AquaMetrix SHARK TXP Multi-Parameter Transmitters Operating Manual

SHARK
TRANSMITTER USER’S MANUAL
TX
& SHARK
TXP
MULTI-PARAMETER
N116-35 Rev 3
AquaMetrix Inc. 1245 Maple Hill Ct., Unit 7 Newmarket, ON Canada, L3Y 9E8
A57-24
Tel: (800) 742-1413
(905) 954-0841
Fax: (905) 954-0415
www.aquametrix.com
MULTI-PARAMETER TRANSMITTER USER’S MANUAL
Table of Contents
Subject Page No.
Introduction 1
Section 1 - Specifications 2
Section 2 - Installation 3 - 5
Section 3 - Electrical Connections & Setup 6 - 11
Differential Probe connection & setup 7 Combination Probe connection & setup 8 Conductivity Cell (Contacting style) connection & setup 9 Paddle Wheel Flow Sensor connection & setup 10 4-20mA Isolated Output 11
Section 4 - Using the Transmitter in pH Mode 12
LCD Menu 13 - 28
Section 5 - Using the Transmitter in ORP Mode 29
LCD Menu 30 - 44
Section 6 - Using the Transmitter in Conductivity Mode 45
LCD Menu 46 - 62
Section 7 - Using the Transmitter in Flow Mode 63
LCD Menu 64 - 78
Appendix A - Probe Configuration Table 79
Return Policy & Warranty Plan 80
MULTI-PARAMETER TRANSMITTER USER’S MANUAL
Introduction
The SHARKTX & SHARKTXP multi-parameter transmitter is a microprocessor based transmitter capable of measuring one of the following parameters, pH, ORP, conductivi­ty or flow.
When shipped from the factory, the Shark is not set to measure any one parameter. When the Shark is powered up for the first time, it will display the meter selection screen where the meter type must be selected. (refer to section 4.5 Meter Selection)
This meter selection screen will only be displayed when the Shark is powered up for the first time.
After the user selects a meter type the Shark will remain set to that meter type until it is changed with the meter selection menu function in the Utilities menu.
To return the Shark to its factory settings, the user must re-select the current meter type from the meter selection menu function. This will override all set-points and return all settings back to the factory settings.
The Shark Users menu has been divided into five main categories
- Calibration, used to calibrate the Shark with the selected sensor
- Utilities, used to manually control or override the outputs.
- Setup, used to configure the Sharks many options
- Diagnostics, used to troubleshoot problems with the Shark or sensor
- Outputs, used to configure the Shark’s 4 to 20mA output.
SHARK
TX is packaged in a rugged NEMA 4X polycarbonate enclosure with a universal
mounting kit for surface, panel and pipe-mount applications. This enclosure is perfect for stand-alone or panel-mount operation.
SHARK
TXP enclosure is also polycarbonate with a NEMA 4X front panel. Panel mount,
and DIN rail mounting hardware are supplied.
Introduction Page 1
MULTI-PARAMETER TRANSMITTER USER’S MANUAL
Section I - Specifications
pH ORP Conductivity Flow
Display 2 x 16 alpha-numeric LCD display
Power Requirements 4 to 20mA, Loop Powered, 16 to 32 VDC
M/cm30 to 19.99 0.01
0 to 2.000 0.01 0 to 20.00 0.1
3
0 to 200.0 0.1 0 to 5000 1.0 0 to 20.00 10
3
0 to 200.0 50
Measuring Range
pH: 0.01 to 14.00 ORP: -1999 to +1999mV
(Dependent on sensor)
uS/cm
Temp: 0 to 100°C or Temp: 0 to 100°C or
32° to +212°F 32° to +212°F (
mS/cm
Temp: 0 to 100°C or 32° to +212°F Automatic or Manual
Temperature Automatic or Manual
Compensation 0 - 100°C (32° to +212°F)
Not r
equired
User selectable temperature compensation slope 0.0 to 10.0%/°C. 0 to 100°C (32° to +212°F)
Temperature Unit °C or °F Not required
Temperature Sensor User selectable: 300NTC Thermistor, 3000NTC Thermistor or Pt. 1000 RTD Not required
Auto-Calibration
Calibration Modes Manual Calibration
Temperature Calibration Temperature Calibration
Manual Calibration T
emperature Calibration
Dry Calibration Sample Calibration K factor Input
Ambient Conditions Temperature:-20°C to +60°C or -4°F to +140°F Humidity: 0 to 90% RH (non-condensing)
Sensor to Transmitter Differential Sensor: 3000 ft
Distance Combination Sensor: 10 ft
Analog Output
4 to 20mA Isolated Output, Range expand 0 to 100% of full scale (min segment 10% of full scale), max. load 800
300 ft
Memory Back-up All user settings are retained indefinitely in memory (EEPROM)
SHARK
TX Enclosure: NEMA 4X,1/4 DIN, polycarbonate enclosure with two 1/2” conduit holes
SHARK
Mechanical
TXP Enclosure: NEMA 4X front panel, 1/4 DIN, polycarbonate
SHARK
TX Mounting: Universal Mounting kit for surface,pipe and panel mount included
SHARKTXP Mounting: Panel and DIN rail mount included
Sensor Input
Probe: -600 to +600mV Probe:-1999 to +1999mV Cell: 0 to 9999 Temp. Sensor:0 to 9999 Temp. Sensor: 0 to 9999 Temp. Sensor:0 to 9999
Invalid Entries Invalid entries cannot be stored
Manual Test Mode Process value can be simulated with arrow keys to verify correct setup of output
Output Hold 4 to 20mA output is placed on hold when the transmitter is in Menu mode
Recall data from last
Calibration Data
Recall data from last calibration, calibration mode, 1st & 2nd accepted buffer value and probe mV output, calibration temperature, calibration slope, and probe efficiency
calibration, calibration buffer accepted v
alue, and cell Recall store K factor. resistance, calibration temperature
Auto Return User selectable auto return if the transmitter is left in menu mode for more than 10 min.
