Service Manual for Volumetric Infusion Pump ARGUS 707 V
Made in Switzerland
ARGUS Medical AG, CH-3627 Heimberg / Switzerland
(a member of the CODAN group)
14.203.B_A707 en.SM.V4.3X |
ARGUS Medical AG 28.03.06 / PJ |
1 / 48 |
|
|
TABLE OF CONTENTS |
|
1. |
INTRODUCTION ........................................................................................................... |
4 |
|
|
1.1. |
General ............................................................................................................................. |
4 |
2. |
PUMP CONFIGURATIONS ........................................................................................... |
4 |
|
|
2.1. |
General ............................................................................................................................. |
4 |
|
2.2. Interrogation mode (without ARGUS service)................................................................... |
5 |
|
|
2.3. Configuration mode (without ARGUS service).................................................................. |
6 |
|
|
2.4. First activation of a PIN code (write protection) ................................................................ |
7 |
|
|
2.5. Changing an existing PIN code......................................................................................... |
8 |
|
|
2.6. Address list of the pump configuration (without ARGUS service) ..................................... |
9 |
|
|
2.7. Index list of the pump configuration (with ARGUS service) ............................................ |
10 |
|
|
2.8. Special configuration options (with ARGUS service) ...................................................... |
14 |
|
|
2.9. |
Medication list ................................................................................................................. |
15 |
3. |
Serial communication of the pump .......................................................................... |
18 |
|
|
3.1. |
General ........................................................................................................................... |
18 |
|
3.2. |
Serial communication protocol........................................................................................ |
18 |
4. |
ARGUS service........................................................................................................... |
19 |
|
|
4.1. |
General ........................................................................................................................... |
19 |
|
4.2. ARGUS service – Configuration...................................................................................... |
20 |
|
|
4.3. ARGUS service - Calibration (volumetric pums only) ..................................................... |
21 |
|
|
4.4. |
ARGUS service - Toolbox............................................................................................... |
22 |
5. |
SOFTWARE UPDATES .............................................................................................. |
26 |
|
|
5.1. |
General ........................................................................................................................... |
26 |
|
5.2. Requirements for a software update............................................................................... |
26 |
|
|
5.3. |
Software update procedure............................................................................................. |
26 |
|
5.4. |
Safety aspects................................................................................................................. |
28 |
6. |
MAINTENANCE .......................................................................................................... |
29 |
|
|
6.1. |
General ........................................................................................................................... |
29 |
|
6.2. |
Recalibration ................................................................................................................... |
29 |
|
6.3. |
Pressure calibration ........................................................................................................ |
30 |
|
6.4. |
Pressure control measurement ....................................................................................... |
35 |
|
6.5. |
Volume calibration .......................................................................................................... |
36 |
|
6.6. |
Pump specifications ........................................................................................................ |
37 |
|
6.7. |
Fault codes ..................................................................................................................... |
38 |
7. |
REPLACEMENT OF PARTS....................................................................................... |
39 |
|
|
7.1. |
General ........................................................................................................................... |
39 |
|
7.2. Disassembling of the ARGUS 707 V............................................................................... |
39 |
|
|
7.3. |
Spare parts ..................................................................................................................... |
43 |
8. |
WIRING DIAGRAMM................................................................................................... |
46 |
|
9. |
BLOC SCHEMATIC .................................................................................................... |
46 |
|
10. SAFETY STANDARD CHECK .................................................................................... |
47 |
||
11. REPAIR ORDER FORM.............................................................................................. |
48 |
14.203.B_A707 en.SM.V4.3X |
ARGUS Medical AG 28.03.06 / PJ |
1 / 48 |
INTRODUCTION
Park area for drop detector
Display
Drop detector
Drop chamber
Global alarm
Air bubble detector
Pressure sensor |
automatic |
Pressure sensor |
„upstream“ |
„free-flow“-clamp |
„downstream“ |
14.203.B_A707 en.SM.V4.3X |
ARGUS Medical AG 28.03.06 / PJ |
2 / 48 |
INTRODUCTION
Universal clamp |
Staff alert and |
|
RS-232-connection |
Drop detector connection
AC power |
Identification plate |
|
with pump serial |
||
connection |
||
number |
||
|
DC-connection & QUICK®100
Docking Station interface
14.203.B_A707 en.SM.V4.3X |
ARGUS Medical AG 28.03.06 / PJ |
3 / 48 |
INTRODUCTION
1.INTRODUCTION
1.1.General
IMPORTANT!
This service manual is intended for the exclusive use of authorized persons who have been trained by ARGUS Medical AG in the maintenance and repair of the
ARGUS 707 V infusion pump.
