Features
• 4 GHz, 2.5 GHz, and 1.5 GHz bandwidth models
• Small size makes probing easier
• 88 ps rise time (on 4 GHz model)
• ±15 V offset
• 5 V peak-to-peak dynamic range
• Accessories designed for minimal loading on your device and optimum response
• Compatible with the Infiniium AutoProbe Interface
As the speeds in your design increase, you may notice more over­shoot, ringing, and other perturba­tions when connecting an oscillo­scope probe. Probes form a resonant circuit where they connect to the device. If this resonance is within the bandwidth of the oscilloscope probe you are using, it will be difficult to determine if the measured perturba­tions are due to your circuit or the probe.
Agilent is the only company that has overcome the resonance formed by the connection of a probe to a device. The Agilent 1156/7/8A probes optimize performance to make your job easier:
•High input impedance of 100 k
and 0.8 pF that never resonates low
•Flat frequency response across the entire bandwidth of the probe
•A full 5 V peak-to-peak dynamic range, and
•±15 V of offset.
Combined with the Agilent 54846A Infiniium scope, the 1158A 4 GHz probe offers you a full 2.25 GHz of system bandwidth, giving you accurate insight into your high-speed devices.
Agilent 1156/7/8A active probe with accessories
Agilent 1156/7/8A active probe with resistive signal pin and ground blade
2
Vin and Vout when driven from a 25source
Response = Vout/Vin
Notice how closely the output matches the input The flat response means the waveform on the scope screen will match
the waveform at the probe tip - across the entire 4 GHz bandwidth.
Small Size
Have you experienced problems with large, clunky probes? If so, you probably found your probe awkward to hold and had difficulty connecting to your signals. With the 1156/7/8A’s small size you can handle the probe expertly and gain access to tight spaces. Plus, the low mass makes the probe more durable. Agilent makes your job easier – giving you perform­ance that is easy to use.
Faithful Reproduction of Your Signal
Now you can accurately measure your high-speed signals without introducing errors from a probe that has a resonant input impedance or non-flat frequency response. Resistance is placed as close as possi­ble to the point being probed, which keeps the input impedance from res­onating low, and it also allows a flat frequency response across the entire bandwidth of the probe. Finally, there is a high bandwidth active probe where the waveform on screen matches the waveform at the probe tip. No other probe currently on the market offers a flat response for the entire bandwidth of a 4 GHz probe!
With the wide dynamic range and ±15 V of offset, the probe can be used in a wide variety of applications.
Superior Accessories
Your device under test (DUT) determines the type of probing accessories you need. Of course, there are electrical trade-offs depending on the type of connection you use. Longer connections from your DUT produce lower performance probing systems.
Agilent offers a variety of accessories optimized to give you the most accurate reproduction of your signal. In addition, the performance of each accessory is characterized for you. Now you can make informed decisions and get the best measurement for your environment. Superior performance combined with the knowledge to use it - that's how Agilent helps you do your job better.
This suite of accessories is properly damped to give you a flat transmitted response and non-resonant input impedance. Use these supplied accessories to get the best performance from your probe:
• Resistive signal pin (orange) E2640A
• Solderable-tip 5 cm resistive signal lead E2638A
• Socket-end 5 cm resistive signal lead 01156-82105
• Socket-end 10 cm resistive signal lead 01156-82108
• Ground blade assembly E2641A
• Solderable SMT ground pin 1253-5054
• Solderable through-hole ground pin 1252-7150
3
Ordering Information
1. With the purchase of a new 54800-series Infiniium oscilloscope:
•Option #012 (54810A, 54815A 54820A, 54825A, 54835A) for 1156A
•Option #013 (54845A) for 1157A
•Option #014 (54846A) for 1158A
2. For existing 54800-Series Infiniium oscilloscope customers order the 1156A**, 1157A**, or 1158A**
Each probe includes:
•Foam Case
•12 Resistive signal pins (orange) (E2640A)
• 2 Solderable-tip 5 cm resistive signal leads
(E2638A)
• 2 Socket-end 5 cm resistive signal leads
(01156-82105)
• 2 Socket-end 10 cm resistive signal leads
(01156-82108)
• 4 Micro clips (E2639A)*
• 12 Ground blade assemblies (E2641A)
• 12 Solderable SMT ground pins
(1253-5054)
• 12 Solderable through-hole ground pins
(1252-7150)
• 12 Offset ground pins (01156-27606)
• 2 Solderable-tip 5 cm ground leads
(E2638A)
• 2 Socket-end 5 cm ground leads
(01156-82103)
• 2 Socket-end 5 cm, 90° pin ground leads
(01156-82104)
• 1 User and Service Guide
• One-year warranty
* The 1156A only comes with 2 micro clips.
