reassemble the pump head assembly in reverse order.
Caution:
Be sure not to pinch wires and gasket between the front housing and the pump head assembly when reinstalling the pump head assembly.
The three black screws (Figure 8-1, item 6) and one Phillips screw (item 5) must be torqued to 9 kgf-cm (7.8 in-lb). The screw (Figure S­2, item 15) for the pump head ground wire must be torqued to 7 kgf­cm (6.1 in-lb).
Route the pump head wiring harnesses as close to the side of the pump head assembly as possible so that the wires are not pinched or damaged by internal components when the housings are closed.
9. Calibrate in accordance with Sections 6.4.1 through 6.4.4.
7-19-Al-688
6-9
Page 76
Disassembly and Calibration
10. Perform the Operational Checkout procedures in Section 7.
6.3.9 Replacement of Upstream Occlusion Sensor
1. Remove two Phillips screws (Figure 8-7, item 16) which secure the sensor (item 9) to the pump head base plate.
2. Pull and rotate the sensor 45 ’ clockwise and carefully remove it from the pump head base plate.
3. Desolder black and white wires from sensor. Use caution to avoid contacting any plas­tic parts with the soldering iron.
4. Remove the screw (Figure 8-7, item 6) securing grounding plate (item 1 1), ground wire and star washer (item 8). Do not lose the plate, washer and screw.
5. Install replacement sensor and reassemble in reverse order. Be sure to attach the ground wire and plate to the sensor, see Figure 8-7. Tightening torque for item 16 is 3
kgf-cm (2.6 in-lb). Tightening torque for item 6 is 1.5 kgf-cm (1.3 in-lb).
6. Calibrate in accordance with Section 6.4.1 and 6.4.4.
7. Perform the Operational Checkout procedures in Section 7.
6.3.10 Replacement of Downstream Occlusion Sensor Assembly
1. Remove two Phillips screws (Figure 8-7, item 16) which secure the sensor to the pump head base plate.
2. Pull and rotate the sensor 45” clockwise and carefully remove it from the pump head base plate.
3. Desolder black and white wires from sensor. Use caution to avoid contacting any plas­tic parts with the soldering iron.
4. Remove the screw (Figure 8-7, item 6), ground wire and star washer (item 8). Do not lose the washer and screw.
5. Install replacement sensor and reassemble in reverse order. Tightening torque for item
16 is 3 kgf-cm (2.6 in-lb). Tightening torque for item 6 is 1.5 kgf-cm (1.3 in-lb).
6-10
7-19-Al-688
-.
Page 77
Page 78
Disassembly and Calibration
7. Perform the calibration in Section 6.4.1 and 6.4.2.
8. Perform the Operational Checkout procedures in Section 7.
6.3.12 Replacement of Force Sensing Resistor* (FSR**) Devices for Tube Misloading Sensors
1. Remove pump door cover in accordance with Section 6.3.4.
2. Separate the front and rear housings (Section 6.3.1).
3. Disconnect the tube misloading sensors (Figure 8-6, items 10 and 11) from the connector at rear top of the pump head assembly by pulling their connector housings. Do not re­move them by pulling the film portion.
4. Open the pump door.
5. Remove two Phillips screws (Figure 8-7, item 16) which secure the upstream occlusion sensor (Figure 8-7, item 9) to the pump head base plate.
Caution: Do not scratch the ribbon connectors located above the sensors.
Scratches can damage the connectors.
6. Pull and rotate the sensor 45” ‘clockwise and carefully remove it from the pump head base plate.
7. Remove the mounting block for the latch pin (Figure 8-6, item 1) if the right sensor will be replaced.
Caution: Do not damage the base plate or the surface of the other tube misload-
ing sensor with the tool. Do not reuse the removed tube misloading sensor.
8. Remove the tube misloading sensor from the pump base plate with a razor blade or simi-
lar tool.
Note: All adhesive residues must be removed from the surface to ensure the
successful attachment of FSRs**.
9. Completely remove any adhesive residue from the base plate and clean the surface with an alcohol-soaked cloth.
* Interlink Electronics **Interlink Electronics
6-12 7-19-Al-688
-
Page 79
Disassembly and Calibration
10. Using a digital multimeter, check that the resistance of the new tube misloading sensor at no load is between 140 KQ and 210 KQ, If not, do not use it.
