Argus 600D User manual

Service Manual for
Syringe pump
green stream® SY-P
ARGUS 600 D
Made in Switzerland
This service manual is intended for the exclusive use of authorized persons who have been trained by ARGUS
Medical AG in the maintenance and repair of the infusion apparatus mentioned above.
ARGUS Medical AG shall not assume any responsibility for any manipulations which have been carried out
on the unit by a non-authorized person.
ARGUS Medical AG, CH-3627 Heimberg/Switzerland
(A member of the CODAN group)
14.135.D_A600D en.SM.V1.50 Argus 10.08.04 / PJ 1
ml
10
100
ml total ml inf. h. min
0,1
START
STOP
MODE
ALARM
1 2 3 4 5 6
BOLUS
ml h
100
10
1
C
7
8..11
1Display "Operation mode" 2Display "Infusion rate" 3Display "Total", Inf-, etc. 4 LED-bar "Pressure" 5Display "Alarm" 6Display "Alarm mode" 7 Key "ON/OFF" 8 Key "100" 9 Key "10" 10 Key "1" 11 Key "0,1" 12 Key "MODE" 13 Key "START/STOP"
ALARM
START
STOP
BOLUS
0,1
ml total ml inf. h. min
MODE
12 13 14 15 16 17 18 19
14 Syringe barrel holder 15 Key "BOLUS" 16 Syringe guide 17 Beak 18 Drive unit 19 Lever for clamp 20 Clutch lever 21 Combination clamp 22 Ext. 12VDC and interface RS 232 23 Staff alert 24 Line plug 25 Equipotential plug
20
14.135.D_A600D en.SM.V1.50 Argus 10.08.04 / PJ 2
Table of contents
IMPORTANT!
This service manual is intended for the exclusive use of authorized persons who have been trained by ARGUS Medical AG in the maintenance and repair of the ARGUS 707 infusion pump.
The service manual is meant to be used together with the user manual.
IMPORTANT!
ARGUS Medical AG shall not assume any responsibility for any manipulations which have been carried out on the ARGUS 707 infusion pump by a non-authorized person.
This manual contains the latest data available. It is subject to further modifications in accordance with technical improvements.
Page
1 Special key inputs and configurations 4
2History and pump configurations printout 13
3 Fault finding 17
4 Replacement of parts 18
5 Safety standard check 23
Mechanical drawings 24
Wiring diagram 29
Bloc schematic 30
14.135.D_A600D en.SM.V1.50 Argus 10.08.04 / PJ 3
1 Special key inputs and configurations
1.1 General
Caution: The configuration possibilities mentioned below constitute a modification of the pump and
may only be carried out by authorized persons.
If the decimal points are flashing in a display, this display is ready to accept an input by means of the keys "100, 10, 1".
Note: Use the interrogation mode to check the present configuration without doing any modification!
1.2 Interrogation mode
MODEMODE
STSTARTART
STSTOPOP
100100
STSTARTART
STSTOPOP
Hold both keys pressed
Important! Do not release the keys "MODE" and "START/STOP"
ml h
ml h
ml total ml inf. h. min
ml total ml inf. h. min
Enter the requested
10
10
1
ml h
0,10,1
address (eg. 52).
ml total ml inf. h. min
The programmed data of address 52 are displayed
Press the key at least 2 seconds.
14.135.D_A600D en.SM.V1.50 Argus 10.08.04 / PJ 4
MODE
MODE
MODE
MODE
ART
OP
ART
OP
ART
OP
ART
OP
ART
OP
100
10
0,1
100
10
0,1
1.3 Configuration mode without PIN code
MODE
STSTART
STSTOP
MODE
STSTART
STSTOP
MODE
STSTART
STSTOP
Hold both keys pressed
Important! Do not release the keys "MODE" and "START/STOP"
ml h
ml h
ml h
ml h
ml total ml inf. h. min
ml total ml inf. h. min
ml total ml inf. h. min
ml total ml inf. h. min
STSTART
STSTOP
100
MODE
100
STSTART
STSTOP
ml total ml inf. h. min
ml total ml inf. h. min
ml total ml inf. h. min
ml total ml inf. h. min
ml total ml inf. h. min
10
10
ml h
ml h
1
0,1
ml h
ml h
1
0,1
ml h
Enter the requested address (eg. 5).
Enter the requested data (eg. 1).
Confirm data entered (acknowledgement)
ml h
ml total ml inf. h. min
14.135.D_A600D en.SM.V1.50 Argus 10.08.04 / PJ 5
1.4 Configuration mode: first input of PIN code IMPORTANT ! Remember to make a note of your code and keep it in a safe place.
Key Description Display (2) Display (3)
1 ”MODE” & Keep both keys pressed before switching
”START/STOP” the unit on. “ 600” “prog”
2 “START/STOP” Acknowledgement (write protection is inactive) “. . . 0.” “ 0”
3 “MODE” Switch over to the display (3) “ 0” “. . . 0.”
4 “START/STOP” Acknowledgement “Cod ” “- - - - ”
5 “MODE” Switch over to the display (3) “Cod ” “ 0”
6 “START/STOP” Acknowledgement “. . . 0.” “ 0”
7 “MODE” Switch over to the display (3) “ 0” “. . . 0.”
8 “100;10;1” Enter 1 to 4 digit code “ 0” “ C.C.C.C.”
9 “START/STOP” Acknowledgement (write protection is active) “. . . 0.” “ 1”
Code is never visible
10 “ON/OFF” End of the programming mode
1.5 Programming mode with active write protection (code) CAUTION! Only the code holder can carry out modifications when the write protection is active.
Key Description Display (2) Display (3)
1 ”MODE” & Keep both keys pressed before switching
”START/STOP” the unit on. “600” “prog”
2 “START/STOP” Acknowledgement (write protection is active) “. . . 0.” “ 1 ”
3 “MODE” Switch over to the display (3) “ 0” “. . . 1.”
4 “START/STOP” Acknowledgement “Cod ” “- - - - ”
5 “100;10;1” Enter code in the display (3) “Cod ” “ 0 ”
6 “START/STOP” Acknowledgement “. . . 0.” “ 1 ”
7 “100;10;1” Enter requested address “A.A.A.A.” “X X X X”
Programmed data appear in the display (3)
8 “MODE” Switch over to the display (3) “ A A A A” “ X.X.X.X.”
9 “100;10;1” Enter requested data “ A A A A” “ Y.Y.Y.Y.”
10 “START/STOP” Acknowledgement. If the data are accepted, “A.A.A.A.” “ Y Y Y Y”
entry changes to the display (2)
11 “ON/OFF” End of the programming mode
14.135.D_A600D en.SM.V1.50 Argus 10.08.04 / PJ 6
1.6 List of the interrogation- and configuration functions:
CAUTION! Before you undertake a new configuration or replace the EPROM or the mainboard, printout the
pump configuration (see chap. 2.2). Afterwards you can reenter the old not write protected values (download).
