TEST EQUIPMENT REQUIRED SECTION 1 Page 3
FAULT FINDING & SERVICING NOTES SECTION 2 Page 3
VOLTAGE CHECKS WITH NO VALVE UNDER TEST SECTION 3 Page 7
SECTION 1 – TEST EQUIPMENT REQUIRED
(a) AVO Electronic Testmeter (or equivalent dc mean valve voltmeter).
(b) Valve CV491 (Standardised for Mutual Conductance at 16mA anode current).
(c) Model 7 or Model 8 AvoMeters (3).
(d) Power Valve capable of passing 100mA anode current
CV428.
(e) Resistor 680kΩ ± 5%
SECTION 2 – FAULT FINDING AND SERVICING NOTES (See WARNING on page 2)
NOTE: All measurements and tolerances stated do not include those of the testing
instrument, and where necessary, these should be ascertained particularly before
commencement of the calibration procedure. Where possible the recommended
instruments should be employed.
1. 500 c/s ac Supply Operation & its Relation to Servicing
Whilst the instrument is suitable for use use on 50 to 500 c/s ac supplies, service
and calibration should normally be carried out using a 220/230V 50 c/s supply.
The following features play a vital part in the correct operation of the instrument
on a 500 c/s supply.
(a) The two electrostatic screens (S1 and S2) on the ht transformer prevent
spurious mA/V readings and care must be taken when replacing a
transformer to ensure that these screens are connected as shown in the
Circuit Diagram (see Fig. 4).
(b) The separate cable forms lying side by side across the instrument ensure
that the grid circuit and its associated wiring is kept well apart from the ht
wiring to prevent the transference of energy from one circuit to the other
at high mains frequencies.
If, at any time, it is necessary to displace wiring within the instrument,
great care must be taken to ensure that it is replaced in its original
position.
(c) The 0.02µF (C1) and 0.02µF (C2) capacitors prevent spurious readings on
insulation ranges when the instrument is used at high mains frequencies.
Page 5
4
2. To Check Accuracy of Instrument
Before commencing servicing the instrument should be checked as follows:–
(a) Ensure that the mains ‘On/Off’ switch is in the ‘Off’ position.
(b) Connect the instrument to an ac voltage supply of 220/250V 50 c/s of known
magnitude.
(c) Set the instrument voltage adjustment to its appropriate position.
(d) Switch on, noting that the panel indicator is illuminated.
(e) Set the Mains Voltage Selector fine control such that the meter pointer lies
as near as possible to the centre of the ‘~’ zone.
(f) Connect the AvoMeter in series with the A1 link.
(g) Using the CV491 (standardised in accordance with Para.3) with 200V dc
anode volts, check that
i. For 16mA anode current the negative grid volts indication is within ±
5% of the standardised value.
ii. The slope (mA/V) is within ± 5% of the standardised value.
A reading of 8mA on the external instrument will be equivalent to a dc
current of 16mA through the valve (this is the value normally indicated by
the anode current controls when the meter is at its null position).
(h) Remove valve and external meter.
(j) Set the Neg.Grid Volts control at ‘40’.
(k) Connect a resistance of 680kΩ ±5% between grid and cathode sockets on
the top cap connector panel.
(m) Set the Circuit Selector switch to position ‘Gas’.
The panel meter should indicate full scale deflection ± 20%.
(n) Disconnect the resistance and switch off.
3. To Obtain Standard Figures for a Valve Using dc Supplies
Using the recommended AvoMeters, the valve should be connected as shown in
Fig. 1.
If unable to use the recommended meters ensure that those used are of
sub-standard accuracy, the current meter having a maximum voltage drop of
100mV and preferably scaled 0–25mA and the voltmeter a sensitivity of 1000Ω/V.
If rectified ac is used for the ht supply, it is essential that steps are taken to
ensure that the supply circuit is adequately smoothed (the Solartron Varipack is a
suitable source). The bias supply should be obtained from a suitable battery (note
polarity of connection). The heater supply for the valve may be ac or dc, but must
within ± 5% of the rated voltage.
(a) Set the grid bias voltmeter to read 9V.
(b) Adjust the ht supply to 200V, then by means of successive adjustments of
the bias and ht voltage controls, set the anode current at 16mA (the anode
voltmeter must read 200V). Note the new grid bias reading.
Page 6
5
(c) The standardised slope for the valve can now be obtained from:—
The difference between the two anode current readings (i.e. 1mA) over the
difference between the two grid voltage readings:—
Ia2 — Ia1
Vg1 — Vg
2
The result will generally be between 4 and 5mA/V (see Final Test Procedure).
