AXEON R1 User Manual

Reverse Osmosis
User’s Manual
Model
R1-1140, R1-2140, R1-3140, R1-4140, R1-5140, R1-6140
R1-6140 Pictured
MKTF - 211 07/26/12
Left Blank
MKTF - 211 07/26/12
2
TABLE OF CONTENTS
INTRODUCTION ................................................................................................................................... 4
SAFETY ................................................................................................................................................ 4
FEED WATER AND OPERATION SPECIFICATIONS .................................................................................. 5
REJECTION, RECOVERY AND FLOW RATES............................................................................................ 5
SYSTEM REQUIREMENTS AND OPERATION GUIDELINES....................................................................... 6
MEMBRANE ELEMENTS ....................................................................................................................... 8
R1-1140, R1-2140, R1-3140, R1-4140, R1-5140, R1-6140 ................................................................... 16
SYSTEM IDENTIFICATION ................................................................................................................... 16
SYSTEM PURGING.............................................................................................................................. 21
INITIAL START-UP .............................................................................................................................. 22
OPERATING DO’s AND DON’Ts........................................................................................................... 24
OPERATION AND MAINTENANCE....................................................................................................... 24
MEMBRANE REMOVAL AND REPLACEMENT ...................................................................................... 26
FLUSHING THE SYSTEM...................................................................................................................... 28
PREPARING UNIT FOR STORAGE OR SHIPMENT ................................................................................. 28
REVERSE OSMOSIS TROUBLESHOOTING ............................................................................................ 29
TEMPERATURE CORRECTION FACTORS FOR MEMBRANE................................................................... 31
OPERATION ....................................................................................................................................... 33
DRAWINGS ........................................................................................................................................ 34
MKTF - 211 07/26/12
3
211
Your R1-Series
system is a durable piece of equipment which, with proper care, will last for many years.
outlines installation, operation
sustained performance of your system.
The test results which are included with this Users Manual indicate your system’s permeate (product) and
If your system is altered at the site of operation
your local dealer or distributor to determine the proper recovery for your application.
NOTE: IN ORDER TO MAINTAIN
MUST BE MAINTAINED AND COPIES WILL NEE
OPERATING OR SERVICING THE
MANUAL MUST BE READ AND FULLY UNDERSTOOD. KEEP THIS AND OTHER ASSOCIATED
INFORMATION FOR FUTURE REFERENCE AND FOR N
PERSONNEL NEAR THE SYSTEM
The Safety section of this User’s Manual outlines the various safety headings used throughout this
manual’s text and are enhanced and defined below:
NOTE: INDICATES STATEMENTS THAT PROVIDE
CAUTION: INDICATES STATEMENTS THAT ARE USED TO IDENTIFY CONDITIONS OR
PRACTICES THAT COULD RESULT IN EQUIPMENT OR OTHER PROPERTY DAMAGE.
WARNING: INDICATES STATEMENTS THAT ARE USED TO IDENTIFY CONDITIONS OR
RACTICES THAT COULD RESULT IN INJURY OR LOSS OF LIFE. FAILURE TO FOLLOW
WARNINGS COULD RESULT IN SERIOUS INJURY OR EVEN DEATH.
DO NOT UNDER ANY CIRCUMSTANCE; REMOVE ANY CA
DESCRIPTIVE LABELS FROM THE SYSTEM.
, maintenance, and troubleshooting
or if the feed water conditions
WARRANTY, AN OPERATING LOG
D TO BE SENT TO YOUR LOCAL DEALER OR
REVERSE OSMOSIS SYSTEM, THIS USERS
EW OPERATORS OR QUALIFIED
FURTHER INFORMATION AND CLARIFICATION.
TION, WARNING, OR OTHER
This User’s Manual
concentrate (waste) test results.
DISTRIBUTOR FOR REVIEW. NOTE: PRIOR TO
INTRODUCTION
details vital to the
change, please contact
THE MANUFACTURER’S
.
SAFETY
P
U
MKTF -
4
07/26/12
FEED WATER & OPERATION SPECIFICATIONS
Nothing has a greater effect on a reverse osmosis system than the feed water quality.
NOTE: IT IS VERY IMPORTANT TO MEET THE MINIMUM FEED WATER REQUIREMENTS. FAILURE TO DO SO WILL CAUSE THE MEMBRANES TO FOUL AND VOID THE MANUFACTURER’S WARRANTY.
NOTE: HIGHER FEED TDS AND/OR LOWER TEMPERATURES WILL REDUCE THE SYSTEM’S
PRODUCTION.
REJECTION, RECOVERY AND FLOW RATES
R1-Series reverse osmosis systems are designed to produce permeate water at the capacities indicated in the “design basis section on page 21 of this manual. For example, the R1-6140 produces 6.25 gallons per minute of permeate water at the listed operating test conditions.
