Argus 600S User manual

Service Manual for Syringe Pump
ARGUS 600 S
Made in Switzerland
(a member of the CODAN group)
14.194.A_A600 en.SM.V4.3X ARGUS Medical AG 28.03.06 / PJ 1 / 47
INTRODUCTION
TABLE OF CONTENTS
1. INTRODUCTION ...............................................................................................................5
1.1. General .................................................................................................................................. 5
2. PUMP CONFIGURATIONS...............................................................................................5
2.1. General .................................................................................................................................. 5
2.2. Interrogation mode (without ARGUS service)........................................................................ 6
2.3. Configuration mode (without ARGUS service) ...................................................................... 7
2.4. First activation of a PIN Code (write protection) .................................................................... 8
2.5. Changing an existing PIN code .............................................................................................9
2.6. Address list of the pump configuration (without ARGUS service)........................................ 10
2.7. Index list of the pump configuration (with ARGUS service) ................................................. 11
2.8. User syringe......................................................................................................................... 14
2.9. Medication list ...................................................................................................................... 15
3. Serial communication of the pump..............................................................................17
3.1. General ................................................................................................................................ 17
3.2. Serial communication protocol.............................................................................................17
4. ARGUS service...............................................................................................................18
4.1. General ................................................................................................................................ 18
4.2. ARGUS service – Configuration .......................................................................................... 19
4.3. ARGUS service - Toolbox.................................................................................................... 20
5. SOFTWARE UPDATES .................................................................................................. 24
5.1. General ................................................................................................................................ 24
5.2. Requirements for a software update.................................................................................... 24
5.3. Software update procedure .................................................................................................24
5.4. Safety aspects ..................................................................................................................... 26
6. MAINTENANCE ..............................................................................................................27
6.1. General ................................................................................................................................ 27
6.2. Recalibration........................................................................................................................ 27
6.3. Final calibration.................................................................................................................... 28
6.4. Pressure control and pump accuracy measurement ........................................................... 34
6.5. Rough alignments................................................................................................................ 35
6.6. Battery capacity calibration.................................................................................................. 37
6.7. Pump specifications............................................................................................................. 37
6.8. Fault codes and "CtrL" message ......................................................................................... 38
7. REPLACEMENT OF PARTS .......................................................................................... 39
7.1. General ................................................................................................................................ 39
7.2. Spare parts .......................................................................................................................... 41
8. WIRING DIAGRAMM ......................................................................................................44
9. BLOC SCHEMATIC ........................................................................................................45
10. SAFETY STANDARD CHECK........................................................................................46
11. REPAIR ORDER FORM..................................................................................................47
14.194.A_A600 en.SM.V4.3X ARGUS Medical AG 28.03.06 / PJ 2 / 47
INTRODUCTION
Clutch lever
Syringe barrel
Beak
holder
pushbutton switch
Syringe guide Syringe presence
Clamp lever
Display panel
Pump serial number
Extension line
Syringes from
10 – 60 ml
Drive unit
14.194.A_A600 en.SM.V4.3X ARGUS Medical AG 28.03.06 / PJ 3 / 47
INTRODUCTION
DC-connection & QUICK®100
Docking Station interface
Equipotential
plug
Staff alert Combination clamp
AC power
connection
14.194.A_A600 en.SM.V4.3X ARGUS Medical AG 28.03.06 / PJ 4 / 47
1. INTRODUCTION
1.1. General
IMPORTANT!
This service manual is intended for the exclusive use of authorized persons who have been trained by ARGUS Medical AG in the maintenance and repair of the ARGUS 600 Syringe pump.
The service manual is meant to be used together with the user manual. IMPORTANT!
ARGUS Medical AG shall not assume any responsibility for any manipulations which have been carried out on the ARGUS 600 Syringe pump by a non-authorized person.
CAUTION!
The ARGUS 600 Syringe pump may only be used with spare parts, accessories, con­sumables and syringes with Luer-Lock connections recommended by ARGUS Medical AG. The functional safety of the pump is not guaranteed if non approved materials are used. The safety of the patient may be endangered.