For greater accuracy it is suggested that readings of grid voltage be plotted
against values of anode current between 10 and 20mA and the slope taken from
the curve at 16mA.
The valve should now be labelled as follows:—
Va = 200V dc
Ia = 16mA dc
Vg = . . . . . . . . . .
Slope = . . . . . . . .mA/V
Date . . . . . . . . . .
CV 491
The valve should be re–standardised daily when in use.
4. Construction
The instrument comprises two units in a hinged transit case, the lid of which is not
detachable. Electrical connection between the two units is effected by means of
two 5–way side by side cables.
5. Removal of the Instrument From its Case (See WARNING on page 2)
To facilitate servicing or calibration of the instrument, it is necessary to remove
both sections from the casing, this being accomplished by the removal of four
hexagonal headed bolts, which form the feet of the control unit, from the
underside of the case. The control panel will then be released. The valve panel can
be withdrawn from its section of the case by the removal of eight fixing screws
around its periphery.
Page 7
6
6. Simple Faults
SYMPTOMS POSSIBLE FAULT ACTION
(a) No dial light indication
No dial light
indication or meter
deflection on SET ~
setting of Circuit
Selector.
(b) No indication of meter
current
No indication of meter
current and protective
relay operates when
testing tetrodes or
penthodes.
7. Relay Operates and Fails to Clear
Should the relay operate due to a suspected faulty valve and fail to clear after
switching of an on again with no valve in panel, set Roller Selector switch to read
000 000 000 and remove top cap connecting lead. Switch instrument off and on
again.
No mains input.
Dial light bulb burnt out.
Fuse blown.
No anode volts at valve pin.
No anode volts at valve pin
but screen volts present.
Check mains connector.
Replace LP1.
Check Mains Voltage
Selector setting and replace
F1 and/or F2.
Check that links A1 & A2 are
tight and making firm
contact.
Check that links A1 & A2 are
tight and making firm
contact
If fault clears the most likely cause of the trouble is a short on the valve panel,
certain pin(s) being shorted out to earth by stray wire or solder, or a breakdown
in insulation.
If the fault still persists, however, check ht line for breakdown to earth between
Roller Selector Switch on Valve Panel and ht transformer on control unit.
8. Adjustment of Protective Relay
The relay should seldom require attention, but if for any reason parts are
replaced, the adjustment is simple, it only being necessary to position two 4BA
screws (see Fig. 3). It should be noted that the bobbins if replaced, should be
positioned such that the flux which they produce is additive.
Operational limits are as follows:–
(a) Anode overload — Relay should operate on 100V short circuit.
(b) Screen overload — Relay should operate on 60V short circuit.
(c) The relay should not arc excessively on a 200V short circuit on
anode or screen.
(d) The relay should not operate when checking a 120mA rectifier.
Before making any adjustments check that the lamp LP1 is operative.
When the instrument is used solely on a 110V supply, it may be preferable
to replace LP1 with a 100V, 15W pigmy lamp.
Page 8
7
9. Servicing the Valve Holder Panel
The Valve Holder Panel is connected electrically to the control panel by means of
two 5–way side by side cables. One of these cables embodies two thicker sections
(16/.012) for H+ and H– leads. Connections to tag boards on either unit are
shown in Fig. 2.
The wiring of the valve holders on the panel is in the form of nine separate loops,
all pins comprising a loop and linking in roller 1 of the Roller Selector Switch. This
form of loop connection is used likewise for pins 2–9, all nine circuits
approximating in length and following a similar route around the panel. These
loops are further loaded with beads of ferroxcube which sufficiently damp the loop
to prevent the valve under test breaking into parasitic oscillation. A diagrammatic
layout is shown in Fig. 2. Ferroxcube is also used on leads feeding the selector
switch, as a precaution against lf oscillation.
Where it is necessary to replace valve holders, these with the exception of the
B8B are fitted to the panel with nuts and bolts, and are thus easily removable.
Care should be taken to replace all wire in its original position.
10. Removal and Replacement of Knobs and Setting of Knob Skirts
To remove any knob, remove 6BA screw and spring washer. To remove knob
spindle an skirt, release locking pin. The switch nut is now accessible. To adjust
skirt, slacken lock nut, rotate skirt to desired position an re–tighten lock nut.
Reverse procedure to replace.
