Section 6 - Using the Transmitter in Conductivity Mode45
LCD Menu46 - 62
Section 7 - Using the Transmitter in Flow Mode63
LCD Menu64 - 78
Appendix A - Probe Configuration Table79
Return Policy & Warranty Plan80
MULTI-PARAMETER TRANSMITTER USER’S MANUAL
Introduction
The SHARKTX & SHARKTXP multi-parameter transmitter is a microprocessor based
transmitter capable of measuring one of the following parameters, pH, ORP, conductivity or flow.
When shipped from the factory, the Shark is not set to measure any one parameter.
When the Shark is powered up for the first time, it will display the meter selection
screen where the meter type must be selected. (refer to section 4.5 Meter Selection)
This meter selection screen will only be displayed when the Shark is powered up for the
first time.
After the user selects a meter type the Shark will remain set to that meter type until it is
changed with the meter selection menu function in the Utilities menu.
To return the Shark to its factory settings, the user must re-select the current meter type
from the meter selection menu function. This will override all set-points and return all
settings back to the factory settings.
The Shark Users menu has been divided into five main categories
- Calibration, used to calibrate the Shark with the selected sensor
- Utilities, used to manually control or override the outputs.
- Setup, used to configure the Sharks many options
- Diagnostics, used to troubleshoot problems with the Shark or sensor
- Outputs, used to configure the Shark’s 4 to 20mA output.
SHARK
TX is packaged in a rugged NEMA 4X polycarbonate enclosure with a universal
mounting kit for surface, panel and pipe-mount applications. This enclosure is perfect for
stand-alone or panel-mount operation.
SHARK
TXP enclosure is also polycarbonate with a NEMA 4X front panel. Panel mount,
and DIN rail mounting hardware are supplied.
IntroductionPage 1
MULTI-PARAMETER TRANSMITTER USER’S MANUAL
Section I - Specifications
pH ORP Conductivity Flow
Display2 x 16 alpha-numeric LCD display
Power Requirements4 to 20mA, Loop Powered, 16 to 32 VDC
MΩ/cm30 to 19.99 0.01
0 to 2.000 0.01
0 to 20.00 0.1
3
0 to 200.0 0.1
0 to 50001.0
0 to 20.00 10
3
0 to 200.0 50
Measuring Range
pH:0.01 to 14.00 ORP: -1999 to +1999mV
(Dependent on sensor)
uS/cm
Temp: 0 to 100°C or Temp: 0 to 100°C or
32° to +212°F 32° to +212°F (
mS/cm
Temp:0 to 100°C or 32° to +212°F
Automatic or Manual
TemperatureAutomatic or Manual
Compensation0 - 100°C (32° to +212°F)
Not r
equired
User selectable temperature
compensation slope 0.0 to 10.0%/°C.
0 to 100°C (32° to +212°F)
Temperature Unit°C or °F Not required
Temperature SensorUser selectable: 300Ω NTC Thermistor, 3000Ω NTC Thermistor or Pt. 1000 RTDNot required
Auto-Calibration
Calibration ModesManual Calibration
Temperature Calibration Temperature Calibration
Manual Calibration
T
emperature Calibration
Dry Calibration
Sample Calibration K factor Input
Ambient ConditionsTemperature:-20°C to +60°C or -4°F to +140°F Humidity: 0 to 90% RH (non-condensing)
Sensor to TransmitterDifferential Sensor: 3000 ft
DistanceCombination Sensor: 10 ft
Analog Output
4 to 20mA
Isolated Output, Range expand 0 to 100% of full scale (min segment 10% of full scale), max. load 800Ω
300 ft
Memory Back-upAll user settings are retained indefinitely in memory (EEPROM)
SHARK
TX Enclosure: NEMA 4X,1/4 DIN, polycarbonate enclosure with two 1/2” conduit holes
SHARK
Mechanical
TXP Enclosure: NEMA 4X front panel, 1/4 DIN, polycarbonate
SHARK
TX Mounting: Universal Mounting kit for surface,pipe and panel mount included
SHARKTXP Mounting: Panel and DIN rail mount included
Sensor Input
Probe: -600 to +600mV Probe:-1999 to +1999mV Cell: 0 to 9999Ω
Temp. Sensor:0 to 9999ΩTemp. Sensor: 0 to 9999ΩTemp. Sensor:0 to 9999Ω
Invalid EntriesInvalid entries cannot be stored
Manual Test ModeProcess value can be simulated with arrow keys to verify correct setup of output
Output Hold4 to 20mA output is placed on hold when the transmitter is in Menu mode
Recall data from last
Calibration Data
Recall data from last calibration, calibration mode,
1st & 2nd accepted buffer value and probe mV output,
calibration temperature, calibration slope, and probe
efficiency
calibration, calibration buffer
accepted v
alue, and cell Recall store K factor.
resistance, calibration
temperature
Auto ReturnUser selectable auto return if the transmitter is left in menu mode for more than 10 min.