Display Damping User can select rate at which the transmitter updates display.Enables display damping of unstable process
Net Weight SHARK
TX: 0.71 lbs (0.32 kg)
SHARKTXP: 0.25 lbs (0.12 kg)
Approvals ULC (pending)
Flow: 0 to 9999 with
flow rate units Volume: 0 - 999 with Auto Range Flow rate units: Gallons (GP), Cubic Feet (CF), Liters (LP), Cubic Meters
CM), custom by entering factor related to Gallons Time units: Seconds (S), Minutes (M) Hours (H)
Not required
2000 ft
P
addle: 0 to 2000Hz
selectable
Section I - Specifications Page 2
MULTI-PARAMETER TRANSMITTER USER’S MANUAL
Section 2 - Installation
2.1 Unpacking
Save the shipping carton and packing material in case the instrument needs to be stored or returned. Inspect the instrument and packing material for shipping damage and report any problems immediately.
2.2 Location
Locate the tr
ansmitter close to the sensor. The list below gives typical maximum distances for
various sensors. Refer to the sensor specifications for exact information.
• Aquametrix Differential PH Probe 3000 ft (914 meters)
• Aquametrix Combination PH Probe 10 ft (3 meters)
• Aquametrix Conductivity Probe 300 ft (91 meters)
• Aquametrix Flow sensor 2000 ft (610 meters)
2.3 Mounting
Fig 2.1 SHARKTX Transmitter dimensions
Dwg# N106-128
Section 2 - Installation Page 3
MULTI-PARAMETER TRANSMITTER USER’S MANUAL
EXTERNAL PANEL GASKET QTY.1
SCREW 4-40 X 3/4" QTY.4
NUT 4-40 QTY.4
PANEL (CUSTOMER SUPPLIED)
PANEL (CUSTOMER SUPPLIED)
EXTERNAL PANEL GASKET QTY.1
SCREW 1/4-20 X 2-1/2" QTY.4
NUT 1/4-20 QTY.4
UNIVERSAL MOUNTING CLAMP QTY.2
SCREW 10-24 X 1/2" QTY.4
UNIVERSAL MOUNTING BRACKET QTY.1
Section 2 - Installation
Fig 2.2 SHARKTXP Transmitter dimensions
Dwg# N106-129
Panel Mount – The transmitters can be panel mounted to a panel using the hardware kit provided. The panel cutout dimensions are shown in fig. 2.1 and 2.2.
Figure 2.3 Panel Mount
Dwg# N105-101 & N105-104
Section 2 - Installation Page 4
MULTI-PARAMETER TRANSMITTER USER’S MANUAL
MIN. PIPE DIA.
1" PIPE (NOMINAL)
MAX. PIPE DIA.
2" PIPE (NOMINAL)
NUT 10-24 QTY.4
SCREW 10-24 X 3-1/2" QTY.4
UNIVERSAL MOUNTING BRACKET QTY.1
MIN. PIPE DIA.
1" PIPE (NOMINAL)
MAX. PIPE DIA.
2" PIPE (NOMINAL)
SCREW 10-24 X 3-1/2" QTY.4
NUT 10-24 QTY.4
UNIVERSAL MOUNTING BRACKET QTY.1
UNIVERSAL MOUNTING BRACKET QTY.1
SCREW 10-24 X 1/2" QTY.4
HOLES Ø1/4" FOR MOUNTING SCREWS QTY.4 (CUSTOMER SUPPLIED)
TS35 DIN RAIL MOUNT (CUSTOMER SUPPLIED)
TS32 DIN RAIL MOUNT (CUSTOMER SUPPLIED)
Section 2 - Installation
Surface Mount – The SHARKTX can be surface mounted using the hardware kit provided with the unit.
