10. SAFETY STANDARD CHECK....................................................................................47
11. REPAIR ORDER FORM..............................................................................................48
14.203.B_A707 en.SM.V4.3X ARGUS Medical AG 28.03.06 / PJ 1 / 48
Drop detector
Drop chamber
„upstream“
Pressure sensor
Park area for
drop detector
automatic
„free-flow“-clamp
INTRODUCTION
Display
Global alarm
Air bubble
detector
Pressure sensor
„downstream“
14.203.B_A707 en.SM.V4.3X ARGUS Medical AG 28.03.06 / PJ 2 / 48
INTRODUCTION
AC power
connection
Universal clamp
Staff alert and
RS-232-connection
Drop detector
connection
Identification plate
with pump serial
number
DC-connection & QUICK®100
Docking Station interface
14.203.B_A707 en.SM.V4.3X ARGUS Medical AG 28.03.06 / PJ 3 / 48
1. INTRODUCTION
1.1. General
IMPORTANT!
This service manual is intended for the exclusive use of authorized persons who
have been trained by ARGUS Medical AG in the maintenance and repair of the
ARGUS 707 V infusion pump.
The service manual is meant to be used together with the user manual.
IMPORTANT!
ARGUS Medical AG shall not assume any responsibility for any manipulations
which have been carried out on the ARGUS 707 V infusion pump by a nonauthorized person.
CAUTION!
The ARGUS 707 V infusion pump may only be used with spare parts, accessories,
consumables and IV-sets with Luer-Lock connections recommended by ARGUS
Medical AG. The functional safety of the pump is not guaranteed if non approved
materials are used. The safety of the patient may be endangered.
This manual contains the latest data available. It is subject to further modifications
in accordance with technical improvements.
INTRODUCTION
2. PUMP CONFIGURATIONS
2.1. General
CAUTION!
The configuration possibilities with the “ARGUS service” PC utility tool and without
PC assistance constitute a modification of the pump and may only be carried out by
authorized persons!
CAUTION!
After changing the configuration a function check and a control measurement has to
be performed!
14.203.B_A707 en.SM.V4.3X ARGUS Medical AG 28.03.06 / PJ 4 / 48
PUMP CONFIGURATIONS
7
o
Pg1
5
2.2. Interrogation mode (without ARGUS service)
With the interrogation mode you can read the present keypad configuration of the
pump without the possibility to modify any configurations. For a complete overview,
please take the “ARGUS service” PC-tool!
4. To quit the interrogation mode press
the “ON/OFF” key
at least 2 seconds.
Flashing decimal points indicate which display is ready to accept an input by the
keys 100, 10 & 1.
In the interrogation mode the left hand display shows the address and the right
hand display shows the according value configured at this address. Please refer
to chapter 2.6. where the meanings of the addresses are explained.
To modify any configuration data you have to go into the configuration mode.
1.To enter into the interrogation mode, switch the pump on while
keeping the keys “MODE” and “START/STOP” pressed.
70
R
....
2. Press the
“START/
STOP” key.
3. Enter the requested address (see chapter 2.6), e.g. 5. The pro-
grammed value of the address 5
appears in the right hand display.
14.203.B_A707 en.SM.V4.3X ARGUS Medical AG 28.03.06 / PJ 5 / 48
2.3. Configuration mode (without ARGUS service)
-
7
p
o
g
The configuration mode permit you to modify the pump keypad configuration
manually using the keypad. Please refer to chapter 2.6. where the meaning of the
adresses are explained. To have access to all configuration options, please use the
“ARGUS service” PC-tool!
2.3.1. Step 1
1.To enter into the configuration mode, switch the pump on
while keeping the keys “MODE” and “START/STOP” pressed.
70
3. Press key “MODE”
r
PUMP CONFIGURATIONS
2. Press key “START”
4. Press key “START” again.
The display will show “Cod ----“
2.3.2. Step 2
You have now access to all addresses in the list of chapter 2.6.
Select therefore any address in the left display (see next page).
14.203.B_A707 en.SM.V4.3X ARGUS Medical AG 28.03.06 / PJ 6 / 48
Cod
---
5. Enter the PIN Code. If no PIN code
was configured, no entry is required
6. Press key “START”
to acknowledge the
entered PIN code
If the PIN was not
accepted, the display
will change back to
point 5.