L = Length at 50, 30, 20, 10ml
B = Barrel ø outside
Adress Write Default Function
(#) prot. value
P = Plunger Length
C = Clamp ø
00 = No PIN code active 10 = No Run indication by running decimalpoint 20 = No Key [On/Off] only at STOP valid 30 = No Rate change only at STOP valid 40 = No Key [STOP] delayed (see #361) 50 = No 2nd entry of rate (#3 = 1, #9 = 0), rate caculation disabled 60 = No Static alarm (staff alerting system) 70 = No Display elapsed- or remaining time in run mode **** 80 = No Select remaining time (#7 = 1) 90 = No Rate change confirmation in stop mode
11 1 = Yes Recall last "ml/h" (rate) at next power on (#9 = 0) 12 0 = No Recall last "ml total" (endvolume) at next power on 13 0 = No Recall "ml inf" (volume infused) at next power on (#20 = 1) 14 0 = No SBS Step By Step function 15 0 = No Display VTBI (Volume To Be Infused) 16 0 = No Syringe type acknowledge at start 17 1 = Yes KVO (KOR), mode see #60 19 1 = Yes Buzzer at start
20 0 = No Menu "Clr" (Clear “ml inf.”) (#15 = 0) **** 23 1 = Yes Menu: "Prl" (Pressure alarm limit) **** 24 1 = Yes Menu: "CAP" (Battery capacity) **** 26 0 = No Menu: "InF" (ml infused since last power up) **** 27 0 = No Menu: "dLo" (Data-lock) **** 28 0 = No Menu: "Stb" (Stand-by) **** 29 0 = No Menu: "MEd" (Medication) ****
30 0 = No Menu: "tM" (Timer alarm) **** 32 1 = Yes Bolus (Prime always possible) 33 1 = Yes Menu: "boLr" (Bolus rate) (#32 = 1) **** 34 1 = Yes Menu: "tot" (Bolus total) (#32 = 1) **** 35 0 = No Display bolus VTBI instead bolus infused 38 1 = Yes Bolus application automatic (#34 = 1; #32 = 1) 39 0 = No Bolus total to be reset after each auto bolus
40 0 = No Demo mode (all menus enabled) **** 41 0 = No Clear "ml/h" after infusion completed 42 0 = No Clear "ml total" after inf. completed (#41 = 1) 43 1 = Yes Syringe clamp diameter outside control 44 1 = Yes Automatic pressure release after occlusion 45 1 = Yes Pressure display 20/40/60/80/100 % (Bargraph ON) 46 0 = No Bargraph with indicator (25% steps, #45 = 1) 47 0 = No Standby- and battery prealarm low volume 48 1 = Yes Flashing numeric display at alarm 49 0 = No Alarm acknowledge only with key [MODE]
55 1 = Yes Med. disp. alternate with rate and ml inf.(#29=1)
14.135.D_A600D en.SM.V1.50 Argus 10.08.04 / PJ 7
Adress Write Default Function
(#) prot. value
60 0 = No KVO (KOR), only after infusion completed 65 0 = No Clear and continue (#15 = 0)
100 0 = No User 10ml [USEr -10-] 101 0 = No B-D Plastipak 10ml [b-d PL10] 102 0 = No Braun Omnifix 10ml [brn OF10] 103 1 = Yes Codan 10ml [Cod -10-] 104 0 = No Fresenius Injectomat 10ml [FrES In10] 105 0 = No Sheerwood Monoject 10ml [Mono -10-] 106 0 = No ONCE 10ml [OnCE -10-] 107 0 = No PIC Indolor 10ml [PIC -10-] 108 0 = No Rymco 10ml [ryco -10-] 109 0 = No Terumo 10ml [tEru -10-] 1100 = No Braun Injekt 10ml (#43=0) [brn In10]
1110 = No Chirana-Prema 10 ml [Chir -10-]
120 0 = No User 20ml [USEr -20-] 121 0 = No B-D Plastipak 20ml [b-d PL20] 122 0 = No Braun Omnifix 20ml [brn OF20] 123 1 = Yes Codan 20ml [Cod -20-] 124 0 = No Sheerwood Monoject 20ml [Mono -20-] 125 0 = No ONCE 20ml [OnCE -20-] 126 0 = No Braun Perfusor 20ml [brn PE20] 127 0 = No Braun Injekt 20ml [brn In20] 128 0 = No Chirana-Prema 20ml [Chir -20-] 129 0 = No Terumo 20 ml [tEru -20-]
140 0 = No User 30ml [USEr -30-] 141 0 = No B-D Plastipak 30ml [b-d PL30] 142 0 = No Codan 30ml [Cod -30-] 143 0 = No ONCE 30ml [OnCE -30-] 144 0 = No Braun Omnifix 30ml [brn OF30] 145 0 = No Terumo 30 ml [tEru -30-]
150 0 = No User 50ml [USEr -50-] 151 0 = No B-D Perfusion 50ml [b-d PE50] 152 0 = No B-D Plastipak 50/60ml [b-d PL50] 153 0 = No Braun Omnifix 50/60ml [brn OF50] 154 0 = No Braun Perfusor 50ml [brn PE50] 155 0 = No Chirana-Prema 50/60ml [Chir -50-] 156 0 = No Codan 50ml [Cod -50-] 157 1 = Yes Codan Perfusion 50ml [Cod PE50] 158 0 = No Dispomed 50/60ml [dISP -50-] 159 0 = No Dispomed Perfusion 50ml [dISP PE50]
160 0 = No Fresenius Injectomat 50/60ml [FrES In50] 161 0 = No Fresenius Perfusion 50/60ml [FrES PE50] 162 0 = No Ivac 50/60ml [IVAC -50-] 163 0 = No JMS 50/60ml [JMS -50-] 164 0 = No Sheerwood Monoject 50/60ml [Mono -50-] 165 0 = No PIC Indolor 50ml [PIC -50-] 166 0 = No PIC Indolor Perfusion 50ml [PIC PE50] 167 0 = No Rymco 50ml [ryco -50-] 168 0 = No Terumo 50/60ml [tEru -50-] 169 0 = No Disoprivan 50ml (ZENECA) [dIPr -50-]
170 0 = No ONCE 50ml [OnCE -50-] 171 0 = No Braun Proinjekt 50ml [brn Pr50]
14.135.D_A600D en.SM.V1.50 Argus 10.08.04 / PJ 8
Adress Write Default Function
(#) prot. value
200 0 Medication enable value 0...8191 201 0 Medication enable value 0...8191 202 0 Medication enable value 0...8191 203 0 Medication enable value 0...8191 204 0 Medication enable value 0...8191 205 0 Medication enable value 0...8191 206 0 Medication enable value 0...8191 207 0 User medication enable value 0...8191 208 0 User medication enable value 0...7
220 5383 Display digit 1 & 2, user medication 1 0...9999 221 6982 Display digit 3 & 4, user medication 1 0...9999 222 4500 Display digit 5 & 6, user medication 1 0...9999 223 17 Display digit 7 & 8, user medication 1 0...9999 224 5383 Display digit 1 & 2, user medication 2 0...9999 225 6982 Display digit 3 & 4, user medication 2 0...9999 226 4500 Display digit 5 & 6, user medication 2 0...9999 227 18 Display digit 7 & 8, user medication 2 0...9999 228 5383 Display digit 1 & 2, user medication 3 0...9999 229 6982 Display digit 3 & 4, user medication 3 0...9999