The amount of total dissolved solids (TDS) rejected by the membrane is expressed as a percentage. For example, a 99% rejection rate means that 99% of total dissolved solids do not pass through the membrane. To calculate the % rejection, use the following formula:
% Rejection = [(Feed TDS – Product TDS) / Feed TDS] x 100
Example:
99% = [(550-5.5)/550] x 100
NOTE: ALL TDS FIGURES MUST BE EXPRESSED IN THE SAME UNITS, TYPICALLY PARTS PER MILLION (PPM) OR MILLIGRAMS PER LITER (MG/L).
R1-Series reverse osmosis systems are designed to reject up to 99% NaCl, unless computer projections have been provided or stated otherwise.
MKTF - 211 07/26/12
5
211
The amount of permeate
water recovered for use is expressed as a percentage. To calculate % recovery,
% Recovery = (Product Water Flow Rate / Feed Water Flow Rate) x 100
% Rejection = [(Feed TDS
NOTE: ALL FLOW RATES MUST BE EXPRESSED IN THE SAME UNITS.
SYSTEM REQUIREMENTS
gh pressure pumps used on R1
feed pressure of 4
FEED WATER CONNECTION
FNPT Sole
inlet piping to the 1
Be certain that all of the
PERMEATE (PRODUCT WATER) CONNECTION
n (Depending on the unit)
sure that the permeate water can flow freely
damage to the membrane elements.
with PVC fittings, or other FDA approved materials
into the permeate water.
H OF THE REVERSE O
LOWER THAN THE FEED WATER PH. A LOW PH CAN BE VERY AGGRESSIVE
TO SOME PLUMBING MATERIALS SUCH AS COPPER PIPING.
Product TDS) / Feed TDS] x 100
OPERATION GUIDELINES
systems require a continuo
the feed water are soluble at the concentrations
labeled permeate and attach to storage tank
Backpressure can cause
1” or 3/4” (Depending on the unit)
.
SMOSIS PERMEATE WATER WILL TYPICA
.
use the following formula:
PLUMBING
Example:
26% = (1.04/4.00) x 100
Example:
99% = [(550-5.5)/550] x 100
AND
The membranes and hi with a minimum
1. Locate the 1 (Figure 1A, Page 14)
2. Attach the
3.
attained in the system.
Locate the 1” or 3/4” connectio En irreversible be run to the holding tank being used does not leach
CAUTION THE p
BE 1-2 pH UNITS
-Series
5 psi, not to exceed 90°F.
noid Valve feed water inlet.
FNPT Solenoid Valve feed water inlet.
dissolved solids within
with no backpressure.
The
us flow of water
.
permeate line can
This is so the material
LLY
MKTF -
6
07/26/12
211
CONCENTRATE (WASTE WATER) CONNECTION
Locate the
1” or 3/4” connection (Depending on the unit)
to an open drain in a free and unrestricted manner (no backpressure).
break be used on the concentrate line to prevent siphoning of water from the pressure vessels
CAUTION: ANY RESTRICTIONS OR BLOCKAGE IN THE DRAIN LINE CAN CAUSE
BACKPRESSURE, WHICH WILL INCREASE THE SYSTEMS OPERATING PRESSURE. THIS CAN
RESULT IN DAMAGE TO THE SYSTEMS MEMBRANES AND COMPONENTS.
Series
hase
nsure that the electrical circuit supplying the system is compatible with the re
ED
ACCORDANCE WITH LOCAL AND NATIONAL ELECTRICAL CODES (NEC).
WARNING: TO REDUCE THE RISK OF ELECTRICAL SHOCK, THE INCOMING
SUPPLY MUST INCLUDE A PROTECTIVE EARTH GROUND.
systems are typically controlled with a liquid level switch in a storage tank. The liquid level
switch turns the system on when the water level in the tank drops, and off when the tank is full
level switches can be obtained by your local dealer or distributor. If a liquid level switch is to be used,
systems are supplied with a
ists between the two pre
Filtration systems, if required.
NOTE: THE SYSTEM MUST BE OPERATED
Series systems is of the
Follow these guidelines to ensure proper operation of the pump:
and
The motor is available
system is equipped with a 5 foot electrical
uirements of the specific
THAT A LICENSED ELECTRICIAN WIRE YOUR SYSTEM IN
. Change the cartridge
. Ask your local deale
WATER ONLY.
stainless
be run dry. Operating the pump without sufficient feed water will damage
uirements of the specific
the concentrate line that an air­when the system is in standby.
ELECTRICAL
labeled concentrate
attach to a drain. Run
It is advised
The motor used on the R1­220/460 Volt, 50/60 Hertz, 1 P cord.
E R1 model you are installing.
NOTE: IT’S RECOMMEND
R1-Series
install it at this time.
PRE-FILTRATION
R1-Series when a 10-15 psi differential ex about Pre-
systems are pump and motor combination.
/3 phase. Each R1-Series
5 micron sediment filter
-filter gauges
in
q
POWER
. Liquid
once a month or
r or distributor
USING FILTERED FEED
PUMP
The pump used on the R1-
The pump must NEVER
the pump.
MKTF -
multi-stage centrifugal
7
steel type.
07/26/12
ALWAYS feed the pump with filtered water. The pump is susceptible to damage from sediment
and debris.