2.3.3. Step 3
5
7
o
P
g
0
8. Enter now the value on the right hand display. The
range of the value is given by the table in chap.2.6
Important remark:
Invalid values entered will be corrected automatically by the pump to the maxima or
minima value allowed for the according address!
....
10. To quit the configuration mode press the
“ON/OFF” key at least 2 seconds.
Changes in configuration become active,
after the pump is switched on normally again.
PUMP CONFIGURATIONS
7. After entering an address e.g. 5 press the
key “MODE”. The flashing decimal points
will change to the right hand display.
9. Press key “START” to acknowledge the value entered. The flashing decimal
points change back to the
left hand display.
2.4. First activation of a PIN code (write protection)
The activation of a PIN code allows you to protect the configuration from unauthor-
ized access. To activate the PIN code, enter into the configuration mode.
5. Press “MODE” key (Code “0” will be set) then “START”
key to acknowledge the entered PIN code “0”.
1.To enter into the configuration mode, switch the pump on
while keeping the keys “MODE” and “START/STOP” pressed.
70
R
2. Press key “START”
4. Press key “START”
again. The display will
show “Cod ----“.
3. Press key “MODE”
14.203.B_A707 en.SM.V4.3X ARGUS Medical AG 28.03.06 / PJ 7 / 48
0
0
7. Enter now the new PIN Code (max. 4 digit number).
Please remember this code, it will never be visible again!
e.g.
CAUTION!
After you switch the pump OFF and ON again you can enter into the configuration
mode only, if you enter the correct PIN code.
Please note: The interrogation mode can always be accessed without the PIN.
....
e.g.
PUMP CONFIGURATIONS
6. Press the “MODE” key.
The flashing decimal
points will change to the
right hand display
8. Press key “START” to
acknowledge to new PIN.
2.5. Changing an existing PIN code
Enter the configuration mode using present PIN, select add. “0” and set the new
code.
Enter actual PIN code and confirm with “START” key.
Press “MODE” key (#0). The flashing decimal points
will change to the right hand display.
Enter the new PIN code and press the “START/STOP”
key to acknowledge the entered code.
14.203.B_A707 en.SM.V4.3X ARGUS Medical AG 28.03.06 / PJ 8 / 48
PUMP CONFIGURATIONS
2.6. Address list of the pump configuration (without ARGUS service)
The following list declares the possible configuration options which can be
performed on the pump keypad without using the PC.
All these options can also be configured by the PC-Software “ARGUS service”.
Address Default Function Unit Range
left display
1 2 No Key ON/OFF only at stop valid - 0=No / 1=Yes
2 10 No Operation without drop detector 3 11 No Recall of the last used ml/h rate 4 19 No Buzzer at start 5 44 Yes Automatic pressure release 6 49 No Alarm acknowledge with key MODE -
14.203.B_A707 en.SM.V4.3X ARGUS Medical AG 28.03.06 / PJ 9 / 48
PUMP CONFIGURATIONS
2.7. Index list of the pump configuration (with ARGUS service)
Index PC Add. Default Function Unit Range
1 Yes Run indication by running decimal point - No / Yes
2 1 No Key ON/OFF only at stop valid - No / Yes
3 No Rate change allowed only at STOP - No / Yes
4 No Key STOP delayed # 361 - No / Yes
5 No Second entry of rate # 3=Yes - No / Yes
6 No Static alarm (staff alerting system) - No / Yes
7 No Display elapsed time in run mode # 8=No - No / Yes
8 No Display remaining time # 7=Yes - No / Yes
9
10 2 No Operation without drop detector - No / Yes
11 3 Yes Recall of ml/h rate at next power on - No / Yes
12 No Recall of ml total at next power on - No / Yes
