AW Gear Meters ZHM Series User Manual

ZHM Series Flow Meter Maintenance Guide
Cleaning, inspecting or repairing a ZHM Series gear flow meter is easily accomplished by following the procedures below.
sensor, IGO6HD-XX, which has 2 built-in pickups, and is attached by two sunken, hex-head screws.
- Note index marks (14). If symbols are absent, mark the body positions.
- Loosen the hex bolts (7) with a 5 mm key wrench - ZHM-04/8mm, ZHM 05+06/10mm, ZHM 07/14mm. Remove the bolts except for two opposing bolts (7)which should remain engaged by a few threads to avoid damaging the shaft when the 2 flow meter housings separate.
- Holding the upper housing (6), gently tap the opposing bolts (7) to separate the lower housing (1). Do not use chisels or screwdrivers
to split or pry apart the housings - this may damage the carbide parts.
- After separation, remove and inspect the gears (4) & shafts (3). Clean with solvent and a small brush. The gears should revolve freely on the shafts.
- After cleaning all parts completely, the gears & shafts can be reinserted. Check for free and easy rotation.
- Replace the O-ring (5) if deformed.
- Keep meter housings parallel during reassembly. Both locating pins must be in place (2). Align index marks (14).
- Replace hex bolts as follows:
ZHM 01-03 - 15 Nm. *(This is hand tight). ZHM 04 - 75 Nm. ZHM 05/06 -130 Nm. ZHM 07 -300 Nm.
- Over tightening will not cause damage but may restrict operation if internal surfaces are not completely clean.
- After reassembly - a brief air blast will spin the gears. This should be clearly audible given a moderate background noise level.
- Clean any debris from the pickup port, insert Hall Sensor all the way, and then back off 1/4 revolution and tighten lock nut. * Do not over tighten *
- With the dual sensor, the electrical connector should point in the flow direction after reassembly. The sensor should fit flush on the meter body with good hex screw alignment.
- Optional PTFE space fillers (10) are available to fill space behind the conversion fittings when multi-colored materials are being metered.
AW Gear Meters 8809 Industrial Drive, Franksville, WI 53126 web: www.awgearmeters.com
Tel: 262-884-9800 Fax: 262-884-9810 Email: awinfo@aw-lake.com
REV. 03/09 ZHM.DOC
Cross Reference Guide for ZHM Series Spare Parts for ZHM Meters
1 - Lower Housing (303 Stainless Steel)* Pt.# LH-**
2 - Stainless Steel Locating Pins Pt.# LP 3 - Tungsten Carbide Shafts Pt.# ST-** 4 - Stainless Steel Gear* Pt.# GR-** 5 - O-ring
Pt.# COV-** Black FKM O-ring COT-** White PTFE O-ring
6 - Upper Housing (303 Stainless Steel)*
Pt.# UH-** 7 - HEX Socket Cap Screw* Pt.# SC-** 8 - Port Adaptor/Conversion Fitting
Pt.# NPT-ZHM 01-03 12mm to 1/4" NPT
GRA-ZHM 01-03 12mm to 3/8" NPT
HEC-ZHM 01-03 12mm to 1/2" NPT
SAE-01-03 12mm to 3/4" SAE
NPT-ZHM 04 20mm to 3/4" NPT NPT-ZHM 05 33mm to 1.0" NPT
9 - Crushable Aluminum Washer
Pt.# WAL-ZHM 10 - Teflon Space Filler
Pt.# SF-ZHM 01
SF-ZHM 02/1-03 11 - Lock Nut for Hall Sensor Pickup Pt.# LN 12 - Hall Sensor Pickup
Pt.# HESKS1 (For ZHM 01-04)
HESLS1 (For ZHM 04 & up)
HDSXS1 (Double Effect Sensor) 13 - 5 Pin Connector for HES & HD Sensors
Pt.# AMP-HES 14 - Index Marks that appear on the Upper and Lower
Housings. This is not a spare part.
* Standard material is 303 stainless steel. A special 316TI
alloy is available and is indicated after the part number as follows: ZHM 02TI or GR 02TI.
** Indicates ZHM Series Flow Meter Size
AW Gear Meters 8809 Industrial Drive, Franksville, WI 53126 web: www.awgearmeters.com
Tel: 262-884-9800 Fax: 262-884-9810 Email: awinfo@aw-lake.com
REV. 03/09 ZHM.DOC