This manual contains the latest data available. It is subject to further modifications in accordance with technical improvements.
INTRODUCTION
2. PUMP CONFIGURATIONS
2.1. General
CAUTION!
The configuration possibilities with the “ARGUS service” PC utility tool and without PC
assistance constitute a modification of the pump and may only be carried out by au­thorized persons!
CAUTION!
After changing the configuration a function check and a control measurement has to be performed!
14.194.A_A600 en.SM.V4.3X ARGUS Medical AG 28.03.06 / PJ 5 / 47
PUMP CONFIGURATIONS
0
o
P
g
1
5
2.2. Interrogation mode (without ARGUS service)
With the interrogation mode you can read the present keypad configuration of the pump without the possibility to modify any configurations. For a complete overview,
please take the “ARGUS service” PC-tool!
1. To enter into the keypad interrogation mode, switch the pump on
while keeping the keys “MODE” and “START/STOP” pressed.
60
R
... .
Flashing decimal points indicate which display is ready to accept an input by the keys 100, 10 & 1.
In the keypad interrogation mode the left hand display shows the address and the right hand display shows the according value configured at this address. Please refer to
chapter 2.6. where the meanings of the addresses are explained.
To modify any configuration data you have to go into the configuration mode.
14.194.A_A600 en.SM.V4.3X ARGUS Medical AG 28.03.06 / PJ 6 / 47
3. Enter the requested address (see chapter 2.6.), e.g. 5. The pro-
grammed value of the address 5 appears in the right hand display.
4. To quit the interrogation mode press the “ON/OFF” key at least 2 seconds.
2. Press the “START/STOP” key.
2.3. Configuration mode (without ARGUS service)
-
0
p
o
g
The configuration mode permit you to modify the pump keypad configuration manually
using the keypad. Please refer to chapter 2.6. where the meaning of the addresses are
explained. To have access to all configuration options, please use the “ARGUS
service” PC-tool!
2.3.1. Step 1
1. To enter into the keypad configuration mode, switch the pump
on while keeping the keys “MODE” and “START/STOP” pressed.
PUMP CONFIGURATIONS
2. Press key “START”
60
2.3.2. Step 2
You have now access to all addresses in the list of chapter 2.6. Select therefore any address in the left display (see next page).
14.194.A_A600 en.SM.V4.3X ARGUS Medical AG 28.03.06 / PJ 7 / 47
3. Press key “MODE” 4. Press key “START” again.
Cod
6. Press key “START” to acknowledge the entered PIN code.
If the PIN was not accepted, the display will change back to 2.3.2. Point 5.
r
5. Enter the PIN Code. If no PIN code was configured, no entry is required
---
2.3.3. Step 3
5
0
o
P
g.0
8. Enter now the value on the right hand display. The
range of the value is given by the table in chapter 2.6.
10. To quit the configuration mode press the “ON/OFF” key at least 2 seconds. Changes in configuration become active, after the pump is switched on normally again.
Important remark:
Invalid values entered will be corrected automatically by the pump to the maxima or minima value allowed for the according address!
...
PUMP CONFIGURATIONS
7. After entering an address e.g. 5 press the key “MODE”. The flashing decimal points will change to the right hand display.
9. Press key “START/STOP” to acknowledge the entered value. The flashing decimal points change back to the left hand display.
2.4. First activation of a PIN Code (write protection)
The activation of a PIN code allows you to protect the configuration from unauthorized access. To activate the PIN code, enter into the configuration mode.
1. To enter into the configuration mode, switch the pump on
5. Press “MODE” key (Code “0” will be set) then “START” key to acknowledge the entered PIN code “0”.
14.194.A_A600 en.SM.V4.3X ARGUS Medical AG 28.03.06 / PJ 8 / 47
while keeping the keys “MODE” and “START/STOP” pressed.
