AXEON R2 User Manual

Reverse Osmosis
User’s Manual
Model
R2-1140, R2-2140, R2-3140,
R2-4140, R2-5140, R2-6140
R2-Series User’s Manual 11/13/12 MKTF - 212
1
This Page Intentionally
Left Blank
TABLE OF CONTENTS
R2-Series User’s Manual 11/13/12 MKTF - 212
2
INTRODUCTION...................................................................................................................................4
SAFETY................................................................................................................................................ 4
FEED WATER AND OPERATION SPECIFICATIONS .................................................................................. 5
REJECTION, RECOVERY AND FLOW RATES............................................................................................ 5
SYSTEM REQUIREMENTS AND OPERATION GUIDELINES.......................................................................6
MEMBRANE ELEMENTS ....................................................................................................................... 8
R2-1140, R2-2140, R2-3140, R2-4140, R2-5140, R2-6140 ................................................................... 13
SYSTEM IDENTIFICATION ................................................................................................................... 13
SYSTEM PURGING.............................................................................................................................. 18
INITIAL START-UP............................................................................................................................... 18
OPERATING DO’s AND DON’Ts........................................................................................................... 20
OPERATION AND MAINTENANCE....................................................................................................... 20
MEMBRANE REMOVAL AND REPLACEMENT ...................................................................................... 22
FLUSHING THE SYSTEM...................................................................................................................... 25
PREPARING UNIT FOR STORAGE OR SHIPMENT.................................................................................. 25
REVERSE OSMOSIS TROUBLESHOOTING ............................................................................................ 27
TEMPERATURE CORRECTION FACTORS FOR MEMBRANE................................................................... 29
OPERATION....................................................................................................................................... 31
DRAWINGS ........................................................................................................................................ 32
R2-Series User’s Manual 11/13/12 MKTF - 212
3
Series User’s Manual
Reverse Osmosis S
This User’s Manual
details vital to the sustained performance of your system.
The test results which are included with this Use
If your system is altered at the site of operation
your local dealer or distributor to determine the proper
NOTE: IN ORDER TO MAINTAIN
MUST BE MAINTAINED AND COPIES WILL NEED TO BE SENT TO YOUR LOCAL DEALER OR
OPERATING OR SERVICING THE
MANUAL MUST BE READ AND FULLY UNDERSTOOD. KEEP THIS AND OTHER ASSOCIATED
INFORMATION FOR FUTURE REFERENCE AND FOR NEW OPERATORS OR QUALIFIED
PERSONNEL NEAR THE SYSTEM
afety section of this User’s Manual
manual’s text and are enhanced and defined below:
NOTE: INDICATES STATEMENTS THAT PROVIDE FURTHER INFORMATION AND CLARIFICATION.
CAUTION: INDICATES STATEMENTS THAT ARE USED TO IDENTIFY CONDITIO
PRACTICES THAT COULD RESULT IN EQUIPMENT OR OTHER PROPERTY DAMAGE.
WARNING: INDICATES STATEMENTS THAT ARE USED TO IDENTIFY CONDITIONS OR
PRACTICES THAT COULD RE
WARNINGS COULD RESULT IN SERIOUS
DO NOT UNDER ANY CIRCUMSTANCE; REMOVE ANY CA
DESCRIPTIVE LABELS FROM THE SYSTEM.
INTRODUCTION
equipment which, with proper c
, maintenance, and troubleshooting
r’s Manual indicate your system’s permeate (product) and
or if the feed water conditions
recovery for your application.
WARRANTY, AN OPERATING LOG
REVERSE OSMOSIS SYSTEM, THIS USERS
outlines the various safety headings used throughout this
FAILURE TO FOLLOW
TION, WARNING, OR OTHER
Your R2-Series last for many years.
concentrate (waste) test results.
DISTRIBUTOR FOR REVIEW.
NOTE: PRIOR TO
The s
ystem is a durable piece of
outlines installation, operation
THE MANUFACTURER’S
.
SAFETY
are, will
change, please contact
R2­MKTF - 212
NS OR
SULT IN INJURY OR LOSS OF LIFE.
INJURY OR EVEN DEATH.
U
4
11/13/12
FEED WATER AND OPERATION SPECIFICATIONS
Nothing has a greater effect on a reverse osmosis system than the feed water quality
NOTE: IT IS VERY IMPORTANT TO MEET THE MINIMUM FEED WATER REQUIREMENTS. FAILURE TO DO SO WILL CAUSE THE MEMBRANES TO FOUL AND VOID THE MANUFACTURER’S WARRANTY.
NOTE: HIGHER TDS AND/OR LOWER FEED WATER TEMPERATURES WILL REDUCE THE SYSTEM’S PRODUCTION.
REJECTION, RECOVERY AND FLOW RATES
R2-Series Reverse Osmosis Systems are designed to produce permeate water at the capacities indicated in the design basis section” on page 19 of this manual. For example, the R2-6140 produces
6.25 gallons per minute of permeate water at the listed operating test conditions. The amount of total dissolved solids (TDS) rejected by the membrane is expressed as a percentage. For
example, a 98.5% rejection rate means that 98.5% of total dissolved solids do not pass through the membrane. To calculate the % rejection, use the following formula:
% Rejection = [(Feed TDS – Product TDS) / Feed TDS] x 100
Example:
99.2% = [(10,000-80)/10,000] x 100
NOTE: ALL TDS FIGURES MUST BE EXPRESSED IN THE SAME UNITS, TYPICALLY PARTS PER MILLION (PPM) OR MILLIGRAMS PER LITER (MG/L).
R2-Series Reverse Osmosis Systems are designed to reject up to 99.2% NaCl, unless computer projections have been provided or stated otherwise.
