Badger Meter ILR 700, ILR 710, ILR 720, ILR 730, ILR 740 Installation And Operation Manual

...
®
Badger Meter Europa GmbH
Industrial oval gear meter series IOG and registers type ILR 700, ILR 710, ILR 720, ILR 730, ILR 740, ILR 741
INSTALLATION AND OPERATION MANUAL
January 2014
OG_IOG_BA_02_1401
Contents Page
1.Basic safety recommendations ........................................................................................ 1
2.Scope of this manual ......................................................................................................... 2
3.Product unpacking and inspection .................................................................................. 2
4.Product identification ........................................................................................................ 2
5.Meter installation ............................................................................................................... 4
6.Meter operation .................................................................................................................. 4
7.Register operation ............................................................................................................. 5
7.1 Normal operation ....................................................................................................... 6
7.2 Status ......................................................................................................................... 6
7.3 Totalizers ................................................................................................................... 6
7.4 Flow rate .................................................................................................................... 6
7.5 Battery ........................................................................................................................ 7
7.6 Checksum .................................................................................................................. 7
7.8 Display scale factor .................................................................................................... 7
8.Register programming ...................................................................................................... 8
8.1 Unit of measure .......................................................................................................... 8
8.2 Scale factor ................................................................................................................ 8
8.3 Changing the meter pulse rate ................................................................................... 9
8.4 Changing the register orientation ............................................................................... 9
8.5 Changing the display mode ..................................................................................... 10
8.6 Exiting Programming Mode ...................................................................................... 10
9.Additional programming: Industrial analog and industrial pulse ............................... 11
Pulse rate out ........................................................................................................... 11
9.1
9.2 Analog minimum flow rate ........................................................................................ 12
9.3 Analog maximum flow rate ....................................................................................... 12
10. Register output specifications & wiring ....................................................................... 13
10.1 Pulse (model ILR 710) ................................................................................................ 13
10.2 Dual pulse (model ILR 720) ....................................................................................... 14
10.3 Analog (model ILR 730) ............................................................................................. 15
10.4 Pulse transmitter (model ILR 740) ............................................................................. 16
10.5 Pulse transmitter (model ILR 741) ............................................................................. 17
10.6 Pulse transmitter (for 1/4’’) ......................................................................................... 18
11. Repair parts .................................................................................................................... 19
12. Return of goods for repair ............................................................................................. 20
OG_IOG_BA_02_1401
Basic safety recommendations Page 1/20
1. Basic safety recommendations
Before installing or using this product, please read this instruction manual thoroughly. Only qualified personnel should install and/or repair this product. If a fault appears, contact your distributor.
Installation
Do not place any unit on an unstable surface that may allow it to fall. Never place the units above a radiator or heating unit. Route all cabling away from potential hazards. Isolate from the mains before removing any covers.
Power connection
Use only the type of power source suitable for electronic equipment. If in doubt, contact your distributor. Ensure that any power cables are of a sufficiently high current rating. All units must be earthed to eliminate risk of electric shock. Failure to properly earth a unit may cause damage to that unit or data stored within it.
Protection class
The device has protection class IP 67 and needs to be protected against dripping water, water, oils, etc.
Setup & operation
Adjust only those controls that are covered by the operating instructions. Improper adjustment of other controls may result in damage, incorrect operation or loss of data.
Cleaning
Switch off all units and isolate from mains before cleaning. Clean using a damp cloth. Do not use liquid or aerosol cleaners.
Repair of faults
Disconnect all units from power supply and have it repaired by a qualified service person if any of the following occurs:
If any power cord or plug is damaged or frayed If a unit does not operate normally when operating instructions are followed If a unit exposed to rain/water or if any liquid has been spilled into it If a unit has been dropped or damaged If a unit shows a change in performance, indicating a need for service.
RoHs
Failure to adhere to these safety instructions may result
in damage to the product or serious bodily injury.
Our products are RoHs compliant.
Battery disposal
The batteries contained in our products need to be disposed of as per your local legislation acc. to EU directive 2006/66/EG.
OG_IOG_BA_02_1401
Scope of this manual / Product unpacking and inspection / Produkt identification Page 2/20
2. Scope of this manual
This manual contains installation and operation instructions for the Badger Meter industrial line of oval gear meters and registers.
Proper performance and reliability of these meters and registers depends upon installation in accordance with these instructions.