11. Loosely attach a new tube misloading sensor to the pump base plate. Ensure that it fits within its recess and does not override the pump head area. Do not attach the sensor firmly to the pump base plate yet.
12. Check that the resistance of the tube misloading sensor at no load is within 140 KR and 210 KR. If not, repeat steps 8 through 12.
13. If the sensor resistance reading is normal, press the tube misloading sensor firmly and en-
sure that it is securely attached to the pump base plate with no gaps.
14. Connect both sensors to the connectors at top rear of the pump base plate through the hole for the upstream occlusion sensor. (See Figure 8-6.)
15. Replace the upstream occlusion sensor to the pump base plate and calibrate in accord-
ance with Section 6.4.1 and 6.4.4.
16. Install the mounting block for the latch pin (Figure 8-6, item 1) if it was removed. Tight-
ening torque is 6 kgf-cm (5.2 in-lb).
17. Reassemble the case halves.
18. Perform the Operational Checkout procedures in Section 7.
6.3.13 Replacement of Pump Motor
1. Remove pump door cover in accordance with Section 6.3.4.
2. Separate the front and rear housings (Section 6.3.1).
3. Remove pump head assembly in accordance with Section 6.3.8.
4. Close the pump door.
5. Carefully place the pump head assembly face down on the work surface.
6. Carefully remove the two springs (Figure 8-8, item 9) above the motor coupling. Per-
form this step carefully to avoid losing springs.
7-19-Al-688
6-13
Page 80
Disassembly and Calibration
7. Remove two screws (Figure 8-9, item 30) that secure the motor holding plate to the safety/slide clamp assembly.
8. Remove two screws, washers and spring washers (Figure 8-8, items 22, 14 and 25) secur­ing the encoder cover (Figure 8-8, item 16) and finger box cover to the finger box. Do not lose the short nut bar (Figure 8-8, item 31).
9. Remove two screws, washers and spring washers (Figure 8-8, items 22, 14 and 25) secur­ing the finger box cover to the finger box (item 15) above the motor coupling. Do not
lose the long nut bar (Figure 8-8, item 32).
10. Remove the finger box cover.
Note: Slide the encoder cover toward the encoder to prevent it from stick-
ing to the encoder screw head.
11. Carefully lift the motor and remove motor and its associated assembly from the pump
head base plate. Take care not to deform the encoder or damage the fingers. Keep the finger box and the fingers away from dirt or dust.
12. Remove the motor holding plate (Figure 8-9, item 8) from the motor shaft bottom and
carefully place on the work surface.
13. Remove two set screws (Figure 8-8, item 27) closest to the motor.
14. To install the motor, push motor shaft into motor coupling (item 8). Install set screws
(item 27). Do not tighten them at this time. Attach the spring retainer (item 2) to the re-
placement motor, as necessary.
15. Place the motor holding plate (Figure 8-9, item 8) onto the motor shaft bottom such that the plate sticks out of the motor housing on the side with the motor wires (Figure 8-8).
16. Install the motor and its associated assembly into the pump head base plate with the fol­lowing cautions:
a. The motor must be positioned such that the wires go out rear of the base plate
(Figure 8-8).
b. Be sure that the fingers are in the finger box and the motor holding plate is properly
installed in the cut of the safety/slide clamp assembly (item 1).
6-14 7-19-Al-688
Page 81
Check that the bearings (item 10) are properly seated on the finger box and install the fin-
17. ger box cover using the four screws, washers and spring washers (items 22, 14 and 25). Before tightening these screws, rotate the motor coupling by hand to ensure smooth rota­tion of the motor shaft and the encoder. Tighten the screws to 6 kgf-cm (5.2 in-lb).
18. Install the motor holding plate onto the safety/slide clamp assembly. Tighten the two screws (Figure 8-9, item 30) to 6 kgf-cm (5.2 in-lb).
19. Apply a trace of Loctite* 211 or 222 to the set screws (Figure 8-8, item 27).
20. Place a 0.3 mm, flat-feeler gauge between the motor bottom surface and the motor end plate. If the flat-feeler gauge cannot be inserted or the gap is wider than .3mm, loosen the set screws (Figure 8-8, item 27) to move the motor. Ensure that the edge of the flat­feeler gauge is touching the side of the motor shaft. Press the motor against the plate and tighten the set screws to 5 kgf-cm (4.3 in-lbs).
