Atec CWS500N1 User Manual

DATA SHEET > CWS 500N1 > 20120604
CWS 500N1
CONTINUOUS WAVE SIMULATOR, 80W
FOR TESTS ACCORDING TO ...
> EN 300329 > EN 300340 > EN 300342-1 > EN 300386 V1.3.2 > EN 301489-1 > EN 301489-17 > EN 301489-24 > EN 301489-7 > EN 55024 > EN 61000-4-6 > EN 61000-6-1 > EN 61000-6-2 > IEC 60601-1-2 > IEC 61000-4-6 > IEC 61326 > IEC 61850-3
CWS 500N1 - THE SINGLE BOX SOLUTION FOR RF CONDUCTED IMMUNITY TESTING
The CWS 500N1 is the most compact single box test equipment for testing conducted rf immunity as per IEC 61000-4-6 and related standards with a frequency range of 100kHz to 300MHz. Apart from the 1kHz 80% AM signal the CWS 500N1 also generates a 2 Hz 80% AM signal to test medical appliances and a 1 Hz PM signal with 50% duty cycle required to test safety equipment like fire alarms. Equipped with a 1 GHz current monitor the CWS 500N1 can be used up to 1 GHz by means of an external amplifier.
EM TEST supplies a large range of CDNs, EM clamp and current injection clamps as well as the corresponding calibration accessories. Full compliant levelling can be run from the front panel of the CWS 500N1 storing the results in 5 memory spaces available in the CWS 500N1.
HIGHLIGHTS
> Signalgenerator 9kHz to 1GHz
APPLICATION AREAS
INDUSTRY
TELECOM
> Amplitude modulation: 1Hz to 3kHz (0% to 95%) > Pulse modulation: 1Hz - 1kHz (10% to 90%)
MEDICAL
RESIDENTIAL
> Built-in amplifier, 100kHz to 300MHz
BROADCAST
RENEWABLE ENERGY
> Built-in current monitor, 9kHz to 1GHz
www.emtest.com © EM TEST > PAGE 1/4
DATA SHEET > CWS 500N1 > 20120604
TECHNICAL DETAILS
TEST LEVEL
Output level 1V - max. 30V (emf)
all standard test levels are guaranteed with all coupling
methods Output power 80W (nominal) Output impedance 50ohm Harmonic
< -20dBc at max.power distortion
TEST FREQUENCIES
Sinus (CW) 100kHz to 300MHz Frequency bands 100kHz to 9.999MHz
10MHz to 99.99MHz
100MHz to 300MHz
in the Quick Start menu the step size
can be selected by the operator Unmodulated
CW (continuous wave) signal
Amplitude modulation
1kHz, 80%AM as per IEC 61000-4-6
2Hz, 80%AM as per IEC 60601-1-2
400Hz, 80%AM Pulse modulation 1Hz, 50% duty cycle as per
EN 50130-4
TEST ROUTINES
Quick Start Immediate test start; easy-to-use
and fast
User Test routines Voltage sweep
Frequency sweep Dwell time sweep
Standard Test routines
Level 1 to Level 3 (IEC 61000-4-6) Automatic Level X to Level Y
Cal procedure Calibration of the complete test
set-up, calibration data saved in internal memory
Service Service, set-up
SIGNAL GENERATOR
Output level -55dBm to 0dBm Frequency range 9kHz to 1GHz Output impedance 50ohm Direct RF output To control an external amplifier
MEASUREMENTS FOR BCI APPLICATION
PM 1000 3-channel power meter up to 1GHz
- to measure the Forward power
- to measure the Reverse power
- to measure the injected current
MEASUREMENTS
Monitor Integrated RF power meter,
measuring input for CDN and clamp
calibration as well as current
monitor for clamp applications RF indicator LED indicating the RF output status LCD Online display of the test level and
the preselected frequency value Cal data F1 - F5 5 internal memories to save
calibration data,
1% step width as required by the
standard
TIME PARAMENTERS
Dwell time for CW
td = 0.3s to 9,999s & AM
Dwell time for PM td = 3s to 9,999s Pause time tr = 0/0.3s to 9,999s
OUTPUT
Direct RF output BNC (on the front panel)
INTERFACE
Serial interface USB Parallel interface IEEE 488, addresses 1 to 30 Fail 1 BNC input; test will be stopped when
active low
Fail 2 BNC input; test status will be saved
(max. 10 events) when active low. Test will continue.
GENERAL DATA
Dimensions,
19"/3HU, approx. 17kg
weight Supply voltage 115V to 230V +10/-15%, 50/60Hz Input power Max. 380W Power factor cos (phi) = 0.98 at max. output
power as per IEC 555
Fuses 2 x 6.3AT (115V) or 2 x 3.15AT
(230V) Cooling Active cooling, air ventilation Temperature 10°C to 40°C Rel. humidity Max. 85%, non condensing
www.emtest.com © EM TEST > PAGE 2/4