Atec PA4000 User Manual

Power Analyzer
PA4000 Datasheet
Applications
Power conversion
Power generation
Inverters
Motor drives
Electrical propulsion
UPS
Frequency converters
Tektronix PA4000 Power Analyzers deliver highly accurate, multi-channel power, energy, and efficiency measurements. Precisely-matched inputs, unique Spiral Shunt™ technology, and advanced signal processing deliver high accuracy even with highly modulated waveforms and crest factors as high as 10. The versatile PA4000 offers comprehensive power measurements. Dual current shunts provide optimal resolution from microwatts to kilowatts. Harmonics analysis up to the 100th harmonic, and motor analysis with torque and speed inputs are included in the standard instrument. Every PA4000 comes with multiple PC interfaces, PC software, and USB flash drive support to help you collect and analyze data.
Key features
1 to 4 Input Modules allows several configurations to match your application
High measurement accuracy of 0.04% (basic voltage & current accuracy) for demanding test requirements
Dual internal current shunts for each module maximize accuracy for high- and low-current measurements
Unique Spiral ShuntTM design maintains stability over variations in current, temperature, etc. (patent applied for)
Proprietary frequency detection algorithms ensure rock-solid frequency tracking even on noisy waveforms
Electric and hybrid vehicle
High-efficiency lighting
Consumer electronics
Standby power
Flexible voltage and current input choices to fit your application
The PA4000 is the only instrument of its kind to feature both high- and low­range internal current shunts in each input module. The 30-Amp shunt is perfect for many applications and can accept up to 200 A peak; but for measuring low-current devices the 1 A shunt provides increased resolution and accuracy for measurements down into the microamp range. For measuring currents over 30 Amps, a selection of matching external current transducers are available in several styles, including high-accuracy transducers up to 1000 A.
The PA4000's voltage inputs can accept up to 1000 V continuous.
RMS
, 2000 V
peak
,
Application-specific test modes simplify instrument setup and reduce the likelihood of user error
Easy data export to USB flash drive or remote PC software, for reporting and/or remote control
Many standard features such as comm ports and harmonic analysis eliminate costly upgrade options
PA4000 Rear Panel - Input Modules
Datasheet
Unique Spiral Shunt™ technology (patent applied for)
The PA4000 employs an innovative Spiral Shunt design that ensures stable, linear response over a wide range of input current levels, ambient temperatures, crest factors, and other variables. This new design is superior to other shunt technologies and contributes to the instrument's reliable accuracy and repeatability over a wide range of signal conditions common in today's power conversion technologies. The spiral construction not only minimizes stray inductance (for optimum high frequency performance) but also provides for high overload capability and improved thermal stability.
Application-specific test modes
Some applications require special instrument settings to ensure proper measurements. The PA4000 simplifies setup for these applications by automatically choosing instrument settings and parameters that are optimized for each type of measurement application, resulting in more reliable measurement results with less opportunity for user setup error.
Selection of application-specific test modes.
Spiral Shunt for 30 A input
PWM motor drive mode
PWM motor mode is designed to overcome the difficulties associated with making measurements on the complex waveforms commonly found on the motor drives. High frequency sampling is combined with digital filtering to reject the carrier frequency and extract the motor frequency while still using prefiltered data for power parameters
Standby power mode
Driven by consumer demand and energy efficiency regulations (such as ENERGY STAR), there is an ever-increasing need to measure power consumption of products while they are in standby mode. One of the most widely used standards for measurement is IEC 62301. Part of this standard requires the measurement of power over a prolonged period of time without missing any short duration power events. The PA4000 standby power mode provides continuous sampling of voltage and current to produce an accurate Watts measurement over the user specified period.
Spiral Shunt for 1 A input
2 www.tektronix.com
PA4000 Power Analyzer
Ballast mode
Ballast mode synchronizes measurements for highly modulated electronic ballast waveforms. In modern electronic lighting ballasts, it is often difficult to make accurate measurements because the output signals are high frequency waveforms that are heavily modulated by the power frequency. Ballast mode provides a way of locking the measurement period to the power frequency.
Integrator mode
Integrator mode is used to provide measurements for determining energy consumption (Watt-hours, Ampere-hours, etc.). In addition, for certain parameters, average values are also available.