230 4500 Display digit 5 & 6, user medication 3 0...9999 231 19 Display digit 7 & 8, user medication 3 0...9999 232 5383 Display digit 1 & 2, user medication 4 0...9999 233 6982 Display digit 3 & 4, user medication 4 0...9999 234 4500 Display digit 5 & 6, user medication 4 0...9999 235 20 Display digit 7 & 8, user medication 4 0...9999 236 5383 Display digit 1 & 2, user medication 5 0...9999 237 6982 Display digit 3 & 4, user medication 5 0...9999 238 4500 Display digit 5 & 6, user medication 5 0...9999 239 21 Display digit 7 & 8, user medication 5 0...9999
240 5383 Display digit 1 & 2, user medication 6 0...9999 241 6982 Display digit 3 & 4, user medication 6 0...9999 242 4500 Display digit 5 & 6, user medication 6 0...9999 243 22 Display digit 7 & 8, user medication 6 0...9999 244 5383 Display digit 1 & 2, user medication 7 0...9999 245 6982 Display digit 3 & 4, user medication 7 0...9999 246 4500 Display digit 5 & 6, user medication 7 0...9999 247 23 Display digit 7 & 8, user medication 7 0...9999 248 5383 Display digit 1 & 2, user medication 8 0...9999 249 6982 Display digit 3 & 4, user medication 8 0...9999
250 4500 Display digit 5 & 6, user medication 8 0...9999 251 24 Display digit 7 & 8, user medication 8 0...9999 252 5383 Display digit 1 & 2, user medication 9 0...9999 253 6982 Display digit 3 & 4, user medication 9 0...9999 254 4500 Display digit 5 & 6, user medication 9 0...9999 255 25 Display digit 7 & 8, user medication 9 0...9999 256 5383 Display digit 1 & 2, user medication 10 0...9999 257 6982 Display digit 3 & 4, user medication 10 0...9999 258 4500 Display digit 5 & 6, user medication 10 0...9999 259 1716 Display digit 7 & 8, user medication 10 0...9999
260 5383 Display digit 1 & 2, user medication 11 0...9999 261 6982 Display digit 3 & 4, user medication 11 0...9999 262 4500 Display digit 5 & 6, user medication 11 0...9999 263 1717 Display digit 7 & 8, user medication 11 0...9999 264 5383 Display digit 1 & 2, user medication 12 0...9999 265 6982 Display digit 3 & 4, user medication 12 0...9999 266 4500 Display digit 5 & 6, user medication 12 0...9999 267 1718 Display digit 7 & 8, user medication 12 0...9999 268 5383 Display digit 1 & 2, user medication 13 0...9999 269 6982 Display digit 3 & 4, user medication 13 0...9999
14.135.D_A600D en.SM.V1.50 Argus 10.08.04 / PJ 9
Adress Write Default Function
(#) prot. value
270 4500 Display digit 5 & 6, user medication 13 0...9999 271 1719 Display digit 7 & 8, user medication 13 0...9999 272 5383 Display digit 1 & 2, user medication 14 0...9999 273 6982 Display digit 3 & 4, user medication 14 0...9999 274 4500 Display digit 5 & 6, user medication 14 0...9999 275 1720 Display digit 7 & 8, user medication 14 0...9999 276 5383 Display digit 1 & 2, user medication 15 0...9999 277 6982 Display digit 3 & 4, user medication 15 0...9999 278 4500 Display digit 5 & 6, user medication 15 0...9999 279 1721 Display digit 7 & 8, user medication 15 0...9999
280 5383 Display digit 1 & 2, user medication 16 0...9999 281 6982 Display digit 3 & 4, user medication 16 0...9999 282 4500 Display digit 5 & 6, user medication 16 0...9999 283 1722 Display digit 7 & 8, user medication 16 0...9999
306 X Infused sum in ml (xxxxyyyy) [xxxx . . . . ml]
307 X Infused sum in ml (xxxxyyyy) [. . . . yyyy ml]
308 X Operating time in min (xxxxyyyy) [xxxx . . . . ml]
309 X Operating time in min (xxxxyyyy) [. . . . yyyy ml]
310 3000 Maxima rate in 1/10ml/h (1.0...300.0 ml/h) 10....3000
311 3000 Prime rate in 1/10ml/h (1.0...300.0 ml/h) 10....3000
312 3000 Max. bolus rate in 1/10ml/h (1.0...300.0 ml/h) 10....3000
313 10 Max. bolus total in 1/10ml (1.0...10.0 ml) 10....100
315 610 *Syringe length (L) in 1/10mm at 10 ml volume 450....800
10 ml
316 160 *Syringe plunger length (P) in 1/10mm at end of infusion 120....300
317 162 *Syringe barrel diameter outside (B) in 1/10mm 150....190
318 187 *Syringe clamp diameter outside (C) in 1/10mm 150....250
320 5000 Maxima rate in 1/10ml/h (1.0...500.0 ml/h) 10....5000
321 5000 Prime rate in 1/10ml/h (1.0...500.0 ml/h) 10....5000
322 5000 Max. bolus rate in 1/10ml/h (1.0...500.0 ml/h) 10....5000
323 20 Max. bolus total in 1/10ml (1.0...20.0 ml) 10....200
325 695 *Syringe length (L) in 1/10mm at 20 ml volume 500....900
20 ml
326 168 *Syringe plunger length (P) in 1/10mm at end of infusion 120....300
327 214 *Syringe barrel diameter outside (B) in 1/10mm 200....240
328 238 *Syringe clamp diameter outside (C) in 1/10mm 200....300
330 5000 Maxima rate in 1/10ml/h (1.0...500.0 ml/h) 10....5000
331 5000 Prime rate in 1/10ml/h (1.0...500.0 ml/h) 10....5000
332 5000 Max. bolus rate in 1/10ml/h (1.0...500.0 ml/h) 10....5000
333 30 Max. bolus total in 1/10ml (1.0...25.0 ml) 10....250
335 820 *Syringe length (L) in 1/10mm at 30 ml volume 700....1000
30 ml
336 160 *Syringe plunger length (P) in 1/10mm at end of infusion 120....300
337 240 *Syringe barrel diameter outside (B) in 1/10mm 220....260
338 260 *Syringe clamp diameter outside (C) in 1/10mm 200....300
340 9999 Maxima rate in 1/10ml/h (1.0...999.9 ml/h) 10....9999
341 9999 Prime rate in 1/10ml/h (1.0...999.9 ml/h) 10....9999
342 9999 Max. bolus rate in 1/10ml/h (1.0...999.9 ml/h) 10....9999
343 50 Max. bolus total in 1/10ml (1.0...25.0 ml) 10....250
345 900 *Syringe length (L) in 1/10mm at 50 ml volume 700....1200
50 ml
346 165 *Syringe plunger length (P) in 1/10mm at end of infusion 120....500
347 294 *Syringe barrel diameter outside (B) in 1/10mm 280....340
348 310 *Syringe clamp diameter outside (C) in 1/10mm 250....370
* ) only active for the respective syringe type "User 10ml" (#100), "User 20ml" (#120), "User 30ml" (#140), "User 50ml" (#150).