SECTION 3 — VOLTAGE CHECKS WITH NO VALVE UNDER TEST
Connect instrument to known 220/230V 50 c/s supply, ensuring that the mains ‘On/Off’
switch is in the ‘Off’ position, and adjust coarse and fine settings of the mains voltage
selector panel to match the supply voltage as accurately as possible. Set the Circuit
Selector to ‘Test’ and the Electrode Selector to ‘A1’ and proceed to check the relevant
electrode voltages as follows:–
1. Heater Voltages
(a) Connect the AvoMeter, switched to its ac voltage ranges, between H+ and H–
sockets on top cap connector panel.
(b) Switch on and rotate the Heater Voltage switch through the full range of values,
the external meter being set to the appropriate voltage range as required.
(c) The heater voltage reading on the meter should conform to the voltage limits
shown in the following table.
Due allowance must be made for the limits of accuracy of the measuring
instrument for each particular reading:–
Nominal
Volts
2
5
10
20
40
117
Actual ac
Volts
2.34
5.5
10.4
21.0
42.0
125.0
Limits
2.2 / 2.5
5.3 / 5.7
10.2 / 10.8
20.5 / 22
41 / 44
120 / 130
(d) Switch off and remove the meter.
Page 9
8
2. Anode Voltages
(a) Connect the AvoMeter, ensuring that the mains ‘On/Off’ switch is in the ‘Off’
position, and set to its ac voltage range, between A en C sockets on the top cap
selector panel.
(b) Switch on and rotate the ‘Anode Voltage’ switch through successive positions, the
meter being set to the appropriate range as required.
(c) The meter readings should be 1.1 × the voltage indicated by the ‘Anode Voltage’
switch –2 +6 per cent.
Due allowance must be made for the limits of accuracy of the measuring
instrument for each particular reading, e.g., with the ‘Anode Voltage’ switch set to
100, the actual voltage reading should be 110V –2 +6%.
(d) Switch off and remove the meter.
3. Screen Voltages
(a) Connect the AvoMeter, ensuring that the mains ‘On/Off’ switch is in the ‘Off’
position, and set to its ac voltage range between S en C on the top cap selector
panel.
(b) Short the anode of V1(a) to cathode (pins 2 and 5. See Fig. 3).
(c) Switch on and rotate the ‘Screen Voltage’ switch through successive positions, the
external meter being set to the appropriate range as required.
(d) The meter readings obtained should be 1.1 × the voltage indicated by the ‘Screen
Voltage’ switch –2 +6%.
Due allowance must be made for the limits of accuracy of the measuring
instrument for each particular reading.
(e) Switch off and remove the meter.
Page 10
9
CHAPTER 2
CALIBRATION
=========
C O N T E N T S
TEST EQUIPMENT REQUIRED SECTION 1 Page 9
CHECKING THE NEG. GRID VOLTS CONTROL SECTION 2 Page 9
CHECKING THE SET mA/V CONTROL SECTION 3 Page 9
CHECKING THE SET ~ INDICATION SECTION 4 Page 10
Ia CALIBRATION CHECK SECTION 5 Page 10
SETTING THE mA/V DIAL SECTION 6 Page 10
THE INDICATING METER SECTION 7 Page 11
SECTION 1 – TEST EQUIPMENT REQUIRED
(a) Avo Electronic Testmeter (or equivalent dc mean Valve Voltmeter)
NOTE: The Electronic Testmeter should be standardised at the appropriate voltages
before making any adjustment mentioned in the following paragraphs.
SECTION 2 – CHECKING THE NEG. GRID VOLTS CONTROL (See WARNING on page 2)
(a) Set mains ‘On/Off’ switch to the ‘Off’ position.
(b) Open LK (Fig. 3) and set the panel controls as follows:–
CIRCUIT SELECTOR to TEST, ELECTRODE SELECTOR to A1 and NEG. GRID
VOLTS to 40.
(c) Connect the Testmeter across RV2 (Fig. 3). Switch on and adjust RV3
(Fig. 3) until a voltage reading of 20.8V is obtained. Remove the Testmeter.
(d) Connect the Testmeter between G1 and C sockets on the top cap connection
panel, or if the panel has been disconnected for servicing, to the G1 and C
positions on the tag board at the back of the Unit.
(e) Check that at the 13 and 4 marks on the dial, readings of 6.75V and 2.1V
± 5% are obtained.