Pipe Mount – The SHARK
TX can be mounted to a
horizontal or vertical pipe with:
a minimum outside diameter of 1.30” (33mm) (for example 1” CPVC pipe)
and a maximum of 2.375” (60mm) (for example 2” CPVC pipe)
DIN Rail Mount – The SHARK
TXP can be DIN rail
mounted or panel mounted. See figure 2.3 & figure 2.7.
Fig. 2.5 Vertical Pipe Mount
Figure 2.6 Horizontal Pipe Mount
Figure 2.4 Surface Mount
Dwg# N105-101
Dwg# N105-101
Dwg# N105-101
Figure 2.7 DIN Rail Mount
Dwg# N105-104
Section 2 - Installation Page 5
3.1 Conduit Connections
MULTI-PARAMETER TRANSMITTER USER’S MANUAL
Section 3 - Electrical Connections and Setup
The SHARK
TX
has two 1/2” conduit holes at the bottom of the enclosure as shown on fig. 2.1. The unit is shipped with these holes plugged with liquid tight conduit seals. These must be left in unused holes to maintain the NEMA 4X integrity. Use approved conduit hubs to connect the conduit, connect these to the conduit before connecting to the enclosure.
3.2 Power Connections
Figure 3.1 Connections
Dwg# N104-41
Section 3 - Electrical Connections and Setup Page 6
MULTI-PARAMETER TRANSMITTER USER’S MANUAL
Section 3 - Electrical Connections and Setup
3.3 pH and ORP Differential Probe connections and setup
The drawing shows the connections for the Aquametrix Differential (5 wire) probe. The cable should be run in a conduit separate from AC power wires, and via a separate conduit hole.
Figure 3.2 Connections for Differential (5 wire) pH or ORP probe
CAUTION:
Always disconnect the loop power before wiring or unwiring a sensor
Dwg# N104-42
Once connected, step through the LCD menus to select the probe in the order shown. The first two steps may be skipped if the meter is already con­figured for pH or ORP and a Differential Probe. When using a pH probe, it is important to ensure that the transmitter is reading the probe tempera­ture correctly for accurate temperature compensa­tion. The ORP probe does not require temperature compensation, although the transmitter can dis­play process temperature measured by the probe. The factory temperature calibration is usually accurate enough that no adjustments are neces­sary.
Section 3 - Electrical Connections and Setup Page 7
MULTI-PARAMETER TRANSMITTER USER’S MANUAL
Section 3 - Electrical Connections and Setup
3.4 pH or ORP Combination Probe connections and setup
The drawing shows the connections for the Aquametrix Combination probe. The cable should be run in a conduit separate from AC power wires, and via a separate conduit hole. The cable length should not exceed 10 feet (3 meters).
The 2 wire version has no temperature sensor and is connected via a coaxial wire. In a pH meter, the user should set the T COMP OVERRIDE menu to ON (Section 4.10) and adjust the temperature setting to the actual probe temper­ature. In an ORP meter, the user should set the
T.DISP OVERRIDE to ON (Section 5.9) to blank
the temperature reading on the display.
The 4 wire version has two additional wires for the probe internal tem­perature sensor. Ensure that the T COMP OVER-
RIDE or T.DISP OVER­RIDE is OFF.
Fig. 3.3 Connections for the 2 and 4 wire Combination Probe
CAUTION:
Always disconnect the loop power before wiring or unwiring a sensor
Dwg# N104-43
Once connected, step through the LCD menus to select the probe in the order shown. The first two steps may be skipped if the meter is already configured for a Combination Probe. If a two wire pH probe is used, which has no temperature sensor, ensure that the Temp. Comp. Override is set to same temper­ature as the buffer before calibrating. If a two wire ORP probe is used, you can blank the Temp display with the T DISP OVERRIDE menu.
Section 3 - Electrical Connections and Setup Page 8
MULTI-PARAMETER TRANSMITTER USER’S MANUAL
Section 3 - Electrical Connections and Setup
3.5 Conductivity Cell (Contacting style) connections and setup
The drawing shows the connections for the Aquametrix Conductivity Cells (Contacting style). The cable should be run in a conduit seperate from the AC power wires, and via a seperate conduit hole. The cell cable length should not exceed 300ft. (91 meters).
Figure 3.4 Connections for Conductivity Cells
CAUTION:
Always disconnect the loop power before wiring or unwiring a sensor
Dwg# N104-44
Once connnected, step through the LCD menus to select the cell in the order shown. The TEMP COMP CURVE setup default is
1.8%/deg C. This is accept­able for most process appli­cations. If your process is significantly different from this, change the setting in the TEMP COMP CURVE menu.