Standard harmonics analysis
The PA4000 features harmonics analysis to the 100th harmonic as a standard feature. Harmonics, THD, and related measurements can all be analyzed simultaneously with other power parameters. Accurate measurement of harmonic amplitude, phase, and harmonic power are included, which are critical to the analysis of losses in rotating machinery.
Standard communication ports
The PA4000 comes standard with USB, Ethernet, and RS-232 communication ports, plus a front-mounted USB port for data export to a flash drive. GPIB is available as a factory-installed option.
Harmonics bar chart display mode.
PA4000 rear panel, with communication ports
www.tektronix.com 3
Datasheet
Specifications
Measurements
Available measurements
Voltage and current ranges
Voltage Ranges 2000 V
Current Ranges (30 A Shunt) 200 A
Current Ranges (1 A Shunt) 5 A
Current Ranges (External Shunt)
V
- Volts RMS VDF - Volts Distortion Factor
RMS
A
- Amps RMS VTIF - Volts Telephone Influence Factor
RMS
WATT - Watts ATHD - Amps Total Harmonic Distortion
VA - Volt-Amps ADF - Amps Distortion Factor
VAR - Volt-Amps Reactive ATIF - Amps Telephone Influence Factor
FRQ - Frequency VF - Fundamental Volts rms
PF - Power factor AF - Fundamental Amps rms
VPK+ - Volts peak (positive) IMP - Impedance
VPK- - Volts peak (negative) RES - Resistance
APK+ - Amps peak (positive) REA - Reactance
APK- - Amps peak (negative) HR - Integrator time
VDC - Volts DC WHR - Watt Hours
ADC - Amps DC VAH - VA Hours
VRMN - Volts rectified mean VRH - VAr Hours
ARMN - Amps rectified mean AHR - Amp Hours
VCF - Voltage crest factor WAV - Average Watts
ACF - Current crest factor PFAV - Average Power Factor
VTHD - Volts Total Harmonic Distortion CVAR - Correction VArs
3 V
peak
peak
peak
peak
, 2 A
, 1.5 V
, 1000 V
, 100 A
peak
, 1 A
peak
peak
, 50 A
peak
peak
, 0.75 V
, 500 V
peak
, 0.5 A
peak
peak
, 20 A
peak
, 0.3 V
, 200 V
peak
, 0.2 A
peak
peak
, 10 A
, 0.1 A
peak
, 0.15 V
, 100 V
, 5 A
peak
peak
, 50 V
peak
peak
, 0.25 A
peak
, 0.075 V
, 2 A
peak
, 20 V
peak
peak
peak
, 0.03 V
peak
, 1 A
peak
, 0.0125 A
peak
, 5 V
peak
, 0.5 A
, 0.005 A
peak
, 0.015 V
peak
, 2 V
, 0.2 A
peak
peak
, 0.1 A
peak
, 0.0025 A
peak
, 0.0075 V
peak
peak
peak
Measurement accuracy - voltage and current
Voltage Accuracy, V
RMS
± 0.04% of Reading ± 0.04% of Range ± 0.02 V
( 45 Hz - 850 Hz )
Voltage Accuracy, V
RMS
± 0.05% of Reading ± 0.05% of Range ± (0.02*F)% of Reading ± 0.02 V (typical)
( 10 Hz - 45 Hz, 850 Hz ­1 MHz )
Voltage Accuracy, DC ± 0.05% of Reading ± 0.1% of Range ± 0.05 V
Voltage Accuracy, V
rmn
Effect of Common Mode (Typical)
Current Accuracy, A
RMS
± 0.2% of Reading ± 0.1% of Range ± 0.1 V
1000 V, 60 Hz < 10 mV
100 V, 100 kHz < 50 mV
± 0.04% of Reading ± 0.04% of Range ± (20 µV/Z
( 45 Hz - 850 Hz )
Current Accuracy, A
RMS
± 0.05% of Reading ± 0.05% of Range ± (0.02*F)% of Reading ± (20 µV/Z
( 10 Hz - 45 Hz, 850 Hz ­1 MHz )
Current Accuracy, DC ± 0.05% of Reading ± 0.1% of Range ± (50 µV/Z
Current Accuracy, A
RMS
± 0.2% of Reading ± 0.1% of Range ± (100 µV/Z
4 www.tektronix.com
)
ext
) (typical)
ext
)
ext
)
ext
Measurement accuracy - power
Watts Accuracy (V
VA Accuracy (V
VAR Accuracy
RMS
RMS
acc. x A
acc. x A
x PF) ± (A
RMS
) ± (A
RMS
RMS
acc. x V
RMS
acc. x V
RMS
RMS
)
x PF) ± (V
RMS
x A
x cos θ -cos { θ± (Vh1
RMS
PA4000 Power Analyzer
ph.err.
