Axis B0.6D, B2D, B3D, B1.5D, B6D User Manual

USER MANUAL
File: R-B-043-04-09 B42_0 GB
2
USER MANUAL
_____________________________________________________________________________________________________________________
Contents:
1. General description....................................................................................................... 3
2. Completeness............................................................................................................... 3
3. Technical data............................................................................................................... 4
4. Keys and indicators....................................................................................................... 5
5. Security rules ................................................................................................................ 6
6. Environment protection ................................................................................................. 6
7. Preparation ................................................................................................................... 7
8. Start-up......................................................................................................................... 8
9. General rules ................................................................................................................ 9
10. Balance checking .................................................................................................... 11
11. Scale adjustment..................................................................................................... 11
12. Connecting a computer or a printer (optional) ......................................................... 11
13. Special functions description................................................................................... 14
14. Normal weighing...................................................................................................... 15
15. Weighing with tare................................................................................................... 15
16. Increased readability ............................................................................................... 15
17. Autozero (F..-AUt) ................................................................................................... 16
18. Funkcja liczenia sztuk (F..-PCS).............................................................................. 17
19. Printer cooperation settings (F..-LPt)....................................................................... 18
20. Serial port parameters setting (F..-rS) ..................................................................... 19
21. Constant tare (F..-tAr) ............................................................................................. 20
22. Function Menu customisation (F..-ACt) ................................................................... 22
23. Maintenance and repairs of small defects ............................................................... 23
Declaration of Conformity................................................................................................... 24
USER MANUAL
3
_____________________________________________________________________________________________________________________
1. General description
B series platform scales are destined for general use. All scales are metrologically tested - calibration or legal verification on demand.
B series scales are made in two main versions: with LED display (basic) and with LCD display (option). Standard indicator may be changed by indicator with additional keys – see appendix B. Scales have following verification features:
- a seal protecting scale casing against opening,
- notified body stamps and green metrological marking placed on the balance name plate.
Legal verification is valid for 3 years unless the seal is broken. NACE classification: : 29.24.23. Certificates:
EC Type
Approval Certificate
No. PL 04 022
Health Quality
Certificate
No. Hś/06458/01
AXIS management System
Certificate
DIN EN ISO 9001:2000
No. 78 100 6386
2. Completeness
Standard set consist of:
1. Scale
2. User Manual
3. Guarantee card
4. Feeder ZN12V/500mA (option)
5. Charger PA6V (option)
4
USER MANUAL
_____________________________________________________________________________________________________________________
3. Technical data
Scale type B15W
B30
B60
B120 B150 B200 B300
Maximum loading 15kg 30kg
60kg
120kg 150kg 200kg 300kg
Minimum loading 100g 200g 400g
1kg
1kg 1kg 2kg
Reading unit (d) 5g
*1g
10g
*5g *1g
20g
*10g
*5g *2g
50g
*10g
50g
*10g
50g
*10g
100g
*10g
Verification unit (e) 5g 10g
20g
50g 50g 50g 100g
Accuracy class III Working temperature -10 ÷ +40°C Tare range -15kg -30kg -60kg
-120kg -150kg -200kg -
300kg
Weighing time <4s Pan dimensions 380x380mm 400x500mm 400x400mm
600x500mm
Scale base dimensions 400x540x100 400x650x125mm 400x550x130
600x650x150mm Scale height 585mm 740mm Power supply 230V, 50Hz, 8VA Internal feeding (option) accumulator 6V/12Ah Accumulator working time ok. 50 godz. Scale weight 13kg 23kg
23kg
21kg
35kg 35kg 35kg
EC Verification
Scale type B33
B63
B151
B152
B301
Maximum loading 30kg
60kg
150kg
150kg
300kg
Minimum loading 200g
400g
1kg
1kg
2kg
Reading unit (d) 10g
*5g *1g
20g
*10g
*5g *2g
50g
*10g
50g
*10g
100g
*10g
Verification unit (e) 10g
20g
50g
50g
100g
Accuracy class III Working temperature -10 ÷ +40°C Tare range -30kg
-60kg
-150kg
-150kg
-300kg Weighing time <4s Pan dimensions 600x500mm 800x800mm 400x400mm 800x800mm Scale base dimensions 600x650x150mm 800x950
x160mm
400x550 x130mm
800x950 x160mm
Scale height 35kg 50kg
21kg
60kg Power supply 740mm Internal feeding (option) 230V, 50Hz, 8VA Accumulator working time accumulator 6V/12Ah Scale weight ok. 50 godz. (opcja) EC Verification
* increased readability for non-EC verification applications
USER MANUAL
5
_____________________________________________________________________________________________________________________
4. Keys and indicators
key I/ - switch-on / switch-off (standby), key
→T←
- tare (subtract package weight from weighed mass),
key
→0←
- zeroing (when the platform is empty),
key
- result printout, key MENU - special function menu, key B/G - gross, key HR - high resolution, indicator
→→→→0←←←←
- zero indicator, indicator
- result stabilization indicator, indicator NET - net weight indicator (indication with subtracted tare)
indicator
- gross mass (after use of B/G key), indicator
- indicator of pieces counting (indications in pieces)
LCD option:
indicator
MODE - special function setting,
bar indicator
- total load indicator (graduated 0-100%) indicator OFF - standby, indicator B/G - gross mass (key B/G), indicator pcs - pieces counting
Max, Min, d, e, III - metrological parameters and accuracy class.
