Atec MP User Manual

MEASURINGPAD
SPECIFICATIONS
Scanning, Monitoring, and Recording
Scan Rate: 10 kHz per analog input; 10 kHz per digital input 16-bit A/D conversion
Overall Accuracy (All Measurements): ±0.1% of full scale or better Overrange (All Measurements): 10% of full scale Logging Modes:
Manual: A single “snapshot” record taken via touchscreen button Timed: Records taken automatically between specified start and
stop times; user-specified rate from one second to 24 hours Event-Triggered: Records taken while any triggering limit violation
or digital input occurs; user-specified rate from one millisecond to 24 hours, or “burst” speed (as fast as possible); overrides timed logging, and may include “pretrigger” data, if desired
User-enabled beeper for setpoint violation alert
Test /Process Interface
Up to four external sensor interface pods per MeasuringPAD unit (ordered separately)
Four DC-coupled differential analog inputs of the same type per pod (see below)
Two digital inputs per pod (see below) EEPROM per pod, for ID and storage of calibration values
Analog Inputs
Low Voltage Pod
Operating Range (Differential and Common-Mode): ±0 to 30 VDC or 0 to 24 VAC RMS
Input Impedance: 4 M(differential); 1 M(common-mode) Accuracy: ±0.1% of full scale up to 400 Hz; ±1.0% of full
scale up to 5 kHz Connection: Five-way binding posts
High Voltage Pod
Operating Range (Differential and Common-Mode): ±0 to 600 VDC or 0 to 480 VAC RMS
Input Impedance: 40 M(differential); 10 M(common-mode) Accuracy: ±0.1% of full scale up to 400 Hz; ±1.0% of full
scale up to 5 kHz Connection: Safety banana jacks
Low Current
Operating Range (Differential): ±0 to 20 mADC or 0 to 20 mAAC RMS
Input Impedance: 250 (differential); 1 M(common-mode) Accuracy: ±0.1% of full scale up to 400 Hz; ±1.0% of full
scale up to 5 kHz Connection: Five-way binding posts
Thermocouple
Accepts Types B, E, J, K, N, R, S, and T
Test /Process Interface (continued)
Resolution: 0.1 degree Input linearization via NIST table Accuracy: ±0.1% of full scale DC ± 0.5° C cold-junction
compensation Connection: Mini jacks
Digital Inputs (All Pods, unless otherwise indicated)
Maximum Frequency: 20 kHz Minimum Time Interval (for timed inputs): 1 msec V
H: 4.5 to 24 VDC
V
L: 0 to 0.6 VDC
R
IN: 6.0 k, approximate
Isolation: Optically isolated to 750 VAC RMS, common-mode Frequency and quadrature inputs configurable for limit logic Connection: Five-way binding posts
DC Strain Gauge
1, 2, 4 & 10 mV/V Accuracy: ±0.1% of full scale up to 400 Hz; ±1.0% of full
scale up to 5 kHz Connection: Screw Terminals
General Specifications
Dimensions (HWD): 12" x 2.5" x 8" (30 cm x 6.4 cm x 20.3 cm) Weight: 4.2 lb. (1.9 kg) Operating Temperature: -10° to 50° C (14° to 122° F) Storage Temperature: -40° to 85° C (-40° to 185° F) Humidity: 95%, noncondensing System Time Clock: Crystal-controlled; 1-second resolution Display: Contrast-adjustable 320 x 240 LCD color touchscreen with
audible feedback and programmable backlight timeout; available languages include English, French, Italian, German, Spanish
Memory Card: Compact flash 256M supplied AC Power Adapter supplied (90-264 VAC, 47-63 Hz) Supporting easel and carry strap supplied
Optional Accessories
Low Voltage and High Voltage Cable Sets Soft-Sided Carrying Case Field-Replaceable Battery Pack External Battery Charger Weatherproof Container Reusable Shipping Container Lockable Portable Case USB Flash Card Reader Current Probes: TR25XX (1-3000A) Clamp-on
RR30XX (30/300 / 3000A) Flexible
M
EASURING
PAD
A versatile, short-term data acquisition solution
for maintenance, troubleshooting and
energy measurement applications
dranetz-bmi.com
1.800.372.6832 sales@dranetz-bmi.com
tel 732.287.3680 • fax 732.248.1834 • 1000 New Durham Road • Edison, New Jersey 08818 USA
©2008 Dranetz-BMI. All rights reserved. Printed in the United States. Specifications are subject to change without notice.
Revised 07/07/08
For temporary but critical on-site data acquisition, the portable, high-speed MeasuringPAD™ offers a convenient, economical alternative
to an elaborate test and measurement installation. This sturdy hand-held instrument can be used in a wide range of manufacturing,
laboratory, plant operations, and quality control applications. Rapidly collecting, monitoring, and displaying multiple sensor inputs, it
can be used for troubleshooting, maintenance, process tuning, fault recording, trend analysis, and much more. Configured and operated
through an easy-to-use color touchscreen interface, MeasuringPAD continuously scans all active inputs, digitizing with 16-bit A/ D while
checking for limit-violation triggers and refreshing its memory buffer of 10 kHz per input.
Real-Time Data Monitoring
MeasuringPAD provides live readout of up to sixteen differential analog inputs and up to eight logic-level digital inputs. As numerical measurements are monitored for exact conformance to user-entered setpoints, the vivid color-coded status display
lets you size up your test or process at a glance. To ensure true “finished-answer” readability, analog signals are easily scalable to desired engineering units.
Convenient input “pods” allow fast and easy hookup of machine- or process-dedicated sensors – including DC/AC Low and High Voltage sources (from 100 mV to 300 V), DC/AC Low Current sources (4 to 20 mA), and most Thermocouple types. Digital inputs are standard TTL-level signals from switches, contacts, encoders, pickups, and other frequency-generating devices. They may be used to trigger the logging of data, to perform programmable counting and timing operations, and to monitor frequency and
quadrature. Up to eight internal cross-channel math calculations can also be set up to display and monitor additional process variables such as horsepower, efficiency, and corrected flow – plus signal max /min and average values.
High-Speed Data Recording
You can instruct MeasuringPAD to sample scaled measurements at a regular time interval – or, immediately on detection of any specified trigger event, to begin recording at a special rate that can be up to a microsecond-level “burst” speed. Instantly stored on a standard CF memory card in a simple and universally compatible ASCII text format, data records are both time- and trigger-stamped, and may be directly imported by all kinds of third-party programs for post-analysis and report generation, without the need for special drivers or version-dependent operating
systems. If desired, records of data
values as they existed just prior to a
triggering event can be included in
the stored data file. The touch of a
button will display a scrollable
spreadsheet report of all currently
logged data, including time and
event indication.
MEASURING
PAD
Using MeasurePAD’s unique
touchscreen pages, setting up the
instrument literally takes seconds. Simple user selections and clear,
intuitive pathing make the setup process both easy to use and easy to
learn. Once created, multiple test-specific setups can be stored for
instant reloading.
Multichannel Data Display
For instant operator interpretation of measurement results, the backlit color LCD furnishes real-time display of measured and calculated values in a number of programmable formats, including
bargraph, for rapid visual comparison of up to sixteen active parameters
trend, for simultaneous, independent timeplotting of up to four active parameters. Ideal for comparing energy and process / production parameters for energy reduction studies.
scope, for live display of up to four active waveform inputs in a conventional oscilloscope mode
discrete Fourier transform (DFT), for harmonic analysis
of a selected analog input.
Loading...