The service manual is meant to be used together with the user manual.
IMPORTANT!
ARGUS Medical AG shall not assume any responsibility for any manipulations which have been carried out on the ARGUS 707 V infusion pump by a nonauthorized person.
CAUTION!
The ARGUS 707 V infusion pump may only be used with spare parts, accessories, consumables and IV-sets with Luer-Lock connections recommended by ARGUS Medical AG. The functional safety of the pump is not guaranteed if non approved materials are used. The safety of the patient may be endangered.
This manual contains the latest data available. It is subject to further modifications in accordance with technical improvements.
2.PUMP CONFIGURATIONS
2.1.General
CAUTION!
The configuration possibilities with the “ARGUS service” PC utility tool and without
PC assistance constitute a modification of the pump and may only be carried out by authorized persons!
CAUTION!
After changing the configuration a function check and a control measurement has to be performed!
14.203.B_A707 en.SM.V4.3X |
ARGUS Medical AG 28.03.06 / PJ |
4 / 48 |
PUMP CONFIGURATIONS
2.2.Interrogation mode (without ARGUS service)
With the interrogation mode you can read the present keypad configuration of the pump without the possibility to modify any configurations. For a complete overview, please take the “ARGUS service” PC-tool!
1.To enter into the interrogation mode, switch the pump on while keeping the keys “MODE” and “START/STOP” pressed.
707 PRog
...5. 1
2. Press the “START/ STOP” key.
4.To quit the interrogation mode press the “ON/OFF” key at least 2 seconds.
3.Enter the requested address (see chapter 2.6), e.g. 5. The programmed value of the address 5 appears in the right hand display.
Flashing decimal points indicate which display is ready to accept an input by the keys 100, 10 & 1.
In the interrogation mode the left hand display shows the address and the right hand display shows the according value configured at this address. Please refer to chapter 2.6. where the meanings of the addresses are explained.
To modify any configuration data you have to go into the configuration mode.
14.203.B_A707 en.SM.V4.3X |
ARGUS Medical AG 28.03.06 / PJ |
5 / 48 |
PUMP CONFIGURATIONS
2.3.Configuration mode (without ARGUS service)
The configuration mode permit you to modify the pump keypad configuration manually using the keypad. Please refer to chapter 2.6. where the meaning of the adresses are explained. To have access to all configuration options, please use the
“ARGUS service” PC-tool!
2.3.1.Step 1
1.To enter into the configuration mode, switch the pump on while keeping the keys “MODE” and “START/STOP” pressed.
2.Press key “START”
707 prog
|
4. Press key “START” again. |
3. Press key “MODE” |
The display will show “Cod ----“ |
|
2.3.2. Step 2
5.Enter the PIN Code. If no PIN code was configured, no entry is required
|
6. Press key “START” |
Cod ---- |
to acknowledge the |
|
entered PIN code |
If the PIN was not accepted, the display will change back to point 5.
You have now access to all addresses in the list of chapter 2.6.
Select therefore any address in the left display (see next page).
14.203.B_A707 en.SM.V4.3X |
ARGUS Medical AG 28.03.06 / PJ |
6 / 48 |
|
|
|
|
PUMP CONFIGURATIONS |
|
2.3.3. Step 3 |
|
|
|||
|
|
|
|
|
|
8. Enter now the value on the right hand display. The |
|
|
7. After entering an address e.g. 5 press the |
|
|
|
|
||||
range of the value is given by the table in chap.2.6 |
key “MODE”. The flashing decimal points |
|
|||
|
|
|
|
will change to the right hand display. |
|
|
|
|
|
|
|
...5. 0
9. Press key “START” to acknowledge the value entered. The flashing decimal points change back to the left hand display.
10.To quit the configuration mode press the “ON/OFF” key at least 2 seconds. Changes in configuration become active,
after the pump is switched on normally again.
Important remark:
Invalid values entered will be corrected automatically by the pump to the maxima or minima value allowed for the according address!
2.4.First activation of a PIN code (write protection)
The activation of a PIN code allows you to protect the configuration from unauthorized access. To activate the PIN code, enter into the configuration mode.
1.To enter into the configuration mode, switch the pump on while keeping the keys “MODE” and “START/STOP” pressed.
