Atec P7313SMA, P7380SMA User Manual

Differential SMA Probe Family
P7313SMA • P7380SMA Data Sheet
Applications
Tektronix continues to demonstrate its proven leadership in differential probing with a signicant addition to our Differential SMA Probe Family. Differential SMA probes are designed for measuring differential signals ina50sign differential SMA signal path to a single oscilloscope input channel. Many of today's high-speed serial data standards employ differential signaling on multiple lanes that are challenging to measure simultaneously on a single oscilloscope. The Tektronix Differential SMA Probes provide the
Features & Benets
>13.0 GHz Bandwidth (P7313SMA only, typical)
>8.0 GHz Bandwidth (P7380SMA only, typical)
50 Termination Network, Differential SMA Inputs
Industry-leading Differential Return Loss and VSWR
High-bandwidth Differential Amplier with Excellen
Internal Termination Voltage Generator can be Controlled Internally by the Oscilloscope*
Phase-matched SMA Cables (38 inch length, <1 ps skew) with Cable Loss Compensation
Switchable Gain for Extended Dynamic Range
Auxiliary (Inverted) Output for use with Spectrum Analyzers, Network Analyzers, or as a Clock Recovery Trigger Source
TekConnect™ Interface
1
, by an External Source, or Automatically by the Probe
ability to measure a high-speed differential s ignal on each channel of a multiple­acquire up to four high-speed differential signals with the use of four differential SMA probes. As an added benet, the SMA inputs on the probes connect to high-quality 50 terminations that offer industry-leading return loss, a critical specication that is very important in compliance testing as frequencies increase.
Tektronix Differential SMA Probes also provide a common mode DC voltage input to the termination network. The termination voltage can be supplied either ext there is also an automatic mode that senses the common mode voltage of the input signal and automatically sets the termination voltage to match. The P7313SMA has an extended te rmination voltage range that ma kes it ideal for testing differential standards with high common mode voltages like HDMI and DVI.
Taken together, the high-speed differential amplier, superior 50 terminations, low return loss, exible te rmination voltage, and quality phase-ma system when used with Tektronix oscilloscopes.
*1Not available on all oscilloscopes.
Validation and Compliance Testing of Serial Data Standards including, but not limited to:
PCI Express I Serial ATA FBDIMM DDR XAUI HDMI/DVI (P7313SMA only)
channel oscilloscope. The right oscilloscope can simultaneously
ernally by the user or internally by the oscilloscope. In addition,
tched SMA cables create a world-class differential acquisition
and II
aling environment, providing the ability to convert from a
Data Sheet
Characteristics
Characteristic P7380SMA P7313SMA
Bandwidth (Typical) >8 GHz >13.0 GHz Rise Time
(10-90%) (Guaranteed) Rise Time
(20-80%) (Typical) Attenuatio
n
Differential Input Range 0.625 V
Common Mode Input Range
tion Voltage
Termin a Range
Noise, Referred to Input
Differential Return Loss <27 dB to 5 GHz
CMRR
ax Voltage (Nondestruct)
M
nterface
I
<55 ps
<35 ps <25 ps
3.0 V
2.5X or 12.5
p-p
(12.5X)
p-p
(2.5X)
X, user selec table
0.800 V
±2.5 V
±2.5 V
Hz (2.5X)
<13 nV/
<40 nV/Hz (12.5X)
(VSWR<1.09:1)
<20dBto8GHz
(VSWR<1.22:1)
>50dBto100MHz
>35dBto1GHz >20dBto5GHz >15dBto8GHz
5 V (DC + peak AC)
±
ekConnect
T
<40 ps
(2.5X) 3.6 V
p-p
(12.5X)
+3.6/-2.5 V
+3.6/-2.5 V
<30dBto500MHz
(VSWR<1.065:1)
<20dBto6.5GHz
(VSWR<1.22:1)
<15dBto10GHz
(VSWR<1.43:1)
<12dBto13GHz
(VSWR<1.67:1)
>50dBto1GHz
>35 dB to 2.5 GHz
>25dBto5GHz >20dBto10GHz >15dBto13GHz
®
p-p