ATD Tools ATD-5540 User Manual

ATD-5540
AUTOMOTIVE
INSTRUCTIONS
INSTRUCTION
MANUAL
Contents
Safety
Meter Basics
Meter Functions
Voltage(V) …………………………………………………
Resistance(
Diode Check( )
Audible Continuity( )
DC Current(A)
RPM( )
Frequency(Freq)
Dwell( )
Duty Cycle(%)
Maintenance
Fuse and Battery Replacement
Trouble Shooting
General Specications
Electrical Specications Warranty Information
)
Ω
1
4
4
6
7
8
9
10
11
12
13
14
15
15
16
16
17
20
-1 0
Safety
Safety Cont ’d…
DANGER
Engines produce carbon monoxide which is
odorless, causes slower reaction time, and can lead to serious injury. When the engine is operating, keep service areas WELL VENTILATED or attach the vehicle exhaust system to the shop exhaust removal system.
Set the parking brake and block the wheels
before testing or repairing the vehicle. It is especially important to block the wheels on front-wheel drive vehicles; the parking brake does not hold the drive wheels.
Wear an eye shield when testing or repair-ing
vehicles.
Exceeding the limits of this meter is dangerous. lt will expose you to serious or possibly fatal injury. Carefully read and understand the cautions and the specification limits of this meter.
Voltage between any terminal and ground
must not exceed 600V DC or 600V AC.
Use caution when measuring voltage above
25V DC or 25V AC.
Circuit tested must be protected by a 10A fuse or
circuit breaker.
Do not use the meter if it has been damaged.
Do not use the test leads if the insulation is damaged or if metal
is exposed.
Danger
Avoid electrical shock: do not touch the test
leads, tips or the circuit being tested.
Do not try a voltage measurement with the test leads in the 10A or the mA terminal.
When testing for the presence of voltage or current, make sure
the meter is functioning correctly. Take a reading of a known voltage or current before accepting a zero reading.
Choose the proper range and function for the measurement.
Do not try voltage or current measurements that may exceed the ratings marked on the Function/Range switch or terminal.
When measuring current,connect the meter in series with the load.
Never connect more than one set of test leads to the meter.
Disconnect the live test lead before disconnecting the common
test lead.
The mA and the 10A terminals are protected by fuses. To avoid
possible injury or damage, use only in circuits limited to DC 10A for 60 seconds.
See also
Fuse Replacement
1 2
Safety Cont’d…
IMPORTANT
To maintain accuracy of the meter, replace the discharged
battery immediately when the battery symbol appears on
the meter display.
Avoid measuring error from outside interference: keep the meter
away from spark plug or coil wires.
Avoid damaging the meter when testing voltage: disconnect the
test leads from the test points before changing functions.
Do not exceed the limits shown in the table below:
FUNCTION Terminal Input limit
DC Volts VHzΩ 600 Volts DC
Frequency 500 Volts AC/DC
Meter Basics
1. Display:
Large 3 1/2 digit (1999 counts) LCD display with low battery
indication.
2. Rotary Selector Switch:
Turn this switch to Select a function or turn the meter OFF.
3. Test Lead Terminals:
The Black test is used in the common(COM) terminal for all
Ohm(resistance)
10 Amp measurement for 60 seconds maximum.
Ohms can not be measured if voltage is present, ohms can be
measured only in a non-powered circuit. However, the meter is
protected to 250 volts
Diode
DC 10A 10A *10A AC/DC
RPM
Duty Cycle(%)
Dwell angle
VHz
Ω
500 Volts AC/DC
VHz
Ω
3 4
250 Volts AC/DC
rests. The Red test Lead is used to measure Amp or Volts.
Meter Functions-Voltage (DCV)
The meter will automatically select the
best voltage (V) range.
Insert:
Black lead in COM terminal.
V-
Red lead in
the Black probe to ground or to the negative (-) circuit.
Touch the Red probe to the circuit coming from the power source
IMPORTANT: voltage must be measured in parallel (Red probe measuring circuit from power source).
Accuracy
Selection of a lower range will move the decimal point one place and increase the accuracy. An “I” display means the range is too low,select the next higher range.
Ω
-RPM
terminal Touch
When measuring voltage,be sure the Red test lead is in the terminal marked “V”.If the test lead is in an Amp (A)or Milliampere (mA) terminal, you may be injured or the meter damaged.
WARNING
5 6
Meter Functions – Resistance
(
Ω
)
Meter Functions – Diode Check ( )
IMPORTANT:
application that has capacitors
in the circuit, be sure to turn the
power OFF on the test circuit and
discharge all capacitors. Accurate
measurement is not possible if
external or residual voltage is
present.
Select the
range with the rotary switch.
Insert:
Black lead in
Red lead in V-
terminal.
Touch the test lead probes across
the resistor to be tested.
