Check the Bair Hugger therapy website to ensure you have the most recent version of this document. www.bairhugger.com |
reorder #200977E |
Bair Hugger*
Temperature Management Unit
Model 505
Operator’s Manual
Betriebshandbuch für Bair Hugger* Temperaturmanagement-Geräte Modell 505
Manuel de l’utilisateur pour les appareils de gestion de la température Bair Hugger* Modèle 505
Manuale d’uso delle unità di gestione della temperatura Bair Hugger* Modello 505
Manual del usuario para las unidades de control de temperatura Bair Hugger*, Modelo 505
Manual do utilizador das unidades de gestão de temperatura Bair Hugger* Modelo 505
Bair Hugger* Temperatuurregelunit Model 505 Gebruikshandleiding
Brugsanvisning til Bair Hugger* temperaturreguleringsenheder, Model 505
Bruksanvisning till Bair Hugger* Modell 505 temperaturstyrenhet
Brukerhåndbok for Bair Hugger*- temperatur-behandlingsenhet, Modell 505
Bair Hugger* -lämpötilansäätölaitteen Malli 505 käyttäjän opas
Εγχειρίδιο Χρήσης Bair Hugger* Μονάδα διαχείρισης θερμοκρασίας, Μοντέλο 505
デル505
Instrukcja Obsługi Dmuchawy
Bair Hugger* Model 505 - urządzenia do aktywnego przeciwdziałania hipotermii
Bair Hugger* 505 sz. Modell hőmérséklet-szabályozó egység használati utasítása
Návod k obsluze jednotky regulace teploty Bair Hugger* typ 505
Operatoriaus Vadovas „Bair Hugger“* Modelis 505 Temperatūros reguliavimo prietaisas
Руководство по эксплуатации
Bair Hugger* Mодель 505 Устройство управления температурой
505
Bair Hugger*
To t a l Te m p e r a t u r e M a n a g e m e n t S y s t e m
Bair Hugger*
Temperature Management Unit
Model 505
Operator’s Manual
English |
1 |
Deutsch |
15 |
Français |
29 |
Italiano |
43 |
Español |
57 |
Português |
71 |
Nederlands |
85 |
Dansk |
99 |
Svenska |
113 |
Norsk |
127 |
Sumoi |
141 |
Ελληνικά |
155 |
|
169 |
Polski |
183 |
Magyar |
197 |
Česky |
211 |
Lietuvių Kalba |
225 |
Русский |
237 |
|
249 |
Table of Contents |
|
|
|
Introduction |
|
|
3 |
Description of the Total Temperature Management* System |
|
|
3 |
Indications |
|
|
3 |
Contraindications |
|
|
3 |
Warnings |
|
|
4 |
Cautions |
|
|
4 |
Notices |
|
|
5 |
Read Before Servicing Equipment |
|
|
5 |
Set up and Operation |
|
|
6 |
Temperature Management Unit |
|
|
7 |
Control Panel Features of the Model 505 Temperature Management Unit |
|
|
7 |
Mounting the Model 505 Temperature Management Unit |
|
|
8 |
Attaching and Storing the Model 505 Unit Hose |
|
|
9 |
General Maintenance |
|
10 |
|
Technical Service and Order Placement |
|
10 |
|
In-Warranty Repair and Exchange |
|
|
10 |
When You Call for Technical Support |
|
|
10 |
Specifications |
|
11 |
Definition of Symbols |
13 |
Bair Hugger Model 505 Temperature Management Unit Operator’s Manual |
English 3 |
The Bair Hugger* brand Total Temperature Management system consists of a Bair Hugger forcedair temperature management unit (with available rolling stand and sheet clip) and disposable components, including Bair Hugger forced-air blankets, Bair Paws* patient warming gowns, and the 241* blood/fluid warming set. You can use the Model 505 temperature management unit in all clinical settings including the operating room to provide patient temperature management.
This manual includes operating instructions and unit specifications for the Model 505 forcedair temperature management unit. For information about using Bair Hugger blankets, Bair Paws gowns, or the 241 blood/fluid warming set with Bair Hugger units, refer to the “Instructions for Use” included with each of these disposable components.
The Bair Hugger Temperature Management system is intended to prevent and treat hypothermia. In addition, the temperature management system can be used to provide patient thermal comfort when conditions exist that may cause patients to become too warm or too cold. The temperature management system can be used with adult and pediatric patients.
Do not apply heat to lower extremities during aortic cross-clamping. Thermal injury may occur if heat is applied to ischemic limbs.
English 200977E
4 English |
Bair Hugger Model 505 Temperature Management Unit Operator’s Manual |
1.Do not leave patients with poor perfusion unmonitored during prolonged warming therapy sessions. Thermal injury may result.
2.The Bair Hugger temperature management unit has been designed to operate safely ONLY with Arizant Healthcare* disposable components. Use with other products may cause thermal injury. (To the full extent permitted by law, the manufacturer and/or importer declines all responsibility for thermal injury resulting from the unit being used in conjunction with nonArizant Healthcare products.)
3.Do not warm patients with the temperature management unit hose alone. Thermal
injury may result. Always connect the hose to a Bair Hugger blanket or Bair Paws gown before providing therapy.
4.Do not place the non-perforated side of the blanket on the patient. Thermal injury may result. Always place the perforated side (that is, with small holes) towards the patient.
5.Do not continue temperature management therapy if the over-temp indicator light illuminates and the alarm sounds. Thermal injury may result. Unplug the unit, and contact a qualified service technician.
6.Do not continue 241 blood/fluid warming therapy if the over-heat indicator light illuminates and the alarm sounds. Immediately stop fluid flow, and discard the blood/fluid warming set. Unplug the temperature management unit, and contact a qualified service technician.
7.Do not use a forced-air warming device over transdermal medication. Increased drug delivery and patient injury or death may occur.
8.Do not allow the patient to lie on the warming unit hose or allow the hose to directly contact the patient’s skin during patient warming; thermal injury may result. Equipment not suitable for use in the presence of a flammable anesthetic mixture with air or with oxygen or nitrous oxide.
9.Resuable blankets made from woven fabric, or blankets without discrete, visible holes, can cause the safety system of this unit to fail, which may result in serious thermal injury. This warming unit has been designed to operate safety ONLY with Bair Hugger blankets or Bair Paws gowns.
1.Except for specific blanket models, Bair Hugger blankets are not sterile and are all intended for single patient use ONLY. Placing a sheet between the Bair Hugger blanket and the patient does not prevent contamination of this product.
2.Monitor the temperature and cutaneous response of patients who are incapable of reacting, communicating and/or who are without a sense of feeling every 10-20 minutes or according to institutional protocol. Monitor the patient’s vital signs regularly. Adjust air temperature or discontinue therapy when the therapeutic goal is reached or if vital sign instability occurs. Notify physician of vital sign instability immediately.
3.Do not leave pediatric patients unattended during therapy.
4.Do not initiate temperature management therapy unless the temperature management unit is safely placed on a hard surface or securely mounted. Otherwise, injury may result.
5.To prevent tipping, clamp the Model 505 temperature management unit to an IV pole at a height that provides stability. We recommend clamping the unit no higher than 44 in. (112 cm) from the floor on an IV pole with a minimum 28 in. (71 cm) diameter wheelbase. Failure to do so may result in IV pole tipping, catheter site trauma, and the patient injury.
Arizant Healthcare Inc.