Aquametrix SHARK-120, SHARK-240 User Manual

4 (1)

S H A R K

SHARK MULTI-PARAMETER CONTROLLER

& ANALYZER USER’S MANUAL

Rev 3

AquaMetrix Inc.

Tel: (800)

742-1413

1245 Maple Hill Ct., Unit 7

(905)

954-0841

Newmarket, ON

Fax: (905)

954-0415

Canada, L3Y 9E8

www.aquametrix.com

MULTI-PARAMETER CONTROLLER & ANALYZER USER’S MANUAL

S H A R K

Table of Contents

Subject

Page No.

Introduction

 

 

1

 

Section 1 - Specifications

2

 

Section 2 - Installation

3 - 5

Section 3 - Electrical Connections & Setup

6 - 14

Differential Probe connection & setup

7

 

Combination Probe connection & setup

8

 

Conductivity Cell (Contacting style) connection & setup

9

 

Paddle Wheel Flow Sensor connection & setup

10

 

Relay connections

11

 

Relay A & B setup

12

 

Alarm relay setup

13

 

Manual test mode & Relay override

14

 

4-20mA Isolated Outputs Channel 1 & Channel 2

15

 

Service & Fuse Replacements

16

 

Section 4 - Using the SHARK in pH Mode

17

 

LCD Menu

18

- 44

LED Menu

45

- 46

Section 5 - Using the SHARK in ORP Mode

47

 

LCD Menu

48

- 73

LED Menu

74

 

Section 6 - Using the SHARK in Conductivity Mode

75

 

LCD Menu

76

- 102

LED Menu

103

Section 7 - Using the SHARK in Flow Mode

104

LCD Menu

105 - 128

Appendix A - Probe Configuration Table

129

Return Policy and Warranty Plan

130

MULTI-PARAMETER CONTROLLER & ANALYZER USER’S MANUAL

S H A R K

Introduction

The SHARK multi-parameter controller is a microprocessor based controller capable of measuring one of the following parameters, pH, ORP, conductivity or flow.

When shipped from the factory, the SHARK is not set to measure any one parameter. When the SHARK is powered up for the first time, it will display the meter selection screen where the meter type must be selected. (refer to section 4.6 Meter Selection)

This meter selection screen will only be displayed when the SHARK is powered up for the first time.

After the user selects a meter type the SHARK will remain set to that meter type until it is changed with the meter selection menu function in the Utilities menu.

To return the SHARK to its factory settings, the user must re-select the current meter type from the meter selection menu function. This will override all set-points and return all settings back to the factory settings.

The SHARK User’s menu has been divided into five main categories

-Calibration, used to calibrate the SHARK with the selected sensor

-Utilities, Used to manually control or override the outputs.

-Setup, used to configure the SHARKs many options

-Diagnostics, used to troubleshoot any problems with the SHARK

-Outputs, used to configure the SHARK’s outputs.

There are two displays on the SHARK. A bright LED numeric display with bar graph on the outside front panel, and a 2-line, 16-character LCD display on the inside. The LED readout on the outside panel can be seen several yards away. The distinctive,colorcoded bar graph will immediately indicate if you are within the process parameters that you set (green), if the control relays are on (yellow) and if you are in alarm condition (red). This makes diagnosing pump and alarm malfunctions easy. All configuration and control functions are performed on the LCD menu on the inside front panel.

A universal mounting kit is included for surface,panel and pipe-mount applications. The 1/4 DIN enclosure makes panel-mount cutouts and engineering simple.

SHARK is packaged in a rugged NEMA 4X polycarbonate enclosure making it ideally suited for heavy-duty applications such as industrial wastewater neutralization, municipal water and wastewater, pulp and paper, and process control.

Introduction

Page 1

MULTI-PARAMETER CONTROLLER & ANALYZER USER’S MANUAL

S H A R K

Section I - Specifications

 

 

pH

ORP

 

 

Conductivity

 

 

Flow

Display

Front Panel: 4 x 7 segment 1/2”LED display, 1 LED indicator 0n-line,7 LED Bar Graph

 

 

 

Inside Panel: 2 x 16 alpha-numeric LCD display

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Power Requirements

120Vac (±10%) 50/60Hz (less than 12VA) or 240Vac (±10%) 50/60Hz (less than 12VA)

 

 

 

 

 

 

 

MΩ/cm3

0 to 19.99

 

0.01

 

Flow: 0 to 9999 with selectable

 

 

 

 

 

 

 

 

flow rate units

 

 

 

 

 

 

 

0 to 2.000

 

0.01

 

 

 

 

 

 

 

 

 

 

Volume: 0 to 9999 with Auto Range

Measuring Range

pH:

0.01 to 14.00

ORP: -1999 to +1999mV

 

 

uS/cm3

0 to 20.00

 

0.1

 

Flow rate units:Gallons (GP),Cubic

 

 

 

(Dependent on sensor)

 

 

 

 

 

 

 

 

 

 

 

0 to 200.0

 

0.1

 

 

 

 

 

 

 

 

 

 

Temp:

0 to 100°C or

Temp: 0 to 100°C or

 

 

 

 

 

Feet (CF),Liters (LP),Cubic Meters

 

 

 

 

 

 

 

 

 

 

 

0 to 2000

 

1.0

 

 

 

32° to +212°F

32° to +212°F

 

 

 

 

 

(CM),custom by entering factor

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

mS/cm3

0 to 20.00

 

10

 

related to Gallons

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

0 to 200.0

 

50

 

Time units:Seconds (S),

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Temp: 0 to 100°C or 32° to +212°F

Minutes (M),Hours (H)

 

 

 

 

 

Automatic or Manual

 

 

 

Temperature

Automatic or Manual

Not required

 

User selectable temperature

Not required

Compensation

0 to 100°C (32° to +212°F)

 

compensation slope 0.0 to 10.0%/°C.

 

 

 

 

 

0 to 100°C (32° to +212°F)

 

Temperature Unit

°C or °F

 

 

 

 

 

 

 

 

 

Not required

Temperature Sensor

User selectable: 300Ω NTC Thermistor, 3000Ω NTC Thermistor or Pt. 1000 RTD

 

 

Not required

Calibration Modes

Auto-Calibration

Manual Calibration

 

 

Dry Calibration

 

 

 

Manual Calibration

 

 

Sample Calibration

 

 

K factor Input

Temperature Calibration

 

 

 

 

 

Temperature Calibration

 

 

Temperature Calibration

 

 

 

 

 

 

Ambient Conditions

Temperature: -20°C to +60°C or -4°F to +140°F Humidity: 0 to 90% RH (non-condensing)

 

Menu Access

Auto-Calibration, Manual

Manual-Calibration,

 

 

Manual Calibration

 

 

 

Calibration,Temperature

 

 

 

 

Not available

Front Panel

Temperature Display

 

 

Temperature Display

 

 

Display

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Menu Access Inside Panel

Full Access to all parameters of operations menu

 

 

 

 

 

 

 

Sensor to SHARK

Differential Sensor: 3000 ft

 

 

 

300 ft

 

 

2000 ft

Distance

Combination Sensor: 10 ft

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Two Control Relays, 10A / NO, 5A / NC @ 240VAC or 28VDC.

 

 

 

 

Mode: Process control, Adjustable parameters: process direction,(rising or falling) on-set-point, off set-point,

Relay Outputs

(0 to 100% of full scale), cycle timer (on / off, 0 to 600 seconds), failsafe (on / off).

 

 

 

One Alarm Relay, 10A / NO, 5A / NC @ 240VAC or 28VDC.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Mode: High / Low Alarm, Adjustable parameters: Low on / Low off set-point (0 to 100% of full scale, low on must be less than

 

low off), High On / High Off set-point (0 to 100% of full scale, High on must be greater than High off).

 

4 to 20mA Channel 1

 

 

 

 

 

 

 

 

 

Analog Outputs

Isolated Output, Range expand 0 - 100% of full scale (min segment 10% of full scale), max. load 800Ω

4 to 20mA Channel 2

 

 

 

 

 

 

 

 

 

 

Isolated Output, Range expand 0 - 100% of full scale (min segment 10% of full scale), max. load 800Ω

 

Can be set to track temperature if sensor is equipped with a temperature sensor

 

 

 

Memory Back-up

All user settings are retained indefinitely in memory (EEPROM)

 

 

 

Mechanical

Enclosure: NEMA 4X, 1/4 DIN, polycarbonate enclosure with four 1/2”conduit holes

 

 

 

Mounting: Universal Mounting kit for surface, pipe and panel mount, is included

 

 

 

 

 

 

 

Sensor Input

Probe: -600 to +600mV

Probe: -1999 to +1999mV

 

 

Cell: 0 to 9999Ω

 

 

Paddle: 0 to 2000Hz

Temp. Sensor: 0 to 9999Ω

Temp. Sensor: 0 to 9999Ω

 

 

Temp. Sensor: 0 to 9999Ω

 

 

 

 

Invalid Entries

Invalid entries cannot be stored

 

 

 

 

 

 

 

Manual Test Mode

Process value can be simulated with arrow keys to verify correct setup of outputs

 

 

 

Manual Relay Override

Relays can be set to on / off / auto, to verify correct wiring of auxiliary devices, or to manually adjust process

Output Hold

All outputs are placed on hold when SHARK is in Menu mode

 

 

 

 

Recall data from last calibration, calibration mode,

 

 

Recall data from last

 

 

 

 

 

 

calibration, calibration buffer

 

 

1st & 2nd accepted buffer value and probe mV output,

 

 

 

Calibration Data

 

 

accepted value, and cell

Recall store K factor.

calibration temperature, calibration slope, and probe

 

 

 

 

 

resistance, calibration

 

 

 

 

efficiency

 

 

 

 

 

 

 

 

 

 

temperature

 

 

 

 

 

 

 

 

 

 

 

 

Auto Return

User selectable auto return if SHARK is left in menu mode or if relays are left in manual override mode for more than 10 min.