To order additional quantities, see Accessories.
Accessories
E2637A Precision Measurement
kit (includes 2 solderable ground sockets with 8 green resistive signal pins)
E2638A Solderable - tip 5 cm
resistive signal leads (10)
with ground leads (3) E2639A Micro clips, Qty. 4 E2640A Resistive signal pins,
(orange) Qty. 8 E2641A Ground blade assembly,
Qty. 8 E2654A EZ-Probe® Positioner
** The Infiniium 54800-Series scope requires version A.04.30 or greater of the application software to work with the 1156A/7A/8A probes. An LS-120 drive is required for this upgrade.
To receive your free Infiniium software update, go to our Infiniium web site:
www.agilent.com/find/Infiniium1
Probe Recommendation Table
Be sure your probe has enough bandwidth to utilize the full performance your scope offers. Use the selection table below for recommended configurations.
Infiniium 54800-Series oscilloscope Probe System Bandwidth
54835A (1.0 GHz) 1156A (1.5 GHz) 1.0 GHz 54845A (1.5 GHz) 1157A (2.5 GHz) 1.5 GHz 54846A (2.25 GHz) 1158A (4.0 GHz) 2.25 GHz
Specifications and Characteristics
Bandwidth
1
(–3dB) 1156A >1.5 GHz 1157A >2.5 GHz 1158A >4 GHz
System Bandwidth
1 GHz 1156A with 54835A
1.5 GHz 1157A with 54845A
2.25 GHz 1158A with 54846A
Rise and Fall time (10% to 90%)
1156A <233 ps 1157A <140 ps 1158A <88 ps
calculated from tr = 0.35/bandwidth Input Capacitance 0.8 pF Input Resistance1100 k1% Flatness , Swept Response
0.2 dB, 100 kHz to 100 MHz
0.4 dB, 100 MHz to 2.5 GHz
2.0 dB, 2.5 GHz to 4.0 GHz
Flatness, Step Response
15% overshoot, 35 ps input edge 10% overshoot, 75 ps input edge 2%, 1 ns after edge
Dynamic Range
2
>5.0 V peak-to-peak
dc Attenuation
1
10:1 ±3% before calibration
4
10:1 ±1% after calibration
4
Zero offset error referred to input
1
<30 mV before calibration <5 mV after calibration
Offset Range
1
±15.0 V
Offset Accuracy
1
<3% of setting before calibration
4
1% of setting after calibration
4
Noise referred to input
3.0 mVrms
Propagation Delay
5.5 ns
Maximum Input Voltage
40 V peak, CAT I
3
ESD Tolerance
>5 kV from 100pF, 300HBM
Temperature Drift
Offset: <1.0 mV/°C Attenuation (Gain):
0.1 %/°C
Environmental Conditions
Temperature
Operating 0 °C to +55 °C Non-operating –40 °C to +70 °C
Humidity
Operating Up to 95% relative
humidity (non-condensing) at +40 °C
Non-operating Up to 90% relative
humidity at +65 °C Power Requirements from AutoProbe Interface
+12 Vdc @ 3 mA typical + 5 Vdc @ 35 mA typical –5 Vdc @ 37 mA typical
Weight Approximately 0.69 kg
1
Denotes warranted specifications, all others are typical
2
For waveforms with edges >3 ns, the dynamic range is 12.0 Vpeak-to-peak
3
Installation category (over voltage category) I: Signal level, special equipment, or parts of equipment, telecommunication, electronic, etc., with smaller transient overvoltage than installation category (overvoltage category) II.