R2-Series User’s Manual 11/13/12 MKTF - 212
5
Series User’s Manual
The amount of permeate
water recovered for us
following formula
% Recovery = (Product Water Flow Rate / Feed Water Flow Rate) x 100
% Rejection = (Feed TDS
NOTE: ALL FLOW RATES MUST BE EXPRESSED IN THE SAME UNITS
SYSTEM REQUIREMENTS
gh pressure pumps used on R
us flow of water with a minimum
FEED WATER CONNECTION
FNPT Sole
tach the inlet piping to the 1
PERMEATE (PRODUCT WATER) CONNECTION
1” or 3/4” connection (d
sure that the permeate water can flow freely
damage to the membrane elements.
with PVC fittings, or other FDA approved materials
into the permeate water.
H OF THE REVERSE O
LOWER THAN THE FEED WATER PH. A LOW PH CAN
PLUMBING MATERIALS SUCH AS COPPER PIPING.
CONCENTRATE (WASTE WATER)
e is expressed as a percentage. To calculate % recovery
x 100
OPERATION GUIDELINES
Reverse Osmosis Sy
exceed 85
labeled permeate and attach to
Backpressure can cause
epending on the unit)
This is so
SMOSIS PERMEATE WATER WILL TYPICA
and % rejection, use the
PLUMBING
s:
Example:
25% = (1.04/4.04) x 100
– Product TDS)/(Feed TDS)
Example:
99.2% = [(10,000-80)/10,000] x 100
.
AND
The membranes and hi continuo
1. Locate the 1 (Figure 1A, Page 13)
2. At
Locate the En irreversible be run to the holding tank being used does not leach
CAUTION: THE P
BE 1-2 PH UNITS
2-Series
feed pressure of 45 psi, not to
noid Valve feed water inlet.
” FNPT Solenoid Valve feed water inlet.
epending on the unit)
with no backpressure.
The 1” or 3/4” (d
stems require a
°F.
.
DAMAGE CERTAIN
the storage tank.
permeate line can
that the material
LLY
CONNECTION
R2­MKTF - 212
6
11/13/12
Series User’s Manual
Locate the
1” or 3/4” connection (D
to an open drain in a free and unrestricted manner (no backpressure).
break be used on the concentrate line to prevent siphoning of water from the pressure vessels
CAUTION: ANY RESTRICTIONS OR BLOCKAGE IN THE DRAIN LINE CAN CAUSE
BACKPRESSURE, WHICH WILL INCREASE THE SYSTEMS OPERATING PRESSURE. THIS CAN
RESULT IN DAMAGE TO THE SYSTEMS MEMBRANES AND COMPONENTS.
pump and
system is equipped with a 5 foot electrical cord
nsure that the electrical circuit supplying the system is compatible with the re
ED
ACCORDANCE WITH LOCAL AND NATIONAL ELECTRICAL CODES (NEC).
WARNING: TO REDUCE THE RISK OF ELECTRICAL SHOCK, THE INCOMING POWER
PROTECTIVE EARTH GROUND.
systems are typically controlled with a liquid level switch in a storage tank. The liquid level
switch turns the system on when the water level in the tank drops, and off when the tank is full
level switches can be obtained by your local dealer or distributor. If a liquid level switch is to be used,
systems are supplied with a
ists between the pre
NOTE: THE SYSTEM MUST BE OPERATED
Series systems is a
Follow these guidelines to ensure proper operation of the pump:
feed the pump with filtered water. The pump is susceptible to damage from sediment
and
50/60 Hertz
uirements of the specific
THAT A LICENSED ELECTRICIAN WIRE YOUR SYSTEM IN
. Change the cartridge
r or distributor about Pre
IN ACCORDANCE TO FEED WATER SPECIFICATIONS.
steel
be run dry. Operating the pump without sufficient feed water will damage
uirements of the specific
the concentrate line that an air­when the system is in standby.
ELECTRICAL
epending on the unit) labeled concentrate
attach to a drain. Run
It is advised
The R2-Series systems Each R2-Series
E R2 model you are installing.
NOTE: IT’S RECOMMEND
SUPPLY MUST INCLUDE A
R2-Series
install it at this time.
PRE-FILTRATION
R2-Series when a 10-15 psi differential ex Filtration systems, if required.
motor are available in 220/460 Volt,
.
5 micron sediment filter
-filter. Ask your local deale
, 1 Phase/3 Phase.
q
. Liquid
once a month or
-
PUMP
The pump used on the R2-
The pump must NEVER
the pump.
ALWAYS
and debris.
R2­MKTF - 212
multi-stage centrifugal stainless
7
type.
11/13/12
If any damage occurs to your system’s pump, a re-build kit is available. Contact your local dealer
or distributor and inform them of your system’s model and pump size.
MOUNTING
The free standing system should be bolted down in compliance with local regulation standards or securely fastened.
MEMBRANE ELEMENTS
R2-Series reverse osmosis systems come pre-loaded with FILMTEC® LC LE membranes, unless otherwise specified. General membrane element performance characteristics are listed on the following pages.