If either or both readings are out of tolerance, the dial should be adjusted
mechanically to split the error. If it is necessary to make an adjustment,
slacken the three counter–sunk screws on the top of the dial which will then
be free to move within the latitude of kidney shaped slots. After adjustment,
re–tighten screws and check readings. The areas marked 0, 5, 15 and 40
should correspond within the indicated area to 0V, 2.6V, 7.8V and 20.8V
± 5% respectively.
(f) Switch off and remove the Testmeter.
SECTION 3 – CHECKING THE SET mA/V CONTROL
(a) With LK still open, connect the Testmeter set to a suitable dc range, across
R5.
(b) Switch on and check that when the dial is advanced to its 10, 5 and 2mA/V
positions, readings of 52.5mV, 105mV and 250mV ± 3% are obtained.
(c) If for any reason the relationship between the dial and the potentiometer has
been upset, the procedure ‘Setting the mA/V Dial’ given in Section 6 should
be adopted.
Page 11
10
(d) Switch off and remove the Testmeter.
SECTION 4 – CHECKING THE SET ~ INDICATION
(a) Standardise the Testmeter at 47V dc. Close the link LK and set the panel
controls as follows:–
CIRCUIT SELECTOR to ‘SET ~’, and ELECTRODE SELECTOR to A1.
(b) Connect the Testmeter across RV2 and switch on.
(c) A reading of 47V should now be obtained, whilst the meter on the the
instrument panel should indicate within the ‘~’ zone.
If voltage reading is correct, but panel meter indication is outside ‘~’ zone,
check Resistors R3 and R4.
(d) Switch off and remove the Testmeter.
SECTION 5 – Ia CALIBRATION CHECK
(a) Open the A1 link on the valve base panel and insert a Model 7 AvoMeter, set
to a suitable dc range, into the circuit.
(b) Set up the instrument and place under test any power valve capable of
passing 100mA anode current, e.g., CV428.
(c) Set the Anode Current controls to 100mA (90mA and 10mA), switch on and
with the instrument set to its test position, allow the valve to warm up.
(d) Set the panel meter pointer to zero by means of the Neg.Grid Volts control.
(e) The external meter should indicate between 47.5mA and 52.5mA
(0.5 × indicated value on Anode Current control ± 5%) the panel instrument
indicating zero. If required repeat this test at any other settings of Anode
Current controls.
(f) Switch off and remove the meter.
SECTION 6 – SETTING THE mA/V DIAL
(a) With the link open and the ‘SET mA/V’ dial at rest, set RV1 at its maximum
anti-clockwise position (viewed from the front panel) and adjust friction tight
the locking nuts of the U-shaped stirrup.
(b) Connect the Testmeter, set to a suitable range across R5.
(c) Switch on and advance the ‘SET mA/V’ dial to a reading of 5.
(d) Rotate the RV1 spindle further, by means of the stirrup, in a clockwise
direction until the Testmeter gives a reading of 105mV.
(e) If this reading is achieved without further clockwise advancement of the
stirrup, or if its procurement necessitates an anti-clockwise movement of the
stirrup, then investigate the accuracy of R1, R2, R5 and RV1.
(f) The locking nuts on the stirrup should now be tightened and the reading of
105mV on the voltmeter checked.
(g) Again check that the dc millivolts developed across R5 at the 2mA/V and
10mA/V settings of the dial are 260mV and 52.5mV ± 3%.
(h) Check that the dial can now be rotated to its 1mA/V position and that the
motion is eventually arrested by the stop screw on the dial and not by the
stop at the end of the potentiometer track.
(j) Switch off and remove the meter.
Page 12
11
SECTION 7 – THE INDICATING METER
This is a self–contained unit which may be withdrawn from the control panel by the
removal of two 2BA screws (see Fig. 3).
When used in the instrument as an anode current null indicator, the meter has a full
scale deflection for approximately 10mA (not critical).
When removed from the instrument, the meter has a full scale deflection of 30µA and
internal resistance of 3,250Ω. When shunted by R9 only (see Circuit Diagram) the meter
has a full scale deflection of 39.8µA
Page 13
12
CHAPTER 3
FINAL TEST PROCEDURE
================
C O N T E N T S
TEST EQUIPMENT REQUIRED SECTION 1 Page 12
FINAL TEST DETAIL SECTION 2 Page 12
NOTE: When all repairs have been carried out and the instrument is re–assembled,
carry out the following final test detail.
SECTION 1 – TEST EQUIPMENT REQUIRED
(a) AVO Electronic Testmeter (or equivalent mean dc Valve Voltmeter).
(b) Model 7 or Model 8 AvoMeter.
