Atec Omni-3, Omni-37i User Manual

n AC Source Line Impedance Network
Works in combination with i Series and
iX Series AC Source models
n Compliant Reference Impedance
Meets IEC 1000-3-3 Flicker test specification
requirements
n Single and Three Phase Models
Supports single and three phase IEC
n Bypass Mode
Low impedance path mode to disconnect
reference impedance
n 18.5 or 37.0 Amps per Phase
Capable of supporting all products that need
to be tested for compliance
n Remote Control Interface
Control of Bypass mode through power source controller for automated test operation
OMNI-1-18i / OMNI-3-18i
OMNI-3-37i
For IEC1000-3-3 (Flicker) Testing
Reference Impedance
Network
Flicker Tests
Existing IEC regulations require flicker testing for a large class of electrical and electronic equipment. Flicker occurs when the current drawn by an electrical device causes a voltage drop on the line due to the line impedance. Flicker measurements require a controlled environment. EN 61000-3-3 / IEC 1000-3-3 deals with flicker measurement requirements. This standard describes the required test setup, which includes a precision AC source having a specific impedance. By measuring the voltage drop across the standard impedance caused by the change in current drawn by the EUT (Equipment Under Test), the amount of flicker caused by the load can be determined. The characteristics of the standard impedance must be precisely controlled at the test frequency of 50 Hz.
OMNI option
The OMNI-1-18i, OMNI-3-18i and OMNI-3-37i were specifically designed to meet the IEC 1000-3-3 reference impedance requirements when used in conjunction with an i Series or iX Series AC power
source. The OMNI-1-18i and OMNI­3-18i are capable of handling the maximum 16 A per phase current required under the IEC standard. The OMNI-3-37i supports 32 A per phase.
Bypass Mode
Bypass mode of operation causes the reference impedance in the EUT power paths to be shunted to create virtually zero impedance. This electrically removes the reference impedance between the source and the EUT. In this mode, testing for other applications such as IEC 1000-3-2 (Harmonics) that don’t require the reference impedance is possible without physically having to disconnect it.
Engage Mode
The Engage mode activates the reference impedance path. This mode should be used for IEC 1000­3-3 (Flicker) test applications. The default mode of operation for the OMNI is the Bypass mode so it is important to select the correct mode of operation when performing IEC 1000-3-3 tests.
Manual Control
When used with a power system that lacks direct control capability, the engage or bypass mode can be selected from the front panel. LED’s clearly indicate the status of the OMNI. At power up, the OMNI defaults to the bypass mode.
Remote Control
When used with the i Series or iX Series of programmable AC power sources, the engage or bypass mode of operation are controlled from the AC Source front panel, or via the Power Source’s RS232 or IEEE 488 interface. The OMNI front panel is disabled in this case; however, the LED’s still indicate the status of the OMNI.
Functional Design
The OMNI-1-18i and OMNI-3-18i are housed in a slim line 3.5” high cabinet that matches the style of the i Series and iX Series AC power source. The OMNI-3-37i is contained in a 5.25” high cabinet. Either cabinet can be rack mounted using the included rack mount handles.
Rack Mounting
For rack mount applications ­especially three phase systems - the optional rack mount slides (-RMS) are recommended. All units provide carrying handles on the front of the unit with integrated rack ears.
Ordering Information
Models:
OMNI-1-18i
Reference Impedance network for 3001i/iX phase AC source.
OMNI-3-18i
Reference Impedance network for 15003i/iX source.
OMNI-3-37i
Reference Impedance network for 30003i/iX source.
Note 1: Models 3001iX, 5001iX and
15003iX provide programmable
impedance (dynamic impedance)
which may be used in lieu of the
OMNI lumped impedance. California
Instruments offers the OMNI option for
those users insisting on the use of a
lumped reference impedance.
Note 2: Model 30003iX does not offer
programmable impedance. The
OMNI-3-37i is required to support IEC
Flicker tests.
Options
-RMS Rack mount Slides. The use of rack mount slides or a rack shelf is recommended when installing the OMNI in a 19” rack.
Line Cord Options:
-PC1Continental Europe
-PC3United Kingdom
* One North American right angle Line cord is included. Optional line cords are straight which adds about 1 inch of depth.
Supplied with:
North American Line Power
Cord
Rack mount handles
Instruction Manual
1
or 5001i/iX1 single
1
three phase AC
2
three phase AC
Specifications
Input
Line Voltage 115 Vac±10% / 230 Vac±10% Line Current (typ.) 200 mA @ 115 Vac / 60 Hz
110 mA @ 230 Vac / 50 Hz Line Frequency 47-63 Hz Fuse Rating 0.25 A slow acting @ 115 Vac
0.125 A slow acting @ 230 Vac
Output OMNI-1-18i OMNI-3-18i OMNI-3-37i
Phases 1 3 3 Engage/Flicker Mode:
Max. Current 18.5 A 18.5 A / ø 37.0 A / ø
Useable range 2.0 - 18.5 A 2.0-18.5 A / ø 2.0 - 37.0 A / ø
Bypass Mode:
Max. Current 37.0 A 37.0 A / ø 74.0 A / ø
Impedance @ 50 Hz Engage/Flicker Mode:
Phase 0.24 + j0.15 0.24 + j0.15 0.24 + j0.15
Neutral 0.16 + j0.10 0.16 + j0.10 0.16 + j0.10
Accuracy < 5 %
Mechanical
Dimensions 3.5” x 19” x 22” 5.25” x 19” x 22”
(H x W x D) 89 x 480 x 560 mm 133x480x560
Weight 31 lbs 37 lbs 62 lbs
14 kg 17 kg 28 kg
Connectors
Input AC IEC 320 Load Input and
Output
Controls and Indicators
Power On/Off toggle switch Flicker On/Off push button Power Led Bypass mode Led Flicker mode Led
Compression terminals
Contact California Instruments:
Web page: http://www.calinst.com
< 5 % < 5 %
mm
Toll-Free: 800-4AC-POWER
800 422-7693
FAX: 858 677-0940
Email: sales@calinst.com
9689 Towne Centre Drive, San Diego CA, 92121-1964 (858) 677-9040 FAX : (858) 677-0940
Copyright 1998, California Instruments Corp. Specifications subject to change without notice Printed in the USA. OMNIDS 10/98
Loading...