Argus 707V User manual

Service Manual for Volumetric Infusion Pump
ARGUS 707 V
Made in Switzerland
ARGUS Medical AG, CH-3627 Heimberg / Switzerland
14.203.B_A707 en.SM.V4.3X ARGUS Medical AG 28.03.06 / PJ 1 / 48
TABLE OF CONTENTS
1. INTRODUCTION ...........................................................................................................4
1.1. General ............................................................................................................................. 4
2. PUMP CONFIGURATIONS...........................................................................................4
2.1. General ............................................................................................................................. 4
2.2. Interrogation mode (without ARGUS service) ................................................................... 5
2.3. Configuration mode (without ARGUS service).................................................................. 6
2.4. First activation of a PIN code (write protection) ................................................................7
2.5. Changing an existing PIN code......................................................................................... 8
2.6. Address list of the pump configuration (without ARGUS service)..................................... 9
2.7. Index list of the pump configuration (with ARGUS service) ............................................10
2.8. Special configuration options (with ARGUS service) ...................................................... 14
2.9. Medication list ................................................................................................................. 15
3. Serial communication of the pump..........................................................................18
3.1. General ........................................................................................................................... 18
3.2. Serial communication protocol........................................................................................ 18
4. ARGUS service...........................................................................................................19
4.1. General ........................................................................................................................... 19
4.2. ARGUS service – Configuration...................................................................................... 20
4.3. ARGUS service - Calibration (volumetric pums only) .....................................................21
4.4. ARGUS service - Toolbox............................................................................................... 22
5. SOFTWARE UPDATES ..............................................................................................26
5.1. General ........................................................................................................................... 26
5.2. Requirements for a software update............................................................................... 26
5.3. Software update procedure............................................................................................. 26
5.4. Safety aspects................................................................................................................. 28
6. MAINTENANCE ..........................................................................................................29
6.1. General ........................................................................................................................... 29
6.2. Recalibration ................................................................................................................... 29
6.3. Pressure calibration ........................................................................................................ 30
6.4. Pressure control measurement....................................................................................... 35
6.5. Volume calibration .......................................................................................................... 36
6.6. Pump specifications ........................................................................................................ 37
6.7. Fault codes ..................................................................................................................... 38
7. REPLACEMENT OF PARTS.......................................................................................39
7.1. General ........................................................................................................................... 39
7.2. Disassembling of the ARGUS 707 V............................................................................... 39
7.3. Spare parts ..................................................................................................................... 43
8. WIRING DIAGRAMM...................................................................................................46
9. BLOC SCHEMATIC ....................................................................................................46
10. SAFETY STANDARD CHECK....................................................................................47
11. REPAIR ORDER FORM..............................................................................................48
14.203.B_A707 en.SM.V4.3X ARGUS Medical AG 28.03.06 / PJ 1 / 48
Drop detector
Drop chamber
„upstream“
Pressure sensor
Park area for drop detector
automatic
„free-flow“-clamp
INTRODUCTION
Display
Global alarm
Air bubble
detector
Pressure sensor
„downstream“
14.203.B_A707 en.SM.V4.3X ARGUS Medical AG 28.03.06 / PJ 2 / 48
INTRODUCTION
AC power
connection
Universal clamp
Staff alert and
RS-232-connection
Drop detector
connection
Identification plate
with pump serial
number
DC-connection & QUICK®100
Docking Station interface
14.203.B_A707 en.SM.V4.3X ARGUS Medical AG 28.03.06 / PJ 3 / 48
1. INTRODUCTION
1.1. General
IMPORTANT!
This service manual is intended for the exclusive use of authorized persons who have been trained by ARGUS Medical AG in the maintenance and repair of the ARGUS 707 V infusion pump.
The service manual is meant to be used together with the user manual.
IMPORTANT!
ARGUS Medical AG shall not assume any responsibility for any manipulations which have been carried out on the ARGUS 707 V infusion pump by a non­authorized person.
CAUTION!
