14.135.D_A600D en.SM.V1.50 Argus 10.08.04 / PJ2
Table of contents
IMPORTANT!
This service manual is intended for the exclusive use of authorized persons who have been trained
by ARGUS Medical AG in the maintenance and repair of the ARGUS 707 infusion pump.
The service manual is meant to be used together with the user manual.
IMPORTANT!
ARGUS Medical AG shall not assume any responsibility for any manipulations which have been
carried out on the ARGUS 707 infusion pump by a non-authorized person.
This manual contains the latest data available. It is subject to further modifications in accordance
with technical improvements.
Page
1Special key inputs and configurations4
2History and pump configurations printout13
3Fault finding17
4Replacement of parts18
5Safety standard check23
Mechanical drawings24
Wiring diagram29
Bloc schematic30
14.135.D_A600D en.SM.V1.50 Argus 10.08.04 / PJ3
1Special key inputs and configurations
1.1General
Caution: The configuration possibilities mentioned below constitute a modification of the pump and
may only be carried out by authorized persons.
If the decimal points are flashing in a display, this display is ready to accept an input by means of the keys
"100, 10, 1".
Note: Use the interrogation mode to check the present configuration without doing any modification!
1.2Interrogation mode
MODEMODE
STSTARTART
STSTOPOP
100100
STSTARTART
STSTOPOP
Hold both keys pressed
Important!
Do not release the keys "MODE" and "START/STOP"
ml
h
ml
h
ml total
ml inf.
h. min
ml total
ml inf.
h. min
Enter the requested
10
10
1
ml
h
0,10,1
address (eg. 52).
ml total
ml inf.
h. min
The programmed data of address 52 are displayed
Press the key at least 2 seconds.
14.135.D_A600D en.SM.V1.50 Argus 10.08.04 / PJ4
MODE
MODE
MODE
MODE
ART
OP
ART
OP
ART
OP
ART
OP
ART
OP
100
10
0,1
100
10
0,1
1.3Configuration mode without PIN code
MODE
STSTART
STSTOP
MODE
STSTART
STSTOP
MODE
STSTART
STSTOP
Hold both keys pressed
Important!
Do not release the keys "MODE" and "START/STOP"
ml
h
ml
h
ml
h
ml
h
ml total
ml inf.
h. min
ml total
ml inf.
h. min
ml total
ml inf.
h. min
ml total
ml inf.
h. min
STSTART
STSTOP
100
MODE
100
STSTART
STSTOP
ml total
ml inf.
h. min
ml total
ml inf.
h. min
ml total
ml inf.
h. min
ml total
ml inf.
h. min
ml total
ml inf.
h. min
10
10
ml
h
ml
h
1
0,1
ml
h
ml
h
1
0,1
ml
h
Enter the requested address (eg. 5).
Enter the requested data (eg. 1).
Confirm data entered (acknowledgement)
ml
h
ml total
ml inf.
h. min
14.135.D_A600D en.SM.V1.50 Argus 10.08.04 / PJ5
1.4Configuration mode: first input of PIN code
IMPORTANT !Remember to make a note of your code and keep it in a safe place.
3 “MODE”Switch over to the display (3)“ 0”“. . . 1.”
4 “START/STOP” Acknowledgement “Cod ”“- - - - ”
5 “100;10;1”Enter code in the display (3)“Cod ”“ 0 ”
6 “START/STOP” Acknowledgement “. . . 0.”“ 1 ”
7 “100;10;1”Enter requested address“A.A.A.A.”“X X X X”
Programmed data appear in the display (3)
8 “MODE”Switch over to the display (3)“ A A A A” “ X.X.X.X.”
9 “100;10;1”Enter requested data“ A A A A”“ Y.Y.Y.Y.”
10 “START/STOP” Acknowledgement. If the data are accepted, “A.A.A.A.”“ Y Y Y Y”
entry changes to the display (2)
11 “ON/OFF”End of the programming mode
14.135.D_A600D en.SM.V1.50 Argus 10.08.04 / PJ6
1.6List of the interrogation- and configuration functions:
CAUTION! Before you undertake a new configuration or replace the EPROM or the mainboard, printout the
pump configuration (see chap. 2.2). Afterwards you can reenter the old not write protected
values (download).
L = Length at 50, 30, 20, 10ml
B = Barrel ø outside
AdressWrite DefaultFunction
(#)prot.value
P = Plunger Length
C = Clamp ø
00 = NoPIN code active
10 = NoRun indication by running decimalpoint
20 = NoKey [On/Off] only at STOP valid
30 = NoRate change only at STOP valid
40 = NoKey [STOP] delayed (see #361)
50 = No2nd entry of rate (#3 = 1, #9 = 0), rate caculation disabled
60 = NoStatic alarm (staff alerting system)
70 = NoDisplay elapsed- or remaining time in run mode****
80 = NoSelect remaining time (#7 = 1)
90 = NoRate change confirmation in stop mode
111 = YesRecall last "ml/h" (rate) at next power on (#9 = 0)
120 = NoRecall last "ml total" (endvolume) at next power on
130 = NoRecall "ml inf" (volume infused) at next power on (#20 = 1)
140 = NoSBS Step By Step function
150 = NoDisplay VTBI (Volume To Be Infused)
160 = NoSyringe type acknowledge at start
171 = YesKVO (KOR), mode see #60
191 = YesBuzzer at start