Display DampingUser can select rate at which the transmitter updates display.Enables display damping of unstable process
Net WeightSHARK
TX: 0.71 lbs (0.32 kg)
SHARKTXP: 0.25 lbs (0.12 kg)
ApprovalsULC (pending)
Flow:0 to 9999 with
flow rate units
Volume: 0 - 999 with Auto Range
Flow rate units: Gallons (GP), Cubic
Feet (CF), Liters (LP), Cubic Meters
CM), custom by entering factor
related to Gallons
Time units: Seconds (S), Minutes (M)
Hours (H)
Not required
2000 ft
P
addle: 0 to 2000Hz
selectable
Section I - SpecificationsPage 2
MULTI-PARAMETER TRANSMITTER USER’S MANUAL
Section 2 - Installation
2.1 Unpacking
Save the shipping carton and packing material in case the instrument needs to be stored or
returned. Inspect the instrument and packing material for shipping damage and report any
problems immediately.
2.2 Location
Locate the tr
ansmitter close to the sensor. The list below gives typical maximum distances for
various sensors. Refer to the sensor specifications for exact information.
• Aquametrix Differential PH Probe3000 ft (914 meters)
• Aquametrix Combination PH Probe10 ft (3 meters)
• Aquametrix Conductivity Probe300 ft (91 meters)
• Aquametrix Flow sensor2000 ft (610 meters)
2.3 Mounting
Fig 2.1 SHARKTX Transmitter dimensions
Dwg# N106-128
Section 2 - InstallationPage 3
MULTI-PARAMETER TRANSMITTER USER’S MANUAL
EXTERNAL
PANEL GASKET
QTY.1
SCREW
4-40 X 3/4"
QTY.4
NUT
4-40
QTY.4
PANEL
(CUSTOMER SUPPLIED)
PANEL
(CUSTOMER SUPPLIED)
EXTERNAL PANEL
GASKET QTY.1
SCREW
1/4-20 X 2-1/2"
QTY.4
NUT
1/4-20
QTY.4
UNIVERSAL
MOUNTING
CLAMP QTY.2
SCREW
10-24 X 1/2"
QTY.4
UNIVERSAL MOUNTING
BRACKET
QTY.1
Section 2 - Installation
Fig 2.2 SHARKTXP Transmitter dimensions
Dwg# N106-129
Panel Mount – The transmitters can be panel mounted to a panel using the hardware kit
provided. The panel cutout dimensions are shown in fig. 2.1 and 2.2.
Figure 2.3 Panel Mount
Dwg# N105-101 & N105-104
Section 2 - InstallationPage 4
MULTI-PARAMETER TRANSMITTER USER’S MANUAL
MIN. PIPE DIA.
1" PIPE (NOMINAL)
MAX. PIPE DIA.
2" PIPE (NOMINAL)
NUT
10-24
QTY.4
SCREW
10-24 X 3-1/2"
QTY.4
UNIVERSAL MOUNTING
BRACKET
QTY.1
MIN. PIPE DIA.
1" PIPE (NOMINAL)
MAX. PIPE DIA.
2" PIPE (NOMINAL)
SCREW
10-24 X 3-1/2"
QTY.4
NUT
10-24
QTY.4
UNIVERSAL MOUNTING
BRACKET
QTY.1
UNIVERSAL MOUNTING
BRACKET
QTY.1
SCREW
10-24 X 1/2"
QTY.4
HOLES Ø1/4"
FOR MOUNTING SCREWS
QTY.4
(CUSTOMER SUPPLIED)
TS35 DIN RAIL MOUNT
(CUSTOMER SUPPLIED)
TS32 DIN RAIL MOUNT
(CUSTOMER SUPPLIED)
Section 2 - Installation
Surface Mount – The SHARKTX can be surface
mounted using the hardware kit provided with the unit.
Pipe Mount – The SHARK
TX can be mounted to a
horizontal or vertical pipe with:
•a minimum outside diameter of 1.30” (33mm) (for
example 1” CPVC pipe)
•and a maximum of 2.375” (60mm) (for example 2”
CPVC pipe)
DIN Rail Mount – The SHARK
TXP can be DIN rail
mounted or panel mounted. See figure 2.3 & figure 2.7.
