FX2P44H
_____
Ordering Information
-
Size
Display digit
Item
Specications
Model
Display digit4-digit6-digit8-digit
Character size (W×H)6×10mm4×8mm3.8×7.6mm
Power supply100-240VACᜠ 50/60Hz
Permissible voltage range90 to 110% of rated voltage
Power consump ion● 1-stage: max. 4.6VA
Max. counting speed of CP1/CP2 Selectable 1cps / 30cps / 2kcps / 5kcps (DIP switch)
Return timeMax. 500ms
Min. signal widthINHIBIT, RESET: approx. 20ms
Mechanical0.75mm amplitude at frequency 10 to 55Hz (for 1 min) in each X, Y, Z direction for 1 hour
Malfunction0.5mm amplitude at frequency 10 to 55Hz (for 1 min) in each X, Y, Z direction for 10 min
Mechanical300m/s
Malfunction100m/s
Ambient temp.-10 to 55℃, storage: -25 to 65
2
(approx. 30G) in each X, Y, Z direction for 3 times
2
(approx. 10G) in each X, Y, Z direction for 3 times
℃
Ambient humi.35 to 85%RH, storage: 35 to 85%RH
(€
1-stage settingApprox. 245g (approx. 180g)
※
1
2-stage settingApprox. 265g (approx. 200g)
,'i\lus
IndicatorApprox. 225g (approx. 160g)
Connections
FX M-1P4
※
1
CP2
9
(INHIBIT)
OUT
2
12VDC
50mA
10
3
CP1
CONTACT
250VAC 3A, 30VDC 3A
RESISTIVE LOAD
SOLID STATE OUT
30VDC 100mA
0VDC
RESET
11
128
5
4
SOURCE
100-240VAC
50/60Hz 4.6VA
13
14
7
6
FX M-2P4
•
□
12VDC
CP2
CP1
CONTACT
250VAC 3A, 30VDC 3A
RESISTIVE LOAD
※
1
9
(INHIBIT)
OUT2
2
50mA
10
3
0VDC
11
OUT1
4
SOURCE
100-240VAC
50/60Hz 5.8VA
Up/Down Counter/Timer
-
FX M-I4
•
CP1
1
※
(INHIBIT)
□
CP2
9
2
12VDC
50mA
10
3
0VDC
RESET
11
128
5
4
SOURCE
100-240VAC
50/60Hz 3.8VA
13
6
SOLID STATE OUT
30VDC 100mA
RESET
OUT1
13
128
5
6
OUT2
14
17
7
SENSORS
CONTROLLERS
MOTION DEVICES
SOFTWARE
(J)
Temperature
Controllers
(K)
SSRs
(L)
Power
Controllers
(M)
14
7
Counters
(N)
Timers
(O)
Digital
Panel Meters
(P)
Indicators
FX4H-1P4
• •
CONTACT
250VAC 3A,
30VDC 3A
RESISTIVE LOAD
SOURCE
100-240VAC
50/60Hz 4.6VA
※
INHIBIT: In case of timer mode, this terminal is for time hold.
1
OUT
2
3
4
5
6
716
817
918
※
(INHIBIT)
CP1
10
11
CP2
12VDC
12
50mA
0VDC
13
0VDC
14
RESET
15
SOLID
STATE
OUT:
30VDC
100mA
SOURCE
100-240VAC
50/60Hz 5.8VA
(voltage input (PNP): connect with 12VDC, no-voltage input (NPN): connect with 0VDC)
Autonics
FX4H-2P4
CONTACT
250VAC 3A,
30VDC 3A
RESISTIVE LOAD
CONTACT
250VAC 3A,
30VDC 3A
RESISTIVE LOAD
1
OUT2
OUT1
(INHIBIT)
2
3
4
5
6
7
8
918
(Q)
Converters
(R)
CP1
10
CP2
11
12VDC
12
50mA
0VDC
13
0VDC
14
RESET
15
16
17
OUT1
OUT2
}
SOLID
STATE
OUT:
30VDC
100mA
Digital
Display Units
(S)
Sensor
Controllers
(T)
Switching
Mode Power
Supplies
(U)
Recorders
(V)
HMIs
(W)
Panel PC
(X)
Field Network
Devices
M-81
FXH/FXM Series
Dimensions
FXM Series Panel cut-out Bracket
14.3
72
□
I
.
.
- .
"
FXH Series
17.6
4813
96
75
11.8
c=J~ii
==
•
c::,J c::,J
==~
=
70
=
=
.