Display Damping

User can select rate at which SHARK updates display. Enables display damping of unstable process

Net Weight

2.2lbs (1kg)

 

 

 

 

 

 

 

 

 

Approvals

ULC (pending)

 

 

 

 

 

 

 

 

 

Section I - Specifications

Page 2

MULTI-PARAMETER CONTROLLER & ANALYZER USER’S MANUAL

S H A R K

Section 2 - Installation

2.1 Unpacking

Save the shipping carton and packing material in case the instrument needs to be stored or returned. Inspect the instrument and packing material for shipping damage and report any problems immediately.

2.2 Location

Locate the controller/analyzer close to the sensor. The list below gives typical maximum distances for various sensors. Refer to the sensor specifications for exact information.

• Aquametrix Differential PH Probe

3000 ft (914 meters)

• Aquametrix Combination PH Probe

10 ft (3 meters)

• Aquametrix Conductivity Probe

300 ft (91 meters)

• Aquametrix Flow sensor

2000 ft (610 meters)

2.3 Mounting

Fig 2.1 Controller dimensions

Dwg# N106-127

Section 2 - Installation

Page 3

MULTI-PARAMETER CONTROLLER & ANALYZER USER’S MANUAL

S H A R K

Section 2 - Installation

Panel Mount – The Shark can be

Figure 2.2 Panel Mount

panel mounted to a panel using the

EXTERNAL PANEL

hardware kit provided. The panel

GASKET QTY.1

cutout dimensions are shown in fig.

PANEL

(CUSTOMER SUPPLIED)

2.1.

UNIVERSAL MOUNTING

 

BRACKET

 

QTY.1

 

 

 

SCREW

 

 

 

10-24 X 1/2"

 

 

 

QTY.4

 

 

 

SCREW

 

 

 

1/4-20 X 6"

 

 

UNIVERSAL

QTY.4

 

 

MOUNTING

 

 

 

CLAMP QTY.2

 

 

 

 

NUT

 

 

 

1/4-20

 

 

 

QTY.4

 

 

Dwg# N105-100

 

Pipe Mount – The Shark

Fig. 2.3

Vertical Pipe Mount

 

can also be mounted to a

 

 

 

horizontal or vertical pipe

 

 

MIN. PIPE DIA.

with:

 

 

1" PIPE (NOMINAL)

 

 

MAX. PIPE DIA.

 

 

 

2" PIPE (NOMINAL)

• a minimum outside

 

 

 

diameter of 1.30” (33mm)

 

 

 

(for example 1” CPVC

 

 

UNIVERSAL MOUNTING

 

 

BRACKET

pipe)

 

 

QTY.1

 

 

 

• and a maximum of

 

 

 

2.375” (60mm) (for

 

 

 

example 2” CPVC pipe)

 

 

NUT

 

 

 

 

 

 

10-24

 

 

 

QTY.4

 

 

 

SCREW

 

 

 

10-24 X 3-1/2"

 

 

 

QTY.4

 

Dwg# N105-100

 

Section 2 - Installation

Page 4

MULTI-PARAMETER CONTROLLER & ANALYZER USER’S MANUAL

S H A R K

Section 2 - Installation

Figure 2.4 Horizontal Pipe Mount

MIN. PIPE DIA. 1" PIPE (NOMINAL) MAX. PIPE DIA. 2" PIPE (NOMINAL)

 

 

NUT

 

 

10-24

 

 

QTY.4

 

 

SCREW

 

 

10-24 X 3-1/2"

 

 

QTY.4

 

 

UNIVERSAL MOUNTING

 

 

BRACKET

 

 

QTY.1

Dwg# N105-100

 

 

Surface Mount – The Shark can be

Figure 2.5

Surface Mount

surface mounted using the hardware

 

 

kit provided with the unit.

 

 

 

 

UNIVERSAL MOUNTING

 

 

BRACKET

 

 

QTY.1

 

 

SCREW

 

 

10-24 X 1/2"

 

 

QTY.4

 

 

HOLES Ø1/4"

 

 

FOR MOUNTING SCREWS

 

 

QTY.4

 

 

(CUSTOMER SUPPLIED)

 

Dwg# N105-100

Section 2 - Installation

Page 5

MULTI-PARAMETER CONTROLLER & ANALYZER USER’S MANUAL

S H A R K

Section 3 - Electrical Connections and Setup

3.1 Conduit Connections

The Shark has four 1/2” conduit holes, 2 on each side of the enclosure as shown on fig. 2.1. The unit is shipped with these holes plugged with liquid tight conduit seals. These must be left in unused holes to maintain the NEMA 4X integrity. Use approved conduit hubs to connect the conduit, connect these to the conduit before connecting to the enclosure.

Wire Specification: Size and fuse wire accroding to local electrical code. Maximum current not to exceed 10A when used to power auxillary decvices powered via internal connections.

3.2 A.C. Power Connections

Caution: This instrument uses 120 or 240 50/60 Hz AC power. Opening the enclosure door exposes you to potentially hazardous line power voltage which might be present on the terminals of plug P3 and P4. Always remove line power before working in this area. If the relay contacts on P4 are powered from a seperate source from the line power on P3, be sure to disconnect that power before proceeding. The Shark flip out door contains only low voltage and is safe to handle.

The Shark is available in two power models.

The Shark-240 is designed to operate at 240 VAC. The Shark-120 is designed to operate at 120 VAC.

To connect power to the Shark, remove the terminal block plug P3 and connect the wiring as shown below.

Figure 3.1 A.C. Power Connections

AC POWER CONNECTIONS FOR 120 VAC

SHARK-120

NEUTRAL LINE GROUND

G L N

AC POWER CONNECTIONS FOR 240 VAC SHARK-240

LINE

LINE

GROUND

P3

G L L

P3

Dwg# N104-33

Section 3 - Electrical Connections and Setup

Page 6

MULTI-PARAMETER CONTROLLER & ANALYZER USER’S MANUAL

S H A R K

Section 3 - Electrical Connections and Setup

3.3 pH and ORP Differential Probe connections and setup

The drawing shows the connections for the Aquametrix Differential (5 wire) probe. The cable should be run in a conduit separate from AC power wires, and via a separate conduit hole.

Note: Leave 4” to 6” slack for all wires connected to the terminals of P6. Slack required so that wires do not interfere with opening or closing of the front door.

If the cable of the differential probe is cut, the blue

wire is not used.

CAUTION:

Always remove line power before unplugging or plugging in the P6 connector

Figure 3.2 Connections for Differential (5 wire) pH or ORP probe

 

 

GREEN (2)

 

 

 

 

 

 

 

 

 

 

 

RED (3)

 

 

 

 

 

 

 

 

 

 

 

 

BLACK (4)

 

 

 

 

 

 

 

 

 

 

 

YELLOW (5)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

WHITE (10)

 

 

 

 

 

 

 

 

 

 

 

 

SHIELD (8)

 

P6

1

2

3

4

5

6

7

8

9

10

11 12

13 14 15 16

17

 

 

 

 

 

 

 

 

 

 

 

 

 

DIFFERENTIAL

Dwg# N104-34

 

 

 

 

 

 

 

 

 

 

 

 

pH or ORP PROBE

 

 

 

 

 

 

 

 

 

 

 

 

 

Once connected, step through the LCD menus to select the probe in the order shown. The first two steps may be skipped if the meter is already configured for pH or ORP and a Differential Probe.

When using a pH probe, it is important to ensure that the Shark is reading the probe temperature correctly for accurate temperature compensation. The ORP probe does not require temperature compensation, although the Shark can display process temperature measured by the probe. The factory temperature calibration is usually accurate enough that no adjustments are necessary.

METER SELECTION

PROBE SELECT

MANUAL CAL PH

7.15pH 25.0C

SELECT pH SEC. 4.6 or ORP SEC. 5.5

(IF NECESSARY)

SELECT DIFFERENTIAL PROBE pH SEC. 4.7 or ORP SEC. 5.6 (IF NECESSARY)

MANUAL CALIBRATE pH PROBE SEC. 4.1

or ORP PROBE SEC. 5.1

RUN MODE

Section 3 - Electrical Connections and Setup

Page 7

MULTI-PARAMETER CONTROLLER & ANALYZER USER’S MANUAL

S H A R K

Section 3 - Electrical Connections and Setup

3.4 pH or ORP Combination Probe connections and setup

The drawing shows the connections for the Aquametrix Combination probe. The cable should be run in a conduit separate from AC power wires, and via a separate conduit hole. The cable length should not exceed 10 feet (3 meters).

The 2 wire version has no

Fig. 3.3 Connections for the 2 and 4 wire Combination Probe

 

 

 

 

 

 

temperature

sensor

and is

 

 

 

GREEN (4)

 

connected via a coaxial wire.

 

 

 

WHITE (5)

 

COAX CENTER (3)

 

 

 

 

In a pH meter,

the

user

 

 

 

These wires are only

should

set

the

T

COMP

COAX SHIELD (4)

 

 

JUMPER 2-4 MUST BE

present with 4-wire

 

 

combination probe.

 

 

 

 

 

 

 

 

 

 

OVERRIDE

menu

to

ON

 

 

 

INSTALLED FOR

 

 

 

 

COMBINATION PROBE

 

(Section 4.11) and adjust the

 

 

 

 

 

 

 

(CUSTOMER SUPPLIED)

 

temperature

setting

to

the

 

 

 

 

 

actual probe temperature.