LC LE-4040-STANDARD
R2-Series User’s Manual 11/13/12 MKTF - 212
8
R2-Series User’s Manual 11/13/12
9
MKTF - 212
LC HR-4040-OPTIONAL
R2-Series User’s Manual 11/13/12 MKTF - 212
10
R2-Series User’s Manual 11/13/12
11
MKTF - 212
R2-Series User’s Manual 11/13/12
12
MKTF - 212
R2-1140, R2-2140, R2-3140, R2-4140, R2-5140, R2-6140
SYSTEM IDENTIFICATION
FIGURE 1A
R2-Series User’s Manual 11/13/12 MKTF - 212
13
NUMBER IDENTIFICATION
1. SOLENOID VALVE – TURNS FEED WATER ON/OFF
2. 5 MICRON SEDIMENT – REDUCES PARTICULATES
3. PRESSURE GAUGE – MEASURES FEED PRESSURE
4. PRESSURE GAUGE – MEASURES PRESSURE AFTER FILTER
5. PRESSURE GAUGE – MEASURES PUMP PRESSURE
6. PRESSURE GAUGE – MEASURES CONCENTRATE PRESSURE
7. COMPUTER CONTROLLER – CONTROLS RO SYSTEM FUNCTIONS
8. CONCENTRATE VALVE– CONTROLS FLOW OF CONCENTRATE (WASTE) WATER
9. RO PUMP AND MOTOR – PRESSURIZES RO SYSTEM
10. THROTTLE VALVE – CONTROLS FLOW OF PUMP
11. FLOW METER – MEASURES FLOW OF PERMEATE WATER
12. FLOW METER – MEASURES FLOW OF CONCENTRATE (WASTE) WATER
13. LOW PRESSURE SWITCH TURNS OFF RO PUMP WHEN FEED PRESSURE FALLS BELOW SET POINT
14. PERMEATE CHECK VALVE PROTECTS MEMBRANE ELEMENTS FROM BACKPRESSURE
15. PRESSURE VESSELS – HOUSES MEMBRANE ELEMENTS
16. HIGH PRESSURE SWITCH – TURNS THE PUMP OFF AT A PRE-DETERMINED HIGH PSI PUMP
PRESSURE
17. PERMEATE SAMPLE VALVE- MEASURES THE QUALITY OF EACH MEMBRANE
18. CLEANING PORT VALVE- PROVIDES ACESS TO CLEAN THE MEMBRANES
R2-Series User’s Manual 11/13/12 MKTF - 212
14
FIGURE 1B
FIGURE 1C
15
R2-Series User’s Manual 11/13/12 MKTF - 212
FIGURE 1D
FIGURE 1E
16
R2-Series User’s Manual 11/13/12 MKTF - 212
FIGURE 1F
Note: A portion of the frame has been removed to expose components.
R2-Series User’s Manual 11/13/12
17
MKTF - 212
SYSTEM PURGING
Carefully inspect your system before initial start-up. Check that all plumbing and electrical connections are not loose or have not come undone during shipment. A User’s Manual, Test Results, and Filter Housing Wrench will accompany your R2-Series Reverse Osmosis System.
NOTE: LEAVE THE POWER TO THE SYSTEM OFF FOR THIS PROCEDURE.
1. Redirect permeate water to the drain for this procedure.
2. Fully open the concentrate valve (Counter Clockwise). (Figure 1B, Page. 15)
3. Fully open the throttle valve (Counter Clockwise). (Figure 1B, Page. 15)
4. Press and hold the reset button on the controller for three seconds to activave the solenoid bypass feature.
INITIAL START-UP
1. Maintain the permeate water line to drain for this
procedure.
2. Fully open the concentrate valve (Counter Clockwise).
(Figure 1B, Page. 15)
3. Adjust the throttle valve at 50% open (Figure 1B,
Page. 15)
4. Turn the RO system on and adjust the concentrate (waste) valve and the throttle valve to the specified system’s production, recovery and operating pressure. (Figure 1B, Page. 15)
5. Inspect the system for leaks.
6. Allow the system to run 30 minutes to flush the preservative solution from the system.
7. After 30 minutes, shut down the system.
8. Re-direct the permeate water back to the tank or point-of-use.
9. Record the readings daily for a week; after one week record the readings regularly.
R2-Series User’s Manual 11/13/12 MKTF - 212
18
DESIGN BASIS FOR R2-1140, R2-2140, R2-3140,
Specifications
R2-4140, R2-5140, R2-6140
WARNING: NEVER EXCEED THE SYSTEM’S MAXIMUM PRESSURE RATING.
R2-Series User’s Manual 11/13/12 MKTF - 212
19
OPERATING DOs AND DON’Ts
DO:
Change the cartridge filters regularly Monitor the system and keep a daily log Adjust the system recovery to the recommended value Always feed the pump with filtered water
DON’T:
Permit chlorine to be present in the feed water Shut down the system for extended periods without preservation Close the control valves completely Operate the system with insufficient feed flow Operate the pump dry
OPERATION AND MAINTENANCE
The reverse osmosis process causes the concentration of impurities. The impurities may precipitate (fall out of solution) when their concentration reaches saturation levels.
NOTE: PRECIPITATION CAN SCALE OR FOUL MEMBRANES AND MUST BE PREVENTED.
Check your feed water chemistry and pre-treatment. If necessary, consult with your local dealer or distributor.
PRE-FILTER PRESSURE GAUGES
Pre-filter gauges measure the feed water pressure when it enters and exits the pre-filter. A pressure differential of 10 - 15 psi or more on the two pressure gauges indicates that the pre-filters require servicing.
PUMP PRESSURE AND CONCENTRATE PRESSURE GAUGES
Pump and concentrate gauges measure the pressure of water exiting the multistage pump and the pressure of concentrate water as it exits the pressure vessel array. Comparison of the pump output and concentrate pressures allows for the establishment of a baseline pressure differential. If the pressure differential increases over time from this baseline, it would be an indication that the reverse osmosis membranes need to be inspected.
R2-Series User’s Manual 11/13/12 MKTF - 212
20
PERMEATE (PRODUCT) FLOW METER AND CONCENTRATE (WASTE) FLOW METER
The permeate and concentrate flow meters indicate the flow rates of the permeate and concentrate water. The measurements, when added together, also indicate the feed water flow rate or total flow rate.