If any damage occurs to your system’s pump a re-build kit may be available. Contact your local
dealer or distributor and inform them of your system’s model and pump size.
MOUNTING
The free standing system should be bolted down in compliance with local regulation standards or securely fastened.
MEMBRANE ELEMENTS
R1-Series reverse osmosis systems come pre-loaded with Thin Film Composite (TFC) HF4 High Flow Extra Low Energy membranes, unless otherwise specified. General membrane element performance characteristics are listed on the following pages:
HF4-STANDARD
MKTF - 211 07/26/12
8
HF5-OPTIONAL
MKTF - 211 07/26/12
9
NF3-OPTIONAL
MKTF - 211 07/26/12
10
NF4-OPTIONAL
MKTF - 211 07/26/12
11
LC LE-4040-OPTIONAL
MKTF - 211 07/26/12
12
MKTF - 211 07/26/12
13
LC HR-4040-OPTIONAL
MKTF - 211 07/26/12
14
MKTF - 211 07/26/12
15
R1-1140, R1-2140, R1-3140, R1-4140, R1-5140, R1-6140
SYSTEM IDENTIFICATION
FIGURE 1A
MKTF - 211 07/26/12
16
NUMBER IDENTIFICATION
1. SOLENOID VALVE – TURNS ON/OFF FEED WATER
2. 5 MICRON SEDIMENT – REMOVES PARTICULATES
3. PRESSURE GAUGE – MEASURES FEED PRESSURE
4. PRESSURE GAUGE – MEASURES PRESSURE AFTER FILTERS
5. PRESSURE GAUGE – MEASURES PUMP PRESSURE
6. PRESSURE GAUGE – MEASURES CONCENTRATE PRESSURE
7. COMPUTER CONTROL – CONTROLS RO SYSTEM FUNCTIONS
8. RECYCLE VALVE – RECYCLES CONCENTRATE BACK TO FEED (IF APPLICABLE)
9. CONCENTRATE VALVE– CONTROLS FLOW OF CONCENTRATE (WASTE) WATER TO THE DRAIN
10. RO PUMP AND MOTOR – PRESSURIZES RO SYSTEM
11. THROTTLE VALVE CONTROLS FLOW OF PUMP
12. FLOW METER MEASURES FLOW OF PERMEATE WATER
13. FLOW METER MEASURES FLOW OF CONCENTRATE (WASTE) WATER
14. FLOW METER MEASURES FLOW OF CONCENTRATE RECYCLE WATER
15. PRESSURE SWITCH – TURNS OFF RO PUMP WHEN FEED PRESSURE FALLS BELOW 15PSI
16. PERMEATE CHECK VALVE – PROTECTS MEMBRANE ELEMENTS FROM BACKPRESSURE
17. PRESSURE VESSELS – HOUSES MEMBRANE ELEMENTS
18. PERMEATE SAMPLE VALVE- MEASURES THE QUALITY OF EACH MEMBRANE
MKTF - 211 07/26/12
17
FIGURE 1B
MKTF - 211 07/26/12
18
FIGURE 1C
FIGURE 1D
MKTF - 211 07/26/12
19
FIGURE 1E
FIGURE 1F
Note: A portion of the frame has been removed to expose components.
MKTF - 211 07/26/12
20
SYSTEM PURGING
Carefully inspect your system before initial start-up. Check that all plumbing and electrical connections are not loose or have not come undone during shipment. A User’s Manual, Test Results, and Filter Housing Wrench will accompany your R1-Series reverse osmosis system.
NOTE: LEAVE THE POWER TO THE SYSTEM OFF FOR THIS PROCEDURE.
1. Redirect permeate water to the drain for this procedure.
2. Fully open the concentrate valve (Counter Clockwise). (Figure 1B, Page. 16)
3. Fully close the recycle valve (Clockwise) (If Applicable). (Figure 1B, Page. 16)
4. Fully open the throttle valve (Counter Clockwise). (Figure 1D, Page. 17)
5. To activate the Solenoid bypass feature, press and hold the momentary push button located on the front panel of the C-22 computer controller.
6. Turn the feed water on and let the system purge until no visible bubbles appear from concentrate flow meter. (Figure 1B, Page. 16)
7. Release the momentary push button.
Note: Units with S-150 controller press and hold the reset button for three seconds to activate the bypass feature.
MKTF - 211 07/26/12
21
FIGURE 2
INITIAL START-UP
1. Keep the permeate water line to drain for this procedure.
2. Fully open the concentrate valve (Counter Clockwise). (Figure 1B, Page. 16)
3. Fully close the recycle valve (Clockwise)(If Applicable). (Figure 1B, Page. 16)
4. Adjust the throttle valve at 50% open (Counter Clockwise). (Figure 1D, Page. 17)
5. Turn the RO system on and adjust the concentrate (waste) valve, recycle valve (If Applicable), and the throttle valve to the designed flow and pressure. (Figure 1B, Page. 16)
6. Inspect the system for leaks.
7. Allow the system to run 30 minutes to flush the preservative solution from the system.
8. After 30 minutes, shut down the system.
9. Re-direct the permeate water back to the tank and then turn the system back on.
10. Record the readings daily for a week; after one week record the readings once a week.
MKTF - 211 07/26/12
22
DESIGN BASIS FOR R1-1140, R1-2140, R1-3140,
Specifications
R1-4140, R1-5140, R1-6140
WARNING: NEVER EXCEED THE SYSTEM’S MAXIMUM PRESSURE RATING
MKTF - 211 07/26/12
23
OPERATING DO’s AND DONT’S
DO:
Change the cartridge filters regularly Monitor the system and keep a daily log Run the system, as much as possible, on a continuous basis. Adjust the system recovery to the recommended value Always feed the pump with filtered water.