 

 

 

 

 

In an ORP meter, the user

 

 

 

 

 

should set the T.DISP OVER-

 

 

 

 

 

RIDE to ON (Section 5.10) to

P6 1

2 3

4 5 6 7 8

9 10 11 12 13 14 15 16 17

 

blank the temperature reading

 

 

 

 

 

on the display.

 

 

 

 

 

 

 

 

The 4 wire version has two

 

 

 

 

 

additional wires for the probe

 

 

 

 

 

internal

temperature

sensor.

 

 

 

 

 

Ensure that the T COMP

 

 

 

 

 

OVERRIDE

or

T.DISP

 

 

 

 

 

OVERRIDE is OFF.

 

 

 

 

 

 

 

Note: Leave 4” to 6” slack for

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

all wires connected to the ter-

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

minals of P6. Slack required

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

COMBINATION

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

so wires do not interfere with

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

pH or ORP PROBE

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

opening/closing of front door.

Dwg# N104-35

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Once connected, step throught the LCD menus to select the probe in the order

 

 

CAUTION:

 

 

 

 

 

 

 

 

shown. The first two steps may be skipped if the meter is already configured for

 

 

Always remove line

 

 

 

 

a Combination Probe. If a two wire pH probe is used, which has no temperature

 

 

power before unplugging

 

 

 

 

sensor, ensure that the Temp. Comp. Override is set to same temperature as

 

 

or plugging in the P6

 

 

 

 

the buffer before calibrating. If a two wire ORP probe is used, you can blank the

 

 

connector

 

 

 

 

 

 

 

 

Temp display with the T DISP OVERRIDE menu.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

pH

 

 

 

 

 

 

 

 

 

 

 

ORP

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

SELECT PH METER

 

 

 

 

 

 

 

 

 

 

 

SELECT ORP METER

 

 

 

 

 

 

METER SELECTION

 

 

 

 

 

 

 

 

 

METER SELECTION

 

 

 

 

 

 

 

 

SEC. 4.6

 

 

 

 

 

 

 

 

 

SEC. 5.5

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

SELECT COMBINATION PROBE

 

 

 

 

SELECT COMBINATION PROBE

 

PROBE SELECT

 

 

PROBE SELECT

 

 

 

SEC. 4.7

 

 

 

 

 

 

 

 

 

SEC. 5.8

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

TWO WIRE PROBE WITH NO TEMP SENSOR

 

 

 

TWO WIRE PROBE WITH NO TEMP SENSOR

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

MANUAL CALIBRATE

 

 

 

 

 

 

 

 

 

 

 

MANUAL CALIBRATE

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

MANUAL CAL PH

 

 

T.COMP OVERRIDE

 

MANUAL CAL PH

 

T.DISP OVERRIDE

 

 

 

 

 

 

PH PROBE

 

 

 

 

 

 

 

 

 

 

 

ORP PROBE

 

 

 

 

 

 

 

 

 

 

SEC. 4.1

 

 

MANUAL PROBE

 

 

 

 

SEC. 5.1

DISABLE TEMPERATURE

 

 

 

 

 

 

 

TEMPERATURE SETUP

 

 

 

 

DISPLAY

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

7.15pH 25.0C

 

RUN MODE

 

 

SEC 4.10

 

7.15pH 25.0C

 

RUN MODE

SEC 5.10

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Section 3 - Electrical Connections and Setup

Page 8

MULTI-PARAMETER CONTROLLER & ANALYZER USER’S MANUAL

S H A R K

Section 3 - Electrical Connections and Setup

3.5 Conductivity Cell (Contacting style) connections and setup

The drawing shows the connections for the Aquametrix Conductivity Cells (Contacting style). The cable should be run in a conduit seperate from the AC power wires, and via a seperate conduit hole. The cell cable length should not exceed 300ft. (91 meters).

Note: Leave 4” to 6” slack for all wires connected to the terminals of P6. Slack required so that wires do not interfere with opening or closing of the front door.

CAUTION:

Always remove line power before unplugging or plugging in the P6 connector

Figure 3.4 Connections for Conductivity Cells

 

 

 

 

 

 

WHITE (CELL) (1)

 

 

 

 

 

 

 

 

 

 

BLACK (CELL) (4)

 

 

 

 

 

 

 

 

 

 

 

RED (TEMP. SENSOR) (4)

 

 

 

 

 

 

 

 

 

 

 

GREEN (TEMP. SENSOR) (5)

P6

1

2

3

4

5

6

7

8

9

10

11 12

13 14 15 16

17

 

 

 

 

 

 

 

 

 

 

 

 

 

CONTACTING

 

 

 

 

 

 

 

 

 

 

 

 

 

CONDUCTIVITY

Dwg# N104-36

 

 

 

 

 

 

 

 

 

 

 

 

CELL

 

 

 

 

 

 

 

 

 

 

 

 

 

Once connnected, step through the LCD menus to select the cell in the order shown. The TEMP COMP CURVE setup default is 1.8%/deg C. This is acceptable for most process applications. If your process is significantly different from this, change the setting in the TEMP COMP CURVE menu.

METER SELECTION

COND RANGE

TEMP SENSOR

TEMP COMP CURVE

OR

SELECT COND METER

SEC. 6.6

SELECT CONDUCTIVITY RANGE SEC. 6.7

ENSURE SENSOR IS CORRECT TYPE

3Kohm NTC (thermistor)

1Kohm RTD

SEC 6.9

SET TEMPERATURE COMPENSATION FOR PROCESS

SEC. 6.13

CALIBRATE WITH

MANUAL CAL COND REFERENCE SOLUTIONS DRY CAL COND

SEC 6.1

1000uS

25.0C

RUN MODE

CALIBRATE WITH FACTORY SPECIFIED CELL CONSTANT SEC 6.2

Section 3 - Electrical Connections and Setup

Page 9

MULTI-PARAMETER CONTROLLER & ANALYZER USER’S MANUAL

S H A R K

Section 3 - Electrical Connections and Setup

3.6 Paddle Wheel Flow Sensor connections and setup

The drawing shows the connnections for a typical paddle wheel flow sensor. The cable to the sensor should not exceed 2000’ (600 meters).

The Shark controller also supports the use of an external “flow switch”. When the flow switch input is grounded, either through a dry contact or solid state input, the flow display will be held at zero. This is useful to ensure the flow reading remains locked at zero when conditions require it. The flow will start reading again when the input is opened. If the flow switch function is not desired, simply leave it disconnected and the flow meter will read as normal.

Figure 3.5 Connections for Flow Sensor

SHIELD

 

 

 

 

 

 

 

 

+12V (6)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

SIGNAL + (14)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

SIGNAL - (GND) (15)

 

 

 

 

 

(FLOW SWITCH INPUT) (13)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

GND (15)

 

 

 

9ma

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

+12V O.C.

 

 

 

 

 

 

 

 

 

 

 

 

 

DRY

 

 

OR

 

OPEN

 

 

 

 

P6 1

2 3

4 5 6 7 8 9 10 11 12 13 14 15 16

17

 

 

CONTACT

 

 

 

COLLECTOR

 

 

 

CLOSING THE FLOW SWITCH INPUT

 

 

FLOW SENSOR

 

 

 

 

TO GROUND WILL ZERO THE FLOW DISPLAY

EXTERNAL POWERED

 

 

 

 

 

 

 

 

 

 

 

 

 

SHIELD

SIGNAL + (14)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

SIGNAL - (15)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

SHIELD (15)

 

 

 

 

1

2

3

4

5

6

7

8

9

10

11 12

13 14 15 16

17

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

FLOW SENSOR

Dwg# N104-37

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

SELF POWERED

Note: Leave 4” to 6” slack for all wires connected to the terminals of P6. Slack required so that wires do not interfere with opening or closing of the front door.

CAUTION:

Always remove line power before unplugging or plugging in the P6 connector

Once connected, step through the LCD menus to select the sensor in the order shown. The Sensor K factor (pulses per U.S. Gallon) is usually printed on the side of the sensor or on a label attached to the sensor cable.

METER SELECTION

SELECT FLOW METER

SEC. 7.4

 

 

ENTER FLOW SENSOR

K FACTOR

CALIBRATION FACTOR

 

 

SEC. 7.1

 

SETUP OF UNITS

UNITS OF VOLUME

OF VOLUME

 

 

SEC 7.5

 

SETUP OF UNITS

UNITS OF TIME

OF TIME

 

 

SEC. 7.6

 

RESET TOTALIZER

TOTALIZER RESET

TO ZERO

 

SEC 7.0

TOTAL

0

RUN MODE

Section 3 - Electrical Connections and Setup

Page 10

MULTI-PARAMETER CONTROLLER & ANALYZER USER’S MANUAL

S H A R K

Section 3 - Electrical Connections and Setup

3.7 Relay connections

The Shark controller has three internal relays. Relays A and B are for control, the Alarm Relay can be configured for alarm functions or as an additional control relay.

The connections to the relays are shown in the drawing. Note that the AC power is internally connected to the relay terminal plug P4. This is used to provide 120V or 240V AC power for the relay contacts.

Wire Specification: Size and fuse wire accroding to local electrical code. Wire size not to exceed 14 AWG.

WARNING

DISCONNECT POWER FROM CONTROLLER AND LOADS WHILE CONNECTING TO THE RELAY OUTPUT TERMINAL PLUG.