LOW PRESSURE SWITCH
The low pressure switch shuts off the system when the feed water pressure drops below 5 PSI, preventing damage to the pump. The system restarts automatically when there is a constant pressure of 35 PSI or more.
PUMP THROTTLE VALVE
The Pump Throttle Valve is installed as a standard feature on the R2-Series reverse osmosis systems. It provides an adjustment for pump pressure. As the feed water temperature decreases, and/or the feed water TDS increases, the system will require a higher operating pressure to produce the specified permeate flow.
ADJUSTING THE THROTTLE VALVE
To decrease the pressure, turn the handle clockwise. To increase the pressure turn the handle counter clockwise. (Figure 3, Page 21)
FIGURE 3
R2-Series User’s Manual 11/13/12 MKTF - 212
21
MEMBRANE REMOVAL AND REPLACEMENT
Replacing membranes in the pressure vessels is an easy process if you have the proper information and tools at hand. Please refer to the following instructions when removing and replacing membrane elements:
WARNING: ALL PRESSURE GAUGES MUST READ ZERO BEFORE PROCEEDING. BEFORE ATTEMPTING, DISCONNECT THE POWER FROM THE SYSTEM AND BLEED ALL WATER PRESSURE FROM THE SYSTEM.
1. Remove the end plugs from the top of the pressure vessels. This is done by removing the two half-moon retaining disks using a #5 Allen wrench; the end plugs should then freely slide out of the pressure vessel.
2. Remove the replacement membrane element(s) from the shipping box; the membrane(s) should be contained within a plastic oxygen barrier bag.
NOTE: WEAR GLOVES FOR THE FOLLOWING STEPS IN ORDER NOT TO CONTAMINATE THE MEMBRANE.
3. Cut the bag open as close as possible to the seal at one end of the bag, so the bag may be re­used if necessary.
4. Make sure that all parts are clean and free from dirt. Examine the brine seal, and permeate tube for nicks or cuts. Replace the O-rings or brine seal if damaged.
5. Flow directions should be observed for installation of each element into their respective pressure vessels.
R2-Series User’s Manual 11/13/12 MKTF - 212
22
Series User’s Manual
REPLACING THE MEMBRANE ELEMENT:
WARNING: THE BRINE SEAL MUST BE IN THE SAME POSITION FOR EACH MEMBRANE
ELEMENT HOUSING, SO MARK EACH HOUSING PRIOR TO REMOVING THE MEMBRANE
ELEMENTS. THE BRINE SEAL IS A RUBBER SEAL THAT PROTRUDES ON ONE SIDE OF THE
MEMBRANE AND IS ALWAYS ON THE FEED SIDE
1.
Remove one membrane element at a time from the pressure vessels, from the top of each
housing. Long nose pliers may be necessary to pull the old membrane element out of the
membrane element housing.
the brine seal with a non
Install membranes with brine seal location depicted in (Figure 4, Page 27).
For brine seals that are on the bottom: At a slight angle insert the membrane while slightly
ng the element being careful not to tear or flip the brine seal. A slow twisting motion should
be used to insert the membrane element, for every couple of turns pull up ½ to make sure the
With a smooth and constant motion, push t
seal enters the housing without coming out of the brine seal groove.
install the end plugs by gently twisting the end cap while pushing it onto the housing. Ensure
that you do not pinch or fatigue any
until the outer diameter of the plug is flush with the outer diameter of the pressure vessel.
moon retaining disks until they are fully seated. Subsequently fasten using a
Reconnect any fittings that may have been disconnected when the membrane pressure vessels
Up the system, please refer to the Initial Start
CAUTION: WET MEMB
T BE FLUSHED FOR AT LEAST 30 MINUTES
FROM THE MEMBRANE. DISCARD ALL OF THE PERMEAT
PRODUCED DURING THE FLUSH PERIOD.
OF THE MEMBRANE ELEMENT.
petroleum based lubricant, such as Dow Corning® 111.
he membrane element into the housing so the brine
installing the end plug. Push the end plug on
ion of this manual. (See page 18
RANES ARE SHIPPED IN A PRESERVATIVE SOLUTION. THE
TO REMOVE THE PRESERVATIVE
E AND CONCENTRATE
2. Lubricate
3.
4. rotati
brine seal doesn’t flip.
5.
6. Re-
7. Insert the two half- #5 Allen wrench.
8. were disassembled.
9. To Start-
-
O-rings while re-
-Up sect
)
MEMBRANES MUS
23
R2­MKTF - 212
, WHICH IS
11/13/12
FIGURE 4
View from the back of R2-1140, R2-2140, R2-3140, R2-4140, R2-5140, R2-6140 reverse osmosis system.
R2-Series User’s Manual 11/13/12
24
MKTF - 212
FLUSHING THE SYSTEM
The system should be flushed weekly to remove sediment from the surface of the membranes. To manually flush the system, follow the preceding steps:
1. The system must be operating during the flush procedure.
2. Fully open the concentrate valve. (Figure 1B, Page. 15)
3. Allow the system to run for 10 to 20 minutes.
4. After 10 to 20 minutes, close the concentrate valve to its previous setting. Ensure the proper concentrate flow rate is going to the drain.
5. The system is now ready to operate.
PREPARING UNIT FOR STORAGE OR SHIPMENT
PRIOR TO SHIPPING OR STORING YOUR SYSTEM, THE SYSTEM SHOULD BE CLEANED WITH AN APPROPRIATE CLEANER, FLUSHED WITH WATER, AND PROTECTED FROM BIOLOGICAL ATTACK WITH AN APPROPRIATE SOLUTION FOR MEMBRANE ELEMENTS. THE MEMBRANE HOUSING(S) AND PLUMBING LINES OF THE SYSTEM MUST BE COMPLETELY DRAINED. ANY WATER REMAINING IN THE PLUMBING OF A SYSTEM MAY FREEZE, CAUSING SERIOUS DAMAGE.