2.Press key “START”
707 |
PRog |
|
||
|
|
|
|
4. Press key “START” |
|
|
|
|
|
|
|
|
|
again. The display will |
|
|
|
|
show “Cod ----“. |
|
|
|
||
|
|
|
|
|
5. Press “MODE” key (Code “0” will be set) then “START” |
|
|||
key to acknowledge the entered PIN code “0”. |
3. Press key “MODE” |
|||
|
|
|
|
|
14.203.B_A707 en.SM.V4.3X |
ARGUS Medical AG 28.03.06 / PJ |
7 / 48 |
PUMP CONFIGURATIONS
|
|
|
|
|
|
|
6. |
Press the “MODE” key. |
|
|
|
|
|
|
|
||
|
|
|
|
|
|
|
|
The flashing decimal |
0 |
...0. |
|
|
|
points will change to the |
|||
|
|
|
right hand display |
|||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
8. |
Press key “START” to |
|||
|
|
|
|
|||||
|
|
|
|
|
|
|
|
acknowledge to new PIN. |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
7.Enter now the new PIN Code (max. 4 digit number).
Please remember this code, it will never be visible again! e.g.
CAUTION!
After you switch the pump OFF and ON again you can enter into the configuration mode only, if you enter the correct PIN code.
e.g.
Please note: The interrogation mode can always be accessed without the PIN.
2.5.Changing an existing PIN code
Enter the configuration mode using present PIN, select add. “0” and set the new code.
Enter actual PIN code and confirm with “START” key.
Press “MODE” key (#0). The flashing decimal points will change to the right hand display.
Enter the new PIN code and press the “START/STOP” key to acknowledge the entered code.
14.203.B_A707 en.SM.V4.3X |
ARGUS Medical AG 28.03.06 / PJ |
8 / 48 |
PUMP CONFIGURATIONS
2.6.Address list of the pump configuration (without ARGUS service)
The following list declares the possible configuration options which can be performed on the pump keypad without using the PC.
All these options can also be configured by the PC-Software “ARGUS service”.
Address |
Index |
Default |
Function |
Unit |
Range |
left display |
PC |
|
|
|
right display |
|
|
|
|
|
|
1 |
2 |
No |
Key ON/OFF only at stop valid |
- |
0=No / 1=Yes |
2 |
10 |
No |
Operation without drop detector |
- |
0=No / 1=Yes |
3 |
11 |
No |
Recall of the last used ml/h rate |
- |
0=No / 1=Yes |
4 |
19 |
No |
Buzzer at start |
- |
0=No / 1=Yes |
5 |
44 |
Yes |
Automatic pressure release |
- |
0=No / 1=Yes |
6 |
49 |
No |
Alarm acknowledge with key MODE |
- |
0=No / 1=Yes |
|
|
|
|
|
|
|
|
|
|
|
|
100 |
361 |
5 |
Key ON/OFF delay time |
• 0.1 s |
0 - 31 |
101 |
362 |
8 |
Display brightness |
level |
2 - 15 |
102 |
363 |
7 |
Buzzer volume |
level |
5 - 10 |
103 |
316 |
9 |
Default pressure limit (IV-set 1) |
• 100 mbar |
1 - 10 |
104 |
318 |
20 |
Drop rate window centre (IV-set 1) |
Drops/ml |
10 - 65 |
|
|
|
|
|
|
|
|
|
|
|
|
200 |
317 |
250 |
Air bubble size (IV-set 1) |
µl |
50 - 1000 |
|
|
|
|
|
|
|
|
|
|
|
|
399 |
- |
707 |
Enter the calibration menu (volume) |
- |
123 |
|
- |
707 |
Enter the calibration menu (pressure) |
- |
1234 |
Note!
The address does not correspond with the index used by the “ARGUS service” tool.