Figure 3.6 Connections for Relay A, B and Alarm

 

RELAY CONNECTIONS FOR 120VAC

RELAY CONNECTIONS FOR 240vac

SHARK-120

SHARK-240

 

N.0

RELAY

 

 

 

 

N.C.

A

 

 

 

 

N.C.

RELAY

 

N.O.

 

 

 

 

 

B

 

N.C.

ALARM

 

N.O.

 

 

Internal

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

connections

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Dwg# N104-38

21 22 23 24 25 26 27 28 29 30 31

P4

N L GNDG

P3

 

W A R N I N G

N.0

A RELAY

 

PIN 21 must be

N.C.

 

connected to relays

 

 

 

for power.

N.C.

RELAY

 

PIN 22 cannot be

 

 

N.O.

 

 

connected to the

 

B

 

 

 

 

relays.

N.C.

ALARM

 

 

 

 

N.O.

 

NEUTRAL OUT (22)

 

 

LINE OUT (21)

Internal connections

NEUTRAL

LINE

GROUND

21 22 23 24 25 26 27 28 29 30 31

P4

L L GNDG

P3

W A R N I N G

PIN 21 must be connected to relays for power.

PIN 22 cannot be connected to the relays.

LINE OUT (22)

LINE OUT (21)

LINE

LINE

GROUND

Caution:

The contacts are rated at 10 amp N.O. and 5 amp N.C. Do not exceed this rating. When switching larger currents, use an auxillary relay switched by the controller relay to extend the controller relay life. If the relays are controlling an inductive load, use appropriate transient suppression at the load.

Section 3 - Electrical Connections and Setup

Page 11

MULTI-PARAMETER CONTROLLER & ANALYZER USER’S MANUAL

S H A R K

Section 3 - Electrical Connections and Setup

3.8 RELAY A and B Setup

(LCD MENU SECTIONS - pH: 4.18 & 4.19, ORP: 5.17 & 5.18,

Conductivity: 6.18 & 6.19, Flow: 7.15 & 7.16)

Relay A & Relay B on the SHARK are SPDT dry contact relays. They are configurable to operate in response to rising or falling process values. Each relay has independently adjustable on and off setpoints, cycle times, and fail-safe options.

The operator would use the control relays if the device to be controlled is a simple on/off device. For example a pump, solenoid valve, fan, or an indicating light.

The control relays have 6 user configurable settings:

DIRECTION: The relay can be set to control either a rising or falling process. If for example the relay is set to control a falling process, the ON set-point must be set lower than the OFF set-point. If the relay is set to control a falling process the SHARK will not allow the RELAY OFF set point to be set lower than the RELAY ON set-point. This rule will also apply to a rising process.

RELAY ON set-point: This is the process value at which the relay will energize. This value can be set anywhere between 0-100% of the range.

RELAY OFF set-point: This is the process value at which the relay will de-energize. Depending on the direction for which the relay is configured, the RELAY OFF set-point will only be setable in a limited range.

CYCLE ON time: To obtain a tighter process control, and limit over-shoot, the control relay can be set with the cycling feature. This feature, if enabled, will cause the control relay to cycle when the process is between the RELAY ON set-point and RELAY OFF set-point. The cycle on time is the amount of time in seconds that the relay will be energized. It can be set between 0 and 600 seconds.

CYCLE OFF time: The CYCLE OFF time is the amount of time in seconds that the relay will be deenergized, it can be set between 0 and 600 seconds. To disable the cycling feature set the cycle off time to 0.

OVERFEED TIMER: The overfeed timer is designed to help safeguard against a process or instrumentation error causing one of the control relays to remain energized for extended periods of time.

When enabled, the overfeed timer will time out if the control RELAY OFF set point is not reached inside the overfeed time out. The control relays will de-energize, the alarm relay will energize and an LED will flash at the front.

FAILSAFE: The FAILSAFE feature is designed to reverse the normal action of the control relay.

When the relay is set to FAILSAFE OFF the relay will operate as a normal relay. When the relay is de-energized the NO contacts are open and the NC contacts are closed. Thus the device connected via the NO contacts will be off. When the relay becomes energized the device will be on.

When the relay is set to FAILSAFE ON, the normal action of the relay is reversed. Thus the NO contacts act as the NC contact and the NC act as the NO. The device connected to the NC contacts will be energized when the RELAY ON set-point is reached. The relay will be de-energized but because it is acting in reverse the device will be energized. When the RELAY OFF set-point is reached the relay will energize and the device connected to the NC contact will de-energize.

The purpose of the Fail Safe option is to have the device turned on in the event of a power interruption.

The factory default for FAILSAFE is OFF.

Section 3 - Electrical Connections and Setup

Page 12

MULTI-PARAMETER CONTROLLER & ANALYZER USER’S MANUAL

S H A R K

Section 3 - Electrical Connections and Setup

3.9 ALARM RELAY Setup

(LCD MENU SECTIONS - pH: 4.20, ORP: 5.19, Conductivity: 6.20, Flow: 7.17)

The third relay (Relay C) is used as an alarm relay. The alarm relay on the SHARK is a SPDT dry contact relay.

This relay will respond to both a rising and falling process. The alarm relay will act as a low alarm (falling process) and a high alarm (rising process). Both relays will have independently adjustable on and off set-points. The ALARM ON set-points will always be set before the ALARM OFF set-points. The shark will not let the user input a value below the ALARM ON set-point. The same rule holds true for the high alarm.

The control relays have 5 user configurable settings:

ALARM LOW ON set-point: This is the low process value that will cause the relay to energize. This value can be set anywhere between 0-100% of the range.

ALARM LOW OFF set-point: This is the value that the process must reach in order to de-energize the alarm relay after it has dropped below the ALARM LOW ON set-point. This value must be higher than the ALARM LOW ON set-point.

ALARM HIGH ON set-point: This is the process value that will cause the relay to energize. This value can be set anywhere between 0-100% of the range.

ALARM HIGH OFF set-point: This is the value that the process must reach in order to de-energize the alarm relay after it has increased over the ALARM HIGH ON set-point. This value must be lower than the ALARM HIGH ON set-point.

FAILSAFE: This option can be turned on or off. It reverses the normal action of the relay. (see description under control relay)

ALARM SET-POINT ERROR: If the ALARM LOW ON set-point is set higher than the factory default ALARM LOW OFF set-point, when the user advances from the ALARM LOW ON set-point to the ALARM LOW OFF set-point the shark will adjust the ALARM LOW OFF set-point to be equal to the ALARM LOW ON set-point. If the user then tries to decrease the ALARM LOW OFF set-point the Shark will display the ALARM LOW ALARM setup error screen.

This screen will be displayed for 10 seconds, then return back to the setup screen that was previously displayed. If the user presses the down key again the error message will be displayed again for 10 seconds. The user must accept the LOW OFF set-point, equal to, or greater than the LOW ON set-point.

The same conditions apply to the ALARM HIGH set-points. Except the ALARM HIGH OFF setpoint must be lower than the ALARM HIGH ON set-point. If the user tries to increase the ALARM HIGH OFF set-point higher than the ALARM HIGH ON set-point the High Alarm setup error screen will be displayed.

ALARM RELAY DISABLE: If the user sets the ALARM LOW ON set-point and the ALARM LOW OFF set-point equal to 0% of the range. It will disable the low alarm relay.

If the user sets the ALARM HIGH ON set-point and the ALARM HIGH OFF set-point equal to 100% of the range. It will disable the high alarm relay.

Section 3 - Electrical Connections and Setup

Page 13

MULTI-PARAMETER CONTROLLER & ANALYZER USER’S MANUAL

S H A R K

Section 3 - Electrical Connections and Setup

3.10 MANUAL TEST MODE

(LCD MENU SECTIONS - pH: 4.4, ORP: 5.3, Conductivity: 6.4, Flow: 7.2)

Once the relays are configured, the setup can be tested using Manual Test Mode to simulate process changes.

MANUAL TEST MODE is used to simulate a process reading in order to verify the correct response of the outputs. When in the MANUAL TEST MODE, the relays and outputs are no longer placed on hold as they are when in the rest of the menu. The relays and outputs will react to the simulated change in process as if the Shark was in RUN MODE.

Note that when the user exits the MANUAL TEST

MODE, the relays and outputs will remain in the

MANUAL TEST MODE state until the user enters

RUN MODE.

3.11 RELAY OVERRIDE

(LCD MENU SECTION - pH: 4.5, ORP: 5.4, Conductivity: 6.5, Flow: 7.3)

Relay Override is used to manually override the state of the relays. The user is able to set the operating mode of the relay as AUTO/ON/OFF (the default and RUN MODE states are AUTO).

This feature can be used to energize or de-energize the relays to manually correct the process, or to shut down an ancillary device to perform maintenance. When in the RELAY OVERRIDE mode, the relays are no longer placed on hold as they are when in the other menus.

Note that if the RELAY AUTO RETURN is set to "ON", the controller will place all the relay settings back to AUTO 10 minutes after the Shark returns to the run mode..

Section 3 - Electrical Connections and Setup

Page 14

MULTI-PARAMETER CONTROLLER & ANALYZER USER’S MANUAL

S H A R K

Section 3 - Electrical Connections and Setup

3.12 4-20 mA Isolated Outputs

Channel 1 and Channel 2

(LCD MENU SECTIONS - pH: 4.21 & 4.22, ORP: 5.20 & 5.21,

Conductivity: 6.21 & 6.22, Flow: 7.18 & 7.19)

The Shark Controller has two 4 to 20mA outputs, electrically isolated from each other and ground. Either output can source current into a maximum of 800 ohms.