DON’T:
Permit chlorine to enter or be present in the feed water Shut down the system for extended periods Close the throttle valve completely Operate the system with insufficient feed flow Operate the pump dry
OPERATION AND MAINTENANCE
The reverse osmosis process causes the concentration of impurities. The impurities may precipitate (fall out of solution) when their concentration reaches saturation levels.
NOTE: PRECIPITATION CAN SCALE OR FOUL MEMBRANES AND MUST BE PREVENTED.
Check your feed water chemistry and pre-treat the water and/or reduce the system’s recovery as required. If necessary, consult with your local dealer or distributor.
PRE-FILTER PRESSURE GAUGES
Pre-filter gauges measure the feed water pressure when it enters and exits the pre-filter. A pressure differential of 10 - 15 psi or more on the two pressure gauges indicates that the pre-filters require servicing.
PUMP PRESSURE AND CONCENTRATE PRESSURE GAUGES
Pump and concentrate gauges measure the pressure of water exiting the multistage pump and the pressure of concentrate water as it exits the pressure vessel array. Comparison of the pump output and concentrate pressures allows for the establishment of a baseline pressure differential. If the pressure differential increases over time from this baseline, it would be an indication that the reverse osmosis membranes need to be inspected.
MKTF - 211 07/26/12
24
PERMEATE (PRODUCT) FLOW METER AND CONCENTRATE (WASTE) FLOW METER
These flow meters indicate the flow rates of the permeate and concentrate water. The measurements, when added together, also indicate the feed water flow rate or (total flow rate) if the recycle valve is not being used. If the recycle valve is being used, add the flow rates for all three flow meters (permeate, concentrate, and recycle) to obtain the total feed flow.
LOW PRESSURE SWITCH
The low pressure switch shuts off the system when the feed water pressure drops below 15 PSI, preventing damage to the pump. The system restarts automatically when there is a constant pressure of 35 PSI or more.
PUMP THROTTLE VALVE
The Pump Throttle Valve is installed as a standard feature on the R1-Series reverse osmosis systems. It provides an adjustment for pump pressure. As the feed water temperature decreases, and/or the feed water TDS increases, the system will require a higher operating pressure to produce the specified permeate flow.
ADJUSTING THE THROTTLE VALVE
To decrease the pressure, turn the handle clockwise. To increase the pressure turn the handle counter
clockwise. (Figure 3, Page 24)
FIGURE 3
MKTF - 211 07/26/12
25
MEMBRANE REMOVAL AND REPLACEMENT
Replacing membranes in the pressure vessels is an easy process if you have the proper information and tools at hand. Please refer to the following instructions when removing and replacing membrane elements:
WARNING: ALL PRESSURE GAUGES MUST READ ZERO BEFORE PROCEEDING. BEFORE ATTEMPTING, DISCONNECT THE POWER FROM THE SYSTEM AND BLEED ALL WATER PRESSURE FROM THE SYSTEM.
1. Remove the end plugs from the top of the pressure vessels. This is done by removing the two half-moon retaining disks using a #5 Allen wrench; the end plugs should then freely slide out of the pressure vessel.
2. Remove the replacement membrane element(s) from the shipping box; the membrane(s) should be contained within a plastic oxygen barrier bag.
NOTE: WEAR GLOVES FOR THE FOLLOWING STEPS IN ORDER NOT TO CONTAMINATE THE MEMBRANE.
3. Cut the bag open as close as possible to the seal at one end of the bag, so the bag may be re­used if necessary.
4. Make sure that all parts are clean and free from dirt. Examine the brine seal, and permeate tube for nicks or cuts. Replace the O-rings or brine seal if damaged.
5. Flow directions should be observed for installation of each element into their respective pressure vessels.
REPLACING THE MEMBRANE ELEMENT:
WARNING: THE BRINE SEAL MUST BE IN THE SAME POSITION FOR EACH MEMBRANE ELEMENT HOUSING, SO MARK EACH HOUSING PRIOR TO REMOVING THE MEMBRANE ELEMENTS. THE BRINE SEAL IS A RUBBER SEAL THAT PROTRUDES ON ONE SIDE OF THE MEMBRANE AND IS ALWAYS ON THE FEED SIDE OF THE MEMBRANE ELEMENT.
1. Remove one membrane element at a time from the pressure vessels, from the top of each housing. Long nose pliers may be necessary to pull the old membrane element out of the membrane element housing.