Section 3 - Electrical Connections and Setup Page 9
MULTI-PARAMETER TRANSMITTER USER’S MANUAL
Section 3 - Electrical Connections and Setup
3.6 Paddle Wheel Flow Sensor connections and setup
The drawing shows the connnections for a typical paddle wheel flow sensor. The cable to the sensor should not exceed 2000’ (600 meters).
Figure 3.5 Connections for Flow Sensor
Dwg# N104-45
CAUTION:
Always disconnect the loop power before wiring or unwiring a sensor
Once connected, step through the LCD menus to select the sensor in the order shown. The Sensor K factor (pulses per U.S. Gallon) is usually printed on the side of the sensor or on a label attached to the sensor cable.
Section 3 - Electrical Connections and Setup Page 10
MULTI-PARAMETER TRANSMITTER USER’S MANUAL
Section 3 - Electrical Connections and Setup
3.10 MANUAL TEST MODE
(LCD MENU SECTIONS - pH: 4.4, ORP: 5.3, Conductivity: 6.4, Flow: 7.2)
The setup can be tested using Manual Test Mode to simulate process changes.
MANUAL TEST MODE is used to simulate a process reading in order to verify the correct response of the output.
3.11 4-20 mA Isolated Output
(LCD MENU SECTIONS - pH: 4.15, ORP: 5.14, Conductivity: 6.16, Flow: 7.14)
The Transmitter has a single 4 to 20mA output, electrically isolated from the ground. The output can source current into a resistive load. Maximum resistance depends on supply voltage. Drawing on page 6 (Fig. 3.1)
The output is dedicated to track the process and has fully independent and fully adjustable 4 & 20 mA output setpoints. This will enable the operator to span the output over the desired range.
The output can be precisely trimmed through the LCD menu for precision applications.
The drawing on page 6 (Fig. 3.1) shows the connections for the output.
Wire Specification
: 22 AWG 7/30, insulation 0.010”
Section 3 - Electrical Connections and Setup Page 11
MULTI-PARAMETER TRANSMITTER USER’S MANUAL
SIMPLE THREE-BUTTON INTERFACE
FOR FAST & EASY SETUP (REFER TO SECTIONS 4.0 TO 4.15)
2 LINE, 16 CHARACTER
LCD DISPLAY
MAIN MENU INTERFACE
SCREEN
1/4" TURN SCREWS
(FOR SHARKTX ONLY)
UNIT LABEL
LABEL TO INDICATE UNIT OF MEASURE SHARK IS CONFIGURED FOR.
Section 4 - Using the Transmitter in pH Mode
Section 4 - Using the Transmitter in pH Mode Page 12
MULTI-PARAMETER TRANSMITTER USER’S MANUAL
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWNDOWN
DOWNDOWN
DOWN
DOWN
DOWNDOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWNDOWN
SELECT DEG C OR DEG F
ENABLE TIME OUT
PROBE WITHOUT TEMP. SENSOR
TEMPERATURE SETUP FOR
SETUP DISPLAY AND OUTPUT RESPONSE TIME
SELECT TYPE OF
FROM MENU
DISPLAY FIRMWAVE REVISION SEC. 4.12
DISPLAY PREVIOUS CALIBRATION DATA SEC 4.13
DISPLAY UNCALIBRATED PROBE DATA SEC 4.14
SEC. 4.9
SEC 4.10
SEC 4.11
SEC. 4.6
SEC 4.7
SELECT TYPE OF TEMP SENSOR SEC 4.8
pH PROBE
RUN MODE
SEC. 4.3
UP
NOTE
PRESS THE AND KEYS
TOGETHER TO GO IMMEDIATELY BACK TO
RUN MODE
DOWN
SELECT TYPE OF METER SEC. 4.5
VERIFY OUTPUTS
SIMULATE PROCESS
SEC. 4.4
AUTO CALIBRATE
MANUAL CALIBRATE
SENSOR IN pH PROBE
CALIBRATE TEMP.
pH PROBE SEC. 4.1
pH PROBE SEC. 4.2
DIAGNOSTICS
EXIT
SENSOR INPUT
CALIBRATION DATA
FIRMWARE REV
EXIT
DISPLAY DAMPING
T.COMP OVERRIDE
EXIT
AUTO RETURN
TEMP UNIT
TEMP SENSOR
PROBE SELECT
7.15pH 25.0C
UTILITIES
SETUP
CALIBRATION
MANUAL TEST MODE
METER SELECTION
EXIT
EXIT
TEMP CALIBRATION
MANUAL CAL pH
AUTO CAL pH
DOWN
DOWN
DOWN
4-20 mA OUTPUT
SETUP
SEC 4.15
EXIT
4-20mA
OUTPUT
pH - Menu Overview 4.0
pH - Menu Overview 4.0 Page 13
MULTI-PARAMETER TRANSMITTER USER’S MANUAL
DOWN
DOWN
DOWN
UP DOWN
DOWN
DOWN
UP DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
UP
DOWN
DOWN
UP
DOWN
DOWN
Please wait for the controller to complete the measurement
After 5 seconds, the controller will compute the slope of the calibration, the estimated probe efficiency and the probe temperature.