± Ah1
ph.err.
)})
PF Accuracy Cos θ -cos [ θ± ( Vh1
Measurement accuracy - Harmonic magnitude and phase
Voltage Harmonics (45 Hz -
1
± 0.08% of Reading ± 0.08% of Range ± (0.02*F)% of Reading ± 0.02 V
850 Hz) (Typical for 10 Hz to 45 Hz and 850 Hz to 1 MHz)
Voltage Harmonics Phase
0.025 ± [ 0.005 * V
(45 Hz - 850 Hz) (Typical for 10 Hz to 45 Hz and 850 Hz to 1 MHz)
Current Harmonics (45 Hz -
± 0.08% of Reading ± 0.08% of Range ± (0.02*F)% of Reading ± (20 µV /Z
850 Hz) (Typical for 10 Hz to 45 Hz and 850 Hz to 1 MHz)
Current Harmonics Phase
0.025 ± [0.005 * A
(45 Hz - 850 Hz) (Typical for 10 Hz to 45 Hz and 850 Hz to 1 MHz)
Physical characteristics
Dimensions
Weight
Height
range
range
ph.err
/ V
/ A
± Ah1
reading
reading
)] ± 0.001
ph.err
)] ± (0.1 / V
)] ± (0.001 / A
) ± ( 0.002 *F )
range
)
ext
)) ± (0.002 *F)
range * Zext
13.2 cm. 5.2 in.
Width 42 cm. 16.5 in.
Depth 31 cm. 12.5 in.
Net (without lead set)
8.8 kg 19.5 lb.
Temperature
Storage
-20 ○ C to +60 ○ C
Operating 0 ○ C to +40 ○ C
1
F is the frequency measured in kHz. In the case of harmonics, F is the harmonic frequency.
Z
is the shunt impedance (0.6 Ω for 1 A shunt, 9.375 mΩ for 30 A shunt). It is assumed that the waveform measured is a sine wave.
ext
Measurement conditions during calibration: Instrument default settings unless otherwise stated, Sine waves applied to V and I inputs, 30 minute warm-up, Temperature 23 °C ±5 °C.
www.tektronix.com 5
Datasheet
Ordering information
PA4000 models
A PA4000 must be ordered with one of the following options:
Opt. 1CH One input module installed
Opt. 2CH Two input modules installed
Opt. 3CH Three input modules installed
Opt. 4CH Four input modules installed
Standard accessories
Accessories
-- Voltage Lead Set (one per input module)
-- country-specific power cord
063-4498-xx CD containing all relevant documentation and translated user manuals
-- USB Host - to - Device interface cable
-- certificate of calibration documenting traceability to National Metrology Institute(s) and ISO9001 Quality System Registration
-- Three year product warranty.
Instrument options
Opt. GPIB GPIB Interface
Opt. 15V Power Output for External Current Transducers
Language options No Language Options - Standard documentation CD includes user manual in English, French, German, Spanish, Japanese,
Portuguese, Simplified Chinese, Traditional Chinese, Korean, and Russian.
