3
Ordering Information
1. With the purchase of a new
54800-series Infiniium oscilloscope:
•Option #012 (54810A, 54815A
54820A, 54825A, 54835A)
for 1156A
•Option #013 (54845A) for 1157A
•Option #014 (54846A) for 1158A
2. For existing 54800-Series Infiniium
oscilloscope customers order the
1156A**, 1157A**, or 1158A**
Each probe includes:
•Foam Case
•12 Resistive signal pins (orange) (E2640A)
• 2 Solderable-tip 5 cm resistive signal leads
(E2638A)
• 2 Socket-end 5 cm resistive signal leads
(01156-82105)
• 2 Socket-end 10 cm resistive signal leads
(01156-82108)
• 4 Micro clips (E2639A)*
• 12 Ground blade assemblies (E2641A)
• 12 Solderable SMT ground pins
(1253-5054)
• 12 Solderable through-hole ground pins
(1252-7150)
• 12 Offset ground pins (01156-27606)
• 2 Solderable-tip 5 cm ground leads
(E2638A)
• 2 Socket-end 5 cm ground leads
(01156-82103)
• 2 Socket-end 5 cm, 90° pin ground leads
(01156-82104)
• 1 User and Service Guide
• One-year warranty
* The 1156A only comes with 2 micro clips.
To order additional quantities, see
Accessories.
Accessories
E2637A Precision Measurement
kit (includes 2 solderable
ground sockets with
8 green resistive
signal pins)
E2638A Solderable - tip 5 cm
resistive signal leads (10)
with ground leads (3)
E2639A Micro clips, Qty. 4
E2640A Resistive signal pins,
(orange) Qty. 8
E2641A Ground blade assembly,
Qty. 8
E2654A EZ-Probe® Positioner
** The Infiniium 54800-Series scope
requires version A.04.30 or greater of
the application software to work with
the 1156A/7A/8A probes. An LS-120
drive is required for this upgrade.
To receive your free Infiniium
software update, go to our Infiniium
web site:
www.agilent.com/find/Infiniium1
Probe Recommendation Table
Be sure your probe has enough bandwidth to utilize the full performance your scope offers. Use the
selection table below for recommended configurations.
Infiniium 54800-Series oscilloscope Probe System Bandwidth
54835A (1.0 GHz) 1156A (1.5 GHz) 1.0 GHz
54845A (1.5 GHz) 1157A (2.5 GHz) 1.5 GHz
54846A (2.25 GHz) 1158A (4.0 GHz) 2.25 GHz
Specifications and Characteristics
Bandwidth
1
(–3dB)
1156A >1.5 GHz
1157A >2.5 GHz
1158A >4 GHz
System Bandwidth
1 GHz 1156A with 54835A
1.5 GHz 1157A with 54845A
2.25 GHz 1158A with 54846A
Rise and Fall time (10% to 90%)
1156A <233 ps
1157A <140 ps
1158A <88 ps
calculated from tr = 0.35/bandwidth
Input Capacitance 0.8 pF
Input Resistance1100 kΩ 1%
Flatness , Swept Response
0.2 dB, 100 kHz to 100 MHz
0.4 dB, 100 MHz to 2.5 GHz
2.0 dB, 2.5 GHz to 4.0 GHz
Flatness, Step Response
15% overshoot, 35 ps input edge
10% overshoot, 75 ps input edge
2%, 1 ns after edge
Dynamic Range
2
>5.0 V peak-to-peak
dc Attenuation
1
10:1 ±3% before calibration
4
10:1 ±1% after calibration
4
Zero offset error referred to input
1
<30 mV before calibration
<5 mV after calibration
Offset Range
1
±15.0 V
Offset Accuracy
1
<3% of setting before
calibration
4
1% of setting after
calibration
4
Noise referred to input
3.0 mVrms
Propagation Delay
5.5 ns
Maximum Input Voltage
40 V peak, CAT I
3
ESD Tolerance
>5 kV from 100pF,
300Ω HBM
Temperature Drift
Offset: <1.0 mV/°C
Attenuation (Gain):
0.1 %/°C
Environmental Conditions
Temperature
Operating 0 °C to +55 °C
Non-operating –40 °C to +70 °C
Humidity
Operating Up to 95% relative
humidity
(non-condensing)
at +40 °C
Non-operating Up to 90% relative
humidity at +65 °C
Power Requirements from
AutoProbe Interface
+12 Vdc @ 3 mA typical
+ 5 Vdc @ 35 mA typical
–5 Vdc @ 37 mA typical
Weight Approximately 0.69 kg
1
Denotes warranted specifications, all others are typical
2
For waveforms with edges >3 ns, the dynamic range is 12.0 Vpeak-to-peak
3
Installation category (over voltage category) I: Signal level, special equipment, or parts of equipment, telecommunication,
electronic, etc., with smaller transient overvoltage than installation category (overvoltage category) II.
4
Probe calibrated to scope channel (under Probes Setup menu)
Related literature
Infiniium 54800 Series Oscilloscope
Probes and Accessories
Data Sheet 5968-7141EUS/EN
Infiniium 54800 Series Oscilloscopes
Product Overview 5980-2397EUS/EN
Optimizing Oscilloscope
Measurement Accuracy on HighPerformance Systems with Agilent
Active Probes Application Note 1385
5988-5021EN