Atec TCPA User Manual

AC/DC Current Measurement Systems
TCPA300 • TCP312 • TCP305 • TCP303 • TCPA400 • TCP404XL Data Sheet
Status Indicators provide Visual Operating Status and Notication of Potential Error Conditions – Degauss, Probe Op en, Overload, Not Terminated into 50 , Noncompatible Probe Type
Lower DC Drift and Noise allows Improved Low-level Current Measurements
Features & Benets
AC/DC Measurement Capabilities
DC – 100 MHz, Current Probe Amplier (TCPA300) uses:
DC – 100 MHz, 30 A DC (TCP312) DC – 50 MHz, 50 A DC (TCP305) DC – 15 MH
DC – 50 MHz, Current Probe Amplier (TCPA400) Use s:
DC – 2 MHz, 750*1A DC (TCP404XL) (500 A D C Continuous)
Automatic Scaling and Units*2– Oscilloscope On-screen Readout of Magnitude and Amps reduces Measurement Errors with No More Hand Calculations
AC/DC In
Low Insertion Impedance reduces Device Under Test Loading
Split-core Construction allows Easy Circuit Connection
put Coupling
Certied for IEC Standards.
use in U.S., Canada, and Europe. C omplies with applicable
Applications
Developmen and Service Personnel in Telecomm, Data Comm, Computer, and Semiconductor P ower Electronics Environments for:
Power S upplies (Switching an d Linear) Semiconductor Devices (SCRs, IGBTs, MOSFETs, CMOS, BJTs) Power Inverters/Converters Electroni Industrial/Consumer Electronics Mobile Communications (Phone, Satellite, Relay Stations) Motor Drives Transportation Systems (Electronic Vehicles, Electric Trains, Locomotives, Avionics)
Increased Performance and Simplicity
The TCP30 is a highly advanced current measurement system for today's current measurement needs. When connected to Tektronix oscilloscopes with TEKPROBE Level II, TekConnect (w/ TCA-BNC), or TekVPI (w/ TPA-BNC) interfaces, current measurements and calculations are simple and easy.
d with duty cycle.
*1Derate
2
*
Requires a TDS TEKPROBE oscilloscope or a TekConnect oscilloscope with TCA-BNC.
t and Analysis Solutions for Designers, Installers,
cBallasts
0 and TCP400 Series AC/DC current measure m ent family
Data Sheet
Meets Today's AC/DC Current Measurement Applications
The TCPA300 amplier, when used w ith TCP312, TCP305, or TCP303 probes, provi
des a wide range of current measurement capability and spans the gap between low-level milliamp measurements to very high current levels. These three probes provide current measurement capabilities of 30 A, 50 A, and 150 A DC continuous. For even higher current levels, the TCPA400 amplier with the TCP404XL current probe measures 500 A D C continuous and 750 A DC continuous, derated with duty cycle.
Higher frequency performance is available with the TCP312 w/TCPA300 providing 100 MHz bandwidth and a maximum current of 30 A DC.
Measurement Errors and Manual Calculations are Now a Thing of the Pa
st
With this new series of current measurement tools, automatic control and on-screen sca
ling and units is provided for users of Tektronix TDS3000, TDS500, T DS600 , TDS700, TDS5000, TDS6000, and TDS7000B Series oscilloscope systems (the DPO3000, DPO4000, and DPO7000 Series oscilloscopes, the TPA-BNC adapter is required).
The TCP300/TCP400 current measurement systems seamlessly integrate with your TDS Series oscilloscope.
Even non-TEKPROBE systems can use the TCPA300/400 Series to make proper current me asurements by simply multiplying the measured output voltage on the oscilloscope by the TCPA300/400 Series range setting.
