Argus 707V User manual

4.5 (13)

Service Manual for Volumetric Infusion Pump ARGUS 707 V

Made in Switzerland

ARGUS Medical AG, CH-3627 Heimberg / Switzerland

(a member of the CODAN group)

14.203.B_A707 en.SM.V4.3X

ARGUS Medical AG 28.03.06 / PJ

1 / 48

 

 

TABLE OF CONTENTS

 

1.

INTRODUCTION ...........................................................................................................

4

 

1.1.

General .............................................................................................................................

4

2.

PUMP CONFIGURATIONS ...........................................................................................

4

 

2.1.

General .............................................................................................................................

4

 

2.2. Interrogation mode (without ARGUS service)...................................................................

5

 

2.3. Configuration mode (without ARGUS service)..................................................................

6

 

2.4. First activation of a PIN code (write protection) ................................................................

7

 

2.5. Changing an existing PIN code.........................................................................................

8

 

2.6. Address list of the pump configuration (without ARGUS service) .....................................

9

 

2.7. Index list of the pump configuration (with ARGUS service) ............................................

10

 

2.8. Special configuration options (with ARGUS service) ......................................................

14

 

2.9.

Medication list .................................................................................................................

15

3.

Serial communication of the pump ..........................................................................

18

 

3.1.

General ...........................................................................................................................

18

 

3.2.

Serial communication protocol........................................................................................

18

4.

ARGUS service...........................................................................................................

19

 

4.1.

General ...........................................................................................................................

19

 

4.2. ARGUS service – Configuration......................................................................................

20

 

4.3. ARGUS service - Calibration (volumetric pums only) .....................................................

21

 

4.4.

ARGUS service - Toolbox...............................................................................................

22

5.

SOFTWARE UPDATES ..............................................................................................

26

 

5.1.

General ...........................................................................................................................

26

 

5.2. Requirements for a software update...............................................................................

26

 

5.3.

Software update procedure.............................................................................................

26

 

5.4.

Safety aspects.................................................................................................................

28

6.

MAINTENANCE ..........................................................................................................

29

 

6.1.

General ...........................................................................................................................

29

 

6.2.

Recalibration ...................................................................................................................

29

 

6.3.

Pressure calibration ........................................................................................................

30

 

6.4.

Pressure control measurement .......................................................................................

35

 

6.5.

Volume calibration ..........................................................................................................

36

 

6.6.

Pump specifications ........................................................................................................

37

 

6.7.

Fault codes .....................................................................................................................

38

7.

REPLACEMENT OF PARTS.......................................................................................

39

 

7.1.

General ...........................................................................................................................

39

 

7.2. Disassembling of the ARGUS 707 V...............................................................................

39

 

7.3.

Spare parts .....................................................................................................................

43

8.

WIRING DIAGRAMM...................................................................................................

46

9.

BLOC SCHEMATIC ....................................................................................................

46

10. SAFETY STANDARD CHECK ....................................................................................

47

11. REPAIR ORDER FORM..............................................................................................

48

14.203.B_A707 en.SM.V4.3X

ARGUS Medical AG 28.03.06 / PJ

1 / 48

INTRODUCTION

Park area for drop detector

Display

Drop detector

Drop chamber

Global alarm

Air bubble detector

Pressure sensor

automatic

Pressure sensor

„upstream“

„free-flow“-clamp

„downstream“

14.203.B_A707 en.SM.V4.3X

ARGUS Medical AG 28.03.06 / PJ

2 / 48

INTRODUCTION

Universal clamp

Staff alert and

 

RS-232-connection

Drop detector connection

AC power

Identification plate

with pump serial

connection

number

 

DC-connection & QUICK®100

Docking Station interface

14.203.B_A707 en.SM.V4.3X

ARGUS Medical AG 28.03.06 / PJ

3 / 48

INTRODUCTION

1.INTRODUCTION

1.1.General

IMPORTANT!

This service manual is intended for the exclusive use of authorized persons who have been trained by ARGUS Medical AG in the maintenance and repair of the

ARGUS 707 V infusion pump.

The service manual is meant to be used together with the user manual.

IMPORTANT!

ARGUS Medical AG shall not assume any responsibility for any manipulations which have been carried out on the ARGUS 707 V infusion pump by a nonauthorized person.

CAUTION!