14.135.D_A600D en.SM.V1.50 Argus 10.08.04 / PJ 10
361 500 Key [ON/OFF] delay in ms, (additional key [STOP] if #4 = 1) 0...3000 362 2 Display brightness 1...3 363 10 Buzzer alarm volume 5...10 365 9 Pressure alarm level (n*100mbar) 2...12 367 3 Time for near empty alarm 1...15 min 368 120 Battery discharge time (min) 45...615 369 5 Automatic menu fall back delay time 5...30 s
370 Clock seconds 00...59 371 Clock minutes 00...59 372 Clock hours 00...23 374 Clock days 01...31 375 Clock months 01...12 376 Clock years 2000...2099
380 X Last failure number (F-XX) 381 X Last infusion rate at failure 382 X 2. last failure number (F-XX) 383 X 2. last infusion rate at failure 384 X 3. last failure number (F-XX) 385 X 3. last infusion rate at failure 386 X 4. last failure number (F-XX) 387 X 4. last infusion rate at failure 388 X 5. last failure number (F-XX) 389 X 5. last infusion rate at failure
390 0 Last Service-date (year and week) yyww 391 X 2. last service-date 392 X 3. last service-date 393 0 Service interval in months (1...24, 0 = disabled) 394 0 Service interval in hours (1...9999, 0 = disabled) 395 0 Own adress for SCI (0 = no adress, or 1...127) 0...127
396 0 Inventory-no. of the pump (xxxx yyyy) [xxxx . . . .]
397 0 Inventory-no. of the pump (xxxx yyyy) [ . . . . yyyy]
399 X 600 Data xxxx -> clears protection key
14.135.D_A600D en.SM.V1.50 Argus 10.08.04 / PJ 11
1.7 Medication list
To display medication names, address 29 (menu "MEd") must be set to "1".
Using the special function "MEd", the following medication names can be displayed. Refer to the EXCEL file supplied by ARGUS or your local distributor to determine medication enable values (adress 200 ... 208).
number
HEX medication
DEC medication
number
Medication name
00 00 "Med " 01 01 "Actilyse" 02 02 "Adrena0.1" 03 03 "Adrena0.2" 04 04 "Ajmalin " 05 05 "Alfentan" 06 06 "Alupent " 07 07 "Ambroxol" 08 08 "Amiodaro" 09 09 "Amphoter" 10 0A "Aprotini" 11 0B "Atracuri" 12 0C "Bretyliu" 13 0D "Bupivaci" 14 0E "Ceruleti" 15 0F "Clonidin" 16 10 "Diltiaze" 17 11 "Dobutami" 18 12 "Dopamine" 19 13 "Dopexami" 20 14 "Esmolol " 21 15 "Fentanyl" 22 16 "Flecaini" 23 17 "Fluimuci" 24 18 "Flumazen" 25 19 "Furosemi" 26 1A "Glucos30"
number
HEX medication
DEC medication
number
Medication name
27 1B "Glucos 5" 28 1C "Heparin " 29 1D "Hydrocor" 30 1E "Insulin " 31 1F "Isoprena" 32 20 "KCL " 33 21 "Ketamin " 34 22 "Labetalo" 35 23 "Lidocain" 36 24 "Liothyro" 37 25 "Magnesiu" 38 26 "Midazola" 39 27 "Milrinon" 40 28 "Morphin " 41 29 "Nacl 0.9" 42 2A "Nalbuphi" 43 2B "Naloxone" 44 2C "Neostigm" 45 2D "Nicardip" 46 2E "Nifedipi" 47 2F "Nimodipi" 48 30 "Nitropru" 49 31 "Noradren" 50 32 "Omeprazo" 51 33 "Pancuron" 52 34 "Pentoxit" 53 35 "Phentola"
number
HEX medication
DEC medication
number
Medication name
54 36 "Phenylep" 55 37 "Procaina" 56 38 "Propafen" 57 39 "Propofol" 58 3A "Rapilysi" 59 3B "Remifent" 60 3C "Risordan" 61 3D "Ropivaca" 62 3E "Salbutam" 63 3F "Somatost" 64 40 "Streptok" 65 41 "Sufentan" 66 42 "Terbutal" 67 43 "Theopyli" 68 44 "Thiopent" 69 45 "Tirofiba" 70 46 "Trinitri" 71 47 "Urapidil" 72 48 "Urokinas" 73 49 "Vasopres" 74 4A "Vecuroni" 75 4B "Verapami" 76 4C "--------" 77 4D "--------" 78 4E "--------" 79 4F "--------" 80 50 "--------"
number
HEX medication
DEC medication
number
Medication name
81 51 "--------" 82 52 "--------" 83 53 "--------" 84 54 "--------" 85 55 "--------" 86 56 "--------" 87 57 "--------" 88 58 "--------" 89 59 "--------" 90 5A "--------" 91 5B "UserM 1" 92 5C "UserM 2" 93 5D "UserM 3" 94 5E "UserM 4" 95 5F "UserM 5" 96 60 "UserM 6" 97 61 "UserM 7" 98 62 "UserM 8" 99 63 "UserM 9" 00 64 "UserM 10" 101 65 "UserM 11" 102 66 "UserM 12" 103 67 "UserM 13" 104 68 "UserM 14" 105 69 "UserM 15" 106 6A "UserM 16"
16 user medication names can be custom defined (UserM 1 ... UserM 16). Refer to the EXCEL file for easy definition.
1.8 Special configuration options
- Configuration of a reminder alarm for the safety standard check: First the service interval has to be configured either in months or in hours of operation, or both (addresses 393, 394). Next the last service date has to be entered on address 390. Any value greater than 0 entered at the adresses 393 and/or 394 will release the reminder alarm after the set service interval has elapsed (check also the correct settings of the internal clock).
- PC configuration tool "AConfig": With this additional software the pump may be configured from a PC over the serial port. This software may be available from your local distributor or our service department.
After changing the configuration a function check and a control measurement has to be made!
14.135.D_A600D en.SM.V1.50 Argus 10.08.04 / PJ 12
2 History and pump configuration printout
2.1 Connecting of the ARGUS600D to the serial interface
Caution: The infusion pump must be disconnected from the patient
over the serial interface is made!
A connection of the ARGUS600D to a computer is useful to the read the present configuration or history of the pump. Even a simple monitoring of the pump can be done over the serial interface RS-232. The connection of the infusion pump with your computer over the interface can be done by connecting the interface cable (part 10.058) and the following steps:
- Connect the RS-232 interface cable to outlet (22) of the infusion pump and to the serial port of your PC. Note in which port (COM1 or COM2) you have pluged in.
-Start your terminal program on your computer. A simple terminal program, e.g. "Hyper Terminal" is included in every MS-Windows 9x and Windows NT systems, but must be installed.
