11970 Series
Harmonic Mixers
For use with the Agilent E4407B, 8560E/EC Series, 8566B,
71000 Series, and PSA Series spectrum analyzers,
plus the N9030A PXA signal analyzer
18 to 110 GHz
11970K*, 11970A, 11970Q, 11970U, 11970V, 11970W
Exceptional performance
employing a dual-diode design to achieve very flat frequency response and low
conversion loss. Each mixer is calibrated across its full band:
11970K* 18 to 26.5 GHz
11970A 26.5 to 40 GHz
This series of mixers has been designed for a local oscillator frequency of 3 to 6.1 GHz.
Accurate absolute amplitude measurements can be made by using the mixer’s conversion loss
calibration chart. The SWR of the waveguide input is typically 2.2:1 to further minimize
measurement uncertainty. The combination of high gain-compression level and low
conversion loss provides the maximum dynamic range for measuring input signals.
* 11970K is excluded from PXA support
11970Q 33 to 50 GHz
11970U 40 to 60 GHz
11970V 50 to 75 GHz
11970W 75 to 110 GHz
Easy to use
The excellent frequency response and low conversion loss are achieved without external dc bias or tuning stubs. Since bias and tuning stubs are not required, manual operation is simplified, and the complexity of hardware and software for automatic systems is greatly reduced. The repeatability of amplitude measurements is also enhanced. The dual-diode design of the mixers further simplifies measurements by suppressing the odd-order harmonics by more than 20 dB, which makes identification of the mixing products easier.
Figure 1. The 11970 Series mixers have separate LO input and IF output ports
Frequency extension for the Agilent E4407B, 8560E/EC Series, 8566B, and 71000 Series and PSA Series spectrum analyzers
The 11970 Series harmonic mixers are fully compatible with the Agilent E4407B, 8560E/EC Series, 8566B (requires a preamplifier), 71000, and PSA Series spectrum analyzers. Accurate frequency and amplitude measurements are made directly from the spectrum analyzer’s display after calibration using the mixer’s calibration chart.
Frequency extension for the Agilent N9030A PXA Series signal analyzer
The 11970 Series harmonic mixers (11970K excluded) are also compatible with the N9030A PXA Series high­performance signal analyzer. The PXA offers optional external mixing (Option EXM), enabling it to work with the 11970 Series and other external mixers for frequency extension. An external diplexer is required because the PXA external mixing has a single SMA LO/IF port, whereas the 11970 Series mixers have separate LO and IF ports. Recommended diplexers can be purchased from Agilent as N9020A-E13, or from OML Inc. as DLP.313B.
Rugged
The rugged Agilent 11970 Series mixers will survive input levels up to 100 milliwatts (+20 dBm) with no damage to the mixer diodes. They will withstand shocks up to 30 G’s and the vibration required by MIL-STD 28800C, Type III, Class 3 tests.
Agilent 8560E/EC Series spectrum analyzer
Extended frequency 8560E/EC Series
Agilent E4407B spectrum analyzer
Extended frequency E4407B ESA-E Series
IF input
1st LO output
Spectrum analyzer
Agilent 11970 Series harmonic mixer
Agilent PSA Series spectrum analyzer
Extended frequency PSA Series
IF Input/1st LO output
Agilent N9030A PXA Series signal analyzer
Extended frequency PXA Series
RF input
SMA cable
SMA cable
SMA cable
SMA cable
Agilent 11970 Series
IF
Mixer
IF
diplexer
LO
SMA barrel
harmonic mixer
LO
2
Specifications
Specifications describe the device’s warranted performance over the temperature range 0 to 55° C (except where noted).
IF range DC to 1300 MHz
LO amplitude range +14 to +18 dBm
Calibration accuracy (with IF of 321.4 MHz or 322.5 MHz) 11970A/Q/U
11970V/W
±2.0 dB with LO amplitude range of +14.5 to +16 dBm; ±3.0 dB with LO amplitude range of +16 to +18 dBm ±2.6 dB with LO amplitude range of +14.5 to +16 dBm; ±3.2 dB with LO amplitude range of +16 to +18 dBm
Nominal RF input SWR 11970A/Q/U 11970V/W
< 2.2:1 < 2.6:1
Bias requirements None
Nominal odd-order suppression 11970A/Q/U 11970V/W
> 20 dB > 15 dB
Maximum CW RF input level +20 dBm (100 mW)
Maximum peak pulse power +24 dBm (250 mW) with < 1 μs
pulse (average power +20 dBm)
1
50
45
40
35
30
Conversion loss, dB
25
20
K
18 26.5 33 40 50 60 75 110
Nominal frequency response & conversion loss
Q U
A
Input frequency, GHz
V
W
Nominal performance, as noted in the following tables, while not warranted, provides information useful in applying the device.
11970 Series tested with Agilent PSA spectrum analyzers
Agilent model number
11970K 18 to 26.5 6– 24 –105 ±1.9 –3
11970A 26.5 to 40 8– 26 –102 ±1.9 –5
11970Q 33 to 50 10– 28 –101 ±1.9 –7
11970U 40 to 60 10– 28 –101 ±1.9 –7
11970V 50 to 75 14– 40 –92 ±2.1 –3
11970W 75 to 110 18– 46 –85 ±3.0 –1
Frequency range (GHz)
LO harmonic number
Maximum conversion loss (dB)
Nominal spectrum analyzer noise (dBm) 1 kHz BW
Frequency response (dB)
Nominal gain compression (dBm)
1. An LO power of between 14.5 to 16 dBm at the mixer’s LO input is necessary to achieve the given frequency response and spectrum analyzer amplitude accuracy specifications. When LO power varies between 14 to 18 dBm at the mixer’s LO input, add ±1 dB to the frequency response and spectrum analyzer amplitude accuracy specification.