Fig. 2.5 Vertical Pipe Mount
Figure 2.6 Horizontal Pipe Mount
Figure 2.4 Surface Mount
Dwg# N105-101
Dwg# N105-101
Dwg# N105-101
Figure 2.7 DIN Rail Mount
Dwg# N105-104
Section 2 - InstallationPage 5
3.1 Conduit Connections
MULTI-PARAMETER TRANSMITTER USER’S MANUAL
Section 3 - Electrical Connections and Setup
The SHARK
TX
has two 1/2” conduit holes at the bottom of the enclosure as shown on fig. 2.1. The
unit is shipped with these holes plugged with liquid tight conduit seals. These must be left in unused
holes to maintain the NEMA 4X integrity. Use approved conduit hubs to connect the conduit, connect
these to the conduit before connecting to the enclosure.
3.2 Power Connections
Figure 3.1 Connections
Dwg# N104-41
Section 3 - Electrical Connections and SetupPage 6
MULTI-PARAMETER TRANSMITTER USER’S MANUAL
Section 3 - Electrical Connections and Setup
3.3 pH and ORP Differential Probe connections and setup
The drawing shows the connections for the Aquametrix Differential (5 wire) probe. The cable
should be run in a conduit separate from AC power wires, and via a separate conduit hole.
Figure 3.2 Connections for Differential (5 wire) pH or ORP probe
CAUTION:
Always disconnect
the loop power
before wiring or
unwiring a sensor
Dwg# N104-42
Once connected, step through the LCD menus to
select the probe in the order shown. The first two
steps may be skipped if the meter is already configured for pH or ORP and a Differential Probe.
When using a pH probe, it is important to ensure
that the transmitter is reading the probe temperature correctly for accurate temperature compensation. The ORP probe does not require temperature
compensation, although the transmitter can display process temperature measured by the probe.
The factory temperature calibration is usually
accurate enough that no adjustments are necessary.
Section 3 - Electrical Connections and SetupPage 7
MULTI-PARAMETER TRANSMITTER USER’S MANUAL
Section 3 - Electrical Connections and Setup
3.4 pH or ORP Combination Probe connections and setup
The drawing shows the connections for the Aquametrix Combination probe. The cable should be run
in a conduit separate from AC power wires, and via a separate conduit hole. The cable length should
not exceed 10 feet (3 meters).
The 2 wire version has
no temperature sensor
and is connected via a
coaxial wire.
In a pH meter, the user
should set the T COMPOVERRIDE menu to ON
(Section 4.10) and adjust
the temperature setting to
the actual probe temperature.
In an ORP meter, the
usershouldsetthe
T.DISPOVERRIDEto
ON (Section 5.9) to blank
the temperature reading
on the display.
The 4 wire version has
two additional wires for
the probe internal temperature sensor. Ensure
that the T COMP OVER-
RIDE or T.DISP OVERRIDE is OFF.
Fig. 3.3 Connections for the 2 and 4 wire Combination Probe
CAUTION:
Always disconnect
the loop power
before wiring or
unwiring a sensor
Dwg# N104-43
Once connected, step through the LCD menus to select the probe in the order shown. The first two steps
may be skipped if the meter is already configured for a Combination Probe. If a two wire pH probe is
used, which has no temperature sensor, ensure that the Temp. Comp. Override is set to same temperature as the buffer before calibrating. If a two wire ORP probe is used, you can blank the Temp display
with the T DISP OVERRIDE menu.
Section 3 - Electrical Connections and SetupPage 8
MULTI-PARAMETER TRANSMITTER USER’S MANUAL
Section 3 - Electrical Connections and Setup
3.5 Conductivity Cell (Contacting style) connections and setup
The drawing shows the connections for the Aquametrix Conductivity Cells (Contacting style). The cable
should be run in a conduit seperate from the AC power wires, and via a seperate conduit hole. The cell
cable length should not exceed 300ft. (91 meters).
Figure 3.4 Connections for Conductivity Cells
CAUTION:
Always disconnect
the loop power
before wiring or
unwiring a sensor
Dwg# N104-44
Onceconnnected,step
through the LCD menus to
select the cell in the order
shown. The TEMP COMP
CURVEsetupdefaultis
1.8%/deg C. This is acceptable for most process applications. If your process is
significantly different from
this, change the setting in
the TEMP COMP CURVE
menu.
Section 3 - Electrical Connections and SetupPage 9
MULTI-PARAMETER TRANSMITTER USER’S MANUAL
Section 3 - Electrical Connections and Setup
3.6 Paddle Wheel Flow Sensor connections and setup
The drawing shows the connnections for a typical paddle wheel flow sensor. The cable to the
sensor should not exceed 2000’ (600 meters).
Figure 3.5 Connections for Flow Sensor
Dwg# N104-45
CAUTION:
Always disconnect
the loop power
before wiring or
unwiring a sensor
Once connected, step through
the LCD menus to select the
sensor in the order shown. The
Sensor K factor (pulses per U.S.
Gallon) is usually printed on the
side of the sensor or on a label
attached to the sensor cable.