Then press to store the value and complete the Manual Calibration
Then press to store the value and move to BUFFER2
Use the and keys to adjust the reading until it agrees with the actual buffer pH value
When complete, the controller will report the measured value
The controller will read the pH value, averaging a number of results to get a stable calibration value.
Place the probe in the second buffer. Be sure to clean and rinse the Probe first with D.I. water and then insert it in the 4.00 buffer.
Press
Use the and keys to adjust the reading until it agrees with the actual buffer pH value
When complete, the controller will report the measured value
Please wait for the controller to complete the measurement
The controller will read the pH value, averaging a number of results to get a stable calibration value.
Place the probe in the first buffer solution, be sure to clean and rinse the Probe first with D.I. water and then insert it in the 7.00 buffer.
Press
If the calibration is OK, use the key to move the cursor over the Y text and press the down key.
If the calibration did not appear to be correct, press the key which will return back to the Manual Cal menu.
RUN MODE
UP
NOTE
PRESS THE AND KEYS
TOGETHER TO GO IMMEDIATELY BACK TO
RUN MODE
DOWN
Press to store the calibration data and return back to the Manual Cal menu so the user can select another function.
Then press to move the cursor to the RH position
Then press to move the cursor to the RH position
STORE? Y N STORE? Y N
IF BUFFER1 READYCALIBRATION MANUAL CAL pH
MANUAL CAL pH
MANUAL CAL pH BUFFER2 4.00 >
SLOPE 61.22MV/pH EFF 95% 24.8C
BUFFER1 7.00 >
MANUAL CAL pH BUFFER2 4.4 0 >
MANUAL CAL pH BUFFER2 4.0 0 >
BUFFER2 WAIT...
RUNNING MANU CAL
IF BUFFER2 READY PRESS 'DOWN'
MANUAL CAL pH
MANUAL CAL pH BUFFER1 7.0 0 >
RUNNING MANU CAL BUFFER1 WAIT...
PRESS 'DOWN'
BUFFER1 7.3 5 >
MANUAL CAL pH
7.15pH 25.0C
MANUAL CAL pH
This example shows a MANUAL calibration with buffers of 7pH & 4pH.
When performing a manual calibration, any two known buffer solutions can be used.
pH - Calibration Menu - Manual Calibrate 4.1
pH - Calibration Menu - Manual Calibrate 4.1 Page 14
MULTI-PARAMETER TRANSMITTER USER’S MANUAL
DOWN
DOWN DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
If an erros occurs, the controller will indicate a "BUFFER ERROR" alarm. This could be caused by
- using a defective buffer
- incorrect probe wiring
- defective probe Press to return to the Auto cal menu, try to
correct the problem and calibrate again, or go to the diagnotics menus to test the probe see Section 4.16
If an erros occurs, the controller will indicate a "BUFFER ERROR" alarm. This could be caused by
- using the same buffer twice
- using a defective buffer
- incorrect probe wiring
- defective probe Press to return to the Auto cal menu, try to
correct the problem and calibrate again, or go to the diagnotics menus to test the probe
If the calibration did not appear to be correct, press the key which will return back to the Auto cal menu.
UP
NOTE
PRESS THE AND KEYS
TOGETHER TO GO IMMEDIATELY BACK TO
RUN MODE
DOWN
This will store the calibration data and return back to the Auto cal setup so the user can select another function.
If the calibration is OK, use the key to move the cursor over the Y text and press the down key.
After 5 seconds, the controller will compute the slope of the calibration, the estimated probe efficiency and the probe temperature.
When complete, the controller will report the measured value
When complete, the controller will report the measured value
Press the key to move to Buffer 2
Please wait for the controller to complete the measurement
The controller will read the pH value, averaging a number of results to get a stable calibration value.
Place the probe in the second buffer solution.Be sure to clean and rinse the Probe first with D.I. water and then insert it in the 4.00 buffer.
Press the key
Please wait for the controller to complete the measurement
The controller will read the pH value, averaging a number of results to get a stable calibration value.
Place the probe in the first buffer solution, be sure to clean and rinse the Probe first with D.I. water and then insert it in the 7.00 buffer.
Press
RUN MODE
AUTO CAL pH
STORE? Y N
STORE? Y N
AUTO CAL pH
AUTO CAL pH
EFF 95% 24.8C
SLOPE 61.22MV/pH
BUFFER2 4.00 >
AUTO CAL pH
AUTO CAL pH
BUFFER ERROR CONTINUE >
RUNNING AUTO CAL BUFFER2 WAIT...