13 No Recall of ml inf. at next power on - No / Yes
14 No SBS (step by step function) - No / Yes
15 No Display VTBI (volume to be infused) - No / Yes
16 No Display “Set-X” if only one set is enabled - No / Yes
17 Yes KVO (KOR) enabled # 60 - No / Yes
18 No Drop alarm only if bottle empty # 10=No, → trA - No / Yes
19 4 No Buzzer at start - No / Yes
20 No Menu Clr (clear ml inf.) # 15=No - No / Yes
21
22 No Menu trA (transport) # 10, 18=No - No / Yes
23 Yes Menu PrL (pressure alarm limit) - No / Yes
24 Yes Menu CAP (battery capacity) - No / Yes
25 No Menu SEt Fill (prime IV set)
26 No Menu InF (ml inf. since last power on) - No / Yes
27 No Menu dLo (data lock) - No / Yes
28 No Menu Stb (stand by) - No / Yes
29 No Menu Med (medication name) - No / Yes
30 No Menu tM (timer alarm) - No / Yes
31
32 No Menu boL (bolus release) - No / Yes
33 No Menu boLr (bolus rate) # 32=Yes - No / Yes
34 No Menu tot (bolus total) # 32=Yes - No / Yes
35
36
37
38 No Automatic bolus application # 32, 34=Yes - No / Yes
39 No Neonatology mode accessible in PrL menu - No / Yes
40
41 No Clear ml/h after infusion completed - No / Yes
42 No Clear ml total after infusion completed # 41=Yes - No / Yes
43 No Air volume accumulated - No / Yes
44 5 Yes Automatic pressure release after occlusion - No / Yes
45 Yes Pressure display ON (LED bar graph - 20/40/60/80/100%) - No / Yes
46 No Pressure display with indicator # 45=Yes - No / Yes
47 No Stand by- and battery pre alarm low volume - No / Yes
48 No Flashing numeric display at alarm - No / Yes
49 6 No Alarm acknowledge only with key MODE - No / Yes
14.203.B_A707 en.SM.V4.3X ARGUS Medical AG 28.03.06 / PJ 10 / 48
PUMP CONFIGURATIONS
50 No Start with ≥ 1 bar pressure allowed - No / Yes
51 Yes Enable upstream occlusion detection - No / Yes
52
53
54
55 No Select RJ-45 connector for serial interface - No / Yes
56 Yes Medication display alternate # 29=Yes - No / Yes
60 No KVO only after infusion completed - No / Yes
65 No Clear and continue # 15=No - No / Yes
101 Yes Set 1 enabled - No / Yes
102 No Set 2 enabled - No / Yes
103 No Set 3 enabled - No / Yes
104 No Set 4 enabled - No / Yes
105 No Automatic return on default set 1 after power up # 101=Yes - No / Yes
201 1065 Correction at 1 ml/h ratio 900 - 1100
202 1054 Correction at 3 ml/h ratio 900 - 1100
203 1044 Correction at 5 ml/h ratio 900 - 1100
204 1035 Correction at 10 ml/h ratio 900 - 1100
205 1027 Correction at 25 ml/h ratio 900 - 1100
206 1020 Correction at 50 ml/h ratio 900 - 1100
207 1014 Correction at 75 ml/h ratio 900 - 1100
208 1008 Correction at 100 ml/h ratio 900 - 1100
209 999 Correction at 150 ml/h ratio 900 -1100
210 988 Correction at 200 ml/h ratio 900 - 1100
211 976 Correction at 250 ml/h ratio 900 - 1100
212 967 Correction at 300 ml/h ratio 900 - 1100
213 952 Correction at 400 ml/h ratio 900 - 1100
214 942 Correction at 500 ml/h ratio 900 - 1100
215 925 Correction at 750 ml/h ratio 900 - 1100
216 910 Correction at 999 ml/h ratio 900 - 1100
217 1000 Correction at 1200 ml/h
221 1065 Correction at 1 ml/h ratio 900 - 1100
222 1054 Correction at 3 ml/h ratio 900 - 1100
223 1044 Correction at 5 ml/h ratio 900 - 1100
224 1035 Correction at 10 ml/h ratio 900 - 1100
225 1027 Correction at 25 ml/h ratio 900 - 1100
226 1020 Correction at 50 ml/h ratio 900 - 1100
227 1014 Correction at 75 ml/h ratio 900 - 1100
228 1008 Correction at 100 ml/h ratio 900 - 1100
229 999 Correction at 150 ml/h ratio 900 -1100
230 988 Correction at 200 ml/h ratio 900 - 1100
231 976 Correction at 250 ml/h ratio 900 - 1100
232 967 Correction at 300 ml/h ratio 900 - 1100
233 952 Correction at 400 ml/h ratio 900 - 1100