0
ii
0
l
FXM Series
•
67.5
FXH Series
•
91.5
'
Min. 90
(FXM, FXH Series
Min. 90
+0.7
□
Min. 65
+0.6
45
Min. 115
0
68
+0.8
0
J
0
92
universal)
46
37.5
(unit: mm)
5
4
40.5
Input Connections
Voltage input (PNP)
● Solid-state input (standard sensor: PNP output type sensor)
Sensor
(PNP output)
※
CP1, CP2 (INHIBIT), RESET input part
Brown
Black
Blue
+12V
※
Inner circuit
10.8kΩ
0V
No-voltage input (NPN)
● Solid-state input (standard sensor: NPN output type sensor)
Sensor
(NPN output)
※
CP1, CP2 (INHIBIT), RESET input part
M-82
Brown
Black
Blue
Counter/Timer
※
+12V
5.4kΩ
Inner circuit
0V
SensorCounter/Timer
(PNP open collector output)
Sensor
(NPN open collector output)
Brown
Black
Blue
Brown
Black
Blue
Counter/Timer
※
Counter/Timer
※
Autonics
10.8kΩ
5.4kΩ
+12V
Inner circuit
0V
+12V
Inner circuit
0V
● Contact input
Counter/Timer
※
※
Counting speed
: Set as 1 or 30cps
+12V
Inner circuit
10.8kΩ
0V
● Contact input
Counter/Timer
※
※
5.4kΩ
Inner circuit
Counting speed
: Set as 1 or 30cps
+12V
0V
Input & Output Connections
When operation load by sensor power
Load1
PR18-5DP
(power for
operating load)
● The sum of operating current capacity of load 1 and
sensor should not be over external power capacity
(50mA).
8 9 10 11 12 13 14
+12VDC
CP1 CP2
N.O.
1 2 3 4 5
Load2
0VDC
N.C.
COM
6
SOURCE
7
Up/Down Counter/Timer
(Q)
When operating load by external power
(+)
Constant-voltage
Load1
(-)
※
8 9 10 11 12 13 14
+12VDC
CP1 CP2
N.O.
0VDC
N.C.
COM
1 2 3 4 5 67
(power for
opera ing load)
Load2
SOURCE
● The capacity of load 1 should not be over transistor
switching capacity (max. 30VDC, 100mA)
● Do not supply the reverse polarity power.
when using inductive load (relay, etc.), connector surge
※
absorber at both ends of the load 1
circuit
SENSORS
CONTROLLERS
MOTION DEVICES
SOFTWARE
(J)
Temperature
Controllers
(K)
SSRs
(L)
Power
Controllers
How to count by external power supply
This unit starts to count when [H] (5-30VDC) is applied at CP1 or CP2 after selecting PNP.
(+)
Limit switch
contact
89 10 11 12 13 14
CP1 CP2
+12VDC
Using 2 counters with one sensor
ID
Please connect as the power of sensor is supplied from only one of counters and design input logic with same way.
Set one-shot output time by front TIME volume switch.
4
ON
Up mode
OFF
RESET
2nd setting
1st setting
0
OUT1
OUT2
(OUT)
RESET
2nd setting
1st setting
0
OUT1
OUT2
(OUT)
RESET
2nd setting
'
1st setting
--Vk12J;-
I I I I
0
OUT1
OUT2
(OUT)
RESET
2nd setting
1st setting
OUT1
OUT2
(OUT)
RESET
2nd setting
1st setting
OUT1
OUT2
(OUT)
RESET
2nd setting
1st setting
OUT1
OUT2
(OUT)
RESET
2nd setting
1st setting
OUT1
OUT2
(OUT)
RESET
2nd setting
1st setting
OUT1
OUT2
(OUT)
RESET
2nd setting
1st setting
OUT1
OUT2
(OUT)
RESET
2nd setting
1st setting
OUT1
OUT2
(OUT)
■
I I I I
-2Ild}([
--~
-~ -~
1 I I I I I I
0
c:i c:i
II II II
■ ■ ■
0
0
I I I I
--
-tzfkd1};-
0
r:±:i r:±:i
I I I I
-
0
0
''
n n
-ktk1dk
0
' '
r::::,
■ ■
-~
c:i
■
''
r:::
--
M-86
p-
Self-holding output
One-shot output of OUT1
(0.5 sec xed)
4
ON
Down mode
OFF
RESET
2nd setting
1st setting
0
OUT1
OUT2
(OUT)
RESET
2nd setting
1st setting
0
OUT1
OUT2
(OUT)
RESET
2nd setting
1st setting
---
--
-~
0
OUT1
OUT2
(OUT)
RESET
2nd setting
1st setting
OUT1
OUT2
(OUT)
RESET
2nd setting
1st setting
OUT1
OUT2
(OUT)
RESET
2nd setting
1st setting
OUT1
OUT2
(OUT)
RESET
2nd setting
1st setting
OUT1
OUT2
(OUT)
RESET
2nd setting
1st setting
OUT1
OUT2
(OUT)
RESET
2nd setting
1st setting
OUT1
OUT2
(OUT)
RESET
2nd setting
1st setting
OUT1
OUT2
(OUT)
' ' '
II II II
■ ■ ■
----
- --
-~~~I
0
' ' '
Ill
1.-:!:J 1.-:!:J 1.-:!:J
II II II
■ ■ ■
0
0
I I I I
--
----
--
0
~
0
0
''
n n
---
- - - - - -
0
~
'
ir:1
t::::::::,
------
----
-
-- --
''
' lei
Self-holding output
Operation
After count-up, counting display value increases
or decreases until reset signal input is applied
and self-holding output is maintained.