14.203.B_A707 en.SM.V4.3X |
ARGUS Medical AG 28.03.06 / PJ |
9 / 48 |
PUMP CONFIGURATIONS
2.7. Index list of the pump configuration (with ARGUS service)
Index PC |
Add. |
Default |
|
Function |
|
Unit |
Range |
|||
|
|
|
|
|
|
|
|
|||
1 |
|
Yes |
|
Run indication by running decimal point |
|
- |
No / Yes |
|||
2 |
1 |
No |
|
Key ON/OFF only at stop valid |
|
- |
No / Yes |
|||
3 |
|
No |
|
Rate change allowed only at STOP |
|
- |
No / Yes |
|||
4 |
|
No |
|
Key STOP delayed |
|
# 361 |
- |
No / Yes |
||
5 |
|
No |
|
Second entry of rate |
|
# 3=Yes |
- |
No / Yes |
||
6 |
|
No |
|
Static alarm (staff alerting system) |
|
|
- |
No / Yes |
||
7 |
|
No |
|
Display elapsed time in run mode |
|
# 8=No |
- |
No / Yes |
||
8 |
|
No |
|
Display remaining time |
|
# 7=Yes |
- |
No / Yes |
||
9 |
|
|
|
|
|
|
|
|
|
|
10 |
2 |
No |
|
Operation without drop detector |
|
- |
No / Yes |
|||
11 |
3 |
Yes |
|
Recall of ml/h rate at next power on |
|
- |
No / Yes |
|||
12 |
|
No |
|
Recall of ml total at next power on |
|
- |
No / Yes |
|||
13 |
|
No |
|
Recall of ml inf. at next power on |
|
- |
No / Yes |
|||
14 |
|
No |
|
SBS (step by step function) |
|
- |
No / Yes |
|||
15 |
|
No |
|
Display VTBI (volume to be infused) |
|
- |
No / Yes |
|||
16 |
|
No |
|
Display “Set-X” if only one set is enabled |
|
- |
No / Yes |
|||
17 |
|
Yes |
|
KVO (KOR) enabled |
|
# 60 |
- |
No / Yes |
||
18 |
|
No |
|
Drop alarm only if bottle empty |
|
# 10=No, → trA |
- |
No / Yes |
||
19 |
4 |
No |
|
Buzzer at start |
|
|
- |
No / Yes |
||
20 |
|
No |
|
Menu Clr (clear ml inf.) |
|
# 15=No |
- |
No / Yes |
||
21 |
|
|
|
|
|
|
|
|
|
|
22 |
|
No |
|
Menu trA (transport) |
|
# 10, 18=No |
- |
No / Yes |
||
23 |
|
Yes |
|
Menu PrL (pressure alarm limit) |
|
|
- |
No / Yes |
||
24 |
|
Yes |
|
Menu CAP (battery capacity) |
|
- |
No / Yes |
|||
25 |
|
No |
|
Menu SEt Fill (prime IV set) |
|
|
|
|
||
26 |
|
No |
|
Menu InF (ml inf. since last power on) |
|
- |
No / Yes |
|||
27 |
|
No |
|
Menu dLo (data lock) |
|
- |
No / Yes |
|||
28 |
|
No |
|
Menu Stb (stand by) |
|
- |
No / Yes |
|||
29 |
|
No |
|
Menu Med (medication name) |
|
- |
No / Yes |
|||
30 |
|
No |
|
Menu tM (timer alarm) |
|
- |
No / Yes |
|||
31 |
|
|
|
|
|
|
|
|
|
|
32 |
|
No |
|
Menu boL (bolus release) |
|
- |
No / Yes |
|||
33 |
|
No |
|
Menu boLr (bolus rate) |
|
# 32=Yes |
- |
No / Yes |
||
34 |
|
No |
|
Menu tot (bolus total) |
|
# 32=Yes |
- |
No / Yes |
||
35 |
|
|
|
|
|
|
|
|
|
|
36 |
|
|
|
|
|
|
|
|
|
|
37 |
|
|
|
|
|
|
|
|
|
|
38 |
|
No |
|
Automatic bolus application |
|
# 32, 34=Yes |
- |
No / Yes |
||
39 |
|
No |
|
Neonatology mode accessible in PrL menu |
|
- |
No / Yes |
|||
40 |
|
|
|
|
|
|
|
|
|
|
41 |
|
No |
|
Clear ml/h after infusion completed |
|
- |
No / Yes |
|||
42 |
|
No |
|
Clear ml total after infusion completed |
|
# 41=Yes |
- |
No / Yes |
||
43 |
|
No |
|
Air volume accumulated |
|
|
- |
No / Yes |
||
44 |
5 |
Yes |
|
Automatic pressure release after occlusion |
|
- |
No / Yes |
|||
45 |
|
Yes |
|
Pressure display ON (LED bar graph - 20/40/60/80/100%) |
- |
No / Yes |