- Be sure that you have selected the right serial port (COM1 or COM2) and set the following communication parameters:
Bits per second: 4800 Baud Data bits: 8 bits Parity: None Stop bits: 1 bit Protocol: None
- Go to the next step in one of the further chapters, depending on your intention.
2.2 Configuration printout
before any connection
- Switch the pump on while keeping the keys "MODE" and "START/STOP" pressed and go in the configura­tion mode.
- Select address 399 on the left hand display.
-Start recording text received over the serial interface, e.g. by selecting "Capture text..." in the menu of the Hyper terminal. A text file which contains the present configuration printout will now be generated.
- Enter the data 3456 on the right display of the pump.
-Press the "START/STOP" key.
- The pump will now transfer the present configuration of the pump in the format mentioned below.
-Stop the recording of the text received over the serial interface; this will also close the generated text file.
- The generated text file can be opened and printed out by any text program.
Pump configuration printout (sample):
/***** Configuration profile *****/ Wed 19-Jan-2002 11:29:55
Pump type : ARGUS600D Inventory number : 0000 0000 Software release : V0.93 (990819-6D5C) Infused sum : 230ml Operating time : 5h32min Last service date : 2000 week 12
00=0 50=0 100=0 150=0 200=0400 250=0000 300=0000 350=0000 01=1 51=0 101=0 151=0 201=0000 251=0000 301=0000 351=0000 02=0 52=0 102=0 152=0 202=0000 252=0000 302=0000 352=0000 03=0 53=1 103=1 153=0 203=0000 253=0000 303=0000 353=0000 etc.
14.135.D_A600D en.SM.V1.50 Argus 10.08.04 / PJ 13
2.3 History printout
- Switch the pump on while keeping the keys "MODE" and "START/STOP" pressed and go in the configu­ration mode.
- Select address 399 on the left display of the pump.
-Start recording text received over the serial interface, e.g. by selecting "Recording text..." in the menu of the Hyper terminal. A text file which contains the history printout will now be generated.
- Enter the data 4567 on the right display of the pump.
-Press the "START/STOP" key.
- The pump will now transfer the last events registered on the pump in the format mentioned below.
-Stop the recording of the text received over the serial interface; this will also close the generated text file.
- The generated text file can be opened and printed out by any text program.
History printout (sample)
/***** History *****/ Mon 06-Mar-2000 08:42:44
Pump off Wed 23-Feb-2000 11:54:38 Rate = 123.0ml/h Syringe = 156 Total = 50.0ml PrLimit = 900mbar Infsum = 5.4ml Status = 0x0000
Pump on Wed 09-Feb-2000 15:01:58 Rate = 10.0ml/h Syringe = 156 Total = 50.0ml PrLimit = 600mbar Infsum = 9.0ml Status = 0x0000 etc.
The possible messages are: Battery defective Battery low prealarm Battery low, pump stop Bolus start Bolus stop External power off External power on Occlusion, pump stop PC configuration done PC configuration failure
PrLimit change Pump has detected failure Pump off Pump on Pump start Pump stop (KVO) Rate change Enter setup mode Exit setup mode Infsum cleared
Syringe barrel, pump stop Syringe clutch, pump stop Syringe empty, pump stop Syringe near empty Syringe switch, pump stop Timer alarm, pump stop (KVO) Total volume reached, pump stop (KVO) Datalock on Datalock off Pump start, ext. changed parameter
14.135.D_A600D en.SM.V1.50 Argus 10.08.04 / PJ 14
2.4 Monitoring of the ARGUS600D Caution: The monitoring of the infusion pump ARGUS600D over the serial interface of a PC is intended for
demonstration purposes only; any connection with patients has not been tested under the conditions of EN 60601-1 and are not allowed.
- Switch the pump on with an inserted filled infusion set.
- Enter one of the following command directly in your terminal window or transmit the corresponding
ASCII code over your own monitoring program. A short sample of a monitoring session is mentioned at the end
of this chapter. Command Keystrokes in terminal ASCII code Description ENQ Ctrl+E 05H Send status (see format below)
SO Ctrl+N 0EH Sets pump in remote mode STX Ctrl+B 02H Start of data entry (see format below)* ‘data‘ Data - Data* ETX Ctrl+C 03H End of data entry* DC2 Ctrl+R 12H Start infusion* DC4 Ctrl+T 14H Stop infusion* SI Ctrl+O 0FH Sets pump in local mode ACK Ctrl+F 06H Alarm suppression (2min)* FS Ctrl+\ 1CH Toggle "KVO mode"* BEL Ctrl+G 07H Toggle "Buzzer at start mode"* CAN Ctrl+X 18H No start test’s at next start* ESC Ctrl+[ 1BH Next character following ESC ("Ctrl+[") will select
the pump with address "addr", if more than one is
DC1 Ctrl+Q 11H Enquiry of inventory number (address 396/397)
connected to the serial interface*
"addr" - 0-127 Address (must be the same as in the pump
Format of "data" entry: STX 0 1 2 0 0 2 0 0 0 0 ETX
Rate 1E1
Rate 1E2
Format of "status", which will be returned by the pump after typing "Ctrl+E" in the terminal:
Rate 1E0
Rate 1E-1
configuration on address 395)* * Only valid in remote mode
Total 1E1
Total 1E2
Total 1E0
Total 1E-1
Med. hex lo digit
Med. hex hi digit
STX 0 1 2 0 0 2 0 0 0 0 1 0 0 9 0 0 0 0 A B C D E ETX
PrL 1E1
Rate 1E2
Rate 1E1
Rate 1E0
Rate 1E-1
Total 1E2
Total 1E1
Total 1E0
Total 1E-1
Infused sum 1E2
Infused sum 1E1
PrL 1E3
PrL 1E2
Infused sum 1E0
Infused sum 1E-1
PrL 1E0
Med. hex hi digit
Med. hex lo digit
Statusbyte-1
Statusbyte-2
Statusbyte-3
Statusbyte-4
Statusbyte-5
14.135.D_A600D en.SM.V1.50 Argus 10.08.04 / PJ 15
Format statusbyte-1:
Format statusbyte-2:
P 1 X X X X X X
Always low
Always high
Pump type (1=A600)
Format statusbyte-3:
Battery active
Buzzer at start
P 1 X X X X X X
Reserved
Always low
Always high
Remote mode active
Global Alarm
Battery defective
Battery low prealarm
Battery low, pump stop
KVO active
Pump stop (KVO)
Syringe global alarm
P 1 X X X X X X
Always low
Always high
Occlusion, pump stop
Syringe barrel, pump stop
Syringe clamp, pump stop
Format statusbyte-4:
Syringe empty, pump stop
Syringe clutch, pump stop
P 1 X X X X X X
Reserved
Always low
Always high
Bolus active
Data lock active
Syringe near empty
Standby alarm active
Timer alarm, pump stop (KVO)
Total volume reached, pump stop (KVO)
Format statusbyte-5:
P 1 X X X X X X
Always low
Always high
Bargraph-LED upper
Bargraph-LED lower+3
Bargraph-LED lower+2
Pressure indicator (Adr. 46)
A sample of a monitoring session:
- Switch the pump on with an inserted filled infusion set.