3
11970 Series tested with Agilent PXA signal analyzer
Agilent model number
11970A 26.5 to 40 6–/8– 26 –110/–108 ±1.9 –5
11970Q 33 to 50 8–/10– 28 –108/–106 ±1.9 –7
11970U 40 to 60 10– 28 –106 ±1.9 –7
11970V 50 to 75 12–/14– 40 –96/–94 ±2.1 –3
11970W 75 to 110 18– 46 –88 ±3.0 –1
Agilent model number
11970K UG-595/U
11970A UG-599/U
11970Q UG-383/U
11970U UG-383/U-M
11970V UG-385/U
11970W UG-387/U
Frequency range (GHz)
Flange Weight X Y Z
WR-42
WR-28
WR-22
WR-19
WR-15
WR-10
LO harmonic number
0.17 kg
0.36 lb
0.14 kg
0.32 lb
0.14 kg
0.32 lb
0.14 kg
0.32 lb
0.14 kg
0.32 lb
0.14 kg
0.32 lb
Maximum
1
36 mm
1.4 in
36 mm
1.4 in
36 mm
1.4 in
36 mm
1.4 in
36 mm 51 mm 76 mm
1.4 in 2.0 in 3.0 in
36 mm
1.4 in
conversion loss (dB)
51 mm
2.0 in
51 mm
2.0 in
51 mm
2.0 in
51 mm
2.0 in
51 mm
2.0 in
Nominal signal analyzer noise (dBm) 1 kHz BW
90 mm
3.5 in
71 mm
2.8 in
76 mm
3.0 in
76 mm
3.0 in
76 mm
3.0 in
2
Frequency response (dB)
Nominal gain compression (dBm)
Ordering Information
Options:
11970K* 18 to 26.5 GHz mixer
11970A 26.5 to 40 GHz mixer
11970Q 33 to 50 GHz mixer
11970U 40 to 60 GHz mixer
11970V 50 to 75 GHz mixer
11970W 75 to 110 GHz mixer
11970X-009
11970X-C01
N9029A-E13**
* Not compatible with N9030A PXA ** Required if the 11970 mixer is used with the Agilent PXA
Mixer connection kit containing three 1-meter SMA cables, wrench, and allen driver. Note: X = K, A, Q, U, V or W
Improved EMC radiated susceptibility. Note: X = K, A, Q, U, V or W
Ordering convenience for one OML INC. LO/IF diplexer with cables and adapter
Warranty and Service
Standard warranty for 11970 mixers is 12 months Standard warranty for N9029A-E13 diplexer is 3 months, provided by OML, Inc.
1. When used with 11970 Series mixer in A-, Q-, or V-band, the PXA’s LO harmonics are automatically switched between two different numbers as listed to optimize conversion loss.
2. If the LO harmonics are switched, the noise levels for the signal analyzer/mixer combination will change, corresponding to the different LO harmonic numbers.
4
www.agilent.com
TM
www.agilent.com/find/externalmixers
Agilent Advantage Services is committed to your success throughout your equipment’s lifetime. To keep you competi­tive, we continually invest in tools and processes that speed up calibration and repair and reduce your cost of ownership. You can also use Infoline Web Services to manage equip­ment and services more effectively. By sharing our measure­ment and service expertise, we help you create the products that change our world.
www.agilent.com/find/advantageservices
Agilent Email Updates
www.agilent.com/find/emailupdates
Get the latest information on the products and applications you select.
www.axiestandard.org
AdvancedTCA® Extensions for Instrumentation and Test (AXIe) is an open standard that extends the Advanced TCA for general purpose and semiconductor test. Agilent is a founding member of the AXIe consortium.
www.lxistandard.org
LAN eXtensions for Instruments puts the power of Ethernet and the Web inside your test systems. Agilent is a founding member of the LXI consortium.
www.pxisa.org
PCI eXtensions for Instrumentation (PXI) modular instrumentation delivers a rugged, PC-based high­performance measurement and automation system.
For more information on Agilent Technologies’ products, applications or services, please contact your local Agilent
office. The complete list is available at:
www.agilent.com/find/contactus
Americas
Canada (877) 894 4414 Brazil (11) 4197 3500 Mexico 01800 5064 800 United States (800) 829 4444
Asia Pacific
Australia 1 800 629 485 China 800 810 0189 Hong Kong 800 938 693 India 1 800 112 929 Japan 0120 (421) 345 Korea 080 769 0800 Malaysia 1 800 888 848 Singapore 1 800 375 8100 Taiwan 0800 047 866 Other AP Countries (65) 375 8100
Europe & Middle East
Belgium 32 (0) 2 404 93 40 Denmark 45 70 13 15 15 Finland 358 (0) 10 855 2100 France 0825 010 700 * 0.125 €/minute Germany 49 (0) 7031 464 6333 Ireland 1890 924 204 Israel 972-3-9288-504/544 Italy 39 02 92 60 8484 Netherlands 31 (0) 20 547 2111 Spain 34 (91) 631 3300 Sweden 0200-88 22 55 United Kingdom 44 (0) 131 452 0200
For other unlisted countries:
www.agilent.com/find/contactus
Revised: June 8, 2011