Atec EHP-200 User Manual

Page 1
Electric and Magnetic Field Measurement
New solution for Isotropic Measurements in the 9 kHz – 30 MHz Range
Electric Fields from 0.02 to 1000 V/m Magnetic Fields from 0.6 mA/m to 300 A/m Built-in Frequency Spectrum Analysis Built-in Rechargeable Battery Optical Fiber Connection to PC
The E-H fields analyzer model EHP-200 has been designed for accurate isotropic mea­surements of both electric and magnetic fields in the 9 kHz - 30 MHz frequency range, with no or minimum perturbation of the fields that are being measured.
The field sensors and the electronic measuring circuitry are contained in a rugged housing, only 3.6 x 3.6 x 4.3 inches in size. Separate 3-axis and total values (peak and average) are measured with exceptional flatness and linearity of ±0.3 dB. Results are expressed in V/m, A/m, μT, mW/cm2, W/m2.
The EHP-200 features built-in spectrum analysis with minimum selectable bandwidth of 1 kHz for detailed measurements of the E and H field intensity vs. frequency, and a dynamic range of 80 dB. The built-in rechargeable Li-Ion battery provides up to 8 hours of continuous operation.
The EHP-200 is controlled by a PC through the optical fiber link, and measurements are displayed in real time. An auxiliary input is available for measuring the frequency spectrum of external applied signals.
36
narda Safety Test Solutions an communications company USA • Germany • Italy
USA TEL: (1) 631 231-1700 • FAX: (1) 631 231-1711 • E-MAIL: NardaSTS@L-3COM.com • www.narda-sts.us
Page 2
Electric and Magnetic Field Measurement
Applications
BROADCASTING SURVEILLANCE
The EHP-200 is particularly useful in measuring the actual fields generated by long, medium and short wave broadcast transmitters, to ensure safety around the sites of large anten­nas, to control the transmitted power in the actual radiation direction, to test the functionality of the transmitting anten­nas and to identify the borders between near and far field regions.
WAVE IMPEDANCE
As a unique feature, the PC program calculates the field wave impedance by dividing the total value of the E-field by that of the H-field. This method is particularly suitable for evaluating the non-linear, scattered near-field region of large broadcast antenna systems.
FIELDS GENERATED BY METAL DETECTORS AND RFID’S
Fields generated by a number of devices using RF to detect the presence of metals, to identify objects, anti-theft systems etc. can now be accurately and easily measured.
EHP-200 Field Analyzer
EHP-200 Control Software
FOR WINDOWS™ OPERATING SYSTEMS
All measuring functions are user-programmable: Resolution Bandwidth Filter, center frequency and frequency span, pre­amplifier, measuring units, etc.
The Marker function is used to measure the frequency and amplitude. It features Highest, Next and Previous Peak func­tions, while the Marker Center function sets the display center frequency at the current marker frequency value.
The Marker also features the Delta Peak function for relative measurements. The Wide Band field value is calculated with reference to the measured frequency span.
To immediately evaluate the measured levels, Limit Lines can be created and displayed on the graphical window. The measured data can be saved as either text or bitmap, and the limits can also be saved and recalled.
Blue Line: total Field / Green Line: X-axis Cyan Line: Y-axis / Magenta Line: Z-axis
narda Safety Test Solutions an communications company USA • Germany • Italy
USA TEL: (1) 631 231-1700 • FAX: (1) 631 231-1711 • E-MAIL: NardaSTS@L-3COM.com • www.narda-sts.us
37
Page 3
Electric and Magnetic Field Measurement
EHP-200 Field Analyzer
Specifi cationsEHP-200 E&H FIELD ANALYSER
EHP200 E AND H FIELD ANALYZER
RF SPECIFICATIONS Electric Field Magnetic Field Mode A Magnetic Field Mode B Auxilllary Input
Frequency Range 9 kHz to 30 MHz 9 kHz to 30 MHz 300 kHz to 30 MHz 9 kHz to 30 MHz
Measurement Range @ 10 KHz RBW
@ Preamp ON
Dynamic Range >80 dB
Sensitivity @ 10 kHz RBW
@ Preamp ON
Resolution 0.01 V/m 1 mA/m 0.1 mA/m 0.01 dB
Flatness
Anisotropy ±0.8 dB at 1 MHz
Linearity 0.5 dB @ 1 MHz from Full Scale to -60 dB Full Scale
Typical Accuracy at 1 MHz ±0.8 dB @ 20 V/m ±0.8 dB @ 53 mA/m ±0.8 dB @ 53 mA/m ±0.3 dB@-10 dBm
Maximum Frequency Span 6 kHz to 30 MHz
Resolution Bandwidths Available 1 kHz, 3 kHz, 10 kHz, 30 kHz, 100 kHz, 300 kHz
Rejection to E-field > 20 dB
Rejection to H-field > 20 dB
Calibration Errors Stored in internal EEPROM
Temperature Error 0.02 dB/
GENERAL SPECIFICATIONS
Preamplifier Selectable ON/OFF, 14 dB gain
Reading Units V/m, A/m, mT, mW/cm
Optical Link Maximum length of 80 m
Internal Battery 3.7 V, 3.6 Ah, Li-ion, rechargeable
Battery Operation Time > 8 hours (recharging time approximately 8 hours)
External Supply 10 – 15 VDC, 500 mA
Firmware Update Via Optical Fiber
Operating Temperature -10
Storage Temperature -20oC to +70oC
Dimensions and Weight 3.6 x 3.6 x 4.3 inches (92 x 92 x 109 mm), 1.22 lb (550 g)
ORDERING INFORMATION
Part Number EHP-200
Accessories Supplied
Optional Accessories EHP-200 Palm Kit
0.1 to 1000 V/m
0.02 to 200 V/m
0.1 V/m
0.02 V/m
±0.5 dB (20 V/m,
from 0.1 to 27 MHz)
10 meter Fiber Optic Cable ( FO-8053/10), Fiber Optic Converter (FO to RS-232, 8053-OC), Soft Carrying
Bag ( 8053-SC), Battery Charger ( 8053-BC), 50 cm Plastic Pole and Tripod, PC Software, Operating Manual,
0.03 to 300 A/m
6.0 mA/m to 60 A//m
30 mA/m
1 mA/m
±0.8 dB (166 A/m,
from 0.15 to 30 MHz)
Calibration Certificate
3.0 mA/m to 30 A/m
0.6 mA/m to 6 A/m
±0.8 dB (53 mA/m,
o
C
o
C to +50oC
3 mA/m
0.1 mA/m
0.3 to 27 MHz)
2
2
, W/m
-80 to 0 dBm
-94 to -14 dBm
-80 dBm
0.01 dB
±0.4 dB (-20 dBm)
www.narda-sts.com
38
narda Safety Test Solutions an communications company USA • Germany • Italy
USA TEL: (1) 631 231-1700 • FAX: (1) 631 231-1711 • E-MAIL: NardaSTS@L-3COM.com • www.narda-sts.us