Channel 1 (the primary output) is located on the flip out door, terminal plug P6. Channel 1 is dedicated to track the process and has fully independent and fully adjustable 4 & 20 mA output setpoints. This will enable the operator to span the output over the desired range.

Channel 2 (the secondary output) is located in the enclosure terminal plug P1. Channel 2 can be selected to track the process value or temperature and has fully independent and adjustable 4 & 20 mA output setpoints.

Both Channel 1 and 2 can be precisely trimmed through the LCD menu for precision applications.

The drawing shows the connections for both outputs.

Wire Specification: 22 AWG 7/30, insulation 0.010”

Figure 3.7 Connections for the 4-20mA outputs

 

 

 

 

 

 

 

 

 

 

 

 

51

 

 

 

30 31

 

 

 

 

 

 

 

5453 52

 

 

 

 

28 29

 

 

 

 

 

 

 

5756 55

 

 

 

 

25 26 27

 

 

 

 

 

 

 

61 60 59 58

Secondary Output

 

 

 

24

 

 

 

 

 

 

 

63 62

CH2 4-20mA +

 

 

 

23

 

 

 

 

 

 

 

64

CH2 4-20mA -

 

 

 

21 22

 

 

 

 

 

 

 

67 66 65

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

P4

 

 

 

P3

 

P1

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Primary Output

 

 

 

 

 

 

 

 

 

 

 

 

 

CH1 4-20mA +

 

 

 

 

 

 

 

 

 

 

 

 

 

CH1 4-20mA -

P6

1

2

3

4

5

6

7

8

9

10

11 12

13 14 15 16

17

Dwg# N104-39

 

 

 

 

 

 

 

 

 

 

 

 

Note: Leave 4” to 6” slack for all wires connected to the terminals of P6. Slack required so that wires do not interfere with opening or closing of the front door.

Section 3 - Electrical Connections and Setup

Page 15

MULTI-PARAMETER CONTROLLER & ANALYZER USER’S MANUAL

S H A R K

Section 3 - Electrical Connections and Setup

3.13 Service

SHARK SERVICE TO BE PERFORMED BY QUALIFIED PERSONNEL ONLY.

3.14 Fuse Replacement

WARNING:

DISCONNECT LINE POWER TO THE UNIT

TO AVOID THE POSSIBILITY

OF ELECTRICAL SHOCK.

1.Proceed after disconnecting line power from the instrument.

2.Open the front panel by rotating the quarter-turn fasteners, using a flat blade screwdriver, to expose the relay board.

3.The fuse, F1, is located in the middle of the relay board, directly above the three terminal connectors.

4.Remove the open fuse and replace it only with a fuse of the same type and rating. REFER TO THE FUSE RATING TABLES BELOW.

5.Close the front panel and secure using the quarter-turn fasteners.

6.Restore power to the unit.

Figure 3.8 Fuse Location

21 22 23 24 25 26 27 28 29 30 31

P4

FUSE

P3

67 66 65 64 63 62 61 60 59 58 57 56 55 54 53 52 51

P1

P6

1

2

3

4

5

6

7

8

9

10

11 12

13 14 15 16

17

Dwg# N104-40

Fuse Rating Table for 120 volt operation

Fuse Type: Slo-Blo Fuse Ratings: 250

fuse 5 x 20mm

VAC, 100mA

Fuse Rating Table for 240 volt operation

Fuse Type: Slo-Blo Fuse Ratings: 250

fuse 5 x 20mm

VAC, 50mA

Section 3 - Electrical Connections and Setup

Page 16

Aquametrix SHARK-120, SHARK-240 User Manual

MULTI-PARAMETER CONTROLLER & ANALYZER USER’S MANUAL

S H A R K

Section 4 - Using the SHARK in pH Mode

4 DIGIT, 7 SEGMENT LED DISPLAY

1/4" TURN SCREWS

DISPLAYS PROCESS VALUE IN RUN MODE.

 

DISPLAYS CALIBRATION DATA IN

 

CALIBRATION MODE.

RELAY C (ALARM RELAY)

 

STATUS LED's (RED)

 

BOTTOM LED WILL ILLUMINATE

 

WHEN ALARM RELAY IS

 

ENERGIZED BECAUSE OF LOW

 

ALARM CONDITION.

 

TOP LED WILL ILLUMINATE WHEN

 

ALARM RELAY IS ENERGIZED

 

BECAUSE OF HIGH ALARM

 

CONDITION.

 

BAR GRAPH LED's (GREEN)

 

LINEAR INDICATOR OF PROCESS

 

VALUE.

 

(REFER TO SECTION 4.13)

 

CALIBRATION FROM THE

FRONT PANEL

RUN LED (GREEN)

LED WILL BE ILLUMINATED WHEN

SHARK IS IN THE RUN MODE.

WILL BE OFF WHEN SHARK IS IN

THE MENU MODE.

OVERFEED LED (RED)

LED WILL FLASH WHEN THE

OVERFEED TIMER IS ACTIVATED.

AUXILLARY LED's

NOT USED IN pH MODE.

UNIT LABEL

LABEL TO INDICATE UNIT OF

MEASURE SHARK IS

CONFIGURED FOR.

RELAY B STATUS LED (YELLOW)

LED WILL BE ILLUMINATED WHEN

RELAY B IS ENERGIZED.

RELAY A STATUS LED (YELLOW)

LED WILL BE ILLUMINATED WHEN

RELAY A IS ENERGIZED.

TEMPERATURE DISPLAY

WHEN THE 'UP' AND 'DOWN' ARROW

KEYS ARE PRESSED

SIMULTANEOUSLY, THE LED WILL

DISPLAY THE CURRENT PROCESS

TEMPERATURE FOR 5 SECONDS.

4 PUSH BUTTONS USED TO

 

 

 

 

FRONT PANEL

 

THIS FEATURE WILL BE DISABLED IF

CALIBRATE THE SHARK FROM THE

 

 

 

 

 

THE 'TEMPERATURE COMPENSATION

FRONT PANEL.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

OVERRIDE' IS TURNED ON IN THE

 

 

 

 

 

 

 

 

 

 

 

 

 

 

SETUP MENU.

(REFER TO SECTIONS 4.22 & 4.23)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

IN THIS CASE THE LED WILL DISPLAY

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

---- IN PLACE OF THE

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

TEMPERATURE.

REMOVABLE TERMINAL BLOCK

P6

 

 

 

 

 

 

 

 

 

 

 

 

 

 

CONNECTORS

1

2

3

4

5

6

7

8

9

10

11 12

13 14 15 16

17

NOTE

"SNAP-ON" CONNECTORS FOR EASY

 

 

 

 

 

 

 

 

 

 

 

 

 

 

WIRING OF YOUR pH SENSOR AND

 

 

 

 

 

 

 

 

 

 

 

 

 

 

WHEN THE SHARK IS TAKEN INTO

PRIMARY 4-20mA OUTPUT.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

MENU MODE VIA THE INSIDE LCD

(REFER TO SECTIONS 3.3, 3.4 & 3.12)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

SCREEN, THE FRONT 7-SEGMENT LED

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

WILL DISPLAY ----. THE STATUS &

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

BAR GRAPH LED's WILL BE TURNED

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

OFF. THE 4 PUSH BUTTONS ON THE

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

FRONT WILL NOT RESPOND.

2 LINE, 16 CHARACTER

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

LCD DISPLAY

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

MAIN MENU INTERFACE

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

SCREEN

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

SIMPLE THREE-BUTTON INTERFACE

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

FOR FAST & EASY SETUP

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

(REFER TO SECTIONS 4.0 TO 4.21)

INSIDE PANEL

Section 4 - Using the SHARK in pH Mode

Page 17

MULTI-PARAMETER CONTROLLER & ANALYZER USER’S MANUAL

S H A R K

pH - Menu Overview 4.0

7.15pH 25.0C

RUN MODE

CALIBRATION

 

UTILITIES

SETUP

DIAGNOSTICS

OUTPUTS

EXIT

FIRMWARE REV

PROBE SELECT

TEMP UNIT

TEMP SENSOR

AUTO RETURN

T.COMP OVERRIDE

DISPLAY DAMPING

BAR GRAPH O/R

EXIT

DISPLAY

FIRMWAVE REVISION SEC. 4.15

MANUAL CAL PH

AUTO CAL PH

TEMP CALIBRATION

EXIT

SELECT TYPE OF

MANUAL TEST MODE

 

pH PROBE

 

SEC. 4.8

RELAY OVERRIDE

SELECT DEG C OR DEG F

 

SEC 4.9

 

SELECT TYPE OF

METER SELECTION

 

TEMP SENSOR

 

SEC 4.10

RESET OVERFD TMR

ENABLE TIME OUT

 

FROM MENU

 

SEC. 4.11

EXIT

TEMPERATURE SETUP FOR

 

PROBE W/O TEMP. SENSOR

 

SEC 4.12

 

SETUP DISPLAY AND

OUTPUT RESPONSE TIME

SEC 4.13

SETUP RESOLUTION

OF FRONT PANEL BAR GRAPH

SEC 4.14

CALIBRATION DATA

DISPLAY PREVIOUS

CALIBRATION DATA

 

 

SEC 4.16

SENSOR INPUT

DISPLAY UNCALIBRATED

PROBE DATA

 

 

SEC 4.17

MANUAL CALIBRATE pH PROBE

SEC. 4.1

AUTO CALIBRATE pH PROBE

SEC. 4.2

CALIBRATE TEMP. SENSOR IN pH PROBE SEC. 4.3

SIMULATE PROCESS VERIFY OUTPUTS SEC. 4.4

MANUALLY OVERRIDE RELAYS

SEC. 4.5

SELECT TYPE OF METER SEC. 4.6

RESETS RELAY OVERFEED TIMER SEC. 4.7

 

EXIT

 

 

RELAY A

SETUP

 

 

RELAY A

 

 

 

 

 

 

SEC 4.18

 

 

RELAY B

SETUP

 

 

RELAY B

 

 

 

 

 

 

SEC 4.19

 

 

ALARM RELAY

SETUP

 

 

ALARM RELAY

 

 

 

 

 

 

SEC 4.20

 

 

4-20mA CH1

SETUP

 

 

4-20 mA CH 1 OUTPUT

 

 

 

 

 

 

SEC 4.21

 

 

4-20mA CH2

SETUP

 

 

4-20 mA CH. 2 OUTPUT

 

 

 

 

 

 

SEC 4.22

NOTE

 

EXIT

 

 

PRESS THE

AND

KEYS

 

TOGETHER TO GO IMMEDIATELY BACK TO

 

 

RUN MODE

 

pH - Menu Overview 4.0

Page 18

MULTI-PARAMETER CONTROLLER & ANALYZER USER’S MANUAL

S H A R K

pH - Calibration Menu - Manual Calibrate 4.1

7.15pH 25.0C

RUN MODE

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Place the probe in the first buffer solution, be

 

CALIBRATION

MANUAL CAL PH

 

IF BUFFER1 READY

 

sure to clean and rinse the Probe first with D.I.