BUFFER1 7.00 >
PRESS 'DOWN'
IF BUFFER2 READY
BUFFER ERROR CONTINUE >
PRESS 'DOWN'
IF BUFFER1 READY
RUNNING AUTO CAL BUFFER1 WAIT...
CALIBRATION
7.15pH 25.0C
MANUAL CAL pH
This example shows an AUTO calibration with buffers of 7pH & 4pH.
When performing an auto calibration, any two standard buffers of 4.00, 7.00 or 10.00 pH can be used. The buffers will be automatically recognized if the probe measures them within +/- 1 pH of their nominal value.
pH - Calibration Menu - Auto Calibrate 4.2
pH - Calibration Menu - Auto Calibrate 4.2 Page 15
MULTI-PARAMETER TRANSMITTER USER’S MANUAL
DOWN
DOWN DOWN
DOWNUP
DOWN
DOWN
DOWN
UP
DOWN
DOWN
DOWN
DOWN
DOWN
If the Temperature Compensation Override is set to ON (see section 4.11), the Shark cannot calibrate the temperature sensor. This display will appear to alert the user to the condition.
If the calibration does not appear to be correct, press the key which will return back to the Temp Calibration menu.
UP
NOTE
PRESS THE AND KEYS
TOGETHER TO GO IMMEDIATELY BACK TO
RUN MODE
DOWN
This will store the calibration data and return back to the Temp Calibration setup so the user can select another function.
Use the and to adjust the reading until it agrees with the known temperature of the probe
Press the key to accept the reading and move to the store function
"T FACTOR SHIFTED" will be displayed for 5 seconds to indicate that the temperature sensor has been calibrated. The Shark will then move to the store function.
The controller displays the current probe temperature. If this incorrect, the controller can be adjusted to compensate.
Press the key once which will move the cursor back to the RH side of temperature display.
Press the key once which will move the cursor over the least sign. digit of the temperature display.
RUN MODE
AUTO CAL pH
STORE? Y N
TEMP CALIBRATION
STORE? Y N
TEMP CALIBRATION
TEMP CALIBRATION
25.0C >
T FACTOR SHIFTED
25. 0 C >
TEMP CALIBRATION
TEMP CALIBRATION
TEMP CALIBRATION
TEMP CALIBRATION
22. 4 C >
22.4C >
TEMP CALIBRATION TEMP O/R ON >
7.15pH 25.0C
CALIBRATION MANUAL CAL pH
pH - Calibration Menu - Temperature Calibration 4.3
pH - Calibration Menu - Temperature Calibration 4.3 Page 16
DOWN
DOWN
DOWN
UP DOWN
DOWN
UP
DOWN
Manual Test Mode is used to simulate a process reading in order to verify the correct response of the output. When in the Manual Test Mode, the output IS no longer placed on hold as it is when in the rest of the menu.
NOTE:
When the user exits the Manual Test Mode, the 4-20mA
output will remain in the Test Mode state until the operator
enters the run mode. The output will then revert back to
the previous On-line state.
The 4-20 mA output will track the change in process value
RUN MODE
UP
NOTE
PRESS THE AND KEYS
TOGETHER TO GO IMMEDIATELY BACK TO
RUN MODE
DOWN
Press the key once to move the cursor to the RH side of the display
The 4 to 20 mA output will follow the process value change.
Press the key to return to the MANUAL TEST MODE menu
Use the and keys to change the simulated process value.
Press the key once which will move the cursor over the least digit of the simulated process value.
TEST 7.0 0 >
TEST 7.00 >
7.15pH 25.0C
CALIBRATION
UTILITIES
19.0mA
19.0mA
TEST 13.1 5 >
TEST 13.15 >
MANUAL TEST MODE
12.0mA
12.0mA
19.0mA
TEST 13.1 5 >
DOWN
MULTI-PARAMETER TRANSMITTER USER’S MANUAL
pH - Utilities Menu - Manual Test Mode 4.4
pH - Utilities Menu - Manual Test Mode 4.4 Page 17
MULTI-PARAMETER TRANSMITTER USER’S MANUAL
DOWNDOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN DOWN
DOWN
If you do not wish to initialize to this type of meter after all, then press key to return back to the Meter Selection menu
Then press The controller will initialize itself for a specific meter
If you do wish to initialize to this type of meter, press the key to highlight the Y character
UP
NOTE
PRESS THE AND KEYS
TOGETHER TO GO IMMEDIATELY BACK TO
RUN MODE
DOWN
Press the key to Exit and return to the Utilities menu
RUN MODE
Press the key to initialize the controller as a Flow meter
Press the key to initialize the controller as a Conductivity meter
The Meter Selection menu is used to select the meter type that the controller is configured for, either pH, ORP, Conductivity or FLow. Once selected, the controller will initialize itself for the selected meter's functionality and move to run mode.
Press the key to initialize the controller as an ORP meter
Press the key to initialize the controller as a pH meter.