After count-up, counting display value and selfholding output are maintained until reset signal
input is applied.
When count-up, counting display value is reset
and it counts simultaneously.
Self-holding output of OUT1 turns OFF after oneshot output time of OUT2.
One-shot output time of OUT1 is regardless of
OUT2 output.
After count-up, counting display value is reset
after one-shot output time of OUT2 and it counts
simultaneously. Self-holding output of OUT1 turns
OFF after one-shot output time of OUT2.
One-shot output time of OUT1 is regardless of
OUT2 output.
After count-up, counting display value increases
or decreases until reset signal input is applied.
Self-holding output of OUT1 turns OFF after oneshot output time of OUT2.
One-shot output time of OUT1 is regardless of
OUT2 output.
After count-up, counting display value
is maintained while OUT2 output is ON.
Counting value is internally reset and it counts
simultaneously. When OUT2 output is OFF,
displays counting value while OUT2 output is ON,
and it increases or decreases. Self-holding output
of OUT1 turns OFF after one-shot output time of
OUT2.
After count-up, counting display value increases
or decreases during one-shot time of OUT2. Selfholding output of OUT1 turns OFF after one-shot
output time of OUT2.
One-shot output time of OUT1 is regardless of
OUT2 output.
● Up, Up/Down-A, B, C input mode
: OUT1 output maintains ON when counting
display value is larger or equal than 1st setting
value. OUT2 output maintains ON when
coun ing display value is larger or equal than
2nd setting value.
● Down, Up/Down-D, E, F input mode
: OUT1 output maintains ON when counting
display value is smaller or equal han 1st
setting value. OUT2 output maintains ON
when counting display value is smaller or
equal than 2nd setting value.
OUT1 and OUT2 turns OFF→ON→OFF
repeatedly (icker).
I::
Up/Down Counter/Timer
Counting & Time Operation For Indicator (FX M-I4)
□
Counting operation
● Input mode: Up
RESET
+Display
value
~
Wv1
0
n
● Input mode: Up / Down-A, B, C
RESET
+Display
value
0
-Display
value
● Input mode: Down
RESET
+Display
value
0
-Display
value
● Input mode: Up / Down-D, E, F
RESET
+Display
value
0
-Display
value
Time operation
● Up mode● Down mode
RESET
NHIBIT
+Max. time
range
0
RESET
INHIBIT
+Max. time
range
-Max. time
range
~ ~
0
~~
SENSORS
CONTROLLERS
MOTION DEVICES
SOFTWARE
(J)
Temperature
Controllers
(K)
SSRs
(L)
r,
Power
Controllers
(M)
Counters
(N)
Timers
Dot for Decimal Point / Hour, Min, Sec
RESET
3 sec
Counter mode
----
RUN mode
RESET
DP
RESET
Set decimal point
RESET
by front
3 sec
Setting mode
Timer mode
CLRSET
Hour, Min, Sec are not
divided wi h dot.
E.g.)
: 59 min 59 sec
5959
RESET
※
In run mode, hold the
enters setting mode [DP].
※
In setting mode, hold the
saves the setting and returns to RUN mode.
※
If here is no
setting mode, it returns to RUN mode.
Hour, Min, Sec are divided with dot.
E.g.)
)5(59
: 59 min 59 sec
RESET
key for over 3 sec, and it
c=i
RESET
key for over 3 sec, and it
c=i
RESET
key input for 60 sec when entering
c=i
Error Display and Output Operation
Error DisplayError descriptionTroubleshooting
ERR0
Setting value is 0.Change the setting value anything but 0.
※
When error occurs, the output turns OFF.
※
When 1st setting value is set as 0 (zero), OUT1 maintains OFF.
When 2nd setting value is smaller than 1st setting value, 1st setting value is ignored and only OUT2 output operates.
※
Indicator model does not have error display function.
(O)
Digital
Panel Meters
(P)
Indicators
(Q)
Converters
(R)
Digital
Display Units
(S)
Sensor
Controllers
(T)
Switching
Mode Power
Supplies
(U)
Recorders
(V)
HMIs
(W)
Panel PC
(X)
Field Network
Devices
Autonics
M-87
FXH/FXM Series
Proper Usage
● Follow instructions in ‘Proper Usage’. Otherwise, it may cause unexpected accidents.
● Use the product, 0.1 sec after supplying power.
● When supplying or turning o the power, use a switch or etc. to avoid chattering.
● Install a power switch or circuit breaker in the easily accessible place for supplying or disconnecting the power.
● In case of contact input, set count speed to low speed mode (1cps or 30cps) to operate.
If set to high speed mode (2kcps or 5kcps), counting error occurs due to chattering.
● Keep away from high voltage lines or power lines to prevent inductive noise.
In case installing power line and input signal line closely, use line lter or varistor at power line and shielded wire at input
signal line.
Do not use near the equipment which generates strong magnetic force or high frequency noise.
● This product may be used in the following environments.
Indoors (in the environment condition rated in ‘Specications’)
①
Altitude max. 2,000m
②
Pollution degree 2
③
Installation category II
④