60
R
3. Press key “MODE”
2. Press key “START”
4. Press key “START” again
0
0
7. Enter now the new PIN Code (max. 4 digit number).
Please remember this code, it will never be visible again!
e.g.
CAUTION!
After you switch the pump OFF and ON again you can enter into the configuration mode only, if you enter the correct PIN code.
e.g.
Please note: The interrogation mode can always be accessed without the PIN.
6. Press the “MODE” key. The flashing decimal points will change to the right hand display.
.
.
..
PUMP CONFIGURATIONS
8. Press key “START/STOP” to acknowledge the new PIN.
2.5. Changing an existing PIN code
Enter the configuration mode using present PIN, select add. “0” and set new code.
Enter actual PIN code and confirm with “START” key.
Press “MODE” key (#0). The flashing decimal points will change to the right hand display.
Enter the new PIN code and press the “START/STOP” key to acknowledge the entered code.
14.194.A_A600 en.SM.V4.3X ARGUS Medical AG 28.03.06 / PJ 9 / 47
PUMP CONFIGURATIONS
2.6. Address list of the pump configuration (without ARGUS service)
The following list declares the possible configuration options which can be performed on the pump keypad without using the PC.
All these options can also be configured by the PC-Software “ARGUS service”.
Address Default Function Unit Range
left display
1 2 No Key ON/OFF only at stop valid - 0=No / 1=Yes 2 11 Yes Recall of the last used ml/h rate ­3 19 Yes Buzzer at start ­4 44 Yes Automatic pressure release ­5 49 No Alarm acknowledge with key MODE -
100 361 5 Key ON/OFF delay time • 0.1 s 0 - 31 101 362 2 Display brightness level 1 - 3 102 363 10 Buzzer volume level 5 - 10 103 365 9 Default pressure limit • 100 mbar 2 - 12
200 368 495 Battery capacity (discharge time) min 45 - 615
399 - 600 Enter the calibration menu - 123
Index
PC
0=No / 1=Yes 0=No / 1=Yes 0=No / 1=Yes 0=No / 1=Yes
Note!
The address does not correspond with the index used by the “ARGUS service” tool.
right display
14.194.A_A600 en.SM.V4.3X ARGUS Medical AG 28.03.06 / PJ 10 / 47
PUMP CONFIGURATIONS
2.7. Index list of the pump configuration (with ARGUS service)
Index PC Add. Default Function Unit Range
1 No Run indication by running decimal point - No / Yes 2 1 No Key ON/OFF only at stop valid - No / Yes 3 No Rate change allowed only at STOP - No / Yes 4 No Key STOP delayed # 361 - No / Yes 5 No Second entry of rate # 3=Yes, # 9=No - No / Yes 6 No Static alarm (staff alerting system) - No / Yes 7 No Display elapsed time in run mode # 8=No - No / Yes 8 No Display remaining time # 7=Yes - No / Yes
9 No Rate change confirmation in stop mode - No / Yes 10 11 2 Yes Recall of ml/h rate at next power on # 9=No - No / Yes 12 No Recall of ml total at next power on - No / Yes 13 No Recall of ml inf. at next power on - No / Yes 14 No SBS (step by step function) - No / Yes 15 No Display VTBI (volume to be infused) - No / Yes 16 No Syringe type acknowledge at start - No / Yes 17 Yes KVO (KOR) enabled # 60 - No / Yes 18 19 3 No Buzzer at start - No / Yes 20 