The ARGUS 707 V infusion pump may only be used with spare parts, accessories, consumables and IV-sets with Luer-Lock connections recommended by ARGUS Medical AG. The functional safety of the pump is not guaranteed if non approved materials are used. The safety of the patient may be endangered.
This manual contains the latest data available. It is subject to further modifications in accordance with technical improvements.
INTRODUCTION
2. PUMP CONFIGURATIONS
2.1. General
CAUTION!
The configuration possibilities with the “ARGUS service” PC utility tool and without
PC assistance constitute a modification of the pump and may only be carried out by authorized persons!
CAUTION!
After changing the configuration a function check and a control measurement has to be performed!
14.203.B_A707 en.SM.V4.3X ARGUS Medical AG 28.03.06 / PJ 4 / 48
PUMP CONFIGURATIONS
7
o
Pg1
5
2.2. Interrogation mode (without ARGUS service)
With the interrogation mode you can read the present keypad configuration of the pump without the possibility to modify any configurations. For a complete overview,
please take the “ARGUS service” PC-tool!
4. To quit the interro­gation mode press the “ON/OFF” key at least 2 seconds.
Flashing decimal points indicate which display is ready to accept an input by the keys 100, 10 & 1.
In the interrogation mode the left hand display shows the address and the right hand display shows the according value configured at this address. Please refer
to chapter 2.6. where the meanings of the addresses are explained.
To modify any configuration data you have to go into the configuration mode.
1. To enter into the interrogation mode, switch the pump on while
keeping the keys “MODE” and “START/STOP” pressed.
70
R
... .
2. Press the “START/
STOP” key.
3. Enter the requested address (see chapter 2.6), e.g. 5. The pro-
grammed value of the address 5 appears in the right hand display.
14.203.B_A707 en.SM.V4.3X ARGUS Medical AG 28.03.06 / PJ 5 / 48
2.3. Configuration mode (without ARGUS service)
-
7
p
o
g
The configuration mode permit you to modify the pump keypad configuration
manually using the keypad. Please refer to chapter 2.6. where the meaning of the
adresses are explained. To have access to all configuration options, please use the
“ARGUS service” PC-tool!
2.3.1. Step 1
1. To enter into the configuration mode, switch the pump on
while keeping the keys “MODE” and “START/STOP” pressed.
70
3. Press key “MODE”
r
PUMP CONFIGURATIONS
2. Press key “START”
4. Press key “START” again. The display will show “Cod ----“
2.3.2. Step 2
You have now access to all addresses in the list of chapter 2.6. Select therefore any address in the left display (see next page).
14.203.B_A707 en.SM.V4.3X ARGUS Medical AG 28.03.06 / PJ 6 / 48
Cod
---
5. Enter the PIN Code. If no PIN code was configured, no entry is required
6. Press key “START” to acknowledge the entered PIN code
If the PIN was not accepted, the display will change back to point 5.
2.3.3. Step 3
5
7
o
P
g
0
8. Enter now the value on the right hand display. The
range of the value is given by the table in chap.2.6
Important remark:
Invalid values entered will be corrected automatically by the pump to the maxima or minima value allowed for the according address!
... .
10. To quit the configuration mode press the “ON/OFF” key at least 2 seconds. Changes in configuration become active,
after the pump is switched on normally again.
PUMP CONFIGURATIONS
7. After entering an address e.g. 5 press the key “MODE”. The flashing decimal points will change to the right hand display.
9. Press key “START” to ac­knowledge the value en­tered. The flashing decimal points change back to the left hand display.
2.4. First activation of a PIN code (write protection)
The activation of a PIN code allows you to protect the configuration from unauthor-
ized access. To activate the PIN code, enter into the configuration mode.
5. Press “MODE” key (Code “0” will be set) then “START” key to acknowledge the entered PIN code “0”.
1. To enter into the configuration mode, switch the pump on
while keeping the keys “MODE” and “START/STOP” pressed.