3. Product unpacking and inspection
Upon receipt of the product, perform the following unpacking and inspection procedures: Note: If there is damage to the shipping container, request the carrier to be present when
unpacking the product. Carefully open the shipping package and follow any instructions marked on the exterior.
Remove all packing material and carefully lift the product from the package. Retain the package and all packing material for possible use in reshipment or storage. Visually inspect the product and applicable accessories for any physical damage such as
scratches, loose or broken parts, or any other sign of damage that may have occurred during shipment.
Note: If you find damage, request an inspection by the carrier’s agent within 48 hours of delivery and file a claim with the carrier.
A claim for equipment damage in transit is the sole responsibility of the purchaser.
4. Product identification
Record the product identification numbers from the nameplate. Model # ____________________________________
Serial Number # ________________________________ Tag # ___________________________________(if applicable)
Disclaimer
The user/purchaser is expected to read and understand the information provided in this manual, follow any listed safety precautions and instructions and keep this manual for future reference.
Misuse, mishandling, and/or inadequate maintenance may impair performance and/or compromise safety.
OG_IOG_BA_02_1401
Product identification Page 3/20
Explosion and fire hazards
Improper grounding, poor ventilation, open flames or sparks can cause a hazardous condition and result in an explosion or fire and cause serious injury.
• Be sure the fluid system is properly grounded. See your pump instruction manual for details.
• If there is static sparking or if you feel an electric shock while using the meter, stop dispensing immediately. Identify and correct the problem before continuing.
• Provide fresh air ventilation. This will avoid the buildup of fumes from the fluid being dispensed.
• Do not smoke while dispensing flammable fluids.
• Keep the dispensing area free of debris including solvents, rags and spilled gasoline.
Meter hazards
Equipment misuse can cause the meter to rupture or malfunction and cause serious injury.
• This equipment is for professional use only.
• Read all instructions, tags and labels before operating the equipment.
• Use the equipment only for its intended purpose.
• Do NOT modify or alter the equipment.
• Do NOT leave equipment unattended while dispensing.
• Check equipment daily. Repair or replace worn or damaged parts immediately.
• Do NOT exceed the maximum working pressure level of the lowest rated system component.
• Use only extensions and nozzles that are designed for use with this equipment.
• Use only fluids and solvents that are compatible with the equipment. Read all fluid and solvent manufacturer’s warnings.
• Tighten all fluid connections before operating this equipment.
• Do NOT stop or deflect leaks with hands, body, gloves or rags.
• Do NOT dispense towards any person or any part of the body.
• Do NOT place hands or fingers over the end of or into the dispense valve.
• Comply with all local, state, and federal fire, electrical and safety regulations.
• Use of this product in a manner other than specified in this manual may result in impaired operation or damage to equipment.
These meters are designed to dispense a wide range of chemicals. Consult the factory for chemical compatibility.
OG_IOG_BA_02_1401
Meter installation / Meter operation Page 4/20
5. Meter installation
Read the following information and have a thorough understanding before proceeding with meter installation. Only qualified personnel should perform meter installation.
• Install a type 60 mesh strainer or Y or basket as close to the inlet side of the meter as possible. Strainers prevent dirt and other fluid contaminants from impeding meter performance. Strainers require periodic cleaning, as clogged strainers also impede meter performance. Contact your local representative for specific information, per your specific application.
Strainer
Strainer
Figure 1: Meter installation
• Turn off any associated pumps to reduce line pressure and slowly fill the line and meter with fluid before restarting pumps. Doing so reduces the possibility of meter damage caused by errant air pressures in the line and meter.
• Make sure all pipe conforms to the same pressure output rating as the pump.
• Make sure to apply thread sealant to all pipe threads.
• Install the meter along the meter shafts in a horizontal plane (see figure 1).
• Check for and repair leaks upon initialization of fluid flow.
6. Meter operation
Fluid enters the inlet port and then passes through the metering chamber. Inside the chamber, fluid forces the internal gears to rotate before exiting through the outlet port. Each rotation of the gears displaces a specific volume of fluid. As the gears rotate, a magnet on each end of the gear pass a reed switch in the top-mounted register's circuit board. The reed switches send pulses to the microprocessor in the register to change the LED display segments.
OG_IOG_BA_02_1401
Meter operation / Register operation Page 5/20
Upon initialization and continuation of fluids through the line and meter, the expected pressures per flow are:
Figure 2: Pressure drop vs. flow
7. Register operation
The following describes register operation and program settings for the industrial oval gear series registers: Industrial Standard (ILR 700), Industrial Pulse (ILR 710), Industrial Quadrature/Dual Pulse (ILR 720) and Industrial Analog (ILR 730).