||||
46 |
|
No |
|
Pressure display with indicator |
|
# 45=Yes |
- |
No / Yes |
||
47 |
|
No |
|
Stand byand battery pre alarm low volume |
- |
No / Yes |
||||
48 |
|
No |
|
Flashing numeric display at alarm |
|
- |
No / Yes |
|||
49 |
6 |
No |
|
Alarm acknowledge only with key MODE |
|
- |
No / Yes |
|||
|
|
|
|
|
|
|
|
|
|
|
|
14.203.B_A707 en.SM.V4.3X |
ARGUS Medical AG 28.03.06 / PJ |
|
10 / 48 |
|
PUMP CONFIGURATIONS
50 |
|
No |
|
Start with ≥ 1 bar pressure allowed |
|
|
|
- |
No / Yes |
|||||
51 |
|
Yes |
|
Enable upstream occlusion detection |
|
|
|
- |
No / Yes |
|||||
52 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
53 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
54 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
55 |
|
No |
|
Select RJ-45 connector for serial interface |
# 29=Yes |
- |
No / Yes |
|||||||
56 |
|
Yes |
|
Medication display alternate |
|
|
- |
No / Yes |
||||||
|
|
|
|
|
|
|
|
|
|
|
|
|||
60 |
|
No |
|
KVO only after infusion completed |
|
|
|
- |
No / Yes |
|||||
|
|
|
|
|
|
|
|
|
|
|
|
|||
65 |
|
No |
|
Clear and continue |
|
|
# 15=No |
|
- |
No / Yes |
||||
|
|
|
|
|
|
|
|
|
|
|
||||
101 |
|
Yes |
|
Set 1 enabled |
|
|
|
- |
No / Yes |
|||||
102 |
|
No |
|
Set 2 enabled |
|
|
|
- |
No / Yes |
|||||
103 |
|
No |
|
Set 3 enabled |
|
|
|
- |
No / Yes |
|||||
104 |
|
No |
|
Set 4 enabled |
|
|
|
- |
No / Yes |
|||||
105 |
|
No |
|
Automatic return on default set 1 after power up |
|
# 101=Yes |
- |
No / Yes |
||||||
|
|
|
|
|
|
|
|
|
|
|
||||
201 |
|
1065 |
|
Correction at 1 ml/h |
|
|
|
|
ratio |
900 - 1100 |
||||
202 |
|
1054 |
|
Correction at 3 ml/h |
|
|
|
|
ratio |
900 - 1100 |
||||
203 |
|
1044 |
|
Correction at 5 ml/h |
|
|
|
|
ratio |
900 - 1100 |
||||
204 |
|
1035 |
|
Correction at 10 ml/h |
|
SET 1 |
|
ratio |
900 - 1100 |
|||||
205 |
|
1027 |
|
Correction at 25 ml/h |
|
Do not change |
ratio |
900 - 1100 |
||||||
206 |
|
1020 |
|
Correction at 50 ml/h |
|
ratio |
900 - 1100 |
|||||||
207 |
|
1014 |
|
Correction at 75 ml/h |
|
these values unless |
ratio |
900 - 1100 |
||||||
|
|
|
|
|
|
|
|
|
explicitly trained |
|
|
|
||
208 |
|
1008 |
|
Correction at 100 ml/h |
|
ratio |
900 - 1100 |
|||||||
|
|
|
|
|
|
|||||||||
209 |
|
999 |
|
Correction at 150 ml/h |
|
|
|
|
ratio |
900 -1100 |
||||
210 |
|
988 |
|
Correction at 200 ml/h |
|
|
|
|
ratio |
900 - 1100 |
||||
211 |
|
976 |
|
Correction at 250 ml/h |
|
|
|
|
ratio |
900 - 1100 |
||||
212 |
|
967 |
|
Correction at 300 ml/h |
|
|
|
|
ratio |
900 - 1100 |
||||
213 |
|
952 |
|
Correction at 400 ml/h |
|
|
|
|
ratio |
900 - 1100 |
||||
214 |
|
942 |
|
Correction at 500 ml/h |
|
|
|
|
ratio |
900 - 1100 |
||||
215 |
|
925 |
|
Correction at 750 ml/h |
|
|
|
|
ratio |
900 - 1100 |
||||
216 |
|
910 |
|
Correction at 999 ml/h |
|
|
|
|
ratio |
900 - 1100 |
||||
217 |
|
1000 |
|
Correction at 1200 ml/h |
|
|
|
|
ratio |
1000-1100 |
||||
|
|
|
|
|
|
|
|
|
|
|||||
221 |
|
1065 |
|
Correction at 1 ml/h |
|
|
|
|
ratio |
900 - 1100 |
||||
222 |
|
1054 |
|
Correction at 3 ml/h |
|
|
|
|
ratio |
900 - 1100 |
||||
223 |
|
