Argus 707 Green Stream User manual

Service Manual for Volumetric Infusion Pump
green stream
ARGUS 707
Made in Switzerland
0120
®
ARGUS Medical AG, CH-3627 Heimberg / Switzerland
(A member of the CODAN group)
INTRODUCTION
Table of contents
1. INTRODUCTION........................................................................................................................................................ 3
1.1. General ................................................................................................................................................ 3
2. PUMP CONFIGURATIONS...................................................................................................................................... 3
2.1. General ................................................................................................................................................ 3
2.2. Interrogation mode .............................................................................................................................. 4
2.3. Configuration mode............................................................................................................................. 5
2.4. First activation of a configuration PIN Code ...................................................................................... 7
2.5. Changing an existing pin code ............................................................................................................ 7
2.6. Address list of the pump configuration ............................................................................................... 8
2.7. Special configuration options.............................................................................................................10
3. CONFIGURATION AND HISTORY PRINTOUT ............................................................................................... 11
3.1. Introduction ........................................................................................................................................11
3.2. Settings of the terminal program ........................................................................................................11
3.3. Configuration printout........................................................................................................................12
3.4. History printout ..................................................................................................................................14
4. SOFTWARE UPDATES ...........................................................................................................................................16
4.1. General ...............................................................................................................................................16
4.2. Requirements for a software update ...................................................................................................16
4.3. Safety aspects .....................................................................................................................................16
5. MAINTENANCE .......................................................................................................................................................17
5.1. General ...............................................................................................................................................17
5.2. Recalibration ......................................................................................................................................17
5.3. Pressure calibration ............................................................................................................................17
5.4. Pressure control measurement............................................................................................................23
5.5. Volume calibration.............................................................................................................................24
5.6. Pump specifications............................................................................................................................25
5.7. Fault codes..........................................................................................................................................26
6. REPLACEMENT OF PARTS.................................................................................................................................. 27
6.1. General ...............................................................................................................................................27
6.2. Disassembling of the ARGUS 707.....................................................................................................27
6.3. Replacements of parts ........................................................................................................................31
6.4. Spare parts ..........................................................................................................................................32
7. SAFETY STANDARD CHECK............................................................................................................................... 36
8. WIRING DIAGRAMM ............................................................................................................................................. 37
9. BLOC SCHEMATIC................................................................................................................................................. 37
10. APPENDIX .................................................................................................................................................................38
14.164.A_A707en.SM.V1.01.doc ARGUS Medical AG 17.02.04 AS 2 / 38
1. INTRODUCTION
1.1. General
IMPORTANT!
This service manual is intended for the exclusive use of authorized persons who have been trained by ARGUS Medical AG in the maintenance and repair of the ARGUS 707 infusion pump.
The service manual is meant to be used together with the user manual.
IMPORTANT!
ARGUS Medical AG shall not assume any responsibility for any manipulations which have been carried out on the ARGUS 707 infusion pump by a non-authorized person.
This manual contains the latest data available. It is subject to further modifications in ac­cordance with technical improvements.
2. PUMP CONFIGURATIONS
INTRODUCTION
2.1. General
CAUTION! The configuration possibilities constitute a modification of the pump and may only be carried out by authorized persons!
CAUTION! After changing the configuration a function check and a control measurement has to be performed!
14.164.A_A707en.SM.V1.01.doc ARGUS Medical AG 10.02.04 AS 3 / 38
PUMP CONFIGURATIONS
7
PRo
g
5.2
0
2.2. Interrogation mode
With the interrogation mode you can read the present configuration of the pump without the possibility to modify any configurations.
Flashing decimal points indicate which display is ready to accept an input by the keys 100, 10 & 1.
In the interrogation mode the left hand display shows the address and the right hand dis­play shows the according value configured at this address. Please refer to chapter 2.6 where the meanings of the addresses are explained.
To modify any configuration data you have to go into the configuration mode.
1. To enter into the interrogation mode, switch the pump on
while keeping the keys “MODE” and START/STOP” pressed.
70
. .
4. To quit the interrogation mode press the “ON/OFF” key at least 2 seconds.
.
3. Enter the requested address (see chapter 2.6), e.g.
52.
The programmed value of the address 52 appears
in the right hand display.
2. Press the “START/ STOP” key.
14.164.A_A707en.SM.V1.01.doc ARGUS Medical AG 10.02.04 AS 4 / 38
2.3. Configuration mode
-
7
g
The configuration mode alows you to modify the pump configuration manually. Please refer to chapter 2.6 where the meaning of the configuration adresses are explained.