Note: When the set screws have been properly tightened, the screw heads
will be recessed in the motor coupling and the screw ends will be in contact with the flat surfaces of the finger shaft.
Install the springs removed in step 6, and remove the flat-feeler gauge.
2 1.
22. Install pump head assembly into case in accordance with Section 6.3.8.
23. Perform the Operational Checkout procedures in Section 7.
6.3.14 Replacement of Safety/Slide Clamp Assembly Remove pump door cover in accordance with Section 6.3.4.
1.
2. Separate the front and rear housings in accordance with Section 6.3.1.
3. Remove the pump head in accordance with Section 6.3.8.
4. Press the safety clamp pin (Figure 8-9, item 6) to close the safety clamp. Remove the screw (Figure 8-7, item 2) on the right side of the safety clamp arm cover (item 10) and
remove the arm cover. Close the pump door.
5. Place the pump head face down and remove the two screws (Figure 8-9, item 35) to re-
move the sensor assembly.
* Loctite, Inc.
7-19-Al-688 6-15
Page 82
Disassembly and Calibration
6. Remove the two screws (Figure 8-9, item 30) on the bottom of the safety/slide clamp as­sembly to release the motor holding plate (item 8).
7. Remove the six screws (Figure 8-8, item 60) on the finger box and lift the finger and mo-
tor assembly off of the base plate.
8. Remove the flexible cable at the back of the slide clamp assembly (Figure 8-9, item 29)
from the connector on the terminal board. Be sure to pull the reinforced part so as not to damage the flexible cable.
9. Remove four screws (Figure 8-8, items 28 and 55) securing the safety/slide clamp assem-
bly and remove it from the pump head base plate.
10. Press the safety clamp pin (Figure 8-9, item 6) on the replacement assembly to close the safety clamp, with the safety clamp seal dangling, install the replacement safety/slide
clamp assembly into rear of the pump head base plate and secure with four screws (Fig­ure 8-8, items 28 and 55). Make sure safety clamp seal (Figure 8-9, item 25) is in place and moves freely side to side with the safety clamp arm. Tightening torque for the four screws is 6 kgf-cm (5.2 in-lb).
11. Connect the flexible cable (Figure 8-9, item 29) to the connector on the terminal board.
12. Place the motor holding plate (Figure 8-9, item 30) on the motor shaft and slide it in the
safety/slide clamp assembly while replacing the finger and motor assembly. Install the six screws (Figure 8-8, item 60). Tightening torque is 6 kgf-cm (5.2 in-lb).
13. Install the two motor holding plate screws (Figure 8-9, item 30). Tightening torque is 4 kgf-cm (3.5 in-lb).
14. Place a 0.3mm, flat-feeler gauge between the motor bottom surface and the motor end plate. If the flat-feeler gauge cannot be inserted, or the gap is wider than .3mm, loosen the set screws (Figure 8-8, item 27) to move the motor. Ensure that the edge of the flat­feeler gauge is touching the side of the motor shaft. Press the motor against the plate and tighten the set screws to Skgf-cm (4.3 in-lb).
15. Install the sensor 2 assembly using the two screws (Figure 8-9, item 35). Tightening torque is 3 kgf-cm (2.6 in-lb).
16. Install safety clamp arm cover. Tightening torque for the arm cover securing screw (Fig­ure 8-7, item 2) is 1 kgf-cm (0.9 in-lb). Remove the flat-feeler gauge.
17. Install pump head assembly into case in accordance with Section 6.3.8.
6-16 7-19-Al-688
Page 83
Disassembly and Calibration
-
18. Install pump door cover in accordance with Section 6.3.4.
19. Perform the Operational Checkout procedures in Section 7.
6.3.15 Spring Retainer Removal The spring retainer is installed into the slide clamp slot at the factory to maintain proper tension
on the springs within the assembly. It must be removed prior to use if the Insert Slide Clamp op-
tion is enabled.