I/
metrological data
6
USER MANUAL
_____________________________________________________________________________________________________________________
5. Security rules
To avoid electrical shock or damage of the scale or
connected peripheral devices, it is necessary to follow
the security rules below.
All repairs and necessary regulations can be made by authorised personnel only.
To avoid fire risk use a feeder of an appropriate type (supplied with the scale). Pay attention that supply voltage is compatible with specified technical data.
Do not use the scale when its cover is opened.
Do not use the scale in explosive conditions.
Do not use the scale in high humidity.
If the scale seems not to operate properly, unplug it from the mains and do
not use until checked by authorised service.
6. Environment protection
According to legal regulations it if forbidden to dispose
wasted electronic equipment in waste containers.
Please return wasted scale to the point of purchase or other company specialised in recycling of wasted electronic components.
USER MANUAL
7
_____________________________________________________________________________________________________________________
7. Preparation
1. Take the scale out of the package removing protective foils.
2. Take a pan off and remove protective elements from below the pan.
3. Place the scale on a stable ground not affected by mechanical vibrations and airflows.
4. Level the scale with the rotating rear legs so that the air bubble in the water-level at the back of the scale is in the middle. Lock the legs with the nut.
5. Put on the pan.
6. Remove everything from a pan and plug a scale to the mains with a ground contact. After self-tests and result stabilisation, the scale displays zero indication.
7. In scales with external feeder connect feeder plug to a connector on a display post. Plug a feeder to the mains (~230V, ground connector is not necessary).
To feed a scale use only socket with a ground connector.
Correct Wrong
8
USER MANUAL
_____________________________________________________________________________________________________________________
8. Start-up
Leave the pan empty, plug a scale to the mains with a ground contact The scale proceeds with following start-up actions:
Display test.
Basic electronic components tests: C1, C-2, ... C-6.
Program version.
The scale is now ready to work.
USER MANUAL
9
_____________________________________________________________________________________________________________________
9. General rules
1. Before each measurement make sure that zero indicator is displayed. If zero
indicator does not displayed or “----“ communicate appears, press 0 key and wait until zero indication and zero indicator appears.
2. The scale is equipped with a tare equal to its range. To tare the scale press
T key (left or right). Storing a tare value does not extend measuring range, but only subtracts it from a load placed on a pan. To make weight control easier and to avoid range overdrawing, the scale is equipped with a load indicator (graduated in percentages).
3. Weighing result should be read when the indicator " " lights, which signalises
stabilisation of a result.
4. When the scale is not used but it is necessary to be ready to work immediately, it
can be switched off by pressing I/ key. The scale reading system is then switched off to "standby" mode (signalled by the indicator "OFF" in version with the LCD display). To switch the scale on press I/ key. The scale is immediately ready to operate maximum accuracy (after self tests).
5. Weighed sample should be placed in the centre of the pan.
6. Protect the scale against dust, aggressive dusts and liquids. To clean the scale
wash it with water with soap and dry it afterwards.
7. A scale equipped with lead-acid accumulator automatically controls accumulator
state, signalises its discharge on LCD display and after around 1h a scale switches itself off to avoid discharging the accumulator below threshold voltage.
Do not discharge an accumulator because it can be damaged.
Place the scale on a platform to avoid dropping weighed objects on the pan.
Do not overload the scale more then 20% of maximum load (Max).
10
USER MANUAL
_____________________________________________________________________________________________________________________
After discharge signalising appears an accumulator should be charged as quickly as possible by connecting external feeder. Charging is more effective when scale is turned off with I/ key, then charging time is about 10 hours. It is also possible to connect accumulator directly to PA6V charger leaving accumulator in a scale or take it out if needed. In order to take accumulator out take off a pan, disconnect accumulator wire and unscrew mounting screws.