(c) Valves CV428, CV491 and U52.
(d) Resistor 1 megohm ± 1%
(e) Resistor 680kohm ± 5%
SECTION 2 – FINAL TEST DETAIL (See WARNING on page 2)
(a) Apply a 500V megger test between the mains input and frame.
(b) Apply a known ac 50 c/s voltage in the range 200–250V to the instrument
and with Circuit Selector at ‘Set ~’, set mains voltage selector until needle
on panel registers as near as possible to the centre of the ‘~ mark’. Check
that selector settings show ± 5V of the actual mains voltage.
It is most important to ensure that the meter reading in the ‘Set ~’ position
is maintained at the centre of the ‘~ mark’ on the scale for all subsequent
tests.
(c) Set the Circuit and Electrode Selector switches as given in the table overleaf
and connect a 1 megohm ± 1% resistor across the electrodes on top cap
board as detailed under the heading ‘Condition’. Check that a leakage of
1 megohm ± 10% is indicated on the panel meter in each case:–
Page 14
13
Circuit Selector
Switch
A/R A1 Resistor connected across
S/R A1 Resistor connected across S
C.H/R A1 Resistor connected across
C/H C/H Resistor connected across C
NOTE: In positions C.H/R and C/H check that heater volts appear across H– and H+.
(d) Check operation of the overload cut–out with:
i. An anode/cathode short at an anode voltage of 100.
(Connect a short across A1 and C top cap board).
Electrode Selector
Switch
Condition
A1 and any of the following
S, H–, C, A2 or G.
and any of the following H–,
C or G.
C, H– or H+ and any of the
following A1, A2, S or G.
and either H+ or H–.
ii. A screen/cathode short at a screen voltage of 60.
(Connect a short across S and C top cap board).
(e) Check the following unloaded anode volts with the Testmeter
(on appropriate ac range) connected across A1 and C on top cap board).
Nominal Volts: 60 100 150 250 400
Actual ac Volts: 66 110 165 275 440
Limits: –2 +6%
(f) Check the following unloaded heater volts with the Testmeter
(on appropriate ac range) connected across H+ and H– on top cap board.
Nominal Volts Actual ac Volts Limits
2
5
10
20
40
117
(g) Insert a CV428 in the appropriate socket, set the instrument to the correct
Electrode Selector switch to A1 and obtain a balance the check that:
2.34
5.5
10.4
21.0
42.0
125.0
2.2 / 2.5
5.3 / 5.7
10.2 / 10.8
20.5 / 22
41 / 44
120 / 130
i. By varying the grid voltage, an anode current of 100mA is obtainable.
ii. The valve does not oscillate.
NOTE: The meter reading should not vary appreciably when the hand is
placed near or on the insulated anode lead.
iii. By varying screen voltage the anode current varies accordingly.
Page 15
14
(h) Insert a CV491 (12AU7) strapped as single triode and standardised for
mutual conductance at 16mA (anode current) with 200V applied to the
anode, connect the H.R. Testmeter (on 10mA ac range) across the A1 link.
Then check that:
i. For 16mA anode current on the CT160 the external Testmeter reading
is 8mA.
ii. Negative grid volts is ± 5% of the standard figure.
iii. Mutual conductance is ± 5% of the standard figure.
NOTE: The mutual conductance figure for the CV491 will normally be in the
order of 4mA per volt at an anode current of 16mA. The tolerance
specified for measurement of mutual conductance on the
instrument applies only to this point on the mutual conductance
control.
(j) Remove the CV491 and with the bias control set at 40 connect a 680kΩ
±5% resistor between G and C connections on top cap board.
Check that, with the Circuit Selector switch set to ‘Gas’ the CT160 meter
shows f.s.d. within 20%.
(k) Insert a U52 in the appropriate socket, set the Circuit Selector switch to test
and Electrode Selector switch to D1. Insert the Testmeter (on appropriate dc
current range) in series with the load from the top cap board (D1) to the
rectifier anode.
Check that for all load switch positions:
i. The CT160 meter reading is approximately in centre of ‘good’ scale.
ii. That the Testmeter readings are within ± 10% of nominal value.
(m) Repeat test as in (j) with anode load connected to D2 on top cap board and
Electrode Selector to D2.
Page 16
Page 17
Page 18
Page 19
Page 20
Page 21
Page 22
Page 23
Page 24
Page 25
Page 26
Page 27
Page 28
Page 29
Page 30
Page 31
Page 32
Page 33
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.