Section 3 - Electrical Connections and SetupPage 10
The Transmitter has a single 4 to 20mA output, electrically isolated from the ground. The output
can source current into a resistive load. Maximum resistance depends on supply voltage.
Drawing on page 6 (Fig. 3.1)
The output is dedicated to track the process and has fully independent and fully adjustable 4 &
20 mA output setpoints. This will enable the operator to span the output over the desired range.
The output can be precisely trimmed through the LCD menu for precision applications.
The drawing on page 6 (Fig. 3.1) shows the connections for the output.
Wire Specification
: 22 AWG 7/30, insulation 0.010”
Section 3 - Electrical Connections and SetupPage 11
MULTI-PARAMETER TRANSMITTER USER’S MANUAL
SIMPLE THREE-BUTTON INTERFACE
FOR FAST & EASY SETUP
(REFER TO SECTIONS 4.0 TO 4.15)
2 LINE, 16 CHARACTER
LCD DISPLAY
MAIN MENU INTERFACE
SCREEN
1/4" TURN SCREWS
(FOR SHARKTX ONLY)
UNIT LABEL
LABEL TO INDICATE UNIT OF
MEASURE SHARK IS
CONFIGURED FOR.
Section 4 - Using the Transmitter in pH Mode
Section 4 - Using the Transmitter in pH ModePage 12
MULTI-PARAMETER TRANSMITTER USER’S MANUAL
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWNDOWN
DOWNDOWN
DOWN
DOWN
DOWNDOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWNDOWN
SELECT DEG C OR DEG F
ENABLE TIME OUT
PROBE WITHOUT TEMP. SENSOR
TEMPERATURE SETUP FOR
SETUP DISPLAY AND
OUTPUT RESPONSE TIME
SELECT TYPE OF
FROM MENU
DISPLAY
FIRMWAVE REVISION
SEC. 4.12
DISPLAY PREVIOUS
CALIBRATION DATA
SEC 4.13
DISPLAY UNCALIBRATED
PROBE DATA
SEC 4.14
SEC. 4.9
SEC 4.10
SEC 4.11
SEC. 4.6
SEC 4.7
SELECT TYPE OF
TEMP SENSOR
SEC 4.8
pH PROBE
RUN MODE
SEC. 4.3
UP
NOTE
PRESS THE AND KEYS
TOGETHER TO GO IMMEDIATELY BACK TO
RUN MODE
DOWN
SELECT TYPE
OF METER
SEC. 4.5
VERIFY OUTPUTS
SIMULATE PROCESS
SEC. 4.4
AUTO CALIBRATE
MANUAL CALIBRATE
SENSOR IN pH PROBE
CALIBRATE TEMP.
pH PROBE
SEC. 4.1
pH PROBE
SEC. 4.2
DIAGNOSTICS
EXIT
SENSOR INPUT
CALIBRATION DATA
FIRMWARE REV
EXIT
DISPLAY DAMPING
T.COMP OVERRIDE
EXIT
AUTO RETURN
TEMP UNIT
TEMP SENSOR
PROBE SELECT
7.15pH 25.0C
UTILITIES
SETUP
CALIBRATION
MANUAL TEST MODE
METER SELECTION
EXIT
EXIT
TEMP CALIBRATION
MANUAL CAL pH
AUTO CAL pH
DOWN
DOWN
DOWN
4-20 mA OUTPUT
SETUP
SEC 4.15
EXIT
4-20mA
OUTPUT
pH - Menu Overview 4.0
pH - Menu Overview 4.0Page 13
MULTI-PARAMETER TRANSMITTER USER’S MANUAL
DOWN
DOWN
DOWN
UPDOWN
DOWN
DOWN
UPDOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
UP
DOWN
DOWN
UP
DOWN
DOWN
Please wait for the controller to complete
the measurement
After 5 seconds, the controller will
compute the slope of the calibration, the
estimated probe efficiency and the probe
temperature.
Then press to store the value and
complete the Manual Calibration
Then press to store the value and move
to BUFFER2
Use the and keys to adjust
the reading until it agrees with the actual
buffer pH value
When complete, the controller will report
the measured value
The controller will read the pH value,
averaging a number of results to get a
stable calibration value.
Place the probe in the second buffer.
Be sure to clean and rinse the Probe first with
D.I. water and then insert it in the 4.00 buffer.
Press
Use the and keys to adjust
the reading until it agrees with the actual
buffer pH value
When complete, the controller will report
the measured value
Please wait for the controller to complete
the measurement
The controller will read the pH value,
averaging a number of results to get a
stable calibration value.
Place the probe in the first buffer solution, be
sure to clean and rinse the Probe first with D.I.
water and then insert it in the 7.00 buffer.
Press
If the calibration is OK, use the key
to move the cursor over the Y text and
press the down key.