 

 

 

 

PRESS 'DOWN'

 

water and then insert it in the 7.00 buffer.

 

 

 

 

 

 

Press

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

RUNNING MANU CAL

The controller will read the pH value,

 

 

 

 

 

 

averaging a number of results to get a

 

 

 

 

 

BUFFER1 WAIT...

 

 

 

 

 

 

 

stable calibration value.

 

 

 

 

 

 

 

 

 

Please wait for the controller to complete

 

 

 

 

 

 

 

the measurement

 

 

 

 

 

 

 

MANUAL CAL PH

 

When complete, the controller will report

 

 

 

 

 

BUFFER1

7.3 5 >

the measured value

 

 

 

 

 

 

 

Use the

and

keys to adjust

 

 

 

 

 

UP

 

 

 

 

 

 

 

the reading until it agrees with the actual

 

 

 

 

 

 

 

 

 

 

 

 

 

 

buffer pH value

 

 

 

 

 

 

 

 

MANUAL CAL PH

 

Then press

to move the

 

 

 

 

 

BUFFER1

7.0 0 >

cursor to the RH position

 

 

 

 

 

 

MANUAL CAL PH

 

Then press

to store the value and move

 

 

 

 

BUFFER1

7.00 >

 

 

 

 

to BUFFER2

 

 

 

 

 

This example shows a MANUAL

 

 

 

 

Place the probe in the second buffer.

 

 

 

 

 

 

 

Be sure to clean and rinse the Probe first with

 

calibration with buffers of 7pH & 4pH.

 

IF BUFFER2 READY

 

 

 

 

D.I. water and then insert it in the 4.00 buffer.

 

 

 

 

 

 

 

 

 

PRESS 'DOWN'

 

 

When performing a manual

 

 

 

Press

 

 

 

 

 

 

 

 

 

 

 

 

 

 

calibration, any two known buffer

 

 

 

 

 

 

 

 

 

 

solutions can be used.

 

 

RUNNING MANU CAL

The controller will read the pH value,

 

 

 

 

 

 

averaging a number of results to get a

 

 

 

 

 

BUFFER2 WAIT...

 

stable calibration value.

 

 

 

 

 

 

 

 

 

Please wait for the controller to complete

 

 

 

 

 

 

 

the measurement

 

 

 

 

 

 

 

MANUAL CAL PH

 

When complete, the controller will report

 

 

 

 

 

BUFFER2

4.4 0 >

the measured value

 

 

 

 

 

 

 

UP

 

Use the

and

keys to adjust

 

 

 

 

 

 

 

the reading until it agrees with the actual

 

 

 

 

 

 

 

buffer pH value

 

 

 

 

 

 

 

 

MANUAL CAL PH

 

Then press

to move the

 

 

 

 

 

BUFFER2

4.0 0 >

cursor to the RH position

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

MANUAL CAL PH

 

Then press

to store the value and

 

 

 

 

 

BUFFER2

4.00 >

complete the Manual Calibration

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

SLOPE 61.22MV/PH

 

After 5 seconds, the controller will

 

 

 

 

 

 

 

compute the slope of the calibration, the

 

 

 

 

 

EFF 95% 24.8C

 

estimated probe efficiency and the probe

 

 

 

 

 

 

 

temperature.

 

 

 

 

 

 

 

 

 

 

If the calibration is OK, use the

key

 

 

 

 

 

 

 

to move the cursor over the Y text and

 

 

 

 

 

 

 

press the down key.

 

 

 

 

 

If the calibration did not appear to be

MANUAL CAL PH

 

MANUAL CAL PH

 

 

Press

to store the calibration

 

correct, press the

key

STORE?

Y N

STORE?

Y

N

data and return back to the Manual

 

which will return back to the Manual

Cal menu so the user can select

 

 

 

 

 

 

 

 

Cal menu.

 

 

 

 

 

 

 

another function.

NOTE

PRESS THE AND KEYS

TOGETHER TO GO IMMEDIATELY BACK TO

RUN MODE

pH - Calibration Menu - Manual Calibrate 4.1

Page 19

MULTI-PARAMETER CONTROLLER & ANALYZER USER’S MANUAL

S H A R K

pH - Calibration Menu - Auto Calibrate 4.2

7.15pH 25.0C

RUN MODE

 

 

 

 

CALIBRATION

 

MANUAL CAL PH

 

 

 

 

 

 

AUTO CAL PH

IF BUFFER1 READY

 

Place the probe in the first buffer solution,

 

 

 

be sure to clean and rinse the Probe first

 

 

 

PRESS 'DOWN'

 

with D.I. water and then insert it in the 7.00

 

 

 

 

buffer.

 

 

 

 

 

 

 

 

 

 

RUNNING AUTO CAL

 

Press

 

 

 

 

 

The controller will read the pH value,

 

 

 

BUFFER1 WAIT...

 

 

 

 

 

averaging a number of results to get a

 

 

 

 

 

 

 

 

 

 

stable calibration value.

 

 

 

 

 

Please wait for the controller to complete

 

If an errors occurs, the controller will indicate a

 

 

the measurement

 

 

 

 

 

 

 

"BUFFER ERROR" alarm. This could be caused by

 

 

 

 

- using a defective buffer

 

 

 

 

 

- incorrect probe wiring

 

 

BUFFER ERROR

 

 

- defective probe

 

 

CONTINUE

>

 

Press

to return to the Auto cal menu, try to

 

 

 

 

 

 

correct the problem and calibrate again, or go to the

 

 

 

 

diagnotics menus to test the probe

 

 

 

 

 

see Section 4.17

 

 

 

 

 

 

 

AUTO CAL PH

 

When complete, the controller will report

 

 

 

BUFFER1

7.00 >

the measured value

 

 

 

 

 

 

This example shows an AUTO

 

 

Press the

key to move to Buffer 2

 

calibration with buffers of 7pH & 4pH.

 

 

 

 

 

When performing an auto calibration,

 

 

Place the probe in the second buffer

 

any two standard buffers of 4.00, 7.00

 

 

 

 

 

solution.Be sure to clean and rinse the

 

or 10.00 pH can be used. The buffers

IF BUFFER2 READY

 

 

 

Probe first with D.I. water and then insert

 

will be automatically recognized if the

PRESS 'DOWN'

 

it in the 4.00 buffer.

 

probe measures them within +/- 1 pH

 

 

 

 

Press the

key

 

of their nominal value.

 

 

 

 

 

RUNNING AUTO CAL

 

The controller will read the pH value,

 

 

 

BUFFER2 WAIT...

 

averaging a number of results to get a

 

 

 

 

stable calibration value.

 

 

 

 

 

 

If an errors occurs, the controller will indicate a

 

 

Please wait for the controller to complete

 

 

 

the measurement

 

 

"BUFFER ERROR" alarm. This could be caused by

 

 

 

 

 

 

 

- using the same buffer twice

 

 

 

 

 

- using a defective buffer

 

 

 

 

 

- incorrect probe wiring

 

 

BUFFER ERROR

 

 

- defective probe

 

 

CONTINUE

>

 

Press

to return to the Auto cal menu, try to

 

 

 

 

 

 

correct the problem and calibrate again, or go to the

 

 

 

 

diagnotics menus to test the probe

 

 

 

 

 

 

 

AUTO CAL PH

 

When complete, the controller will report the

 

 

 

 

measured value

 

 

 

 

BUFFER2

4.00 >

 

 

 

 

 

 

If the calibration did not appear to be correct, press the key which will return back to the Auto cal menu.

SLOPE 61.22MV/PH

EFF 95% 24.8C

AUTO CAL PH

STORE? Y N

After 5 seconds, the controller will compute the slope of the calibration, the estimated probe efficiency and the probe temperature.

If the calibration is OK, use the key to move the cursor over the Y text and press the down key.

AUTO CAL PH

 

This will store the calibration

STORE?

Y N

data and return back to the

Auto cal setup so the user

 

 

 

 

can select another function.