Characteristics
Characteristic TCP312 w/ TCPA300 TCP305 w/ TCPA300 TCP303 w / TCPA300 TCP404XL w/ TCPA400
Bandwidth (–3 dB) DC – 100 MHz DC – 50 MHz DC – 15 MHz DC – 2 MHz* Rise Time 3.5 ns Maximum Current Ratings: High-current Sensitivity Range 10 A/V Range 10 A/V Range 50 A/V Range 1 A/mV Range
DC (Continuous) RMS (Sinusoidal) Peak Pulse 50 A 50 A 500 A 750 A
Low-current Sensitivity Range 1 A/V Range 5 A/V Range 5 A/V Range N/A
DC (Continuous) RMS (Sinusoidal) Peak Pulse 50 A 50 A 500 A
DC Accuracy (Operating temp 0 °C to 50 °C)
DC Accuracy, Typical (Operating temp 23 °C ±5 °C)
l
Nomina
Maximum Bare Wire Voltage For use with insulated wires only
t Measurable Current
Lowes
3% accuracy at DC)
(at ±
Insertion Impedance (See curves below)
30 A 50 A 150 A
21.2 A 35.4 A 150 A 500 A
5A 25A 25A
3.5 A 17.7 A 17.7 A
±3% of reading ±3% of reading ±3% of reading (10 °C to 50 °C)
±1% of reading ±1% of reading ±1% of reading ±1% of reading
1mA 5mA 5mA 1A
e set to 1 mV/div and 20 MHz BW Limited
Scop
at 1 MHz
0.08
at 10 MHz
0.15
at 50 MHz
0.27 at 100 MHz
0.7
7ns
5 at 1 MHz
0.03
at 10 MHz
0.12
0.4
at 50 MHz
23 ns 175 ns
+3%/–6% of reading (0 °C
to <10 °C)
600 V
300 V
e set to 1 mV/div and 20 MHz BW Limited
Scop
at 1 MHz
0.01
5 at 5 MHz
0.02 at 15 MHz
0.1
CAT I & II
RMS
RMS
CAT III
500 A (750 A*4)
±3% of reading
at 10 kHz
0.1 m
at 100 kHz
0.6 m
8mΩ a
16 m
3
N/A N/A N/A
t1MHz
at 2 MHz
2 www.tektronix.com
AC/DC Current Measurement Systems — TCPA300 • TCP312 • TCP305 • TCP303 • TCPA400 • TCP404XL
Characteristic TCP312 w/ TCPA300 TCP305 w/ TCPA300 TCP303 w / TCPA300 TCP404XL w/ TCPA400
Typical
Maximum Amp-Second Product (Based on Amplier Range setting)
AC-coupling Low-frequency Bandwidth (Low Pass – 3 dB point)
Displayed RMS Noise (at 20 MHz bandwidth limit)
Signal Delay (Delay to output BNC)
Inputs (Probe amplier) Probe Open Indicator Overload Indicator Termination Indicator Yes Noncompatible Probe Indicator Yes
Safety Certications
U.S. NRTL Listing
Canadian Certication CAN/CSA C22.2 No.1010.1
European Union Compliance EN61010-1/A2
Other IEC61010-1/A2 (Amplier) IEC61010-2-032
Electromagnetic Compatibility EC Council Directive 89/336/EEC, FCC Part 15, Subpart B Class A, AS/NZS 2064.1/2. Power Requirements
(TCPA300/TCPA400 Ampliers) Power Requirements (Probes) Requires TCPA300 Amplier Requires TCPA400 Amplier Probe Model Warranty 1Year
Probe Mechanical Characteristics
Probe Cable Length 1.5 meters (60 inches) 2 meters (78.7 inches) 8 meters (315 inches) Probe Jaw Size (Max conductor
size) Length Width Height Weight
TCPA300 an d TCPA400 Mechanical Characteristics
Length Width Height Weight
50 A*μS–1A/V
500 A*μS–10A/V
500 A*μS–5A/V
NA–10A/V
3,000 A*μS–5A/V
15,000 A*μS–50A/V
<7 Hz
250 μA
RMS
1.25 mA
RMS
2.5 mA
RMS
17 ns 19 ns 40 ns 80 ns
1 Yes Yes
UL3111-1
(Amplier)
UL 3111-2 -03 2 ; UL 3111-1
(Probe and Amplier)
CAN/CSA C22.2 No.1010.2.032
(Amplier)
CAN/CSA C22.2 No.1010.1
(Probe and Amplier)
EN61010-1/A2; EN61010-2-032
(Amplier)
EN61010-1/A2
(Probe and Amplier)
IEC61010-1/A2
(Probe and Amplier)
90 V to 264 V; 47 to 440 Hz; 50 W
Maximum CAT II (Auto Switch)
TCP312 TCP305 TCP303 TCP404XL
3.8 mm (0.15 inches) 21 mm × 25 mm (0.83 × 1.0 inches)
20 cm (7.87 inches) 26.8 cm (10.55 inches) 26.8 cm (10.55 inches)
1.6 cm (0.625 inches) 4.1 cm (1.60 inches) 4.1 cm (1.60 inches)
3.2 cm (1.25 inches) 15.6 cm (6.13 inches) 15.6 cm (6.13 inches)
0.15 kg (0.33 lb.) 0.66 kg (1.45 lb.) 0.88 kg (1.90 lb.)
17.3 cm (6.8 inches)
16.7 cm (6.6 inches)
9.14 cm (3.6 inches)
1.14 kg (2.5 lb.)
NA – 1 A/mV
250 mA
RMS
www.tektronix.com 3