11. SAFETY STANDARD CHECK..............................................................................................22
14.147.B_QUICK100M en.SM.V1.00.doc ARGUS Medical AG 11.08.04 / PJ 2 / 22
1. INTRODUCTION
1.1. General
IMPORTANT!
This service manual is intended for the exclusive use of authorized persons who
have been trained by ARGUS Medical AG in the maintenance and repair of the
QUICK
®
100 M docking station.
The service manual is meant to be used together with the user manual.
IMPORTANT!
ARGUS Medical AG shall not assume any responsibility for any manipulations
which have been carried out on the QUICK
authorized person.
IMPORTANT!
The QUICK
®
100 M docking station may only be used with the ARGUS 600 D
Software V1.5X and/or ARGUS 414 D Software V1.3X, and/or ARGUS 707,
ARGUS 708 Software V1.XX.
IMPORTANT!
The reading of the barcode (medicament name and infusion data's on a syringe or
a fluid bag) an d t h e d ata transfer to t he p u mp are fail save. The data transfer from
the QUICK
®
100 M docking station to a PC is not fail safe.
This manual contains the latest data available. It is subject to further modificatio ns
in accordance with technical improvements.
INTRODUCTION
®
100 M docking station by a non-
14.147.B_QUICK100M en.SM.V1.00.doc ARGUS Medical AG 11.08.04 / PJ 3 / 22
BARCODE READER AND LABEL DATA
2. BARCODE READER AND LABEL DATA
2.1. General
The barcode reader "Scan Plus 1800" (Intermec) including a gender changer may
be ordered from ARGUS Medical AG (Art. no. 90.150). It should be
connected to the appropriate plug of the QUICK
marked with the sign.
CAUTION!
The barcode reader "ScanPlus 1800" has been preconfigured and tested by
ARGUS Medical AG for the use with the QUICK
2.2. Label layout
ARGUS Medical recommends to use its own Excel tool named “Medication Barcode Labels” to generate these barcode labels and bar-coded prescri pt i on sheets.
A medication barcode label holds the following relevant information:
1. The Medication concentration in the programmed unit
2. The dose (ml total) in ml
3. The dose in the programmed unit
4. The infusion rate in ml/h
5. The medication name
6. The initial rate in the
programmed unit
7. The patient name
8. The patient weight in kg
9. A software version and an ARGUS check number
®
100 M docking station
®
100 M docking station.
2.3. Barcode format
The used barcode standard corresponds to code 3 of 9 including check digit.
The barcode format is:
* T Rate
1E3
Rate
1E2
Rate
1E1
Rate
1E0
Rate
1E-1
Dose
1E3
Dose
1E2
Dose
1E1
Dose
1E0
Dose
1E-1
Med.
HEX
High
digit
Med.
HEX
low
digit
CD *
T = Pump Type: “S” for syringe pump, “V” for volumetric pump
Med. = Medication number in HEX format (see pump service manual for details)
CD = Modulo 43 Check Digit
14.147.B_QUICK100M en.SM.V1.00.doc ARGUS Medical AG 11.08.04 / PJ 4 / 22
CONFIGURATIONS
3. CONFIGURATIONS
3.1. Configuration of the docking station
A special software tool for the configuration of the docking station will be released
soon. Please contact your local distributor or ARGUS Medical AG.
3.2. Pump configuration
CAUTION!
Every pump used with the QUICK
®
100 M docking station has to be preconfigured with
its corresponding serial number!
Thus the addresses #396 and #397 in the configuration of each pump must be programmed with the corresponding serial number. Please refer to the pumps respectiv e
service manual to change these addresses if necessary. In the case of an inv alid serial
number, the docking station will generate an alarm and the pump can not be programmed via barcode.
14.147.B_QUICK100M en.SM.V1.00.doc ARGUS Medical AG 11.08.04 / PJ 5 / 22
4. SOFTWARE UPDATES
4.1. General
4.1.1. Introduction
This chapter describes the procedure to perform a software update on the
QUICK
®
100 M docking station. To check the installed software release in
your docking station you can either read out the history (chapter 5) or enquiry the docking station’s status (chapter 6) over the serial interface.