The register display consists of two rows of seven-segment digits, status, unit of measures, flow rate, and battery indicators. Operating function settings and programming are provided using the TOTAL and RESET buttons.
Figure 3: Register display and buttons
OG_IOG_BA_02_1401
Register operation Page 6/20
7.1 Normal operation
(for models ILR 700, 710, 720 and 730) To enter normal operation mode - when the screen is blank after exiting programming mode, or upon initial use, press either the TOTAL or RESET button once.
7.2 Status
The status indicators are RESET and TOTAL.
7.3 Totalizers
The top row of indicators is the batch totalizer. This totalizer displays the cumulative volume of flow through the meter with six digits. The batch totalizer totalizes in selected units of measure.
To reset the batch totalizer, after 2 seconds of no flow, press and release the RESET button.
NOTE: For the ILR 720 model only, the batch totalizer can be reset by a low pulse on the external reset input.
The bottom row of indicators display the resettable totalizer with five digits or the five least significant digits of the non-resettable totalizer. RESET and TOTAL is indicated when the resettable total is displayed in the five-digit lower row. Only TOTAL is indicated when the non-resettable total is displayed.
To toggle between the non-resettable totalizer and the resettable totalizer, press and release the TOTAL button.
To reset the resettable totalizer, press and hold the TOTAL button and then press and release the RESET button.
To display 11-digit non-resettable totalizer, while the non-resettable total is displayed, press and hold the TOTAL button for seconds. The top row displays the 6 most significant digits; the bottom row displays five least-significant digits.
NOTE: The non-resettable totalizer normally displays 5 least-significant digits.
7.4 Flow rate
PER MIN is displayed in conjunction with the unit of measure. All flow rates are calculated in volume unit per minute.
OG_IOG_BA_02_1401
Register operation Page 7/20
7.5 Battery
The "LBat" indicator will indicate when the battery is approaching end of life. When the indicator is illuminated, the 2/3AA, 3.0 VDC lithium battery is drained to 10% of its total capacity and should be changed. Normal battery life is four years. "Normal" assumes operating conditions of an ambient temperature of 25°C (77° F) and a throughput of 60,000 liters [15,850 gallons, 63,400 quarts, 126,800 pints (US)].
Note: A 2/3AA, 3.6 VDC battery may also be used as a replacement.
Figure 4: Low battery indicator
7.6 Checksum
To display the firmware checksum, press and hold the RESET button for three seconds. To return to normal display, release the RESET button.
7.8 Display scale factor
To display the scale factor: At the same time, press and hold the TOTAL and RESET buttons for two seconds to
display the programmed scale factor. To return to the normal display, release both buttons.
OG_IOG_BA_02_1401
Register programming Page 8/20
8. Register programming
In programming mode only, pressing and releasing the TOTAL button advances to the next parameter on the current screen. Pressing and releasing the RESET button changes the current flashing selection to another selection (such as “L” to “GAL”).
To enter the programming mode, press the TOTAL button three times and then press the RESET button three times (the time lag between pressing both buttons six times must be within two seconds).
Changing the unit of measure and scale factor
(for models ILR 700, 710, 720 and 730)
8.1 Unit of measure
Figure 5: Unit of measure & scale factor programming
1. Press and release the RESET button to change the unit of measure (L, GAL, QT, PT).
2. Press and release the TOTAL button to select desired the unit of measure (the selected unit of measure will flash).
3. When the appropriate unit of measure is selected, press the TOTAL button to advance to the scale factor programming.
8.2 Scale factor
(for models ILR 700, 710, 720 and 730) The register collects input pulses from the oval gear meter and then determines the
appropriate display output using the scale factor. This scale factor varies depending upon the viscosity of the liquid being measured, therefore calibrating the meter and register in the appropriate liquid will affect the scale factor. The scale factor is displayed as 5 digits (on the top row) next to the unit of measure. The scale factor consists of 1 integer digit and 4 decimal digits (see figure 5).
1. Press the TOTAL button to select a digit (selected digits flash). After cycling through all 5 digits of the scale factor, the register will return to the unit of measure selection.
2. Press RESET to change the selected digit. The scale factor must fall between the values of 0.5000 and 2.0000. The Badger Meter factory preset is set between those values at 1.0000.
3. When finished adjusting the unit of measure and scale factor, press and hold the TOTAL button for one second to advance to the Pulse Rate section.