No Menu Clr (clear ml inf.) # 15=No, # 65 - No / Yes 21 22 23 Yes Menu PrL (pressure alarm limit) - No / Yes 24 Yes Menu CAP (battery capacity) - No / Yes 25 26 No Menu InF (ml inf. since last power on) - No / Yes 27 No Menu dLo (data lock) - No / Yes 28 No Menu Stb (stand by) - No / Yes 29 Yes Menu Med (medication name) - No / Yes 30 No Menu tM (timer alarm) - No / Yes 31 32 Yes Menu boL "bolu Man" / "bolu Auto" (bolus always possible) - No / Yes 33 Yes Menu boLr (bolus rate) # 32=Yes - No / Yes 34 Yes Menu tot (bolus total) # 32=Yes - No / Yes 35 No Display BOLUS-VTBI - No / Yes 36 37 38 Yes Automatic bolus application # 32, 34=Yes - No / Yes 39 No Bolus total to be reset after each auto bolus - No / Yes 40 41 No Clear ml/h after infusion completed - No / Yes 42 No Clear ml total after infusion completed # 41=Yes - No / Yes 43 Yes Syringe clamp diameter outside control - No / Yes 44 4 Yes Automatic pressure release after occlusion - No / Yes 45 Yes Pressure display ON (LED bar graph - 20/40/60/80/100%) - No / Yes 46 No Pressure display with indicator # 45=Yes - No / Yes 47 No Stand by- and battery pre alarm low volume - No / Yes 48 Yes Flashing numeric display at alarm - No / Yes 49 5 No Alarm acknowledge only with key MODE - No / Yes
14.194.A_A600 en.SM.V4.3X ARGUS Medical AG 28.03.06 / PJ 11 / 47
PUMP CONFIGURATIONS
55 Yes Med. disp. alternate with rate and ml inf. # 29=Yes - No / Yes
60 No KVO only after infusion completed - No / Yes
65 No Clear and continue # 15=No - No / Yes
75 No Select binder connector for serial interface - No / Yes
100 No
User syringe 10 ml
101 No BD Plastipak 102 No Braun Omnifix 103 Yes Codan 104 No Fresenius Injectomat 105 No Sherwood Monoject 106 No ONCE 107 No PIC Indolor 108 No Rymco 109 No Terumo 110 No Braun Injekt (#43=No) 111 No Chirana-Prema
120 No
User syringe 20 ml
121 No BD Plastipak 122 No Braun Omnifix 123 Yes Codan 124 No Sherwood Monoject 125 No ONCE 126 No Braun Perfusor 127 No Braun Inject 128 No Chirana-Prema 129 No Terumo 130 No Penta Ferte
140 No
User syringe 30 ml
141 No BD Plastipak 142 No Codan 143 No ONCE 144 No Braun Omnifix 145 No Terumo 146 No Penta Ferte
150 No
User syringe 50 ml
151 No BD Perfusion 152 No BD Plastipak 153 No Braun Omnifix 154 No Braun Perfusor 155 No Chirana-Prema 156 No Codan 157 Yes Codan Perfusion 158 No Dispomed 159 No Dipomed Perfusion 160 No Fresenius Injectomat
USEr -10-
b-d PL10
brn OF10
Cod -10­FrES In10 Mono -10­OnCE -10­PIC -10­ryco -10­tEru -10­brn In10 Chir -10-
USEr -20-
b-d PL20
brn OF20
Cod -20­Mono -20­OnCE -20-
brn PE20
brn In20 Chir -20­tEru -20-
PF -20-
USEr -30-
b-d PL30
Cod -30­OnCE -30-
brn OF30
tEru -30­PF -30-
USEr -50­ b-d PE50 b-d PL50
brn OF50
brn PE50
Chir -50-
Cod -50-
Cod PE50
dISP -50-
dISP PE50
FrES In50
10 ml 10 ml 10 ml 10 ml 10 ml 10 ml 10 ml 10 ml 10 ml 10 ml 10 ml 10 ml
20 ml 20 ml 20 ml 20 ml 20 ml 20 ml 20 ml 20 ml 20 ml 20 ml 20 ml
30 ml 30 ml 30 ml 30 ml 30 ml 30 ml 30 ml
50 ml
50 ml 50/60 ml 50/60 ml
50 ml 50/60 ml
50 ml
50 ml 50/60 ml
50 ml 50/60 ml