2.3.1. Step 1
1. To enter into the configuration mode, switch the pump on
while keeping the keys “MODE” and START/STOP” pressed.
PUMP CONFIGURATIONS
70
pro
2. Press key “START”.
3. Press key “MODE”.
4. Press key “START” again. The
3.
display will show “Cod ----“.
Pre
2.3.2. Step 2
If the PIN code was accepted the pump will show the display of chapter 2.3.1 again, but you have now access to all writeable addresses in the list of chapter 2.6.
Select therefore any address in the left display (see next side):
5. Enter the PIN Code. If no PIN code was configured before, default PIN 0 will be active. For PIN 0 press the key “10” while keeping the key “1” pressed.
Cod
---
6. Press key “START” to acknowledge the en­tered PIN code.
If the PIN was not accepted, the display will change back to step 2.
14.164.A_A707en.SM.V1.01.doc ARGUS Medical AG 10.02.04 AS 5 / 38
PUMP CONFIGURATIONS
0
0.
2.3.3. Step 3
8. Enter now the value on the right hand
Important remark:
Invalid values entered will be corrected automatically by the pump to the maxima or minima value allowed for the according address!
display. The range of the value is given by the table in chapter 2.6.
1
10. To quit the configuration mode press the “ON/OFF” key at least 2 seconds. Changes in configuration become active, if the pump is switched on normally again.
7. After entering an address e.g. 10 press the key “MODE”. The flashing decimal points will change to the right hand display.
. . .
9. Press key “START” to acknowledge the value entered. The flashing decimal points change back to the left hand display.
14.164.A_A707en.SM.V1.01.doc ARGUS Medical AG 10.02.04 AS 6 / 38
PUMP CONFIGURATIONS
g
7
g
0
0
2.4. First activation of a configuration PIN Code
The activation of a PIN code allows you to protect the configuration from unauthorized persons. To activate the PIN code, enter the configuration mode as described in the previous chapter. For this use the default PIN code “0”:
After you switch the pump OFF and ON again you can enter the configuration mode only, if you enter the correct PIN code in display of chapter 2.3.2.
Please note: The interrogation mode can always be accessed without the PIN.
1. To enter into the configuration mode, switch the pump on
while keeping the keys “MODE” and START/STOP” pressed.
70
5. Enter the default PIN Code “0”. To do this press the key “10” while keeping the key “1” pressed. Press “START” key to acknowled
e the entered PIN code.
pro
. . .
7. Enter now the new PIN Code (max. 4 digit number). Please remember this code, it will never be visible again!
2. Press key “START”.
3. Press key “MODE”.
4. Press key “START” again. The
3.
display will show “Cod ----“.
Pre
6. Press the “MODE” key. The flashing decimal points will change to the right hand display
8. Press key “START” to acknowledge to new PIN.
2.5. Changing an existing pin code
Enter the configuration mode (see chapter 2.3), select address 0 and enter the new pin code.
14.164.A_A707en.SM.V1.01.doc ARGUS Medical AG 10.02.04 AS 7 / 38
2.6. Address list of the pump configuration
Add.: Address Def.: Default value of the manufacturer P: Protection grade (W = Write enabled)
Add. Def. P Function Range
0 0 W PIN code active 0=No; 1=Yes 1 1 W Run indication by running decimal point 0=No; 1=Yes 2 0 W Key [ON/OFF] only at STOP valid 0=No; 1=Yes 3 0 W Rate change only at STOP valid 0=No; 1=Yes 4 0 W Key [STOP] delayed (time at #361) 0=No; 1=Yes 5 6 0 W Static alarm (staff alerting system) 0=No; 1=Yes 7 0 W Display elapsed time in run mode (#8=0) 0=No; 1=Yes 8 0 W Select remaining time (#7=1) 0=No; 1=Yes 9
10 0 W Operation without drop detector 0=No; 1=Yes 11 1 W Recall "ml/h" (rate) at next power on 0=No; 1=Yes 12 0 W Recall "ml total" (end volume) at next power on 0 = N o ; 1=Yes 13 0 W Recall "ml inf." (volume inf.) at next power on 0 = N o ; 1=Yes 14 0 W SBS Step by step function 0=No; 1=Yes 15 0 W Display VTBI (Volume To Be Infused) 0=No; 1=Yes 16 17 1 W KVO (KOR, mode see (#60) 0=No; 1=Yes 18 0 W Drop alarm only if bottle empty