PREPARING SYSTEM FOR STORAGE:
1. Totally immerse the elements in the membrane housing in a solution of 2% membrane preservative, venting the air outside of the pressure vessels. Use the overflow technique: circulate the Memstor solution in such a way that the remaining air in the system is minimized after the recirculation is completed. After the pressure vessel is filled, the Memstor solution should be allowed to overflow through an opening located higher than the upper end of the highest pressure vessel being filled.
2. Separate the preservation solution from the air outside by closing all valves.
3. Repeat this process at least once a month.
During the shutdown period, the plant must be kept frost-free, or the temperature must not exceed 113°F (45°C).
PREPARING UNIT FOR SHIPMENT:
4. Disconnect the inlet, concentrate, pre-filter, and permeate plumbing.
5. Drain all water from the pre-filter cartridge housings by unscrewing the housings, removing the pre-filter cartridges, and drain the water from the housings.
6. Disconnect the tubing from the connectors on the permeate and concentrate inlets and outlets.
R2-Series User’s Manual 11/13/12 MKTF - 212
25
7. Fully open the concentrate valve.
8. Drain the flow meters.
9. Allow the system to drain for a minimum of eight hours or until the opened ports quit dripping.
10. After draining is complete, reconnect all of the plumbing.
R2-Series User’s Manual 11/13/12
26
MKTF - 212
REVERSE OSMOSIS TROUBLESHOOTING
SYMPTOMS POSSIBLE CAUSES CORRECTIVE ACTION
LOW INLET PRESSURE
LOW PERMEATE FLOW
HIGH PERMEATE FLOW
POOR PERMEATE QUALITY
MEMBRANE FOULING
Low inlet pressure Increase inlet pressure
Cartridge filters plugged Change filters
Solenoid valve malfunction Replace sol. valve and/or coil
Concentrate valve might be
damaged
Leaks Fix any visible leaks
Cold feed water See temperature correction sheet
Low operating pressure
Defective membrane brine seal/
membrane installed backwards
Fouled or scaled membrane Clean membranes
Damaged product tube o-rings Inspect and/or replace
Damaged or oxidized membrane Replace membrane
Exceeding maximum feed water
temperature
Low operating pressure Adjust concentrate valve
Damage product tube o-rings Inspect and/or replace
Damaged or oxidized membrane Replace membrane
Metal Oxide Fouling
Colloidal Fouling
Scaling (CaSO4, CaSO3,
BaSO4, SiO2)
Biological Fouling
Replace needle valve
Adjust throttle and concentrate
valve
Replace brine seal and / or
reposition membranes
See temperature correction sheet
Improve pretreatment to remove
metals. Clean with Acid
Cleaners.
Optimize pretreatment for colloid
removal. Clean with high pH
anionic cleaners.
Increase acid addition and
antiscalant dosage for CaVO3
and CaCO4. Reduce recovery.
Clean with Acid Cleaners
Shock dosage of Sodium Bi-
Sulfate. Continuous feed of
Sodium Bi-Sulfate at reduced pH.
Chlorination and de-chlorination.
Replace cartridge filters.
Activated Carbon or other
Organic Fouling
Chlorine Oxidation
Abrasion of membrane by
Crystalline Material
27
R2-Series User’s Manual 11/13/12 MKTF - 212
pretreatment. Clean with high pH
cleaner.
Check Chlorine feed equipment
and de-chlorination system.
Improve pretreatment. Check all
filters for media leakage.
ABNORMAL PERMEATE FLOW
As time progresses, the efficiency of the membrane will be reduced. In general, the salt rejection does not change significantly until two or three years after installation when operated on properly pretreated feed water. The permeate flow rate will begin to decline slightly after one year of operation, but can be extended with diligent flushing and cleaning of the system. A high pH and/or precipitation of hardness can cause premature loss in rejection.
Permeate flow should be within 20% of the rated production, after correcting the feed water temperatures above or below 77°F. Check your permeate flow meter to determine the permeate flow rate.
NOTE: TO DETERMINE THE TEMPERATURE CORRECTION FACTOR, LOCATE THE TEMPERATURE CORRECTION TABLE ON THE FOLLOWING PAGE.
R2-Series User’s Manual 11/13/12 MKTF - 212
28
TEMPERATURE CORRECTION FACTORS FOR MEMBRANE
Find the temperature correction factor (TCF) from the table below. Divide the rated permeate flow at 77°F by the temperature correction factor. The result is the permeate flow at the desired temperature. (See
example on the next page). If a system is rated to produce 5 gpm of permeate water @ 77˚ F. The same
system will produce more water at a higher temperature. It will also produce less water at a lower temperature. Use the temperature correction table to obtain the correct flow.
Example:
R2-Series User’s Manual 11/13/12 MKTF - 212
29
5 gpm @ 59˚ F (5÷1.42=3.52 gpm)
5 gpm @ 77˚ F (5÷1=5 gpm)
5 gpm @ 84˚ F (5÷0.89=5.62 gpm)
SERVICE ASSISTANCE
If service assistance is required, please complete the following process: Contact your local dealer or distributor. Prior to making the call, have the following information available:
system installation date, serial number, daily log sheets, current operating parameters (e.g. flow, operating pressures, pH, etc.), and a detailed description of the problem.