1. Open the pump door. See Figure l-2 for the location of the spring retainer.
2. Press the safety clamp arm cover in until the safety clamp latches open.
3. Using E-ring pliers, insert the tips into the holes in the spring retainer.
4. Gently remove the spring retainer. Keep the removed spring retainer in case it is needed in the future.
6.3.16 Spring Retainer Installation The spring retainer is installed into the slide clamp slot at the factory to maintain proper tension
on the springs within the assembly. If the Insert Slide Clamp option is not going to be used, the spring retainer should remain in the assembly during pump use to maintain proper spring ten­sion as well as to help prevent the ingress of fluids into the mechanism. Instructions for install­ing the spring retainer are provided below.
1. Open the pump door and insert the Slide Clamp Resetting Tool into the slide clamp mechanism until it stops (this tool cannot be pushed in fully).
2. Press the safety clamp pin to release the slide mechanism. This pushes out the tool slightly.
3. Carefully remove the tool. If the slide clamp mechanism was activated properly, the
ridges on both sides of the tool should rub against the claws of the mechanism.
4. Insert the spring retainer until it locks in place.
7-19-Al-688
6-17
Page 84
Disassembly and Calibration
6.3.17 Replacement of Back Plate
1. Remove the pump door cover in accordance with Section 6.3.4.
2. Remove the back plate cover (Figure 8-5, item 16) by removing the screw inside the door (item 27). Then remove the 7 screws (Figure 8-5, item 26) from the outside of the door. Perform this step carefully to avoid losing springs (item 18).
3. Remove the 5 springs (item 18) and the back plate (item 17).
Install the replacement back plate. Ensure that the back plate is clean and not worn. En-
4. sure that the notch in the back plate aligns with the tongue in the door. Ensure that all 5
springs (item 18) are in place and perpendicular to the back plate cover.
5. Install the back plate cover (item 16) in reverse order. Apply pressure to the back plate
cover and install the screws (items 26 and 27) working from the outside edges to the cen­ter. That is, install the top and bottom corner screws first, followed by the screws in the
center of the back plate cover. Before tightening these screws, press the back plate a few times from the inside of the door to ensure smooth operation. Tighten these screws to 6 kgf-cm (5.2 in-lb).
Replace pump door cover in accordance with Section 6.3.4.
6.
Perform the Operational Checkout procedures in Section 7.
7.
6.3.18 Replacement of Battery
1. Turn off and unplug the pump.
2. Remove the 4 screws (Figure 8-4, item 27) on the battery compartment cover.
Slide the battery out of the rear housing.
3.
Note: Lead batteries must be recycled or disposed of properly. Contact Fed-
eral or State environmental agencies for information on recycling and disposal options.
4. Disconnect the two female connectors from the battery terminals.
Reassemble the battery into the pump in reverse order. Tightening torque for the four
5. screws is 9 kgf-cm (7.8 in-lb).
6-18
7-19-Al-688
Page 85
Disassembly and Calibration
Caution: Be sure to match the polarities of the female connectors.
Red wires (+) must be connected to red marked (+) battery terminal. See the battery label on the underside of the battery cover. The bat­tery must be placed with its connectors up. (See Figure S-4). The wires must be routed in the battery compartment so that they run be­tween the battery and the boss for the housing screw.
Be sure to place the cushion (Figure 8-4, item 17) around the boss such that its surface becomes flush with the boss.
Be careful not to pinch wires between the front housing and battery compartment cover.
6. Perform the Self Test (Section 7.3.2.2) and Battery Check (Section 7.3.8).
6.3.19 Replacement of Power Supply Board
1. Turn off and unplug the pump.
2. Separate front and rear housings completely in accordance with Section 6.3.1.
3. Separate the boards in accordance with Section 6.3.2.
4. Unplug CN301 from the sensor board (Figure 8-2, item 4).
5. Stand the boards upright and remove the screws (Figure 8-2, item 18) securing the lower corners of the power supply board.
6. Squeeze the board spacer (Figure 8-2, item 11) tops with needle nose pliers to release the
power supply board from the spacers.
7. Install the power supply board. Tightening torque for the two screws (item 18) securing
the board is 7 kgf-cm (6.1 in-lb).
8. Reassemble in reverse order. Be sure to remove any dirt or dust on the LCD windows be-
fore tightening the sensor board to the front housing. Be sure to reconnect all the re-
moved connectors.