Power cord options
Opt. A0 North America power plug (115 V, 60 Hz)
Opt. A1 Universal Euro power plug (220 V, 50 Hz)
Opt. A2 United Kingdom power plug (240 V, 50 Hz)
Opt. A3 Australia power plug (240 V, 50 Hz)
Opt. A4 North America power plug (240 V, 50 Hz)
Opt. A5 Switzerland power plug (220 V, 50 Hz)
Opt. A6 Japan power plug (100 V, 110/120 V, 60 Hz)
Opt. A10 China power plug (50 Hz)
Opt. A11 India power plug (50 Hz)
Opt. A12 Brazil power plug (60 Hz)
Opt. A99 No power cord
6 www.tektronix.com
PA4000 Power Analyzer
Service options
Opt. C3 Calibration Service 3 Years
Opt. C5 Calibration Service 5 Years
Opt. D1 Calibration Data Report
Opt. D3 Calibration Data Report 3 Years (with Opt. C3)
Opt. D5 Calibration Data Report 5 Years (with Opt. C5)
Opt. G3 Complete Care 3 Years (includes loaner, scheduled calibration, and more)
Opt. G5 Complete Care 5 Years (includes loaner, scheduled calibration, and more)
Opt. R5 Repair Service 5 Years (including warranty)
Opt. R5DW Repair Service Coverage 5 Years (includes product warranty period). 5-year period starts at time of instrument purchase
Recommended accessories
CT-60-S Fixed-core current transducer, high accuracy, up to 60 A
CT-200-S Fixed-core current transducer, high accuracy, up to 200 A
CT-400-S Fixed-core current transducer, high accuracy, up to 400 A
CT-1000-S Fixed-core current transducer, high accuracy, up to 1000 A
CT-100-M Fixed-core current transducer, Hall effect, up to 100 A
CT-200-M Fixed-core current transducer, Hall effect, up to 200 A
CT-500-M Fixed-core current transducer, Hall effect, up to 500 A
CT-1000-M Fixed-core current transducer, Hall effect, up to 1000 A
CL200 Current clamp, 1 A - 200 A, for Tektronix Power Analyzers
CL1200 Current clamp, 0.1 A - 1200 A, for Tektronix Power Analyzers
PA-LEADSET Replacement lead set for Tektronix Power Analyzers (One channel lead set)
Tektronix is registered to ISO 9001 and ISO 14001 by SRI Quality System Registrar.
Product(s) complies with IEEE Standard 488.1-1987, RS-232-C, and with Tektronix Standard Codes and Formats.
Product Area Assessed: The planning, design/development and manufacture of electronic Test and Measurement instruments.
www.tektronix.com 7
Datasheet
ASEAN / Australasia (65) 6356 3900 Austria 00800 2255 4835* Balkans, Israel, South Africa and other ISE Countries +41 52 675 3777 Belgium 00800 2255 4835* Brazil +55 (11) 3759 7627 Canada 1 800 833 9200 Central East Europe and the Baltics +41 52 675 3777 Central Europe & Greece +41 52 675 3777 Denmark +45 80 88 1401 Finland +41 52 675 3777 France 00800 2255 4835* Germany 00800 2255 4835* Hong Kong 400 820 5835 India 000 800 650 1835 Italy 00800 2255 4835* Japan 81 (3) 6714 3010 Luxembourg +41 52 675 3777 Mexico, Central/South America & Caribbean 52 (55) 56 04 50 90 Middle East, Asia, and North Africa +41 52 675 3777 The Netherlands 00800 2255 4835* Norway 800 16098 People's Republic of China 400 820 5835 Poland +41 52 675 3777 Portugal 80 08 12370 Republic of Korea 001 800 8255 2835 Russia & CIS +7 (495) 7484900 South Africa +41 52 675 3777 Spain 00800 2255 4835* Sweden 00800 2255 4835* Switzerland 00800 2255 4835* Taiwan 886 (2) 2722 9622 United Kingdom & Ireland 00800 2255 4835* USA 1 800 833 9200
* European toll-free number. If not accessible, call: +41 52 675 3777 Updated 10 February 2011
For Further Information. Tektronix maintains a comprehensive, constantly expanding collection of application notes, technical briefs and other resources to help engineers working on the cutting edge of technology. Please visit www.tektronix.com.
Copyright © Tektronix, Inc. All rights reserved. Tektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication supersedes that in all previously published material. Specification and price change privileges reserved. TEKTRONIX and TEK are registered trademarks of Tektronix, Inc. All other trade names referenced are the service marks, trademarks, or registered trademarks of their respective companies.
09 Apr 2013 55W-28940-0