NOTE: Error checking will not allow the user to advance to the next screen.
OG_IOG_BA_02_1401
Register programming Page 9/20
8.3 Changing the meter pulse rate
(for models ILR 700, 710, 720 and 730) The meter pulse rate (screen is indicated by the “I” on the top row, on the left side) is
the number of pulses per unit of measure as detected by the register. The pulse rate varies according to the type of attached meter. The bottom row consists of the 5-digit integer value of the meter pulse rate, whereas the top row consists of the 2-digit decimal value of the meter pulse rate.
The meter pulse rate is entered in pulses per liter if the selected unit of measure is liters. The meter pulse rate is entered in pulses per gallon if the selected unit of measure is gallons, quarts or pints.
Figure 6: Meter Pulse Rate
1. Press the TOTAL button to select a digit (selected digits flash). Press RESET to change the selected digit. The pulse rate can be any value between 00000.01 and
99999.99 on the top row; integer values are displayed on the bottom row. Example: 10.45 would display .45 on the top row and 10 would be displayed on the bottom row.
2. When finished adjusting the pulse rate, press and hold the TOTAL button for one second to advance to the “register orientation” section.
NOTE: Error checking will not allow the user to advance to the next screen.
8.4 Changing the register orientation
(for models ILR 700, 710, 720 and 730) Depending on the orientation (perpendicular or inline on the meter); this setting may
need to be changed.
Figure 7: Register orientation
Flow direction
Inline
Perpendicular
OG_IOG_BA_02_1401
Register programming Page 10/20
1. Press the RESET button to toggle between available options (“I, for an inline-to- flow orientation and “P” for a perpendicular-to-flow orientation).
2. When finished adjusting the register orientation, press and hold the TOTAL button for one second to advance to the “Default Display” section.
8.5 Changing the display mode
(for models ILR 700, 710, 720 and 730) The display mode screen (indicated by a “d” on the top row, on the left side)
determines the information displayed on the top line of the register during normal operation. The display mode may be either the totalizer screen or the flow rate screen.
“C,” indicates the totalizer screen and “F” indicates the flow rate screen. The totalizer screen is depicted below:
Figure 8: Default display
1. While a letter is flashing on the display, press the RESET button to select either totalizer or flow rate.
2. Upon completion of this setting, the programming of the Industrial Standard Register and the Industrial Dual Pulse Output is complete. For ILR 710 and ILR 730 models, see additional programming parameters.
Note: For ILR 710 and ILR 730 models, see "Additional Programming: Industrial Analog and Industrial Pulse (ILR 710 & ILR 730)" on page 11.
8.6 Exiting Programming Mode
(for models ILR 700, 710, 720 and 730) To exit the programing mode:
1. On any screen, press and hold the both the TOTAL and RESET buttons. The screen will revert back to the programmed scale factor, and then flash. Following the three flashes, the register display will be blank.
Note: Pressing the TOTAL or RESET buttons wil turn the display back on.
OG_IOG_BA_02_1401
Additional programming: Industrial analog and industrial pulse Page 11/20
9. Additional programming: Industrial analog and industrial pulse (ILR 710 & ILR 730) Output pulse length
(for models ILR 710 only) Indicated by a “P” on the left hand side of the display, this screen allows the selection of
the low duration of the output pulse.
• “0” for zero milliseconds (Pulse Output is disabled)
• “2” for 2 milliseconds
• “10” for 10 milliseconds
• “20” for 20 milliseconds
• “40” for 40 milliseconds
• “100” for 100 milliseconds
To advance to the next programming screen, hold the TOTAL button.
Figure 9: Ouput pulse length screen
About Output Pulse Length: The pulse rate duration should take into account the "Pulse Rate Out" and maximum meter flow rate, to prevent an output pulse duration greater than the required time between pulses. The Output Pulse Length should be set to less than the value of “t.”
Per the equation: Maximum meter flow rate (in GPM or l/m)
t = ----------------------------------------------------------------- x 1000 60X output pulse rate
where t = the required pulse rate in milliseconds. The output pulse rate = the programmed parameter (default = 1.00 PPL/PPG)
The maximum meter flow rate = the maximum flow rate of the meter for the application.
9.1 Pulse rate out
(for model ILR 710 only) Indicated by an “o” on the left hand side of the display, this screen allows selection of
the of pulses output per liter or per gallon depending on unit of measure (0.01 PPL/PPG to 100 PPL/PPG).