2. Lubricate the brine seal with a non-petroleum based lubricant, such as Dow Corning® 111.
3. Install membranes with brine seal location depicted in (Figure 4, Page 26).
4. For brine seals that are on the bottom: At a slight angle insert the membrane while slightly rotating the element being careful not to tear or flip the brine seal. A slow twisting motion should be used to insert the membrane element, for every couple of turns pull up ½” to make sure the brine seal doesn’t flip.
5. With a smooth and constant motion, push the membrane element into the housing so the brine seal enters the housing without coming out of the brine seal groove.
MKTF - 211 07/26/12
26
211
6. Re-
install the end plugs by gently twisting the end cap while pushing it onto the housing. Ensure
that you do not pinch or fatigue any O
diameter of the plug is flush with the outer diameter of the pressure vessel.
moon retaining disks until they are fully seated. Subsequently fasten using a
Reconnect any fittings that may have been disconnected when
Up the system, please refer to the Initial Start
CAUTION: WET MEMBRANES ARE SHIPPED IN A PRESERVATIVE SOLUTION.
T BE FLUSHED FOR AT LE
PRESERVATIVE FROM THE MEMBRANE. DISCARD ALL OF THE PERMEAT
PRODUCED DURING THE FLUSH PERIOD.
1500, R1
installing the end plug. Push the end plug on
the membrane pressure vessels
ion of this manual. (See page 2
AST 30 MINUTES
7500, R1
until the outer
7. Insert the two half­#5 Allen wrench.
8. were disassembled.
-rings while re-
9. To Start-
THE MEMBRANES MUS
-Up sect
8)
TO REMOVE THE
E, WHICH IS
FIGURE 4
View from the back of R1­system.
MKTF -
-3000, R1-4500, R1-6000, R1-
27
-9000 reverse osmosis
07/26/12
FLUSHING THE SYSTEM
The system should be flushed weekly to remove sediment from the surface of the membranes. To manually flush the system, follow the preceding steps:
1. The system must be operating during the flush procedure.
2. Fully open the concentrate valve. (Figure 1B, Page. 16)
3. Allow the system to run for 10 to 20 minutes.
4. After 10 to 20 minutes, close the concentrate valve to its previous setting. Ensure the proper concentrate flow rate is going to the drain.
5. The system is now ready to operate.
PREPARING UNIT FOR STORAGE OR SHIPMENT
PRIOR TO SHIPPING OR STORING YOUR SYSTEM, THE SYSTEM SHOULD BE CLEANED WITH AN APPROPRIATE CLEANER, FLUSHED WITH WATER, AND PROTECTED FROM BIOLOGICAL ATTACK WITH AN APPROPRIATE SOLUTION FOR MEMBRANE ELEMENTS. THE MEMBRANE HOUSING(S) AND PLUMBING LINES OF THE SYSTEM MUST BE COMPLETELY DRAINED. ANY WATER REMAINING IN THE PLUMBING OF A SYSTEM MAY FREEZE, CAUSING SERIOUS DAMAGE.
PREPARING SYSTEM FOR STORAGE:
1. Fully immerse the elements in the membrane housing in a solution of 2% M-100, venting the air outside of the pressure vessels. Use the overflow technique: circulate the M-100 solution in such a way that the remaining air in the system is minimized after the recirculation is completed. After the pressure vessel is filled, the M-100 solution should be allowed to overflow through an opening located higher than the upper end of the highest pressure vessel being filled.
2. Separate the preservation solution from the air outside by closing all valves.
3. Repeat this process at least once a month. During the shutdown period, the plant must be kept frost-free, or the temperature must not exceed 113°F (45°C).
PREPARING UNIT FOR SHIPMENT:
1. Disconnect the inlet, concentrate, pre-filter, and permeate plumbing.
2. Drain all water from the pre-filter cartridge housings by unscrewing the housings, removing the pre-filter cartridges, and drain the water from the housings.
3. Disconnect the tubing from the connectors on the permeate and concentrate inlets and outlets.
4. Fully open the concentrate valve.
5. Drain the flow meters.
6. Allow the system to drain for a minimum of eight hours or until the opened ports quit dripping.
7. After draining is complete, reconnect all of the plumbing.
MKTF - 211 07/26/12
28
REVERSE OSMOSIS TROUBLESHOOTING
temperature
SYMPTOMS POSSIBLE CAUSES CORRECTIVE ACTION
LOW INLET PRESSURE
LOW PERMEATE FLOW
HIGH PERMEATE FLOW
POOR PERMEATE QUALITY
MEMBRANE FOULING
Low supply pressure Increase Inlet Pressure
Cartridge filters plugged Change Filters
Solenoid valve malfunction Replace Sol. Valve and/or Coil
Concentrate valve might be
damaged Replace Needle Valve
Leaks Fix any visible leaks
Cold feed water See temperature correction sheet
Low operating pressure See low inlet pressure
Defective membrane brine seal/
Membrane installed backwards
Fouled or Scaled membrane Clean membranes
Damaged product tube o-rings Inspect and/or replace
Damaged or oxidized membrane Replace membrane
Exceeding maximum feed water
Low operating pressure Adjust concentrate valve
Damage product tube o-rings Inspect and/or replace
Damaged or oxidized membrane Replace membrane
Metal Oxide Fouling
Colloidal Fouling
Scaling (CaSO4, CaSO3,
BaSO4, SiO2)
Biological Fouling
Replace brine seal and / or
Reposition membranes
See temperature correction sheet
Improve pretreatment to remove
metals. Clean with Acid
Cleaners.