1044 |
|
Correction at 5 ml/h |
|
|
|
|
ratio |
900 - 1100 |
||||
224 |
|
1035 |
|
Correction at 10 ml/h |
|
SET 2 |
|
ratio |
900 - 1100 |
|||||
225 |
|
1027 |
|
Correction at 25 ml/h |
|
Do not change |
ratio |
900 - 1100 |
||||||
226 |
|
1020 |
|
Correction at 50 ml/h |
|
ratio |
900 - 1100 |
|||||||
227 |
|
1014 |
|
Correction at 75 ml/h |
|
these values unless |
ratio |
900 - 1100 |
||||||
|
|
|
|
|
|
|
|
|
explicitly trained |
|
|
|
||
228 |
|
1008 |
|
Correction at 100 ml/h |
|
ratio |
900 - 1100 |
|||||||
|
|
|
|
|
|
|||||||||
229 |
|
999 |
|
Correction at 150 ml/h |
|
|
|
|
ratio |
900 -1100 |
||||
230 |
|
988 |
|
Correction at 200 ml/h |
|
|
|
|
ratio |
900 - 1100 |
||||
231 |
|
976 |
|
Correction at 250 ml/h |
|
|
|
|
ratio |
900 - 1100 |
||||
232 |
|
967 |
|
Correction at 300 ml/h |
|
|
|
|
ratio |
900 - 1100 |
||||
233 |
|
952 |
|
Correction at 400 ml/h |
|
|
|
|
ratio |
900 - 1100 |
||||
234 |
|
942 |
|
Correction at 500 ml/h |
|
|
|
|
ratio |
900 - 1100 |
||||
235 |
|
925 |
|
Correction at 750 ml/h |
|
|
|
|
ratio |
900 - 1100 |
||||
236 |
|
910 |
|
Correction at 999 ml/h |
|
|
|
|
ratio |
900 - 1100 |
||||
237 |
|
1000 |
|
Correction at 1200 ml/h |
|
|
|
|
ratio |
1000-1100 |
||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
14.203.B_A707 en.SM.V4.3X |
ARGUS Medical AG 28.03.06 / PJ |
|
|
11 / 48 |
|
PUMP CONFIGURATIONS
238 |
|
|
|
|
|
|
|
|
|
|
239 |
|
|
|
|
|
|
|
|
|
|
240 |
|
|
|
|
|
|
|
|
|
|
241 |
|
1065 |
|
Correction at 1 ml/h |
|
ratio |
900 |
- 1100 |
||
242 |
|
1054 |
|
Correction at 3 ml/h |
|
ratio |
900 |
- 1100 |
||
243 |
|
1044 |
|
Correction at 5 ml/h |
|
ratio |
900 |
- 1100 |
||
244 |
|
1035 |
|
Correction at 10 ml/h |
SET 3 |
ratio |
900 |
- 1100 |
||
245 |
|
1027 |
|
Correction at 25 ml/h |
Do not change |
ratio |
900 |
- 1100 |
||
246 |
|
1020 |
|
Correction at 50 ml/h |
ratio |
900 |
- 1100 |
|||
247 |
|
1014 |
|
Correction at 75 ml/h |
these values unless |
ratio |
900 |
- 1100 |
||
|
|
|
|
|
|
explicitly trained |
|
|
|
|
248 |
|
1008 |
|
Correction at 100 ml/h |
ratio |
900 |
- 1100 |
|||
|
|
|
||||||||
249 |
|
999 |
|
Correction at 150 ml/h |
|
ratio |
900 |
- 1100 |
||
250 |
|
988 |
|
Correction at 200 ml/h |
|
ratio |
900 |
- 1100 |
||
251 |
|
976 |
|
Correction at 250 ml/h |
|
ratio |
900 |
- 1100 |
||
252 |
|
967 |
|
Correction at 300 ml/h |
|
ratio |
900 |
- 1100 |
||
253 |
|
952 |
|
Correction at 400 ml/h |
|
ratio |
900 |
- 1100 |
||
254 |
|
942 |
|
Correction at 500 ml/h |
|
ratio |
900 |
- 1100 |
||
255 |
|
925 |
|
Correction at 750 ml/h |
|
ratio |
900 |
- 1100 |
||
256 |
|
910 |
|
Correction at 999 ml/h |
|
ratio |
900 |
- 1100 |
||
257 |
|
1000 |
|
Correction at 1200 ml/h |
|
ratio |
1000-1100 |
|||
258 |
|
|
|
|
|
|
|
|
|
|
259 |
|
|
|
|
|
|
|
|
|
|
260 |
|
|
|
|
|
|
|
|
|
|
261 |
|
1065 |
|
Correction at 1 ml/h |
|
ratio |
900 |
- 1100 |
||
262 |
|
1054 |
|
Correction at 3 ml/h |
|
ratio |
900 |
- 1100 |
||
263 |
|
1044 |
|
Correction at 5 ml/h |
|
ratio |
900 |
- 1100 |
||
264 |
|
1035 |
|
Correction at 10 ml/h |
SET 4 |
ratio |
900 |
- 1100 |
||
265 |
|
1027 |