NOTE

PRESS THE AND KEYS TOGETHER TO GO IMMEDIATELY BACK TO RUN MODE

pH - Calibration Menu - Auto Calibrate 4.2

Page 20

MULTI-PARAMETER CONTROLLER & ANALYZER USER’S MANUAL

S H A R K

pH - Calibration Menu - Temperature Calibration 4.3

7.15pH 25.0C

RUN MODE

 

 

 

CALIBRATION

MANUAL CAL PH

 

If the Temperature Compensation Override is set to ON

 

 

 

 

AUTO CAL PH

 

(see section 4.12), the Shark cannot calibrate the

 

 

temperature sensor. This display will appear to alert the

 

 

 

user to the condition.

 

 

TEMP CALIBRATION

 

TEMP CALIBRATION

 

 

 

TEMP O/R ON

> DOWN

 

 

 

The controller displays the current probe

 

 

 

temperature.

 

 

 

 

If this incorrect, the controller can be

 

 

 

adjusted to compensate.

 

 

TEMP CALIBRATION

 

Press the

key once which will move

 

22.4C

>

 

 

 

the cursor over the least sign. digit of the

 

 

 

temperature display.

If the calibration did not appear to be correct, press the key which will return back to the Temp Calibration menu.

TEMP CALIBRATION

22. 4 C >

Use the and to adjust UP the reading until it agrees with the

known temperature of the probe

TEMP CALIBRATION

25. 0 C > Press the key once which will move the cursor

back to the RH side of temperature display.

TEMP CALIBRATION

 

 

Press the

key to accept the

 

25.0C

 

>

 

 

reading and move to the store function

 

 

 

T FACTOR SHIFTED

 

 

"T FACTOR SHIFTED" will be

 

 

 

displayed for 5 seconds to indicate that

 

 

 

 

 

 

the temperature sensor has been

 

 

 

 

calibrated.

 

 

 

 

 

The Shark will then move to the store

 

 

 

 

function.

 

 

TEMP CALIBRATION

 

 

TEMP CALIBRATION

This will store the calibration data

STORE?

Y N

 

STORE?

Y N

and return back to the Temp

 

 

 

 

 

Calibration setup so the user can

 

 

 

 

 

select another function.

NOTE

PRESS THE AND KEYS TOGETHER TO GO IMMEDIATELY BACK TO RUN MODE

pH - Calibration Menu - Temperature Calibration 4.3

Page 21

MULTI-PARAMETER CONTROLLER & ANALYZER USER’S MANUAL

S H A R K

pH - Utilities Menu - Manual Test Mode 4.4

 

 

 

Manual Test Mode is used to simulate a process reading

 

 

 

in order to verify the correct response of the outputs.

7.15pH 25.0C

RUN MODE

 

When in the Manual Test Mode, the outputs are no

 

 

 

longer placed on hold as they are when in the rest of the

CALIBRATION

 

 

menu.

 

 

 

 

 

 

 

UTILITIES

MANUAL TEST MODE

TEST

7.00 >

Press the

key once which will

 

 

a b c

12.0mA

move the cursor over the least digit of the

 

 

simulated process value.

 

 

 

 

When a relay is energized, the characters a,b,c will change to upper case A,B,C

The CH1 4-20 mA output will track the change in process

TEST

7.0 0 >

Use the

and

keys to change

the simulated process value.

a b c

12.0mA

 

 

 

UP

 

As the value is changed, the state of the relays

 

 

will change depending on their settings in the

TEST

13.1 5 >

OUTPUTS menu.

 

a/A - state of Relay A

 

a B c

19.0mA

 

b/B - state of Relay B

 

 

 

 

 

 

c/C - state of Alarm relay

 

 

As well, Channel 1 4-20 ma output will also

 

 

follow the process value change. The actual

 

 

change will depend on how the CH1 output

 

 

was scaled in the OUTPUTS menus.

 

 

(See Section 4.21)

 

TEST

13.1 5 >

Press the

 

key once to

a b c 19.0mA

move the cursor to the RH side

of the display

 

 

 

 

TEST

13.15 >

Press the

key to return to the

a b c 19.0mA

MANUAL TEST MODE menu

 

 

NOTE:

When the user exits the Manual Test Mode, the relays and 4-20mA outputs will remain in the Test Mode state until the operator enters the run mode. The relays and outputs will then revert back to the previous On-line state.

NOTE

PRESS THE AND KEYS

TOGETHER TO GO IMMEDIATELY BACK TO

RUN MODE

pH - Utilities Menu - Manual Test Mode 4.4

Page 22

MULTI-PARAMETER CONTROLLER & ANALYZER USER’S MANUAL

S H A R K

pH - Utilities Menu - Relay Override 4.5

7.15pH

25.0C

RUN MODE

 

 

Relay Override is used to manually override the state of the relays, the user

 

 

 

 

 

 

is able to set the operating mode of the relay as AUTO/ON/OFF (the default

 

 

 

 

 

 

and run mode state are AUTO).

 

 

 

CALIBRATION

 

 

 

This feature can be used to turn the relays ON or OFF to manually correct

 

 

 

 

the process, or to shut down an ancillary device to perform maintenance.

 

 

 

 

 

 

 

 

 

 

 

 

When in the Relay Override mode, the outputs are no longer placed on hold

 

UTILITIES

 

MANUAL TEST MODE

 

 

as they are when in the other menus.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

To use the Relay Override

 

CAUTION:

 

 

 

 

 

 

Note that if the Relay Auto Return Menu is set to

 

 

 

RELAY OVERRIDE

 

 

function, press the

key

 

 

 

 

 

"ON", the controller will place all these settings

 

 

 

 

 

 

 

 

 

back to AUTO 10 minutes after the Shark returns to

 

 

If you wish to Override Relay

 

 

 

the run mode.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

A, Press the

key once

 

 

 

 

 

 

 

 

 

which will move the cursor over

 

 

 

 

 

 

 

RELAY OVERRIDE

to the AUTO/ON/OFF text

 

 

 

 

 

 

 

 

 

Use the

or

 

key to

 

 

 

 

 

RELAY A

AUTO

>

 

 

 

 

 

 

change the Override state from

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

AUTO - controlled by process

 

 

 

 

 

 

 

 

settings

 

 

 

 

 

 

 

 

 

 

 

ON - relay forced on

 

 

 

 

 

 

 

If you do not wish to

OFFrelay forced off

If you wish to accept the

 

 

 

 

Override Relay A, then

The relay will react

Override setting,

 

 

 

 

 

 

use the

key to

immediately.

 

Press the

key once

 

 

 

 

move to Relay B

 

 

 

to move the cursor to the

 

 

 

 

 

RELAY OVERRIDE

 

 

RH of the display

 

RELAY OVERRIDE

 

Then use the

key

 

 

 

 

 

 

 

 

 

RELAY A

A UTO

>

 

 

 

RELAY A AUTO

>

to move to the Override

 

 

UP

 

 

 

 

 

 

 

setting for Relay A

 

 

 

 

 

 

 

 

 

 

 

 

 

 

RELAY OVERRIDE

 

 

 

 

RELAY OVERRIDE

 

 

 

 

 

RELAY A

O N

>

 

 

 

RELAY A ON

>

 

 

 

 

UP

 

 

 

 

 

 

 

 

 

 

 

RELAY OVERRIDE

 

 

 

 

RELAY OVERRIDE

 

 

 

 

 

RELAY A

O FF

>

 

 

 

RELAY A OFF

>

 

 

 

 

UP

 

 

 

 

 

 

 

 

 

RELAY OVERRIDE

RELAY B AUTO

>

If you do not wish to Override Relay B, then use the key to the Alarm Relay

If you wish to Override Relay

 

 

 

 

 

B, Press the

key once

 

 

 

 

 

which will move the cursor over

 

 

NOTE

 

 

to the AUTO/ON/OFF text

 

 

 

 

 

 

 

PRESS THE

AND

KEYS

 

 

 

 

 

 

 

 

 

Use the

or

key to

TOGETHER TO GO IMMEDIATELY BACK TO

 

 

 

change the Override state from

RUN MODE

 

 

 

 

AUTO - controlled by process

 

 

 

 

 

 

settings

 

 

 

 

 

 

 

 

ON - relay forced on

If you wish to accept the

 

 

 

 

 

OFFrelay forced off

 

 

 

 

 

Override setting,

 

 

 

 

 

 

The relay will react

 

 

 

 

 

 

Press the

key once

 

 

 

 

 

immediately.

 

 

 

 

 

to move the cursor to the

 

 

 

 

 

 

 

 

 

 

 

 

 

 

RH of the display

 

 

 

 

RELAY OVERRIDE

 

 

RELAY OVERRIDE

 

Then use the

key

RELAY B

A UTO

>

 

RELAY B AUTO

> to move to the Override

UP

 

 

 

 

 

 

setting for Relay B

 

 

 

 

 

 

 

 

 

RELAY OVERRIDE

 

 

RELAY OVERRIDE

 

 

 

RELAY B

O N

>

 

RELAY B ON

>

 

 

UP

 

 

 

 

 

 

 

 

RELAY OVERRIDE

 

 

RELAY OVERRIDE

 

 

 

RELAY B

O FF

>

 

RELAY B OFF

>

 

 

UP

 

 

 

 

 

 

 

 

To Alarm Relay (Relay C) Override

pH - Utilities Menu - Relay Override 4.5

Page 23

MULTI-PARAMETER CONTROLLER & ANALYZER USER’S MANUAL

S H A R K

pH - Utilities Menu - Relay Override 4.5

FROM RELAY B

OVERRIDE MENU

 

 

If you wish to Override Relay

 

 

 

 

 

 

 

C, Press the

key

 

 

NOTE

 

 

 

RELAY OVERRIDE

once which will move the

 

 

Relay C is the Alarm Relay

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

RELAY C

AUTO

>

 

Use the

or

key to

 

 

 

 

 

 

 

 

change the Override state from

 

 

 

 

 

 

 

 

AUTO - controlled by process

 

 

 

 

 

 

 

 

settings

 

 

 

 

 

 

 

If you do not wish to

 

ON - relay forced on

 

 

 

 

 

 

 

OFFrelay forced off

If you wish to accept the

 

 

 

 

Override Relay C, then

 

 

 

 

 

 

The relay will react

Override setting,

 

 

 

 

 

use the

key to

 

 

 

 

 

 

 

immediately.

