Aquametrix SHARK TX Datasheet

Description
SUMMARY
Complete and versatile, SHARK-TX is the only 1/4 DIN and DIN rail mount­able two wire transmitter on the market that allows the user to select one of four measuring parameters.
FOUR MEASURING PARAMETERS
Select the parameter you wish to mea­sure from the easy-to-use LCD menu on the front cover. Choose Conductivity, pH, ORP or Flow.
COMPLETE - NO EXTRA CARDS OR OPTIONS REQUIRED
Each SHARK-TX comes complete. There are no extra costs associated with buying boards for different applications.
TWO MOUNTING OPTIONS
SHARK-TX comes complete with a uni­versal mounting kit for surface, panel and pipe-mount applications. The NEMA 4X 1/4 DIN enclosure is perfect for stand­alone or panel-mount operation. SHARK-TXP is NEMA 4X for front panel mounting and comes complete with DIN rail mounting hardware for mounting in a control panel.
DISPLAY
2-line, 16-character LCD on the front panel.
Aquametrix SHARK TX / TXP
Multi-Parameter Transmitter
ANALOG OUTPUTS
The SHARK-TX provides an isolated and fully scalable 4-20 mA output.
RELAYS WITH CYCLE TIMERS
The instrument also provides control of external devices using its two indepen­dent control relays. A third relay is pre­set to act as an alarm relay, but can be used as a process control relay. It has both high and low on setpoints with ad­justable off setpoints. Factory set for bidirectional control, both control relays can be set for either a rising or falling process, with easily programmed on setpoints and off setpoints. Each control relay has a built-in independent cycle timer, with eld-set on and off times. This feature enables tighter control of batch processes by eliminating chemi­cal overshoot.
ENCLOSURE
SHARKTX is packaged in a rugged NEMA 4X polycarbonate enclosure making it ideally suited for heavy-duty applications such as industrial wastewa­ter neutralization, municipal water and wastewater, pulp and paper, and pro­cess control. The SHARKTXP enclo­sure is also polycarbonate with a NEMA 4X front panel, and DIN rail mounting hardware on the back.
Features
pH, ORP, Conductivity &
Flow parameters available
24 VDC / 24 VDC Loop
Easy to read 2 X 16
character LCD display
Quick and easy to calibrate
Single 4-20mA output
with range expandability
1/4 DIN size, NEMA 4X
polycarbonate housing
Shark-TX: Universal
mounting hardware pro­vided for surface, panel and pipe mounting
Shark-TXP: Panel
or DIN rail mounting hardware provided
Applications
Process Control
Industrial and Municipal
Water Treatment
Industrial and Municipal
Waste Treatment and
Neutralization
Fume Scrubbers
Suitable for the Plating,
Circuit Board Manufactur­ing, Food and Beverage, Chemical Processing, Pulp & Paper, Mining, Nuclear Energy and Phar­maceutical Industries
makers of AquaMetrix instruments
100 School Street
Andover, MA 01810
978-749-9949 887-749-9099
www.WaterAnalytics.net
SHARK TX/P pH, ORP, Conductivity, Flow Transmitter
Technical Data
pH ORP Conductivity Flow
Display
Measuring Range
Temperature
Compensation
Temperautre Unit
Temperature Sensor
Calibration modes
Ambient Conditions
Sensor to Trans-
mitter Distance
AnalogOutput
Memory Back-up
Mechanical
Sensor Input
Invalid Entries
Manual Test Mode
Output Hold
Calibration Data
Auto Return
Display Damping
Net Weight
pH: 0.01 to 14.00 Temp: 0 to 100°C
or 32° to 212°F
Automatic or Manual
0 to 100°C (32° to +212°F)
User selectable: 300Ω NTC Thermistor, 3000Ω NTC Thermistor or Pt. 1000 RTD
Auto-Calibration,
Manual Calibration,
Temperature Display
Isolated Output, Range expand 0 to 100% of full scale (min segment 10% of full scale),max. load 800Ω
SHARKTX Enclosure:NEMA 4X, 1/4 DIN, polycarbonate enclosure with two 1/2” conduit holes
Probe: -600 to +600 mV
Temp. Sensor: 0 to 9999 Ω
Recall data from last calibration, calibra­tion mode, 1st&2nd accepted buffer value and probe mV output, calibration tempera-
ture, calibration slope, and probe efciency
User can select rate at which SHARK updates display. Enables display damping of unstable process
Front Panel: 4 x 7 segment 1/2” LED display, 1 LED indicator 0n-line,7 LED
Bar Graph. Inside Panel: 2 x 16 alpha-numeric LCD display
MΩ/
3
cm
µS/
3
ORP: -1999 to +1999mV
(Dependent on sensor)
Temp: 0 to 100°C
or 32° to +212°F
Not required
°C or °F Not required
Manual Calibration
Temperature Calibration
Temperature: -20°C to +60°C or -4°F to +140°F Humidity: 0 to 90% RH (non-condensing)
Differential Sensor: 3000 ft
Combination Sensor:10 ft
All user settings are retained indenitely in memory (EEPROM)
SHARKTXP Enclosure:NEMA 4X front panel, 1/4 DIN, polycarbonate
SHARKTX Mounting: Universal Mounting kit for sur face, pipe and panel mount included
SHARKTXP Mounting: Panel and DIN rail mount included
Probe: -1999 to
+1999 mV
Temp. Sensor:
0 to 9999 Ω
Invalid entries cannot be stored
Process value can be simulated with arrow keys to verify correct setup of outputs
4 to 20 mA output is placed on hold when the transmitter is in Menu mode
User selectable auto return if the transmitter is left in menu mode for more than 10 min.
SHARK-TX: 0.71 lbs (0.32 kg)
SHARK-TXP: 0.25 lbs (0.12 kg)
cm
mS/
3
cm
Automatic or Manual
User selectable tempera-
ture compensation slope
0.0 to 10.0% / °C.
0 to 100°C (32° to 212°F)
Sample Calibration
Temperature Calibration
4 to 20 mA
Cell: 0 to 9999 Ω
Temp. Sensor: 0 to 9999 Ω
Recall data from last calibra-
tion, calibration buffer ac-
cepted value and cell resistance,
calibration temperature.
0 to 19.9 0.01
0 to 2.00 0.01
0 to 20.0 0.1
0 to 200 0.1
0 to
2000
0 to 20 10
0 to 200 50
Dry Calibration
300 ft 2000 f t
1.0
Flow: 0 to 9999 with
selectable ow rate units
Volume: 0 to 9999 with
Auto Range Flow rate
units: Gallons (GP),
Cubic Feet (CF), Liters
(LP), Cubic Meters (CM),
custom by entering factor
related to Gallons.
Time units: Seconds (S),
Minutes (M), Hours (H)
Paddle: 0 to 2000 Hz
Recall store K factor.
Not required
K factor input
www.WaterAnalytics.net 978-749-9949 888-749-9099