The ARGUS 707 V infusion pump may only be used with spare parts, accessories, consumables and IV-sets with Luer-Lock connections recommended by ARGUS Medical AG. The functional safety of the pump is not guaranteed if non approved materials are used. The safety of the patient may be endangered.

This manual contains the latest data available. It is subject to further modifications in accordance with technical improvements.

2.PUMP CONFIGURATIONS

2.1.General

CAUTION!

The configuration possibilities with the “ARGUS service” PC utility tool and without

PC assistance constitute a modification of the pump and may only be carried out by authorized persons!

CAUTION!

After changing the configuration a function check and a control measurement has to be performed!

14.203.B_A707 en.SM.V4.3X

ARGUS Medical AG 28.03.06 / PJ

4 / 48

PUMP CONFIGURATIONS

2.2.Interrogation mode (without ARGUS service)

With the interrogation mode you can read the present keypad configuration of the pump without the possibility to modify any configurations. For a complete overview, please take the “ARGUS service” PC-tool!

1.To enter into the interrogation mode, switch the pump on while keeping the keys “MODE” and “START/STOP” pressed.

707 PRog

...5. 1

2. Press the “START/ STOP” key.

4.To quit the interrogation mode press the “ON/OFF” key at least 2 seconds.

3.Enter the requested address (see chapter 2.6), e.g. 5. The programmed value of the address 5 appears in the right hand display.

Flashing decimal points indicate which display is ready to accept an input by the keys 100, 10 & 1.

In the interrogation mode the left hand display shows the address and the right hand display shows the according value configured at this address. Please refer to chapter 2.6. where the meanings of the addresses are explained.

To modify any configuration data you have to go into the configuration mode.

14.203.B_A707 en.SM.V4.3X

ARGUS Medical AG 28.03.06 / PJ

5 / 48

PUMP CONFIGURATIONS

2.3.Configuration mode (without ARGUS service)

The configuration mode permit you to modify the pump keypad configuration manually using the keypad. Please refer to chapter 2.6. where the meaning of the adresses are explained. To have access to all configuration options, please use the

“ARGUS service” PC-tool!

2.3.1.Step 1

1.To enter into the configuration mode, switch the pump on while keeping the keys “MODE” and “START/STOP” pressed.

2.Press key “START”

707 prog

 

4. Press key “START” again.

3. Press key “MODE”

The display will show “Cod ----“

 

2.3.2. Step 2

5.Enter the PIN Code. If no PIN code was configured, no entry is required

 

6. Press key “START”

Cod ----

to acknowledge the

 

entered PIN code

If the PIN was not accepted, the display will change back to point 5.

You have now access to all addresses in the list of chapter 2.6.

Select therefore any address in the left display (see next page).

14.203.B_A707 en.SM.V4.3X

ARGUS Medical AG 28.03.06 / PJ

6 / 48

Argus 707V User manual

 

 

 

 

PUMP CONFIGURATIONS

2.3.3. Step 3

 

 

 

 

 

 

 

8. Enter now the value on the right hand display. The

 

 

7. After entering an address e.g. 5 press the

 

 

 

range of the value is given by the table in chap.2.6

key “MODE”. The flashing decimal points

 

 

 

 

 

will change to the right hand display.

 

 

 

 

 

 

 

...5. 0

9. Press key “START” to acknowledge the value entered. The flashing decimal points change back to the left hand display.

10.To quit the configuration mode press the “ON/OFF” key at least 2 seconds. Changes in configuration become active,

after the pump is switched on normally again.

Important remark:

Invalid values entered will be corrected automatically by the pump to the maxima or minima value allowed for the according address!

2.4.First activation of a PIN code (write protection)

The activation of a PIN code allows you to protect the configuration from unauthorized access. To activate the PIN code, enter into the configuration mode.

1.To enter into the configuration mode, switch the pump on while keeping the keys “MODE” and “START/STOP” pressed.

2.Press key “START”

707

PRog

 

 

 

 

 

4. Press key “START”

 

 

 

 

 

 

 

 

again. The display will

 

 

 

 

show “Cod ----“.

 

 

 

 

 

 

 

 

5. Press “MODE” key (Code “0” will be set) then “START”

 

key to acknowledge the entered PIN code “0”.

3. Press key “MODE”

 

 

 

 

 

14.203.B_A707 en.SM.V4.3X

ARGUS Medical AG 28.03.06 / PJ

7 / 48

PUMP CONFIGURATIONS

 

 

 

 

 

 

 

6.

