AquaMetrix P65C, R65C Data sheet

Water Analytics
Aqua
Description
by
Water Analytics
Aqua
Metrix
AquaMetrix pH and ORP differential probes stay in service and provide accu­rate measurements under conditions that often render conventional pH probes in­operable. Now for added versatility, these probes, eld-proven in hundreds of instal­lations, are available with an integral en­capsulated 4-20 mA two-wire transmitter to feed directly to a PLC or a DCS.
The P65 pH and R65 ORP probes em­ploy a differential measurement tech­nique. Unlike conventional combination probes, the differential probe has two high impedance measurement circuits containing a common metallic return electrode. One circuit includes the pro­cess measurement electrode which gen­erates a potential E1 proportional to the process pH. The second circuit includes an internal measurement electrode im­mersed in a stable buffer solution which generates a standard reference potential, E2. Both circuitshave a common potential E3 developed at the return electrode. The two circuits are fed into ampliers which provide an output representing the differ­ential between them:
(E1 - E 3) - (E2 - E3). The common po­tential E3 is cancelled out electronically, greatly reducing inaccuracies caused by ground loops which may exist between process and instrument grounds. Ground loop current willow through the low im­pedance path of the return electrode, af­fecting the potential E3, but not the dif­ferential measurement.
Aquametrix Model P65C/R65C
Probes with 4 to 20 mA Output
The differential probe maintains its accu­racy and stability in aggressive process applications long after a combination­style probe’s performance begins to de­teriorate.
Maintenance costs are reduced and the life of the probe is increased. The internal reference electrode is electrically con­nected to the process solution by means of a eld-replaceable double junction salt bridge which greatly reduces the possibil­ity of contamination of the buffer solution in the reference circuit. Although seldom required, the reference solution may be easily replaced by removing the screw­out salt bridge. A salt bridge and buffer kit is available for this purpose.
Another advantage of the 65 series is the semi-ush face which is easily cleaned and avoids solution materials gather­ing on protrusions found in competitive probes. The domed glass electrode, the protective metal electrode and the tem­perature sensor protrude only about 1/8 inch while the salt bridge is ush. A at­face version of the pH probe is also avail­able.
Features
Two-Wire Transmitter built in
Integral two-wire 4-20
mA transmitter can be fed directly to PLC, DCS
Differential Measurement
Replaceable Salt Bridge
Long-lasting
Low Maintenance Cost
Field-proven
Reduces ground loop problems
Easy Cleaning
• Semi-ush face: re-
duced buildup
Temperature Compensation
Automatic temperature compen-
sation on pH versions
Options
Flow-through and submersion
Hot-Tap available
Flat-faced available (pH)
Gold electrode available (ORP)
Differential pH & ORP ver-
sions without 2 wire transmitter available (see P/R60 series)
Applications
Process Control
Industrial and Municipal Wa-
ter Treatment
Industrial and Municipal
Waste Treatment and Neutral­ization
Fume Scrubbers, Plating,
Circuit Board Manufacturing, Food and Beverage, Chemi­cal Processing, Pulp and Pa­per, Mining, Power Generation, Pharmaceutical Industry
by
Andover, MA 01810
Toll free - 855-747-7623
www.WaterAnalytics.net
Metrix
100 School Street
978-749-9949
Model P65C/R65C pH or ORP Probes with 4-20 mA Output
18”
1.25” NPT
Compression
Fitting
Technical Data
Measuring Range
pH 0 to 14 pH (Consult factory for applications
below 2 or above 12).
ORP -500 to +500 mV or
0 mV to +1000 mV, eld selectable
Flow Rate 10 ft./sec maximum (3 metres/sec)
Flow should be as low as possible in low conductivity water and in solutions with high suspended solids
Wetted Materials CPVC, kynar, glass, titanium palladium al-
loy and EPDM (platinum for ORP probe); opt. viton and ceramic
Transmission Distance Dependant upon transmission distance
and supply voltage
Sensitivity
pH 0.001 pH
ORP 0.1 mV
Stability
pH 0.03 pH per day, non- cumulative
ORP 2 mV per day, non-cumulative
Temperature Copensation
Automatic -5 to 95°C (23 to 203°F)
Pressure Limit 100 psig at 65°C maximum
Temperature Limits
CPVC -5 to 95°C (23 to 203°F)
The temperature limit of probes in ow­through applications is limited by pressure and by the pipe tting material.
Probe Cable 5 Conductor plus shield, 15 ft. (4.6m) long
Dimensions
Related Products
CABLES & ACCESORIES
JB1 NEMA 4X junction box
C42-1-XXX Interconnect cable; dressed both ends - specify length
C35-17(K) Salt Bridge Kit for P/R65C- 4/- 6/-7 probes, package of 3
AM60-9765K Salt Bridge Kit, all P/R65C-8 probes, package of 3
P60-HTC Hot Tap Ball Valve assembly
CALIBR ATION SOLUTIONS
A35-13 pH 4 Buffer, 500 mL.
A35-14 pH 7 Buffer, 500 mL
A35-24 pH 10 Buffer, 500 mL
A35-40 ORP Buffer, 200 mV, 500 mL
A35-41 ORP Buffer, 600 mV, 500 mL
Ordering Information
P/R65C-8 Flow-through or submersion applications; body threaded 1-1/2” both ends (P-pH; R- ORP)
P/R65C-6 “Easy-in Easy out” variable insertion depth, 1-1/4” tting (P-pH; R-ORP)
P/R65C-7 For use with Hot-Tap hardware (P-pH; R-ORP)
P65C-8-F Flat-Face pH probe
P65C-8-A Antimony pH probe
R65C-8-G ORP probe with Gold electrode
www.WaterAnalytics.net 978-749-9949 Toll free - 855-747-7623 (855-pHProbe)