Optimize pretreatment for colloid
removal. Clean with high pH
anionic cleaners.
Increase acid addition and
antiscalant dosage for CaVO3
and CaCO4. Reduce recovery.
Clean with Acid Cleaners
Shock dosage of Sodium Bi-
Sulfate. Continuous feed of
Sodium Bi-Sulfate at reduced pH.
Chlorination and de-chlorination.
Replace cartridge filters.
MKTF - 211 07/26/12
Organic Fouling
Chlorine Oxidation
Abrasion of membrane by
Crystalline Material
29
Activated Carbon or other
pretreatment. Clean with high pH
cleaner.
Check Chlorine feed equipment
and de-chlorination system.
Improve pretreatment. Check all
filters for media leakage.
ABNORMAL PERMEATE FLOW
As time progresses, the efficiency of the membrane will be reduced. In general, the salt rejection does not change significantly until two or three years after installation when operated on properly pretreated feed water. The permeate flow rate will begin to decline slightly after one year of operation, but can be extended with diligent flushing and cleaning of the system. A high pH and/or precipitation of hardness can cause premature loss in rejection.
Permeate flow should be within 20% of the rated production, after correcting the feed water temperatures above or below 77°F. Check your permeate flow meter to determine the permeate flow rate.
NOTE: TO DETERMINE THE TEMPERATURE CORRECTION FACTOR, LOCATE THE TEMPERATURE CORRECTION TABLE IN THIS USER’S MANUAL AND FOLLOW THE DIRECTIONS
MKTF - 211 07/26/12
30
TEMPERATURE CORRECTION FACTORS FOR MEMBRANE
Find the temperature correction factor (TCF) from the table below. Divide the rated permeate flow at 77°F by the temperature correction factor. The result is the permeate flow at the desired temperature. (See example on the next page)
MKTF - 211 07/26/12
31
If a system is rated to produce 5 gpm of permeate water @ 77˚ F. The same system will produce more
water at a higher temperature. It will also produce less water at a lower temperature. Use the temperature correction table to obtain the correct flow.
Example:
5 gpm @ 59˚ F (5÷1.42=3.52 gpm)
5 gpm @ 77˚ F (5÷1=5 gpm)
5 gpm @ 84˚ F (5÷0.89=5.62 gpm)
SERVICE ASSISTANCE
If service assistance is required, please complete the following process: Contact your local dealer or distributor. Prior to making the call, have the following information available:
system installation date, serial number, daily log sheets, current operating parameters (e.g. flow, operating pressures, pH, etc.), and a detailed description of the problem.
MKTF - 211 07/26/12
32
OPERATION
Date of Start
-
FILTER
OUTLET PRESSURE
CONCENTRATE PRESSURE
PUMP DISCHARGE PRESSURE
Company:
Location:
Week Of:
System Serial #:
HOUR OF OPERATION
FILTER INLET PRESSURE (PSI)
PERMEATE FLOW (GPM)
____________________
____________________ ____________________ ____________________
DATE
TIME
(PSI)
(PSI)
(PSI)
FEED FLOW (GPM)
Up: Date of Last
Cleaning:
___________________
___________________
CONCENTRATE FLOW (GPM)
RECYCLE FLOW (GPM)
RECOVERY %
FEED TEMPERATURE
FEED TDS (PPM)
PERMEATE TDS (PPM)
REJECTION %
FEED PH
PERMEATE PH
SCALE INHIBITOR FEED (PPM)
IRON (mg/L)
FREE CHLORINE (mg/L)
HARDNESS (GPG CaCO3)
MKTF - 211 07/26/12
33
DRAWINGS
FIGURE 5A
FIGURE 5B
MKTF - 211 07/26/12
34
FIGURE 6
MKTF - 211 07/26/12
35
FIGURE 7
MKTF - 211 07/26/12
36
FIGURE 8
Note: A portion of the frame has been removed to expose components.