If the calibration did not appear to be
correct, press the key
which will return back to the Manual
Cal menu.
RUN MODE
UP
NOTE
PRESS THE AND KEYS
TOGETHER TO GO IMMEDIATELY BACK TO
RUN MODE
DOWN
Press to store the calibration
data and return back to the Manual
Cal menu so the user can select
another function.
Then press to move the
cursor to the RH position
Then press to move the
cursor to the RH position
STORE? Y NSTORE? Y N
IF BUFFER1 READYCALIBRATIONMANUAL CAL pH
MANUAL CAL pH
MANUAL CAL pH
BUFFER2 4.00 >
SLOPE 61.22MV/pH
EFF 95% 24.8C
BUFFER1 7.00 >
MANUAL CAL pH
BUFFER2 4.4 0 >
MANUAL CAL pH
BUFFER2 4.0 0 >
BUFFER2 WAIT...
RUNNING MANU CAL
IF BUFFER2 READY
PRESS 'DOWN'
MANUAL CAL pH
MANUAL CAL pH
BUFFER1 7.0 0 >
RUNNING MANU CAL
BUFFER1 WAIT...
PRESS 'DOWN'
BUFFER1 7.3 5 >
MANUAL CAL pH
7.15pH 25.0C
MANUAL CAL pH
This example shows a MANUAL
calibration with buffers of 7pH & 4pH.
When performing a manual
calibration, any two known buffer
solutions can be used.
pH - Calibration Menu - Manual Calibrate 4.1
pH - Calibration Menu - Manual Calibrate 4.1Page 14
MULTI-PARAMETER TRANSMITTER USER’S MANUAL
DOWN
DOWNDOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
If an erros occurs, the controller will indicate a
"BUFFER ERROR" alarm. This could be caused by
- using a defective buffer
- incorrect probe wiring
- defective probe
Press to return to the Auto cal menu, try to
correct the problem and calibrate again, or go to the
diagnotics menus to test the probe
see Section 4.16
If an erros occurs, the controller will indicate a
"BUFFER ERROR" alarm. This could be caused by
- using the same buffer twice
- using a defective buffer
- incorrect probe wiring
- defective probe
Press to return to the Auto cal menu, try to
correct the problem and calibrate again, or go to the
diagnotics menus to test the probe
If the calibration did not appear to
be correct, press the key
which will return back to the Auto cal
menu.
UP
NOTE
PRESS THE AND KEYS
TOGETHER TO GO IMMEDIATELY BACK TO
RUN MODE
DOWN
This will store the calibration
data and return back to the
Auto cal setup so the user
can select another function.
If the calibration is OK, use the key to
move the cursor over the Y text and press the
down key.
After 5 seconds, the controller will
compute the slope of the calibration, the
estimated probe efficiency and the probe
temperature.
When complete, the controller will report the
measured value
When complete, the controller will report
the measured value
Press the key to move to Buffer 2
Please wait for the controller to complete
the measurement
The controller will read the pH value,
averaging a number of results to get a
stable calibration value.
Place the probe in the second buffer
solution.Be sure to clean and rinse the
Probe first with D.I. water and then insert
it in the 4.00 buffer.
Press the key
Please wait for the controller to complete
the measurement
The controller will read the pH value,
averaging a number of results to get a
stable calibration value.
Place the probe in the first buffer solution,
be sure to clean and rinse the Probe first
with D.I. water and then insert it in the 7.00
buffer.
Press
RUN MODE
AUTO CAL pH
STORE? Y N
STORE? Y N
AUTO CAL pH
AUTO CAL pH
EFF 95% 24.8C
SLOPE 61.22MV/pH
BUFFER2 4.00 >
AUTO CAL pH
AUTO CAL pH
BUFFER ERROR
CONTINUE >
RUNNING AUTO CAL
BUFFER2 WAIT...
BUFFER1 7.00 >
PRESS 'DOWN'
IF BUFFER2 READY
BUFFER ERROR
CONTINUE >
PRESS 'DOWN'
IF BUFFER1 READY
RUNNING AUTO CAL
BUFFER1 WAIT...
CALIBRATION
7.15pH 25.0C
MANUAL CAL pH
This example shows an AUTO
calibration with buffers of 7pH & 4pH.
When performing an auto calibration,
any two standard buffers of 4.00, 7.00
or 10.00 pH can be used. The buffers
will be automatically recognized if the
probe measures them within +/- 1 pH
of their nominal value.
pH - Calibration Menu - Auto Calibrate 4.2
pH - Calibration Menu - Auto Calibrate 4.2Page 15
MULTI-PARAMETER TRANSMITTER USER’S MANUAL
DOWN
DOWNDOWN
DOWNUP
DOWN
DOWN
DOWN
UP
DOWN
DOWN
DOWN
DOWN
DOWN
If the Temperature Compensation Override is set to ON
(see section 4.11), the Shark cannot calibrate the
temperature sensor. This display will appear to alert the
user to the condition.
