Atec Netwave-3-Phase User Manual

DATA SHEET > NetWave Series 3-phase > 20130606
NETWAVE SERIES (3-PHASE)
THREE-PHASE MULTIFUNCTIONAL AC/DC POWER SOURCES
FOR TESTS ACCORDING TO ...
> AIRBUS > BOEING > DO 160 Section 16 > EN 61000-3-11 > EN 61000-3-12 > EN 61000-3-2 > EN 61000-3-3 > EN 61000-4-13 > EN 61000-4-14 > EN 61000-4-17 > EN 61000-4-27 > EN 61000-4-28 > EN 61000-4-29 > IEC 61000-3-11 > IEC 61000-3-12 Ed.2:2011 > IEC 61000-3-2 > IEC 61000-3-3 > IEC 61000-4-13 > IEC 61000-4-14 > IEC 61000-4-17 > IEC 61000-4-27 > ...
NETWAVE - SIMULATION OF THE MOST REQUIRED POWER SUPPLY PHENOMENON
The NetWave Series (3-phase) are three-phase AC/DC power source, specifically designed to meet the requirements as per the standards IEC/EN 61000-4-13, -4-14, -4-27(*) and -4-28. It is also used as a DC power source to cover the requirements as per the standards IEC/EN 61000-4-17 (Ripple on DC) and IEC/EN 61000-4-29 for voltage dips and interruptions on DC supplies. The NetWave series is well suited for testing inverters (e.g. solar power, wind power) and e-vehicles. Additionally, the NetWave series (3-phase) offers the necessary capabilities for avionics testing as per DO-160, Airbus ABD0100 and Boeing as well as per MIL-STD-704. Optionally the NetWave 3-phase series can be equipped with a power-recovery module to absorb fed-back power (AC/DC) up to nominal power of the NetWave. (*) pre-compliant
HIGHLIGHTS
> Wide Power Bandwidth; DC - 5kHz
APPLICATION AREAS
INDUSTRY
AVIONICS
> Output Power up to 60,000VA AC / 72,000W DC > Output Voltage up to 3*360V AC (p-n), +/-500V DC
MEDICAL
MILITARY
> High Inrush Current Capability
RESIDENTIAL
RENEWABLE ENERGY
> Power-recovery up to nominal power (optional)
www.emtest.com © EM TEST > PAGE 1/6
DATA SHEET > NetWave Series 3-phase > 20130606
TECHNICAL DETAILS
MODEL OVERVIEW
3-PHASE NETWAVE-MODELS
NetWave 20.x 3-phase Multifunction AC/DC source,
22,500VA AC / 27,000W DC
NetWave 30.x 3-phase Multifunction AC/DC source,
30,000VA AC / 36,000W DC
NetWave 60.x 3-phase Multifunction AC/DC source,
60,000VA AC / 72,000W DC
TECHNICAL DETAILS
NETWAVE 20
Output voltage 0V - 3*300V AC (p-n)
0V - +/- 425V DC
Output current 26A (RMS) continuous
47A (RMS) short-term (max. 3s) 200A repetitive peak
TECHNICAL DETAILS
NETWAVE 60
Output voltage 0V - 3*300V AC (p-n)
0V - +/- 425V DC
Output current 66A (RMS) continuous
100A (RMS) short-term (max. 3s) 400A repetitive peak
NETWAVE 60.2
Output voltage 0V - 3*360V AC (p-n)
0V - 3*620V AC (p-p) 0V - +/- 500V DC
Output current (@ max. 300V AC)
EXTENDED CAPABILITIES FOR NETWAVE
Simple mode Optimized control for integration of
66A (RMS) continuous 100A (RMS) short-term (max. 3s) 400A repetitive peak
the Netwave into existing automation environments (for example Matlab)
NETWAVE 20.2
Output voltage 0V - 3*360V AC (p-n)
0V - 3*620V AC (p-p) 0V - +/- 500V DC
Output current (@ max. 300V AC)
NETWAVE 30
Output voltage 0V - 3*300V AC (p-n)
Output current 33A (RMS) continuous
NETWAVE 30.2
Output voltage 0V - 3*360V AC (p-n)
Output current (@ max. 300V AC)
26A (RMS) continuous 47A (RMS) short-term (max. 3s) 200A repetitive peak
0V - +/- 425V DC
66A (RMS) short-term (max. 3s) 250A repetitive peak
0V - 3*620V AC (p-p) 0V - +/- 500V DC
33A (RMS) continuous 66A (RMS) short-term (max. 3s) 250A repetitive peak
SourceAC mode PLL synchronization with other
voltage sources Trigger channel Extended trigger functions Segment "Step" Ramping of voltage and/or frequency
in constant time windows Extern mode Control of the NetWave by an
external control signal
www.emtest.com © EM TEST > PAGE 2/6