Please refer to your local distributor or ARGUS Medical AG to determine
the latest software release able to run on your device hardware.
®
The QUICK
100 M contains a flash memory which can be updated easily
via the serial interface of a PC. This flash memory is divided in 2 sectors:
- User Program Sector - Boot loader Sector
The User Program Sector contains the program which runs when the device is switched on normally. The Boot loader Sector contains the program
which runs when the device is forced by the user, to enter this mode in order to update the flash memory.
SOFTWARE UPDATES
4.1.2. Requi rements for a software update
If you want to update an ARGUS Medical device please be sure you have
the following items available:
1. Standard RS232 null modem serial interface cable, connected to the
barcode interface.
2. File "AFlash_Dongle.txt" from your local distributor or from ARGUS
Medical on your PC.
3. File "Q100_xxx.txt, if you want to update the user program. The file
name contains the software release of the user program.
4. Terminal emulator e.g. Hyper Terminal included in Microsoft
dows™.
4.1.3. Safety aspects
Be aware of the following points:
• For medical device traceability your local distributor or ARGUS
Medical AG needs to be informed about every device updates
rial number) you performed!
• Do not make any software updates when the device is used and/or
connected to a patient!
®
Win-
(se-
• Do not disconnect the data cable during the software update process!
14.147.B_QUICK100M en.SM.V1.00.doc ARGUS Medical AG 11.08.04 / PJ 6 / 22
SOFTWARE UPDATES
4.1.4. Settin g s of the terminal program
A software update can be done without additional software running on the
PC. The only tool you need is a terminal program like the Hyper Terminal
included in every Microsoft
®
Windows® Operating System.
Please ensure you
have selected the
right COM Port of
your PC.
Please make t he
following settings in
the Terminal Program
to enable the
communication with
the pump and to
ensure the right
display information:
Please set al so the following
ASCII character settings to make
the correct display settings in the
terminal program.
4.1.5. How to enter the flash mod e o f
the docking station
To enter the boot loader mode of
the device please follow the
procedure listed below:
14.147.B_QUICK100M en.SM.V1.00.doc ARGUS Medical AG 11.08.04 / PJ 7 / 22
SOFTWARE UPDATES
®
1. Unplugthe mains power cord from the QUICK
100 M docking station
to reset it and remove all pumps.
2. Connect the device to your PC.
3. Start the terminal program and make the settings as described in chapter 4.1.4.
4. Use the menu "Send text file..." in the menu "Transfer" of the Hyper
Terminal program to send the dongle file "AFlash_Dongle.txt" to the
QUICK
®
100 M docking station (this file will send synchronisation char-
acters for approx. 5 seconds).
5. After the start of the transmission, within 5 seconds plug in the mains power cord to the QUICK
®
100 M docking station.
6. Wait several seconds until the terminal program d is plays the following
text (if the device entered the boot loader mode correctly):
7. Follow the steps described in the next chapters based on the kin d of
update you want to perform.
ARGUS Medical AG
CH-3627 Heimberg
Bootloader V0.21
20030328 (C) AM
Loader for flashing ARGUS Medical devices.
1. Erase user program.
2. Update user program.
3. Erase bootloader.
4. Update bootloader.
14.147.B_QUICK100M en.SM.V1.00.doc ARGUS Medical AG 11.08.04 / PJ 8 / 22
4.2. Updating the User Program
To update the user program please go through the following steps to ensure a
complete software download:
1. Select menu 1 in the main menu of the boot loader by typing a "1" in the terminal program.
The boot loader will now clear the sectors containing the user program!
2. Select menu 2 in the main menu. The QUICK
be ready to receive flash programming data.
3. Send the file "Q100_xxx.txt" as text to the QUICK
using the menu "Send text file..." in the menu "Transfer" of the Hyper Terminal
program.
This part will take some time until the entire data is flashed. The QUICK
M docking station will flash the red LED as long as it' s rec eiving data from the
PC.