Press the

key once

 

 

 

 

the Store function

 

 

 

 

 

 

 

 

to move the cursor to the

 

 

 

 

 

 

 

 

 

 

 

 

 

 

RELAY OVERRIDE

 

RH of the display

RELAY OVERRIDE

 

Then use the

key

 

 

 

 

 

 

 

RELAY C

A UTO

>

 

RELAY C AUTO

>

to move to the Override

 

 

UP

 

 

 

 

 

 

setting for Relay C

 

 

 

 

 

 

 

 

 

 

 

 

 

RELAY OVERRIDE

 

 

RELAY OVERRIDE

 

 

 

 

 

RELAY C

O N

>

 

RELAY C ON

>

 

 

 

 

UP

 

 

 

 

 

 

 

 

 

 

RELAY OVERRIDE

 

 

RELAY OVERRIDE

 

 

 

 

 

RELAY C

O FF

>

 

RELAY C OFF

>

 

 

 

 

UP

 

 

 

 

 

 

 

 

 

 

If you wish to store the state

 

 

 

 

 

 

 

 

of the Relay Overrides,

 

 

 

 

 

 

 

 

press the

 

key to

 

 

 

 

 

 

RELAY OVERRIDE

highlight the Y character

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

STORE?

Y

N

 

 

 

 

 

 

 

 

NOT STORED

With "N" highlighted, pressing the key will NOT store the Override states of the relays, but they will revert to the state they were in before the menu was entered. Note that, if any of the relays were in a specific Override state before the menu was entered (i.e. not in AUTO), they will revert to that state again unless the changes are STORED.

RELAY OVERRIDE

STORE?

Y

N

STORED

Then press the key to store the Override states of all the relays and return back to the Override menu

TO RELAY

OVERRIDE MENU

NOTE

PRESS THE AND KEYS

TOGETHER TO GO IMMEDIATELY BACK TO

RUN MODE

pH - Utilities Menu - Relay Override 4.5

Page 24

MULTI-PARAMETER CONTROLLER & ANALYZER USER’S MANUAL

S H A R K

pH - Utilities Menu - Meter Selection 4.6

7.15pH 25.0C

CALIBRATION

UTILITIES

RUN MODE

The Meter Selection menu is used to select the meter type that the controller is configured for, either pH, ORP, Conductivity or FLow. Once selected, the controller will initialize itself for the selected meter's functionality and move to run mode.

MANUAL TEST MODE

RELAY OVERRIDE

Press the

key to initialize the

 

controller as a pH meter.

METER SELECTION

To reset the controller back to all the factory default values, reselect the current meter type

P H

PH

 

ORP

INITIALIZE?

Y N

O RP

CONDUCTIVITY

C ONDUCTIVITY

FLOW

F LOW

EXIT

E XIT

PH

Press the key to initialize the controller as an ORP meter

ORP

INITIALIZE? Y N

Press the key to initialize the controller as a Conductivity meter

CONDUCTIVITY INITIALIZE? Y N

Press the key to initialize the controller as a Flow meter

FLOW

INITIALIZE? Y N

Press the key to Exit and return to the Utilities menu

If you do not wish to initialize to this type of meter after all, then press key to return back to the Meter Selection menu

If you do wish to initialize to this type of meter, press the key to highlight the Y character

Then press

PH

 

The controller will initialize itself for a

INITIALIZE?

Y N

specific meter

 

 

NOTE

PRESS THE AND KEYS TOGETHER TO GO IMMEDIATELY BACK TO RUN MODE

pH - Utilities Menu - Meter Selection 4.6

Page 25

MULTI-PARAMETER CONTROLLER & ANALYZER USER’S MANUAL

S H A R K

pH - Utilities Menu - Overfeed Timer Reset 4.7

7.15pH 25.0C

RUN MODE

CALIBRATION

 

UTILITIES

MANUAL TEST MODE

 

RELAY OVERRIDE

 

METER SELECTION

 

RESET OVERFD TMR

If the control relay overfeed timer has been enabled, the relay will Òtime outÓ after the specified overfeed time. When the relay times out, it must be manually reset. The time out will be signaled by the time out LED flashing on the front panel.

The reset overfeed timer will reset the timers for both relay A & relay B at the same time.

Press the key to choose between Y or N.

RESET OVERFD TMR

RESET OVERFD TMR

Y N

Y N

If you wish to reset the overfeed timer, then press the key to proceed and return back to the Utilities menu.

If you do not wish to reset the overfeed timer after all, press the key to highlight the N character and press the

key to return back to the Utilities menu.

NOTE

PRESS THE AND KEYS TOGETHER TO GO IMMEDIATELY BACK TO RUN MODE

pH - Utilities Menu - Overfeed Timer Reset 4.7

Page 26

MULTI-PARAMETER CONTROLLER & ANALYZER USER’S MANUAL

S H A R K

pH - Setup Menu - Probe Select 4.8

7.15pH 25.0C

RUN MODE

PROBE SELECT will allow the user to select

 

 

CALIBRATION

 

whether the probe is a 2 or 4 wire

 

 

combination probe, or a 5 wire differential

 

 

 

probe.

 

UTILITIES

 

 

 

SETUP

PROBE SELECT

PROBE SELECT

 

 

 

DIFFERENTIAL

>

Press the key once which will move the cursor over the first character of the probe type.

Note:

 

 

PROBE SELECT

Use the

or

keys to scroll through

 

 

D IFFERENTIAL

the probe types available.

If using the 2 wire Combination Probe (which

 

 

In this case, the user can select

 

 

 

doesn't have a temperature sensor), T COMP

 

UP

 

 

 

- the 2 or 4 wire combination probe

OVERRIDE must be set to ON, and the actual

 

 

 

 

 

 

- the 5 wire differential probe

probe temperature set through the T COMP

 

 

 

 

 

 

 

 

 

 

OVERRIDE menu.

 

 

 

 

 

Once the correct probe type is selected,

(see sec. 4.12)

 

 

 

 

 

 

 

 

 

 

 

move to the store function to save the

 

 

 

 

 

 

selection

 

 

 

 

 

PROBE SELECT

Press the

 

key once

 

 

 

C OMBINATION

 

 

 

 

which will move the cursor

 

 

 

 

 

 

 

 

 

 

back to the RH side of the

 

 

 

 

 

display.

 

 

 

 

 

PROBE SELECT

 

 

 

 

 

 

 

COMBINATION

>

 

 

 

 

 

 

 

 

Press the

key to accept the

 

 

 

 

 

setting and move to the store function

With "N" selected, pressing the

 

 

Or press the

key to highlight

 

 

 

 

 

the Y character.

 

 

 

 

 

key will NOT store the selection, but

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

simply return to the PROBE SELECT PROBE SELECT

 

PROBE SELECT

 

 

 

 

Menu. This function is useful if you

STORE?

Y N

STORE?

 

Y N

 

 

 

wish to view the current selection

 

 

 

 

 

 

 

 

 

 

 

 

without making any changes.

 

 

 

 

Then press the

key to store the

 

 

 

 

 

 

 

 

Not stored

 

Stored

selection and return to the PROBE

 

 

 

 

SELECT Menu.

 

 

 

 

 

 

 

 

 

 

 

Note: Refer to Appendix A - Probe

Configuration Table

NOTE

PRESS THE AND KEYS

TOGETHER TO GO IMMEDIATELY BACK TO

RUN MODE

pH - Setup Menu - Probe Select 4.8

Page 27

MULTI-PARAMETER CONTROLLER & ANALYZER USER’S MANUAL

S H A R K

pH - Setup Menu -Temp Unit 4.9

7.15pH 25.0C

RUN MODE

 

TEMP UNIT allows the user to select either

 

 

 

Degrees Centigrade or Fahrenheit units for

CALIBRATION

 

 

display

 

 

 

 

 

 

 

UTILITIES

 

 

 

 

 

SETUP

PROBE SELECT

 

 

 

 

 

TEMP UNIT

TEMP UNIT

 

Press the

key once

 

 

DEGREE C

>

which will move the cursor

 

 

 

 

over the unit type, C or F.

TEMP UNIT

DEGREE

C

>

UP

 

 

TEMP UNIT

 

DEGREE

F

>

Use the

or

to select C or F

Press the

 

key once

which will move the cursor

back to the RH side of the

display.

 

 

TEMP UNIT

DEGREE F

>

 

With "N" selected, pressing the

TEMP UNIT

 

key will NOT store the selection, but

STORE?

Y N

simply return to the TEMP UNIT

 

 

Selection Menu. This function is useful if

 

 

you wish to view the current selection

Not stored

 

without making any changes.

 

 

 

Press the key to accept the change and move to the store function

Or press the key to highlight the Y character.

TEMP UNIT

 

 

STORE?

Y N

 

Stored

Then press the

key to store the

selection and return to the TEMP UNIT

 

Selection Menu.

 

NOTE

PRESS THE AND KEYS TOGETHER TO GO IMMEDIATELY BACK TO RUN MODE

pH - Setup Menu - Temp Unit 4.9

Page 28

Loading...
+ 102 hidden pages