70
R
2. Press key “START”
4. Press key “START” again. The display will show “Cod ----“.
3. Press key “MODE”
14.203.B_A707 en.SM.V4.3X ARGUS Medical AG 28.03.06 / PJ 7 / 48
0
0
7. Enter now the new PIN Code (max. 4 digit number).
Please remember this code, it will never be visible again!
e.g.
CAUTION!
After you switch the pump OFF and ON again you can enter into the configuration mode only, if you enter the correct PIN code.
Please note: The interrogation mode can always be accessed without the PIN.
... .
e.g.
PUMP CONFIGURATIONS
6. Press the “MODE” key. The flashing decimal points will change to the right hand display
8. Press key “START” to acknowledge to new PIN.
2.5. Changing an existing PIN code
Enter the configuration mode using present PIN, select add. “0” and set the new code.
Enter actual PIN code and confirm with “START” key.
Press “MODE” key (#0). The flashing decimal points will change to the right hand display.
Enter the new PIN code and press the “START/STOP” key to acknowledge the entered code.
14.203.B_A707 en.SM.V4.3X ARGUS Medical AG 28.03.06 / PJ 8 / 48
PUMP CONFIGURATIONS
2.6. Address list of the pump configuration (without ARGUS service)
The following list declares the possible configuration options which can be performed on the pump keypad without using the PC.
All these options can also be configured by the PC-Software “ARGUS service”.
Address Default Function Unit Range
left display
1 2 No Key ON/OFF only at stop valid - 0=No / 1=Yes 2 10 No Operation without drop detector ­3 11 No Recall of the last used ml/h rate ­4 19 No Buzzer at start ­5 44 Yes Automatic pressure release ­6 49 No Alarm acknowledge with key MODE -
100 361 5 Key ON/OFF delay time • 0.1 s 0 - 31 101 362 8 Display brightness level 2 - 15 102 363 7 Buzzer volume level 5 - 10 103 316 9 Default pressure limit (IV-set 1) • 100 mbar 1 - 10 104 318 20 Drop rate window centre (IV-set 1) Drops/ml 10 - 65
200 317 250 Air bubble size (IV-set 1) µl 50 - 1000
399 - 707 Enter the calibration menu (volume) - 123
- 707 Enter the calibration menu (pressure) - 1234
Index
PC
Note! The address does not correspond with the index used by the “ARGUS service” tool.
right display
0=No / 1=Yes 0=No / 1=Yes 0=No / 1=Yes 0=No / 1=Yes 0=No / 1=Yes
14.203.B_A707 en.SM.V4.3X ARGUS Medical AG 28.03.06 / PJ 9 / 48
PUMP CONFIGURATIONS
2.7. Index list of the pump configuration (with ARGUS service)
Index PC Add. Default Function Unit Range
1 Yes Run indication by running decimal point - No / Yes 2 1 No Key ON/OFF only at stop valid - No / Yes 3 No Rate change allowed only at STOP - No / Yes 4 No Key STOP delayed # 361 - No / Yes 5 No Second entry of rate # 3=Yes - No / Yes 6 No Static alarm (staff alerting system) - No / Yes 7 No Display elapsed time in run mode # 8=No - No / Yes 8 No Display remaining time # 7=Yes - No / Yes