Press the “MODE” key.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

The flashing decimal

0

...0.

 

 

 

points will change to the

 

 

 

right hand display

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

8.

Press key “START” to

 

 

 

 

 

 

 

 

 

 

 

 

acknowledge to new PIN.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

7.Enter now the new PIN Code (max. 4 digit number).

Please remember this code, it will never be visible again! e.g.

CAUTION!

After you switch the pump OFF and ON again you can enter into the configuration mode only, if you enter the correct PIN code.

e.g.

Please note: The interrogation mode can always be accessed without the PIN.

2.5.Changing an existing PIN code

Enter the configuration mode using present PIN, select add. “0” and set the new code.

Enter actual PIN code and confirm with “START” key.

Press “MODE” key (#0). The flashing decimal points will change to the right hand display.

Enter the new PIN code and press the “START/STOP” key to acknowledge the entered code.

14.203.B_A707 en.SM.V4.3X

ARGUS Medical AG 28.03.06 / PJ

8 / 48

PUMP CONFIGURATIONS

2.6.Address list of the pump configuration (without ARGUS service)

The following list declares the possible configuration options which can be performed on the pump keypad without using the PC.

All these options can also be configured by the PC-Software “ARGUS service”.

Address

Index

Default

Function

Unit

Range

left display

PC

 

 

 

right display

 

 

 

 

 

 

1

2

No

Key ON/OFF only at stop valid

-

0=No / 1=Yes

2

10

No

Operation without drop detector

-

0=No / 1=Yes

3

11

No

Recall of the last used ml/h rate

-

0=No / 1=Yes

4

19

No

Buzzer at start

-

0=No / 1=Yes

5

44

Yes

Automatic pressure release

-

0=No / 1=Yes

6

49

No

Alarm acknowledge with key MODE

-

0=No / 1=Yes

 

 

 

 

 

 

 

 

 

 

 

 

100

361

5

Key ON/OFF delay time

• 0.1 s

0 - 31

101

362

8

Display brightness

level

2 - 15

102

363

7

Buzzer volume

level

5 - 10

103

316

9

Default pressure limit (IV-set 1)

• 100 mbar

1 - 10

104

318

20

Drop rate window centre (IV-set 1)

Drops/ml

10 - 65

 

 

 

 

 

 

 

 

 

 

 

 

200

317

250

Air bubble size (IV-set 1)

µl

50 - 1000

 

 

 

 

 

 

 

 

 

 

 

 

399

-

707

Enter the calibration menu (volume)

-

123

 

-

707

Enter the calibration menu (pressure)

-

1234

Note!

The address does not correspond with the index used by the “ARGUS service” tool.

14.203.B_A707 en.SM.V4.3X

ARGUS Medical AG 28.03.06 / PJ

9 / 48

PUMP CONFIGURATIONS

2.7. Index list of the pump configuration (with ARGUS service)

Index PC

Add.

Default

 

Function

 

Unit

Range

 

 

 

 

 

 

 

 

1

 

Yes

 

Run indication by running decimal point

 

-

No / Yes

2

1

No

 

Key ON/OFF only at stop valid

 

-

No / Yes

3

 

No

 

Rate change allowed only at STOP

 

-

No / Yes

4

 

No

 

Key STOP delayed

 

# 361

-

No / Yes

5

 

No

 

Second entry of rate

 

# 3=Yes

-

No / Yes

6

 

No

 

Static alarm (staff alerting system)

 

 

-

No / Yes

7

 

No

 

Display elapsed time in run mode

 

# 8=No

-

No / Yes

8

 

No

 

Display remaining time

 

# 7=Yes

-

No / Yes

9

 

 

 

 

 

 

 

 

 

10

2

No

 

Operation without drop detector

 

-

No / Yes

11

3

Yes

 

Recall of ml/h rate at next power on

 

-

No / Yes

12

 

No

 

Recall of ml total at next power on

 

-

No / Yes

13

 

No

 

Recall of ml inf. at next power on

 

-

No / Yes

14

 

No

 

SBS (step by step function)

 

-

No / Yes

15

 

No

 

Display VTBI (volume to be infused)

 

-

No / Yes

16

 

No

 

Display “Set-X” if only one set is enabled

 

-

No / Yes

17

 

Yes

 

KVO (KOR) enabled

 

# 60

-

No / Yes

18

 

No

 

Drop alarm only if bottle empty

 

# 10=No, → trA

-

No / Yes

19

4

No

 

Buzzer at start

 

 

-

No / Yes

20

 

No

 

Menu Clr (clear ml inf.)