-Type "Ctrl+N" to set the pump in remote mode.
-Type "Ctrl+B", then "01200200" and then "Ctrl+C" which sets the rate to 12.0 ml/h and an infusion total of 200 ml. The rate should now be shown in the left display of the pump.
-Type "Ctrl+R" to start the infusion.
-Type "Ctrl+T" to stop the infusion.
Bargraph-LED lower
Bargraph-LED lower+1
14.135.D_A600D en.SM.V1.50 Argus 10.08.04 / PJ 16
3 Fault finding
The fault code in case of a failure is displayed by pressing "MODE" key (12). (F-XX) appears in display (3), and the source of the failure is listed in the table below:
Error Error reason Assembly group
F-20 Internal watchdog Mainboard F-21...22 ROM test Mainboard F-23...24 RAM test Mainboard F-25 CPU test Mainboard F-26 Invalid function menu F-27 EEPROM data invalid Mainboard F-28 RTC (real time clock) data invalid Mainboard F-29 Stepper motor power test Mainboard F-30 Plunger position calculation failed F-31 Check for near empty F-32 5Volt supply out of range Mainboard F-33 20Volt supply out of range Mainboard F-34 Pressure reference out of range (LM385 2.5V) Mainboard F-35 Pressure signal out of range F-36 Pressure result invalid Mainboard F-37 Pressure sensor test failed F-38 Barrel diameter signal test failed F-39 Barrel diameter signal out of range F-40 Clamp diameter signal out of range F-44...45 Address invalid for EEPROM Mainboard F-46 Frequency from µC or RTC out of range Mainboard F-47 Displayboard not present F-48 Key(s) too long active Displayboard F-54 Movement result invalid F-55 Frequency calculation Mainboard F-56...59 Volume control Mainboard
Exceptionally a fault code may appear, which is not included in this list. In this case we recommend to change the main board.
14.135.D_A600D en.SM.V1.50 Argus 10.08.04 / PJ 17
4 Replacement of parts
4.1 Disassembly of the pump NOTE: The exploded views in the appendix show the individual assembling steps.
CAUTION! Switch the unit off and disconnect the mains cable from the power outlet before opening the housing. Observe the antistatic protection rules when disassembling the ARGUS infusion pump (the use of an anti­static table mat and a grounded clip are recommended, for example). Mind the battery voltage!
aDisassembly of the housing: Remove the pole clamp at the rear side. Remove 7 screws at the bottom side
(6 x M4 and 1 x M3), the 2 screws at the left side cannot be removed completely. Place the upper part behind the bottom.
b Remove the main board:
Remove the battery connector and all other cables of the main board.
c Remove the syringe drive:
Remove the fixing plate and syringe guide. Unsolder the connecting leads of the strain gauge (DMS) on the power board. Move the drive head to the left and remove the syringe drive out from the housing.
d Remove the driving head:
Remove the 3 screws on the cover. Important: To disassemble the unit, open the cogs by hand one third (or put a coin between the cogs) then pull the cover with the levers out of the housing.
e Remove the power board:
Important: Removing the power board requires a new basic alignment. Unsolder the connecting leads of the strain gauge on the power board. Remove all cables from the board.
fInsert the power board:
Syringe holder must be in the closed position (no syringe inserted). Loosen the lock screw of the cogwheel on the syringe holder potentiometer axle. Fix the power board with the 4 screws. Note that the lock screw is accessible from above. Solder the connecting leads of the strain gauge and connect the other cables. Make sure to remove the AC power cord and operate the pump (with open housing) on battery power.
g Replace the sidewall (motor)
After each disassembly or replacing of the sidewall the basic alignment of the strain gauge must be executed to guarantee a perfect pressure monitoring.
h Replacement of the EPROM or main board
Software updates may reset automatically the configured values in the EEPROM. You are requested to upload and save or write down the present contents of the none write protected addresses before you replace the EPROM or the main board. Afterwards you have to re-enter these values in the program mode or download your saved file. If a PIN code has been used before, the same code has to be programmed again. Charge the batterie for more than 16 hours! Note: A pressure sensor calibration is necessary when replacing the pressure sensor, a pressure sensor calibration and a volume calibration are necessary when replacing the EEPROM!
4.2 Check the pump accuracy and the pressure alarm level a Check the pump accuracy (Select the syringe type [Cod -50-])
Insert a new syringe "Codan 50ml" filled with distilled water and pump into a cup. Pump settings: set rate at 200 ml/h, set "total" at 20 ml Net weight result: 20 g +/-2%
b Check the pressure alarm level
Adjust the pressure alarm limit to 900mbar (90kPa). Insert a syringe filled with water and connect the tube on the patient side to a manometer.
14.135.D_A600D en.SM.V1.50 Argus 10.08.04 / PJ 18
Set rate of 100 ml/h, start and pump till the occlusion-alarm goes on and immediately read the pressure on
the manometer: 900mbar (90kPa) ± 200mbar
4.3 Rough alignments
Plunger length (P):
- Go into the programming mode (see chapters 1.3 or 1.5)
- Select address 399
-Press key "MODE"
- Enter data 123
-Press key "SRART/STOP", the display indicates "17bd xxxx"
-Press key "MODE" until "120P xxxx " is displayed
- Loosen the lock screw of the cogwheel on the plunger potentiometer axle
- Move syringe drive (without syringe) fully to the left
-Turn the potentiometer axle in clockwise direction up to the final position and afterwards in the counter clockwise direction until approx. 700 is displayed
-Fix the lock screw!
- Control whether the full stroke can be made
Syringe clamp diameter (cd):
- Go into the programming mode (see chapters 1.3 or 1.5)
- Select address 399
-Press key "MODE"
- Enter data 123
-Press key "SRART/STOP", the display indicates "17bd xxxx"
-Press key "MODE" until "20cd xxxx" is displayed
- Remove the syringe and make sure the clamp is fully closed
- Remove the cover of the driving head (11.212)
- Remove the clamp spring (11.210)
- Loosen the lock screw of the position lever (11.208)
-Turn carefully the potentiometer axle (R2) in counter clockwise direction up to the final position
-Turn position lever (11.208) counter clockwise until it touches the housing (see picture below)
-Fix the lock screw (make sure the position lever touches the housing)
- Re-install the clamp spring, then a value of approx. 700 is displayed
- Control whether the clamp stroke can be made
R2
Syringe barrel diameter (bd):
- Go into the programming mode (see chapters 1.3 or 1.5)
- Select address 399
-Press key "MODE"
- Enter data 123
-Press key "START/STOP", the display indicates "17bd xxxx"
- Loosen the lock screw of the potentiometer R14 on the power board
-Turn the potentiometer axle (R14, on the power board) in the counter clockwise direction up to the final position and afterwards in the clockwise direction until approx. 700 appears in the display.
-Fix the lock screw
- Control whether the syringe barrel can make the full stroke.
Strain gauge (b): Important: No syringe is inserted and the syringe drive is positioned on the right side.