9 10 2 No Operation without drop detector - No / Yes 11 3 Yes Recall of ml/h rate at next power on - No / Yes 12 No Recall of ml total at next power on - No / Yes 13 No Recall of ml inf. at next power on - No / Yes 14 No SBS (step by step function) - No / Yes 15 No Display VTBI (volume to be infused) - No / Yes 16 No Display “Set-X” if only one set is enabled - No / Yes 17 Yes KVO (KOR) enabled # 60 - No / Yes 18 No Drop alarm only if bottle empty # 10=No, trA - No / Yes 19 4 No Buzzer at start - No / Yes 20 No Menu Clr (clear ml inf.) # 15=No - No / Yes 21 22 No Menu trA (transport) # 10, 18=No - No / Yes 23 Yes Menu PrL (pressure alarm limit) - No / Yes 24 Yes Menu CAP (battery capacity) - No / Yes 25 No Menu SEt Fill (prime IV set) 26 No Menu InF (ml inf. since last power on) - No / Yes 27 No Menu dLo (data lock) - No / Yes 28 No Menu Stb (stand by) - No / Yes 29 No Menu Med (medication name) - No / Yes 30 No Menu tM (timer alarm) - No / Yes 31 32 No Menu boL (bolus release) - No / Yes 33 No Menu boLr (bolus rate) # 32=Yes - No / Yes 34 No Menu tot (bolus total) # 32=Yes - No / Yes 35 36 37 38 No Automatic bolus application # 32, 34=Yes - No / Yes 39 No Neonatology mode accessible in PrL menu - No / Yes 40 41 No Clear ml/h after infusion completed - No / Yes 42 No Clear ml total after infusion completed # 41=Yes - No / Yes 43 No Air volume accumulated - No / Yes 44 5 Yes Automatic pressure release after occlusion - No / Yes 45 Yes Pressure display ON (LED bar graph - 20/40/60/80/100%) - No / Yes 46 No Pressure display with indicator # 45=Yes - No / Yes 47 No Stand by- and battery pre alarm low volume - No / Yes 48 No Flashing numeric display at alarm - No / Yes 49 6 No Alarm acknowledge only with key MODE - No / Yes
14.203.B_A707 en.SM.V4.3X ARGUS Medical AG 28.03.06 / PJ 10 / 48
PUMP CONFIGURATIONS
50 No Start with ≥ 1 bar pressure allowed - No / Yes 51 Yes Enable upstream occlusion detection - No / Yes 52 53 54 55 No Select RJ-45 connector for serial interface - No / Yes 56 Yes Medication display alternate # 29=Yes - No / Yes
60 No KVO only after infusion completed - No / Yes
65 No Clear and continue # 15=No - No / Yes
101 Yes Set 1 enabled - No / Yes 102 No Set 2 enabled - No / Yes 103 No Set 3 enabled - No / Yes 104 No Set 4 enabled - No / Yes
105 No Automatic return on default set 1 after power up # 101=Yes - No / Yes
201 1065 Correction at 1 ml/h ratio 900 - 1100 202 1054 Correction at 3 ml/h ratio 900 - 1100 203 1044 Correction at 5 ml/h ratio 900 - 1100 204 1035 Correction at 10 ml/h ratio 900 - 1100 205 1027 Correction at 25 ml/h ratio 900 - 1100 206 1020 Correction at 50 ml/h ratio 900 - 1100 207 1014 Correction at 75 ml/h ratio 900 - 1100 208 1008 Correction at 100 ml/h ratio 900 - 1100 209 999 Correction at 150 ml/h ratio 900 -1100 210 988 Correction at 200 ml/h ratio 900 - 1100 211 976 Correction at 250 ml/h ratio 900 - 1100 212 967 Correction at 300 ml/h ratio 900 - 1100 213 952 Correction at 400 ml/h ratio 900 - 1100 214 942 Correction at 500 ml/h ratio 900 - 1100 215 925 Correction at 750 ml/h ratio 900 - 1100 216 910 Correction at 999 ml/h ratio 900 - 1100 217 1000 Correction at 1200 ml/h