 

# 15=No

-

No / Yes

21

 

 

 

 

 

 

 

 

 

22

 

No

 

Menu trA (transport)

 

# 10, 18=No

-

No / Yes

23

 

Yes

 

Menu PrL (pressure alarm limit)

 

 

-

No / Yes

24

 

Yes

 

Menu CAP (battery capacity)

 

-

No / Yes

25

 

No

 

Menu SEt Fill (prime IV set)

 

 

 

 

26

 

No

 

Menu InF (ml inf. since last power on)

 

-

No / Yes

27

 

No

 

Menu dLo (data lock)

 

-

No / Yes

28

 

No

 

Menu Stb (stand by)

 

-

No / Yes

29

 

No

 

Menu Med (medication name)

 

-

No / Yes

30

 

No

 

Menu tM (timer alarm)

 

-

No / Yes

31

 

 

 

 

 

 

 

 

 

32

 

No

 

Menu boL (bolus release)

 

-

No / Yes

33

 

No

 

Menu boLr (bolus rate)

 

# 32=Yes

-

No / Yes

34

 

No

 

Menu tot (bolus total)

 

# 32=Yes

-

No / Yes

35

 

 

 

 

 

 

 

 

 

36

 

 

 

 

 

 

 

 

 

37

 

 

 

 

 

 

 

 

 

38

 

No

 

Automatic bolus application

 

# 32, 34=Yes

-

No / Yes

39

 

No

 

Neonatology mode accessible in PrL menu

 

-

No / Yes

40

 

 

 

 

 

 

 

 

 

41

 

No

 

Clear ml/h after infusion completed

 

-

No / Yes

42

 

No

 

Clear ml total after infusion completed

 

# 41=Yes

-

No / Yes

43

 

No

 

Air volume accumulated

 

 

-

No / Yes

44

5

Yes

 

Automatic pressure release after occlusion

 

-

No / Yes

45

 

Yes

 

Pressure display ON (LED bar graph - 20/40/60/80/100%)

-

No / Yes

46

 

No

 

Pressure display with indicator

 

# 45=Yes

-

No / Yes

47

 

No

 

Stand byand battery pre alarm low volume

-

No / Yes

48

 

No

 

Flashing numeric display at alarm

 

-

No / Yes

49

6

No

 

Alarm acknowledge only with key MODE

 

-

No / Yes

 

 

 

 

 

 

 

 

 

 

 

 

14.203.B_A707 en.SM.V4.3X

ARGUS Medical AG 28.03.06 / PJ

 

10 / 48

 

PUMP CONFIGURATIONS

50

 

No

 

Start with ≥ 1 bar pressure allowed

 

 

 

-

No / Yes

51

 

Yes

 

Enable upstream occlusion detection

 

 

 

-

No / Yes

52

 

 

 

 

 

 

 

 

 

 

 

 

 

53

 

 

 

 

 

 

 

 

 

 

 

 

 

54

 

 

 

 

 

 

 

 

 

 

 

 

 

55

 

No

 

Select RJ-45 connector for serial interface

# 29=Yes

-

No / Yes

56

 

Yes

 

Medication display alternate

 

 

-

No / Yes

 

 

 

 

 

 

 

 

 

 

 

 

60

 

No

 

KVO only after infusion completed

 

 

 

-

No / Yes

 

 

 

 

 

 

 

 

 

 

 

 

65

 

No

 

Clear and continue

 

 

# 15=No

 

-

No / Yes

 

 

 

 

 

 

 

 

 

 

 

101

 

Yes

 

Set 1 enabled

 

 

 

-

No / Yes

102

 

No

 

Set 2 enabled

 

 

 

-

No / Yes

103

 

No

 

Set 3 enabled

 

 

 

-

No / Yes

104

 

No

 

Set 4 enabled

 

 

 

-

No / Yes

105

 

No

 

Automatic return on default set 1 after power up

 

# 101=Yes

-

No / Yes

 

 

 

 

 

 

 

 

 

 

 

201

 

1065

 

Correction at 1 ml/h

 

 

 

 

ratio

900 - 1100

202

 

1054

 

Correction at 3 ml/h

 

 

 

 

ratio

900 - 1100

203

 

