M-Series® M2000
M2000-to-M2000 Interchangeability Procedure
SCOPE
This document defines the process for updating an
M2000 amplifier.
LIMITATIONS
For M2000-to-M2000 replacement, all detector sizes are allowed
for exchange.
M2000-TO-M2000 ACCURACY STATEMENT
The M2000 amplifier will be within ±0.10% of Full Scale.
M2000-TO-M2000 UPDATE PROCEDURE
1. Record the amplier pipe diameter: ______________
2. Record the amplier detector factor: ______________
3. Record the amplier detector oset: ______________
4. Record all applicable setup parameters in the tables provided in
this document.
5. Disconnect power to the amplier before
removing connections.
6. Remove all wiring connections from the amplier.
7. Remove amplier PCB/enclosure and replace with new
PCB/enclosure.
8. Reconnect Coil/Electrode connections. Reconnect all other
pluggable inputs and outputs.
9. Apply power to the amplier.
10. Congure Pipe Diameter with original value recorded in step 1.
11. Congure Detector Factor with original value recorded in
step 2.
12. Congure Detector Oset with original value recorded in step 3.
13. Congure M2000 to remaining parameters recorded
in step 4.
RECORD SETUP PARAMETERS
Main Menu
METER SETUP
Scale Factor
Empty Pipe
Power Line Freq
Excitation Freq
Pipe Diameter
Amplifier Factor
Detector Factor
Detector Current
Detector Offset
MEASUREMENTS
Flow Unit
Totalizer Unit
Full Scale Flow
Low Flow Cut Off
Flow Direction
Damping Factor
INPUTS/OUTPUTS
Analog Output
Digital Input
Digital Outputs #1 #2 #3 #4
Pulses/Unit N/A N/A
Pulse Width N/A N/A
Full Scale Freq N/A N/A N/A
Preset Amount
Set Point Min
Set Point Max
Output Type
Select Function
MAG-AS-01086-EN-02 (August 2014)
OTE:N Tables continue on the next page.
Application Brief
Communications Menu
Advanced Menu
Interface
Modbus RTU (def)
Remote Menu
Disable Port
Port Address
Baud Rate
Parity
Data Bits
Stop Bits
Port A Settings
Unit Multiplier
Backlight Control
Analog Output
Software Filter
Activation
Filter Delay
Acceleration
Constant Flow
Peak Detect
Empty Pipe Cal
Cal Empty Pipe
Cal Full Pipe
Security
Set Admin Pin
Set Service Pin
Set User Pin
Language Select
Control. Manage. Optimize.
M-SERIES is a registered trademark of Badger Meter, Inc. Other trademarks appearing in this document are the property of their respective entities. Due to continuous research,
product improvements and enhancements, Badger Meter reserves the right to change product or system specications without notice, except to the extent an outstanding
contractual obligation exists. © 2014 Badger Meter, Inc. All rights reserved.
www.badgermeter.com
The Americas | Badger Meter | 4545 West Brown Deer Rd | PO Box 245036 | Milwaukee, WI 53224-9536 | 800-876-3837 | 414-355-0400
México | Badger Meter de las Americas, S.A. de C.V. | Pedro Luis Ogazón N°32 | Esq. Angelina N°24 | Colonia Guadalupe Inn | CP 01050 | México, DF | México | +52-55-5662-0882
Europe, Middle East and Africa | Badger Meter Europa GmbH | Nurtinger Str 76 | 72639 Neuen | Germany | +49-7025-9208-0
Europe, Middle East Branch Oce | Badger Meter Europe | PO Box 341442 | Dubai Silicon Oasis, Head Quarter Building, Wing C, Oce #C209 | Dubai / UAE | +971-4-371 2503
Czech Republic | Badger Meter Czech Republic s.r.o. | Maříkova 2082/26 | 621 00 Brno, Czech Republic | +420-5-41420411
Slovakia | Badger Meter Slovakia s.r.o. | Racianska 109/B | 831 02 Bratislava, Slovakia | +421-2-44 63 83 01
Asia Pacic | Badger Meter | 80 Marine Parade Rd | 21-06 Parkway Parade | Singapore 449269 | +65-63464836
China | Badger Meter | 7-1202 | 99 Hangzhong Road | Minhang District | Shanghai | China 201101 | +86-21-5763 5412 Legacy Document Number: IAB-188-01-EN