4. After a successful download, the QUICK
edge the end by ringing the internal buzzer twice.
SOFTWARE UPDATES
®
100 M docking station will now
®
100 M docking station by
®
100
®
100 M docking station will acknowl-
5. Select menu 5 to quit the boot loader mode of the device. You will be asked if
you want to quit. After confirmation the QUICK
run in the normal mode.
Important Remark:
Updating the user program does not have any influence on the boot loader software.
If you quit the boot loader mode without any valid user program in the flash memory, the device will always enter the boot loader mode.
4.3. Updating the Boot loader Program
We recommend updating the boot loader only if it's re quired by your local distributor or by ARGUS Medical.
To update the boot loader program please go through the following steps to ensure a complete software download:
1. Select menu 3 in the main menu of the boot loader by typing a "3" in the terminal program.
The boot loader will now clear the sector which contains the boot loader
program!
®
100 M docking station will now
2. Select menu 4 in the main menu.
3. Send the file "AMLQ100_xxx.tx t" as text to the QUICK
®
100 M docking station
by using the menu "Send Text File..." in the menu "Transfer" of the Hyper Terminal program. This part will take some time until the entire data is flashed.
The QUICK
®
100 M docking station will flash the red LED as long as it is re-
ceiving data from the PC.
14.147.B_QUICK100M en.SM.V1.00.doc ARGUS Medical AG 11.08.04 / PJ 9 / 22
4. After a successful download, the QUICK
edge the end by ringing the internal buzzer twice.
5. Type a "5" in the main menu to quit the boot loader mode of the QUICK
docking station. You will be asked if you want to quit. After confirmation the
QUICK
Important Remark:
Updating the boot loader program will not have any influence on the user program.
Do not unplug the mains of the device during updating the boot loader or quit this
mode without any valid boot loader in the flash memory. The device must contain
a valid boot loader to operate correctly. If it is not the case, it will no longer be possible to enter the flash mode. In this case the main board M has to be replaced!
4.4. Important Notes
The user program performs an internal ROM-Test during the self test phase of the
docking station. If an error occurred during the update process the device will enter the fail safe state (fault codes please refer to chapter 7.2).
Please refer to your local distributor or ARGUS Medical AG to determine the latest
software release able to run on your device hardware.
SOFTWARE UPDATES
®
100 M docking station will acknowl-
®
100 M docking station will now run in the normal mod e.
®
100 M
14.147.B_QUICK100M en.SM.V1.00.doc ARGUS Medical AG 11.08.04 / PJ 10 / 22
5. HISTORY
5.1. General
The connection of the QUICK
®
100 M docking station to a computer allows to read
the history file stored in the flash memory.
5.2. Entering the docking station history read mode
To enter the history read mode please follow the procedure listed below:
1. Switch off all docked in pumps. Unplug the mains power cord from
the QUICK
®
100 M docking station and plug in the mains power cord
again to reset it.
2. Connect the device to your PC.
3. Start the terminal program and make the settings as described in chapter 4.1.4.
4. Type "QUI;100" + ENTER. The history read mode program will be activated and the red LED lights up.
HISTORY
5. Type "HIS;" + Enter. The history will now be displayed on the screen
(see examples below).
6. Exit the history read mode by unplugging the mains power cord again.
The events registered will be displayed in a chronological order; the most recent
entry is displayed on top of the list. Please be aware that the time displayed in
minutes refers always to the elapsed time since last power up of the QUICK
M docking station.
5.3. History header
At the top of the history, a header will appear showing the device type, the software release and if configured, the inventory number and the last service date .
5.4. History events
Each registered event starts with a message line. See the list mentioned in chapter 5.6.
In case the docking station detects a technical failure, the last event line registered
includes the failure number (fault code) which generated that technical failure.
Please refer to chapter 7.2 for the complete list.
