NSG 4070 ApplicAtioN for Mil-StD 461f cS 114
RF out
Amp in
< +10 dBm
Amp out
ch.1 < +27 dBm
ch.2 < +20 dBm
ch.3 < +20 dBm
Power meter
NSG 4070
Power
USB
Tuning
StSizeStSize StSize
Local
Back
Stop
Run
Hold
0
.
1
4
7 8
5
2
3
kHz
dBm
6
9
MHz
dBµV
Hz
V
Enter
Step
1
STO
FRQ
LVL
RCL
Step2Step
3
MOD
2nd
Help
RF
ON/OFF
NSG 4070
Power meter
ch. 1 stress level
ch. 2 forward power
ch. 3 reverse power
Amp in
Amp out
50 Ω Termination
NSG 4070
RF output
NSG 4070
User port
Amp in
Amp out
Power amplifier
inputs
Calibration jig MD 4070 Ground plane
Interlock
Switch (part of SW 4070)
Switch (part of SW 4070)
2
A
A
1
B
B
DIRECTIONAL COUPLER
Test parameter
Standards: MIL-STD 461F CS 114
Frequency range: 10 kHz to 200 MHz
Curve 1 to 5: see diagrams
Modulation: 1 kHz pulse modulation, 50%
duty cycle
Test method: Substitution method with
monitoring probe
Monitoring probe: only for information
WARNING: The power meter inputs are very sensitive. It is the user’s responsibility to ensure
that the selected test levels does not damage the equipment. Any hardware/setup changes
should be calculated before starting the test.
Equipment
Signal generation: NSG 4070 -0
Modulator: included in NSG 4070-0
Power meter: 3x included in NSG 4070-0
Power amplier: max. 30 W required
RF Switch: SW 4070
Directional coupler: DCP 0100
Current injection probe: CIP 9136A
Monitoring probe: MD 4070
Calibration jig: PCJ 9201B
Termination: 50 Ω 10 W
Attenuation: 1x 20 dB 30 W, 2x 10 dB 30 W, 1x
6 dB 30 W
Software: incl. in NSG 4070 or optional C3I
or WIN 6000
Calibration set-up for monitoring probe
Teseq GmbH
Lands berger Str. 255 12623 Ber lin Germany
T + 49 30 56 59 88 35 F + 49 30 56 59 88 34
desales@teseq.com www.teseq.com
Remarks:
The monitoring probe MD 4070
needs to be calibrated in the way
of its use (active, passive or with
switching at a specic frequency
from active to passive).
1
RF out
Amp in
< +10 dBm
Amp out
ch.1 < +27 dBm
ch.2 < +20 dBm
ch.3 < +20 dBm
Power meter
NSG 4070
Power
USB
Tuning
StSizeStSize StSize
Local
Back
Stop
Run
Hold
0
.
1
4
7 8
5
2
3
kHz
dBm
6
9
MHz
dBµV
Hz
V
Enter
Step
1
STO
FRQ
LVL
RCL
Step2Step
3
MOD
2nd
Help
RF
ON/OFF
NSG 4070
Power meter
ch. 1 stress level
ch. 2 forward power
ch. 3 reverse power
Amp in
Amp out
50 Ω Termination
NSG 4070
RF output
NSG 4070
User port
Amp in
Amp out
Power amplifier
inputs
Calibration jig BCI probe Ground plane
Interlock
Switch (part of SW 4070)
Switch (part of SW 4070)
2
A
A
1
B
B
DIRECTIONAL COUPLER
Attenuator #1 Attenuator #2
RF out
Amp in
< +10 dBm
Amp out
ch.1 < +27 dBm
ch.2 < +20 dBm
ch.3 < +20 dBm
Power meter
NSG 4070
Power
USB
Tuning
StSizeStSize StSize
Local
Back
Stop
Run
Hold
0
.
1
4
7 8
5
2
3
kHz
dBm
6
9
MHz
dBµV
Hz
V
Enter
Step
1
STO
FRQ
LVL
RCL
Step2Step
3
MOD
2nd
Help
RF
ON/OFF
NSG 4070
Power meter
ch. 1 stress level
ch. 2 forward power
ch. 3 reverse power
Amp in
Amp out
NSG 4070
RF output
NSG 4070
User port
Amp in
Amp out
Power amplifier
inputs
BCI probe Ground plane
Interlock
Switch (part of SW 4070)
Switch (part of SW 4070)
2
A
A
1
B
B
DIRECTIONAL COUPLER
Attenuator #1 Attenuator #3
Auxiliary
equipment
AE
EUT
Equipment
under test
Insulating
MD 4070
Test level for MIL STD 461F CS114 curve #1
30
40
50
60
70
80
90
100
110
0,01 1 30 200
Calibration set-up (for curve #1)
Attenuator #1: 10 dB, 30 W
Attenuator #2: not in use
Attenuator #3: not in use
Termination: 50 Ω 10 W
Remarks:
Calibration with a 20 dB increased
level.
Test set-up with monitoring probe (for curve #1)
Teseq GmbH
Lands berger Str. 255 12623 Ber lin Germany
T + 49 30 56 59 88 35 F + 49 30 56 59 88 34
desales@teseq.com www.teseq.com
2
Attenuator #1: 10 dB, 30 W
Attenuator #2: not in use
Attenuator #3: 20 dB, 30 W
Termination: 50 Ω 10 W
Remarks:
Test with additional 20 dB attenuator (attenuator #3) between
directional coupler and BCI probe.
Use of MD 4070 on PM ch.1:
The monitoring probe is in the
linear measuring range above 54
dBμA stress level (above 70 kHz).
The MD 4070 needs to be used in
the active mode.