(#10=0, => 'trA' as default)
19 1 W Buzzer at start 0=No; 1=Yes
20 0 W Menu 'CLr' (Clear "ml inf.") (#15=0) 0=No; 1=Yes 21 22 0 W Menu 'trA' (Transport)(#10=0) 0=No; 1=Yes 23 1 W Menu 'PrL' (Pressure alarm limit) 0=No; 1=Yes 24 1 W Menu 'CAP' (Battery Capacity) 0=No; 1=Yes 25 0 W Menu 'SEt Fill' 0=No; 1=Yes 26 27 0 W Menu 'dLo' (Data-lock) 0=No; 1=Yes 28 0 W Menu 'Stb' (Stand-by) 0=No; 1=Yes 29 0 W Menu 'MEd' (Medication-no.) 0=No; 1=Yes
30 0 W Menu 'tM ' (Timer alarm) 0=No; 1=Yes 32 0 W Menu 'boL ' (Release Bolus) 0=No; 1=Yes 33 0 W Menu 'boLr' (Bolus rate)(#32=1) 0=No; 1=Yes 34 0 W Menu 'tot ' (Bolus total)(#32=1) 0=No; 1=Yes 38 0 W BOLUS application automatic (#32=1,#34=1) 0=No; 1=Yes
40 0 W Demo mode (menus addr. 20..34 and 7 enabled) 0=No; 1=Yes 41 0 W Clear "ml/h" after infusion completed 0=No; 1=Yes 42 0 W Clear "ml total" after inf. completed (#41=1) 0=No; 1=Yes 43 0 W Air volume accumulated, see #350 and 351 0=No; 1=Yes 44 1 W Automatic pressure release after occlusion 0=No; 1=Yes 45 1 W Pressure display (bar graph) 0=No; 1=Yes 46 0 W Pressure display with indicator (#45=1) 0=No; 1=Yes 47 0 W Standby- and battery pre alarm low volume 0=No; 1=Yes 48 0 W Flashing numeric display at alarm 0=No; 1=Yes 49 0 W Alarm acknowledge only with key [MODE] 0=No; 1=Yes
PUMP CONFIGURATIONS
0=No; 1=Yes
14.164.A_A707en.SM.V1.01.doc ARGUS Medical AG 10.02.04 AS 8 / 38
PUMP CONFIGURATIONS
Add. Def. P Function Range
50 0 W Start with >= 1bar downstream pressure allowed 0=No; 1=Yes 55 0 W Set default serial interface to RJ-45 connector 0=No; 1=Yes
60 0 W KVO only after infusion completed 0=No; 1=Yes
101 1 Set 1 enabled (always enabled) 0=No; 1=Yes
201 1091 W 202 1071 W 203 1061 W 204 1051 W 205 1031 W 206 1023 W 207 1014 W 208 1006 W 209 999 W
210 988 W
IV-Set 1 definitions IV-Set 1 parameters
211 974 W 212 967 W 213 947 W 214 939 W 215 917 W 216 900 W 217 1000 W
306 0 Infused sum in ml (xxxxyyyy) [xxxx.... ml]
307 0 Infused sum in ml (xxxxyyyy) [....yyyy ml]
308 0 Operating time in min (xxxxyyyy) [xxxx.... min]
309 0 Operating time in min (xxxxyyyy) [....yyyy min]
310 9999 W 311 999 W Prime rate in ml/h (1...999) 312 1200 W Max. bolus rate in ml/h (1...1200) 313 10 W Max. bolus total in ml (1...999) 314 10 W Upstream occlusion sensitivity
316 7 W Downstream pressure limit "PrL" default
317 250 W Air bubble size [microliters] (50...1000) 318 20 W Drop-rate window center in [drops/ml] (10...65) 319 1000 W Correction factor ((actual/nominal)*1000)
350 20 W Air summation volume, n * 50 [microliters] (2...40) 351 4 W Air summation time, n * 8 [min] (8..64)
361 500 W Key [ON/OFF] delay
362 8 W Display brightness (2...15) 363 10 W Buzzer alarm volume (5...10) 368 300 W Battery discharge time incl. 15min pre-
369 5 W Automatic Menu fall back delay time in [s] (5...30)
Max. rate in (1/10)ml/h
(0...60, 60 is highest sensitivity)
value
refer to chapter 5.5
(additional key [STOP] if #4=1)
alarm in [min]
(1...9999)
(0...60)
(1...10)
(850...1150)
(0...3000)
(45...300)
14.164.A_A707en.SM.V1.01.doc ARGUS Medical AG 10.02.04 AS 9 / 38
PUMP CONFIGURATIONS
Add. Def. P Function Range
370 0 W Clock seconds (0...59) 371 0 W Clock minutes (0...59) 372 0 W Clock hours (0...23) 374 0 W Clock days (1...31) 375 0 W Clock months (1...12) 376 0 W Clock years (2000...2099)
380 0 Last failure number (F-XX) Refer to chapter 5.7 381 0 Last infusion rate at failure 382 0 . 383 0 . 384 0 . 385 0 . 386 0 . 387 0 . 388 0 Oldest failure number (F-XX) 389 0 Oldest infusion rate at failure
390 0 W Last service date (yyww, year and week) 391 0 2. last service date (yyww, year and week) 392 0 3. last service date (yyww, year and week) 393 0 W Service interval in months
(0 = disabled)
394 0 W Service interval in hours of
operation (0 = disabled)
395 0 W Own address for SCI
(0 = no address)
396 xxx W Serial number of the pump
(xxxx yyyy)
397 8yyy W Serial number of the pump
(xxxx yyyy)
399 707 Data xxxx -> clears protection key
(0...24)
(0...9999)
(0...127)
[xxxx....]