234 942 Correction at 500 ml/h ratio 900 - 1100
235 925 Correction at 750 ml/h ratio 900 - 1100
236 910 Correction at 999 ml/h ratio 900 - 1100
237 1000 Correction at 1200 ml/h
SET 1
Do not change
these values unless
explicitly trained
ratio 1000-1100
SET 2
Do not change
these values unless
explicitly trained
ratio 1000-1100
14.203.B_A707 en.SM.V4.3X ARGUS Medical AG 28.03.06 / PJ 11 / 48
PUMP CONFIGURATIONS
238
239
240
241 1065 Correction at 1 ml/h ratio 900 - 1100
242 1054 Correction at 3 ml/h ratio 900 - 1100
243 1044 Correction at 5 ml/h ratio 900 - 1100
244 1035 Correction at 10 ml/h ratio 900 - 1100
245 1027 Correction at 25 ml/h ratio 900 - 1100
246 1020 Correction at 50 ml/h ratio 900 - 1100
247 1014 Correction at 75 ml/h ratio 900 - 1100
248 1008 Correction at 100 ml/h ratio 900 - 1100
249 999 Correction at 150 ml/h ratio 900 - 1100
250 988 Correction at 200 ml/h ratio 900 - 1100
251 976 Correction at 250 ml/h ratio 900 - 1100
252 967 Correction at 300 ml/h ratio 900 - 1100
253 952 Correction at 400 ml/h ratio 900 - 1100
254 942 Correction at 500 ml/h ratio 900 - 1100
255 925 Correction at 750 ml/h ratio 900 - 1100
256 910 Correction at 999 ml/h ratio 900 - 1100
257 1000 Correction at 1200 ml/h
258
259
260
261 1065 Correction at 1 ml/h ratio 900 - 1100
262 1054 Correction at 3 ml/h ratio 900 - 1100
263 1044 Correction at 5 ml/h ratio 900 - 1100
264 1035 Correction at 10 ml/h ratio 900 - 1100
265 1027 Correction at 25 ml/h ratio 900 - 1100
266 1020 Correction at 50 ml/h ratio 900 - 1100
267 1014 Correction at 75 ml/h ratio 900 - 1100
268 1008 Correction at 100 ml/h ratio 900 - 1100
269 999 Correction at 150 ml/h ratio 900 - 1100
270 988 Correction at 200 ml/h ratio 900 - 1100
271 976 Correction at 250 ml/h ratio 900 - 1100
272 967 Correction at 300 ml/h ratio 900 - 1100
273 952 Correction at 400 ml/h ratio 900 - 1100
274 942 Correction at 500 ml/h ratio 900 - 1100
275 925 Correction at 750 ml/h ratio 900 - 1100
276 910 Correction at 999 ml/h ratio 900 - 1100
277 1000 Correction at 1200 ml/h
390 0 Last service date in year year 0 - 99
391 0 Last service date in month month 0 - 12
392 0 Last service date in day day 0 - 31
393 0 Service interval in months (0 = disabled) month 0 - 24
394 0 Service interval in hours of operation (0 = disabled) hour 0-10000
395
396 0 Pump serial number xxx 8 yyy xxx 8 yyy
397
398
- 399 707 Enter the calibration menu / clears protection key - 123/4
Set 2 parameters
Set 3 parameters
Set 4 parameters
14.203.B_A707 en.SM.V4.3X ARGUS Medical AG 28.03.06 / PJ 13 / 48
PUMP CONFIGURATIONS
518 2
522 No Allow change of ml total while infusing # 65=No - No / Yes
523 15 Alarm lamp brightness level 2 - 15
524 Display a department info text (after power up) char
Permissions for serial communication
0 = none, 1 = query only, 2 = parametrising, 3 = remote control
Enum 0 - 3
1-16 ASCII
Using the “ARGUS service” tool, the complete and detailed pump configuration
can be done.
2.8. Special configuration options (with ARGUS service)
Enabling a new IV set
Follow the steps below to enable and configure a new IV set:
a) write “Yes” in the corresponding index 101, 102, 103, 104
b) make sure the set definitions (index 201 to 277) correspond to the IV sets used.
In case of doubt, contact ARGUS Medical AG or your local dealer.
c) configure the set parameters (index 310 to 349 and 351)
d) perform a complete set calibration as per chapters 6.3. to 6.5.
14.203.B_A707 en.SM.V4.3X ARGUS Medical AG 28.03.06 / PJ 14 / 48
Loading...
+ 34 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.