If the calibration does not appear to
be correct, press the key
which will return back to the Temp
Calibration menu.
UP
NOTE
PRESS THE AND KEYS
TOGETHER TO GO IMMEDIATELY BACK TO
RUN MODE
DOWN
This will store the calibration data
and return back to the Temp
Calibration setup so the user can
select another function.
Use the and to adjust
the reading until it agrees with the
known temperature of the probe
Press the key to accept the
reading and move to the store function
"T FACTOR SHIFTED" will be
displayed for 5 seconds to indicate that
the temperature sensor has been
calibrated.
The Shark will then move to the store
function.
The controller displays the current probe
temperature.
If this incorrect, the controller can be
adjusted to compensate.
Press the key once
which will move the cursor
back to the RH side of
temperature display.
Press the key once which will move
the cursor over the least sign. digit of the
temperature display.
RUN MODE
AUTO CAL pH
STORE? Y N
TEMP CALIBRATION
STORE? Y N
TEMP CALIBRATION
TEMP CALIBRATION
25.0C >
T FACTOR SHIFTED
25. 0 C >
TEMP CALIBRATION
TEMP CALIBRATION
TEMP CALIBRATION
TEMP CALIBRATION
22. 4 C >
22.4C >
TEMP CALIBRATION
TEMP O/R ON >
7.15pH 25.0C
CALIBRATIONMANUAL CAL pH
pH - Calibration Menu - Temperature Calibration 4.3
pH - Calibration Menu - Temperature Calibration 4.3Page 16
DOWN
DOWN
DOWN
UPDOWN
DOWN
UP
DOWN
Manual Test Mode is used to simulate a process reading
in order to verify the correct response of the output.
When in the Manual Test Mode, the output IS no longer
placed on hold as it is when in the rest of the menu.
NOTE:
When the user exits the Manual Test Mode, the 4-20mA
output will remain in the Test Mode state until the operator
enters the run mode. The output will then revert back to
the previous On-line state.
The 4-20 mA output will track
the change in process value
RUN MODE
UP
NOTE
PRESS THE AND KEYS
TOGETHER TO GO IMMEDIATELY BACK TO
RUN MODE
DOWN
Press the key once to
move the cursor to the RH side
of the display
The 4 to 20 mA output will follow the process
value change.
Press the key to return to the
MANUAL TEST MODE menu
Use the and keys to change
the simulated process value.
Press the key once which will
move the cursor over the least digit of the
simulated process value.
TEST 7.0 0 >
TEST 7.00 >
7.15pH 25.0C
CALIBRATION
UTILITIES
19.0mA
19.0mA
TEST 13.1 5 >
TEST 13.15 >
MANUAL TEST MODE
12.0mA
12.0mA
19.0mA
TEST 13.1 5 >
DOWN
MULTI-PARAMETER TRANSMITTER USER’S MANUAL
pH - Utilities Menu - Manual Test Mode 4.4
pH - Utilities Menu - Manual Test Mode 4.4Page 17
MULTI-PARAMETER TRANSMITTER USER’S MANUAL
DOWNDOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWNDOWN
DOWN
If you do not wish to initialize to this
type of meter after all, then press
key to return back to the Meter
Selection menu
Then press
The controller will initialize itself for a
specific meter
If you do wish to initialize to this
type of meter, press the
key to highlight the Y character
UP
NOTE
PRESS THE AND KEYS
TOGETHER TO GO IMMEDIATELY BACK TO
RUN MODE
DOWN
Press the key to Exit and
return to the Utilities menu
RUN MODE
Press the key to initialize the
controller as a Flow meter
Press the key to initialize the
controller as a Conductivity meter
The Meter Selection menu is used to select the meter type that the
controller is configured for, either pH, ORP, Conductivity or FLow.
Once selected, the controller will initialize itself for the selected meter's
functionality and move to run mode.
Press the key to initialize the
controller as an ORP meter
Press the key to initialize the
controller as a pH meter.
INITIALIZE? Y N
INITIALIZE? Y N
INITIALIZE? Y N
INITIALIZE? Y N
INITIALIZE? Y N
PH
PH
E XIT
MANUAL TEST MODE
7.15pH 25.0C
UTILITIES
CALIBRATION
METER SELECTION
C ONDUCTIVITY
FLOW
EXIT
F LOW
CONDUCTIVITY
O RP
CONDUCTIVITY
FLOW
ORP
P H
ORP
PH
To reset the controller back to all
the factory default values, reselect
the current meter type
pH - Utilities Menu - Meter Selection 4.5
pH - Utilities Menu - Meter Selection 4.5Page 18
MULTI-PARAMETER TRANSMITTER USER’S MANUAL
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
UPDOWN
DOWN
DOWN
DOWN
UP
DOWN
Then press the key to store the
selection and return to the PROBE
SELECT Menu.