|
Correction at 25 ml/h |
Do not change |
ratio |
900 |
- 1100 |
||
266 |
|
1020 |
|
Correction at 50 ml/h |
ratio |
900 |
- 1100 |
|||
267 |
|
1014 |
|
Correction at 75 ml/h |
these values unless |
ratio |
900 |
- 1100 |
||
|
|
|
|
|
|
explicitly trained |
|
|
|
|
268 |
|
1008 |
|
Correction at 100 ml/h |
ratio |
900 |
- 1100 |
|||
|
|
|
||||||||
269 |
|
999 |
|
Correction at 150 ml/h |
|
ratio |
900 |
- 1100 |
||
270 |
|
988 |
|
Correction at 200 ml/h |
|
ratio |
900 |
- 1100 |
||
271 |
|
976 |
|
Correction at 250 ml/h |
|
ratio |
900 |
- 1100 |
||
272 |
|
967 |
|
Correction at 300 ml/h |
|
ratio |
900 |
- 1100 |
||
273 |
|
952 |
|
Correction at 400 ml/h |
|
ratio |
900 |
- 1100 |
||
274 |
|
942 |
|
Correction at 500 ml/h |
|
ratio |
900 |
- 1100 |
||
275 |
|
925 |
|
Correction at 750 ml/h |
|
ratio |
900 |
- 1100 |
||
276 |
|
910 |
|
Correction at 999 ml/h |
|
ratio |
900 |
- 1100 |
||
277 |
|
1000 |
|
Correction at 1200 ml/h |
|
ratio |
1000-1100 |
|||
|
|
|
|
|
|
|
|
|||
|
|
|
|
|
|
|
|
|||
|
|
|
|
|
|
|
|
|||
310 |
|
999.9 |
|
Max. flow rate |
|
ml/h |
1 - 999.9 |
|||
311 |
|
999 |
|
Max. prime rate |
|
ml/h |
1 - 999 |
|||
312 |
|
1200 |
|
Max. bolus rate |
Set 1 parameters |
ml/h |
1 - 1200 |
|||
313 |
|
10 |
|
Max. bolus total |
ml |
1 - 999 |
||||
|
|
|
||||||||
314 |
|
10 |
|
Upstream occlusion sensitivity |
|
factor |
0 |
- 60 |
||
315 |
|
20 |
|
Air summation volume |
|
• 50 µl |
2 |
- 40 |
||
316 |
103 |
7 |
|
Downstream default pressure limit PrL |
|
• 100 mbar |
1 |
- 10 |
||
317 |
200 |
250 |
|
Air bubble size |
|
µl |
50 - 1000 |
|||
318 |
104 |
20 |
|
Drop rate window center |
|
drops |
10 - 65 |
|||
319 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
14.203.B_A707 en.SM.V4.3X |
ARGUS Medical AG 28.03.06 / PJ |
|
12 / 48 |
|
PUMP CONFIGURATIONS
320 |
|
999.9 |
|
Max. flow rate |
|
ml/h |
1 - 999.9 |
|||
321 |
|
999 |
|
Max. prime rate |
|
ml/h |
1 - 999 |
|||
322 |
|
1200 |
|
Max. bolus rate |
Set 2 parameters |
ml/h |
1 - 1200 |
|||
323 |
|
10 |
|
Max. bolus total |
ml |
1 - 999 |
||||
|
|
|
||||||||
324 |
|
10 |
|
Upstream occlusion sensitivity |
|
factor |
0 |
- 60 |
||
325 |
|
20 |
|
Air summation volume |
|
• 50 µl |
2 |
- 40 |
||
326 |
|
7 |
|
Downstream default pressure limit PrL |
|
• 100 mbar |
1 |
- 10 |
||
327 |
|
250 |
|
Air bubble size |
|
µl |
50 - 1000 |
|||
328 |
|
20 |
|
Drop rate window center |
|
drops |
10 - 65 |
|||
329 |
|
|
|
|
|
|
|
|
|
|
330 |
|
999.9 |
|
Max. flow rate |
|
ml/h |
1 - 999.9 |
|||
331 |
|
999 |
|
Max. prime rate |
|
ml/h |
1 - 999 |
|||
332 |
|
1200 |
|
Max. bolus rate |
Set 3 parameters |
ml/h |
1 - 1200 |
|||
333 |
|
10 |
|
Max. bolus total |
ml |
1 - 999 |
||||
|
|
|
||||||||
334 |
|
10 |
|
Upstream occlusion sensitivity |
|
factor |
0 |
- 60 |
||
335 |
|
20 |
|
Air summation volume |
|
• 50 µl |
2 |
- 40 |
||
336 |
|
7 |
|
Downstream default pressure limit PrL |
|
• 100 mbar |
1 |
- 10 |
||
337 |
|
250 |
|
Air bubble size |
|
µl |
50 - 1000 |
|||
338 |
|
20 |
|
Drop rate window center |
|
drops |
10 - 65 |
|||
339 |
|
|
|
|
|
|
|
|
|
|
340 |
|
999.9 |
