AW-Lake RT-50 Installation Manual

RT-50 Flow Transmitter
Installation, Operating &
Maintenance Manual
©2020 AW-Lake Company. All rights reserved. Doc ID: RT50M AN010720
Table of Contents
Unpacking................................................................................................ 3
Quick Start Guide .................................................................................... 3
Power ...................................................................................................... 3
Installation ............................................................................................... 3
Product Description ................................................................................. 4
Technical Specifications .......................................................................... 4
Battery Replacement ............................................................................... 5
Navigational Buttons ............................................................................... 5
Run Mode Screens .................................................................................. 6
Programming Menu Navigation ............................................................... 6
K-Factor ....................................................................................... 6
Linearizer ...................................................................................... 8
Input/Output ..................................................................................8
System ......................................................................................... 8
Bluetooth ......................................................................................9
Mobile Bluetooth App ............................................................................ 10
Menu Navigation ...........................................................................10
System Settings .............................................................................11
Linearizer ..................................................................................... 12
Device Log ................................................................................... 13
Date/Time Settings .......................................................................13
Display Settings ............................................................................13
Compter Toolkit ..................................................................................... 14
Connecting ...................................................................................14
Upgrading Firmware ....................................................................... 15
2
Safety Denitions and Information
Do not attempt to install or use your AW Gear Meters product until you have read the safety instructions in this section. Save this manual and keep it in an easily accessible place.
WARNING means that failure to follow this safety statement may result in extensive product damage, serious personal injury, or death.
CAUTION means that failure to follow this safety statement may result in minor or moderate personal injury, property or equipment damage.
NOTICE is a statement that informs about installation, operation, maintenance, performance issues, or general tips that are important but do not create a hazard or safety concern.
Unpacking
Separate the RT-50 Flow Display from packaging materials and check for any visual signs of damage. If you determine there are damages caused by shipping, file a claim with the shipping company. If the flow monitor appears to have been improperly assembled or does not operate properly, Contact AW Lake’s returns department for replacement or repair.
Quick Start Guide
CAUTION: As with any precision-engineered device, always use the RT-50 in accordance with the manufacturer’s instructions.
Power
3 lithium batteries provide the power to the RT-50. Units must be shipped with the battery unplugged. To power On, plug in the battery to the designated socket on the board marked BAT (See Figure 1). To replace the battery, refer to Changing the Battery on page 4.
Installation
Secure display to flow meter by screwing sensor into designated sensor hole or with proper fittings.
NOTICE: Before aaching the RT-50 to the ow meter, note any potenal clearance issues.
User must enter in the K-factor for the RT-50 before use. K-factors are found on meter calibration sheet. For detailed instructions refer to programming instructions in this manual. Once the K-factor is entered the unit is ready to read flow.
3
Product Description
The RT-50 is a Battery-operated flow sensor that shows Rate and Totals on an easy to read LCD display. With a new, user-friendly menu structure, this unit is simple to setup and use. A new feature with the RT-50 is its Bluetooth capabilities. User can do the setup in the palm of their hand and have a second display for checking flow information. The basic function of this display is to read the incoming frequency from the attached sensor and convert it to a flow reading.
Technical Specications
Flow Meter Compatibility
JV series positive displacement gear meters and TRG, TR and TW turbine meters.
Power
3.6V “C” cell battery, field replaceable; 3-year continuous use estimated average life, depending on options enabled.
Backup Battery
Lithium coin battery 3V. (CR2032) Stores user data and time when battery is disconnected.
Response Time
25 mS response, frequency dependent.
Temperature Rating
-4 to 140°F (-20 to 60°C) ambient,
175˚F (80˚C) maximum fluid temp.
Input Options
Inductive and Carrier frequency sensors
Setup Options
• Local display
• Computer connection (cable shown below)
• Mobile app through Bluetooth connection
4
Battery Replacement
To replace the battery, first open the unit and unplug battery from BAT connector. Be sure to push down on the locking tab which holds the connector in place. Next pull back on the battery bracket and carefully pull battery pack straight out of the unit. Place new battery pack into the unit and plug it into BAT connector. The RT-50 will initialize and power back on. Unit has a small coin cell backup battery that will keep totalizer values saved while main battery is disconnected.
Navigational Buttons
Figure 1: Battery Replacement
Left and Right Arrows
Allows user to switch between flow monitoring displays (ex. Rate, Total, Grand total screen) Also allows user to move left and right when entering values.
Up and Down Arrows
Allows users to move up and down when in the menu structure. Also used to increase and decrease values in programmable fields.
Enter Used to Enter into a selected menu option
F1 (Dynamic button with 2 functions)
-Takes user to Main Menu
-Go Back to previous menu
F2 (Dynamic button with 3 functions)
-Soft Key to go directly into specific menus
-Saves user parameters
-Go Home to Main Screen
RT-50
F1 F2
5