With "N" selected, pressing the
key will NOT store the selection, but
simply return to the PROBE SELECT
Menu. This function is useful if you
wish to view the current selection
without making any changes.
Not stored
Stored
Or press the key to highlight
the Y character.
Note:
If using the 2 wire Combination Probe (which
doesn't have a temperature sensor), T COMP
OVERRIDE must be set to ON, and the actual
probe temperature set through the T COMP
OVERRIDE menu.
(see sec. 4.11)
PROBE SELECT will allow the user to select
whether the probe is a 2 or 4 wire
combination probe, or a 5 wire differential
probe.
RUN MODE
UP
NOTE
PRESS THE AND KEYS
TOGETHER TO GO IMMEDIATELY BACK TO
RUN MODE
DOWN
Use the or keys to scroll through
the probe types available.
In this case, the user can select
- the 2 or 4 wire combination probe
- the 5 wire differential probe
Press the key to accept the
setting and move to the store function
Press the key once
which will move the cursor
back to the RH side of the
display.
Once the correct probe type is selected,
move to the store function to save the
selection
Press the key once
which will move the cursor
over the first character of the
probe type.
STORE? Y NSTORE? Y N
PROBE SELECTPROBE SELECT
7.15pH 25.0C
UTILITIES
SETUP
CALIBRATION
PROBE SELECT
COMBINATION >
C OMBINATION
PROBE SELECT
D IFFERENTIAL
PROBE SELECT
PROBE SELECTPROBE SELECT
DIFFERENTIAL >
Note: Refer to Appendix A - Probe
Configuration Table
pH - Setup Menu - Probe Select 4.6
pH - Setup Menu - Probe Select 4.6Page 19
MULTI-PARAMETER TRANSMITTER USER’S MANUAL
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
UPDOWN
DOWN
UP
DOWN
With "N" selected, pressing the
key will NOT store the selection, but
simply return to the TEMP UNIT
Selection Menu. This function is useful if
you wish to view the current selection
without making any changes.
RUN MODE
Or press the key to highlight
UP
NOTE
PRESS THE AND KEYS
TOGETHER TO GO IMMEDIATELY BACK TO
RUN MODE
DOWN
Then press the key to store the
selection and return to the TEMP UNIT
Selection Menu.
Not stored
Stored
the Y character.
Use the or to select C or F
Press the key to accept the
change and move to the store function
Press the key once
which will move the cursor
over the unit type, C or F.
Press the key once
which will move the cursor
back to the RH side of the
display.
TEMP UNIT allows the user to select either
Degrees Centigrade or Fahrenheit units for
display
7.15pH 25.0C
CALIBRATION
UTILITIES
SETUP
TEMP UNIT
PROBE SELECT
STORE? Y N
STORE? Y N
TEMP UNITTEMP UNIT
DEG
REE
C >
DEGREE F >
TEMP UNIT
DEGREE F >
DEG
REE
C >
TEMP UNIT
TEMP UNIT
TEMP UNIT
pH - Setup Menu -Temp Unit 4.7
pH - Setup Menu - Temp Unit 4.7Page 20
MULTI-PARAMETER TRANSMITTER USER’S MANUAL
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
UP
UP
DOWN
DOWN
DOWN
UP
DOWN
the Y character.
Or press the key to highlight
With "N" selected, pressing the
key will NOT store the selection, but
simply return to the TEMP SENSOR
Selection Menu. This function is useful if
you wish to view the current selection
without making any changes.
Not stored
UP
NOTE
PRESS THE AND KEYS
TOGETHER TO GO IMMEDIATELY BACK TO
RUN MODE
DOWN
Press the key to accept the
change and move to the store function
Then press the key to store the
selection and return to the TEMP
SENSOR Selection Menu.
Stored
TEMP SENSOR allows the user to select the
type of temperature sensor used the in the
probe.
The factory default for pH is a 300 NTC
Thermistor. The user can also select a 3000
NTC Thermistor or a 1000 RTD.
RUN MODE
Use the or to select the
sensor type
Once the correct Sensor has
been selected press the
key once which will move the
cursor back to the RH side of
the display.
Press the key once
which will move the cursor
over to the sensor type.