9. Perform the Operational Checkout procedures in Section 7.
7-19-Al-688 6-19
Page 86
Page 87
Disassembly and Calibration
6. Reassemble in reverse order. Be sure to remove any dirt or dust on the LCD windows be­fore tightening the sensor board to the front housing. Be sure to reconnect all the re-
moved connectors.
7. Perform the Operational Checkout procedures in Section 7.
6.3.22 Replacement of Sensor Board
1. Separate front and rear housings in accordance with Section 6.3.1.
2. Separate printed circuit boards in accordance with Section 6.3.2.
3. Separate display board in accordance with Section 6.3.21.
4. Unplug CN301 from the sensor board (Figure 8-2, item 4).
5. Remove the three screws (item 18) securing the center of the sensor board.
6. Carefully disconnect connector CNDISP at the center and remove the sensor board from the display board.
7. Install the sensor board. Tightening torque for the three screws (item 18) securing the board is 1.5 kgf-cm (1.3 in-lb). Be sure to insert the standoffs of the spacers into the holes of the sensor board.
8. Reassemble in reverse order. Be sure to remove any dirt or dust on the LCD windows be­fore tightening the sensor board to the front housing. Be sure to reconnect all the re­moved connectors.
9. Calibrate in accordance with Sections 6.4.1 through 6.4.7.
10. Perform the Operational Checkout procedures in Section 7.
6.3.23 Replacement of Audible Alarm/Alert Board
1. Separate the front and rear housings in accordance with Section 6.3.1.
2. Remove one screw (Figure 8-3, item 23) securing the audible alarm/alert board (item 4) to the rear housing.
7-19-Al-688 6-21
Page 88
Disassembly and Calibration
3. Remove the board and unplug connector CN75 1.
4. Reassemble in reverse order. Tightening torque for item 23 is 7 kgf-cm (6.1 in-lb).
5. Perform the Operational Checkout procedures in Section 7.
6.3.24 Replacement of Backup Audible Alarm/Alert Board
1. Separate the front and rear housings in accordance with Section 6.3.1.
2. Unplug connectors CN75 1 and CN752.
3. Remove one screw (Figure 8-3, item 23) securing the backup audible alarm/alert board (item 3) to the rear housing.
4. Remove the board.
5. Reassemble in reverse order. Tightening torque for item 23 is 7 kgf-cm (6.1 in-lb).
6. Perform the Operational Checkout procedures in Section 7.
6.3.25 Replacement of Terminal Board
1. Separate the front and rear housings in accordance with Section 6.3.1,
Note: Be sure to unlock the connector flange before disconnecting the cable.
2. Unplug connectors CN401, CN701, and ribbon cable connector CNI/F2.
3. Remove three screws (Figure 8-3, item 23) securing the terminal board (item 2) to the rear housing.
4. Remove the board.
Note: Be sure to unlock the connector flange before connecting the cable
and then push it in to lock.
5. Reassemble in reverse order. Tightening torque for item 23 is 7 kgf-cm (6.1 in-lb).
6. Perform the Operational Checkout procedures in Section 7.
6-22
7-19-Al-688
Page 89
Disassembly and Calibration
6.3.26 Replacement of Power Transformer
1. Separate the front and rear housings completely in accordance with Section 6.3.1.
2. Disconnect the 2-pin CNAC connector from the power supply board (Figure 8-2, item
13).
3. Desolder the black and white wires from the transformer at the AC receptacle (Figure 8­4, item 22).
4. Remove the transformer mounting bracket by removing four securing screws (Figure 8­3, items 21 and 24).
5. Remove the transformer by removing two securing screws (Figure 8-3, item 22).
6. Reassemble in reverse order. Tightening torque for items 22 and 24 is 9 kgf-cm (7.8 in­lb) and for item 21 is 5 kgf-cm (4.3 in-lb).
Note: Be sure to match the color of the wire when resoldering the black and
white wires to the fuse holders.
7. Perform the Operational Checkout procedures in Section 7.
6.3.27 Replacement of PANEL LOCK Switch
1. Separate front and rear housings in accordance with Section 6.3.1.
2. Pull out black pushbutton (Figure 8-4, item 8) on rear of pump while holding the rear housing.
3. Desolder pink and brown wires from switch.
4. Remove hex nut and lockwasher securing the switch and remove it.
5. Reassemble in reverse order.
6. Perform the Operational Checkout procedures in Section 7.
7-19-Al-688
6-23
Page 90
Disassembly and Calibration
6.3.28 Replacement of Lithium Backup Battery The lithium backup battery provides backup voltage to the pump’s memory when the main bat-
tery is disconnected. To avoid loss of configuration option settings, record the settings and ensure that the main battery is connected before beginning this procedure.