4
Probe calibrated to scope channel (under Probes Setup menu)
Related literature
Infiniium 54800 Series Oscilloscope Probes and Accessories
Data Sheet 5968-7141EUS/EN
Infiniium 54800 Series Oscilloscopes
Product Overview 5980-2397EUS/EN
Optimizing Oscilloscope Measurement Accuracy on High­Performance Systems with Agilent Active Probes Application Note 1385
5988-5021EN
Agilent Technologies’ Test and Measurement Support, Services, and Assistance
Agilent Technologies aims to maximize the value you receive, while minimizing your risk and problems. We strive to ensure that you get the test and measurement capabilities you paid for and obtain the sup­port you need. Our extensive support resources and services can help you choose the right Agilent products for your applications and apply them successfully. Every instrument and system we sell has a global warranty. Support is available for at least five years beyond the production life of the product. Two concepts underlie Agilent's overall support policy: "Our Promise" and "Your Advantage."
Our Promise
Our Promise means your Agilent test and measurement equipment will meet its advertised perform­ance and functionality. When you are choosing new equipment, we will help you with product informa­tion, including realistic performance specifications and practical recommendations from experienced test engineers. When you use Agilent equipment, we can verify that it works properly, help with prod­uct operation, and provide basic measurement assistance for the use of specified capabilities, at no extra cost upon request. Many self-help tools are available.
Your Advantage
Your Advantage means that Agilent offers a wide range of additional expert test and measurement serv­ices, which you can purchase according to your unique technical and business needs. Solve problems efficiently and gain a competitive edge by contracting with us for calibration, extra-cost upgrades, out­of-warranty repairs, and on-site education and training, as well as design, system integration, project management, and other professional engineering services. Experienced Agilent engineers and techni­cians worldwide can help you maximize your productivity, optimize the return on investment of your Agilent instruments and systems, and obtain dependable measurement accuracy for the life of those products. By internet, phone, or fax, get assistance with all your test & measurement needs
By internet, phone, or fax, get assistance with all your test & measurement needs
Phone or Fax
United States:
(tel) 800 829 4444
Canada:
(tel) 877 894 4414 (fax) 905 282 6495
China:
(tel) 800 810 0189 (fax) 800 820 2816
Europe:
(tel) (31 20) 547 2323 (fax) (31 20) 547 2390
Japan:
(tel) (81) 426 56 7832 (fax) (81) 426 56 7840
Korea:
(tel) (82 2) 2004 5004 (fax) (82 2) 2004 5115
Latin America:
(tel) (305) 269 7500 (fax) (305) 269 7599
Taiwan: (tel) 0800 047 866 (fax) 0800 286 331
Other Asia Pacific Countries:
(tel) (65) 6375 8100 (fax) (65) 6836 0252 Email: tm_asia@agilent.com
Online Assistance:
ww.agilent.com/find/assist
Product specifications and descriptions in this document subject to change without notice.
© Agilent Technologies, Inc. 2004 Printed in USA, March 5, 2004
5988-3361EN
www.agilent.com/find/emailupdates
Get the latest information on the products and applications you select.
Agilent T&M Software and Connectivity
Agilent's Test and Measurement software and connectivity products, solutions and developer network allows you to take time out of connecting your instruments to your computer with tools based on PC standards, so you can focus on your tasks, not on your connections. Visit www.agilent.com/find/connectivity for more information.
www.agilent.com/find/agilentdirect
Quickly choose and use your test equipment solutions with confidence.
Agilent Email Updates
Agilent Direct
Loading...