14.194.A_A600 en.SM.V4.3X ARGUS Medical AG 28.03.06 / PJ 12 / 47
PUMP CONFIGURATIONS
161 No Fresenius Perfusion 162 No Ivac 163 No JMS 164 No Sherwood Monoject 165 No PIC Indolor 166 No PIC Indolor Perfusion 167 No Rymco 168 No Terumo 169 No Disoprivan (ZENECA) 170 No ONCE 171 No Braun Proinjekt 172 No Penta Ferte
FrES PE50
IVAC -50-
JMS -50­Mono -50­PIC -50-
PIC PE50
ryco -50­tEru -50­dIPr -50­OnCE -50-
brn Pr50
PF -50-
50/60 ml 50/60 ml 50/60 ml 50/60 ml
50 ml 50 ml 50 ml
50/60 ml
50 ml 50 ml 50 ml 50 ml
310 300.0 Max. rate ml/h 1 - 300 311 300.0 Prime rate ml/h 1 - 300 312 300.0 Max. bolus rate ml/h 1 - 300 313 1.0 Max. total ml 1 - 10 314 315 61.0 Syringh length mm 45 - 70 316 16.0 Plunger length mm 12 - 30 317 16.2 Barrel diameter mm 15 - 19
user syringe
318 18.7 Clamp diameter
10 ml syringe
parameters
mm 15 - 25
319
320 500.0 Max. rate ml/h 1 - 500 321 500.0 Prime rate ml/h 1 - 500 322 500.0 Max. bolus rate ml/h 1 - 500 323 2.0 Max. total ml 1 - 20 324 325 69.5 Syringh length mm 50 - 80 326 16.8 Plunger length mm 12 - 30 327 21.4 Barrel diameter mm 20 - 24
user syringe
328 23.8 Clamp diameter
20 ml syringe
parameters
mm 20 - 30
329
330 500.0 Max. rate ml/h 1 - 500 331 500.0 Prime rate ml/h 1 - 500 332 500.0 Max. bolus rate ml/h 1 - 500 333 3.0 Max. total ml 1 - 25 334 335 82.0 Syringh length mm 70 - 90 336 16.0 Plunger length mm 12 - 30 337 24.0 Barrel diameter mm 22 - 26
user syringe
338 26.0 Clamp diameter
30 ml syringe
parameters
mm 20 - 30 339 340 999.9 Max. rate ml/h 1 - 999.9 341 999.9 Prime rate ml/h 1 - 999.9 342 800.0 Max. bolus rate ml/h 1 - 800.0 343 5.0 Max. total ml 1 - 25 344 345 90.0 Syringh length mm 70 - 100 346 16.5 Plunger length mm 12 - 50 347 29.4 Barrel diameter mm 28 - 34
user syringe
348 31.0 Clamp diameter
50 ml syringe
parameters
mm 25 - 37
14.194.A_A600 en.SM.V4.3X ARGUS Medical AG 28.03.06 / PJ 13 / 47
PUMP CONFIGURATIONS
361 100 5 Key ON/OFF delay time # 4 • 1/10 s 0 - 31 362 101 2 Display brightness level 1 - 3 363 102 7 Buzzer alarm volume level 5 - 10 364 365 103 9 Default pressure limit (PrL levels) 100mbar 2 - 12 366 1 Pressure display unit (mbar / mmHg / kPa / cmH2O / Psi) Enum 1 - 5 367 3 Time for near empty alarm min 1 - 15 368 200 495 Battery capacity (discharge time) min 45 - 615 369 5 Automatic menu fall back delay time sec 5 - 30
390 0 Last service date in year year 0 - 99 391 0 Last service date in month month 0 - 12 392 0 Last service date in day day 0 - 31 393 0 Service interval in months month 0 - 24 394 0 Service interval in hours of operation hour 0 -10000 395
396 0 Pump serial number xxx 6 yyy xxx 6 yyy
397 398
- 399 600 Enter the calibration menu / clears protection key - 123
518
522 No Allow change of ml total while infusing # 65=No - No / Yes 523 524 Display a department info text (after power up) char
Permissions for serial communication
2
0 = none, 1 = query only, 2 = parametrising, 3 = remote control
Enum 0 - 3
1-16 ASCII
Using the “ARGUS service” tool, the complete and detailed pump configuration can be done.