- Go into the programming mode (see chapters 1.3 or 1.5)
- Select address 399
-Press key "MODE"
14.135.D_A600D en.SM.V1.50 Argus 10.08.04 / PJ 19
- Enter data 123
-Press key "MODE" several times until " 0.2b xxxx " is displayed.
- Adjust the screw of the trimmer (R13, on the power board) until approx. 1500 is displayed.
R13
Lock
R14
Power Board
4.4 Final calibration of the assembled pump
- Go into the program mode (see chapters 1.3 or 1.5).
- Select address 399.
-Press key "MODE".
- Enter data 123.
-Press key "START/STOP". The display indicates "17bd xxxx".
-Please verify that the displayed values in the right hand display are within the following ranges (without calibration part):
Valid ranges in the right hand display : Syringe barrel holder unpulled: 700 ± 300 Syringe barrel holder pulled: 4'200 ± 300
Note: Please refer to chapter "Rough alignments" if the displayed value is out of range!
-Press key "MODE". The display indicates "120P xxxx". Please verify that the displayed values in the right hand display are within the following ranges (without calibration part):
Valid ranges in the right hand display (without calibration part): Drive unit totally left: 600 ± 200 Drive unit totally right: 4'400 ± 200
Note: Please refer to chapter "Rough alignments" if the displayed value is out of range!
-Press key "MODE". The display indicates "20cd xxxx". Please verify that the displayed values in the right hand display are within the following ranges (without calibration part):
Valid ranges in the right hand display (without calibration part): Clamp fully closed: 700 ± 300 Clamp fully opened (clutch & clamp lever fully pressed): 2'000 ± 300
Note: Please refer to chapter "Rough alignments" if the displayed value is out of range!
-Press key “MODE” until the display indicates "17bd xxxx".
- Put the calibration part-1 (no.11-194) (d=17mm ; l=120mm) in place
-Press key "START/STOP". (Barrel diameter for 17mm will be stored). Note: Each stored value will be acknowledged by a sound.
-Press key "MODE". The display indicates "120P xxxx".
-Press key "START/STOP". (Plunger length for 120mm will be stored)
-Press key "MODE". The display indicates "20cd xxxx".
-Press key "START/STOP". (Clamp diameter for 20mm will be stored)
- Remove the calibration part-1.
14.135.D_A600D en.SM.V1.50 Argus 10.08.04 / PJ 20
- Put the calibration part-2 (no.11-195) (d=31mm ; l=20mm) in place.
-Press key "MODE". The display indicates "31bd xxxx".
-Press key "START/STOP". (Barrel diameter for 31mm will be stored).
-Press key "MODE". The display indicates "20P xxxx".
-Press key "START/STOP". (Plunger length for 20mm will be stored).
-Press key "MODE". The display indicates "32cd xxxx".
-Press key "START/STOP". (Clamp diameter for 32mm will be stored).
- Remove the calibration part-2. Insert a filled 50 ml syringe and connect the patient line to the pressure
measurement system.
-Press key "MODE". The display indicates "0.2b xxxx" and the pump starts to run.
with a low rate. Close the line (occlusion).
-Wait until 0.2 bar is reached on the scale. At this point press the key "START/STOP" immediately
to register the value for 0.2 bar. Note: To speed up the process increase the infusion rate in steps by pressing the key "1".
-Press key "MODE". The display indicates "1.2b xxxx".
-Wait until 1.2 bar is reached on the scale. At this point press the key "START/STOP" immediately
to register the value for 1.2 bar. Note: To speed up the process increase the infusion rate in steps by pressing the key "1".
- Switch the pump off and on again.
- Remove the pressure in the system by opening the line.
- Make a control measurement for a pressure of 0.9 bar. To do this, select 900 mbar in the function "PrL"
The pressure alarm should be released within ± 200 mbar.
4.5 Calibration of the battery capacity Each battery is subject to a chemical process with a slowly decreasing running time. After many charge and discharge cycles the battery may not have the capacity to provide the running time shown in the menu "CAP". To adjust the running time of the used battery please follow the steps mentioned below:
- Go in the configuration mode of the pump (see chapters 1.3 or 1.5).
- Select address "368" in the left display.
- Enter the data "615" in the right display and press the "START/STOP" key to accept the data. This will set
the battery discharge time to the maximum of >10 hours.
- Switch the pump off.
- Be sure you have unplugged the line connection.
- Switch the pump on and run the pump on battery until it switches off.
- Charge the battery for more than 16 hours.
- Switch on the pump and start an infusion with a rate of 5 ml/h. The infused sum at this rate multiplied by
12 is now equal to battery operating time in minutes.
- Leave the pump running on battery until it switches off again.
- Connect pump to the AC line.
14.135.D_A600D en.SM.V1.50 Argus 10.08.04 / PJ 21
- Switch the pump on while keeping the key "1" pressed. Multiply the value in the right display by 12, this gives the capacity of the battery in minutes. Multiply this time by 0.8 and enter the result on address "368" in the configuration mode. This time defines from now on, the running time of the pump including a 15 minutes pre-alarm (valid after a full charge).
-Standard battery 6V/1.2 Ah If this time is less than 2 hrs, you should replace the battery (part 12.032). If the specified time > 2 hrs is not required, the battery has to be changed only if the time less than 1.5 hrs, to respect to environmental pollution.
-High energie battery 6V/4 Ah If this time is less than 8 hrs, you should replace the battery (part 12.026). If the specified time > 8 hrs is not required, the battery has to be changed only if the time less than 5 hrs, to respect to environmental pollution.
14.135.D_A600D en.SM.V1.50 Argus 10.08.04 / PJ 22
Safety standard check ARGUS 600 ARGUS Medical AG
5 Safety standard check
Serial-no:
Hospital/Dept./Customer:...............................................................................................................................................
The safet The check has to be done in accordance to the operation- and service manuals.
1 Visual check for dama
2 Keep key "MODE" pressed while switching on the pump - Display shows the software release: V ......…
3 Hold the barrel switch pressed and move the lever for
clamp (bottom) up and down - Alarm "Syringe" lights on and off
4 Hold the lever for clamp (bottom) in its upper position
and actuate the barrel switch - Alarm "Syringe" lights on and off
5 Hold the barrel switch pressed and the lever fo
clamp (bottom) in its upper position, then actuat
the lever for clutch (top) - Alarm "Syringe" lights on and off
6 Check the prime function (press the key "BOLUS"
twice within 1 sec.) - See service manual "Prime"
7 Set rate to 111.1 ml/h and the total to 0.1 ml,
press "START STOP" released
8 Test the pump at max. rate (999.9 ml/h) - Running smooth? 9 Check of the occlusion-alarm pressure: - See service manual "Final calibration"
50 ml syringe: CODAN: ........................... Other: ................................................................
Pressure increase to ≥1.2 bar?
