MAINTENANCE GUIDE FOR THE JV-XXCG FLOW METER
Cleaning, inspecting or repairing a JV-CG Series gear flow meter is easily
accomplished by following the procedures below.
1. Remove the Sensor (1) from the flow meter body Upper Housing (4).
Using a 3mm hex key, loosen the 2 mounting screws inside the sensor.
2. The JV-30 size and smaller use M6 Bolts (2), loosen them using a 5mm
hex key. The JV-60 uses M10 Bolts (2), loosen them using an 8mm hex
key. Keep the 2 opposing bolts near
few threads to avoid stress on the shafts and the locating pins during
housing separation. Please note the orientation of the Upper and
Lower housings with respect to each other so that the meter is
reassembled the same way.
3. Holding the Upper Housing (4), gently tap on the top of the 2 opposing
bolt heads to separate the Upper Housing (4) from the Lower Housing
(9). Do NOT use chisels or screwd
housings. This can cause damage to the meter bodies and meter
internal parts.
4. After separat
Clean out the o-ring groove, shaft holes and meter cavity.
5. After cleaning all parts completely, the Gears (6), Shafts (5) and
Locating Pins (7) can be reinserted. Check for free and easy rotation of
the gears.
6. Replace the O-Ring (8
FKM o-rings can be reused if they are not damaged.
Note: JVM-60CG Flow meters have 2 o-rings per meter.
7. During reassembly keep the meter housings as parallel as possible.
Make sure the housings are orientated the same way they were prior to
disassembly.
8. Replace the Washers (3) and the Bolts (2). Torque the bolts to 15Nm
for the12, 20 and 30 size meters and 30Nm for the 60
not force the meter housings together. Do NOT use a hammer or other
such device. Over tightening will not cause damage to the meter, but
may fatigue the bolts and/or restrict the operation if internal surfaces are
not completely clean.
9. After reassembly, gently blow
This should be clearly audible given a moderate background noise level.
ion, remove, clean and inspect the gears (6) and shafts (5).
). PTFE o-rings should always be replaced.
air through the meter so the gears spin.
the Locating Pins (7) engaged by a
rivers to split and pry apart the
size meters. Do
1
2
3
4
5
6
7
8
9
10. Clean any debris from the pickup holes before remounting the sensor (1)
to the Upper Housing (4). Use a 3mm hex key to tighten the mounting
screws in the sensor to mount the sensor on the housing. Note that
there is a correct and incorrect orientation for the sensor to be
mounted on the flow meter. Please see the AW Gear Meters website
for documentation if you do not have it. The orie
sensor’s connector with respect to
the outlet port of the w meter.
ntation is noted by the
2440 W. Corporate Preserve Dr. #600, Oak Creek, WI 53154 | www.aw-lake.com | Rev. 06/2017
JV-20CG w/ DH-B Shown
CROSS REFERENCE GUIDE
Table Format:
AG15030 AW Part #
Part - Qty. per meter
[Drawing Ref. #]
FOR JV-XXCG SERIES SPARE PARTS
Locating Pin - 2
[7]
Shaft - 2
[5]
Gear w/ Insert - 2
[6]
PTFE O-Ring - 1
[8]
FKM O-Ring - 1
[8]
Bolts - 6
[2]
Washers - 6
[3]
JVM-01CG JVM-10CG JVM-15CG
AG15030
AG01020 AG10020 AG15020 AG20020 AG30020 AG60020
ST-01CG Shaft ST-10CG Shaft ST-15CG Shaft ST-20CG Shaft ST-30CG Shaft ST-60CG Shaft
AG01010 AG10010 BG10021 BG10020 AG30010 AG60010
GR-01CG GR-10CG GR-02/1 GR-02 GR-30CG GR-60CG
AG01040 AG10040 AG60080 - 2 pcs
COT-01CG COT-10CG COT-60KG
AG01045 AG10045 AG60085 - 2 pcs
COV-01CG COV-10CG COV-60KG
SC-M6x1-30mm
Washer - M6 Hardened Washer
LP-CG Locating Pin AW Part Description
JVM-20CG JVM-30CG JVM -60CG
AG60030
LP-60CG Locating Pin
AG15040
COT-15CG
COV-15CG
FAS-SM10-1 - 12 pcs
M10x70mm
AG60055 - 12 pcs
r - M10 Hardened Washer
Washe
2440 W. Corporate Preserve Dr. #600, Oak Creek, WI 53154 | www.aw-lake.com | Rev. 06/2017