®
100
14.147.B_QUICK100M en.SM.V1.00.doc ARGUS Medical AG 11.08.04 / PJ 11 / 22
5.5. History example
/***** History *****/
Docking type: QUICK(R)100 M
Inventory number: 0
Software release: V0.90 (030730-130A)
Last service date:
Pump docked out at place 2
Place 1 = A600 Status = 40 Elapsed time = 0:01:12
Place 2 = No pump Status = 00
Place 3 = A600 Status = 40
Place 4 = No pump Status = 00
Place 5 = A414 Status = 40
Fail = None
Pump in place 3 has alarm
Place 1 = A600 Status = 40 Elapsed time = 0:01:01
Place 2 = A600 Status = 40
Place 3 = A600 Status = 4E
Place 4 = No pump Status = 00
Place 5 = A414 Status = 40
Fail = None
Pump docked in at place 3
Place 1 = A600 Status = 40 Elapsed time = 0:00:14
Place 2 = A600 Status = 40
Place 3 = A600 Status = 42
Place 4 = No pump Status = 00
Place 5 = No pump Status = 00
Fail = None
Pump in place 2 has failure
Place 1 = A600 Status = 40 Elapsed time = 0:00:10
Place 2 = A600 Status = 42
Place 3 = No pump Status = 00
Place 4 = No pump Status = 00
Place 5 = No pump Status = 00
Fail = None
Pump docked in at place 2
Place 1 = A600 Status = 40 Elapsed time = 0:00:08
Place 2 = A600 Status = 42
Place 3 = No pump Status = 00
Place 4 = No pump Status = 00
Place 5 = No pump Status = 00
Fail = None
Power up on docking station
Place 1 = No pump Status = 00 Elapsed time = 0:00:00
Place 2 = No pump Status = 00
Place 3 = No pump Status = 00
Place 4 = No pump Status = 00
Place 5 = No pump Status = 00
Fail = None
HISTORY
14.147.B_QUICK100M en.SM.V1.00.doc ARGUS Medical AG 11.08.04 / PJ 12 / 22
5.6. History messages
Possible messages appearing in the first line of each history event:
No information available
Power up on docking station
Power off on docking station
Reserved
Reserved
Pump with invalid SN at place 1
Pump with invalid SN at place 2
Pump with invalid SN at place 3
Pump with invalid SN at place 4
Pump with invalid SN at place 5
Docking station has detected failure
Com. troubled with pump place 1
Com. troubled with pump place 2
Com. troubled with pump place 3
Com. troubled with pump place 4
Com. troubled with pump place 5
Pump in place 1 has alarm
Pump in place 2 has alarm
Pump in place 3 has alarm
Pump in place 4 has alarm
Pump in place 5 has alarm
Pump in place 1 has failure
Pump in place 2 has failure
Pump in place 3 has failure
Pump in place 4 has failure
HISTORY
Pump in place 5 has failure
Low pump supply voltage detected
Enter PC configuration mode
PC configuration done
PC configuration failure
Pump docked in at place 1
Pump docked out at place 1
Pump docked in at place 2
Pump docked out at place 2
Pump docked in at place 3
Pump docked out at place 3
Pump docked in at place 4
Pump docked out at place 4
Pump docked in at place 5
Pump docked out at place 5
Pump 1 programmed by barcode
Invalid barcode programming at pump 1
Pump 2 programmed by barcode
Invalid barcode programming at pump 2
Pump 3 programmed by barcode
Invalid barcode programming at pump 3
Pump 4 programmed by barcode
Invalid barcode programming at pump 4
Pump 5 programmed by barcode
Invalid barcode programming at pump 5
14.147.B_QUICK100M en.SM.V1.00.doc ARGUS Medical AG 11.08.04 / PJ 13 / 22
e
ceived?
(
6. MONITORING
6.1. General
The RS 232 PC interface protocol of software release V1.00 is intended to be
used only in combination with a PDMS (Patient Data Management System). For
visualization and documentation purposes, the status of each docked-in pump and
the status of the docking station itself can be requested through the serial interface.
CAUTION!
Those PC transmitted data are not intended to be used for diagnostic or remote
controlling of the pumps or other devices.
6.2. Serial communication protocol
6.2.1. Principle of t he communication
The following flow diagram shows how a serial communication procedure could be
implemented on the PC side to poll each pump placed on the docking station.