1044

 

Correction at 5 ml/h

 

 

 

 

ratio

900 - 1100

204

 

1035

 

Correction at 10 ml/h

 

SET 1

 

ratio

900 - 1100

205

 

1027

 

Correction at 25 ml/h

 

Do not change

ratio

900 - 1100

206

 

1020

 

Correction at 50 ml/h

 

ratio

900 - 1100

207

 

1014

 

Correction at 75 ml/h

 

these values unless

ratio

900 - 1100

 

 

 

 

 

 

 

 

 

explicitly trained

 

 

 

208

 

1008

 

Correction at 100 ml/h

 

ratio

900 - 1100

 

 

 

 

 

 

209

 

999

 

Correction at 150 ml/h

 

 

 

 

ratio

900 -1100

210

 

988

 

Correction at 200 ml/h

 

 

 

 

ratio

900 - 1100

211

 

976

 

Correction at 250 ml/h

 

 

 

 

ratio

900 - 1100

212

 

967

 

Correction at 300 ml/h

 

 

 

 

ratio

900 - 1100

213

 

952

 

Correction at 400 ml/h

 

 

 

 

ratio

900 - 1100

214

 

942

 

Correction at 500 ml/h

 

 

 

 

ratio

900 - 1100

215

 

925

 

Correction at 750 ml/h

 

 

 

 

ratio

900 - 1100

216

 

910

 

Correction at 999 ml/h

 

 

 

 

ratio

900 - 1100

217

 

1000

 

Correction at 1200 ml/h

 

 

 

 

ratio

1000-1100

 

 

 

 

 

 

 

 

 

 

221

 

1065

 

Correction at 1 ml/h

 

 

 

 

ratio

900 - 1100

222

 

1054

 

Correction at 3 ml/h

 

 

 

 

ratio

900 - 1100

223

 

1044

 

Correction at 5 ml/h

 

 

 

 

ratio

900 - 1100

224

 

1035

 

Correction at 10 ml/h

 

SET 2

 

ratio

900 - 1100

225

 

1027

 

Correction at 25 ml/h

 

Do not change

ratio

900 - 1100

226

 

1020

 

Correction at 50 ml/h

 

ratio

900 - 1100

227

 

1014

 

Correction at 75 ml/h

 

these values unless

ratio

900 - 1100

 

 

 

 

 

 

 

 

 

explicitly trained

 

 

 

228

 

1008

 

Correction at 100 ml/h

 

ratio

900 - 1100

 

 

 

 

 

 

229

 

999

 

Correction at 150 ml/h

 

 

 

 

ratio

900 -1100

230

 

988

 

Correction at 200 ml/h

 

 

 

 

ratio

900 - 1100

231

 

976

 

Correction at 250 ml/h

 

 

 

 

ratio

900 - 1100

232

 

967

 

Correction at 300 ml/h

 

 

 

 

ratio

900 - 1100

233

 

952

 

Correction at 400 ml/h

 

 

 

 

ratio

900 - 1100

234

 

942

 

Correction at 500 ml/h

 

 

 

 

ratio

900 - 1100

235

 

925

 

Correction at 750 ml/h

 

 

 

 

ratio

900 - 1100

236

 

910

 

Correction at 999 ml/h

 

 

 

 

ratio

900 - 1100

237

 

1000

 

Correction at 1200 ml/h

 

 

 

 

ratio

1000-1100

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

14.203.B_A707 en.SM.V4.3X

ARGUS Medical AG 28.03.06 / PJ

 

 

11 / 48

 

PUMP CONFIGURATIONS

238

 

 

 

 

 

 

 

 

 

239

 

 

 

 

 

 

 

 

 

240

 

 

 

 

 

 

 

 

 

241

 

1065

 

Correction at 1 ml/h

 

ratio

900

- 1100

242

 

1054

 

Correction at 3 ml/h

 

ratio

900

- 1100

243

 

1044

 

Correction at 5 ml/h

 

ratio

900

- 1100

244

 

1035

 

Correction at 10 ml/h

SET 3

ratio

900

- 1100

245

 

1027

 

Correction at 25 ml/h

Do not change

ratio

900

- 1100

246

 

1020

 

Correction at 50 ml/h

ratio

900

- 1100

247

 

1014

 

Correction at 75 ml/h

these values unless

ratio

900

- 1100

 

 

 

 

 

 

explicitly trained

 