R2-Series User’s Manual 11/13/12 MKTF - 212
30
OPERATION
Date of Start
-
Company:
Location:
Week Of:
System Serial #: ____________________
FILTER INLET PRESSURE (PSI)
FILTER OUTLET PRESSURE (PSI)
CONCENTRATE PRESSURE (PSI)
PUMP DISCHARGE PRESSURE (PSI)
PERMEATE FLOW (GPM)
____________________
____________________
____________________
DATE
TIME
HOUR OF OPERATION
FEED FLOW (GPM)
Up: ___________________
Date of Last Cleaning: ___________________
CONCENTRATE FLOW (GPM)
RECYCLE FLOW (GPM)
RECOVERY %
FEED TEMPERATURE
FEED TDS (PPM)
PERMEATE TDS (PPM)
REJECTION %
FEED PH
PERMEATE PH
SCALE INHIBITOR FEED (PPM)
IRON (mg/L)
FREE CHLORINE (mg/L)
HARDNESS (GPG CaCO3)
R2-Series User’s Manual 11/13/12
31
MKTF - 212
DRAWINGS
FIGURE 5
R2-Series User’s Manual 11/13/12 MKTF - 212
32
FIGURE 6
R2-Series User’s Manual 11/13/12
33
MKTF - 212
FIGURE 7
Note: A portion of the frame has been removed to expose components.
R2-Series User’s Manual 11/13/12
34
MKTF - 212
R2-1140 SYSTEM PART LIST
3SV20PA30/V09722
Item No. Quantity Part Number Description
1 1 206420 2 1 204914 VALVE, SOLENOID, N/C, UL, 220V, 1” FNPT, 3 METER CORD 3 1 205098 SWITCH, PRESSURE, HIGH, N/C 450PSI, ¼” MNPT, NASON 4 1 205903 VALVE, GLOBE, 316, 1" FNPT, 800LB 5 2 205113 VALVE, BALL, REDUCED PORT, ¾” FNPT, 316SS 6 1 203649 HOUSING, FILTER, BLK/BLU, 4.5”x 20”, 1” FNPT 7 1 200640 CARTRIDGE, SEDIMENT, POLYPRO, 4.5" X 20", 5 MIC 8 1 205097 SWITCH, PRESSURE, LOW, N/O 5PSI, ¼ MNPT, NASON
9 1 205589 15 1 206350 16 1 200899 METER, FLOW, PM, 0-10 GPM, 1” MNPT x 1” MNPT 17 1 200898 METER, FLOW, PM, 0-5 GPM, 1/2" MNPT x 1/2" MNPT 18 1 200965 VALVE, CHECK, PP, 1/2" FNPT X 1/2" FNPT
PUMP, MULTISTAGE, 3 HP, 115-230V, 1PH, TEFC,
HOUSING, MEMBRANE, 450PSI, FRP, 4040, 1" VICTAULIC, P-18, ROPV CONTROLLER, COMPUTER, ROTROL S-150, 220V, 1PH, 12 X 10 BOX
19 2 205099 GAUGE, BKM, FILL, 0-100PSI/BAR, 2.5” DIA, ¼” MNPT, SS 20 2 205100 GAUGE, BKM, FILL, 0-600PSI/BAR, 2.5” DIA, ¼” MNPT, SS 21 1 205949 MEMBRANE, LC LE, 4040, FILMTEC 22 1 203296 VALVE, NEEDLE, SS 316L, 3/4" FNPT
R2-Series User’s Manual 11/13/12
MKTF - 212
35
R2-2140 SYSTEM PART LIST
3SV20PA30/V09722
X 10 BOX
Item No. Quantity Part Number Description
1 1 206420
2 1 204914 VALVE, SOLENOID, N/C, UL, 220V, 1” FNPT, 3 METER CORD
3 1 205098 SWITCH, PRESSURE, HIGH, N/C 450PSI, ¼” MNPT, NASON
4 1 205903 VALVE, GLOBE, 316, 1" FNPT, 800LB
5 2 205113 VALVE, BALL, REDUCED PORT, ¾” FNPT, 316SS
6 1 203649 HOUSING, FILTER, BLK/BLU, 4.5”x 20”, 1” FNPT
7 1 200640 CARTRIDGE, SEDIMENT, POLYPRO, 4.5" X 20", 5 MIC
8 1 205097 SWITCH, PRESSURE, LOW, N/O 5PSI, ¼ MNPT, NASON
9 1 205589 10 1 205593 15 1 206350 16 1 200899 METER, FLOW, PM, 0-10 GPM, 1” MNPT x 1” MNPT 17 1 200898 METER, FLOW, PM, 0-5 GPM, 1/2" MNPT x 1/2" MNPT
PUMP, MULTISTAGE, 3 HP, 115-230V, 1PH, TEFC,
HOUSING, MEMBRANE, 450PSI, FRP, 4040, 1" VICTAULIC, P-18, ROPV HOUSING, MEMBRANE, 450PSI, FRP, 4040, 1" VICTAULIC, P-46, ROPV CONTROLLER, COMPUTER, ROTROL S-150, 220V, 1PH, 12
18 1 200965 VALVE, CHECK, PP, 1/2" FNPT X 1/2" FNPT 19 2 205099 GAUGE, BKM, FILL, 0-100PSI/BAR, 2.5” DIA, ¼” MNPT, SS 20 2 205100 GAUGE, BKM, FILL, 0-600PSI/BAR, 2.5” DIA, ¼” MNPT, SS 21 2 205949 MEMBRANE, LC LE, 4040, FILMTEC 22 1 203296 VALVE, NEEDLE, SS 316L, 3/4" FNPT
R2-Series User’s Manual 11/13/12