Charging an accumulator through scale feeder:
230V
feeder
scale
Charging an accumulator directly through charger:
230V charger
feeder
scale
USER MANUAL
11
_____________________________________________________________________________________________________________________
10. Balance checking
It is advised to check scale indication accuracy before and after series of measurement using any load with known weight. To check the scale with legal verification use a calibration weight with valid calibration certificate. In case permissible error is exceeded it is advised to contact the nearest service to calibrate the scale.
11. Scale adjustment
To adjust a balance it is necessary to break protective marks, therefore to calibrate the scale please contact the nearest service.
12. Connecting a computer or a printer (optional)
The scale is equipped with RS232C or RS485 (optional), which can be used to connect external devices such as a computer or a printer.
When cooperating with a printer data is send automatically after result stabilisation, but next transmission is possible after removing previously weighted sample.
The scale sends following information: successive number of weighing and weighing result (see: printer cooperation mode setting).
The scale may be equipped with the second interface (optional) used for example for continuous weighing result transmission to an additional, external display.
When cooperating with the scale, a computer should be equipped with a program which enables processing data from the scale – available in AXIS with extra charge.
RS485
(option)
RxD (receiver)
TxD (transmitter)
GND
Case
1
2
3
4
5
6
7
8
9
RS-232C
A
Input
Output
A
B
B
Case
1
2
3
4
5
6
7
8
9
}
}
12
USER MANUAL
_____________________________________________________________________________________________________________________
Computer cooperation protocol:
A scale sends data with following settings: Communication parameters: 8 bits, 1 stop bit, no parity, baud rate 4800bps.
Available orders and scale responses: initialising signal (data send order):
ComputerScale: S I CR LF (53h 49h 0Dh 0Ah), ScaleComputer: scale response according to description below (16 bytes),
Description of response bytes:
Byte 1 - sign or space Byte 2 - space Byte
3÷4
- digit or space
Byte 5÷9 - digit, decimal point or space Byte 10 - digit Byte 11 - space Byte 12 - k, l, c, p or space Byte 13 - g, b, t, c or % Byte 14 - space Byte 15 - CR Byte 16 - LF
scale tarring (calling →T← key press):
ComputerScale: S T CR LF (53h 54h 0Dh 0Ah), ScaleComputer: without response,
scale turning on / off (calling I/ key press):
Computer Scale: S S CR LF (53h 53h 0Dh 0Ah), Scale Computer: without response,
scale zeroing (calling →0← key press):
Computer Scale: S Z CR LF (53h 5Ah 0Dh 0Ah), Scale Computer: without response,
entering to special function menu (calling MENU key press):
Computer Scale: S F CR LF (53h 46h 0Dh 0Ah), Scale Computer: without response,
setting low threshold value:
Computer Scale: S L D1...DN CR LF (53h 4Ch D1...DN 0Dh 0Ah) D1...DN – threshold value, maximum 8 characters („-” – negative value, digits, dot – decimal separator), number of digits after dot should be the same as on
USER MANUAL
13
_____________________________________________________________________________________________________________________
scale display, Scale Computer: without response, Example: in order to set low threshold 1000g in scale B1.5 (d=0.5g) the following order should be sent: S L 1 0 0 0 . 0 CR LF (53h 4Ch 31h 30h 30h 30h 2Eh 30h 0Dh 0Ah), in order to set low threshold 100kg in scale B150 (d=50g) the following order should be sent: S L 1 0 0 . 0 0 CR LF (53h 4Ch 31h 30h 30h 2Eh 30h 30h 0Dh 0Ah),
setting high threshold value:
Computer Scale: S H D1...DN CR LF (53h 48h D1...DN 0Dh 0Ah), D1...DN – threshold value (see ) Scale Computer: without response.
Note: Network number different than zero (F..-rS / nr function) changes scale working mode: communication with a computer is possible after logging the scale in with 02h scale_number command. To log the scale out use 03h command.
Connecting cable WK-1 (scale – computer / 9-pin interface):
14
USER MANUAL
_____________________________________________________________________________________________________________________
13. Special functions description
All scales, beside basic functions like weighing and tare, are equipped with the set of additional functions. Basic set of user special functions consists of:
- autozero (AUt),
- pieces counting (PCS),
- constant tare (tAr),
- serial port working mode setting (LPt),
- serial port parameters setting (rS),
- menu customisation (ACt).
Other special functions can be enabled as an option on customer request (all functions are described in additional brochure when ordered).
Press MENU key to enter the function menu. Functions are displayed successively with consecutive numbers: F1-PCS, F2-AUt, etc.
To customise the function menu with most frequently used functions use Menu Customisation function (ACt).
