Classic Tenney Series - Wide Range Testing
Classic Tenney Series
The Tenney Classic Series of temperature and humidity test chambers deploys environmentally friendly technology that uses CFC-free refrigerants. The time­proven mechanical refrigeration systems are available in single or cascade systems and allow for years of worry-free use.
The Classic Series temperature and humidity test chambers are available in 10, 20, 30, 40, and 64 cubic feet models, all of which are available with single or cascading refrigeration systems.
• Semi hermetic refrigeration system
• Vertical recirculation conditioning system
• Automatic defrost for door gaskets
• Control tolerance between ±0.3°C and ±2% RH typical after stabilization
• Non-corroding parts on all humidity capable units
• Low water protection
10, 20, and 30 cubic foot models
Model Non-
humidifi ed
T10S T10C T20S1.5 T20C-3 T20C-4 T30S T30C
Humidifi ed T10RS T10RC T20RS1.5 T20RC-3 T20RC-4 T30RS T30RC
Workspace W 24(610) 30(762) 36(914)
D 27(686) 32(813) 40(1016) H28(711) 36(914) 36(914)
Exterior W 30(762) 41(1041) 46(1168)
D 47(1194) 55(1397) 64(1625)
H 74(1880) 84(2134) 84(2134) Temperature & Humidity Low °C -40°C -73°C -40°C -73°C -73°C -40°C -73°C High °C +200°C +200°C +200°C +200°C +200°C +200°C +200°C Change Rates in Minutes Time starting at
Ambient to listed temperature. Based on empty chamb
er.
200°C 30 30 60 60 60 70 70 150°C 20 20 40 40 40 50 50
93°C 15 15 25 25 25 25 25 65°C 10 10 15 15 15 15 15
C 105401054010
-18°C 20 10 60 15 10 60 15
-40°C 60 20 90 30 20 90 25
-54°C —25—4025—35
-65°C —35—5035—50 Live Load Capacity in Watts (Humidity System Off) Temperature +10°C 600 1275 600 1200 1800 600 1600
-18°C 450 1000 450 1000 1500 500 1400
-40°C 100 850 100 850 1275 250 1200
-54°C 600 700 1050 1000
-65°C 400 500 750 700 Utilities, etc Refrigeration (1)1.5HP (2)1.5HP (1) 1.5HP (2) 1.5HP (2) 2HP (1) 2HP (2) 2HP Heater Capacity 2.5 KW 2.5 KW (3 KW If 208 Option) Humidifi er KW 1 KW 2 KW
GPH 0.5 1 AMPS @ 230V. 1 Ph 28 30 32 34 35 34 35 AMPS Fuse 35 40 45 Unit Weight LBS. 850 950 1250 1350 1400 1500 1500
Overall dimensions in inches (mm) for the 10, 20, and 30 cubic foot models:
Options:
• IEEE interface
• LinkTenn software for Windows
• RS-422, 423, 232 or 485 interface assemblies
• 6-event relay board at 1 amp each
• Water demineralizer
• Water reservoir for humidity system (5 gallon)
• Recirculating system for humidity water
• Viewing window, thermally insulated and heated
• Interior light
• Shelving, adjustable and removable
• Water-cooled refrigeration system
• Alternate power supply wiring
• Automatic CO2 or LN2 cooling boost system
• Additional ports, gloves, etc.
• 5” or 6” rubber tire casters (3” phenolic casters are standard)
• GN2 purge system
• Refrigeration taps and pressure gauges
• Remote console for instrumentation
• Recording instruments
• Redundant thermal protection and alarm system
• External dryer for obtaining humidity as low as 5% (to 20°C)
• Alternate refrigeration and/or heating systems for faster temperature change rates or increased product load handling capabilities
Humidity capability: 20% to 98% RH in the dry bulb range of +20°C or +85°C as limited by a 3°C dew point. Test data based on 24°C ambient, sea level, 60 Hz. On 50 Hz or higher than 24°C, ambient performance may be reduced. Consult factory regarding any special cooling requirements.
®