MKTF - 212
36
R2-3140 SYSTEM PART LIST
Item No. Quantity Part Number Description
1 1 206420
2 1 204914 VALVE, SOLENOID, N/C, UL, 220V, 1” FNPT, 3 METER CORD
3 1 205098 SWITCH, PRESSURE, HIGH, N/C 450PSI, ¼” MNPT, NASON
4 1 205903 VALVE, GLOBE, 316, 1" FNPT, 800LB
5 2 205113 VALVE, BALL, REDUCED PORT, ¾” FNPT, 316SS
6 1 203649 HOUSING, FILTER, BLK/BLU, 4.5”x 20”, 1” FNPT
7 1 200640 CARTRIDGE, SEDIMENT, POLYPRO, 4.5" X 20", 5 MIC
8 1 205097 SWITCH, PRESSURE, LOW, N/O 5PSI, ¼ MNPT, NASON
9 1 205589 10 1 205593 11 1 205591 15 1 206350 16 1 200899 METER, FLOW, PM, 0-10 GPM, 1” MNPT x 1” MNPT
PUMP, MULTISTAGE, 3 HP, 115-230V, 1PH, TEFC, 3SV20PA30/V09722
HOUSING, MEMBRANE, 450PSI, FRP, 4040, 1" VICTAULIC, P-18, ROPV HOUSING, MEMBRANE, 450PSI, FRP, 4040, 1" VICTAULIC, P-46, ROPV HOUSING, MEMBRANE, 450PSI, FRP, 4040, 1" VICTAULIC, P-27, ROPV CONTROLLER, COMPUTER, ROTROL S-150, 220V, 1PH, 12 X 10 BOX
17 1 200898 METER, FLOW, PM, 0-5 GPM, 1/2" MNPT x 1/2" MNPT 18 1 200965 VALVE, CHECK, PP, 1/2" FNPT X 1/2" FNPT 19 2 205099 GAUGE, BKM, FILL, 0-100PSI/BAR, 2.5” DIA, ¼” MNPT, SS 20 2 205100 GAUGE, BKM, FILL, 0-600PSI/BAR, 2.5” DIA, ¼” MNPT, SS 21 3 205949 MEMBRANE, LC LE, 4040, FILMTEC 22 1 203296 VALVE, NEEDLE, SS 316L, 3/4" FNPT
R2-Series User’s Manual 11/13/12
MKTF - 212
37
R2-4140 SYSTEM PART LIST
Item No. Quantity Part Number Description
1 1 206421
2 1 204914 VALVE, SOLENOID, N/C, UL, 220V, 1” FNPT, 3 METER CORD
3 1 205098 SWITCH, PRESSURE, HIGH, N/C 450PSI, ¼” MNPT, NASON
4 1 205903 VALVE, GLOBE, 316, 1" FNPT, 800LB
5 2 205113 VALVE, BALL, REDUCED PORT, ¾” FNPT, 316SS
6 1 203649 HOUSING, FILTER, BLK/BLU, 4.5”x 20”, 1” FNPT
7 1 200640 CARTRIDGE, SEDIMENT, POLYPRO, 4.5" X 20", 5 MIC
8 1 205097 SWITCH, PRESSURE, LOW, N/O 5PSI, ¼ MNPT, NASON
9 1 205589 10 1 205593 11 1 205591 12 1 205589 15 1 203263 16 1 200899 METER, FLOW, PM, 0-10 GPM, 1” MNPT x 1” MNPT
PUMP, MULTISTAGE, 5HP,208­230/460V,3HP,TEFC,3SV22PB30/V10742A
HOUSING, MEMBRANE, 450PSI, FRP, 4040, 1" VICTAULIC, P-18, ROPV HOUSING, MEMBRANE, 450PSI, FRP, 4040, 1" VICTAULIC, P-46, ROPV HOUSING, MEMBRANE, 450PSI, FRP, 4040, 1" VICTAULIC, P-27, ROPV HOUSING, MEMBRANE, 450PSI, FRP, 4040, 1" VICTAULIC, P-18, ROPV CONTROLLER, COMPUTER, ROTROL S-150, 220V, 3PH, 5 HP
17 1 200899 METER, FLOW, PM, 0-10 GPM, 1” MNPT x 1” MNPT 18 1 200965 VALVE, CHECK, PP, 1/2" FNPT X 1/2" FNPT 19 2 205099 GAUGE, BKM, FILL, 0-100PSI/BAR, 2.5” DIA, ¼” MNPT, SS 20 2 205100 GAUGE, BKM, FILL, 0-600PSI/BAR, 2.5” DIA, ¼” MNPT, SS 21 4 205949 MEMBRANE, LC LE, 4040, FILMTEC 22 1 203296 VALVE, NEEDLE, SS 316L, 3/4" FNPT
R2-Series User’s Manual 11/13/12
MKTF - 212
38
R2-5140 SYSTEM PART LIST
HP
Item No. Quantity Part Number Description
1 1 206421
2 1 204914 VALVE, SOLENOID, N/C, UL, 220V, 1” FNPT, 3 METER CORD
3 1 205098 SWITCH, PRESSURE, HIGH, N/C 450PSI, ¼” MNPT, NASON
4 1 205903 VALVE, GLOBE, 316, 1" FNPT, 800LB
5 2 205113 VALVE, BALL, REDUCED PORT, ¾” FNPT, 316SS
6 1 203649 HOUSING, FILTER, BLK/BLU, 4.5”x 20”, 1” FNPT
7 1 200640 CARTRIDGE, SEDIMENT, POLYPRO, 4.5" X 20", 5 MIC
8 1 205097 SWITCH, PRESSURE, LOW, N/O 5PSI, ¼ MNPT, NASON
9 1 205589 10 1 205593 11 1 205591 12 1 205589 13 1 205593 15 1 203263
PUMP, MULTISTAGE, 5HP,208­230/460V,3HP,TEFC,3SV22PB30/V10742A