When switching between special functions MODE indicator is displayed. To make clear how to manage with each function, in further part of instruction
descriptions are replaced with pictures. Every time the picture of a hand appears, press the indicated key.
Legend:
- put a load on the pan
- remove the load from the pan
-
press the key when indication is displayed
- forced change
- automatic change
2.30
T
USER MANUAL
15
_____________________________________________________________________________________________________________________
14. Normal weighing
→0←
key, which zeros the scale,
operates only when the pan is empty.
Weighing result should be read when the indicator " " lights.
15. Weighing with tare
The scale is equipped with tare equal to its range.
To display gross weight press B/G key.
Note:
Press B/G key to return to net weight indication..
16. Increased readability
Press HR key to display the weighing result (for 5s.) with the highest readability possible. This function is especially helpful in scales with legal verification with d=e.
The weighing result with increased readability can be used for informational purposes only and cannot be printed or sent to a computer with key.
0.00
2.30
0
2.30
2.29
2.30
2.30
0.00
40.01
42.31
T
NET
NET
16
USER MANUAL
_____________________________________________________________________________________________________________________
17. Autozero (F..-AUt)
When F..-Aut function is activated, the scale automatically ensures stable zero indication if the pan is empty or if zero indication was acquired by pressing →T← key. To leave the function press MENU
key, then with →T← key chose F..-AUt and F..-0.
Note: Autozeroing function is activated automatically for 10 min. after switching-on.
0.01
F.. -AUT
F.. -1
F.. -0
0.00
T
T
USER MANUAL
17
_____________________________________________________________________________________________________________________
18. Pieces counting (F..-PCS)
This function enables to count identical pieces, e.g. turnbuckles or buttons. A measurement is performed in two phases:
- first phase - single piece weight calculation on the basis of defined pieces amount (5, 10, 20, 50, 100, 200 or 500 pieces),
- second phase – pieces counting.
It is advised that single piece weight is not less than one reading unit and sample weight used in first phase is bigger than 100 reading units.
To leave function press MENU key and then with
→Τ←
key chose
F..-PCS and F..-0.
Notes:
1. Err-3 communicate signalises that a sample was not put on the pan. The same communicate appears if single piece weight is less than one reading unit (it is possible to count pieces but measuring error is bigger).
2. To chose previously used pieces amount select “..” in first phase (in case no value was chosen, error communicate appears).
3. During pieces counting →T← key function does not change.
4.In scales equipped with LCD display, weighing unit is visible and "
" sign is replaced with
"pcs".
0
0.00
23.40
F.. -PCS
F.. -0
F.. -1
F.. - --
5
5
12
12
T
T
T
T
18
USER MANUAL
_____________________________________________________________________________________________________________________
19. Printer cooperation settings (F..-LPt)
Activate the function for automatic serial port working mode (cooperation with a printer). After activation the scale prints a header. Weighing result with a successive measurement number is printed automatically after result stabilisation (without using key).
To select computer cooperation mode ( key activated and weighing results without successive numbers) press MENU key, then with →T← key chose F..-LPt and F..-0.
0.00
F.. -LPt
F.. -1
F.. -0
0.00
T
T
USER MANUAL
19
_____________________________________________________________________________________________________________________
20. Serial port parameters setting (F..-rS)
The function enables to set the following transmission parameters (standard parameters underlined):
- transmission speed (bAud: 1200, 4800, 9600),
- the number of bits in a byte (bit: 7,
8),
- parity control (PArItY: 0, 1; Odd: 0,
1),
- network number when working in multistand computer system (when working as a single scale the value should be “0”)
- continuous transmission – without using key, approx. 10 results per second (SEnd: 0, 1).
Default parameters underlined.
To set desired transmission parameters activate F.._rS function, choose appropriate parameter and press →T← key to accept needed parameter value. The example at the left presents how to set transmission speed value to 9600bps.
To leave the function choose out option.
0.00
F.. -rS
F.. -1
F.. -0
baud
-9600
-4800
bIt
nr
out
Odd
PArItY
Send
T
T
T
T
T
T
T
T
T
T
20
USER MANUAL
_____________________________________________________________________________________________________________________
21. Constant tare (F..-tAr)
This function enables to measure gross weight of a sample placed in a container of a known weigh value (stored in the memory) and to display calculated net weight of the sample. Tare value is recalled from the memory with →0← key when the pan is empty. Tare value may be entered using the keypad or by sampling container weight from the pan.
Operation sequence:
The following options are possible:
- F-0 – leave the function,
- F-1 – activate the function with the previous tare value,
- F-2 – sample tare value from the pan,
- F-3 – enter tare value with keys:
→0←
,
,
→T←
and MENU,
If the function is active, NET indicator is displayed.