Depending on which status is required by the PC, the docking station will transmit
either the pump status or the status of the docking station itself.
Send preamble address
ESC (0x1B)
Send pump place or
docking station addr.
0x’Addr‘)
Send ENQ (0x05)
Complete
Status
r
N
Status decoding
and handling.
Displaying of the rele-
vant data.
Y
Select next address in
the docking station
The protocol syntax of the status transmitted is part of the next chapters.
MONITORING
The pump places are labelled on the docking
station housing and have the following value:
Pump place 5, ‘Addr’ 0x35
Pump place 4, ‘Addr’ 0x34
Pump place 3, ‘Addr’ 0x33
Pump place 2, ‘Addr’ 0x32
Pump place 1, ‘Addr’ 0x31
Docking station, ‘Addr’ 0x30
No pump docked at
this place.
14.147.B_QUICK100M en.SM.V1.00.doc ARGUS Medical AG 11.08.04 / PJ 14 / 22
pump
(
))
y
MONITORING
6.2.2. Docking station status
Software release: V1.00
STXCS
P4P5
1. B
DS SWETXP1P2P3
All pump places equal to place 1
P1
Pump place 1
te2. Byte
Reserved
Communication error occurred
21765410765
6343207
STX: Start of text
P1...P5 : Word boundaries for each pump place (16 Bit)
DS : Docking Station status (8 bit)
SW: Docking Station software release (24Bit)
CS: Checksum CRC (16 Bit, not implemented yet)
ETX: End of text (0x03)
000B = None
100B = A414 D
010B = A600 D
110B = Reserved (A02)
XX1B = Reserved
Always 1
14.147.B_QUICK100M en.SM.V1.00.doc ARGUS Medical AG 11.08.04 / PJ 15 / 22
STXP1P2P3
CS ETXP4 P5DS SW
7654
1. Byte2. Byte3. Byte
1. Byte
P4 P5DS SWSTXP1P2P3
CS ETX
MONITORING
732106
Reserved
Reserved
Any pump alarm or pump failure active
Low pump power supply (+20V)
Reserved
Always 1
Main release X (ASCII character '0'-'9')
Sub release Y (ASCII character '0'-'9')
Sub release Z (ASCII character '0'-'9')
Software release X.YZ
6.2.3. Pump status
The transmitted pump status will be equal to the one described in the pump service manual.
An example of an ARGUS 600 D syringe pump status is shown below:
14.147.B_QUICK100M en.SM.V1.00.doc ARGUS Medical AG 11.08.04 / PJ 16 / 22
7. MAINTENANCE AND FAULT FINDING
7.1. Maintenance
CAUTION!
Only authorized persons who have been instructed by ARGUS Medical AG or by
the trained local distributor are allowed to service the QUICK
tion. In case of repair request, send the unit to the local distributor with a report
outlining the exact nature of the failure. More information is available from:
ARGUS Medical AG
CH-3627 Heimberg / Switzerland
E-mail: info@argusmedical.com
CAUTION!
The safety standard check has to be performed at least every 24 month or after
10'000 hours of operation. The check has to be done in accordance to the chapter
11.
No special maintenance of the QUICK
are no wear and t ear parts.
Interferences through external high frequency electromagnetic fields (e.g. in combination with surgical equipment) are unknown. In case of doubt we suggest that
you contact your local distributor.
MAINTENANCE AND FAULT FINDING
®
100 M docking sta-
®
100 M docking station is necessary. There
7.2. Fault code
The fault code in case of a technical failure is displayed in the history file, refer to
chapter 5. The possible fault codes registered in the history event log file are listed
in the table below:
Fault Failure:
Code:
21 ROM test
22 ROM check (Runtime)
23 RAM test/check
24 XRAM test/check
25 CPU test
26 Invalid state (in PC-Configuration mode)
27 Eeprom data invalid
32 5Volt supply out of range
33 20Volt supply out of range (delayed 5s)
44 Address invalid for config-eeprom
45 Address invalid for history-eeprom
Exceptionally a fault code not listed above may appear. In this case we recommend changing the main board.
