OCS 500N6
OSCILLATORY WAVE SIMULATOR
FOR TESTS ACCORDING TO ...
> ANSI/IEEE C37.90 > ANSI/IEEE C62.41 > EN 61000-4-10 > EN 61000-4-12 > EN 61000-4-18 > IEC 60255-22-1 > IEC 61000-4-10 > IEC 61000-4-12 > IEC 61000-4-18 > IEC 61850-3
The OCS 500N6 includes test capabilities for ringwave up to 6kV (as per EN/IEC 61000-4-12) and the damped oscillatory waves at 100kHz and 1MHz up to 3kV (as per EN/IEC 61000-4-18). The Ringwave is a non-repetitive damped oscillatory transient occurring in low-voltage power, control and signal lines supplied by public and non-public networks. Damped Oscillatory Waves are repetitive transients mainly occurring in power, control and signal cables installed in high voltage and medium voltage stations. The OCS 500N6 can also be used to perform magnetic field tests as required in EN/IEC 61000-4-10 using a magnetic field coil such as the MS 100N.
HIGHLIGHTS
> Standalone test generator
APPLICATION AREAS
INDUSTRY
RESIDENTIAL
> Includes Ringwave up to 6kV > Includes Damped Oscillator Waves up to 3kV
COMPONENTS
RENEWABLE ENERGY
> Built-in CDN, single phase or three-phase up to
32A
> Front panel operation
MEDICAL
BROADCAST
www.emtest.com © EM TEST > PAGE 1/4
DATA SHEET > OCS 500N6 > 20111021
TECHNICAL DETAILS
OCS 500N6 MODELS
OCS 500N6.5 With built-in CDN 250V / 16A OCS 500N6.6 With built-in CDN 250V / 32A OCS 500N6.7 With built-in CDN 3x440V / 16A OCS 500N6.8 With built-in CDN 3x440V / 32A
RINGWAVE AS PER EN/IEC 61000-4-12 AND ANSI/IEEE C62.41
Voltage (o.c.) 250V - 6,000V ± 10% Rise time 0.5us ± 30% Oscillation
100kHz ± 10%
frequency Decaying Ratio of peak 2 to peak 1: 0.4 - 1.1
Ratio of peak 3 to peak 2: 0.4 - 0.8 Ratio of peak 4 to peak 3: 0.4 - 0.8
Source impedance 12ohm and 30ohm ± 20%
200ohm for I/O lines (CNV 504M)
Peak current (s.c.) Max. 500A @ 12ohm or
Max. 200A @ 30ohm internally; Max. 30A @ 200ohm, with CNV 504M
TRIGGER CIRCUIT
Release of pulses Automatic, manual, external Synchronization 0° - 360°, resolution 1°
OUTPUT
Direct Via HV-safety lab connectors Coupling mode Line to line
Line(s) to ground (PE)
DUT supply 1ph-16A
DUT supply 1ph-32A
DUT supply 3ph-16A
DUT supply 3ph-32A
AC: 250V/16A; 50/60Hz DC: 250V/10A
AC: 250V/32A; 50/60Hz DC: 250V/20A
AC: 3x440V/16A; 50/60Hz DC: 250V/10A
AC: 3x440V/32A; 50/60Hz DC: 250V/20A
CRO trigger 5V trigger signal for oscilloscope
MEASUREMENTS
Ring wave Peak voltage and peak current in LCD
Rise time < 1us Oscillation
100kHz ± 10%
frequency Polarity Positive, negative Repetition rate 1 to 60 transients per minute
SLOW DAMPED OSCILLATORY WAVES AS PER EN/IEC 61000-4-18
Voltage (o.c.) at
250V - 3,000V ± 10%
HV output Voltage (o.c) at
250V - 2,500V ± 10%
line output Rise time 75ns ± 20% Oscillation
100kHz and 1MHz ± 10%
frequency Decaying Peak 5 to be > 50% of peak 1 value
Peak 10 to be < 50% of peak 1 value Source impedance 200ohm ± 20% Polarity Positive, negative Repetition rate Max. 50/s for 100kHz and
Max. 500/s for 1MHz Burst duration At least 2s
TEST ROUTINES
Quick Start Immediate start; easy-to-use and
fast
Standard Test routines
As per IEC 61000-4-12, Level 1 - 4 Manual Standard Test routine As per ANSI/IEEE C62.41 As per IEC 61000-4-10, Level 1 - 5
User Test routines Change polarity after n pulses
Change coupling after n pulses Change voltage after n pulses Change phase angle after n pulses
INTERFACE
Serial interface USB Parallel interface IEEE 488, addresses 1 - 30
GENERAL DATA
Dimensions, weight
19"/6HU, approx. 28kg (1-phase)
19"/6HU, approx. 33kg (3-phase) Supply voltage 115/230V +10/-15% Fuses 2 x T 2AT (230V); 2 x T4AT (115V)
www.emtest.com © EM TEST > PAGE 2/4