For use with the Agilent E4407B, 8560E/EC Series, 8566B,
71000 Series, and PSA Series spectrum analyzers,
plus the N9030A PXA signal analyzer
18 to 110 GHz
11970K*, 11970A, 11970Q,
11970U, 11970V, 11970W
Exceptional performance
The Agilent Technologies 11970 Series harmonic mixers are general purpose mixers
employing a dual-diode design to achieve very flat frequency response and low
conversion loss. Each mixer is calibrated across its full band:
11970K* 18 to 26.5 GHz
11970A 26.5 to 40 GHz
This series of mixers has been designed for a local oscillator frequency of 3 to 6.1 GHz.
Accurate absolute amplitude measurements can be made by using the mixer’s conversion loss
calibration chart. The SWR of the waveguide input is typically 2.2:1 to further minimize
measurement uncertainty. The combination of high gain-compression level and low
conversion loss provides the maximum dynamic range for measuring input signals.
* 11970K is excluded from PXA support
11970Q 33 to 50 GHz
11970U 40 to 60 GHz
11970V 50 to 75 GHz
11970W 75 to 110 GHz
Easy to use
The excellent frequency response and low conversion
loss are achieved without external dc bias or tuning stubs.
Since bias and tuning stubs are not required, manual
operation is simplified, and the complexity of hardware and
software for automatic systems is greatly reduced. The
repeatability of amplitude measurements is also enhanced.
The dual-diode design of the mixers further simplifies
measurements by suppressing the odd-order harmonics by
more than 20 dB, which makes identification of the mixing
products easier.
Figure 1. The 11970 Series mixers have separate LO input
and IF output ports
Frequency extension for the Agilent E4407B,
8560E/EC Series, 8566B, and 71000 Series
and PSA Series spectrum analyzers
The 11970 Series harmonic mixers are fully compatible
with the Agilent E4407B, 8560E/EC Series, 8566B (requires
a preamplifier), 71000, and PSA Series spectrum analyzers.
Accurate frequency and amplitude measurements are
made directly from the spectrum analyzer’s display after
calibration using the mixer’s calibration chart.
Frequency extension for the Agilent
N9030A PXA Series signal analyzer
The 11970 Series harmonic mixers (11970K excluded)
are also compatible with the N9030A PXA Series highperformance signal analyzer. The PXA offers optional
external mixing (Option EXM), enabling it to work with
the 11970 Series and other external mixers for frequency
extension. An external diplexer is required because the
PXA external mixing has a single SMA LO/IF port, whereas
the 11970 Series mixers have separate LO and IF ports.
Recommended diplexers can be purchased from Agilent
as N9020A-E13, or from OML Inc. as DLP.313B.
Rugged
The rugged Agilent 11970 Series mixers will
survive input levels up to 100 milliwatts (+20 dBm)
with no damage to the mixer diodes. They will
withstand shocks up to 30 G’s and the vibration
required by MIL-STD 28800C, Type III, Class 3 tests.
Agilent 8560E/EC
Series spectrum
analyzer
Extended
frequency
8560E/EC
Series
Agilent E4407B
spectrum analyzer
Extended frequency
E4407B ESA-E Series
IF input
1st LO output
Spectrum analyzer
Agilent 11970 Series
harmonic mixer
Agilent PSA Series spectrum analyzer
Extended frequency PSA Series
IF Input/1st LO output
Agilent N9030A
PXA Series signal analyzer
Extended frequency PXA Series
RF input
SMA cable
SMA cable
SMA cable
SMA cable
Agilent 11970 Series
IF
Mixer
IF
diplexer
LO
SMA barrel
harmonic mixer
LO
2
Specifications
Specifications describe the device’s warranted performance over the temperature range 0 to 55° C (except where noted).
IF rangeDC to 1300 MHz
LO amplitude range +14 to +18 dBm
Calibration accuracy
(with IF of 321.4 MHz or 322.5 MHz)
11970A/Q/U
11970V/W
±2.0 dB with LO amplitude range of +14.5 to +16 dBm;
±3.0 dB with LO amplitude range of +16 to +18 dBm
±2.6 dB with LO amplitude range of +14.5 to +16 dBm;
±3.2 dB with LO amplitude range of +16 to +18 dBm
Nominal RF input SWR
11970A/Q/U
11970V/W
< 2.2:1
< 2.6:1
Bias requirements None
Nominal odd-order suppression
11970A/Q/U
11970V/W
> 20 dB
> 15 dB
Maximum CW RF input level +20 dBm (100 mW)
Maximum peak pulse power +24 dBm (250 mW) with < 1 μs
pulse (average power +20 dBm)
1
50
45
40
35
30
Conversion loss, dB
25
20
K
18 26.5 33 40 50 60 75 110
Nominal frequency response & conversion loss
QU
A
Input frequency, GHz
V
W
Nominal performance, as noted in the following tables, while not warranted, provides information useful in applying the device.