HOUSING, MEMBRANE, 450PSI, FRP, 4040, 1" VICTAULIC, P-18, ROPV HOUSING, MEMBRANE, 450PSI, FRP, 4040, 1" VICTAULIC, P-46, ROPV HOUSING, MEMBRANE, 450PSI, FRP, 4040, 1" VICTAULIC, P-27, ROPV HOUSING, MEMBRANE, 450PSI, FRP, 4040, 1" VICTAULIC, P-18, ROPV HOUSING, MEMBRANE, 450PSI, FRP, 4040, 1" VICTAULIC, P-46, ROPV CONTROLLER, COMPUTER, ROTROL S-150, 220V, 3PH, 5
16 1 200899 METER, FLOW, PM, 0-10 GPM, 1” MNPT x 1” MNPT 17 1 200899 METER, FLOW, PM, 0-10 GPM, 1” MNPT x 1” MNPT 18 1 200965 VALVE, CHECK, PP, 1/2" FNPT X 1/2" FNPT 19 2 205099 GAUGE, BKM, FILL, 0-100PSI/BAR, 2.5” DIA, ¼” MNPT, SS 20 2 205100 GAUGE, BKM, FILL, 0-600PSI/BAR, 2.5” DIA, ¼” MNPT, SS 21 5 205949 MEMBRANE, LC LE, 4040, FILMTEC 22 1 203296 VALVE, NEEDLE, SS 316L, 3/4" FNPT
R2-Series User’s Manual 11/13/12
MKTF - 212
39
R2-6140 SYSTEM PART LIST
P-27, ROPV
Item No. Quantity Part Number Description
1 1 206422
2 1 204914 VALVE, SOLENOID, N/C, UL, 220V, 1” FNPT, 3 METER CORD
3 1 205098 SWITCH, PRESSURE, HIGH, N/C 450PSI, ¼” MNPT, NASON
4 1 205903 VALVE, GLOBE, 316, 1" FNPT, 800LB
5 2 205113 VALVE, BALL, REDUCED PORT, ¾” FNPT, 316SS
6 1 203649 HOUSING, FILTER, BLK/BLU, 4.5”x 20”, 1” FNPT
7 1 200640 CARTRIDGE, SEDIMENT, POLYPRO, 4.5" X 20", 5 MIC
8 1 205097 SWITCH, PRESSURE, LOW, N/O 5PSI, ¼ MNPT, NASON
9 1 205589 10 1 205593 11 1 205591 12 1 205589 13 1 205593 14 1 205591
PUMP, MULTISTAGE, 5HP,208­230/460V,3HP,TEFC,3SV26PB30/V10742A
HOUSING, MEMBRANE, 450PSI, FRP, 4040, 1" VICTAULIC, P-18, ROPV HOUSING, MEMBRANE, 450PSI, FRP, 4040, 1" VICTAULIC, P-46, ROPV HOUSING, MEMBRANE, 450PSI, FRP, 4040, 1" VICTAULIC, P-27, ROPV HOUSING, MEMBRANE, 450PSI, FRP, 4040, 1" VICTAULIC, P-18, ROPV HOUSING, MEMBRANE, 450PSI, FRP, 4040, 1" VICTAULIC, P-46, ROPV HOUSING, MEMBRANE, 450PSI, FRP, 4040, 1" VICTAULIC,
15 1 203263 16 1 200899 METER, FLOW, PM, 0-10 GPM, 1” MNPT x 1” MNPT 17 1 200899 METER, FLOW, PM, 0-10 GPM, 1” MNPT x 1” MNPT 18 1 200965 VALVE, CHECK, PP, 1/2" FNPT X 1/2" FNPT 19 2 205099 GAUGE, BKM, FILL, 0-100PSI/BAR, 2.5” DIA, ¼” MNPT, SS 20 2 205100 GAUGE, BKM, FILL, 0-600PSI/BAR, 2.5” DIA, ¼” MNPT, SS 21 6 205949 MEMBRANE, LC LE, 4040, FILMTEC 22 1 203296 VALVE, NEEDLE, SS 316L, 3/4" FNPT
CONTROLLER, COMPUTER, ROTROL S-150, 220V, 3PH, 5 HP
R2-1140 FLOW DIAGRAM
R2-Series User’s Manual 11/13/12
MKTF - 212
40
R2-2140 FLOW DIAGRAM
R2-Series User’s Manual 11/13/12
MKTF - 212
41
R2-3140 FLOW DIAGRAM
R2-Series User’s Manual 11/13/12
MKTF - 212
42
R2-4140 FLOW DIAGRAM
R2-Series User’s Manual 11/13/12
MKTF - 212
43
R2-5140 FLOW DIAGRAM
R2-Series User’s Manual 11/13/12
MKTF - 212
44
R2-6140 FLOW DIAGRAM
R2-Series User’s Manual 11/13/12
MKTF - 212
45
R2-Series User’s Manual 11/13/12
46
MKTF - 212
R2-Series User’s Manual 11/13/12
47
MKTF - 212
R2-Series User’s Manual 11/13/12
48
MKTF - 212
R2-Series User’s Manual 11/13/12
49
MKTF - 212
R2-Series User’s Manual 11/13/12
50
MKTF - 212
R2-Series User’s Manual 11/13/12
51
MKTF - 212
R2-Series User’s Manual 11/13/12
52
MKTF - 212
R2-Series User’s Manual 11/13/12
53
MKTF - 212
R2-Series User’s Manual 11/13/12
54
MKTF - 212
R2-Series User’s Manual 11/13/12
55
MKTF - 212
R2-Series User’s Manual 11/13/12
56
MKTF - 212
This Page Intentionally
Left Blank
R2-Series User’s Manual 11/13/12
MKTF - 212
57
Loading...