1. Separate front and rear housings in accordance with Section 6.3.1.
2. Cut the tie wrap holding the lithium battery on the CPU board (Figure 8-2, item 3).
3. Disconnect connector CN304 and remove the lithium battery.
4. Install replacement battery in reverse order.
5. Verify that the configuration option is set as desired and modify if necessary (See Sec­tion 1.5.2).
6. Check the backup battery voltage reading per Section 5.3.1, steps 11 and 12.
7. Perform the Operational Checkout in Section 7.
6.3.29 Tube Guide Shim Installation Required Parts
Air Sensor Tubing guide shim (Part no. 722003396) Loctite* Prism 454 adhesive (Part no. 722003397) Cellophane tape
Note: Open the pump head door and check for a white tubing guide shim or
a letter “A” on the door, opposite the air sensor. If the shim or letter
“A” is present, it is NOT necessary to perform this procedure.
Note:
Make sure the pump is NOT plugged into AC power. Clean the tube guide area on the door with a small paint brush soaked in a solution of dishwashing detergent and rinse with clean water. Use an air source that is free of oil to dry the tube guide or allow it to air dry.
1. Lay the pump on its side opposite the pump head. Place the tube guide shim on the tube guide and note the surfaces making contact with the tubing guide for adhesive applica­tion. To handle the shim, press a 3-inch piece of cellophane tape onto the shim length­wise. Use the tape to lift the shim from the door. Save the tape and shim combination.
*Loctite, Inc.
6-24
7-19-Al-688
Page 91
Disassembly and Calibration
2. Use the applicator on the tube to apply a thin, uniform layer of Loctite* Prism 454 adhe­sive to the tube guide surfaces to which the shim is to be bonded. To achieve a good bond you must apply adhesive to all of the shaded areas (See Figure 6-l).
6-l. Adhesive Application
3. Immediately place the shim back onto the tube guide using the tape as a handling aid. Apply finger pressure to the comer of the shim at a 45” angle as indicated in Figure 6-2.
Apply pressure at the ends, as well as, at the middle of the shim. The tape acts as a bar-
rier preventing adhesive from contacting the fingers (See Figure 6-2).
U’
6-2.
Pressure Application Points
4. Hold the shim motionless for 60 seconds while continuing to apply pressure.
5. Remove the tape and visually inspect the shim for excess adhesive. Remove excess adhe-
*Loctite, Inc.
7-19-Al-688
sive using a dry cotton swab.
6-25
Page 92
Page 93
Page 94
Disassembly and Calibration
Note: There are two EPROMs available for replacement/upgrade in this de-
vice. Although these ICs are the same, the software stored in each is different. Therefore, ensure that the correct part is installed into the correct location.
Note: This device contains CMOS circuitry. Adequate steps to prevent dam-
age cause by electrostatic discharge should he taken such as ground straps and grounded work surfaces.
EPROM Installation
1. With the pump powered off, press and hold the SILENCE key then press the ON/OFF CHARGE key. Record the software version number shown on the lower right corner of the main display.
2. Turn off the pump and disconnect it from AC power.
3. Remove the battery from the pump (Section 6.3.18).
4. Remove the screw (item 30) from the EPROM access cover (item 4) located inside the battery compartment, and remove the cover.
5. Remove the EPROM(s) using an IC removal tool.
Note: The Master EPROM is located above the Slave EPROM (designated
on the board). Also, note that the orientation of both EPROMs is with the notch toward the bottom of the PC board.
6. Install the new version EPROM(s) in its (their) socket(s). Make sure that all of the pins are seated in the socket correctly, and that the notched end(s) of the EPROM(s) is/are to­ward the bottom of the PC board. An IC insertion tool makes this step much easier.
7. Replace the EPROM access cover along with the cover screw.
8. Replace the battery (Section 6.3.18). Connect the battery wires to the battery (check po-
larity) and carefully slide the battery into the battery compartment.