Options F-1 enables to activate the function with previous tare value after leaving the function with F-0 option.
Note:
Tare value is stored in memory also after unplugging the scale from the mains.
-
1.7
0.00
F.. -tAr
1.
1
F.. -3
1.0
T
T
T
NET
...
...
1.70
USER MANUAL
21
_____________________________________________________________________________________________________________________
Weighing with constant tare:
When tAr function is activated, press
→0←
key to zero the indication and to recall tare value from the memory. Tare value is displayed with ”-” sign.
B/G key enables for instant switching between net and gross weight.
Note:
When the pan is empty →T← key does not operate – to tare the scale use →0← key.
-1.70
13.70
10.00
0
B/G
NET
.
22
USER MANUAL
_____________________________________________________________________________________________________________________
22. Function Menu customisation (F..-ACt)
This function enables to select special functions that will be displayed after pressing MENU key. Easy access to the most useful functions will shorten operation time and make work more comfortable.
Operation sequence shown on the picture, presents how to add RS232C parameters setting function (F..-rS) to the Function Menu.
To remove a function from the Function Menu choose F..-0 in the last operation.
0.00
F.. -ACt
F.. -1
F.. -0
F..-AUt
F..-rS
T
T
T
USER MANUAL
23
_____________________________________________________________________________________________________________________
23. Maintenance and repairs of small defects
1. The scale should be kept clean.
2. Take care that no dirt gets between the platform and the scale base. If found any, remove the pan (lift it up), remove dirt and then replace the pan.
3. In case of improper operation caused by short-lasting power supply decay, unplug the scale from the mains and then plug it again after few seconds.
4. If the scale is switched on with empty pan and “Err-b” communicate appears, the load cell has been mechanically damaged.
5. It is forbidden to make any repairs by unauthorised persons.
6. To repair the scale, please contact our nearest service.
Error communicates:
Communicate
Possible cause Remedy
C-1 ... 6
(over 1 min.)
selftests failed
if displayed more than 1 minute, contact an authorised service
the scale was switched on with loaded pan remove a load from the pan mechanical damage of the load cell contact an authorised service
L
pan missing
put the pan on
mechanical damage
contact an authorised service
H overloading
remove the load from the pan
mechanical damage
contact an authorised service
unstable ground vibrations air flows
place the scale on a stable ground not affected by mechanical vibrations and airflows
indicator does
not appear
scale is damaged
contact an authorised service
- - - - - -
taring in progress
as above
- - taring could not be finished (the load is too small or B\G key was used)
zero the scale or press B\G key again
-
-
the load is too big to be zeroed
tare the scale (T)
24
USER MANUAL
_____________________________________________________________________________________________________________________
Declaration of Conformity
We:
AXIS Spółka z o.o. 80-125 Gdańsk, ul. Kartuska 375B
confirm with all responsibility that scales:
B15W, B30, B60, B120, B150, B200 i B300
oraz B33, B63, B151, B152, B301
marked with CE mark comply with the following:
1. EN 61010-1 standard Safety requirements for electrical equipment for measurement, control and laboratory use. General requirements harmonized with the directive 73/23/EEC (Low Voltage Directive).
2. EN 55022:2000 standard Limits and methods of measurement of radio disturbance characteristics of information technology equipment and IEC 61000-4-3 Electromagnetic compatibility (EMC) - Part 4-3: Testing and measurement techniques - Radiated, radio­frequency, electromagnetic field immunity test, harmonised with the Council Directive 89/336/EEC.
Additionally scales with the following markings on the name plate:
- a sticker with two-digit number of the year in which the mark was affixed and the number of the Notified Body responsible for EC verification
- a green metrology sticker with “M” mark,
- a protective seal affixed by the Notified Body
comply with requirements stipulated on the EC Type-Approval Certificate No. PL 04 022 and was verified by Notified Body No. 1440 to comply with:
3. EN 45501 Metrological aspects of non-automatic weighing instruments harmonised with the Council Directive 90/384/EEC amended with 93/68/EEC.
Additional information
- Conformity evaluation for the Council Directive 73/23/EEC and 89/336/EEC were carried out
by Laboratorium Badawcze Oddziału Instytutu Elektrotechniki in Gdańsk, accredited by PCA
- EC Type-Approval Certificate No. PL 04 022 was issued by Główny Urząd Miar in Warsaw
(Notified Body no. 1440).
Gdańsk, 5.09.2006 r.
Per pro Director of AXIS Sp. z o.o.:
Production Manager Jan Kończak _________________
Signature
xx