221 1065 Correction at 1 ml/h ratio 900 - 1100 222 1054 Correction at 3 ml/h ratio 900 - 1100 223 1044 Correction at 5 ml/h ratio 900 - 1100 224 1035 Correction at 10 ml/h ratio 900 - 1100 225 1027 Correction at 25 ml/h ratio 900 - 1100 226 1020 Correction at 50 ml/h ratio 900 - 1100 227 1014 Correction at 75 ml/h ratio 900 - 1100 228 1008 Correction at 100 ml/h ratio 900 - 1100 229 999 Correction at 150 ml/h ratio 900 -1100 230 988 Correction at 200 ml/h ratio 900 - 1100 231 976 Correction at 250 ml/h ratio 900 - 1100 232 967 Correction at 300 ml/h ratio 900 - 1100 233 952 Correction at 400 ml/h ratio 900 - 1100 234 942 Correction at 500 ml/h ratio 900 - 1100 235 925 Correction at 750 ml/h ratio 900 - 1100 236 910 Correction at 999 ml/h ratio 900 - 1100 237 1000 Correction at 1200 ml/h
SET 1
Do not change these values unless explicitly trained
ratio 1000-1100
SET 2
Do not change these values unless explicitly trained
ratio 1000-1100
14.203.B_A707 en.SM.V4.3X ARGUS Medical AG 28.03.06 / PJ 11 / 48
PUMP CONFIGURATIONS
238 239 240 241 1065 Correction at 1 ml/h ratio 900 - 1100 242 1054 Correction at 3 ml/h ratio 900 - 1100 243 1044 Correction at 5 ml/h ratio 900 - 1100 244 1035 Correction at 10 ml/h ratio 900 - 1100 245 1027 Correction at 25 ml/h ratio 900 - 1100 246 1020 Correction at 50 ml/h ratio 900 - 1100 247 1014 Correction at 75 ml/h ratio 900 - 1100 248 1008 Correction at 100 ml/h ratio 900 - 1100 249 999 Correction at 150 ml/h ratio 900 - 1100 250 988 Correction at 200 ml/h ratio 900 - 1100 251 976 Correction at 250 ml/h ratio 900 - 1100 252 967 Correction at 300 ml/h ratio 900 - 1100 253 952 Correction at 400 ml/h ratio 900 - 1100 254 942 Correction at 500 ml/h ratio 900 - 1100 255 925 Correction at 750 ml/h ratio 900 - 1100 256 910 Correction at 999 ml/h ratio 900 - 1100 257 1000 Correction at 1200 ml/h 258 259 260 261 1065 Correction at 1 ml/h ratio 900 - 1100 262 1054 Correction at 3 ml/h ratio 900 - 1100 263 1044 Correction at 5 ml/h ratio 900 - 1100 264 1035 Correction at 10 ml/h ratio 900 - 1100 265 1027 Correction at 25 ml/h ratio 900 - 1100 266 1020 Correction at 50 ml/h ratio 900 - 1100 267 1014 Correction at 75 ml/h ratio 900 - 1100 268 1008 Correction at 100 ml/h ratio 900 - 1100 269 999 Correction at 150 ml/h ratio 900 - 1100 270 988 Correction at 200 ml/h ratio 900 - 1100 271 976 Correction at 250 ml/h ratio 900 - 1100 272 967 Correction at 300 ml/h ratio 900 - 1100 273 952 Correction at 400 ml/h ratio 900 - 1100 274 942 Correction at 500 ml/h ratio 900 - 1100 275 925 Correction at 750 ml/h ratio 900 - 1100 276 910 Correction at 999 ml/h ratio 900 - 1100 277 1000 Correction at 1200 ml/h
310 999.9 Max. flow rate ml/h 1 - 999.9 311 999 Max. prime rate ml/h 1 - 999 312 1200 Max. bolus rate ml/h 1 - 1200 313 10 Max. bolus total ml 1 - 999 314 10 Upstream occlusion sensitivity factor 0 - 60 315 20 Air summation volume • 50 µl 2 - 40 316 103 7 Downstream default pressure limit PrL • 100 mbar 1 - 10 317 200 250 Air bubble size µl 50 - 1000 318 104 20 Drop rate window center drops 10 - 65 319
SET 3
Do not change these values unless explicitly trained
ratio 1000-1100
SET 4
Do not change these values unless explicitly trained
ratio 1000-1100
Set 1 parameters
14.203.B_A707 en.SM.V4.3X ARGUS Medical AG 28.03.06 / PJ 12 / 48
PUMP CONFIGURATIONS