The meter pulse rate is entered in pulses per liter if the selected unit of measure is liters. The meter pulse rate is entered in pulses per gallon if the selected unit of measure is gallons, quarts or pints. To advance to the next programming screen, hold the TOTAL button. NOTE: Error checking will not allow the user to advance to the next screen.
Figure 10: Pulse rate out screen
OG_IOG_BA_02_1401
Additional programming: Industrial analog and industrial pulse Page 12/20
9.2 Analog minimum flow rate
(for models ILR 730 only) Indicated by a “L” on the left hand side of the display, this screen allows the setting of
the flow rate that corresponds to the 4mA output:
NOTE: The minimum flow rate value must be less that the maximum flow rate value.
• Minimum 0.0 LPM/GPM
• Maximum 100.0 LPM/GPM
• Default 0.0 LPM/GPM
NOTE: Error checking will not allow the user to advance to the next screen. To advance to the next programming screen, hold the TOTAL button for one second.
Figure 11: Analog minimum flow rate screen
9.3 Analog maximum flow rate
(for models ILR 730 only) Indicated by a “H” on the left hand side of the display, this screen allows the setting of
the flow rate that corresponds to the 20mA output: NOTE: The maximum flow rate value must be greater than the minimum flow rate
value.
• Minimum 0.0 LPM/GPM
• Maximum 100.0 LPM/GPM
• Default 30 LPM / 8 GPM To advance to the next programming screen, hold the TOTAL button. NOTE: Error checking will not allow the user to advance to the next screen.
Figure 12: Analog maximum flow rate screen
OG_IOG_BA_02_1401
Register output specifications & wiring Page 13/20
10. Register output specifications & wiring
10.1 Pulse (model ILR 710)
Register wiring
External DC+: Yellow External ground: Brown Pulse output: White DC Input: 6 to 24 VDC; 10 to 20mA Outputs: Pulse output with internal pull-up resistor; optional open collector output
with output jumper removal; pulse output is scalable in pulses per liter or pulses per gallon.
ILR 710
8-24V DC input
Output jumper
Figure 13: ILR 710 wiring
Pulse output
Yellow
White
Brown Ground
OG_IOG_BA_02_1401
Register output specifications & wiring Page 14/20
jumper
10.2 Dual pulse (model ILR 720)
Register wiring
External DC+: Yellow External ground: Brown Pulse output 1: White Pulse output 2: Green External reset: Grey DC input: 6 to 24 VDC; 10 to 20mA Outputs: Dual-pulse output with internal pull-up resistor; optional open collector
output with output jumper removed; dual pulse output forms a quadrature signal for direction of flow.
Inputs: External reset pulled low to reset the batch totalizer.
ILR 720
8-24V DC input
Output
+3.0V DC
Ground
Figure 14: ILR 720 wiring
Pulse output
Output jumper
External reset
Yellow
White
Green
Grey Brown
OG_IOG_BA_02_1401
Register output specifications & wiring Page 15/20
r
10.3 Analog (model ILR 730)
Register wiring
External DC+: Yellow External ground: Brown Analog output: White DC input: 6 to 24 VDC; 10 to 20mA Outputs: Analog 4 to 20mA output in loop powered configuration; external load of 50
ohms to 250 ohms; flow rate is linear scaled between 4mA minimum and 20mA maximum set points; flow rates below programmed minimum read 4mA.
ILR 730
8-24V DC input
4-20 mA transmitte
Figure 15: ILR 730 wiring
Analog
output
Yellow
White
Brown Mass
OG_IOG_BA_02_1401
Register output specifications & wiring Page 16/20
10.4 Pulse transmitter (model ILR 740)
Figure 16: Pulse transmitter
Orientation: The register must be mounted as shown above (mounted in-line with the flow) with the flow going left to right or right to left as shown on this page. The transmitter will not function if mounted perpendicular to the flow.
Transmitter wiring
Reed switch outputs: Green and white. Ratings: Max power: 10W (not to exceed!); max. voltage: 200 VDC/peak AC; max. current: 0.5A DC/peak AC. Outputs: Raw reed switch output with no signal conditioning.
ILR 740
Green
White
Figure 17: ILR 730 wiring
Pulse per unit of measure
Meter Pulse
per gallon
Pulse
per liter
½” 378.5 100 ¾” 249.8 66
1” 249.8 66
1” HF 162.8 43
1 ½” 64.4 17
2” 34.1 9 3” 11.4 3
Note: Actual pulses per unit of measure are listed o n the certific ate pro vided with the meter.