MKTF - 211 07/26/12
37
R1-1140 SYSTEM PART LIST
Item No. Quantity Part Number Description
1 1 200906 SWITCH, PRESSURE, LOW, N/O 15-30, ¼” FNPT 2 1 200795 3 1 205903 VALVE, GLOBE, SS, 1” FNPT 4 1 204914 VALVE, SOLENOID, N/C, UL, 220V, 1” FNPT 5 1 200640 CART, SEDIMENT, PLEATED, 4.5” x 20”, 5 MIC 6 1 203649 HOUSING, FILT, BLK/BLU, 4.5” x 20”, 1” FNPT 7 1 202929 HOUSING, MEM, FRP, 4040, ½P x ¾”C FNPT 8 1 204207 9 1 200899 METER, FLOW, PM, 1-10 GPM, 1x 1”
10 2 200898 METER, FLOW, PM, 0-5 GPM, ½x ½ MNPT
PUMP, MULTI-STAGE, 1.5 HP, 110/220V, 1PH 10GBS1514Q4, GOULDS
CONTROLLER, COMPUTER, C22, 120/220V, 1PH
11 1 200965 VALVE, CHECK, PP, ½ FNPT x ½” FNPT 12 2 200904 GAUGE, BKM, FILL, 0-300 PSI/BAR, 2.5” DIA 13 2 204165 GAUGE, BKM, FILL, 0-100 PSI/BAR, 2.5” DIA 14 1 200391 MEMBRANE, HF4, 4040 15 2 201006 VALVE, NEEDLE, SS 316, ½” FNPT 16 1 203606 VALVE, BALL, 1/4" MNPT X 1/4" QC
MKTF - 211 07/26/12
38
R1-2140 SYSTEM PART LIST
Item No. Quantity Part Number Description
1 1 200906 SWITCH, PRESSURE, LOW, N/O 15-30, ¼” FNPT 2 1 200795 3 1 205903 VALVE, GLOBE, SS, 1” FNPT 4 1 204914 VALVE, SOLENOID, N/C, UL, 220V, 1” FNPT 5 1 200640 CART, SEDIMENT, PLEATED, 4.5” x 20”, 5 MIC 6 1 203649 HOUSING, FILT, BLK/BLU, 4.5” x 20”, 1” FNPT 7 2 202929 HOUSING, MEM, FRP, 4040, ½P x ¾”C FNPT 8 1 204207 9 1 200899 METER, FLOW, PM, 1-10 GPM, 1x 1”
10 2 200898 METER, FLOW, PM, 0-5 GPM, ½x ½ MNPT
PUMP, MULTI-STAGE, 1.5 HP, 110/220V, 1PH 10GBS1514Q4, GOULDS
CONTROLLER, COMPUTER, C22, 120/220V, 1PH
11 1 200965 VALVE, CHECK, PP, ½ FNPT x ½” FNPT 12 2 200904 GAUGE, BKM, FILL, 0-300 PSI/BAR, 2.5” DIA 13 2 204165 GAUGE, BKM, FILL, 0-100 PSI/BAR, 2.5” DIA 14 2 200391 MEMBRANE, HF4, 4040 15 2 201006 VALVE, NEEDLE, SS 316, ½” FNPT 16 2 203606 VALVE, BALL, 1/4" MNPT X 1/4" QC
MKTF - 211 07/26/12
39
R1-3140 SYSTEM PART LIST
Item No. Quantity Part Number Description
1 1 200906 SWITCH, PRESSURE, LOW, N/O 15-30, ¼” FNPT 2 1 200795 3 1 205903 VALVE, GLOBE, SS, 1” FNPT 4 1 206688 5 1 200640 CART, SEDIMENT, PLEATED, 4.5” x 20”, 5 MIC 6 1 203649 HOUSING, FILT, BLK/BLU, 4.5” x 20”, 1” FNPT 7 3 202929 HOUSING, MEM, FRP, 4040, ½P x ¾”C FNPT 8 1 204207 9 1 200899 METER, FLOW, PM, 1-10 GPM, 1x 1”
10 2 200898 METER, FLOW, PM, 0-5 GPM, ½x ½ MNPT
PUMP, MULTI-STAGE, 1.5 HP, 110/220V, 1PH 10GBS1514Q4, GOULDS
VALVE, SOLENOID, 2-WAY,BRASS, 100–240V, 1” FNPT, ASCO
CONTROLLER, COMPUTER, C22, 120/220V, 1PH
11 1 200965 VALVE, CHECK, PP, ½ FNPT x ½” FNPT 12 2 200904 GAUGE, BKM, FILL, 0-300 PSI/BAR, 2.5” DIA 13 2 204165 GAUGE, BKM, FILL, 0-100 PSI/BAR, 2.5” DIA 14 3 200391 MEMBRANE, HF4, 4040 15 2 201006 VALVE, NEEDLE, SS 316, ½” FNPT 16 3 203606 VALVE, BALL, 1/4" MNPT X 1/4" QC
MKTF - 211 07/26/12
40
R1-4140 SYSTEM PART LIST
Item No. Quantity Part Number Description