7
Note: If the secondary alarm is activated, press the ON/OFF CHARGE key
to stop the alarm. If a failure code appears, perform steps l-5 in the Set Configuration section.
9. Set the configuration to its original settings per Section 1.5.2
6-28 7-19-Al-688
Page 95
Disassembly and Calibration
10. Perform the Operational Checkout procedures in Section 7.

6.4 Calibration

Read all steps in this procedure prior to calibration of the pump. Be sure that all of the necessary equipment is available before beginning calibration procedures. During calibration it is impor­tant to keep the connectors clean.
The pump should be powered off for at least two hours before attempting to perform cali­bration. All calibration procedures should be performed in ambient and solution tempera­tures of 22°C to 28°C (72°F to 82°F).
6.4.1 DC Line Voltages
Verify DC line voltages prior to performing other calibrations. (Further calibrations will not be reliable if DC line voltages are out of specification.) If the DC voltages are not within specifica­tion, check the appropriate DC power supply circuit.
The following section may be used to verify various voltages and signals without disassembling the pump. Measurements should be made with the pump on AC power.
1. Remove the battery access cover (Section 6.3.18).
-
2. Remove the EPROM access cover by removing 1 securing screw.
3. Locate the CNTEST connector beneath the EPROM access cover.
Note: Connections called out in the following steps refer to the CNTEST
connector. Pin 8 of the CNTEST connector is always used as signal ground.
4. Connect a voltmeter to Vmain (pin 6). With the pump powered off, the reading should be
13.65 f 0.15 VDC. The ripple voltage should be measured using an oscilloscope or us­ing a 0.1 pF capacitor with the lead connected in series and the voltmeter set to AC. The ripple voltage should be 5 100 mVp-p at 105 VAC.
5. Connect the voltmeter to Vref (pin 1). Power up the pump. While in STOP mode, the voltage should be 5.000 f 0.050 VDC.
6. Measure the VcpU voltage at pin 3. The voltage should be 5.00 f 0.30 VDC.
7. Measure the Vkey voltage at pin 5. The voltage should be 5.00 f 0.30 VDC.
8. Measure the Voth voltage at pin 4. The voltage should be 12.90 to 13.90 VDC.
7-19-AI-688
6-29
Page 96
Page 97
Page 98
Disassembly and Calibration
b. Slowly turn VRl clockwise so that the voltage between pin 2 (+, red) and pin 3 (-,
black) of CN301 is 11.5 f 0.5 V.
c. Adjust the variable transformer output to 135 VAC and check that the voltage
between pin 2 (+, red) and pin 3 (-, black) of CN301 is 13.65 f 0.1 V.
d. Turn off the pump and disconnect the external circuit.
18. Reassemble the pump.
19. Perform the Operational Checkout procedures in Section 7 if any calibration was per­formed.
6.4.2 Air Sensor Calibration
Note:
This section must be performed if the pump head, sensor board, air sensor, door/door assembly, door latch, or latch pin is replaced.
Verify if a modification to the pump head door is necessary to com­plete this calibration procedure. Open the door and check for a white tubing guide shim or a letter “A” on the door, opposite the air sensor. If the shim or letter “A” is present, the modification was already per-
formed. Proceed to step 1 below. If the white tubing guide shim or 1etter”A” is not present, go to Sec-
tion 6.3.29 to install the tubing guide shim. For Flo-GardB 6201 pumps with serial numbers less than 83030018 but not ending in FA, a modification to the air sensor assembly is also necessary. Replace the existing air sensor assembly with a new assembly, Part no. F049120023.
1. Modify an I.V. tubing set by cutting the tubing about 6 inches below the first Y-site and splicing in a piece of calibration tubing (Part no. 3-2-92-479) using tubing connectors. Fully prime the set with distilled water.
Note: This fluid filled tube, along with a dry calibration tubing segment,
will be used to calibrate the pump. These sections of tubing should
only be used to calibrate 5 pumps with the tubing moved to an un­used section for each pump tested. After the calibration of the fifth pump, discard the segments and replace them with new ones.
2. Place the pump upright. If the case is not already opened, separate the front and rear housings (see section 6.3. l), but do NOT remove any connectors. Plug the pump into AC power.
6-32
7-19-Al-688
Page 99
Disassembly and Calibration
Warning: When working on an open pump, an electrical hazard exists when
the pump is plugged into AC power.