11970 Series tested with Agilent PSA spectrum analyzers
Agilent model
number
11970K 18 to 26.56– 24 –105 ±1.9 –3
11970A 26.5 to 40 8– 26 –102 ±1.9 –5
11970Q 33 to 50 10– 28 –101 ±1.9 –7
11970U 40 to 60 10– 28 –101 ±1.9 –7
11970V 50 to 75 14– 40 –92 ±2.1 –3
11970W 75 to 110 18– 46 –85 ±3.0 –1
Frequency
range
(GHz)
LO
harmonic
number
Maximum
conversion
loss (dB)
Nominal spectrum
analyzer noise (dBm)
1 kHz BW
Frequency
response
(dB)
Nominal gain
compression
(dBm)
1. An LO power of between 14.5 to 16 dBm at the mixer’s LO input is necessary to achieve the given frequency response and spectrum analyzer amplitude
accuracy specifications. When LO power varies between 14 to 18 dBm at the mixer’s LO input, add ±1 dB to the frequency response and spectrum analyzer
amplitude accuracy specification.
3
11970 Series tested with Agilent PXA signal analyzer
Agilent model
number
11970A26.5 to 406–/8–26–110/–108±1.9–5
11970Q33 to 50 8–/10–28–108/–106±1.9–7
11970U40 to 6010–28–106±1.9–7
11970V 50 to 7512–/14–40 –96/–94±2.1–3
11970W 75 to 11018–46 –88±3.0–1
Agilent model
number
11970K UG-595/U
11970A UG-599/U
11970Q UG-383/U
11970U UG-383/U-M
11970V UG-385/U
11970WUG-387/U
Frequency
range
(GHz)
Flange Weight X Y Z
WR-42
WR-28
WR-22
WR-19
WR-15
WR-10
LO
harmonic
number
0.17 kg
0.36 lb
0.14 kg
0.32 lb
0.14 kg
0.32 lb
0.14 kg
0.32 lb
0.14 kg
0.32 lb
0.14 kg
0.32 lb
Maximum
1
36 mm
1.4 in
36 mm
1.4 in
36 mm
1.4 in
36 mm
1.4 in
36 mm 51 mm 76 mm
1.4 in 2.0 in 3.0 in
36 mm
1.4 in
conversion
loss (dB)
51 mm
2.0 in
51 mm
2.0 in
51 mm
2.0 in
51 mm
2.0 in
51 mm
2.0 in
Nominal signal
analyzer noise (dBm)
1 kHz BW
90 mm
3.5 in
71 mm
2.8 in
76 mm
3.0 in
76 mm
3.0 in
76 mm
3.0 in
2
Frequency
response
(dB)
Nominal gain
compression
(dBm)
Ordering Information
Options:
11970K* 18 to 26.5 GHz mixer
11970A 26.5 to 40 GHz mixer
11970Q 33 to 50 GHz mixer
11970U 40 to 60 GHz mixer
11970V 50 to 75 GHz mixer
11970W 75 to 110 GHz mixer
11970X-009
11970X-C01
N9029A-E13**
* Not compatible with N9030A PXA
** Required if the 11970 mixer is used with the Agilent PXA
Mixer connection kit containing three
1-meter SMA cables, wrench, and allen
driver. Note: X = K, A, Q, U, V or W
Improved EMC radiated susceptibility.
Note: X = K, A, Q, U, V or W
Ordering convenience for one OML INC.
LO/IF diplexer with cables and adapter
Warranty and Service
Standard warranty for 11970 mixers is 12 months
Standard warranty for N9029A-E13 diplexer is 3 months,
provided by OML, Inc.
1. When used with 11970 Series mixer in A-, Q-, or V-band, the PXA’s LO harmonics are automatically switched between two different numbers as listed
to optimize conversion loss.
2. If the LO harmonics are switched, the noise levels for the signal analyzer/mixer combination will change, corresponding to the different LO harmonic numbers.
4
www.agilent.com
TM
www.agilent.com/find/externalmixers
Agilent Advantage Services is committed to your success
throughout your equipment’s lifetime. To keep you competitive, we continually invest in tools and processes that speed
up calibration and repair and reduce your cost of ownership.
You can also use Infoline Web Services to manage equipment and services more effectively. By sharing our measurement and service expertise, we help you create the products
that change our world.
www.agilent.com/find/advantageservices
Agilent Email Updates
www.agilent.com/find/emailupdates
Get the latest information on the products and
applications you select.
www.axiestandard.org
AdvancedTCA® Extensions for Instrumentation and Test
(AXIe) is an open standard that extends the Advanced
TCA for general purpose and semiconductor test. Agilent
is a founding member of the AXIe consortium.
www.lxistandard.org
LAN eXtensions for Instruments puts the power of
Ethernet and the Web inside your test systems. Agilent
is a founding member of the LXI consortium.
www.pxisa.org
PCI eXtensions for Instrumentation (PXI) modular
instrumentation delivers a rugged, PC-based highperformance measurement and automation system.
For more information on Agilent Technologies’ products,
applications or services, please contact your local Agilent
office. The complete list is available at:
www.agilent.com/find/contactus
Americas
Canada (877) 894 4414
Brazil (11) 4197 3500
Mexico 01800 5064 800
United States (800) 829 4444
Asia Pacific
Australia 1 800 629 485
China 800 810 0189
Hong Kong 800 938 693
India 1 800 112 929
Japan 0120 (421) 345
Korea 080 769 0800
Malaysia 1 800 888 848
Singapore 1 800 375 8100
Taiwan 0800 047 866
Other AP Countries (65) 375 8100
Product specifications and descriptions in this document
subject to change without notice.
Get the best of both worlds: Agilent’s measurement
expertise and product breadth, combined with channel
partner convenience.