OG_IOG_BA_02_1401
Register output specifications & wiring Page 17/20
A
10.5 Pulse transmitter (model ILR 741)
The meter size selector switch must be set to correspond to the size of the meter to properly detect fluid flow:
Position 1 (top): 1/2", 3/4", 1"  Position 2 (center): 1 1/2"  Position 3 (bottom): 2", 3"
Meter size
selector switch
Figure 18: ILR 741 transmitter switch positions
Transmitter wiring
ILR 741
White
Green
Reedswitch bank
Blue
uxiliary
reedswitch bank
OG_IOG_BA_02_1401
Register output specifications & wiring Page 18/20
10.6 Pulse transmitter (for 1/4’’)
Figure 19: 1/4" and 1/4" low flow wiring
Hall effect switch
Rating: Power Supply: Supply input range:5-24V DC Supply current: 3.5 mA Pulse Output Output current: 30 mA, max.
Wiring: Yellow: Hall effect DC+ Brown: Hall effect ground Green: Hall pulse output
Reed switch
Rating: Power rating: 10W Switching voltage: 100V (DC or peak AC) Switching current: 500 mA (DC or peak AC)
Wiring: Grey: Reedswitch White: Reedswitch Green: Hall effect pulse output
Pulses per liter (PPL)
Meter size Pulses per liter
1/4 “ 390
1/4 “ LF 2170
OG_IOG_BA_02_1401
Repair parts Page 19/20
11. Repair parts
5
1, 1A
2, 2A
3
Item Description
1 Aluminum cover w/screws 66885-
1A SST cover w/screws 66885-
2 SST gear service kit w/aflas o-
ring
2A LCP gear service kit w/aflas o-
ring
3 ILR-700 industrial register
w/screws
4 ILR-710 industrial register single
pulse output w/screws
4A ILR-720 industrial register quad
output with ext reset w/screws
4B ILR-730 industrial register analog
output 4-20 mA w/screws
5 ILR-740 transmitter w/screws 66885-
6 Battery (not shown) 66299-
015
016 66885-
005 66885-
006 66885-
001 66885-
002 66885-
003 66885-
004
012
001
Articel n°.
1/2” 3/4” 1” 1“ HF 1-1/2“ 2“ 3“
66885­013
66885­014
66885­007
66885­008
66885­001
66885­002
66885­003
66885­004
66885­012
66299­001
66885­013
66885­014
66885­007
66885­008
66885­001
66885­002
66885­003
66885­004
66885­012
66299­001
66885­026
66885­027
66885­035
66885­034
66885­001
66885­002
66885­003
66885­004
66885­012
66299­001
66885­028
66885­029
6685­037
66885­036
66885­001
66885­002
66885­003
66885­004
66885­012
66299­001
66885­030
66885­031
66885­039
66885­038
66885­001
66885­002
66885­003
66885­004
66885­012
66299­001
66885­032
66885­033
66885­041
66885­040
66885­001
66885­002
66885­003
66885­004
66885­012
66299­001
OG_IOG_BA_02_1401
Return of goods for repair Page 20/20
12. Return of goods for repair
Please copy, fill in and sign hereafter harmlessness declaration and enclose it for any return of goods you may send back for repair.
No repair will be performed prior to receiving the harmlessness declaration duly filled and signed.
Harmlessness declaration
To : __________________________________________________________ Attn. : __________________________________________________________ From : __________________________________________________________ Dept. : __________________________________________________________
Please note that no repair will be performed prior to receiving of this declaration duly signed by you!
Please send all parts clean from medium and inform us about possible medium wastes remaining in the part. For this purpose, please use this form. A security specification sheet of the medium must accompany this declaration in the following cases: Toxical, dangerous or objectionable media, or media belonging to any dangerous materials class. We inform you that uncleaned parts lead to additional costs. Extra clean costs will be charged to you. Furthermore, we reserve us the right to send the parts back to you for cleaning!
Declaration
We herewith confirm that the part(s) sent for repair has/have been cleaned and is/are free of any liquid and/or solid wastes of the medium and/or cleaning medium: Any eventually remaining wastes are:
harmless dangerous, toxic, etc. – Security specifications are attached
Signature of person in charge: ______________________________ Name of the person in charge in capital letters: ______________________________ Date: ______________________________
Company stamp: ______________________________
OG_IOG_BA_02_1401