If you are testing an
resistance(
COM
terminal.
-RPM
Ω
Ω
IMPORTANT: Turn the power OFF to the test circuit
Select the
setting with the rotary switch.
Insert:
Black lead in COM terminal. Red lead in
Touch the Black test probe to the negative (-) side of the diode.
Touch the Red test probe to the
)
positive (+) side of the diode.
Reverse the probes : Black to the positive (+) side and Red to the negative (-) side.
Note:
good〃 diode will read low in
A one direction and high in the other direction when the probes are reversed (or vice versa). A defective diode will have the same reading in both directions or read between 1.0 to 3.0V. in both directions
Diode - to +
Good
Bad
Diode Check
-RPM
V-
Ω
.4 to .9V OL
OL .4 to .9V
OL 1.0 to 3.0V
1.0 to
.4 to .9V .4 to .9V OL
.000V
( )
terminal.
Reverse
Probes
+ to -
OL
OL .000V
7 8
Meter Functions – Audible Continuity ( )
r
Meter Functions DC Current (A)
IMPORTANT:
OFF on the test circuit
Select the
( ) range with the rotary
switch.
Insert:
Black lead in COM terminal.
Red lead in
terminal.
Connect one test probe to
each end of the circuit to be
tested.
Circuit complete, the meter
will beep continuously.
Circuit open, there is no beep
and the display shows to
“ I ” (over limit)
Turn the power
Audible Continuity
- RPM
V -
Ω
IMPORTANT :
flows through the meter. It is important that you do not: Measure current greater than
600 Volts AC or DC, with respect to ground.
Exceed 60 seconds when
measuring continuous current between 1A-10A. Allow five minutes for cool down before continuing.
Select the
rotary switch.
Insert:
Black lead in Red lead in the
IMPORTANT: Turn OFF all power to the circuit or disconnect the circuit from the power source.
Connect: The Red probe to the side of the
circuit closest to the power source.
The Black probe to the side of the
circuit to ground.
Turn the power ON and test.
All current measured
10A
range with the
COM
terminal.
10A
terminal
Note: Current must always be measured with the mete test probes connected in series, as described.
-
9 10
Meter Functions-RPM/x10RPM
Select the
rotary switch.
or Select the the rotary switch (1,000 to 12,000 RPM). Multiply the displayed reading times by ten to get actual RPM.
Insert the inductive pickup connecting terminal into the meter.
Ground lead in COM terminal.
Output lead in
terminal.
Connect the inductive pickup to a spark plug wire. lf no reading is received, unhook the clamp, turn it over and connect again.
Note:
·
Position the inductive pick-up as far away from the distributor and the exhaust manifold as possible.
·
Position the inductive pick-up to within six inches of the spark plug or move it to another plug wire if no reading or an erratic reading is received.
RPM
range with the
x 10RPM
- RPM
V -
Ω
range with
Meter Functions-Frequency(Hz)
Select the Frequency(Freq)
setting with the rotary switch.
Set the rotary switch to the
Frequency range that gives the
most accurate measurement
reading.
Insert:
Black lead in
Red lead in
terminal. Connect the
Black test probe to
ground.
Connect the Red test probe
to the ”signal out” wire of the
sensor to be tested.
COM
V-
Ω
terminal.
-RPM
11 12
Meter Functions-Dwell
Select the proper
range with the rotary switch.
Insert:
·Black lead in COM terminal.
Dwell
Meter Functions-Duty Cycle (%)
Select the %
range with the rotary switch.
Insert:
Duty Cycle
RPM
-
·Red lead in V-
Connect the Black test probe
to ground.
Connect the Red test probe to the
wire that connects to the breaker
points(see illustration).
Ω
terminal.
·Black lead in
·Red lead in V-
Connect the Black test probe to
ground.
Connect the Red test probe to the
signal wire circuit.
The illustration for a mixture control
solenoid is shown with the metering
rod in the closed position. The
meter will display the percentage of
time the plunger is in the closed
position (low duty cycle) during one
duty cycle.
COM
Ω
terminal.
RPM
-
terminal.
13 14
Maintenance
Trouble Shooting
Fuse and Battery Replacement
WARNING:
Avoid electrical shock: remove test leads before opening
case.
Do not operate the meter or rotate the meter switch when
the case is open.
1. To replace a battery or fuse, loosen the three screws in the case
back and remove the case by lifting up and forward .
2. Replace the battery with an 9 Volt alkaline battery.
Important:
To prevent contamination of the circuits, your hands must
be clean and the printed circuit board must be held by the
edges.
Replace the fuses with the same type of fuse.
10 A is a F 10 A ,250V high energy, fast acting fuse.
Make sure the replacement fuse is centered in the fuse
holder.
3.Re-assemble the case. Fasten the three screws.
1. Meter wilI not turn ON.
Check the battery contacts for a tight fit.