7.3. Specifications
Please refer to the user manual for the specifications (cap. 7).
14.147.B_QUICK100M en.SM.V1.00.doc ARGUS Medical AG 11.08.04 / PJ 17 / 22
8. REPLACEMENT OF PARTS
8.1. Disassembling of the Docking Station
CAUTION!
Disconnect the mains cable from the power outlet before opening the housing!
Observe the antistatic protection rules when disassembling the QUICK
docking station (the use of an antistatic table mat and a grounded clip are recommended).
8.2. Replacement parts
CAUTION!
The QUICK
®
100 M docking station may only be used with accessories and spare
parts which have been approved by ARGUS Medical AG for safe technical use.
11.220 Centring fixation
10.114 Cover
11.215 Casing
REPLACEMENT OF PARTS
®
100 M
10.096 Voltage regulator
14.147.B_QUICK100M en.SM.V1.00.doc ARGUS Medical AG 11.08.04 / PJ 18 / 22
REPLACEMENT OF PARTS
11.218 Coil spring
11.216 Handle with axle
11.217 Locking device
11.214 Transformer holder
12.008 Mains plug with filter
10.095 Extended board
10.097 Transformer 230V
10.087 Combination clamp
10.115 Main board Quick 100M
10.130 Barcode holder with clamp 12.041 Interface cable
90.041 Rail set (25x10mm)
90.042 Rail set (35x 8mm)
10.116 Flat cable set Quick 100M
12.043 Gender Charger
14.147.B_QUICK100M en.SM.V1.00.doc ARGUS Medical AG 11.08.04 / PJ 19 / 22
9. WIRING DIAGRAM
Main-Board_2
(30.022.A)
X15
10
WIRING DIAGRAM
X13
Docking place 5
X12
Docking place 4
Voltage regulato r
20V
Operation LED
Trafo
2
3
3
X16
X03
X14
X11
X10
Docking place 3
Docking place 2
Main-Board_1
(30.020.A)
X09
2
X01
X05
X07
Docking place 1
AC Main
10
X06
Barcode Reader
Nurse Call
10
X08
Interface
RS 232
14.147.B_QUICK100M en.SM.V1.00.doc ARGUS Medical AG 11.08.04 / PJ 20 / 22
10. BLOC SCHEMATIC
BLOC SCHEMATIC
AC Main
Nurse-call
AC
DC
RS232
RS232
DC
Alarm
Relay
DC
+ 20V
(Docking-
places)
Controller
2
I C
Buzzer
Operation/
Alarm LED
Reset/
Watchdog
Latch
EEPROM
FLASH
DC
+ 5V
DC
(Logic)
14.147.B_QUICK100M en.SM.V1.00.doc ARGUS Medical AG 11.08.04 / PJ 21 / 22
p
y pump
SAFETY STANDARD CHECK
11. SAFETY STANDARD CHECK
Safety standard checkQUICK® 100 MARGUS M edical A G
Serial-No.:
ital/Dept/Customer:
Hos
The safety standard check has to be performed at least every 24 months or after 10000 hours of operation.
The check has to be done in accordance to the operation and service manual.
1 Visual check for damage, cleanness and completeness:- Housin g, labe ls, accessories, con n ectors, cables, etc..?
2 Plug in the mains of the docking station- The green LED o n th e d o ckin g station ligh ts up ?
3 Check each docking position with a switched
4 Check each docking position with a switched on pump:- Buzzer sounds twice, ok?
5 Chech the extern al co nn ectio ns:- Co mp uter interface ?
6 Electrical test acco rd ing to EN60601-1:- Measurements attached?
Do not install an
All measurements made with a 2.5 m power cord
!
off
pump:- Mechanical locking and unlocking ok ?
- After docking the pump, its green power LED lights up?
- LED on docking station flashes orange-red, ok?
- Barcod e reader in terface ?
- S taff alerting system ?
14.147.B_QUICK100M en.SM.V1.00.doc ARGUS Medical AG 11.08.04 / PJ 22 / 22