 

 

 

248

 

1008

 

Correction at 100 ml/h

ratio

900

- 1100

 

 

 

249

 

999

 

Correction at 150 ml/h

 

ratio

900

- 1100

250

 

988

 

Correction at 200 ml/h

 

ratio

900

- 1100

251

 

976

 

Correction at 250 ml/h

 

ratio

900

- 1100

252

 

967

 

Correction at 300 ml/h

 

ratio

900

- 1100

253

 

952

 

Correction at 400 ml/h

 

ratio

900

- 1100

254

 

942

 

Correction at 500 ml/h

 

ratio

900

- 1100

255

 

925

 

Correction at 750 ml/h

 

ratio

900

- 1100

256

 

910

 

Correction at 999 ml/h

 

ratio

900

- 1100

257

 

1000

 

Correction at 1200 ml/h

 

ratio

1000-1100

258

 

 

 

 

 

 

 

 

 

259

 

 

 

 

 

 

 

 

 

260

 

 

 

 

 

 

 

 

 

261

 

1065

 

Correction at 1 ml/h

 

ratio

900

- 1100

262

 

1054

 

Correction at 3 ml/h

 

ratio

900

- 1100

263

 

1044

 

Correction at 5 ml/h

 

ratio

900

- 1100

264

 

1035

 

Correction at 10 ml/h

SET 4

ratio

900

- 1100

265

 

1027

 

Correction at 25 ml/h

Do not change

ratio

900

- 1100

266

 

1020

 

Correction at 50 ml/h

ratio

900

- 1100

267

 

1014

 

Correction at 75 ml/h

these values unless

ratio

900

- 1100

 

 

 

 

 

 

explicitly trained

 

 

 

 

268

 

1008

 

Correction at 100 ml/h

ratio

900

- 1100

 

 

 

269

 

999

 

Correction at 150 ml/h

 

ratio

900

- 1100

270

 

988

 

Correction at 200 ml/h

 

ratio

900

- 1100

271

 

976

 

Correction at 250 ml/h

 

ratio

900

- 1100

272

 

967

 

Correction at 300 ml/h

 

ratio

900

- 1100

273

 

952

 

Correction at 400 ml/h

 

ratio

900

- 1100

274

 

942

 

Correction at 500 ml/h

 

ratio

900

- 1100

275

 

925

 

Correction at 750 ml/h

 

ratio

900

- 1100

276

 

910

 

Correction at 999 ml/h

 

ratio

900

- 1100

277

 

1000

 

Correction at 1200 ml/h

 

ratio

1000-1100

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

310

 

999.9

 

Max. flow rate

 

ml/h

1 - 999.9

311

 

999

 

Max. prime rate

 

ml/h

1 - 999

312

 

1200

 

Max. bolus rate

Set 1 parameters

ml/h

1 - 1200

313

 

10

 

Max. bolus total

ml

1 - 999

 

 

 

314

 

10

 

Upstream occlusion sensitivity

 

factor

0

- 60

315

 

20

 

Air summation volume

 

• 50 µl

2

- 40

316

103

7

 

Downstream default pressure limit PrL

 

• 100 mbar

1

- 10

317

200

250

 

Air bubble size

 

µl

50 - 1000

318

104

20

 

Drop rate window center

 

drops

10 - 65

319

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

14.203.B_A707 en.SM.V4.3X

ARGUS Medical AG 28.03.06 / PJ

 

12 / 48

 

PUMP CONFIGURATIONS

320

 

999.9

 

Max. flow rate

 

ml/h

1 - 999.9

321

 

999

 

Max. prime rate

 

ml/h

1 - 999

322

 

1200

 

Max. bolus rate

Set 2 parameters

ml/h

1 - 1200

323

 

10

 

Max. bolus total

ml

1 - 999

 

 

 

324

 

10

 

Upstream occlusion sensitivity

 

factor

0

- 60

325

 

20

 

Air summation volume

 

• 50 µl

2

- 40

326

 

7

 

Downstream default pressure limit PrL

 

• 100 mbar

1

- 10

327

 

250

 

Air bubble size

 

µl

50 - 1000

328

 

20

 

Drop rate window center

 

drops

10 - 65

329

 

 

 

 

 

 

 

 

 

330

 

999.9

 

Max. flow rate

 

ml/h

1 - 999.9

331

 

999

 

Max. prime rate

 

ml/h

1 - 999

332

 