3. Set the pump to Test Mode 2. Press and hold the CLEAR TOT VOL and “2” keys, then press the ON/OFF CHARGE key. Release the ON/OFF CHARGE key without releasing the CLEAR TOT VOL and “2” keys for at least three seconds. The pump is in Test Mode 2 if NORM and MIN are displayed on the left hand side of the message (upper) display. If the Pump is not in Test Mode 2, turn off the pump and repeat this step.
4. Using the set created in Step 1, load the fluid filled calibration tubing segment, into the pump head and close the pump door. Open and close the pump door two more times and wait 2 minutes.
5. Adjust the NORM value (PRI RATE window) and the MIN value (PRI VTBI window) using VR80 1. Take the greater of the NORM and MIN values and calibrate to 450 + 20 (430 - 470). If the lesser value is less than 410, adjust it up so that it is just above 410. Recheck that the greater of the NORM and MIN values is still within 430 - 470. If both
the NORM and MIN air sensor values are within the specified ranges, continue to the next step. If these ranges cannot be achieved, replace the air sensor per Section 6.3.11. After the air sensor is replaced, begin this section, starting with step 2.
6. Remove the fluid filled tubing from the pump head. Load an air filled piece of calibra-
tion tubing into the pump head and close the pump door. Open and close the pump door two more times and wait 2 minutes.
7. Proceed to the next step if the values are 5 or less. If the values are greater than 5, adjust
VR801 so that the higher of the NORM or MIN values is 4 f 1. If adjustment was re­quired, load the fluid filled piece of calibration tubing into the pump head and close the pump door. Open and close the pump door two more times and wait 2 minutes. Verify that both NORM and MIN values are within the range of 410 and 470.
Note: Apply Red Glpt* to VR801.
8. Exit Test Mode 2 by pressing the ON/OFF CHARGE key.
9. Prime and load a Baxter’s “s” suffix standard administration set. See Section 7.3.2.1, steps l-4.
Using the precision syringe, introduce an air bubble (25 ltL for MIN setting, or 50 pL for
10. NORM setting) into tubing just above the pump head door. Press and hold the safety clamp arm cover slightly and allow the air bubble to travel to just above the air sensor. Then release the safety clamp arm cover and close the pump door.
*GC Electronics
7-19-Al-688 6-33
Page 100
Disassembly and Calibration
11. Turn on the pump, set the Primary Rate to 125 mL/hr and Primary VTBI to 1000 mL and start the pump.
12. Ensure that this bubble is allowed to pass with no alarm.
13. If the air bubble is detected, repeat the calibration procedure. If the air sensor cannot be calibrated, turn off the pump. Replace the sensor in accordance with Section 6.3.11.
14. Perform the Operational Checkout procedures in Section 7. You may wish to perform the Air Alarm Test (Section 7.3.4) prior to the other tests.
6.4.3 Downstream Occlusion Sensor Calibration
1. Check calibration as described in Section 5.3.1, steps 1 through 5.
2. If the value is out of specification, calibrate the sensor as follows:
a. Remove the pump head door cover, see Section 6.3.4. Loosen and remove the
occlusion detector stop (Figure 8-5, item 19). If the stop is stuck, apply a small amount of isopropyl alcohol to the threads. Wipe the threads clean.
b. Apply Loctite* 211 or 222 to the screw threads of the stop.
c. Repeat Section 5.3.1, steps 1 through 4. Reassemble the stop into the pump head
door and slowly turn it with a screwdriver until the value displayed in the PRI VTBI window of the Main Display is between 2998 and 3002. Knock on the center of the door and check that the reading does not change more than 2 counts. Allow the
Loctite* 2 11 or 222 to set.
d. Remove the thickness gauge and place the pump upright. Turn off the pump.
3. The following procedure confirms that the calibration is correct.
a. Load a Baxter’s “s” suffix standard administration set, prime it with fluid and close
the pump head door. Ambient and liquid temperatures must be between 72” and 82” F (22”and 28°C).
b. Occlude the tubing downstream of the pump.
c. Turn the pump on and set the Primary Rate to 125 mL/hr and Primary VTBI to 50
mL.
*Loctite, Inc.
6-34 7-19-Al-688
Loading...