[....yyyy]
2.7. Special configuration options
1. Configuration of a reminder alarm for the safety standard check:
If a safety standard check reminder alarm is configured, the pump will display “CtrL” after power up when the time has come to perform a standard safety check. To configure a reminder alarm please follow next steps:
First the service interval has to be entered either in months or in hours of operation,
or both (address 393, 394).
Next the last service date has to be entered on address 390. Any value higher than 0
entered at the address 393 and/or 394 will release the reminder alarm after the ser­vice interval has elapsed (check also the correct settings of the internal clock).
2. PC configuration tool “AConfig”:
With this additional software the pump may be configured from a PC over the serial
port. This software may be available from your local distributor or ARGUS Medi­cal AG.
Caution: AConfig may only be used with software versions greater or equal to
1.01.
14.164.A_A707en.SM.V1.01.doc ARGUS Medical AG 10.02.04 AS 10 / 38
CONFIGURATION AND HISTORY PRINTOUT
3. CONFIGURATION AND HISTORY PRINTOUT
3.1. Introduction
CAUTION! The infusion pump has to be disconnected from the patient before and while the serial interface cable is connected to the pump.
The connection of the ARGUS 707 over the serial interface RS 232 can be done by con­necting the interface cable no.10.093 to the serial interface outlet of the serial COM port of a PC. A data transfer between ARGUS 707 and a PC can be done without any additional soft­ware running on the PC.
3.2. Settings of the terminal program
3.2.1. RS232 connection settings of the terminal program
Open the terminal program on the PC, e.g. HyperTerminal, which has been included in every Microsoft
Windows Operating System.
In the “Connection Description” menu, type for e.g. “ARGUS” and click OK.
Please make sure that the correct COM port of the PC has been selected.
Make the appropriate changes on the “COM Proper­ties” menu according to the right hand picture:
Click OK, then a connection to the pump should now be established.
14.164.A_A707en.SM.V1.01.doc ARGUS Medical AG 10.02.04 AS 11 / 38
CONFIGURATION AND HISTORY PRINTOUT
(
6
3.2.2. Capturing text from the pump transmitted
If you intent to print out the configuration or the history events transmitted from the pump, you should capture the text transmit­ted by the pump into a text file. To do this please select “Capture Text” in the hyper terminal’s menu.
A “File box” will open; type a descriptive name for the file, and then click “Start”. The terminal program starts to capture each text received over the serial interface.
Stop the capturing after you did the re­quired pump settings. To do this select “Stop” in the menu “Cap­ture Text” of the terminal program. You will then have a text file including the data transmitted by the pump.
3.3. Configuration printout
Connect the ARGUS 707 to your PC (see chap. 3.1) and setup the terminal program (see chap. 3.2). Start to capture text on the PC side before you switch the pump on.
By saving the transmitted configuration text, you may print it using any text program. See the following example of a configuration printout:
2. Select address 399 in the left hand display.
3. Enter value 3456 in the
1. Enter into the configuration mode
see chapter2.3).
.3.9.9.
right hand display.
345
4. Press the “START/STOP” key. The pump will now transfer the actual pump configuration as text.
14.164.A_A707en.SM.V1.01.doc ARGUS Medical AG 10.02.04 AS 12 / 38
+ 26 hidden pages