1200

 

Max. bolus rate

Set 3 parameters

ml/h

1 - 1200

333

 

10

 

Max. bolus total

ml

1 - 999

 

 

 

334

 

10

 

Upstream occlusion sensitivity

 

factor

0

- 60

335

 

20

 

Air summation volume

 

• 50 µl

2

- 40

336

 

7

 

Downstream default pressure limit PrL

 

• 100 mbar

1

- 10

337

 

250

 

Air bubble size

 

µl

50 - 1000

338

 

20

 

Drop rate window center

 

drops

10 - 65

339

 

 

 

 

 

 

 

 

 

340

 

999.9

 

Max. flow rate

 

ml/h

1 - 999.9

341

 

999

 

Max. prime rate

 

ml/h

1 - 999

342

 

1200

 

Max. bolus rate

Set 4 parameters

ml/h

1 - 1200

343

 

10

 

Max. bolus total

ml

1 - 999

 

 

 

344

 

10

 

Upstream occlusion sensitivity

 

factor

0

- 60

345

 

20

 

Air summation volume

 

• 50 µl

2

- 40

346

 

7

 

Downstream default pressure limit PrL

 

• 100 mbar

1

- 10

347

 

250

 

Air bubble size

 

µl

50 - 1000

348

 

20

 

Drop rate window center

 

drops

10 - 65

349

 

 

 

 

 

 

 

 

 

350

 

 

 

 

 

 

 

 

 

351

 

4

 

Air volume summation time

 

• 8 min

1 - 8

 

 

 

 

 

 

 

 

 

361

100

5

 

Key ON/OFF delay time

# 4

• 1/10 s

0

- 31

362

101

8

 

Display brightness

 

level

2

- 15

363

102

10

 

Buzzer alarm volume

 

level

5

- 10

364

 

 

 

 

 

 

 

 

 

365

103

150

 

Pressure step for neonatology mode

# 39=Yes

mbar

60

- 250

366

 

1

 

Pressure display unit (mbar / mmHg / kPa

/ cmH2O / Psi)

Enum

1 - 5

367

 

 

 

 

 

 

 

 

 

368

200

300

 

Battery capacity (discharge time)

 

min

45

- 300

369

 

5

 

Automatic menu fall back delay time

 

sec

5

- 30

 

 

 

 

 

 

 

 

 

390

 

0

 

Last service date in year

 

year

0

- 99

391

 

0

 

Last service date in month

 

month

0

- 12

392

 

0

 

Last service date in day

 

day

0

- 31

393

 

0

 

Service interval in months

(0 = disabled)

month

0

- 24

394

 

0

 

Service interval in hours of operation

(0 = disabled)

hour

0 -10000

395

 

 

 

 

 

 

 

 

 

396

 

0

 

Pump serial number

 

xxx 8 yyy

xxx 8 yyy

397

 

 

 

 

 

 

 

 

 

398

 

 

 

 

 

 

 

 

 

-

399

707

 

Enter the calibration menu / clears protection key

-

123/4

 

 

 

 

 

 

 

 

 

 

 

 

14.203.B_A707 en.SM.V4.3X

ARGUS Medical AG 28.03.06 / PJ

 

13 / 48

 

PUMP CONFIGURATIONS

 

 

 

 

 

 

 

518

 

2

Permissions for serial communication

 

Enum

0 - 3

 

0 = none, 1 = query only, 2 = parametrising, 3 = remote control

 

 

 

 

 

522

 

No

Allow change of ml total while infusing

# 65=No

-

No / Yes

523

 

15

Alarm lamp brightness

 

level

2 - 15

524

 

 

Display a department info text (after power up)

char

1-16 ASCII

Using the “ARGUS service” tool, the complete and detailed pump configuration can be done.

2.8.Special configuration options (with ARGUS service)

Enabling a new IV set

Follow the steps below to enable and configure a new IV set:

a)write “Yes” in the corresponding index 101, 102, 103, 104

b)make sure the set definitions (index 201 to 277) correspond to the IV sets used. In case of doubt, contact ARGUS Medical AG or your local dealer.

c)configure the set parameters (index 310 to 349 and 351)

d)perform a complete set calibration as per chapters 6.3. to 6.5.

14.203.B_A707 en.SM.V4.3X

ARGUS Medical AG 28.03.06 / PJ

14 / 48

Loading...
+ 34 hidden pages