7.15pH 25.0C
SETUP
UTILITIES
CALIBRATION
STORE? Y N
STORE? Y N
TEMP SENSOR
300 NTC >
3 000 NTC >
TEMP SENSOR
Ω
TEMP SENSOR
TEMP UNIT
TEMP SENSOR
1 000 RTD >
TEMP SENSOR
TEMP SENSOR
3 00 NTC >
TEMP SENSOR
Ω
Ω
300 NTC >
TEMP SENSOR
PROBE SELECT
Ω
Ω
Ω
Ω
pH - Setup Menu - Temp. Sensor 4.8
pH - Setup Menu - Temp. Sensor 4.8Page 21
MULTI-PARAMETER TRANSMITTER USER’S MANUAL
DOWN
DOWN
DOWN
DOWN
Not stored
DOWN
Stored
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
UPDOWN
DOWN
UP
UP
NOTE
PRESS THE AND KEYS
TOGETHER TO GO IMMEDIATELY BACK TO
RUN MODE
DOWN
With "N" selected, pressing the
key will NOT store the selection, but
simply return to the AUTO RETURN
Selection Menu. This function is useful if
you wish to view the current selection
without making any changes.
Then press the key to
store the selection and return
to the Auto Return Selection
Menu.
Press the key to
accept the change and move
to the STORE function
Or press the key to highlight
the Y character.
RUN MODE
To change the MENU
RETURN setting,
Press the key once
which will move the cursor to
the ON or OFF text
Use the or
keys to select either ON or
OFF.
Press the key
once which will move
the cursor back to the
RH side of the display.
MENU ON will cause the controller to exit the menu and
revert back to the online run mode after 10 minutes with no
buttons pressed. This feature ensures that if a user forgets
to return back to run mode, the controller will not be left in
an offline state. If for some reason, the user would like to
remain in the menu mode for extended periods of time, the
AUTO RETURN function can be set to "OFF".
AUTO RETURN is used to select what conditions will
cause the controller to time-out of the operations
menu
7.15pH 25.0C
SETUP
UTILITIES
CALIBRATION
STORE? Y N
STORE? Y N
AUTO RETURNAUTO RETURN
AUTO RETURNAUTO RETURN
MENU ON >
TEMP SENSOR
TEMP UNIT
PROBE SELECTION
AUTO RETURN
MENU OFF >
AUTO RETURN
AUTO RETURN
MENU O FF >
MENU O N >
pH - Setup Menu - Auto Return 4.9
pH - Setup Menu - Auto Return 4.9Page 22
MULTI-PARAMETER TRANSMITTER USER’S MANUAL
UP
DOWN
UP
DOWN
DOWN
DOWN
DOWNDOWN
DOWN
DOWNUP
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWN
DOWNUP
DOWN
DOWN
DOWN
Use the or keys to
select either ON or OFF.
Press the key to accept the
change and move to the Temperature
Entry function
Press the key once
which will move the cursor
back to the RH side of the
display.
Press the key once
which will move the cursor
over the least significant digit
of the temperature display.
Use the and to
adjust the reading until it agrees
with the known temperature of the
probe or process
Press the key to accept the
reading and move to the store function
Press the key once
which will move the cursor
back to the RH side of
temperature display.
The user can now enter the actual
probe or process temperature
Then press the key to store the
selection and return to the T. COMP
OVERRIDE Menu.
UP
NOTE
PRESS THE AND KEYS
TOGETHER TO GO IMMEDIATELY BACK TO
RUN MODE
DOWN
With "N" selected, pressing the
key will NOT store the selection, but
simply return to the T. Comp Override
Menu. This function is useful if you wish
to view the current selection without
making any changes.
If you wish to save the setting, press the
key to highlight the Y character
Not stored
Stored
RUN MODE
Press the key once
which will move the cursor to
the ON or OFF text
Temperature Compensation Override is used to manually set
the actual probe or process temperature. This is useful if the
probe does not have a temperature sensor or if the process
temperature is constant. When the override is enabled, the
controller will use the selected temperature when performing
temperature compensation calculations.
T.COMP OVERRIDE
STORE? Y N
STORE? Y N
TEMP 35. 7 C >
TEMP 35.7C >
T.COMP OVERRIDE
T.COMP OVERRIDET.COMP OVERRIDE
TEMP 25. 4 C >
T.COMP OVERRIDE
T.COMP OVERRIDE
TEMP 25.4C >
AUTO RETURN
T.COMP OVERRIDE
CALIBRATION
UTILITIES
SETUP
7.15pH 25.0C
TEMP UNIT
TEMP SENSOR
PROBE SELECTION
COMPENSATE O FF >
COMPENSATE O N >
T.COMP OVERRIDE
T.COMP OVERRIDE
COMPENSATE O N >
T.COMP OVERRIDE
COMPENSATE OFF >
T.COMP OVERRIDE
pH - Setup Menu - T.Comp Override 4.10
pH - Setup Menu - T.Comp Override 4.10Page 23
Loading...
+ 57 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.