Check for a minimum battery voltage of 8.0 volts.
2. Ampere reading is erratic or there is no reading at all.
Disassemble the meter back cover and test the fuses for
continuity.
3. Meter reading is erratic.
Printed circuit board contaminated from handling with hands.
Low battery.
Open circuit in a test lead (frayed or broken wire).
Wrong range selected.
“Blown” fuse.
General Specifications
GENERAL SPECIFICATIONS
Display :
Polarity : Automatic,(-)negative polarity indication.
Overrange lndication :“
Low Battery lndication:
Measurement Rate :
Operating Environment :
3 1/2digit(1999 counts)liquid crystal display(LCD)
1” mark indication.
The is displayed when the battery voltage drops below the operating level.
2 times per second ,nominal.
0to 50 (320
0
F to 122 0F)
15 16
at<70% R.H.
Storage Environment : -20to 60℃ (-4 at<80%R.H
Temperature Coefficient :
Power : Single standard 9 Volt battery (NEDA 1604 or lEC6F22).
Battery Life :
Fuse :
Dimensions : 147mm(H)×70mm(W)×39mm(D).
Weight :
10A/600V,6.3×25mm fast acting ceramic type.
200 hours typical with alkaline battery.
Approx.222g(Meter Only),355g(With Holster).
0.2×(specified accuracy ) / ( < 18 ℃ or > 28 ).
0
F to140 0F )
%DUTY CYCLE
Ranges:1.0%-90.0%
Resolution:0.1%
Pulse width:>100us,<100ms
Accuracy:±(2.0%rdg+5dgts)
Overload protection: 500VDC or RMS AC
DWELL ANGLE
No.of cylinders:3,4,5,6,8
Ranges:0-120.0
0-60.0
Resolution:0.1
Accuracy:±(2.0%rdg+5dgts)
0
(3CYL),0-90.00(4CYL),0-72.00(5CYL),
0
(6CYL),0-45.00(8CYL)
Electrical Specifications
ELECTRICAL SPECIFICATIONS*
Accuracy is given as±([%of reading]+[number of least
significant digits])at 18 humidity up to 70%.
RPM(Tach)
Ranges:600-2000, 6000-12000(x10RPM)
Resolution:1 RPM
Effect Reading:>600RPM
Accuracy:±(2%rdg+4 dgts)
Overload protection:500VDC or RMS AC
0
C to 280C(650F to 830F),with relative
17 18
Overload protection: 500VDC or RMS AC
DC VOLTAGE
Ranges : 200mV, 2V, 20V, 200V, 600V Resolution :100
Accuracy :±(0.5%rdg+1dgt)on 200mV
±(0.8%rdg+1dgt) 2V to 600V ranges
Input impedence :10M
Overload protection:600VDC or AC rms.
DC CURRENT
Ranges:10A
Resolution:10mA
Accuracy:±(3.0%rdg+3dgts)
μ
V
Ω
Input protection:10A/250V fast acting ceramic fuse on 10A input
RESISTANCE
Ranges : 200
Resolution : 100m
Accuracy :±(0.8%rdg+3dgts) on 200
±(0.8%rdg+2dgts) on 2K
±(1.5%rdg+5dgts) on 20M
Overload protection:250VDC or RMS AC
, 2KΩ, 20KΩ, 2MΩ, 20MΩ
Ω
Ω
to 2MΩ ranges
Ω
Ω
ranges
FREQUENCY
Ranges : 200Hz, 2000Hz, 20KHz
Resolution : 0.1Hz
Accuracy :±(1.0%rdg+4dgts) on all ranges
Sensitivity : 3.5V RMS min. at>20% and <80% duty cycle
Effect Reading : More than 100 digits at pulse width>2 Overload protection : 500 VDC or RMS AC
DIODE TEST
Test current : 0.6mA typical(Vf=0.6V)
Resolution :1mV
Accuracy :±(10%rdg+3dgts)
Open circuit voltage : 3.0Vdc typical
Overload protection : 250VDC or RMS AC
μ
Sec
AUDIBLE CONTINUITY
Audible threshold : Less than 50
Resolution:100m
Overload protection:250VDC or RMS AC
Ω
Ω
Warranty Information
This product is warranted to be free from defects for one year. If this product fails during the first 12 months due to faulty materials or workmanship, it will be replaced free of charge, at the discretion of the manufacturer. NOTE: This one year warranty does not cover dead batteries and blown fuses.
For warranty and service coverage, please return this product to your dealer for processing and evaluation. OR, return it directly to:
Electronic Specialties, Inc. 139 Elizabeth Ln. Genoa City, WI 53128 262-279-1400 WWW.ESITEST.COM
Defective units being returned to your dealer or to the factory should include proof of purchase date.
Any testers that do not function due to misuse or abuse will be subject to “out of warranty service charges.”
19 20
Loading...