2.8. User syringe
The ARGUS 600 Syringe pump uses syringes from various manufacturers (see user manual list of recommended syringes). If you want to use any other brand you must be sure that the syringe is CE marked and is specified by the syringe manufacturer to be pressure resistant and/or safe to be used with infusion pumps, the syringe must be made out of 3 parts (barrel, plunger, sealing) and have Luer-Lock connection (same applies to extension lines). When all these points are met you are allowed to configure
your own "USER" syringe (one per size). Use the “ARGUS service” tool to enter the
syringe parameters into index 340 – 348 (50 ml syringe).
14.194.A_A600 en.SM.V4.3X ARGUS Medical AG 28.03.06 / PJ 14 / 47
PUMP CONFIGURATIONS
2.9. Medication list
2.9.1. General
To display medication names, index 29 (menu "MEd") must be set to "Yes". The se­lected medication name can be displayed also in alternate mode (rate, ml inf./ med. name) during infusion, for that set index 56 to “Yes”.
After enabled special function “Med” the following medication names can be selected via pump keypad (see user manual).
2.9.2. User medication
32 user medication names can be custom defined. Choose between capital and small letters for a better displayed medication name. Because of the 7-segment pump dis­play some characters maybe difficult to read.
Index
PC
561 UserM 1 User medication name 1 char 1 - 8 ASCII
562 UserM 2 User medication name 2 char 563 UserM 3 User medication name 3 564 UserM 4 User medication name 4 565 UserM 5 User medication name 5 566 UserM 6 User medication name 6 567 UserM 7 User medication name 7 568 UserM 8 User medication name 8 569 UserM 9 User medication name 9 570 UserM 10 User medication name 10 571 UserM 11 User medication name 11 572 UserM 12 User medication name 12 573 UserM 13 User medication name 13 574 UserM 14 User medication name 14 575 UserM 15 User medication name 15 576 UserM 16 User medication name 16 577 UserM 17 User medication name 17 578 UserM 18 User medication name 18 579 UserM 19 User medication name 19 580 UserM 20 User medication name 20 581 UserM 21 User medication name 21 582 UserM 22 User medication name 22 583 UserM 23 User medication name 23 584 UserM 24 User medication name 24 585 UserM 25 User medication name 25 586 UserM 26 User medication name 26 587 UserM 27 User medication name 27 588 UserM 28 User medication name 28 589 UserM 29 User medication name 29 590 UserM 30 User medication name 30 591 UserM 31 User medication name 31 592 UserM 32 User medication name 32
Default Function Unit Range
1 - 8 ASCII
char
1 - 8 ASCII 1 - 8 ASCII
char char
1 - 8 ASCII 1 - 8 ASCII
char
1 - 8 ASCII
char
1 - 8 ASCII
char char
1 - 8 ASCII 1 - 8 ASCII
char char
1 - 8 ASCII 1 - 8 ASCII
char
1 - 8 ASCII
char char
1 - 8 ASCII 1 - 8 ASCII
char
1 - 8 ASCII
char
1 - 8 ASCII
char
1 - 8 ASCII
char
1 - 8 ASCII
char
1 - 8 ASCII
char
1 - 8 ASCII
char
1 - 8 ASCII
char
1 - 8 ASCII
char
1 - 8 ASCII
char
1 - 8 ASCII
char
1 - 8 ASCII
char
1 - 8 ASCII
char
1 - 8 ASCII
char
1 - 8 ASCII
char
1 - 8 ASCII
char
1 - 8 ASCII
char
1 - 8 ASCII
char
14.194.A_A600 en.SM.V4.3X ARGUS Medical AG 28.03.06 / PJ 15 / 47
Loading...
+ 32 hidden pages