INITIALIZE? Y N
INITIALIZE? Y N
INITIALIZE? Y N
INITIALIZE? Y N
INITIALIZE? Y N
PH
PH
E XIT
MANUAL TEST MODE
7.15pH 25.0C
UTILITIES
CALIBRATION
METER SELECTION
C ONDUCTIVITY FLOW
EXIT
F LOW
CONDUCTIVITY
O RP
CONDUCTIVITY
FLOW
ORP
P H
ORP
PH
To reset the controller back to all the factory default values, reselect the current meter type
pH - Utilities Menu - Meter Selection 4.5
pH - Utilities Menu - Meter Selection 4.5 Page 18
MULTI-PARAMETER TRANSMITTER USER’S MANUAL
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
UP DOWN
DOWN
DOWN
DOWN
UP
DOWN
Then press the key to store the selection and return to the PROBE SELECT Menu.
With "N" selected, pressing the key will NOT store the selection, but simply return to the PROBE SELECT Menu. This function is useful if you wish to view the current selection without making any changes.
Not stored
Stored
Or press the key to highlight the Y character.
Note:
If using the 2 wire Combination Probe (which doesn't have a temperature sensor), T COMP OVERRIDE must be set to ON, and the actual probe temperature set through the T COMP OVERRIDE menu. (see sec. 4.11)
PROBE SELECT will allow the user to select whether the probe is a 2 or 4 wire combination probe, or a 5 wire differential probe.
RUN MODE
UP
NOTE
PRESS THE AND KEYS
TOGETHER TO GO IMMEDIATELY BACK TO
RUN MODE
DOWN
Use the or keys to scroll through the probe types available. In this case, the user can select
- the 2 or 4 wire combination probe
- the 5 wire differential probe
Press the key to accept the setting and move to the store function
Press the key once which will move the cursor back to the RH side of the display.
Once the correct probe type is selected, move to the store function to save the selection
Press the key once which will move the cursor over the first character of the probe type.
STORE? Y N STORE? Y N
PROBE SELECT PROBE SELECT
7.15pH 25.0C
UTILITIES
SETUP
CALIBRATION
PROBE SELECT COMBINATION >
C OMBINATION
PROBE SELECT
D IFFERENTIAL
PROBE SELECT
PROBE SELECT PROBE SELECT
DIFFERENTIAL >
Note: Refer to Appendix A - Probe Configuration Table
pH - Setup Menu - Probe Select 4.6
pH - Setup Menu - Probe Select 4.6 Page 19
MULTI-PARAMETER TRANSMITTER USER’S MANUAL
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
UP DOWN
DOWN
UP
DOWN
With "N" selected, pressing the key will NOT store the selection, but simply return to the TEMP UNIT Selection Menu. This function is useful if you wish to view the current selection without making any changes.
RUN MODE
Or press the key to highlight
UP
NOTE
PRESS THE AND KEYS
TOGETHER TO GO IMMEDIATELY BACK TO
RUN MODE
DOWN
Then press the key to store the selection and return to the TEMP UNIT Selection Menu.
Not stored
Stored
the Y character.
Use the or to select C or F
Press the key to accept the change and move to the store function
Press the key once which will move the cursor over the unit type, C or F.
Press the key once which will move the cursor back to the RH side of the display.
TEMP UNIT allows the user to select either Degrees Centigrade or Fahrenheit units for display
7.15pH 25.0C
CALIBRATION
UTILITIES
SETUP
TEMP UNIT
PROBE SELECT
STORE? Y N
STORE? Y N
TEMP UNIT TEMP UNIT
DEG
REE
C >
DEGREE F >
TEMP UNIT
DEGREE F >
DEG
REE
C >
TEMP UNIT
TEMP UNIT
TEMP UNIT
pH - Setup Menu -Temp Unit 4.7
pH - Setup Menu - Temp Unit 4.7 Page 20
MULTI-PARAMETER TRANSMITTER USER’S MANUAL
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
UP
UP
DOWN
DOWN
DOWN
UP
DOWN
the Y character.
Or press the key to highlight
With "N" selected, pressing the key will NOT store the selection, but simply return to the TEMP SENSOR Selection Menu. This function is useful if you wish to view the current selection without making any changes.
Not stored
UP
NOTE
PRESS THE AND KEYS
TOGETHER TO GO IMMEDIATELY BACK TO
RUN MODE
DOWN
Press the key to accept the change and move to the store function
Then press the key to store the selection and return to the TEMP SENSOR Selection Menu.
Stored
TEMP SENSOR allows the user to select the type of temperature sensor used the in the probe.
The factory default for pH is a 300 NTC Thermistor. The user can also select a 3000 NTC Thermistor or a 1000 RTD.
RUN MODE
Use the or to select the sensor type
Once the correct Sensor has been selected press the key once which will move the cursor back to the RH side of the display.