320 999.9 Max. flow rate ml/h 1 - 999.9 321 999 Max. prime rate ml/h 1 - 999 322 1200 Max. bolus rate ml/h 1 - 1200 323 10 Max. bolus total ml 1 - 999 324 10 Upstream occlusion sensitivity factor 0 - 60 325 20 Air summation volume • 50 µl 2 - 40 326 7 Downstream default pressure limit PrL • 100 mbar 1 - 10 327 250 Air bubble size µl 50 - 1000 328 20 Drop rate window center drops 10 - 65 329 330 999.9 Max. flow rate ml/h 1 - 999.9 331 999 Max. prime rate ml/h 1 - 999 332 1200 Max. bolus rate ml/h 1 - 1200 333 10 Max. bolus total ml 1 - 999 334 10 Upstream occlusion sensitivity factor 0 - 60 335 20 Air summation volume • 50 µl 2 - 40 336 7 Downstream default pressure limit PrL • 100 mbar 1 - 10 337 250 Air bubble size µl 50 - 1000 338 20 Drop rate window center drops 10 - 65 339 340 999.9 Max. flow rate ml/h 1 - 999.9 341 999 Max. prime rate ml/h 1 - 999 342 1200 Max. bolus rate ml/h 1 - 1200 343 10 Max. bolus total ml 1 - 999 344 10 Upstream occlusion sensitivity factor 0 - 60 345 20 Air summation volume • 50 µl 2 - 40 346 7 Downstream default pressure limit PrL • 100 mbar 1 - 10 347 250 Air bubble size µl 50 - 1000 348 20 Drop rate window center drops 10 - 65 349 350 351 4 Air volume summation time • 8 min 1 - 8
361 100 5 Key ON/OFF delay time # 4 • 1/10 s 0 - 31 362 101 8 Display brightness level 2 - 15 363 102 10 Buzzer alarm volume level 5 - 10 364 365 103 150 Pressure step for neonatology mode # 39=Yes mbar 60 - 250 366 1 Pressure display unit (mbar / mmHg / kPa / cmH2O / Psi) Enum 1 - 5 367 368 200 300 Battery capacity (discharge time) min 45 - 300 369 5 Automatic menu fall back delay time sec 5 - 30
390 0 Last service date in year year 0 - 99 391 0 Last service date in month month 0 - 12 392 0 Last service date in day day 0 - 31 393 0 Service interval in months (0 = disabled) month 0 - 24 394 0 Service interval in hours of operation (0 = disabled) hour 0 -10000 395
396 0 Pump serial number xxx 8 yyy xxx 8 yyy
397 398
- 399 707 Enter the calibration menu / clears protection key - 123/4
Set 2 parameters
Set 3 parameters
Set 4 parameters
14.203.B_A707 en.SM.V4.3X ARGUS Medical AG 28.03.06 / PJ 13 / 48
PUMP CONFIGURATIONS
518 2
522 No Allow change of ml total while infusing # 65=No - No / Yes 523 15 Alarm lamp brightness level 2 - 15 524 Display a department info text (after power up) char
Permissions for serial communication 0 = none, 1 = query only, 2 = parametrising, 3 = remote control
Enum 0 - 3
1-16 ASCII
Using the “ARGUS service” tool, the complete and detailed pump configuration
can be done.
2.8. Special configuration options (with ARGUS service)
Enabling a new IV set
Follow the steps below to enable and configure a new IV set:
a) write “Yes” in the corresponding index 101, 102, 103, 104 b) make sure the set definitions (index 201 to 277) correspond to the IV sets used.
In case of doubt, contact ARGUS Medical AG or your local dealer.
c) configure the set parameters (index 310 to 349 and 351)
d) perform a complete set calibration as per chapters 6.3. to 6.5.
14.203.B_A707 en.SM.V4.3X ARGUS Medical AG 28.03.06 / PJ 14 / 48
Loading...
+ 34 hidden pages