Test of the occlusion-alarm-pressure Preset level: ......... mbar Measured: ......... mbar
10 Check of the
Rate: 200 ml/h Preset volume: 20 ml Measured volume: ............ml
11 Batter
disconnect the line and start the pump: - Battery prealarm after typ. 1 hr 45 min or 7 hrs 45 min
Running time: ........ hrs ........ min - Battery alarm 15 min. after prealarm
(If the specified typical 2hrs (8 hrs) of operation are not (Red LED battery alarm + ALARM + alarm acoustically) required, the battery has to be changed only if the time - After 6 min. the pump switches off is <1.5 hrs (5hrs), due to environmental pollution)
12 Charge the battery min. 16 hrs.
13 Check the external connections: - Staff alertin
14 Electrical test according to EN60601-1 - Measurements attached
(all measurements made with a power cable 2.5 m) The pump has passed the safety standard check and is safe for use.
.........................
y standard check has to be performed at least every 24 months or after 10000 hours of operation.
ge, cleanness and completeness: - Housing, labels, accessories, connectors,
pump accuracy: - See service manual "Check pump accuracy..."
y check by setting the rate to 5 ml/h, - Green LED battery light?
power cable, etc.
play of 2, 4, 7, F., ml total, ml inf., h.min
- Dis
- Test of the battery, KVO
- Test of the red alarm LEDs: Occlusion, near empty,
empty, syringe, battery, ALARM
- "END" reached, the acoustical alarm + LED ALARM
(Red LED battery alarm + acoustical alarm)
- Computer interface RS232 / External DC
green operation LEDs: Run indicator, line,
g system
Date: ...........................Name: ...............................Signature: ....................................
14.135.D_A600D en.SM.V1.50 Argus 10.08.04 / PJ 23
10.072
10.069
10.073
10.065
11.200 Identification plate
11.201 Short instruction english
11.206 Protection (display board)
11.213 Frontpanel universal
12.030 Male plug 3 c
12.031 Male plug 7 c
12.035 Manometer 0-2,5 bar
12.036 AM pressure system
Moduls
10.065 Syringe drive compl.
10.069 Thread spindle compl.
10.072 Driving head compl.
10.073 Carrier (incl. nut) compl.
General spare parts (without drawings)
10.058 RS-232 interface cable
10.060 Power board A600
10.061 Display board A600
10.071 Main board A600 Index C
11.189 Casing cover (without fittings)
11.190 Casing base (without fittings)
11.194 Calibration part 1
11.195 Calibration part 2
14.135.D_A600D en.SM.V1.50 Argus 10.08.04 / PJ 24
Basement
11.002 Spindle for clamp
11.150 Potentiometer holder
11.152 Combination clamp
11.153 Handle
11.154 Battery cover
11.155 Clamp holder
11.190
12.026 (12.032 + 11.207)
11.002
11.205
11.165 Fixing plate
11.168 Syringe barrel holder
11.169 Cog shaft
11.171 Switch holder
11.176 Cog wheel 24 c
11.188 Syringe guide
11.190 Casing bottom part without fittings
11.192 Pressure spring
11.196 Screw M4x50
11.197 Support
11.205 Squeeze screw
11.207 Battery holder for part-no 12.032
12.021 PCB potentiometer 10K
11.188
11.154
11.171
12.022 Pushbutton switch
12.023 Mains plug
12.024 Female receptacle 3 c
12.025 Female receptacle 7 c
12.026 Lead-acid battery 6V/4Ah
11.168
(Switch 12.022)
11.192
12.032 Lead-acid battery 6V/1.2Ah
11.169
11.165
11.153
11.176
12.021
11.150
11.152
11.155
12.025
12.024
11.197
11.196
12.023
14.135.D_A600D en.SM.V1.50 Argus 10.08.04 / PJ 25
10.068
11.193
11.157
11.156
11.175
11.177
12.028
11.166
11.198
11.159
11.170
11.185
12.029
11.202
11.177 Spindle cog wheel
11.185 Main support
11.193 Distance bolt
11.198 Fixing plate
Drive
10.068 Motor and gear compl.
11.156 Bearing shaft
11.157 Cog rail
11.159 Axle
11.202 Sealing
12.028 Ball bearing
12.029 Glide bearing ø3/4,5x3
11.166 Thread spindle
11.170 Sidewall motor incl. DMS
11.175 Cog wheel 36 c
14.135.D_A600D en.SM.V1.50 Argus 10.08.04 / PJ 26
12.020
11.172
11.174
10.074
11.164
12.018
11.161
10.066
11.160
11.162
11.173
Carrier
10.066 Spindle nut compl.
10.074 Flex board with cable
11.160 Sleeve (in front)
11.161 Sleeve (in back)
11.162 Wing
11.164 Tension spring
11.172 Housing (Carrier)
11.173 Cover (Carrier)
11.174 Cog wheel 23 c
12.018 PCB micro switch
12.019 Flex cable
12.020 Potentiometer 10-turns 10K
14.135.D_A600D en.SM.V1.50 Argus 10.08.04 / PJ 27
(11.191 torsions spring pair)
11.211
11.183
11.178
11.182
11.163
11.158
11.184
11.181
11.209
11.187
11.212
11.186
11.167
11.208
11.210
11.212 Cover (Driving head)
11.211 Housing (Driving head)
11.210 Clamp spring
11.209 Potentiometer small
11.208 Position lever
11.191 Torsion spring (pair)
11.187 Working lever
11.186 Lever for beaks
11.184 Cog segment (upper)
11.183 Cog segment (under)
11.182 Clamp (bottom)
11.181 Clamp (top)
11.178 Bearing
11.167 Clutch lever
11.163 Carrier pipe
11.158 Connection pipe
Drive head
14.135.D_A600D en.SM.V1.50 Argus 10.08.04 / PJ 28
AC
Input
LN
X1
X2
X4
1
1
1
X5
DMS
1
X16
30.010
Motor
1
Nurse
Call
1
X12X14
EPROM
X3X13
Ext.-DC
SCI
X17
1
30.011
Barrel-
switch
Battery
X15
1
X11
1
1
X9
X7
X10
X8
30.007
X6
30.008
Wiring diagram
14.135.D_A600D en.SM.V1.50 Argus 10.08.04 / PJ 29
Bolus
Start/Stop
Mode
0,1
10
100
1
Pressure
Alarm
Occlusion
Empty
Near empty
ml total
ml inf.
Syringe
Battery
h. min
KVO
On/Off
Driver
Display
EEPROM
EPROM
Latch
2
RTC
I C
3 Volt
Li-Batt.
SQW-Frequency
+20V Control
+5V Control
Syringe pressure
Barrel size
Syringe position
Clutch + barrel switch
Clamp size
Controller
Driver
Buzzer
Nurse-call
Relay
Driver
Relay
Alarm
RS-232
Isolation
Reset +
Watchdog
Sin/cos
generator
Motor
Driver
Motor
Power test
AC Main
DC
AC
+ 20V
(Motor +
DC
DC
Buzzer)
Ext. DC
DC
+ 5V
(Logic)
DC
Line voltage control
6V
Pb-Batt.
Battery voltage control
Bloc schematic
14.135.D_A600D en.SM.V1.50 Argus 10.08.04 / PJ 30
14.135.D_A600D en.SM.V1.50 Argus 10.08.04 / PJ 31