Press the key once which will move the cursor over to the sensor type.
7.15pH 25.0C
SETUP
UTILITIES
CALIBRATION
STORE? Y N
STORE? Y N
TEMP SENSOR
300 NTC >
3 000 NTC >
TEMP SENSOR
TEMP SENSOR
TEMP UNIT
TEMP SENSOR
1 000 RTD >
TEMP SENSOR
TEMP SENSOR
3 00 NTC >
TEMP SENSOR
300 NTC >
TEMP SENSOR
PROBE SELECT
pH - Setup Menu - Temp. Sensor 4.8
pH - Setup Menu - Temp. Sensor 4.8 Page 21
MULTI-PARAMETER TRANSMITTER USER’S MANUAL
DOWN
DOWN
DOWN
DOWN
Not stored
DOWN
Stored
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
UP DOWN
DOWN
UP
UP
NOTE
PRESS THE AND KEYS
TOGETHER TO GO IMMEDIATELY BACK TO
RUN MODE
DOWN
With "N" selected, pressing the key will NOT store the selection, but simply return to the AUTO RETURN Selection Menu. This function is useful if you wish to view the current selection without making any changes.
Then press the key to store the selection and return to the Auto Return Selection Menu.
Press the key to accept the change and move to the STORE function
Or press the key to highlight the Y character.
RUN MODE
To change the MENU RETURN setting, Press the key once which will move the cursor to the ON or OFF text
Use the or keys to select either ON or OFF.
Press the key once which will move the cursor back to the RH side of the display.
MENU ON will cause the controller to exit the menu and revert back to the online run mode after 10 minutes with no buttons pressed. This feature ensures that if a user forgets to return back to run mode, the controller will not be left in an offline state. If for some reason, the user would like to remain in the menu mode for extended periods of time, the AUTO RETURN function can be set to "OFF".
AUTO RETURN is used to select what conditions will cause the controller to time-out of the operations menu
7.15pH 25.0C
SETUP
UTILITIES
CALIBRATION
STORE? Y N
STORE? Y N
AUTO RETURN AUTO RETURN
AUTO RETURNAUTO RETURN MENU ON >
TEMP SENSOR
TEMP UNIT
PROBE SELECTION
AUTO RETURN
MENU OFF >
AUTO RETURN
AUTO RETURN
MENU O FF >
MENU O N >
pH - Setup Menu - Auto Return 4.9
pH - Setup Menu - Auto Return 4.9 Page 22
MULTI-PARAMETER TRANSMITTER USER’S MANUAL
UP
DOWN
UP
DOWN
DOWN
DOWN
DOWN DOWN
DOWN
DOWNUP
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWNUP
DOWN
DOWN
DOWN
Use the or keys to select either ON or OFF.
Press the key to accept the change and move to the Temperature Entry function
Press the key once which will move the cursor back to the RH side of the display.
Press the key once which will move the cursor over the least significant digit of the temperature display.
Use the and to adjust the reading until it agrees with the known temperature of the probe or process
Press the key to accept the reading and move to the store function
Press the key once which will move the cursor back to the RH side of temperature display.
The user can now enter the actual probe or process temperature
Then press the key to store the selection and return to the T. COMP OVERRIDE Menu.
UP
NOTE
PRESS THE AND KEYS
TOGETHER TO GO IMMEDIATELY BACK TO
RUN MODE
DOWN
With "N" selected, pressing the key will NOT store the selection, but simply return to the T. Comp Override Menu. This function is useful if you wish to view the current selection without making any changes.
If you wish to save the setting, press the key to highlight the Y character
Not stored
Stored
RUN MODE
Press the key once which will move the cursor to the ON or OFF text
Temperature Compensation Override is used to manually set the actual probe or process temperature. This is useful if the probe does not have a temperature sensor or if the process temperature is constant. When the override is enabled, the controller will use the selected temperature when performing temperature compensation calculations.
T.COMP OVERRIDE
STORE? Y N
STORE? Y N
TEMP 35. 7 C >
TEMP 35.7C >
T.COMP OVERRIDE
T.COMP OVERRIDE T.COMP OVERRIDE
TEMP 25. 4 C >
T.COMP OVERRIDE
T.COMP OVERRIDE TEMP 25.4C >
AUTO RETURN
T.COMP OVERRIDE
CALIBRATION
UTILITIES
SETUP
7.15pH 25.0C
TEMP UNIT
TEMP SENSOR
PROBE SELECTION
COMPENSATE O FF >
COMPENSATE O N >
T.COMP OVERRIDE
T.COMP OVERRIDE
COMPENSATE O N >
T.COMP OVERRIDE
COMPENSATE OFF >
T.COMP OVERRIDE
pH - Setup Menu - T.Comp Override 4.10
pH - Setup Menu - T.Comp Override 4.10 Page 23
Loading...
+ 57 hidden pages