|
Max. flow rate |
|
ml/h |
1 - 999.9 |
|||
341 |
|
999 |
|
Max. prime rate |
|
ml/h |
1 - 999 |
|||
342 |
|
1200 |
|
Max. bolus rate |
Set 4 parameters |
ml/h |
1 - 1200 |
|||
343 |
|
10 |
|
Max. bolus total |
ml |
1 - 999 |
||||
|
|
|
||||||||
344 |
|
10 |
|
Upstream occlusion sensitivity |
|
factor |
0 |
- 60 |
||
345 |
|
20 |
|
Air summation volume |
|
• 50 µl |
2 |
- 40 |
||
346 |
|
7 |
|
Downstream default pressure limit PrL |
|
• 100 mbar |
1 |
- 10 |
||
347 |
|
250 |
|
Air bubble size |
|
µl |
50 - 1000 |
|||
348 |
|
20 |
|
Drop rate window center |
|
drops |
10 - 65 |
|||
349 |
|
|
|
|
|
|
|
|
|
|
350 |
|
|
|
|
|
|
|
|
|
|
351 |
|
4 |
|
Air volume summation time |
|
• 8 min |
1 - 8 |
|||
|
|
|
|
|
|
|
|
|
||
361 |
100 |
5 |
|
Key ON/OFF delay time |
# 4 |
• 1/10 s |
0 |
- 31 |
||
362 |
101 |
8 |
|
Display brightness |
|
level |
2 |
- 15 |
||
363 |
102 |
10 |
|
Buzzer alarm volume |
|
level |
5 |
- 10 |
||
364 |
|
|
|
|
|
|
|
|
|
|
365 |
103 |
150 |
|
Pressure step for neonatology mode |
# 39=Yes |
mbar |
60 |
- 250 |
||
366 |
|
1 |
|
Pressure display unit (mbar / mmHg / kPa |
/ cmH2O / Psi) |
Enum |
1 - 5 |
|||
367 |
|
|
|
|
|
|
|
|
|
|
368 |
200 |
300 |
|
Battery capacity (discharge time) |
|
min |
45 |
- 300 |
||
369 |
|
5 |
|
Automatic menu fall back delay time |
|
sec |
5 |
- 30 |
||
|
|
|
|
|
|
|
|
|
||
390 |
|
0 |
|
Last service date in year |
|
year |
0 |
- 99 |
||
391 |
|
0 |
|
Last service date in month |
|
month |
0 |
- 12 |
||
392 |
|
0 |
|
Last service date in day |
|
day |
0 |
- 31 |
||
393 |
|
0 |
|
Service interval in months |
(0 = disabled) |
month |
0 |
- 24 |
||
394 |
|
0 |
|
Service interval in hours of operation |
(0 = disabled) |
hour |
0 -10000 |
|||
395 |
|
|
|
|
|
|
|
|
|
|
396 |
|
0 |
|
Pump serial number |
|
xxx 8 yyy |
xxx 8 yyy |
|||
397 |
|
|
|
|
|
|
|
|
|
|
398 |
|
|
|
|
|
|
|
|
|
|
- |
399 |
707 |
|
Enter the calibration menu / clears protection key |
- |
123/4 |
||||
|
|
|
|
|
|
|
|
|
|
|
|
14.203.B_A707 en.SM.V4.3X |
ARGUS Medical AG 28.03.06 / PJ |
|
13 / 48 |
|
PUMP CONFIGURATIONS
|
|
|
|
|
|
|
518 |
|
2 |
Permissions for serial communication |
|
Enum |
0 - 3 |
|
0 = none, 1 = query only, 2 = parametrising, 3 = remote control |
|||||
|
|
|
|
|
||
522 |
|
No |
Allow change of ml total while infusing |
# 65=No |
- |
No / Yes |
523 |
|
15 |
Alarm lamp brightness |
|
level |
2 - 15 |
524 |
|
|
Display a department info text (after power up) |
char |
1-16 ASCII |
Using the “ARGUS service” tool, the complete and detailed pump configuration can be done.
2.8.Special configuration options (with ARGUS service)
Enabling a new IV set
Follow the steps below to enable and configure a new IV set:
a)write “Yes” in the corresponding index 101, 102, 103, 104
b)make sure the set definitions (index 201 to 277) correspond to the IV sets used. In case of doubt, contact ARGUS Medical AG or your local dealer.
c)configure the set parameters (index 310 to 349 and 351)
d)perform a complete set calibration as per chapters 6.3. to 6.5.
14.203.B_A707 en.SM.V4.3X |
ARGUS Medical AG 28.03.06 / PJ |
14 / 48 |