1 1 200906 SWITCH, PRESSURE, LOW, N/O 15-30, ¼” FNPT 2 1 200795 3 1 205903 VALVE, GLOBE, SS, 1” FNPT 4 1 206688 VALVE, SOLENOID, N/C, UL, 220V, 1” FNPT 5 1 200640 CART, SEDIMENT, PLEATED, 4.5” x 20”, 5 MIC 6 1 203649 HOUSING, FILT, BLK/BLU, 4.5” x 20”, 1” FNPT 7 4 202929 HOUSING, MEM, FRP, 4040, ½P x ¾”C FNPT 8 1 204207 9 2 200899 METER, FLOW, PM, 1-10 GPM, 1x 1”
PUMP, MULTI-STAGE, 1.5 HP, 110/220V, 1PH 10GBS1514Q4, GOULDS
CONTROLLER, COMPUTER, C22, 120/220V, 1PH
10 1 200898 METER, FLOW, PM, 0-5 GPM, ½”x ½” MNPT 11 1 200965 VALVE, CHECK, PP, ½” FNPT x ½” FNPT 12 2 200904 GAUGE, BKM, FILL, 0-300 PSI/BAR, 2.5” DIA 13 2 204165 GAUGE, BKM, FILL, 0-100 PSI/BAR, 2.5” DIA 14 4 200391 MEMBRANE, HF4, 4040 15 2 201006 VALVE, NEEDLE, SS 316, ½” FNPT 16 4 203606 VALVE, BALL, 1/4" MNPT X 1/4" QC
MKTF - 211 07/26/12
41
R1-5140 SYSTEM PART LIST
Item No. Quantity Part Number Description
1 1 200906 SWITCH, PRESSURE, LOW, N/O 15-30, ¼” FNPT 2 1 200795 3 1 205903 VALVE, GLOBE, SS, 1” FNPT 4 1 204914 VALVE, SOLENOID, N/C, UL, 220V, 1” FNPT 5 1 200640 CART, SEDIMENT, PLEATED, 4.5” x 20”, 5 MIC 6 1 203649 HOUSING, FILT, BLK/BLU, 4.5” x 20”, 1” FNPT 7 5 202929 HOUSING, MEM, FRP, 4040, ½P x ¾”C FNPT 8 1 204207 9 2 200899 METER, FLOW, PM, 1-10 GPM, 1x 1”
10 1 200898 METER, FLOW, PM, 0-5 GPM, ½x ½” MNPT
PUMP, MULTI-STAGE, 1.5 HP, 110/220V, 1PH 10GBS1514Q4, GOULDS
CONTROLLER, COMPUTER, C22, 120/220V, 1PH
11 1 200965 VALVE, CHECK, PP, ½” FNPT x ½” FNPT 12 2 200904 GAUGE, BKM, FILL, 0-300 PSI/BAR, 2.5” DIA 13 2 204165 GAUGE, BKM, FILL, 0-100 PSI/BAR, 2.5” DIA 14 5 200391 MEMBRANE, HF4, 4040 15 2 201006 VALVE, NEEDLE, SS 316, ½” FNPT 16 5 203606 VALVE, BALL, 1/4" MNPT X 1/4" QC
MKTF - 211 07/26/12
42
R1-6140 SYSTEM PART LIST
Item No. Quantity Part Number Description
1 1 200906 SWITCH, PRESSURE, LOW, N/O 15-30, ¼” FNPT 2 1 206427 3 1 205903 VALVE, GLOBE, SS, 1” FNPT 4 1 204914 VALVE, SOLENOID, N/C, UL, 220V, 1” FNPT 5 1 200640 CART, SEDIMENT, PLEATED, 4.5” x 20”, 5 MIC 6 1 203649 HOUSING, FILT, BLK/BLU, 4.5” x 20”, 1” FNPT 7 6 202929 HOUSING, MEM, FRP, 4040, ½P x ¾”C FNPT 8 1 204207 9 2 200899 METER, FLOW, PM, 1-10 GPM, 1x 1”
PUMP, MULTISTAGE,2HP,115/208­230V,1PH,TEFC,18GBS2014N4
CONTROLLER, COMPUTER, C22, 120/220V, 1PH
10 1 200898 METER, FLOW, PM, 0-5 GPM, ½”x ½” MNPT 11 1 200965 VALVE, CHECK, PP, ½” FNPT x ½” FNPT 12 2 200904 GAUGE, BKM, FILL, 0-300 PSI/BAR, 2.5” DIA 13 2 204165 GAUGE, BKM, FILL, 0-100 PSI/BAR, 2.5” DIA 14 6 200391 MEMBRANE, HF4, 4040 15 2 201006 VALVE, NEEDLE, SS 316, ½” FNPT 16 6 203606 VALVE, BALL, 1/4" MNPT X 1/4" QC
MKTF - 211 07/26/12
43
R1-1140 FLOW DIAGRAM
MKTF - 211 07/26/12
44
R1-2140 FLOW DIAGRAM
MKTF - 211 07/26/12
45
R1-3140 FLOW DIAGRAM
MKTF - 211 07/26/12
46
R1-4140 FLOW DIAGRAM
MKTF - 211 07/26/12
47
R1-5140 FLOW DIAGRAM
MKTF - 211 07/26/12
48
R1-6140 FLOW DIAGRAM
MKTF - 211 07/26/12
49
MKTF - 211 07/26/12
50
MKTF - 211 07/26/12
51
MKTF - 211 07/26/12
52
MKTF - 211 07/26/12
53
MKTF - 211 07/26/12
54
MKTF - 211 07/26/12
55
MKTF - 211 07/26/12
56
MKTF - 211 07/26/12
57
This Page Intentionally
Left Blank
MKTF - 211 07/26/12
58
Loading...